White Paper. The Advantages of Hybrid Gamma PXL. 1 Introduction Outline of Hybrid Gamma PXL Study Results...

Similar documents
Technical Paper CONSISTENT PRESENTATION OF MEDICAL IMAGES

Image Display and Perception

SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES.

UP-D74XRD. DICOM Digital Imager.

A Novel 12-bit Grayscale Topology PACSmate Revolutionary GrayBoost Technology

Things you may want to know

Computer applications, PACS, Instrumentation, Equipment, Technology assessment /ecr2012/C-0655

A comparison between medical-grade liquid crystal display (LCD) and ipad color imaging

Digital radiography (DR) post processing techniques for pediatric radiology

Practical guidelines for color calibration and quality assurance of medical displays

MEDICAL DISPLAYS. NEC 90 Series. NEC MD Series

Barco medical display systems. Product catalog

For Your Diagnostic Confidence. Product Brochure

Increasing the Number of Gray Shades in Medical Display Systems How Much is Enough?

Clinical and management aspects of digital imaging and PACS

IHE Radiology Technical Framework Supplement. Stereotactic Mammography Image (SMI) Trial Implementation

Explanation, Benefits, and Comparison to Traditional Proofing Systems

160-slice CT SCANNER / New Standard for the Future

Digital Image Management: the Basics

Outline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control

ABSTRACT 1. PURPOSE 2. METHODS

CXDI-70C WIRELESS SPECIFICATIONS

Yagi Digital Microscope Calibration

8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey

TDI2131 Digital Image Processing

Effects of Pixel Density On Softcopy Image Interpretability

PHOTOTUTOR.com.au Share the Knowledge

Infrared Screening. with TotalVision anatomy software

Black & White and colouring with GIMP

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Solutions Page Content ImagePilot. Primary keywords: Digital radiography Computed radiography Image viewing and storage

Small Animal Radiographic Techniques and Positioning COPYRIGHTED MATERIAL

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities

Display Profiling. How to Calibrate a. Introduction. 4 C s of Color Management. Step 1: Consistency. Step 2: Calibration

Chin J Radiol 2003; 28:

153 Shimokashiwano, Hakusan, Ishikawa Japan Phone Fax:

CONSOLE ADVANCE The He art of Your FDR/ FCR System

CR Retrofit Kit Package Configuration Integrated PACS

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.

k m = 683 lumens/watt Display QC QC Measures for Color Displays joules watts lumens SAM Imaging Education Course

Image evaluation method for digital mammography with soft-copy reading using a digital phantom

DICOM Correction Proposal Form

CONSOLE ADVANCE The He art of Your FDR/ FCR System

DR Retrofit Kit Package Configuration DICOM

PACS Fundamentals. By: Eng. Valentino T. Mvanga Ministry of Health and Social Welfare Tanzania

CR35-SP+ Computed Radiography (CR) System. Innovations in Digital Imaging

H34: Putting Numbers to Colour: srgb

Digital Film Imager UP-DF550

Explain what is meant by a photon and state one of its main properties [2]

A Study On Preprocessing A Mammogram Image Using Adaptive Median Filter

A Module for Visualisation and Analysis of Digital Images in DICOM File Format

GE Healthcare. Senographe 2000D Full-field digital mammography system

Color Matching Beginning with the Basics. EIZO's Guide to enjoyment. a further. of digital. I see. photography

Vascular. Development of Trinias Series unity edition Angiography Systems. 1. Introduction. 3. Three Concepts of "unity"

SIMPLE. VERSATILE. VERSATILE PRINTING ON FILM OR PAPER. DRYVIEW CHROMA Imaging System

Recovering highlight detail in over exposed NEF images

4.0 How to Turn On the Selenia Dimensions

COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS

Imaging Informatics. Standards Support for Color in Medical Imaging. SAM Imaging Education Course. Photometric Standards Colorimetric Standards

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?

Digital Image Processing

Guidance on Using Scanning Software: Part 6. Canon ScanGear

ISO/IEC TS TECHNICAL SPECIFICATION

Medical Images Analysis and Processing

Brief Introduction to Vision and Images

Instant DR in Jordan

This document is a preview generated by EVS

DR Retrofit Kit Package Configuration Integrated PACS

Ideal for display mura (nonuniformity) evaluation and inspection on smartphones and tablet PCs.

Digital Imaging and Communications in Medicine (DICOM) Supplement 39: Add Stored Print Media Storage - Retire Normalized Print Media Storage

you can Portable DR with High Sensitivity. Less is more. CXDI-501C DIGITAL RADIOGRAPHY SYSTEM

Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques

Better Light ViewFinder Repro Curves

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

FFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May Carl R. Keener, Ph.D., DABMP, DABR

ISSN (PRINT): ,(ONLINE): ,VOLUME-4,ISSUE-3,

Open. the Digitized world. Fuji Computed Radiography

Camera Overview. Digital Microscope Cameras for Material Science: Clear Images, Precise Analysis. Digital Cameras for Microscopy

Working with Wide Color Gamut and High Dynamic Range in Final Cut Pro X. New Workflows for Editing

HP DREAMCOLOR USE CASES

DELWORKS DR MEDICAL. take the next step

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

PSSA Calibration and Colour Management

ISO/IEC JTC 1/SC 29 N 16019

Quick Start Guide to Printing on the EPSON 9800

Conformance Statement for DICOM Viewer

Black and White Photoshop Conversion Techniques

A SURVEY ON DICOM IMAGE COMPRESSION AND DECOMPRESSION TECHNIQUES

Performance and care. all in one

Multi-Access Biplane Lab

you can Portable Digital Radiography. Going places. CXDI-501G DIGITAL RADIOGRAPHY SYSTEM

Details that Make a Difference

Scalable Data Storage Analysis and Solution for Picture Archiving and Communication Systems (PACS)

COLORMUNKI DISPLAY & i1display PRO

DICOM Conformance. DICOM Detailed Specification for Diagnostic Labs and Radiology Center Connectivity

Camera Overview. Digital Microscope Cameras for Material Science: Clear Images, Precise Analysis. Digital Cameras for Microscopy

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

FCR XL-2 and FCR XC-2. High-quality digital x-ray that's perfect for your private practice.

Maximum Performance, Minimum Space

PHOTOTUTOR.com.au Share the Knowledge

Transcription:

White Paper The Advantages of Hybrid Gamma PXL CONTENTS 1 Introduction... 2 2 Outline of Hybrid Gamma PXL... 4 3 Study... 5 4 Results... 7 5 Conclusions... 10 No.17-001 Revision A November 2017 Product & Business Development Dept. Product Technology section White Paper (Q17B014-AS-10002) 1/11

1 Introduction With advances in medical imaging technology over the years, hospitals are now handling a wider variety and larger volume of image data. As a result, it is increasingly common for medical facilities to use a variety of color and monochrome imaging modalities to interpret images. Advanced display technology such as high brightness, high contrast color LCD devices also accelerates this trend of reading various color and monochrome imaging modalities on one screen. However, there is a fundamental technical challenge when it comes to displaying both monochrome and color images on one screen, while ensuring image display conditions to be appropriate for each modality. In other words, monochrome images must be displayed with standardized gamma characteristics, which are quite different from these of color images. DICOM (Digital Imaging and Communications in Medicine) part 14 defines specific gamma characteristics for monochrome images called GSDF (Grayscale Standard Display Function). It is commonly used for Monitor QC assessment and judgment criteria, and is standardized by organizations worldwide such as AAPM, DIN, JIRA, and IEC. On the other hand, gamma characteristics called Gamma 2.2, which are srgb defined, are very commonly used for displaying color images across the medical imaging workflow from acquisition to viewing.* (Fig. 1) Currently some medical grade monitor technologies are available to switch between DICOM Grayscale images and colored Gamma 2.2 values in order to display both monochrome and color images with accurate and appropriate gamma curve. (Fig.2) It normally requires manual operation and it is still not possible to display monochrome and color images at the same time on one screen. The Hybrid Gamma PXL is designed to solve this problem. The Hybrid Gamma PXL enables the display of both monochrome and color images with appropriate gamma characteristics without any manual operation. The Hybrid Gamma PXL function automatically distinguishes between monochrome and color images pixel by pixel, creating a hybrid display where each pixel has optimum grayscale; giving it greater accuracy and reliability. In this technical white paper, we outline this innovative technology, and discuss how this function fully complies with the DICOM part 14 standard while optimizing color medical images, by introducing a recent study conducted by Kumamoto Chuo Hospital in Japan. * The American Association of Physicists in Medicine (AAPM) and International Color Consortium (ICC) is working to standardize criteria to display color medical images because there is not any criteria and standard for them. White Paper (Q17B014-AS-10002) 2/11

GSDF 2.2 Fig. 1 Characteristics of GSDF and Gamma 2.2 CAL Switch Cal Switch Fig. 2 Multiple modes White Paper (Q17B014-AS-10002) 3/11

2 Outline of Hybrid Gamma PXL Although traditional color medical monitors can switch between GSDF and Gamma 2.2, they are mainly used with the GSDF gamma curve due to the inconvenience of manual operation. Therefore, there are cases where the color images are not ideally displayed. Fig. 3 shows an example of the issue. On the GSDF gamma curve, the mid tone of the color image gets darker and it might affect the readability of color images. GSDF γ2.2 GSDF Gamma 2.2 Fig. 3 Visual difference (Blue: monochrome images, Red: color images) The Hybrid Gamma PXL solves this problem. The Hybrid Gamma PXL detects monochrome and color pixel by pixel and assigns appropriate gamma curve characteristics in real time. This means that by enabling this function, all the monochrome pixels displayed on screen have accurate GSDF characteristics, while a Gamma 2.2-based curve is applied to the rest of the pixels detected as color pixels. Fig. 4 Algorithm of the Hybrid Gamma PXL On this Hybrid Gamma PXL function, this blending technique will be applied to some of the color images containing monochrome pixels, which eliminates unwanted noise effects on the color image and natural color image presentation. This white paper also validates this gamma 2.2-based curve used for color images. White Paper (Q17B014-AS-10002) 4/11

3 Study The goal of this study conducted by Dr. Kazuhiro Katahira, M.D., radiologist at the Kumamoto Chuo Hospital, was to assess the effectiveness and safety of the Hybrid Gamma PXL. There were three aspects of the Hybrid Gamma PXL that were assessed: 1) DICOM part 14 standard conformance on reading monochrome studies. 2) Diagnosis effectiveness of utilizing the Gamma2.2 curve on reading color studies. 3) Impact of utilizing gamma 2.2-based blended curve vs. true 2.2 Gamma curve. Period, place and subjects Period : May 23, 2017 to July 3, 2017 Place : A reading room in Kumamoto Chuo Hospital Subjects : Two radiologists with experiences of 12 and 27 years. Devices Monitor : Two 5MP color monitors (EIZO RadiForce RX560) * With the settings outlined in 'Setup' below. Test images Table 2 Test images Modality Num. Region X-ray 3 Breast, lumbus, bone Monochrome CT 3 Head, breast, abdomen MRI 3 Head, spine, knee 3D-CT 3 Heart, abdomen, spine 3D-MRI 1 Heart Color MRI fusion 2 Colon, prostate RI 50 Cardiac muscle, brain blood flow, ect. ES 20 Superior division, inferior division US 10 Mammary gland White Paper (Q17B014-AS-10002) 5/11

Setup Table 1 Monitor settings Setting No. Details (A) Hybrid Gamma PXL Activated (Lmin = 0.6, Lmax = 500 [cd/m 2 ]) (B) Calibrated to GSDF (Lmin = 0.6, Lmax = 500 [cd/m 2 ]) (C) Calibrated to Gamma 2.2 (Lmin = 0.6, Lmax = 500 [cd/m 2 ]) (A) Hybrid Gamma PXL (B)GSDF (A) Hybrid Gamma PXL (C)G2.2 Fig 5. Combination of the monitors Rating method and scale Three alternative forced choice (3-AFC) (A) > (B) or (A) > (C) : (A) is better for diagnosis. (A) = (B) or (A) = (C) : There is no significant difference. (A) < (B) or (A) < (C) : (B) or (C) is better for diagnosis. Table 3 Evaluation sheet Reader 1 (A)>(B) (A)=(B) (A)<(B) (A)>(C) (A)=(C) (A)<(C) Comment Image 1 (Region, case) 1 0 0 0 1 0 Image 2 (Region, case) 0 1 0 0 1 0 Image 3 (Region, case) 0 1 0 1 0 0 Test procedure 1) Display a test image both on (A) and (B). (Fig. 5, Left) 2) Choose only one out of two alternative options on (A) vs. (B) comparison 3) Display a test image both on (A) and (C). (Fig. 5, Right) 4) Choose only one out of two alternative options on (A) vs. (C) comparison Analysis Calculate total sum score by the following three conditions and compare the results: All modalities Monochrome modalities Color modalities * The total scores are average values of the observers. White Paper (Q17B014-AS-10002) 6/11

4 Results The following graphs show the total scores of all modalities, monochrome modalities and color modalities. Fig. 6 Hybrid Gamma PXL and GSDF (All modalities) Fig. 7 Hybrid Gamma PXL and 2.2 (All modalities) Fig. 8 Hybrid Gamma PXL and GSDF (Monochrome modalities) Fig. 9 Hybrid Gamma PXL and 2.2 (Monochrome modalities) Fig. 10 Hybrid Gamma PXL and GSDF (Color modalities) Fig. 11 Hybrid Gamma PXL and 2.2 (Color modalities) White Paper (Q17B014-AS-10002) 7/11

The followings are examples of comparison images. Fig. 12 Examples of RI images (Left: Hybrid Gamma PXL, Center: GSDF, Right: 2.2) White Paper (Q17B014-AS-10002) 8/11

Fig. 13 Examples of ES images (Left: Hybrid Gamma PXL, Center: GSDF, Right: 2.2) Fig. 14 Examples of CT images (Left: Hybrid Gamma PXL, Center: GSDF, Right: 2.2) White Paper (Q17B014-AS-10002) 9/11

The summaries for the three aspects are as follows. 1) DICOM part 14 standard conformance on reading monochrome studies. Fig. 8 shows that in 8.5 out of 9 monochrome image readings (about 94%), the Hybrid Gamma PXL mode (DICOM GSDF applied) is rated as equivalent to the DICOM GSDF calibrated display, and there was one case recorded where a subject rated Hybrid Gamma PXL (DICOM GSDF applied) as superior to DICOM GSDF mode. There was no case where DICOM GSDF mode was rated as superior to Hybrid Gamma PXL mode. As explained earlier, Hybrid Gamma PXL is designed to apply DICOM GSDF for all the monochrome pixels, this result confirms that Hybrid Gamma PXL for monochrome images is functioning properly and is performing as designed. 2) Diagnosis effectiveness of utilizing Gamma 2.2 curve on reading color studies. Fig. 10 shows that in 31.5 out of 86 color image readings (about 37%), the Hybrid Gamma PXL mode (Blended Gamma 2.2 applied) is rated as superior to the GSDF gamma curve, and these cases are either ES or RI color images, no US images. There was no case where the subjects rated DICOM GSDF as superior to Hybrid Gamma mode (Blended Gamma 2.2 applied) on color image reading. 3) Impact of utilizing gamma 2.2-based blended curve vs. true 2.2 Gamma curve. Fig. 11 shows that in 66 out of 86 color image readings (about 77%) the Hybrid Gamma PXL mode (Blended Gamma 2.2 applied) is rated as equivalent to the Gamma 2.2 calibrated display. In 8.5 out of 86 color image readings (about 10%), the subjects rated Hybrid Gamma PXL (Blended Gamma 2.2 applied) as superior to Gamma 2.2 mode. There is no negative impact by utilizing blended 2.2 Gamma curve on color image reading. 5 Conclusions < Safety > As confirmed on the summary (1), Hybrid Gamma PXL for monochrome images functions properly to apply DICOM GSDF to all monochrome images on screen and is equivalent to a GSDF calibrated display. This shows that the Hybrid Gamma PXL has DICOM part 14 standard conformance on monochrome images. Also from the result of the summary (3), there was no significant difference in diagnosis performance between the gamma 2.2-based curve and the true 2.2 gamma curve when reading color images. < Effectiveness > The result of the summary (2) revealed a significant difference (37% of total 86 color image readings) in diagnosis with the Hybrid Gamma PXL (Blended Gamma 2.2 applied) compared to the GSDF gamma curve. It indicates that reading color images with the Hybrid Gamma PXL is expected to be more effective in diagnostics than a GSDF gamma curve alone. We conclude that the Hybrid Gamma PXL maintains the same level as DICOM GSDF conformance for diagnosis while improving color image diagnosis performance without any risk. White Paper (Q17B014-AS-10002) 10/11

Reference 1. Digital Imaging and Communication in Medicine (DICOM) (2004), Part14: Grayscale Standard Display Function, http://dicom.nema.org/dicom/2004/04_14pu.pdf 2. IEC 61966-2-1:1999, Multimedia systems and equipment - Colour measurement and management - Part 2-1: Colour management - Default RGB colour space srgb https://webstore.iec.ch/publication/6169 Note: All figures are not based on the measurement result but simplified images. All product names are trademarks or registered trademarks of their respective companies. Copyright 2017 EIZO CORPORATION. All rights reserved. White Paper (Q17B014-AS-10002) 11/11