Module 5: Experimental Modal Analysis for SHM Lecture 36: Laser doppler vibrometry. The Lecture Contains: Laser Doppler Vibrometry

Similar documents
Tutorial: 3D Scanning Vibrometry for. IMAC XXVII D. E. Oliver, Polytec, Inc.

3D Optical Motion Analysis of Micro Systems. Heinrich Steger, Polytec GmbH, Waldbronn

Modal analysis: a comparison between Finite Element Analysis (FEA) and practical Laser Doppler Vibrometer (LDV) testing.

648. Measurement of trajectories of piezoelectric actuators with laser Doppler vibrometer

SCANNING LASER VIBROMETRY FOR DETECTION NOISE SOURCES WITH HIGH SPATIAL RESOLUTION

Fiber Pigtailed Variable Frequency Shifters Acousto-optic products

CALIBRATION OF LASER VIBROMETER STANDARDS ACCORDING TO ISO

Unit-23 Michelson Interferometer I

Characterization of Silicon-based Ultrasonic Nozzles

VIBROMET 500V: SINGLE POINT LASER DOPPLER VIBROMETER

SPRAY DROPLET SIZE MEASUREMENT

This document is downloaded from DR-NTU, Nanyang Technological University Library, Singapore.

MEASUREMENT OF SURFACE ACOUSTIC WAVE USING AIR COUPLED TRANSDUCER AND LASER DOPPLER VIBROMETER

1241. Efficiency improvement of energy harvester at higher frequencies

Laser Telemetric System (Metrology)

684. Remote sensing of vibration on induction motor and spectral analysis

la. Smith and C.P. Burger Department of Mechanical Engineering Texas A&M University College Station Tx

ULTRASONIC TRANSDUCER PEAK-TO-PEAK OPTICAL MEASUREMENT

Laser Doppler sensing in acoustic detection of buried landmines

Applications of Acoustic-to-Seismic Coupling for Landmine Detection

NEW LASER ULTRASONIC INTERFEROMETER FOR INDUSTRIAL APPLICATIONS B.Pouet and S.Breugnot Bossa Nova Technologies; Venice, CA, USA

Very High Frequency Calibration of Laser Vibrometer up to 350 khz

430. The Research System for Vibration Analysis in Domestic Installation Pipes

Scanning laser Doppler vibrometry

Development of a Low Cost 3x3 Coupler. Mach-Zehnder Interferometric Optical Fibre Vibration. Sensor

HSV-2000 High Speed Vibrometer

CHAPTER 5 FINE-TUNING OF AN ECDL WITH AN INTRACAVITY LIQUID CRYSTAL ELEMENT

Underwater Acoustics Research

VibSoft Data Acquisition Software

Setup of the four-wavelength Doppler lidar system with feedback controlled pulse shaping

VibSoft Data Acquisition Software

LASER DOPPLER VELOCIMETRY

Swept Wavelength Testing:

A STUDY ON THE VIBRATION CHARACTERISTICS OF CFRP COMPOSITE MATERIALS USING TIME- AVERAGE ESPI

Autocorrelator MODEL AA- 10DM

visibility values: 1) V1=0.5 2) V2=0.9 3) V3=0.99 b) In the three cases considered, what are the values of FSR (Free Spectral Range) and

This document is downloaded from DR-NTU, Nanyang Technological University Library, Singapore.

Single Photon Interference Laboratory

INDIAN INSTITUTE OF TECHNOLOGY BOMBAY

7. Michelson Interferometer

Fibre Optic Sensors: basic principles and most common applications

Correction for Synchronization Errors in Dynamic Measurements

Proton Induced Thermal Stress Wave Measurements in. Solid Targets

Clarification of the Effect of High-Speed Train Induced Vibrations on a Railway Steel Box Girder Bridge Using Laser Doppler Vibrometer

LASER VIBROMETER CALIBRATION AT HIGH FREQUENCIES USING CONVENTIONAL CALIBRATION EQUIPMENT

Fibre Laser Doppler Vibrometry System for Target Recognition

VibSoft Data Acquisition Software

In House Developments

Physics 3340 Spring 2005

VibroFlex Connect Configurable core of the modular vibration sensing system Preliminary datasheet. VibroFlex Connect

Experiments with wave, using low-cost amplitude modulated ultrasonic techniques

Imaging Systems Laboratory II. Laboratory 8: The Michelson Interferometer / Diffraction April 30 & May 02, 2002

Microscopic Laser Doppler Vibrometer

Installation and Characterization of the Advanced LIGO 200 Watt PSL

Universal and compact laser stabilization electronics

attosnom I: Topography and Force Images NANOSCOPY APPLICATION NOTE M06 RELATED PRODUCTS G

Absolute distance interferometer in LaserTracer geometry

First Observation of Stimulated Coherent Transition Radiation

Roland Kruse, Tiedo Meyer: Vibration platform for the calibration of optical sensors

NCSL International 2995 Wilderness Place, Suite 107 Boulder, Colorado Office: (303) Fax: (303)

Dynamic beam shaping with programmable diffractive optics

ECE 185 ELECTRO-OPTIC MODULATION OF LIGHT

Periodic Error Correction in Heterodyne Interferometry

The 34th International Physics Olympiad

Development of C-Mod FIR Polarimeter*

Bending vibration measurement on rotors by laser vibrometry

Vibration measurements by tracking laser doppler vibrometer on automotive components

VDD PC-Based Digital Vibrometer

Fumiaki UEHAN, Dr.. Eng. Senior Researcher, Structural Mechanics Laboratory, Railway Dynamics Div.

Jones matrix analysis of high-precision displacement measuring interferometers

Modal testing circuit board assembly of an electronic apparatus by laser vibrometry

Variable Configuration Fiber Optic Laser Doppler Vibrometer System

Advances in Acoustic Landmine Detection

Acoustic-Laser Vibrometry for Standoff Detection of Defects in Materials

Use of Computer Generated Holograms for Testing Aspheric Optics

(51) Int Cl.: G01B 9/02 ( ) G01B 11/24 ( ) G01N 21/47 ( )

REAL TIME THICKNESS MEASUREMENT OF A MOVING WIRE

Studies with a small gamelan gong

PHYS 3153 Methods of Experimental Physics II O2. Applications of Interferometry

Chapter 7. Optical Measurement and Interferometry

Enhancing the capability of primary calibration system for shock acceleration in NML

Department of Electrical Engineering and Computer Science

Advanced Optical Communications Prof. R. K. Shevgaonkar Department of Electrical Engineering Indian Institute of Technology, Bombay

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science

R. J. Jones College of Optical Sciences OPTI 511L Fall 2017

Components of Optical Instruments

EXPERIMENTAL AND NUMERICAL ANALYSIS OF THE MUSICAL BEHAVIOR OF TRIANGLE INSTRUMENTS

Microwave Optics. Department of Physics & Astronomy Texas Christian University, Fort Worth, TX. January 16, 2014

Designing Optical Layouts for AEI s 10 meter Prototype. Stephanie Wiele August 5, 2008

LECTURE 26: Interference

Dynamic Phase-Shifting Microscopy Tracks Living Cells

Experiments with wave, using low-cost amplitude modulated ultrasonic techniques

Wave Front Detection for Virgo

INTEGRATED ACOUSTO-OPTICAL HETERODYNE INTERFEROMETER FOR DISPLACEMENT AND VIBRATION MEASUREMENT

The Virgo detector. L. Rolland LAPP-Annecy GraSPA summer school L. Rolland GraSPA2013 Annecy le Vieux

A detailed experimental modal analysis of a clamped circular plate

Holography (A13) Christopher Bronner, Frank Essenberger Freie Universität Berlin Tutor: Dr. Fidder. July 1, 2007 Experiment on July 2, 2007

AIR FORCE INSTITUTE OF TECHNOLOGY

High-Speed 3D Sensor with Micrometer Resolution Ready for the Production Floor

3.003 Lab 3 Part A. Measurement of Speed of Light

Transcription:

The Lecture Contains: Laser Doppler Vibrometry Basics of Laser Doppler Vibrometry Components of the LDV system Working with the LDV system file:///d /neha%20backup%20courses%2019-09-2011/structural_health/lecture36/36_1.html [4/4/2013 4:02:17 PM]

Laser Doppler Vibrometry Laser Doppler Vibrometer (LDV) is a Laser based non-contact vibration measurement system. It consists of three measuring scan heads which are capable of measuring the movements in all the three orthogonal directions yielding full information of the three dimensional movements [Figure 36.1]. The system works on the principle of Doppler Effect and interferometry for vibration measurement. Figure 36.1 Laser Doppler Vibrometer with composite test plate The minimum detectable vibration speed using this system is 5 µm/s at1hz resolution while the maximum speed is of 10 m/s. The LDV system software controls the entire measurement process with graphical user interface. The PSV system also has the provision for input channels which can be used for simultaneous acquisition of data from accelerometers, load cells etc. Transfer function between any of the input channels connected to the system can be obtained. Signal generator card (NI-671x) contained in the system is used for generating excitation signals in the frequency range of 0-80 khz. LDVs can measure vibrations up to 30 MHz range with very linear phase response and high accuracy. Applications of LDV include modal analysis of automotive parts, car bodies and aircraft panels etc. file:///d /neha%20backup%20courses%2019-09-2011/structural_health/lecture36/36_2.html [4/4/2013 4:02:17 PM]

Basics of Laser Doppler Vibrometry Doppler Effect Heterodyne Interferometry Doppler Effect Doppler Effect is the change in the frequency (or wavelength) of emitted waves as the source of the wave approaches or moves away from an observer. This effect was named after the Austrian physicist Christian Johann Doppler who first stated this physical principle in 1842. The change or shift in frequency observed depends on the speed and direction of travel of both source and observer. Helium-Neon (He-Ne) Laser beam is made to incident on the vibrating surface and the reflected Laser light from the surface is detected by the vibrometer scanning unit. Incident and reflected beams are made to interfere on the detector by suitable arrangement. A moving surface induces a frequency shift on the light received by Vibrometer optics. (36.1) where f D is the frequency shift in the reflected beam, V is the velocity of the surface and λ is the wavelength of the He-Ne Laser. file:///d /neha%20backup%20courses%2019-09-2011/structural_health/lecture36/36_3.html [4/4/2013 4:02:18 PM]

Heterodyne Interferometry The Laser Doppler Vibrometer works on the basis of optical interference requiring two coherent light beams. The interference term relates to the path difference between both the beams. If the path difference between the interfering beams is integral multiplier of Laser wavelength, constructive interference occurs. (36.2) I total is the resultant intensity, I 1 and I 2 are the intensities of two interfering Laser beams and ( r 1 - r 2 ) is the path difference. In this case, overall intensity becomes four times the single intensity. If the path difference is odd multiplier of half the wave length, destructive interference occurs where the overall intensity becomes zero. The interference phenomenon is exploited technically in Laser Doppler Vibrometer as shown in the Figure 36.2. Figure 36.2 Schematic system setup for measuring vibration using LDV A He-Ne Laser beam is split by a beam splitter BS1 into a reference beam and a measurement beam. After passing the beam splitter BS2, the measurement beam is focused onto the object to be measured. The object to be investigated must be reflective. Surface of the object may be made reflective by applying Ardox spray coating or retro reflective tape. The reflected beam is deflected by BS2 and is merged with the reference beam by the third beam splitter BS3 and is then directed on to the detector. As the path length of the reference beam is constant over time, a movement of object under consideration generates a dark and file:///d /neha%20backup%20courses%2019-09-2011/structural_health/lecture36/36_4.html (1 of 2) [4/4/2013 4:02:18 PM]

bright fringe pattern on the detector. One complete dark bright cycle corresponds to an object displacement of exactly half the wavelength of the light used. For a He-Ne Laser, this displacement is 316 nanometers. Change in the optical path length per unit time causes the Doppler frequency shift of the measured beam. The modulation frequency of the interferometer pattern is exactly proportional to the velocity of the object. Same interference patterns (and frequency shifts) are generated as the object moves towards or moves away from the interferometer. A Bragg cell is placed in the reference beam to distinguish the direction of movement as it shifts the Laser frequency by 40 MHz. A modulation frequency of the fringe pattern of 40 MHz is generated when the object is at rest. Movement of the object towards the interferometer reduces the modulation frequency while it increases when the object moves away from the Vibrometer. The detector receives a frequency lower or higher than 40 MHz indicating the direction and amplitude of movement of the object. file:///d /neha%20backup%20courses%2019-09-2011/structural_health/lecture36/36_4.html (2 of 2) [4/4/2013 4:02:18 PM]

Components of the LDV system The main components of the Polytec Scanning Laser Doppler Vibrometer are the scan heads (top, left and right), Vibrometer controllers (OFV-5000) for each scan head, junction box connecting all these controllers and a personal computer (PC) containing the PSV software. PSV-I-400 Scan head Each scanning head has four major inbuilt units. They are video, scan electronics, scanning mirrors and photo sensor (OFV-505) units as shown in the Figure 36.3. Figure 36.3 Scan head and its inbuilt units. Salient features of the scanning head are Scan angle ± 20 o with 0.02 o resolution 72 x Zoom video camera Large working distances 0.5 m to 50 m Close up unit for scanning small objects in millimeter range Laser focus, alignment procedures can be remote controlled using PDA which connects directly to the PSV software Vibrometer Controller (OFV-5000) The Vibrometer controller provides signals and power for the sensor head which is present in the scan head and processes the vibration signals. They are electronically converted by specially developed analog and digital decoders within the controller to obtain velocity and displacement information about the test structure. Information provided by controller is available in analog or digital form for further evaluation of data. The analog output is provided at standard BNC connectors. file:///d /neha%20backup%20courses%2019-09-2011/structural_health/lecture36/36_5.html (1 of 2) [4/4/2013 4:02:18 PM]

Junction Box The junction box acts as an interface between the Vibrometer controller and the PSV software. Vibrometer controller of each scan head is connected to the junction box. Input for up to 8 analog signals and trigger are available on BNC connectors. file:///d /neha%20backup%20courses%2019-09-2011/structural_health/lecture36/36_5.html (2 of 2) [4/4/2013 4:02:18 PM]

Working with the LDV system The sequence of steps followed while workings with the PSV system are shown in Figure 36.4. Details of the test setup Figure 36.4 Workflow during the experiment on LDV Composite plates are coated with white spray for the purpose of better reflection. Plates are experimented in cantilever position as shown in the Figure 36.5 (a). file:///d /neha%20backup%20courses%2019-09-2011/structural_health/lecture36/36_6.html (1 of 2) [4/4/2013 4:02:19 PM]

Figure 36.5 (a) Composite test plate with dynamic shaker For the dynamic excitation of the composite plate an electro-dynamic shaker is used. A power amplifier (make: LDS, PA500L series) is connected to the shaker for the purpose of amplifying the excitation signal generated by the LDV system. During the experiment, a pseudo random signal in the frequency range of 0-2000 Hz is used for the excitation of the composite plate. Experiment is carried out in two steps. In the first step the plate is excited from 0-800 Hz with a signal voltage of 0.2 V. In the second step the plate is excited from 800-2000 Hz with increased amplifier gain in order to excite the high frequency modes. file:///d /neha%20backup%20courses%2019-09-2011/structural_health/lecture36/36_6.html (2 of 2) [4/4/2013 4:02:19 PM]

Alignment procedure The 2-D alignment step is carried out to relate the video and scanner coordinate system. This is done by selecting 10-15 points for each Laser head on the scan area. PSV software stores the video coordinates and the scan angles of these points on the scan area and calculates the polynomial interpolation. 2-D alignment step performed for one of the scan heads is shown in Figure 36.5(b). Figure 36.5 (b) 2-D alignment step performed for the composite plate Alignment procedure for 3D In 3D alignment, coordinate system for the scan area is defined. This is done by selecting origin, point on X-axis, and point on X- Y plane on the scan area. After this step, all the three Lasers are made to coincide at other four to five points. The purpose of using these points is to back calculate the position and angle between the Laser heads by PSV software. After 3-D alignment procedure, the software is able to provide the coordinates (X, Y, and Z) of any point on the scan area with respect to the chosen coordinate system. Geometry scan All the scan points in the mesh grid are accessed by the three Lasers in order to estimate the surface of the test object. The three Lasers coincide at each and every scan point on the mesh grid [Figure 36.5(c)]. The accuracy of the surface estimated depends on the previous steps. This step is crucial to obtain the exact simulations of the test surface during the test. file:///d /neha%20backup%20courses%2019-09-2011/structural_health/lecture36/36_7.html (1 of 3) [4/4/2013 4:02:19 PM]

Figure 36.5 (c) Status points on the scanned composite plate Rectangular mesh grid of size 250 mm x 200 mm was done in ABAQUS. This grid was imported into the PSV software to define the scan area as shown in Figure 6.5(d). The grid has 356 scan points in the present case. Figure 36.5 (d) Mesh grid with the scan points on the rectangular composite plate. Data acquisition parameters Data acquisition properties such as the excitation signal to be used, frequency range, parameters for the FFT analysis (number of FFT lines, bandwidth etc.), averages, windows, velocity decoder are given in this step. Time required for the complete scan depends on the number of scan points defined and FFT parameters. In the present analysis, 356 scan points were defined on the composite plate and it took about one hour to complete the scan. Response plots, mode shapes animations are visualized after the scan. file:///d /neha%20backup%20courses%2019-09-2011/structural_health/lecture36/36_7.html (2 of 3) [4/4/2013 4:02:19 PM]

file:///d /neha%20backup%20courses%2019-09-2011/structural_health/lecture36/36_7.html (3 of 3) [4/4/2013 4:02:19 PM]