Induction coil Magnetometer LEMI-120 User Manual

Similar documents
User manual for LEMI-029 digital fluxgate sensor system with KMS820 USER MANUAL. for LEMI 029 DIGITAL FLUXGATE SENSOR SYSTEM

Installation & Operation Manual SAGA1-K Series Industrial Radio Remote Control

HIGH VOLTAGE AND CURRENT CUT-OFF CAPACITY IN A COMPACT PACKAGE FEATURES

EA3012 TRANSCONDUCTANCE AMPLIFIER OPERATION MANUAL

INDEX PREFACE... 1 CAUTIONS... 2 OPERATION ON SITE(9) STANDARD INSTRUMENT... 3 OPTIONAL ACCESSORIES... 4 OPERATION OF TRANSMITTER(3)...

DC Current Transducers CT-200 CT-300 CT-400 User s Manual All Rights Reserved CAEN ELS d.o.o. Rev. 1.0 November 2014

Operation manual. Model 24 Infrasound Sensor

HIGH VOLTAGE AND CURRENT CUT-OFF CAPACITY IN A COMPACT PACKAGE

HIGH VOLTAGE AND CURRENT CUT-OFF CAPACITY IN A COMPACT PACKAGE

Broadband Power Amplifier

SL300 Snow Depth Sensor USL300 SNOW DEPTH SENSOR. Revision User Manual

SI-125 Power Amplifier Manual 6205 Kestrel Road; Mississauga, Ontario; Canada; L5T 2A1 November 2016, Rev 0.5

WIRELESS INSULATOR POLLUTION MONITORING SYSTEM

For ultra-high precision measurement of current: DC, AC, pulsed..., with galvanic separation between primary and secondary. Applications.

Single Channel Loop Vehicle detector User Manual

USER. manual. Falco Systems WMA-100. High Voltage Amplifier DC - 500kHz

NEW APPROACH TO THE EXACT DESIGN OF LOW NOISE SEARCH-COIL MAGNETOMETERS

Inductive Proximity Detectors Technical Guide

Model 7000 Low Noise Differential Preamplifier

MAGNETIC FIELD METER Operator s Manual

Order/Technical Support Tel: (800) / FAX: (800) /

USER MANUAL. DC 13kHz High Voltage Amplifier WMA V to +175V output. Stable with all capacitive loads, generates no overshoot

34134A AC/DC DMM Current Probe. User s Guide. Publication number April 2009

AMP-13 OPERATOR S MANUAL

PRECISION CURRENT TRANSDUCERS. DC Current Transducers CT-100 CT-150. User s Manual. All Rights Reserved CAEN ELS d.o.o. Rev. 1.

USER MANUAL. Ultra-Low Noise High Voltage Amplifier WMA V to +150V output. 300µV rms output noise. 2mV output offset voltage

Current Probes. User Manual

M1 Series. Humidity - Temperature Transmitter INSTRUCTION MANUAL

Roline L1 Series. Humidity - Temperature Transmitters INSTRUCTION MANUAL

V6 V10 V35 V45 V218-S V221-S

AC/DC Current Probe CT6844/CT6845/CT6846

For ultra-high precision measurement of current: DC, AC, pulsed..., with galvanic separation between primary and secondary. Applications.

GOLDEN MASK DEEP HUNTER LE

World-Class Accuracy & Measurement Range (40 Conventional Models)

Broadband Current Probe Series Operation Manual

For ultra-high precision measurement of current: DC, AC, pulsed..., with galvanic separation between primary and secondary. Applications.

TECHNICAL MANUAL. SERIES AP5103 DIN-Rail DC Strain Gage Conditioner ISO 9001/AS9100

For the electronic measurement of current: DC, AC, pulsed..., with galvanic separation between the primary and the secondary circuit.

STAFF User Manual. Manual Part #

2015 RIGOL TECHNOLOGIES, INC.

Application Note (Revision NEW) Original Instructions. EMI Control in Electronic Governing Systems

CD770 DIGITAL MULTIMETER INSTRUCTION MANUAL

Loop-Powered Isolators for Standard Signals

Advanced Test Equipment Rentals ATEC (2832)

87.5 TO MHz BAND II 2 WAY 4.8dBi STACKED DIPOLE ANTENNA

TVA 080 / 100 / 120 / 180 / 200 / 220 Isolation amplifier active

Amplifier Series BASIC. Installation & Operations Manual

AC/DC Current Oscilloscope Probe Model SL261

DC Current Transducers 4XCT-100-C-HP User s Manual All Rights Reserved CAEN ELS s.r.l. Rev. 1.0 February 2017

Overtravel of 3.5 mm max. Power source DC D5C-1DS0 D5C-1DP0 D5C-1DA0 AC D5C-1AS0 D5C-1AP0 D5C-1AA0 Antenna only D5C-00S0 D5C-00P0 D5C-00A0

Telemetrie-Messtechnik Schnorrenberg

New INTERMAGNET Fluxgate Magnetometer

Technical Specification Temperature-Humidity-Pressure Sensor

DUAL OUTPUT AC CURRENT/VOLTAGE TRANSDUCER

Programmable transmitter of temperature, relative humidity and other derived humidity values with 4-20 ma outputs Instruction Manual

SAS-563B Active Loop Antenna Operation Manual

For ultra-high precision measurement of current: DC, AC, pulsed..., with galvanic separation between primary and secondary. Applications.

Instruction Manual MSC710 MSC710-U MSC710-I

Current Probe Fixture Instruction Manual

OPERATION AND MAINTENANCE MANUAL TRIAXIAL ACCELEROMETER MODEL PA-23 STOCK NO

Broadband Current Probe Series Operation Manual

Inductive Sensors. Fig. 1: Geophone

PixController, Inc. Wireless Vibration Sensor For Indoor and Outdoor Use

High Performance Current Transducer IT 200-S ULTRASTAB = A. ε L

Programmable transmitter of temperature, relative humidity and other derived humidity values with 4-20 ma outputs Instruction Manual

IsoAmp 3000/4000. The Advantages The dynamic range extends into the negative, allowing for strictly linear transmission

For the electronic measurement of current: DC, AC, pulsed..., with galvanic separation between the primary and the secondary circuit.

DM-45 Digital Multimeter

Programmable transmitter of temperature, relative humidity and other derived humidity values with 4-20 ma outputs Instruction Manual

Wireless Transceivers for Fuel Level Sensors RUd-DF-232 RUd-DF-485. Instruction Manual Version # 2.1 VTAS IM

Operating manual. TruTool N 500 (1A1) english

Programmable transmitter of temperature, relative humidity and other derived humidity values with 4-20 ma outputs Instruction Manual

COMPACT MOLECULAR-ELECTRONIC SEISMIC SENSORS

DUAL%CHANNEL%LINEAR%AMPLIFIER %WITH%PHASE%INVERTER Model&A400DI

MAGNETIC FIELD METER Operator s Manual

PCT200 SERIES CT/PT TEST SYSTEM USER MANUAL

Bulk Current Injection Probe Test Procedure

SITRANS F flowmeters. SITRANS F M System information MAGFLO electromagnetic flowmeters. 4/18 Siemens FI

For ultra-high precision measurement of current: DC, AC, pulsed..., with galvanic separation between primary and secondary. Applications.

DUNHAM & MORROW By Schonstedt Instrument Company Tel: Fax:

SMART Position Sensor, 100 and 180 Arc Configurations

TA 450 MK-X TA 600 MK-X TA 1050 MK-X TA 1400 MK-X TA 2400 MK-X power amplifier. user manual

For ultra-high precision measurement of current: DC, AC, pulsed..., with galvanic separation between primary and secondary. Applications.

Series 500. Owner s Manual. Analog Transmitters by Data Industrial. Data Industrial. Data Industrial 2/95 PN 72806

CAPSULE CONTACT MECHANISM AND HIGH-CAPACITY CUTOFF COMPACT RELAY FEATURES

Geotagger N3. User Manual (V1.0) Revised by Geosolve.be (Pol F. Gillard) with personal updates and help. Solmeta Technology Co.

DUAL CHANNEL BROADBAND LINEAR AMPLIFIER Model A800D

AK-18G Antenna Kit Operation Manual

Manostar Electronic Low Differential Pressure Measurement System

5. Transducers Definition and General Concept of Transducer Classification of Transducers

SENSOR AND MEASUREMENT EXPERIMENTS

Stake-less earth / ground testing

USER S MANUAL. Rail-mounted temperature transmitter ATL

Magnetic absolute encoder GEL 2035 Serial interface (SSI), 24 Bit

AMP-12 OPERATOR S MANUAL

PROXIMITY SENSOR TERMINOLOGY

SAS-562B Active Loop Antenna Operation Manual

Operating instructions D-37

TLE4990 TLE4990-E6782

EMI Installation Guidelines

Transcription:

EUSA UKRAIN Induction coil Magnetometer LEMI-120 User Manual LVIV

Disclaimer notice All information in this manual is accurate as of July 2014. However, the Laboratory for Electromagnetic Innovations cannot be held liable for any damage caused by possible inaccuracies in the Manual. The Laboratory for Electromagnetic Innovations reserves the right to make corrections, modifications, enhancements, improvements, and other changes to its products at any time and to discontinue any product or service without notice. The Customer is advised to obtain the latest relevant information before placing orders and verify that such information is current and complete. LEMI-120 NN555-557 User Manual 2

Contents 1. Introduction..... 2. Main Technical Parameters... 3. Delivery Set... 4. Service and Guarantee 5. Structure and Operation... 6. Preparation for Operation... 7. Utilization and Transportation... Appendix A. Typical Transformation factor values... Appendix B. Typical noise spectral density... Appendix C. Selecting the measurements site and conditions. 4 5 6 7 8 10 11 11 16 17 LEMI-120 NN555-557 User Manual 3

1. Introduction The induction coil magnetometer LEMI-120 is intended to be used for the study of magnetic field fluctuations in land conditions, in the frequency band 0.0001-1000 Hz. It can be used both autonomously with any analog registration unit and as a part of a magnetotelluric station. Extremely high sensitivity (Fig.1) ensures excellent signal-tonoise ratio for ground measurements. 10 4 pt/sqrt(hz) 10 2 10 0 natural noise 10-2 LEMI-120 noise 10-4 10-2 10 10 0 frequency, Hz 10 2 Fig.1. LEMI-120 noise level and natural electromagnetic noise background Low power consumption and wide power supply range allows for long period measurements at remote sites. A rugged and waterproof housing combined with a waterproof output connector prolong the sensor s active lifetime and ensure overall reliability. The circuit design and magnetometer construction use several industry specific procedures as well as new technological processes that provide the best possible combination of metrological and operational parameters. Each individual LEMI sensor is experimentally tested and certified. LEMI-120 NN555-557 User Manual 4

2. Main Technical Parameters Frequency band of received signals Shape of transfer function Transfer function corner frequency Transformation factor 1 at differential output (pins A and G) at the flat part 2 at the linear part 3 Transformation factor error Magnetic noise level at 0.001 Hz, at 0.01 Hz, at 1 Hz, at 100 Hz Length of connecting cable Power supply voltage (upper limit recommended) Maximum output voltage Current consumption (nominal) Temperature range of operation Outer dimensions Design Weight 0.0001-1000 Hz linear - flat 1 Hz 100 mv/nt 100* f mv/nt < 1 db 100 pt/ Hz 10pT/ Hz 0.1 pt/ Hz 0.01pT/ Hz 200 m ± (6 12) V ± 2.5 V + 14 ma -10 ma minus 20 + 50 o C l = 1290 mm d = 85 mm Rugged and waterproof 6 kg Notes: 1 The transfer function experimentally derived values are given in Appendix A. 2 For frequency band between 1 and 1000 Hz transformation factor remains the same for the measured signal within this frequency band. 3 For frequency band between 0.0001 and 1 Hz transformation factor depends on the frequency f of the measured signal. LEMI-120 NN555-557 User Manual 5

3. Delivery Set The magnetometer delivery set includes: Item Quantity 1 Induction magnetometer LEMI-120 1 3 2 Technical Description and Operation Manual 1 1 NN555-557 Fig.2. LEMI-120 induction magnetometers LEMI-120 NN555-557 User Manual 6

4. Service and Guarantee 4.1. The term of warranty is 18 months after delivery if all requirements of the present instruction such as applied voltage, weather conditions, vibrations and shocks are observed. During this term the manufacturer is liable to repair any defects that may occur through no fault of the consumer or force majeure, or in the event repair is not possible, to change the device by other equivalent specimen. 4.2. The manufacturer will make free service and repair calls to repair the magnetometer as necessary for two years. However, the user must cover the cost of all necessary spare parts and transportation/visit fees. 4.3. The repair time at the manufacturer must be no longer than 45 days. Delivery date: 10.07.2014 V. Pronenko Signature All questions as to the magnetometer structure and operation have to be addressed to the designer: pron@isr.lviv.ua Vira Pronenko. LEMI-120 NN555-557 User Manual 7

5. Structure and Operation 5.1. The induction coil magnetometer consists of an induction coil sensor and an electronic unit both of which are located inside a common protective housing. The front panel of the magnetometer has a MS3112E12-8S connector for cable coupling. 5.2. The sensor part consists of a magnetic core, the main winding W2 and a magnetic feedback winding W1 (Fig.3). The magnetic core is made of a number of µ-metal tapes, which are insulated one from another and installed inside the protective tube on which both windings are made. A set of electrostatic screens are installed to reduce the interference to a negligible minimum value. 5.3. The electronic unit consists of two circuit boards which are fixed to the internal part of the magnetometer front panel. A simplified functional diagram of the unit is presented in Fig. 3. R3 C1 R2 A2 X1 W1 W2 R1 A1 R4 C2 R5 LPF 1 100 Hz circuit OUT+ CAL + V - V POWERGND and SHIELD pin A B C D E 3 kohm CODE OUT- OUTGND F G H Fig.3. Simplified functional diagram of induction coil magnetometer LEMI-120 The output signal of the main winding W2 is coupled to the input of ultra lownoise modulator-demodulator amplifier A1. The local feedback loop consisting of R1, R4, R5, and C2 of amplifier A1 fixes its total amplification factor at low frequencies to approximately 200. Output of amplifier A1 through the correction circuit R2, R3, and C1 is coupled to the magnetic flux feedback winding W1. Magnetic and local feedbacks circuits help develop flat part of the transfer function of the magnetometer within the frequency band from 1 to ~ 1000 Hz. The output of A1 is also connected to a 6-th order LPF filter (1100 Hz cutoff frequency) with a gain of 15. Symmetrical output voltage is formed from the output of inverter A2. LEMI-120 NN555-557 User Manual 8

CAUTION: The A1 amplifier input is protected from damage by overloading signals but not from strong ones, such as nearby lightning! 5.4. The amplifier A1 was developed using the principle of modulationdemodulation (M-DM). The processing of input signals uses three stages: modulation of high frequency signal (f m =6000Hz) by input low frequency signal (DC to 1000Hz); amplification of the modulated high-frequency signal; demodulation of this modulated signal. NOTE: The amplifier has extremely low noise at the low frequency end, but each M-DM amplifier is sensitive to all signals in the band ±1000Hz around the frequencies n*f m. The LEMI-120 magnetometer suppresses these signals by 40dB, but it is not recommended to use the LEMI-120 in noisy environment with high content of upper frequencies, e. g., close to powerful LF transmitters. 5.5. A reference signal may be applied to pin B (CAL) of the output connector to test the device. Applying a 10 Hz 5V ptp signal to pin B (with pin H as the return wire) produces output signal about 2 nt ptp. NOTE: This procedure serves exclusively for IM operation checking. The IM calibration can not be done by giving signals to pins B and H, but only in specialized calibration facility! 5.6. The external housing of the induction coil magnetometer has a tubular shape and is made of a fiberglas tube with built-in electrostatic screen. The housing is also covered by a protective plastic tube to withstand environmental exposure, shocks and vibrations within allowed limits without damage. The front panel which includes the cable connector, electronic unit and induction coil sensor with all windings can be removed in order to access the housing interior. NOTE: Opening the sensor housing before the end of the warranty term without written permission from the manufacturer cancels all guarantee obligations. LEMI-120 NN555-557 User Manual 9

6. Preparation for Operation 6.1. Read this entire manual, check the delivery set and inspect the exterior or the unit for damages that might have occurred during shipping. 6.2. Couple the sensor to a measuring device and a power supply. See Fig.4 for recommended connection layout. X1 circuit OUT+ CAL + V - V POWERGND and SHIELD CODE OUT- OUTGND pin A B C D E F G H 100mV/nT ϕ=0 100mV/nT ϕ=π instrumental amplifier + - ADC test 10Hz 5Vptp To PC LEMI-120 Power Fig.4. Connection diagram for precise measurements. CAUTION: Be careful when connecting the magnetometer to the power supply! The magnetometer is protected from incorrect polarity, but is not protected against overvoltage of power supply! 6.3. If a performance check is required, then a test signal source must be connected as shown in Fig. 4. In regular operation the mode test signal source must be disconnected. 6.4. The LEMI-120 magnetometer is ready for operation within three minutes after power is applied. In order to obtain extremely low noise level at the lowest frequency it is recommended to increase the time interval between the installation of the sensor and measurement. LEMI-120 NN555-557 User Manual 10

7. Utilization and Transportation 7.1. The utilization conditions of LEMI-120 magnetometer are: temperature -10 ºC to +50ºC humidity up to 100% pressure 700 to 1100 hpa 7.2. Although the magnetometer housing is hermetical it is not advisable to expose the device to heavy rain or immerse it in water. 7.3. Avoid shocks to the magnetometer of more than 10 g and vibrations in a frequency band between 1 1000 Hz with accelerations more than 5 g. 7.4. Do not pull the cable to retrieve the magnetometer as the connector or cable can be damaged. LEMI-120 NN555-557 User Manual 11

Appendix A. Transformation factor values The first column contains frequency in Hz, the following columns contain transformation factor(s) values in mv/nt and phase shift in degrees of angle between input and output signals. Both are measured with error 2 % For all magnetometers the parameters at 0.0001 Hz, 0.001 Hz and 0.01 Hz are obtained by calculation. f, Hz 0.0001 0.001 0.01 S, mv/nt 0.02 0.2 2 Phase, 90 90 89 LEMI-120 555 f, Hz S, mv/nt Phaze, º 2.1000E-2 1.877E0 8.7022E1 3.1770E-2 2.912E0 8.8247E1 4.8062E-2 4.403E0 8.7471E1 7.2711E-2 6.642E0 8.5759E1 1.1000E-1 1.007E1 8.4529E1 1.3249E-1 1.234E1 8.7859E1 1.5959E-1 1.397E1 8.0891E1 1.9222E-1 1.729E1 8.0095E1 2.3153E-1 2.063E1 7.7175E1 2.7888E-1 2.461E1 7.4771E1 3.3590E-1 2.987E1 7.1209E1 4.0459E-1 3.551E1 6.9186E1 4.8733E-1 4.056E1 6.6878E1 5.8698E-1 4.789E1 6.1667E1 7.0702E-1 5.490E1 5.7426E1 8.5160E-1 6.045E1 5.2269E1 1.0257E0 6.870E1 4.6523E1 1.2355E0 7.405E1 4.1201E1 1.4881E0 8.065E1 3.5923E1 1.7925E0 8.575E1 3.0723E1 2.1590E0 8.980E1 2.6044E1 2.6005E0 9.220E1 2.2004E1 3.1323E0 9.475E1 1.8258E1 3.7728E0 9.620E1 1.4997E1 4.5443E0 9.735E1 1.2225E1 5.4736E0 9.820E1 9.7740E0 6.5929E0 9.890E1 7.5350E0 7.9411E0 9.915E1 5.6269E0 9.5649E0 9.945E1 3.8884E0 1.1521E1 9.970E1 2.2603E0 1.3877E1 9.975E1 6.5976E-1 LEMI-120 NN555-557 User Manual 12

1.6715E1 9.985E1-8.7464E-1 2.0132E1 9.990E1-2.4690E0 2.4249E1 9.995E1-4.1553E0 2.9208E1 9.990E1-5.9520E0 3.5181E1 1.000E2-7.9490E0 4.2375E1 1.001E2-1.0157E1 5.1041E1 1.008E2-1.2756E1 6.1478E1 1.001E2-1.6073E1 7.4050E1 1.001E2-1.9681E1 8.9192E1 9.980E1-2.4092E1 1.0743E2 9.985E1-2.9270E1 1.2940E2 9.965E1-3.5494E1 1.5586E2 9.940E1-4.2906E1 1.8773E2 9.925E1-5.1929E1 2.2612E2 9.880E1-6.2839E1 2.7236E2 9.810E1-7.6053E1 3.2806E2 9.700E1-9.2000E1 3.9514E2 9.510E1-1.1167E2 4.7595E2 9.185E1-1.3544E2 5.7328E2 8.700E1-1.6418E2 6.9050E2 8.085E1 1.6102E2 8.3171E2 7.530E1 1.1756E2 1.0018E3 7.145E1 5.2120E1 LEMI-120 556 f, Hz S, mv/nt Phaze, º 2.1000E-2 1.930E0 8.8941E1 3.1770E-2 2.915E0 8.8169E1 4.8062E-2 4.412E0 8.7295E1 7.2711E-2 6.639E0 8.6058E1 1.1000E-1 1.002E1 8.4086E1 1.3249E-1 1.225E1 8.1622E1 1.5959E-1 1.489E1 8.2228E1 1.9222E-1 1.722E1 7.9310E1 2.3153E-1 2.045E1 7.9072E1 2.7888E-1 2.508E1 7.5888E1 3.3590E-1 2.970E1 7.2319E1 4.0459E-1 3.356E1 7.1409E1 4.8733E-1 4.060E1 6.5826E1 5.8698E-1 4.716E1 6.1796E1 7.0702E-1 5.504E1 5.6888E1 8.5160E-1 6.172E1 5.2370E1 1.0257E0 6.874E1 4.6557E1 1.2355E0 7.500E1 4.1311E1 1.4881E0 8.033E1 3.5945E1 1.7925E0 8.565E1 3.1042E1 2.1590E0 8.940E1 2.6362E1 LEMI-120 NN555-557 User Manual 13

2.6005E0 9.208E1 2.2085E1 3.1323E0 9.459E1 1.8469E1 3.7728E0 9.620E1 1.5231E1 4.5443E0 9.730E1 1.2367E1 5.4736E0 9.831E1 9.8539E0 6.5929E0 9.881E1 7.6248E0 7.9411E0 9.920E1 5.6951E0 9.5649E0 9.946E1 3.9068E0 1.1521E1 9.973E1 2.2709E0 1.3877E1 9.985E1 7.3379E-1 1.6715E1 9.995E1-7.5158E-1 2.0132E1 1.000E2-2.4009E0 2.4249E1 1.000E2-4.0361E0 2.9208E1 1.000E2-5.8204E0 3.5181E1 1.000E2-7.8261E0 4.2375E1 1.001E2-1.0081E1 5.1041E1 1.004E2-1.2757E1 6.1478E1 1.001E2-1.5795E1 7.4050E1 1.002E2-1.9380E1 8.9192E1 1.001E2-2.3748E1 1.0743E2 1.001E2-2.8822E1 1.2940E2 1.000E2-3.4997E1 1.5586E2 1.000E2-4.2373E1 1.8773E2 9.993E1-5.1306E1 2.2612E2 9.976E1-6.2139E1 2.7236E2 9.939E1-7.5310E1 3.2806E2 9.868E1-9.1318E1 3.9514E2 9.718E1-1.1106E2 4.7595E2 9.425E1-1.3503E2 5.7328E2 8.949E1-1.6405E2 6.9050E2 8.307E1 1.6084E2 8.3171E2 7.695E1 1.1710E2 1.0018E3 7.164E1 5.1749E1 LEMI-120 557 f, Hz S, mv/nt Phaze, º 2.1000E-2 1.932E0 8.8675E1 3.1770E-2 2.911E0 8.8336E1 4.8062E-2 4.413E0 8.7296E1 7.2711E-2 6.674E0 8.6047E1 1.1000E-1 1.003E1 8.4029E1 1.3249E-1 1.163E1 8.2991E1 1.5959E-1 1.382E1 8.2393E1 1.9222E-1 1.772E1 8.1144E1 2.3153E-1 2.098E1 7.7756E1 2.7888E-1 2.472E1 7.5202E1 3.3590E-1 2.983E1 7.3010E1 4.0459E-1 3.404E1 7.0004E1 LEMI-120 NN555-557 User Manual 14

4.8733E-1 4.020E1 6.6565E1 5.8698E-1 4.720E1 6.2203E1 7.0702E-1 5.434E1 5.7369E1 8.5160E-1 6.156E1 5.2190E1 1.0257E0 6.860E1 4.6901E1 1.2355E0 7.533E1 4.1190E1 1.4881E0 8.054E1 3.5152E1 1.7925E0 8.557E1 3.0944E1 2.1590E0 8.956E1 2.5879E1 2.6005E0 9.229E1 2.2077E1 3.1323E0 9.443E1 1.8114E1 3.7728E0 9.628E1 1.5131E1 4.5443E0 9.744E1 1.2278E1 5.4736E0 9.820E1 9.8185E0 6.5929E0 9.886E1 7.5665E0 7.9411E0 9.927E1 5.6472E0 9.5649E0 9.972E1 3.9459E0 1.1521E1 9.982E1 2.2511E0 1.3877E1 9.993E1 6.8165E-1 1.6715E1 1.000E2-8.7940E-1 2.0132E1 1.001E2-2.4179E0 2.4249E1 1.001E2-4.0430E0 2.9208E1 1.002E2-5.8735E0 3.5181E1 1.001E2-7.8606E0 4.2375E1 1.001E2-1.0189E1 5.1041E1 1.012E2-1.3198E1 6.1478E1 1.001E2-1.5833E1 7.4050E1 1.002E2-1.9433E1 8.9192E1 1.000E2-2.3804E1 1.0743E2 1.000E2-2.8866E1 1.2940E2 9.999E1-3.5046E1 1.5586E2 9.990E1-4.2410E1 1.8773E2 9.982E1-5.1371E1 2.2612E2 9.958E1-6.2216E1 2.7236E2 9.916E1-7.5474E1 3.2806E2 9.711E1-9.1584E1 3.9514E2 9.613E1-1.1073E2 4.7595E2 9.325E1-1.3458E2 5.7328E2 8.854E1-1.6344E2 6.9050E2 8.231E1 1.6162E2 8.3171E2 7.645E1 1.1796E2 1.0018E3 7.176E1 5.1974E1 LEMI-120 NN555-557 User Manual 15

Appendix B. Typical noise spectral density Noise spectral density was measured in the field by subtraction of output signals of the two magnetometers placed in similar conditions. Green and blue curves are output signals spectra of two LEMI-120 magnetometers. The red curve is noise level of LEMI- 120 magnetometer. LEMI-120 NN555-557 User Manual 16

Appendix C. Selecting the measurements site and conditions Important! The induction sensors are very sensitive to the mechanical movements and vibrations. It is recommended to follow the advices below in order to realize their full sensitivity. 1) The measurement should be carried out in electromagnetically clean area and users should take into account possible magnetic signals from, e.g., nearby thunderstorms, solar bursts etc. The sensor should be located as far as possible from mains power lines and large ferromagnetic objects. It is difficult to give accurate recommendations about the distance to them as it depends on magnetic moment of such objects; the maximal distance between the sensors and registration unit should be no more than 200 m (recommended at least about 100 m). For example, we can clearly observe the process of car doors opening and closing at distances of about 25 meters. Do not select measurement sites near large trees - strong wind will produce soil oscillations and consequently sensor vibration. The distances between each of the three sensors in a three-component array have to be not less than 2 meters. 2) Every effort should be made to provide a stable position for the sensor (FYI, only 1" rotation of the sensor in the Earth magnetic field may produce ~250 pt of parasitic signal). In order to ensure a stable position it is necessary to make beds (holes) with concrete walls for permanent installations. 200mm 200mm l+200mm induction -coil sensor concrete walls and bottom >20mm thick ground Fig. 5. LEMI sensor permanent installation site All dimensions of the sensor hole are given for orientation - any deviations in bigger side are possible. The hole has to be covered with any hut against rain and wind, the external dimensions of which have to be at least 20 mm bigger than the hole dimensions. The hut has to be made from non-magnetic and nonconductive material. It is necessary to foresee small gaps between inner part of LEMI-120 NN555-557 User Manual 17

the hut and the edges of concrete walls in order to allow the outer air inside freely. The edges of the concrete walls have to be at least 50-100 mm above the ground level in order to protect from water when covered by the hut. To avoid water and moisture penetration into connector, we recommend encapsulate connector and adjacent part of the cable (~10-15 cm) together into heat-shrinkage insulation tube. If there are microseisms supposed in the region it is advisable to fill the concrete bed to half of depth with elastic material such as soft polyurethane, and to put the sensor on top of it, then to press from above by the heavy bag with sand. The first three to five meters of cable connected to the sensor must also be buried. 3) When installing a temporary station it is necessary to bury the sensor in the ground at the depth not less than 2-3 sensor diameters, to pile up about 2-3 cm layer of the sand on the bottom, put then the sensor and press from above by the bag with sand. Such an installation procedure allows for avoiding influence due to blowing wind and decreases sensor temperature fluctuations. If burying the sensor and cable is not possible (e.g., rocky ground) it is advisable that you put the heavy bag with sand on the sensor and cover against the rain. 4) Proceed as follows to connect the sensor to the acquisition system. Prepare a sensor cable, preferably shorter than 200 m, which should contain three twisted wires for the power supply, a shielded twisted pairs for the signal and one wire for analog ground. 5) It is important that a good quality power supply be used, especially when working in high-frequency range. The best way to ensure this is to use a battery placed on an insulating cover such as polyethylene film. When using DC-DC transformers in the receiving equipment, ensure that they do not produce noise, especially spikes, in their input circuits. 6) Earthing of the analog ground and common power supply lines is necessary and it is advisable that this grounding be made according to "star" system - in one point, using a buried copper stick to connect to it all ground wires together. LEMI-120 NN555-557 User Manual 18