Optimized CT metal artifact reduction using the Metal Deletion Technique (MDT)

Similar documents
Yinsheng Li 1, Peter Bannas 2, M.D., Perry Pickhardt M.D. 2, Meghan Lubner M.D. 2, Ke Li Ph.D. 1,2, and Guang-Hong Chen Ph.D. 1,2

Metal Artifact Reduction for Orthopedic Implants (O-MAR)

DICOM Conformance. DICOM Detailed Specification for Diagnostic Labs and Radiology Center Connectivity

Maximum Performance, Minimum Space

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

Delivering Better Patient Care with SIGNA Architect

Model Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI

1571. Artifacts in computer tomography imaging: how it can really affect diagnostic image quality and confuse clinical diagnosis?

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

MRI Metal Artifact Reduction

Brilliance in everything Philips CT products and services

Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques

Iterative Reconstruction

METAL artifact reduction (MAR) is a major problem

COMPUTED TOMOGRAPHY 1

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

Aim. Images for this section: Page 2 of 13

CT Data Storage Reduction by Means of Compressing Projection Data Instead of Images: Feasibility Study 1

160-slice CT SCANNER / New Standard for the Future

Advanced Noise Reduction Processing for X-ray CT System with Iterative Processing. Koichi Hirokawa MEDIX VOL. 56 P.43 P.46

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging

CT Basics: Data Acquisition Module 3

PACS Fundamentals. By: Eng. Valentino T. Mvanga Ministry of Health and Social Welfare Tanzania

Detector technology in simultaneous spectral imaging

GE Healthcare. Senographe 2000D Full-field digital mammography system

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11

Q3D. Speak to a 3D Specialist. CBCT 3D / Panoramic Imaging GENERAL DIMENSIONS. Suni Imaging Product Lines GET.

Software and Hardware in CCTA. Elly Castellano PhD

BRINGING DEEP LEARNING TO ENTERPRISE IMAGING CLINICAL PRACTICE

Introduction, Review of Signals & Systems, Image Quality Metrics

QC Testing for Computed Tomography (CT) Scanner

Computerized Medical Imaging and Graphics

nothing is more important than health

SIMPLE. VERSATILE. VERSATILE PRINTING ON FILM OR PAPER. DRYVIEW CHROMA Imaging System

2D, 3D CT Intervention, and CT Fluoroscopy

RADspeed Pro. EDGEpackage C501-E041C

Maximizing clinical outcomes

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Translating Protocols Between Scanner Manufacturer and Model

diagnostic examination

NeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results

The Versatile and Powerful ACLxy. ACLxy

MUSICA Nerve Center. Artificial Intelligence. Intelligent tools for your Digital Radiography workflow. Fluoroscopy. Workflow Optimization

make it easy, with Ray

Digital Imaging CT & MR

Scopis Hybrid Navigation with Augmented Reality

An Activity in Computed Tomography

Data. microcat +SPECT

TORNIER BLUEPRINT. 3D Planning + PSI SCAN PROTOCOL

Aquilion Precision Ultra-High Resolution CT: Quantifying diagnostic image quality

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

The future of nuclear imaging is clear

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

CR Basics and FAQ. Overview. Historical Perspective

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose

Alternative lossless compression algorithms in X-ray cardiac images

Advanced digital image processing for clinical excellence in fluoroscopy

An Activity in Computed Tomography

Barco medical display systems. Product catalog

PET: New Technologies & Applications, Including Oncology

Project Abstract Submission : Entry # 456. Part 1 - Team. Part 2 - Project. Team Leader Name. Maroua Filali. Team Leader .

Architecture of Quality Imaging Mary K. Henne, MS, CNMT, RDMS, RVT Ultrasound Education Specialist GE Healthcare

PD233: Design of Biomedical Devices and Systems

Digital Image Management: the Basics

Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016

Iterative Reconstruction in Image Space. Answers for life.

radiography detector

imagespectrum ADVANCED DIGITAL IMAGE MANAGEMENT SYSTEM Get a Better Handle on the Big Picture

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

SAFIRE. Sinogram Affirmed Iterative Reconstruction. Answers for life.

THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY

QC by the MPE in Belgium

CT Basics: Image Quality Module 6

Wide-Detector CT for TAVR Planning:

Going Live with the Aquilion VISION Volume 4D CT

Outline ASRT Changes Impact on current curriculum Potential new courses WECM Changes Last update Resources and needs

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

A Module for Visualisation and Analysis of Digital Images in DICOM File Format

GE Healthcare. Discovery MR T. Simply powerful. Powerfully simple.

SPECT Reconstruction & Filtering

2 nd generation TOMOSYNTHESIS

UGEO H60. Performance in Style. Features

KODAK Dental Imaging Software. Quick Start Guide

Electronic Noise in CT Detectors: Impact on Image Noise and Artifacts

Performance and care. all in one

SIGNA Explorer Lift revives our MR

Suppression of metal artifacts using image-based monoenergetic DECT imaging

Four-dimensional Computed Tomography (4D CT) Concepts and Preliminary Development

Focused RF Hyperthermia Using Ultra-High Field MRI Joshua de Bever, PhD

Local ROI Reconstruction via Generalized FBP and BPF Algorithms along More Flexible Curves

COMPUTED RADIOGRAPHY (CR)

A Basic Overview of Veterinary Computed Tomography

Improvement of CT image quality with iterative reconstruction idose4

MRI Phase Mismapping Image Artifact Correction

Introduction. MIA1 5/14/03 4:37 PM Page 1

Transcription:

Optimized CT metal artifact reduction using the Metal Deletion Technique (MDT) F Edward Boas, Roland Bammer, and Dominik Fleischmann Extended abstract for RSNA 2012 Purpose CT metal streak artifacts are extremely common: 21% of scans in one series (1). They are caused by multiple mechanisms, including beam hardening, scatter, Poisson noise, motion, and edge effects. We developed an iterative method called the Metal Deletion Technique (MDT) (1, 2), which is based on the principle that projection data involving or near metal is less accurate. MDT starts with either raw projection data from the scanner, or simulated projection data from a DICOM image. It then only uses high quality non metal data to reconstruct the non metal portions of the image. Metal pixels are erased from the reconstructed image, and on each iteration, the inaccurate metal data are replaced with forward projected values from the previous iteration. This means that, instead of trying to look through the metal to see soft tissue, we look around the metal. An initial evaluation of MDT showed that it had the best image quality when compared against FBP and two metal artifact reduction methods (1). In 2 of 11 scans, the improved image quality revealed important new findings. This included a case of rectal cancer (in a patient with bilateral hip replacements) that was originally missed when reviewing only the images produced by the scanner. A follow up study of 80 patients showed that MDT improved image quality 73% of the time for small metal implants, and 75% of the time for large metal implants (3). In that study, MDT had better image quality than all three other metal artifact reduction techniques tested. MDT reduces metal artifacts due to beam hardening, motion, edge effects, and Poisson noise (1, 4). We have found it to be particularly useful for radiation oncology, interventional radiology, orthopedics, and neurosurgery applications (4). The main limitations of MDT are the long processing time, decreased resolution for large or long implants, and new artifacts introduced in certain cases. In this paper, we develop a variation on the MDT algorithm that addresses some of these limitations. Methods The institutional review board approved this HIPAA compliant study; informed patient consent was waived. We optimized the speed and image quality of the Metal Deletion Technique (MDT) for metal artifact reduction, by testing 84 variations of the MDT algorithm on a diverse test set of 22 DICOM images reconstructed by the scanner. Raw projection data was estimated by forward projecting the DICOM images.

The optimized metal artifact reduction algorithm was then set up as a scalable (tested up to 40 CPU cores in parallel) cloud application, which was integrated with our clinical PACS. The DICOM send function is used to send scans to a server that automatically reduces artifacts and sends the processed images back to PACS as a new series under the same accession. This procedure works with images from any scanner, and it does not require any manual drawing of regions of interest, or tuning of parameters. Radiologists and clinicians can submit studies for metal artifact reduction. 90 full scans referred for metal artifact reduction at a single institution were then evaluated for image quality, and whether the metal artifact reduction images affected diagnosis, visualization of key findings, or diagnostic confidence. Results The optimized MDT algorithm includes changes to virtually all steps of the algorithm, resulting in an improvement in both speed and image quality (shown below). Improvements in processing speed were primarily due to optimized automatic determination of the sampling resolution. Improvements in image quality were primarily due to the following changes: simulated raw data is iteratively corrected to optimize fidelity, the level of streak suppression tapers off with distance from metal, and the algorithm detects cases where it performs poorly, and automatically switches to a different algorithm. The optimized MDT algorithm can process a single image in 15 minutes on a single CPU core, compared to 97 minutes previously (Intel Xeon Nehalem X5575 2.93 GHz), a speed up of more than 6 fold. With a 40 CPU core server, 40 slices containing metal artifact can be processed in a total of 15 minutes. (Note that our previously reported processing time of 28 minutes per slice (1) used raw projection data from the scanner. In this paper, we use projection data that is simulated from the DICOM image generated by the scanner. Simulated projection data requires more processing time due to both the time required to forward project the DICOM image, and the higher sampling resolution required to prevent blurring and artifacts introduced by the forward projection step. The advantage of simulated projection data is that it does not require access to proprietary raw data, and thus works with any manufacturer s scanner.) In 18 of 22 DICOM images, the optimized MDT algorithm had better subjective image quality. In particular, image quality was substantially improved compared to the original algorithm for aneurysm coils in the same slice as the mastoids (Figure 1). This is a particularly difficult case for metal artifact reduction algorithms, due to the motion of the coil and the fine bony details in the mastoids.

FBP MDT (original) MDT (optimized) Figure 1. Filtered backprojection (FBP) images generated by the scanner, compared to the original and optimized Metal Deletion Technique (MDT) algorithms. Of the 90 scans referred for metal artifact reduction by radiologists or clinicians, image quality was improved compared to the image generated by the scanner in 86% (77 scans). The most common cause of worsened image quality was resolution loss around implants measuring > 5 cm in the axial plane, such as pedicle screws. In 13 scans (14%), MDT changed the diagnosis, improved visualization of key findings, or improved diagnostic confidence (Table 1 and Figure 2). Result Finding Metal implant Number of cases Changed diagnosis Obstructing ureteral stone Bilateral hip replacements 1 Improved visuaization Hemorrhage or infarct Intracranial aneurysm coil or clip 6 Improved visualization Pelvic mass Bilateral hip replacements 1 Improved visualization Periprosthetic fluid collection Hip replacment 1 Improved visualization Traumatic urethral injury Bilateral hip hardware 1 Improved confidence No parotid duct stone Dental fillings 1 Improved confidence No ureteral stone Hip replacement 1 Improved confidence No PE or mesenteric ischemia Biventricular assist device 1 Table 1. Cases where MDT changed the diagnosis, improved visualization of key findings, or improved diagnostic confidence.

FBP MDT (optimized) Hemorrhage Obstructing stone Figure 2. Filtered backprojection (FBP) images generated by the scanner, compared to Metal Deletion Technique (MDT) post-processed images. First row: FBP image (curved planar reformat) shows left hydroureteronephrosis. MDT image reveals an obstructing ureteral stone, which was originally obscured by streaks from bilateral hip replacements. Second row: FBP image shows motion artifact around a metal aneurysm coil. MDT image reduces the artifact, and more clearly shows the hemorrhage. Conclusion MDT can be helpful in the clinical evaluation of patients with metal implants. The optimized version of MDT is 6x faster, and has further improved image quality. In 14% of cases, it changed the diagnosis, improved visualization of key findings, or improved diagnostic confidence.

Integration with PACS allows any clinician or radiologist to easily perform metal artifact reduction. MDT works with scanners from all major CT manufacturers, and is now used in 55 hospitals around the world. The speed of MDT could be further improved by implementing the algorithm on a graphics processing unit (GPU) or field programmable gate array (FPGA), rather than running on the CPU. MDT completely discards metal data. The advantage of this approach is that motion, edge, and scatter artifacts are reduced, not just beam hardening artifacts. The disadvantage is loss of resolution around large or long implants. Thus, MDT images are not a replacement for FBP, but must be viewed in conjuction with the FBP images generated by the scanner. References 1. Boas FE, Fleischmann D. Evaluation of two iterative techniques for reducing metal artifacts in computed tomography. Radiology 2011; 259:894 902. 2. Boas FE. Iterative reduction of artifacts in computed tomography images using forward projection and an edge preserving blur filter. US patent 8233586, issued 2012. 3. Golden C, Mazin SR, Boas FE, et al. A comparison of four algorithms for metal artifact reduction in CT imaging. Proc. SPIE 2011; 7961:79612Y. 4. Boas FE, Fleischmann D. Computed tomography artifacts: Causes and reduction techniques. Imaging in Medicine 2012; 4:229 240.