PXI Contents. PXI-2584 Specifications DEVICE SPECIFICATIONS. 300 V CAT II, 600 V CAT I Multiplexers

Similar documents
PXI Contents. PXI-2585 Specifications. Topology DEVICE SPECIFICATIONS. 10-Channel Multiplexer

NI PXIe Contents. About These Specifications DEVICE SPECIFICATIONS. 3 GHz Ω Multiplexer (SP16T)

NI PXI Contents. About These Specifications DEVICE SPECIFICATIONS GHz Dual Transfer Switch 50 Ω Relay Module

Refer to the NI Switches Help at ni.com/manuals for detailed topology information.

NI PXI Contents DEVICE SPECIFICATIONS. RF Attenuator

Related Information Refer to the NI Switches Help for detailed topology information.

USRP Contents. Definitions. Conditions SPECIFICATIONS. Software Defined Radio Device

USRP Contents. Definitions. Conditions SPECIFICATIONS. Software Defined Radio Device

USRP Contents. Definitions. Conditions SPECIFICATIONS. Software Defined Radio Reconfigurable Device

USRP Contents. Definitions SPECIFICATIONS. Software Defined Radio Reconfigurable Device

NI PXI/PXIe Contents DEVICE SPECIFICATIONS. 40 A Fault Insertion Unit

NI PXI-2521 Specifications 40-Channel DPST Relay Module

USRP Contents. Definitions SPECIFICATIONS. 50 MHz to 2.2 GHz Tunable RF Transceiver

NI PXI/PXIe Contents DEVICE SPECIFICATIONS. 10 A Fault Insertion Unit

1- or 2-Port, High-Speed, Flexible Data Rate PXI CAN Interface Module

PXI Contents SPECIFICATIONS. ±20 V, 40 W PXI Source Measure Unit

USRP Contents. Definitions SPECIFICATIONS. 10 MHz to 6 GHz Tunable RF Transceiver

PXI Contents. Definitions SPECIFICATIONS. ±100 V, 2 W Precision PXI Source Measure Unit. These specifications apply to the PXI-4132.

CDA Contents. Definitions SPECIFICATIONS. Clock Distribution Device

PXIe Contents SPECIFICATIONS. RF Amplifier

1- or 2-Port, Software-Selectable PXI CAN Interface Module

High-Speed/Flexible Data-Rate CAN Characteristics 5 VDC

NI PXI-2520 Specifications 80-Channel SPST Relay Module

PXIe Contents SPECIFICATIONS. RF Preselector Module

100-SPST (latching), 50-DPST. Refer to the NI Switches Help at ni.com/manuals for detailed topology information.

NI PXI/PXIe Contents. About These Specifications DEVICE SPECIFICATIONS. Quad 2 1 Terminated Multiplexer

PXI Contents SPECIFICATIONS. 1.3 GHz RF Analog Signal Generator

100 MHz Bandwidth, 12-Bit PXI Waveform Generator

PXI Contents SPECIFICATIONS. 6.6 GHz RF Analog Signal Generator

PXIe Contents SPECIFICATIONS. 3.3 GHz RF Analog Signal Generator

NI PXI/PXIe-2529 Specifications

This document lists the specifications for the NI PXI The PXI-5900 is a differential amplifier.

Related Information Refer to the NI Switches Help for detailed topology information.

PXIe Contents SPECIFICATIONS. 250 khz to 20 GHz RF Analog Signal Generator

PXIe Contents. Definitions SPECIFICATIONS. 8.5 GHz Vector Network Analyzer

NI PXI-2797 Specifications 40 GHz 6 1 Terminated 50 Ω Multiplexer

NI PXI/PXIe-2540 Specifications

PXIe Contents. Definitions SPECIFICATIONS. 8.5 GHz Vector Network Analyzer

NI PXI/PXIe-2527 Specifications

SA1500X. Operating Basics. Cautions DEVICE SPECIFICATIONS. High Impedance Active Probe

4-Channel, ±6 V, 500 ma Precision PXI Source Measure Unit

4-Channel, ±24 V, 150 ma PXI Source Measure Unit

4-Channel ±10 V, 100 ma, Precision PXI Source Measure Unit

4-Channel ±10 V, 100 ma, Precision PXI Source Measure Unit

NI PXI/PXIe-2541 Specifications 300 MHz Ω Matrix

Note Using the NI 5751/5751B in a manner not described in this document may impair the protection that the NI 5751/5751B provides.

4-Channel, ±24 V, 150 ma Precision PXI Source Measure Unit

NI PXI-2557 Specifications

NI USB Contents DEVICE SPECIFICATIONS. 50 MS/s Bus-Powered USB Oscilloscope

SCXI MHz Quad 8x1 50 Ω Multiplexer

USRP Contents. Definitions SPECIFICATIONS. USRP Software Defined Radio Stand-Alone Device

NI 2865A 0.3 A Matrix Cards for NI SwitchBlock

NI PXI/PXIe-2532 Specifications

NI PXI-2555 Specifications

NI PXI-2530 Specifications

16-Bit, 120 MS/s, 12 Channel DC/16 Channel AC Receiver Adapter Module

PXIe Contents SPECIFICATIONS. 2-Channel, 60 V, 1 A PXI Programmable Power Supply. These specifications apply to the PXIe-4112.

NI roborio. Contents. Processor SPECIFICATIONS. RIO Device for Robotics

PCI Express, 6½-Digit, ±300 V Digital Multimeter Device

PXIe, 6½-Digit, ±300 V, Onboard 1.8 MS/s Isolated Digitizer,PXI Digital Multimeter

PXIe Contents SPECIFICATIONS. ±60 V, 3 A System PXI Source Measure Unit. These specifications apply to the PXIe-4138.

6½-Digit, ±300 V, 1.8 MS/s Isolated Digitizer Included, PCI Digital Multimeter Device

PXIe Contents SPECIFICATIONS. ±200 V, 1 A System PXI Source Measure Unit. These specifications apply to the PXIe-4136.

Specifications are typical and valid from -40 C to +85 C unless otherwise noted. Galvanic isolation between channels and to chassis

PXIe Contents SPECIFICATIONS. ±200 V, 1 A Precision System PXI Source Measure Unit. These specifications apply to the PXIe-4137.

NI SCXI Specifications

PXIe Contents SPECIFICATIONS. ±60 V, 3 A Precision System PXI Source Measure Unit. These specifications apply to the PXIe-4139.

Current ON state current ma minimum, 6 ma typical, 10 ma maximum OFF state current ma maximum

mmwave Transceiver System

80 MHz Bandwidth, 16-Bit PXI Waveform Generator. These specifications apply to the one-channel and two-channel PXIe-5433.

Power Amplifier. Input Characteristics. SPECIFICATIONS NI PXIe-4610 Power Amplifier. Signal Range

8-Ch 24 V Sinking DI, 8-Ch 24 V Sourcing DO SIL3 Capable

NI 9207 DATASHEET. 16 AI, ±20 ma/±10 V, 24 Bit, 500 S/s Aggregate

PXI Contents SPECIFICATIONS. 8-Channel, 12-Bit, 60 MHz PXI Oscilloscope

PCI Contents SPECIFICATIONS. 8-Channel, 12-Bit, 60 MHz PCI Oscilloscope Device

NI 9220 DATASHEET. 16 AI, ±10 V, 16 Bit, 100 ks/s/ch Simultaneous. DSUB or push-in springterminal. 250 Vrms, CAT II, channelto-earth

PXIe, 40 MHz Bandwidth, 16-Bit PXI Waveform Generator. These specifications apply to the one-channel and two-channel PXIe-5423.

16 AI (16-Bit, 250 ks/s), 4 DI, 4 DO USB Multifunction I/O Device

8-Channel, 12-Bit, 60 MHz PXI Express Oscilloscope

PXI Contents SPECIFICATIONS. 6½-Digit, ±300 V, Onboard 1.8 MS/s Isolated Digitizer, L and C Measurement Support, PXI Digital Multimeter

Input Characteristics. Measurement Range. SPECIFICATIONS NI PXIe-4357

PXIe Contents SPECIFICATIONS. Precision, 10 fa Sensitivity Low-Current PXI Source Measure Unit. These specifications apply to the PXIe-4135.

PXIe, 80 MHz Bandwidth, 16-Bit PXI Waveform Generator. These specifications apply to the one-channel and two-channel PXIe-5433.

NI 9216 DATASHEET. 8 RTD, 0 Ω to 400 Ω, 24 Bit, 400 S/s Aggregate, PT100

4-Slot, Extended Temperature, Ethernet CompactDAQ Chassis

PXI Contents SPECIFICATIONS. 1 GHz Bandwidth, 2 GS/s, 8-Bit PXI Oscilloscope

NI 9213 DATASHEET. 16 TC, ±78 mv, 24 Bit, 75 S/s Aggregate

8-Slot, Extended Temperature, Ethernet CompactDAQ Chassis

100 MHz, 32-Channel, 5 V PXI Digital Waveform Instrument

NI 9770 DATASHEET. 30 khz to 100 MHz, 50 Ω, AC Coupled RF Receiver Module

R Series for USB Multifunction RIO with Kintex-7 70T FPGA. may impair the protection the NI USB-7855R provides.

NI PXI-2532 Specifications

GETTING STARTED GUIDE NI Port DeviceNet Module

PXI Contents SPECIFICATIONS. 300 MHz Bandwidth, 2 GS/s, 8-Bit PXI Oscilloscope

PXIe-6570 DEVICE SPECIFICATIONS. 32-Channel Digital Pattern Instrument

NI 9881 GETTING STARTED GUIDE. 1-Port CANopen Module

NI 9213 DATASHEET. 16 TC, ±78 mv, 24 Bit, 75 S/s Aggregate

R Series for USB Multifunction RIO with Kintex-7 70T FPGA

GETTING STARTED GUIDE NI Port, High-Speed CAN Module. Français Deutsch 日本語한국어简体中文. ni.com/manuals

NI 9853 GETTING STARTED GUIDE. 2-Port, High-Speed CAN Module. This document explains how to connect to the NI Français Deutsch 日本語한국어简体中文

Transcription:

DEVICE SPECIFICATIONS PXI-2584 300 V CAT II, 600 V CAT I Multiplexers This document lists specifications for the PXI-2584 high-voltage multiplexer switch module. All specifications are subject to change without notice. Visit ni.com/manuals for the most current specifications. Contents PXI-2584 Specifications... 1 Topology... 2 Input Characteristics... 2 Dynamic Characteristics... 3 Trigger Characteristics... 4 Physical Characteristics... 4 Environment...4 Shock and Vibration...5 Compliance and Certifications...5 Diagrams... 7 Accessories...9 PXI-2584 Specifications Specifications characterize the warranted performance of the instrument under the stated operating conditions. Data in this document are Specifications unless otherwise noted. Typical Specifications are specifications met by the majority of the instrument under the stated operating conditions and are tested at 23 C ambient temperature. Typical specifications are not warranted. All voltages are specified in DC, AC pk, or a combination unless otherwise specified. Caution The protection provided by the PXI-2584 can be impaired if it is used in a manner not described in this document.

Topology Refer to the NI Switches Help at ni.com/manuals for detailed topology information. Topologies 1-wire 12 1 multiplexer 1-wire dual 6 1 multiplexers 2-wire 6 1 multiplexer 2-wire 11 1 interleaved multiplexer Independent Input Characteristics Caution The PXI-2584 is not EMC protected and may generate emissions interference or disturbance. Relay contact actuation can cause unwanted emission disturbance in which case the installer or user will be required to take suppression measures such as, shielded cables, metal conduits, grounding, filtering, and protection techniques necessary to mitigate the source of interference or disturbance. The PXI-2584 is intended for use in industrial installations in which the user provides EMC controls. Caution When hazardous voltages (>42.4 Vpk/60 V DC) are present on any channel, safety low-voltage ( 42.4 Vpk/60 V DC) cannot be connected to any other channel. Caution This module is rated for 300 V Category II and 600 V Category I. This module can withstand up to 2,500 V impulse voltage. Do not use this module for connection to signals or for measurements within Categories III or IV. Do not connect this module to MAINs Category II circuits when operated above 300 V. Maximum switching voltage Channel-to-ground Channel-to-channel 300 V, Measurement Category II 600 V DC, V ACpk, Measurement Category I 1 300 V Caution The switching power is limited by the maximum switching current and the maximum voltage. Switching power must not exceed 10 W. Maximum switching power (per channel) DC systems Maximum current (switching or carry, per channel or common) 10 W 0.5 A 1 Measurement Categories CAT I and CAT O are equivalent. These test and measurement circuits are not intended for direct connection to the MAINS building installations of Measurement Categories CAT II, CAT III, or CAT IV. 2 ni.com PXI-2584 Specifications

Note Switching inductive loads (for example, motors and solenoids) can produce high voltage transients in excess of the module s rated voltage. Without additional protection, these transients can interfere with module operation and impact relay life. For more information about transient suppression, visit ni.com/info and enter the Info Code relayflyback. DC path resistance (1-wire path) Initial End-of-life 1.1 Ω >2 Ω DC path resistance typically remains low for the life of the relay. At the end of relay life, the path resistance rapidly rises above 2 Ω. Load ratings apply to relays used within the specification before the end of relay life. Minimum switching capacity Bandwidth (-3 db, 50 Ω termination) Isolation (50 Ω termination) Open channel 0.5 V 1 ma 900 khz, typical 10 khz >86 db, typical 100 khz >76 db, typical 1 MHz >58 db, typical DC open channel isolation Thermal EMF (1-wire path, channel-tocommon) Dynamic Characteristics Relay operate time >1.0 10 11 Ω, typical <50 μv, typical 0.4 ms, typical 0.81 ms, maximum Note Certain applications may require additional time for proper settling. Refer to the NI Switches Help at ni.com/manuals for more information about including additional settling time. Maximum scan rate Expected relay life Mechanical 600 channels/s 1 V at 10 ma resistive 3 10 8 cycles PXI-2584 Specifications National Instruments 3

Electrical 600 V at 7 ma 90 pf capacitive 1.8 10 7 cycles 600 V at 16.5 ma 90 pf capacitive 7 10 6 cycles Note Reed relays are highly susceptible to damage from in-rush currents. Switching capacitive loads without resistive or inductive protection can weld the relay contacts in less than 5 10 5 cycles. Note Relays are field replaceable. Refer to the NI Switches Help at ni.com/manuals for more information about replacing a failed relay. Trigger Characteristics Input trigger Sources PXI trigger lines <0...7> Minimum pulse width Output trigger 150 ns Note The PXI-2584 can recognize trigger pulse widths less than 150 ns if you disable digital filtering. Refer to the NI Switches Help at ni.com/manuals for information about disabling digital filtering. Destinations PXI trigger lines <0...7> Pulse width Physical Characteristics Relay type Relay contact material I/O connector PXI power requirement Dimensions (L W H) Weight Environment Maximum altitude Pollution Degree 2 Programmable (1 µs to 62 µs) Reed Rhodium MINI-COMBICON, 3.81 mm (16 position) 1.6 W at 5 V 0.2 W at 3.3 V 3U, one slot, PXI/cPCI module, 21.6 cm 2.0 cm 13.0 cm (8.5 in. 0.8 in. 5.1 in.) 212 g (7.5 oz) 2,000 m (at 25 C ambient temperature) 4 ni.com PXI-2584 Specifications

Indoor use only. Operating Environment Ambient temperature range Relative humidity range Storage Environment Ambient temperature range Relative humidity range Shock and Vibration Operational shock Random vibration Operating Nonoperating Compliance and Certifications 0 C to 55 C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.) 10% to 90%, noncondensing (Tested in accordance with IEC 60068-2-56.) -20 C to 70 C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.) 5% to 95%, noncondensing (Tested in accordance with IEC 60068-2-56.) 30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.) 5 Hz to 500 Hz, 0.31 g rms (Tested in accordance with IEC 60068-2-64.) 5 Hz to 500 Hz, 2.46 g rms (Tested in accordance with IEC 60068-2-64. Test profile exceeds the requirements of MIL-PRF-28800F, Class 3.) Safety This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use: IEC 61010-1, EN 61010-1 UL 61010-1, CSA 61010-1 Note For UL and other safety certifications, refer to the product label or the Online Product Certification section. PXI-2584 Specifications National Instruments 5

Electromagnetic Compatibility This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use: EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity EN 55011 (CISPR 11): Group 1, Class A emissions AS/NZS CISPR 11: Group 1, Class A emissions FCC 47 CFR Part 15B: Class A emissions ICES-001: Class A emissions Note In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia, and New Zealand (per CISPR 11), Class A equipment is intended for use only in heavy-industrial locations. Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes. Note For EMC declarations and certifications, refer to the Online Product Certification section. CE Compliance This product meets the essential requirements of applicable European Directives, as follows: 2014/35/EU; Low-Voltage Directive (safety) 2014/30/EU; Electromagnetic Compatibility Directive (EMC) Online Product Certification Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/ certification, search by model number or product line, and click the appropriate link in the Certification column. Environmental Management NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers. For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document. 6 ni.com PXI-2584 Specifications

Waste Electrical and Electronic Equipment (WEEE) EU Customers At the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/environment/weee. 电子信息产品污染控制管理办法 ( 中国 RoHS) 中国客户 National Instruments 符合中国电子信息产品中限制使用某些有害物质指令 (RoHS) 关于 National Instruments 中国 RoHS 合规性信息, 请登录 ni.com/environment/rohs_china (For information about China RoHS compliance, go to ni.com/environment/rohs_china.) Diagrams Refer to the following figure for the PXI-2584 power-on state: PXI-2584 Specifications National Instruments 7

Figure 1. PXI-2584 Power-On State CH0 K0 CH1 K1 K2 CH2 COM0 K5 CH5 REF REF K6 CH6 K7 CH7 K8 CH8 COM1 K11 CH11 Refer to the following figure for the PXI-2584 front panel connector pinout: 8 ni.com PXI-2584 Specifications

Figure 2. PXI-2584 Connector Pinout REF CH0 CH6 CH1 CH7 CH2 CH8 CH3 CH9 CH4 CH10 CH5 CH11 COM0 COM1 REF Note For topology-specific connection information, refer to your device in the NI Switches Help at ni.com/manuals and associated cable or terminal block installation instructions. Accessories Visit ni.com for more information about the following accessories. Caution You must ensure that devices and circuits connected to the PXI-2584 are properly insulated from human contact. Use the PXI-2584 BACKSHELL custom cable kit to ensure that the terminals are not accessible. Table 1. NI Accessories for the PXI-2584 Accessory Part Number PXI-2584 BACKSHELL custom cable kit 779168-02 Use 28 AWG to 16 AWG signal wires with the appropriate voltage rating, safety certifications, and wiring methods for the applied load. PXI-2584 Specifications National Instruments 9

Caution You must install mating connectors according to local safety codes and standards and according to the specifications provided by the connector manufacturer. You are responsible for verifying safety compliance of third-party connectors and their usage according to the relevant standard(s), including UL and CSA in North America and IEC and VDE in Europe. Plug screw terminal torque Wire strip length Tighten the screws to 0.5 to 0.6 N m (4.4 to 5.3 lb in.) of torque. Strip 7 mm (0.25 in.) of insulation from the ends of the signal wires. Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on NI trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/ legal/export-compliance for the NI global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015. 2006 2016 National Instruments. All rights reserved. 371667G-01 Sep16