NON-DESTRUCTIVE TESTING OF ARTIFICIAL JOINTS WITH DEFECTS BY EDDY CURRENT METHOD

Similar documents
A Numerical Study of Depth of Penetration of Eddy Currents

Steam Generator Tubing Inspection

Modelling of Pulsed Eddy Current Testing of wall thinning of carbon steel pipes through insulation and cladding

Effect of fatigue crack orientation on the sensitivity of eddy current inspection in martensitic stainless steels

FEM SIMULATION FOR DESIGN AND EVALUATION OF AN EDDY CURRENT MICROSENSOR

Maximizing the Fatigue Crack Response in Surface Eddy Current Inspections of Aircraft Structures

Corrosion Steel Inspection under Steel Plate Using Pulsed Eddy Current Testing

INVESTIGATION OF IMPACT DAMAGE OF CARBON FIBER- RAINFORCED PLASTIC (CFRP) BY EDDY CURRENT NON- DESTRUCTIVE TESTING

Modelling III ABSTRACT

2.5D Finite Element Simulation Eddy Current Heat Exchanger Tube Inspection using FEMM

Giant Magnetoresistance Based Eddy-Current Sensor for High-Speed PCB Defect Detection

54. IWK Internationales Wissenschaftliches Kolloquium International Scientific Colloquium

Hybrid of Eddy Current Probe Based on Permanent Magnet and GMR Sensor

Array Eddy Current for Fatigue Crack Detection of Aircraft Skin Structures

DEVELOPMENT OF EDDY CURRENT PROBES BASED ON MAGNETORESISTIVE ARRAY SENSORS

Weld gap position detection based on eddy current methods with mismatch compensation

DEVELOPMENT OF VERY LOW FREQUENCY SELF-NULLING PROBE FOR INSPECTION OF THICK LAYERED ALUMINUM STRUCTURES

IMPROVEMENT OF DETECTION OF SMALL DEFECTS LOCATED NEAR OR FAR FROM WELDS OF MAGNETIC STEAM GENERATOR TUBES USING REMOTE FIELD EDDY CURRENT

Enhance the Sensibility of the Eddy Current Testing

Novel Demagnetization Method after Magnetic Particle Testing

Application of Ultrasonic Guided Wave to Heat Exchanger Tubes Inspection

PULSED REMOTE FIELD TECHNIQUE IN FERROMAGNETIC TUBE WALL THICKNESS AND INNER DIAMETER MEASUREMENT

Equivalent current models and the analysis of directional ECT signals

New portable eddy current flaw detector and application examples

D DAVID PUBLISHING. Eddy Current Test for Detection of Foreign Material using Rotating Probe. 2. Theory. 1. Introduction

EDDY CURRENT INSPECTION FOR DEEP CRACK DETECTION AROUND FASTENER HOLES IN AIRPLANE MULTI-LAYERED STRUCTURES

Detecting 1 st and 2 nd Layer Simulated Cracks in Aircraft Wing Spanwise Splice Standards Using Remote-Field Eddy Current Technique

Spatial detection of ferromagnetic wires using GMR sensor and. based on shape induced anisotropy

Eddy Current Nondestructive Evaluation Based on Fluxgate Magnetometry Umberto Principio Sponsored by: INFM

Detection of Surface and Sub-surface Defects in Aluminium Plate Using Pulsed Eddy Current Technique

A Novel Self Calibrating Pulsed Eddy Current Probe for Defect Detection in Pipework

EDDY-CURRENT MODELING OF FERRITE-CORED PROBES

The Battle of Carbon Steel

DEEP FLAW DETECTION WITH GIANT MAGNETORESISTIVE (GMR) BASED SELF-NULLING PROBE

Flexible PCB-Based Eddy Current Array Probes for the Inspection of Turbine Components

Global Journal of Engineering Science and Research Management

EXTREME LOW FREQUENCY MAGNETIC IMAGING METHOD FOR DETECTION OF. Katsumi INOUE 2)

Imaging for 3D Eddy Current Nondestructive Evaluation Pasquale Buonadonna Sponsored by: INFM

Steam Generator Tube Inspection I

MEASUREMENT OF SURFACE DISPLACEMENT EXCITED BY EMAT TRANSDUCER

Depth of Penetration Effects in Eddy Current Testing

ANALYSIS OF TRANSIENT ACTIONS INFLUENCE IN POWER TRANSFORMER

DEVELOPMENT OF A PROBE OF EDDY CURRENT TESTING FOR DETECTION OF IN-PLANE WAVINESS IN CFRP CROSS-PLY LAMINATES

Research and Application of Pulsed Eddy Current Testing for Ferromagnetic Metallic Components

ELECTROMAGNETIC ON-LINE TESTING OF ROLLED PRODUCTS AND TUBES. PROBLEMS AND SOLUTIONS.

MultiScan MS Tube Inspection System. Multi-technology System Eddy Current Magnetic Flux Leakage Remote Field IRIS Ultrasound

Magnetic Eddy Current (MEC) Inspection Technique

MultiScan MS Tube Inspection System. Multi-technology System Eddy Current Magnetic Flux Leakage Remote Field IRIS Ultrasound

NDT-PRO Services expands service offering

Multiple Frequency Eddy Current Technique

High-Precision Internal Diameter Measurements Using Eddy Current Arrays

tas&m, INC., 107 Research Drive, Hampton, VA 23666

Eddy Current Modelling for Fasteners Inspection in Aeronautic

Eddy Current Testing (ET) Technique

EDDY CURRENT EXAM SIMULATION USING COUPLED FINITE ELEMENT/ VOLUME INTEGRAL OR FINITE ELEMENT/BOUNDARY ELEMENT METHOD

Reduction stray loss on transformer tank wall with optimized widthwise electromagnetic shunts

2014 EDDY CURRENT BENCHMARK

Signal Processing in an Eddy Current Non-Destructive Testing System

AFRL-RX-WP-TP

Influence of Scanning Velocity and Gap Distance on Magnetic Flux Leakage Measurement

Testing Critical Medical Tubing Using High Frequency Eddy Current Coils

Quantitative Crack Depth Study in Homogeneous Plates Using Simulated Lamb Waves.

DEEP PENETRATING EDDY CURRENT for DETECTING VOIDS in COPPER

VD3-71 universal eddy current flaw detector application for field inspection of aeronautical engineering

warwick.ac.uk/lib-publications

Evaluation of Crack Depth Using Eddy Current Techniques with GMR-based Probes

ANALYSIS OF TRANSIENT ACTIONS INFLUENCE IN POWER

Exercise 7. Inductive Proximity Switches EXERCISE OBJECTIVE

Detection and Imaging of Internal Cracks by Tangential Magnetic Field Component Analysis using Low-Frequency Eddy Current Testing

Usage of Magnetic Field Sensors for Low Frequency Eddy Current Testing

Skin Effect in Eddy Current Testing with Bobbin Coil and Encircling Coil

Research Article Flexible GMR Sensor Array for Magnetic Flux Leakage Testing of Steel Track Ropes

Efficient Electromagnetic Analysis of Spiral Inductor Patterned Ground Shields

An instrument for detecting corrosion in anchorage zones of bridge cables using guided waves

Analysis of metallic ropes magnetisation during magneto-inductive testing

DETECTION OF THE SUBSURFACE CRACKS PROPAGATED FROM RIVET HOLES IN THE MULTILAYER AIRCRAFT STRUCTURES

Research on High Resolution Stress Corrosion Crack Detection Based on Remote Field Eddy Current Non-Destructive Testing

Heat Exchanger & Boiler Tube Inspection Techniques

Qualitative Measurement of Moisture Absorption in GFRP Utilizing Electromagnetic Induction

Induction thermography for automatic crack detection in automotive components

Eddy Current Array for Aerospace

In-line eddy current testing of wire rod

ABSTRACT 1. INTRODUCTION

Data Processing at the Flaw Detector with Combined Multisector Eddy-Current Transducer

A FPGA Based Platform for Multi-Frequency Eddy Current Testing

Ultrasonic Guided Wave Testing of Cylindrical Bars

THE UNDER HUNG VOICE COIL MOTOR ASSEMBLY REVISITED IN THE LARGE SIGNAL DOMAIN BY STEVE MOWRY

Target Temperature Effect on Eddy-Current Displacement Sensing

AA&S Conference 2018 Eddy Current Array for Aircraft

High Resolution Eddy Current Testing of Superconducting Wires using GMR-Sensors

R&D of Multi-Frequency ECT Algorithms for FBR SG Tubes

EDDY CURRENT TESTING

Defect detection in stainless steel tubes with AMR and GMR sensors using remote field eddy current inspection

FLAW DETECTION USING ENCIRCLING COIL EDDY CURRENT SYSTEMS

Dual Core Differential Pulsed Eddy Current Probe to Detect the Wall Thickness Variation in an Insulated Stainless Steel Pipe

Name: Lab Partner: Section: The purpose of this lab is to study induction. Faraday s law of induction and Lenz s law will be explored. B = B A (8.

EMAT Application on Incoloy furnace Tubing Ramamohan Reddy M (ASNT Level III UT, PCN Level III UT,PAUT&TOFD)

Research & Development of Integrated MMM/EC/MFL Testing System

A COMBINED ACFM-SMFM SYSTEM FOR REAL-TIME DETECTION AND. S. H. H. Sadeghi and D. Mirshekar-Syahkal

MULTI-PARAMETER ANALYSIS IN EDDY CURRENT INSPECTION OF

Transcription:

NON-DESTRUCTIVE TESTING OF ARTIFICIAL JOINTS WITH DEFECTS BY EDDY CURRENT METHOD Andrea Stubendekova 1, Ladislav Janousek 1 1 Department of Electromagnetic and Biomedical Engineering, Faculty of Electrical Engineering, University of Zilina, Zilina, Slovakia Abstract This work deals with non-destructive testing of artificial joints made from the austenitic steel SUS316L based on eddy current method. The technique is appropriate for final investigation of artificial joints before insertion the prosthesis into human body to avoid repeated operations. The most commonly implanted joints - knee and hip joint are evaluated in this study. The eddy current probe consists of one coil, which has both the excitation and sensing functions. Five detects are inspected using three different frequencies. The numerical simulations from Opera software with experiments are compared in this article to obtain the most appropriate frequency. The aim of this work is evaluation which frequency provides the biggest magnitude difference between individual simulated signals with various depths if surface defects are inspected. Keywords artificial joints, austenitic steel, non-destructive testing, eddy currents Introduction Nowadays, the investigation of artificial prostheses is very significant field. Non-destructive method is type of technique which evaluates an object without damaging it. Eddy current testing is one of the electromagnetic methods which is useful for testing of biomaterials with nonzero conductivity. This method successfully detects the defects inside an inspected product. The value of testing frequency depends on the type of a detected defect (surface or subsurface). This work concerns on a reliable frequency selection for detection of surface defects using probe of absolute type, it means the eddy currents are excited and sensed by a same coil. This probe continuously moves over a tested material containing defects and resulting signal is displayed. In the world there are a lot of repeated operations of artificial prostheses. In Slovakia, the number of above mentioned operations represents around four hundred of hip joint and one hundred of knee joint. One of the reasons of the repeated operations may be inserting a damaged implant. Therefore, the artificial joints should be investigated before implantation and the eddy current method is suitable for it, [1]. Artificial prostheses of joints The endoprosthesis is the term used to describe the artificial joint intended to implantation. The hip and knee joints are the most often implanted types of artificeal joints. The materials which are used to manufacture them must be non-toxic, non-carcinogenic, corrosion resistant and therefore they must meet the requirements of biocompatibility. Consequently the stainless steel, cobalt alloys, titan and its alloys are suitable for the purpose. The lifetime of the artificial joints depends on several factors, for example a patient weight or too big load in some sports. Then the exchange of endoprosthesis is essential, [2], [3]. Non-destructive testing of materials Non-destructive testing of material is included among the successful diagnostic methods. The test measurements are created by manual testing or using automated systems. The task of evaluation are the selection of damaged products, specifying of the 5

character and size of defects, their location and eventually making decision about the applicability of an object, [4]. The visual, penetrant, magnetic particle, ultrasonic, radiographic, potential drop method and eddy current method belong to the methods of nondestructive testing, [5]. Eddy current method Eddy current method is based on the phenomenon of electromagnetic induction and therefore it belongs to the electromagnetic methods of non-destructive testing. If a conductor is situated in a time-varying magnetic field, electromagnetic induction arises and it results in the induction of electrical voltage within the mentioned conductor. Currents are induced in planes, which are perpendicular to the vector of the magnetic induction. The shape of these currents resemble the whirlpool, consequently they are called eddy currents, [6], [7]. The coil is supplied with a time-varying electrical current and generates a time-varying electromagnetic field. If a conductive object is situated in the vicinity of the coil it causes the induction of alternating voltage and the flow of eddy currents is formed in the object. The direction of EM field of coil is opposite as the direction of electromagnetic field of eddy currents. The resulting field of this system is the connection of mentioned fields. Defects in the evaluated object impact the resulting electromagnetic field and cause a change of induced voltage in the coil. These changes give information about a presence of crack in a test material with nonzero conductivity, [6], [7]. The objective of eddy current testing (ECT) is to record the signal of the coil, which moves above surface of a material and evaluate whether the object contains a defect, [6]. Numerical simulations and experiments by means of ECT The purpose is to simulate the process of nondestructive testing by eddy current method. Numerical simulation of artificial knee and hip joint with various defects are made in the Opera software. Shape of nonconductive defects is rectangular. Results of these simulations are verified experimentally on the material SUS316L with the same properties. Dimension of prostheses models and location of defects are created on the basis of real implants. All simulations are carried out by absolute type probe, which is positioned axially perpendicular to the material at a distance of h = 0.5 mm. The both mentioned joints contain five surface defects with the same thickness w = 0.2 mm and width l = 10 mm. The depth is different for each defect. Depth of defects in the knee joint (Fig. 1) are d = 1, 3, 5, 7, 10.5 mm and those in the hip joint are 1, 2, 3, 4 and 5mm deep. Fig. 1: Evaluated knee joint and positioning of the probe. The probe is positioned over intact place, meaning without defect to obtain reference value at first. Subsequently the probe movement over investigated material is defined over a distance <-10 mm; 10 mm>, hence 21 numerical models are created for the hip joint. In the knee join, there are made 31 numerical models in the range <-15 mm; 15 mm>. The position 0 represents to the centre of defect. The measurements are performed by three various frequencies f = 1 khz, 20 khz and 100 khz. The frequency displaying the biggest magnitude difference between individual simulated depth signals is selected. The simulations are created with the same mesh resolution in order to be able to compare results and to get on for the least possible error rate. The mesh of part of hip joint is shown in Fig. 2 and knee joint is displayed in Fig. 3. Numerical simulation The OPERA simulation package is based on the finite element method (FEM) and it is employed for simulations of electromagnetic field issues. The created models represent the artificial joints and contained defects are modeled in said software. The conductivity of the plate made of austenitic steel SUS316L is σ = 1.4x10 6 S/m and conductivity of the defect is σ d = 0. Fig. 2: The used mesh of hip joint. 6

compared. It can be concluded that the measurements made by numerical simulations are able to detect these surface defects using all the three frequencies. In fact, the experimental detection is successful only with the highest frequency applied. All results are viewed for the highest of the applied frequencies, therefore f = 100 khz. Results of numerical simulations Fig. 3: The mesh of knee joint. Experimental verification The results of knee joint are verified by commercial eddy current equipment ELOTEST B300 manufactured by Rohmann and LOCK-IN amplifier. The experiments are carried out using a plate made of austenitic steel contained artificial electro-discharge machined (EDM) slits. The dimensions of this plate are the following: length l p = 0.12 m, width w p = 0.5 mm and height h p = 0.01 mm. Depths of defects are d = 1, 3, 5, 7 and 9 mm. Other parameters, meaning conductivity of material and defects, thickness and width of defects and used frequencies are same as in the numerical simulations. The distance between individual defects is 0.06 m, so there is no overlap between resulting signals of neighboring cracks. A 3-axis manipulator controlled by a PC, with an accuracy of 10-6 m is used for precise regulation of the probe movement. The measurement probe of absolute type is located over the individual defects. The distance from the inspected object is h = 0.5 mm due to encapsulation of probe. This probe is moved by the manipulator fluently from the place with the shallowest defect towards the deepest defect and the measurement is repeated for the three frequencies f = 1 khz, 20 khz and 100 khz. The probe is fed with harmonic voltage generator and output signal is obtained using Lock-in amplifier and evaluated parameter is the voltage output of magnetic sensor. Lock-in amplifier in the realization of experiments serves as a filtering element whose aim is to filter out all signals barring the signal with a frequency that is probed supplied. The real and imaginary values of magnetic flux density are obtained from Opera software. Consequently, the dependence of magnetic flux density on the probe position is created by them. In Fig. 4 and Fig. 5 it is shown mentioned dependences of five different defects. B mod is module of magnetic flux density and the following expression can be used: B 2 2 mod Re{ B z } Im{ Bz} (1) Fig. 4: Dependence of magnetic flux density on the probe position <-15 mm, 15 mm> for the knee joint. Results Results of the numerical simulations and experiments for artificial prostheses by three exciting frequencies are Fig. 5: Dependence of magnetic flux density on the probe position <-10 mm, 10 mm> for the hip joint. 7

Fig. 4 displays the results of numerical simulations of the knee joint with depth of the defect d = 1, 3, 5, 7 and 10.5 mm. Fig. 5 shows the numerical simulation results of the hip joint with defect depth d = 1, 2, 3, 4 and 5mm. The both graphs displayed the smallest deflection when the defect with depth d = 1 mm and conversely the biggest deflection represents the deepest defect. Next experiments are designed for comparison the results obtained. The results of measurement by Lock-in amplifier are displayed in Fig. 8 and Fig. 9. These results confirm that the frequency f = 100 khz is suitably selected and the smallest frequency is not usable. Experimental results The results from commercial eddy current equipment are displayed in Fig. 6 and Fig. 7, which represent complex plane display and probe position signal dependences. Three frequencies are used in this study but only frequency of f = 100 khz detects all defects and it displays them as five different defects. In the Fig. 6 it is displayed the result of detection with frequency of f = 1 khz, for comparison. This frequency can not to detect any defects and it is not able to trace that measurement is made for five different defects. Therefore, it can be concluded that these types of defects are not possible to detect by the smallest used frequency. Fig. 8: Dependence of the voltage output of magnetic sensor and probe position <50mm; 50 mm> at a frequency f = 100 khz. Fig. 6: Dependence of differential response signal on probe movement for f = 100 khz. Fig. 9: Dependence of the voltage output of magnetic sensor and probe position <50mm; 50 mm> at a frequency f = 1 khz. Conclusion Fig. 7: Dependence of differential response signal on probe movement for f = 1 khz. Eddy current testing is powerful tool for evaluation of conductive biomaterials. The knee and hip joints belong to the most often implanted joints but they are the most frequently exchanged, too. Therefore, the verification of their quality is necessary before they are inserted into the body as damaged materials may cause some problems. The purpose of this article was to find the frequency which successfully detects five different defect with various depths and displays differences between them. Results were obtained using numerical simulations from Opera software and they were experimentally verified subsequently. 8

The most suitable frequency f = 100 khz was determined based on these measurements. Detections by numerical simulation were successful with all frequencies and their graphs provide opportunity to determine the depth of each defect inside the evaluated material. The experimental measurements representing the reality have demonstrated that these types of defect, it means surface defects can be detected only with higher frequencies. It can be concluded that chosen suitable excitation frequency and type of used probe is very important for non-destructive testing by eddy current method. Eddy current testing requires a very small distant between the probe and evaluated material, hence this method is possible to use only for applications outside the human body, it means as the final inspection of prostheses before their implantation. [3] Endoprotézy kloubní náhrady, [online], [2014/05/07], http://www.lekari-online.cz/ortopedie/novinky/endoprotezynahrady [4] Kreidl M., Šmíd R. Technická diagnostika, vyd. BEN-technická literatúra, Praha 2006. IBSN 80-7300-158-6. [5] NDT Method Summary. [online], [2014/05/07], http://www.ndted.org/generalresources/methodsummary/methodsummary.ht m [6] Kopec B. Nedestruktivní zkoušení materiálů a konstrukcí, vyd. Akademické nakladatelství CERM, Brno 2008. ISBN 978-80- 7204-591-4. [7] Rao, B. P. C. Practical eddy current testing. Alpha Science International Limited, 2007. ISBN: 978-1-84265-299-2. Acknowledgement This work was supported by the Slovak Research and Development Agency under the contracts No. APVV- 0349-10 and APVV-0194-07. References [1] Výročná správa za rok 2011. [online], [2014/05/07], http://sar.mfn.sk/dokumenty.327.html [2] Pernisova V., Capova K., Smetana M., Strapacova T., Biomaterial real defects evaluation using advanced detection sensors, ENDE 2013, Proceedings, ISBN 978-80-554-0713-5. Ing. Andrea Stubendekova Department of Electromagnetic and Biomedical Engineering Faculty of Electrical Engineering University of Zilina Univerzitna 1, 010 26 Žilina andrea.stubendekova@fel.uniza.sk Phone: +421 415 135 062 9