TECHNICAL INFORMATION DFO (1,8-Diazafluoren-9-One) Catalog Nos. LV500, LV5001, DFS200P

Similar documents
TECHNICAL INFORMATION Hungarian Red Catalog Nos. LV503, LV5031

TECHNICAL INFORMATION Crystal Violet (Methyl Violet) Catalog Nos. LV502, LV5021

TECHNICAL INFORMATION GoldPanther Forensic Light Source Kit (Patent Pending) Cat. Nos. FAL2000, FAL

TECHNICAL INFORMATION BLUEMAXX Digital Latent Evidence Photo System Catalog No. BML100

TECHNICAL INFORMATION BLUEMAXX Portable Illumination Systems Catalog Nos. BM600, BM600220

TECHNICAL INFORMATION BLUEMAXX Rechargeable Forensic Lights Catalog Nos. BM500, BM500220

TECHNICAL INFORMATION Trace Metal Detection Kit Catalog No. TMDT100

Latent Print Development TECHNIQUES

SIRCHIE. 1 of 20. Products Vehicles Training

TECHNICAL INFORMATION LumaVision Luminol Companion Illuminator Kits Catalog Nos. LVL1000, LVL2000

PAPER No. 3; FINGERPRINTS AND OTHER IMPRESSIONS Module No. 19; Silver Nitrate Method

Labino Nova Torch 8 LAMPS SMALL LIGHT WEIGHT PERFECT LIGHT BEAM TECHNICAL SPECIFICATION.

User's Manual BLUESTAR FORENSIC TABLETS

Lighting Techniques 18 The Color of Light 21 SAMPLE

JY Division I nformation

TECHNICAL INFORMATION Tactical MAX Forensic Light Kit Catalog No. TMX100

Unit 2 Review-Fingerprints. 1. Match the definitions of the word on the right with the vocabulary terms on the right.

Fingerprints. Sierra Kiss

Alternate Light Sources

TECHNICAL INFORMATION Portable Fingerprint Taking Outfits Catalog Nos. PFT500PF, PFT500CM, PFT500, PFT600, PFT600PIP

Alternate Light Sources

Crime-lite 82 NEW. Infrared model UV-Vis-IR camera Wheeled laboratory cart. foster + freeman

The 20 th Conference of Korean Society of Forensic Science. (Hemascein TM ) Validation of novel latent bloodstain detection kit Hemascein TM , 2

SER: Biological Stains Visualization with Alternate Light Sources

GLS can. Gellifter imaging system. Easy to operate and results in less then a minute. Detect the weakest of traces excellent illumination

The study of fingerprints for identification purposes is known as dactylography or dactyloscopy.

Crime-lite 2 A complete range of handheld LED light sources for crime scene investigation and forensic laboratory examination

KODAK PRIMETIME 640T Teleproduction Film / 5620,7620

TECHNICAL DATA / COLOR NEGATIVE FILM February 1998 E-44

SmartDoc 2.0 E5001-SDB Instruction Manual

Technical Procedure for the UltraLite ALS

KODAK PROFESSIONAL ELITE Chrome 200 Film

KODAK EKTACHROME 64 Professional Film

EASTMAN EXR 200T Film 5287, 7287

KODAK PROFESSIONAL T-MAX P3200 Black & White Negative Film

ELITE Chrome 400 Film

EASTMAN EXR 500T Film 5298

KODAK High Definition 400 Film

Fingerprints - Formation - Fingerprints are a reproduction of friction skin ridges that are on the palm side of fingers and thumbs

FORENSIC SCIENCE Fingerprints

KODAK EKTACHROME 160T Professional Film / EPT

Crime-lite. Bench top magnifier with multi-wavelength illumination for general search and fluorescence examinations

KODAK PROFESSIONAL T-MAX P3200 Black & White Negative Film

1 MPTVI DATA SHEET XXXXXXXXXXX

KODAK PROFESSIONAL EKTACHROME Films E100G and E100GX Grain as Fine as Your Art.

EXAMINER+ Manual V.1. ARROWHEAD FORENSICS Strang Line Road Lenexa, Kansas PHONE FAX

ELITE Chrome 100 Film

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I

KODAK VISION Expression 500T Color Negative Film / 5284, 7284

EASTMAN EXR 200T Film / 5293, 7293

Friction ridge detail of the fingers, palms and feet is among the

KODAK EKTACHROME 100 Plus Professional Film

Recording The Incident. Forensic Science

KODAK EKTACHROME 64 Professional Film

FORMULARY P.O. Box 950 Condon MT FAX

KODAK EKTACHROME RADIANCE III Paper

Fluorescing mud prints

TECHNICAL DATA / COLOR REVERSAL FILM January 1998 E-154

LITESTAGE USER'S GUIDE

Mr. Futrell is a supervisory fingerprint specialist in the Latent Fingerprint Section of the FBI Laboratory in Washington, D.C.

Failure is a crucial part of the creative process. Authentic success arrives only after we have mastered failing better. George Bernard Shaw

TECHNICAL DATA / COLOR TRANSPARENCY FILM

KODALITH Ortho Films 2556,6556, Type 3

Guide to Manual Processing of NDT Films

Understanding f-stopsf

KODAK EKTACHROME 64T Professional Film

Education and Training

Aperture: Circular hole in front of or within a lens that restricts the amount of light passing through the lens to the photographic material.

FC-2500 Quick Reference Guide

400BW TECHNICAL DATA / BLACK-AND-WHITE FILM. BLACK and WHITE FILM. February 2001 F-4012

Basics of Footwear/Tire Tracks Impressions March 5, 2015

OHIO STATE UNIVERSITY EXTENSION

ULTRAVIOLET and INFRARED Photography Summarized

TECHNICAL DATA / COLOR PAPER July 2003 E-23

Footwear & Tire Track Evidence

TECHNICAL DATA / COLOR PAPER July 2003 E-19

Tools in Forensic Science Student Notes

EASTMAN TRI-X Reversal Film 7278

Department of Pathobiology, Faculty of Science, Mahidol University, Bangkok 10400, Thailand

Detection of Pretreated Fingerprint Fluorescence Using an LED-based Excitation System

KODAK PROFESSIONAL T400 CN Film (Process C-41)

Density vs. Contrast

KODAK PROFESSIONAL TRI-X 320 and 400 Films

TECHNICAL INFORMATION

KODAK PROFESSIONAL EKTACHROME Film E100VS

Technical Procedure for Liquid Adhesive Lifters

TECHNICAL INFORMATION

TECHNICAL INFORMATION Tire and Footprint Plaster Casting Kit Catalog No. 639C

ISO 400/27 o BLACK AND WHITE PROFESSIONAL FILM FOR HIGH PRINT QUALITY AND FLEXIBILITY IN USE

ISO 400/27º, BLACK AND WHITE PROFESSIONAL FILM FOR HIGH PRINT QUALITY AND FLEXIBILITY IN USE

DEVELOPMENT ON MULTICOLOURED SURFACES, A PROBLEM RESOLVED USING FLUORESCENT DYES

Microscope (and The Cell) Lab Exercise #1

Unit 5- Fingerprints and Other Prints (palm, lip, shoe, tire)

RUVIS & ALS Systems.

Wimberley Developer #2, Version D+ 1 liter kit (Dry) CATALOG NO TO MAKE 50 LITERS OF WORKING SOLUTION

SFX FACT SHEET BLACK AND WHITE CAMERA FILM WITH EXTENDED RED SENSITIVITY FOR CREATIVE PHOTOGRAPHY. April 2004

:AVIPHOT COLOR N400 PE1

FUJICHROME PROVIA 1600 Professional [RSP]

DELTA 400 PROFESSIONAL

Glossary of Terms (Basic Photography)

Transcription:

SIRCHIE Products Vehicles Training Copyright 2011 by SIRCHIE All Rights Reserved. TECHNICAL INFORMATION DFO (1,8-Diazafluoren-9-One) Catalog Nos. LV500, LV5001, DFS200P INTRODUCTION DFO is a ninhydrin analog used for developing latent prints on porous surfaces. Its fluorochrome properties make it especially useful on multi-colored surfaces where the background would interfere with subsequent photography. DFO reacts with the amino acid content of fingerprint residue, and it will not interfere with subsequent analysis using ninhydrin, silver nitrate or physical developer. In fact, some labs report that DFO use improves the results gained with ninhydrin. According to the Manual of Fingerprint Development Techniques published by the Police Scientific Development Branch of the British Home Office, DFO will develop more fingerprints than ninhydrin. Testing in the U.S. indicates that DFO will develop 2 1/2 times as many prints as ninhydrin. Other testing indicates that DFO is also effective for developing latent blood prints. DFO developed latent 100 HUNTER PLACE, YOUNGSVILLE, NC 27596 USA Ph: (919) 554-2244, (800) 356-7311 Fax: (919) 554-2266, (800) 899-8181 Web: www.sirchie.com Email: sirchieinfo@sirchie.com 1 of 8

On occasion, DFO will produce lightly colored (pink) latent prints, but its real strength is in its fluorescent properties. DFO fluorescence may be produced using a UV light source such as BLUEMAXX, FAL3000 Forensic Alternative Light Source Kit, and other alternative light sources. DFO is now packaged in a pump spray which contains 100ml of pre-mixed reagent for better control and less waste. DFO in crystal form is available for those labs preparing their own formulations. CAUTIONS Before using this kit, consult the appropriate Material Safety Data Sheets (MSDS) found on our website at www.sirchie.com/support. Use only in well-ventilated area or with a fume hood. Wear latex gloves, safety glasses, and an organic vapor respirator. Some DFO formulations are flammable and highly toxic. Use DFO prior to applications of ninhydrin. DFO can interfere with the forensic examination of handwriting (inks may run), indented impressions, body fluids including DNA profiling, fibers, hairs, paint and other types of evidence. Handle objects with gloved hands before and after treatment. PREPARATION DFO Pump Spray Catalog No. DFS200P is a DFO formula that is delivered from a pump spray bottle. Be certain that the subject paper is dry and that all preliminary examinations of ink or handwriting have been completed. To apply the reagent, shake the bottle well, and then spray the surface from a distance of from six to nine inches. Cover the entire surface with an even spray. Apply the reagent in a well-ventilated area or within the confines of a fuming hood. DFS200P has a useful life of up to one year. 2 of 8

To accelerate development, treat the object in a 212ºF (100ºC) oven for 10 to 20 minutes. Examine the object under UV light such as the BLUEMAXX or FAL3000 Forensic Alternative Light Source Kit. Note: See the section on photography for recording the evidence. DFO Powder Formulations Special Considerations: Due to the limited shelf life of some solutions prepared in the laboratory, DFO should be stored in its powder form until needed. Working solutions should not be prepared until you are ready to develop prints. FBI Working Solutions have a useful life of about 6 months. Wear gloves to prevent weakening or contamination of the solutions with sweat components from hands and possible staining of skin. Solution should be prepared in a well-ventilated area or within the confines of a fuming hood. FORMULA NO. 1 FBI FORMULATION (Stock Solution) 1 gram DFO Crystals 200ml Ethyl Acetate 200ml Methanol 40ml Glacial Acetic Acid Combine and stir with a magnetic stirrer until ALL the ingredients are dissolved. CAUTION: Highly flammable! (Working Solution) Add Petroleum Ether to the Stock Solution until total volume is 2 liters. FORMULA NO. 2 STANDARD (Stock Solution).25 gram DFO 20ml Glacial Acetic Acid Crystals 40ml Methanol 1- Dissolve.25g of DFO in 250ml beaker with 40ml of Methanol. 2- Then, and 20ml of Glacial Acetic Acid and mix until all DFO is desolved into the solution. CAUTION: Highly flammable! (Working Solution) Add 940ml of 3M NOVEC Fluid HFE-7100 to the Stock Solution. Cover and allow to settle for approx. 30 minutes.* 3 of 8

*FORMULA NO. 2 NOTE: A thin, oily film may form on top of the working solution. It consists of water, excess Methanol and DFO it must be removed prior to use following one of these methods: 1. Use a separatory funnel to process the working solution. Allow to settle for 30 min. and drain the bottom phase into a storage container. Stop draining when separate, clear-looking solution nears the bottom of the funnel, or approx. 50-100ml of solution remains in funnel. This remainder contains undissolved ethanol, water and DFO and should be discarded in a proper waste container for flammable solvents. 2. Use a pipette to skim the oily film from the top of the working solution. Again, discard this in a proper waste receptacle as described above. 3. Use a squirt bottle to contain the working solution. This will help insure a clean solution. When the solution level is below the straw, discard the remainder in a proper waste receptacle as described above. PROCEDURE Dip the specimen to be processed into the working solution for about ten seconds and then allow it to air dry. It is possible to apply the reagents mixed according to Formula No. 1 and Formula No. 2, using a pump spray, but care should be taken. The technician should spray only in a fume hood (Fig. 1). After drying the specimen, development may be accelerated by applying heat at 212ºF (100ºC) for 10-20 minutes. For the optimum acceleration environment, SIRCHIE designed several heat chambers including the No. DFC100 and No. DFC200. After acceleration is complete, remove the object from the heat chamber and examine under UV light. Suggested wavelengths for viewing DFO prints are 450nm, 485nm, 525nm and 530nm with orange filter goggles for most porous surfaces such as paper (white or pastel) and glassine envelopes. Yellow legal pad papers are best viewed at 570nm to 590nm with a red filter. 4 of 8 FIGURE 1 DFO will often produce visible prints using the No. AC632 Forensic Workstation.

Photography Note: Video and film cameras may not exhibit spectral sensitivity identical to the human eye. Put another way, what you see you may not be what you get. As a consequence, it is difficult to develop hard and fast rules for the use of these instruments in recording fluorescence. It is possible, however, to develop general rules for fluorescence photography. Use of Light Meters Unless you are fortunate enough to have a highly specialized spot meter, you will find that normal hand-held or in-camera light meters are of no use in fluorescence photography. Do not rely on the camera s automatic mode or upon recommendations that its light meter may offer. Use bracketing techniques and expect exposure times of a half-minute or longer. Steady The Camera Because of the extremely low light levels involved with fluorescence, photography involves long exposures. Photography must be done with the camera mounted on a sturdy tripod and the room should be totally dark or you must use a light-tight enclosure. Illumination Techniques The only illumination present must be from a suitable UV light such as the BLUEMAXX or FAL3000 Forensic Alternative Light Source Kit. Stray light must be eliminated and illumination must be as even as possible. Good practice suggests that you paint the light onto the surface being photographed by moving the illumination spot around during exposure. This will tend to even out the illumination over the period of the exposure. Fill The Frame Fluorescence intensity falls off with distance so it is important to position the camera as close to the subject as possible while filling the viewfinder frame with the image. If you use a commercial lab for film processing, allow a 20% border around the subject to allow for their cropping of the photos. If you are using a 35mm camera, specify 4" x 6" prints. 5 of 8

Use A Barrier Filter A barrier filter on the camera is necessary because of the same reason you needed one for your eyes. If you are using standard UV light sources, equip the camera with a filter that is the same color as the goggles used to view the fluorescence. Although it is possible to photograph through the barrier filters supplied with BLUEMAXX lights, we recommend filters specifically designed for use with your camera. Please note that some commercially available filters may fluoresce weakly when exposed to the BLUEMAXX beam. This will result in hazy or foggy photos. Specify non-fluorescing threaded filters or Cokin -type sheet barrier filters available from your local camera distributor. Films All films are sensitive to blue light. If a barrier filter were not used, the blue light would overpower the fluorescence and mask it. In this application Daylight films are superior to specialty films such as Tungsten or similar light compensated films. Daylight films generally have red, green and blue layers of equal sensitivity. Tungsten films are manufactured to be especially sensitive to blue and are not suitable for UV photography. Use the highest speed film you can find. Reduce the effect of graininess by filling the frame, thus reducing the enlargement factor. Use a large format camera for the same reason. Black & white films are excellent for recording fluorescence due to their high speed and relatively fine grain. A barrier filter is still necessary with black and white film. Aperture Settings The camera s aperture setting controls the size of the opening in the lens iris and thus controls the amount of light reaching the film. Because fluorescence is so weak, you may be tempted to open the aperture as wide as possible to reduce exposure times. But keep in mind that the aperture also controls depth of field or depth of focus. Use of a small aperture means sharpest focus over an extended object. Focus is especially critical when photographing at close range. We recommend using longer exposures and smaller apertures (f/8 to f/22). Reciprocity Failure If you experiment with the relationship between the amount of light present from a 6 of 8

7 of 8 TI02-28ENG-REV7 scene, the length of time that the camera shutter is open, and the density of the exposure created on film, you will find that for normal photography, a simple relationship exists: exposure=light intensity x time. When exposures are long or light intensities are either very low or very high, this relationship fails to hold together. This is known as reciprocity failure. The consequences of this in fluorescence photography are seen in color shifts and in a general decrease in film speed. Color shifts can be compensated for with color-compensating filters if it is felt necessary. (Consult photography texts for assistance.) You will find that camera settings no longer scale in a linear fashion (i.e., doubling the aperture size or exposure time does not double the density of the exposure). What all of this leads up to is that fluorescence photography is an art. Keeping Records Successful fluorescence photography requires practice, experience and patience. As you gain experience, your judgment of exposure times will improve. The learning curve can be greatly reduced if you record your experiences in a notebook. Data worth recording will include the subject being photographed, aperture settings, film used, exposure times, and a copy of the resultant photograph. Starting Point The table shown to the right was developed to provide a starting point for your photographic efforts: Using Digital Camera Equipment Digital photography is rapidly replacing the use of film cameras. Among the many advantages digital equipment offers to the law enforcement photographer are: 1. The LCD screen provides instant viewing of the photos taken. 2. Portable digital printers permit quick printouts of photos at the crime scene. 3. All photo records can be stored on a computer and easily accessed. 4. Pricing is now within the range of most agencies. Digital cameras are used in the same manner as film cameras in that they are available with adjustable set-

tings for ISO/ASA, shutter speeds, and aperture settings. Using the suggested starting points given for film cameras, the evidence photographer can determine the proper settings to use. FILM SPEED: ASA/ISO400 Camera-to-Subject Distance = 8" (20.3cm) SUBJECT APERATURE TIME Latent Fingerprints f/8 10 seconds We recommend that the digital camera be set up for an ISO/ASA of 100. Most digital cameras offer a range of quality settings. Use the highest quality setting available. For crime scene use, we recommend a camera with a high resolution. 8 of 8