FEA of Prosthetic Lens Insertion During Cataract Surgery

Similar documents
EXCHANGE. Financial Disclosure. Clinical pearls In advanced anterior segment surgery being able to do a IOL exchange is a must. Why Do an Exchange

IOL Types. Hazem Elbedewy. M.D., FRCS (Glasg.) Lecturer of Ophthalmology Tanta university

Unique Aberration-Free IOL: A Vision that Patients

Long-term quality of vision is what every patient expects

The Dysphotopsia Mystery. John J. Bussa, M.D.

THE ASPHERIC PRELOADED INJECTION

The complete choice in refractive lens solutions

Multifocal and Accommodative

(12) United States Patent (10) Patent No.: US 7,156,854 B2

What s New in Ocular Biomechanics?

FOR PRECISE ASTIGMATISM CORRECTION.

Clinical Evaluation 3-month Follow-up Report

ROTATIONAL STABILITY MAKES THE DIFFERENCE

Glistening-Free Hydrophobic Acrylic IOL. Glistening-Free Hydrophobic Acrylic IOL

Improving Lifestyle Vision. with Small Aperture Optics

Sulcoflex. For when perfection is the only option! Pseudophakic Sulcus Fixated Secondary IOLs. Sulcoflex Aspheric. Sulcoflex Toric

Evolution of Diffractive Multifocal Intraocular Lenses

Product Portfolio. Sulcoflex Pseudophakic Supplementary IOLs. Your skill. Our vision.

Product Portfolio. Sulcoflex Pseudophakic Supplementary IOLs

Customized intraocular lenses

Patient information. Your options for cataract treatment Enjoy clear vision at all distances with multifocal IOLs

Not everyone can do this. Introducing RayOne with patented Lock & Roll TM technology for the smallest fully preloaded IOL incision

HYDROPHILIC ACRYLIC ADVANCED POLYMER

FOR EXCELLENT VISION QUALITY TO SUPPORT AN ACTIVE LIFESTYLE

RayOne Hydrophobic IOL. New design. New standard MADE IN UK

Technicians & Nurses Program

The design is distinctive. The outcomes are clear. Defocus tolerance 1 Glistening-free performance 1,2 Predictable outcomes 1

Getting More From Abaqus by Tackling Noisy and Challenging Data

Dynamic Approach to Quasi-static Nonlinear Problems for Sub-Sea Applications

Corporate Perspective Alcon Unanswered Technical Challenges that Still Need to be Overcome

Intraocular lens Difficulties

Cataract Information. victoriaeye.com

WHY EDOF INTRAOCULAR LENSES? FOR EXCELLENT VISION QUALITY TO SUPPORT AN ACTIVE LIFESTYLE PATIENT INFORMATION. Cataract treatment

Crystalens AO: Accommodating, Aberration-Free, Aspheric Y. Ralph Chu, MD Chu Vision Institute Bloomington, MN

I N M A D E F R A N C E. Hydrophobic monofocal NEW QUALITY SAFETY STABILITY TECHNICAL BROCHURE

Dr. Magda Rau Eye Clinic Cham, Germany

NOW. Approved for NTIOL classification from CMS Available in Quar ter Diopter Powers. Accommodating. Aberration Free. Aspheric.

Influence of Lubrication and Draw Bead in Hemispherical Cup Forming

*Simulated vision. **Individual results may vary and are not guaranteed. Visual Performance When It s Needed Most

Paper Flow Simulation Using Abaqus

Sutureless, Glueless, Scleral Fixation of Single-Piece and Toric Intraocular Lens: A Novel Technique

CONTENTS. HYDROPHILIC IOLs. Bi-Flex PIL-MA hydrophilic preloaded P / PY 677PT / PTY 677PMY 677PMTY. Bi-Flex HL...

Experimental and numerical investigation of tube sinking of rectangular tubes from round section

The Behaviour Of Round Timber Sections Notched Over The Support On The Tension Face. Justin Dewey

AHEAD Superior technology, thoughtfully designed with you in mind for an intelligent approach to cataract surgery.

NEW THE WORLD S FIRST AND ONLY SINUSOIDAL TRIFOCAL IOL

AN INNOVATIVE FEA METHODOLOGY FOR MODELING FASTENERS

Corso di Studi di Fabbricazione

day night convinced supreme contrast sensitivity THE IOL FOR DAY & NIGHT

AT LISA tri 839MP and AT LISA tri toric 939MP from ZEISS The innovative trifocal IOL concept providing True Living Vision to more patients

New Materials for Perfect Vision

Retinal stray light originating from intraocular lenses and its effect on visual performance van der Mooren, Marie Huibert

Application Research on BP Neural Network PID Control of the Belt Conveyor

NEW. AT LISA tri 839MP and AT LISA tri toric 939MP from ZEISS The innovative trifocal IOL concept providing True Living Vision to more patients

Sheep Eye Dissection

Choosing the Proper Power for the IOL. Brannon Aden, MD Miles H. Friedlander, MD, FACS

Process optimised FEA- Calculation for Hydroforming Components

INFLUENCE OF PILES ON LOAD- SETTLEMENT BEHAVIOUR OF RAFT FOUNDATION

Trouble Shooting Guide for Ortho-K lenses

GETTING STARTED WITH DMEK

FINITE ELEMENT SIMULATIONS OF THE EFFECT OF FRICTION COEFFICIENT IN FRETTING WEAR

Special Publication: Ophthalmochirurgie Supplement 2/2009 (Original printed issue available in the German language)

Maximum Light Transmission. Pupil-independent Light Distribution. 3.75D Near Addition Improved Intermediate Vision

OPHTHALMIC SURGICAL MODELS

THIN-WALLED HOLLOW BOLTS

SEE BEYOND WITH FULLRANGE OPTICS. Developed by Hanita Lenses

ifs 150 khz IntraLase Technology

Numerical Study of Cornea Applanation by Using a Portable. Force-Displacement Sensor for Intraocular Pressure Measurements

Borehole vibration response to hydraulic fracture pressure

THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening

LASIK & Refractive Surgery

Advanced Machining Processes Professor Vijay K. Jain Department of Mechanical Engineering Indian Institute of Technology, Kanpur Lecture 06

ROBUST PARAMETER DESIGN AND FINITE ELEMENT ANALYSIS FOR A NON- PNEUMATIC TIRE WITH LOW VIBRATION

Stress Analysis of Flanged Joint Using Finite Element Method

(12) United States Patent (10) Patent No.: US 6,334,862 B1

Is LASIK the best choice for me? LASIK AND LASIK-LIKE PROCEDURES

Corporate Subscription. NAFEMS reference library at the click of a button

PreciSAL Preloaded Lens System

Module 2 WAVE PROPAGATION (Lectures 7 to 9)

Modified Peripheral Iris Fixation of a Foldable Intraocular Lens Following Phacoemulsification

The procedure was nearly complete. I was

Prying of a Large Span Base Plate Undergoing a Moment Load Applied by a Round Pier

Choices and Vision. Jeffrey Koziol M.D. Friday, December 7, 12

(12) Patent Application Publication (10) Pub. No.: US 2004/ A1

An Interesting Use of Bausch and Lomb s KeraSoft IC Lens

Basic Principles of the Surgical Microscope. by Charles L. Crain

Choices and Vision. Jeffrey Koziol M.D. Thursday, December 6, 12

12 10) Patent No.: US 8.663,235 B2. Tassignon (45) Date of Patent: Mar. 4, 2014 (54) BAG-IN-THE-LENS INTRAOCULAR LENS FOREIGN PATENT DOCUMENTS

Files required to restart an analysis (Abaqus/Explicit)

HIGH DEFINITION VISION SOLUTIONS THE COMPLETE GUIDE TO LASER VISION CO R R EC T I O N

Sutureless Trocar-Cannula Based Transconjunctival Flanged Intrascleral Intraocular Lens Fixation

PRODUCT CATALOGUE Sterile Single Use Instruments for Cataract and Vitreoretinal Surgery

Treatment of Presbyopia during Crystalline Lens Surgery A Review

Module 3 Selection of Manufacturing Processes

Contents. Corneal Markers, Scleral Depressors

THE XTRAFOCUS IS AN ELEGANT SOLUTION TO COMPLEX CASES.

Table of contents. 5 Swiss precision for clear vision. 6 What types of defective eyesight are there? 8 What is refractive eye surgery?

November 14, 2017 Vision: photoreceptor cells in eye 3 grps of accessory organs 1-eyebrows, eyelids, & eyelashes 2- lacrimal apparatus:

Indian Journal of Engineering An International Journal ISSN EISSN Discovery Publication. All Rights Reserved

Financial Disclosure. Acufocus. Presbyopia Surgery. Inlay Concept 8/14/17. Presbyopia Correction: The Holy Grail of Ophthalmology

Transcription:

Visit the SIMULIA Resource Center for more customer examples. FEA of Prosthetic Lens Insertion During Cataract Surgery R. Stupplebeen, C. Liu, X. Qin Bausch + Lomb, SIMULIA, SIMULIA Abstract: Cataract surgery is the most common surgery in America today. Modern surgeries require the opacified crystalline lens to be removed and for a prosthetic lens to be inserted through a suture-less incision during a 5-10 minute outpatient procedure. The industry is driving for smaller incisions by redesigning the lens and insertion device geometry in addition to new materials. Typical lens dimensions are 6mm diameter with a center thickness of 1mm which is inserted through a 2.8mm incision. For the insertion the lens is folded and elongates while advancing down a tapering tube. Abaqus Explicit was chosen for this analysis because of its capability to solve large deformations and difficult self contact. During the insertion the lens can experience strains in excess of 60%. The purpose of this model is to increase our understanding of the mechanical response of existing products to aide in the design of our future products. Keywords: Biomechanics, Biomedical, Contact, Self Contact, Abaqus/Explicit, Hyperelasticity, Implantable Medical Device, Large Deformation, Optics, Visualization, quasi-static. 1. Introduction Cataracts develop as part of the natural aging process. The crystalline lens which is located behind the iris and is responsible for the focusing of light becomes hardened and opaque. If left untreated blindness would inevitably result. Insertion of a prosthetic lens following cataract removal began with a rigid PMMA intraocular lens (IOL) which required an incision roughly half way around the cornea. The medical community has made significant strides in the improvement of this procedure by improving outcomes, reducing recovery time and simplifying surgical techniques. Current surgeries employ a highly deformable silicone or acrylic lens utilizing an inserter instead of forceps. The lens is loaded into the inserter and pushed with a plunger down a tapering tube. Internal areas of the inserter are filled with viscoelastic (viscous aqueous substance) to lubricate the surface and to prevent the introduction of air bubbles into the eye. The end of the tube is placed through an incision in the eye and the lens is allowed to bloom inside of the eye behind the iris. The driving factors in reducing the incision size to 3mm and below removes the need for sutures to seal the wound and reduces surgically induced astigmatism. Now the driving factor is to further improve visual outcomes. The size of the incision is directly related to optical aberrations that will remain after healing. Incisions less than 1mm have been proposed which is currently the limit of the additional surgical instruments used in the procedure. 2010 SIMULIA Customer Conference 1 Visit the SIMULIA Resource Center for more customer examples.

The design of the inserter and the lens directly affects the incision size during surgery. A smaller lens allows for a smaller incision however a smaller optic can create glare and halos for the patient if the pupil dilates larger than the optic diameter. Because of this and many similar tradeoffs lens and inserter design ideally are created in tandem. A convenient metric for assessing the performance of the inserter lens system is to divide the cross sectional area of the lens by the inserter's area. The currently available lens inserter system which is presented in this paper has a 2.2 compression ratio. To achieve a 1mm incision a 3.3 compression ratio is needed. 2. Model SolidWorks was used to create all of the solid CAD geometry. This data was imported utilizing the SolidWorks Associative Translator. Abaqus/CAE was used for finite element model creation. Abaqus/Explicit was selected as the solver to employ it's robust contact algorithms and tolerance of extremely large deformations. Geometry The IOL geometry consists of 2 discrete parts, the optic body and the haptics as shown in Figure 1. The haptics are used to stabilize the lens in the eye. They press onto the equator of the capsular bag which held the now removed cataract. Figure 1. Lens geometry. The critical features of the inserter are the plunger, loading area and tip as shown in Figure 2. The IOL is placed in the loading area. The tip is snapped over the loading area. The inserter tip is inserted through a corneal incision. Then the IOL is then advanced by the plunger down the tip. 2 2010 SIMULIA Customer Conference

Materials Figure 2. Inserter geometry. The lens material has been tested by compressing buttons with a load frame. The material has been fitted with a hyperelastic Neo Hooke material model. Rayleigh damping which utilizes Equation 1 has been applied to the material model to reduce low frequency oscillations. For this application only (mass proportional) damping was utilized. β (stiffness proportional) damping was not used because it reduces the stable time increment and in turn increases the solution time. Since β damping was not used, the second half of the equation can be negated. The first natural frequency of the lens was predicted with a separate modal analysis. The damping value selected was 5% which is very conservative for this application because of the visco-elastic fluid filling the loading area and tip. Linear elastic vendor data is being utilized for all inserter parts. Equation 1. Rayleigh damping. α R β Rωi ξ i = + 2ω 2 i Boundary Conditions In use the inserter remains relatively stationary while the surgeon pushes on the plunger. The insertion velocity was chosen to deliver in 2 seconds, which is the fastest that delivery would be performed in practice. The choice of the fastest delivery allows for a reduced step time and in turn less computational time. General contact has been applied to the model with zero friction because of the smooth surfaces and the lubricant used. A nonlinear pressure-overclosure relationship was used to reduce contact penetration. Step Properties Due to the quasi-static nature of this analysis mass scaling was utilized to greatly reduce the solve time. This is accomplished by artificially increasing the mass of all elements with stable time 2010 SIMULIA Customer Conference 3

increments below the specified value. The kinetic energy of the lens was less than 4% of its internal energy throughout the step. This value is deemed acceptable and is a desirable trade off for the 10x reduction in solution time. 3. Results Several images of the deformed geometry can be viewed in Figure 3 - Figure 6. Figure 3. Strain of sectioned geometry (0-60% ). Figure 4. Strain of lens and plunger (0-60%). 4 2010 SIMULIA Customer Conference

Figure 5. Strain of final deformed geometry (0-60%). Figure 6. Strain of final deformed geometry (0-60%). A critical design feature of an inserter system is the force required by the surgeon to deliver the IOL. Test data is plotted with the FE predicted data in Figure 7. The test results are the thin lines. The unfiltered data is the diamonds. The thick line is the filtered data. The FE data needs to be filtered with a moving average to remove some noise inherent in an Explicit analysis. It can be seen that the FE results predict within roughly 10% of the test data. This disparity is likely due to the absence of friction and visco-elastic in the model. At the end of the data set there is a significant divergence which attributed to a spring that retracts the plunger which was not modeled. 2010 SIMULIA Customer Conference 5

Figure 7. Test and FE comparison of Plunger Force vs. Displacement. There are several extremely rare, failure modes of the system: tip fracture, lens tear and lens scratches. Tip fracture occurs in the most compressed region of the inserter. The fracture usually propagates from just inbound of the inserter end. The maximum strain in the tip across all time steps is shown in Figure 8. The deformed geometry with the lens fully compressed is shown in Figure 9. This location agrees well with historical failure modes. Figure 8. Strain of maximum compression (0-0.3%). Figure 9. Strain at maximum compression of lens (0-60% ). Lens tear typically occurs where the plunger contacts the IOL or on the trailing haptics. Figure 10 shows the maximum stress in the lens across all time steps. There is a significant stress concentration and localized deformation where the plunger is pushing the IOL. 6 2010 SIMULIA Customer Conference

Figure 10. Maximum strain in IOL across all time points, anterior and posterior respectively (0-100%). The optic body can become scratched during delivery. This is expected to occur where contact pressures are highest as shown in Figure 11. Large strain values also could increase the susceptibility of scratching as shown previously in Figure 9. These locations agree well with historical locations where scratches occur. Figure 11. Contact pressure on the IOL in fully compressed state with contours (0-1.5MPa). 4. Conclusion Abaqus/Explicit is able to model this highly nonlinear event with sliding self contact and large deformations. The model has accurately predicted insertion force and likely failure locations where peak stresses are developing. This validated model is being used to help develop the next generation of IOLs and inserters which will experience much more demanding loading conditions. 2010 SIMULIA Customer Conference 7 Visit the SIMULIA Resource Center for more customer examples.