NI PXIe Contents. Required Software CALIBRATION PROCEDURE. Dual-Output Programmable DC Power Supply

Similar documents
NI PXIe Contents CALIBRATION PROCEDURE. Four Channel Precision Source-Measure Unit (SMU)

NI PXIe Contents CALIBRATION PROCEDURE. Single-Channel Precision Source-Measure Unit (SMU)

Contents. Software Requirements

PXIe, 7½-Digit, ±1,000 V, Onboard 1.8 MS/s Isolated Digitizer, PXI Digital Multimeter

PXIe Contents. Required Software CALIBRATION PROCEDURE

CALIBRATION PROCEDURE PXIe-4302/4303 and TB-4302C 32 Ch, 24-bit, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module. ni.

Calibrating the NI 5653 requires you to install one of the following packages on the calibration system. NI-RFSA 2.4 or later NI-RFSG 1.

Contents. CALIBRATION PROCEDURE NI PXIe-5698

NI PXIe-5171R. Contents. Required Software CALIBRATION PROCEDURE

Contents. Software. Documentation. CALIBRATION PROCEDURE NI PXIe-6674T

Contents. CALIBRATION PROCEDURE NI PXIe-4463 DSA Analog Output

CALIBRATION PROCEDURE NI PXIe-4330/4331. Contents

Contents. Software Requirements. Documentation Requirements. Calibration Interval CALIBRATION PROCEDURE NI TB-4300B

PXIe Contents CALIBRATION PROCEDURE. 10 GHz or 20 GHz RF Analog Signal Generator

GETTING STARTED GUIDE NI khz to 100 MHz, 50 Ω, AC Coupled RF Receiver Module

Required Software. Related Documentation. Password. Calibration Interval

GETTING STARTED GUIDE NI AI, ±5 V, 24 Bit, 51.2 ks/s/ch Simultaneous, AC/DC Coupling, IEPE AC Coupling

PXIe, 24-Channel, ±24 V, 50 ma Precision PXI Source Measure Unit

Contents. Software Requirements. CALIBRATION PROCEDURE NI PXIe-5663E

NI 6624 Calibration Procedure

PXIe Contents. Required Software CALIBRATION PROCEDURE

Contents. Software Requirements CALIBRATION PROCEDURE NI PXI-5663

Power Amplifier. Input Characteristics. SPECIFICATIONS NI PXIe-4610 Power Amplifier. Signal Range

GETTING STARTED GUIDE NI AI, ±10 V, 24 Bit, 50 ks/s/ch Simultaneous

GETTING STARTED GUIDE NI AI, ±60 V, 24 Bit, 1 ks/s/ch Simultaneous

if it is used in a manner not described in the user documentation.

Contents. CALIBRATION PROCEDURE PXIe-5673 Vector Signal Generator

is used in a manner not described in the user documentation.

GETTING STARTED GUIDE NI AI Differential/32 AI Single-Ended, ±200 mv to ±10 V, 16 Bit, 250 ks/s Aggregate, Fuel-Cell Measurements

Input Characteristics. Measurement Range. SPECIFICATIONS NI PXIe-4357

DEVICE SPECIFICATIONS NI PXIe Ch, 24-bit, 25.6 ks/s Simultaneous AC LVDT Input Module

NI sbrio-9632/9642 Verification Procedure

SCXI Terminal Block

Analog Characteristics

4-Channel, ±24 V, 150 ma Precision PXI Source Measure Unit

USRP-2940/2942/2943/2944/2945

4-Channel ±10 V, 100 ma, Precision PXI Source Measure Unit

These specifications apply to the PXIe-5113 with 64 MB and 512 MB of memory.

4-Channel, ±6 V, 500 ma Precision PXI Source Measure Unit

4-Channel ±10 V, 100 ma, Precision PXI Source Measure Unit

4-Channel, ±24 V, 150 ma PXI Source Measure Unit

Current ON state current ma minimum, 6 ma typical, 10 ma maximum OFF state current ma maximum

NI PXI-2521 Specifications 40-Channel DPST Relay Module

NI PXI-4110 Calibration Procedure Addendum

PXIe Contents SPECIFICATIONS. 32-Channel Digital Pattern Instrument

DEVICE SPECIFICATIONS NI PXIe-4304/ Ch, 24-bit, ±42 V, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module. ni.

Contents. CALIBRATION PROCEDURE NI PXIe-6555/6556. ni.com/manuals

SPECIFICATIONS PXIe Ch (8 ADC), 2 MS/s, bit, Flexible Resolution PXI Analog Input Module

PXIe Contents SPECIFICATIONS. 100 MVector/s PXI Digital Pattern Instrument

OPERATING INSTRUCTIONS AND SPECIFICATIONS. NI 9211E 4-Channel Thermocouple Input Module

NI PXI Contents DEVICE SPECIFICATIONS. RF Attenuator

PXI Contents. Definitions SPECIFICATIONS. ±100 V, 2 W Precision PXI Source Measure Unit. These specifications apply to the PXI-4132.

Note Using the PXIe-5785 in a manner not described in this document might impair the protection the PXIe-5785 provides.

DEVICE SPECIFICATIONS NI PXIe-4302/ Ch, 24-bit, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module

Contents. CALIBRATION PROCEDURE NI 5421/ MS/s Arbitrary Waveform Generator

NI PXIe-5630 Specifications

Contents CALIBRATION PROCEDURE NI PXI-5422

NI PXI-2520 Specifications 80-Channel SPST Relay Module

PXIe Contents SPECIFICATIONS. RF Preselector Module

PXIe Contents SPECIFICATIONS. RF Amplifier

Contents. CALIBRATION PROCEDURE NI PXIe-5668R 14 GHz and 26.5 GHz Signal Analyzer

SPECIFICATIONS PXIe channel, 16-bit, 400 ks/s/ch, Ch-Ch Isolated Analog Input Module. ni.com/manuals

OPERATING INSTRUCTIONS AND SPECIFICATIONS. NI 9207E 16-Channel, ±20 ma/±10 V, 24-Bit Analog Input Module

OPERATING INSTRUCTIONS NI Port, Low-Speed/Fault Tolerant CAN Module

PXIe Contents SPECIFICATIONS. ±200 V, 1 A Precision System PXI Source Measure Unit. These specifications apply to the PXIe-4137.

SCXI MHz Quad 8x1 50 Ω Multiplexer

PCI Express, 6½-Digit, ±300 V Digital Multimeter Device

PXIe Contents SPECIFICATIONS. ±200 V, 1 A System PXI Source Measure Unit. These specifications apply to the PXIe-4136.

This document lists the specifications for the NI PXI The PXI-5900 is a differential amplifier.

PXI Contents SPECIFICATIONS. ±20 V, 40 W PXI Source Measure Unit

PXIe Contents SPECIFICATIONS. ±60 V, 3 A Precision System PXI Source Measure Unit. These specifications apply to the PXIe-4139.

PXIe Contents SPECIFICATIONS. ±60 V, 3 A System PXI Source Measure Unit. These specifications apply to the PXIe-4138.

NI PXI ½ Digit FlexDMM Calibration Procedure

Contents CALIBRATION PROCEDURE NI 5412

PXIe Contents SPECIFICATIONS. Precision, 10 fa Sensitivity Low-Current PXI Source Measure Unit. These specifications apply to the PXIe-4135.

NI PXI-2797 Specifications 40 GHz 6 1 Terminated 50 Ω Multiplexer

PXIe Contents CALIBRATION PROCEDURE

DEVICE SPECIFICATIONS Phase Matrix PXI GHz PXI Vector Signal Analyzer

NI PXIe Contents. About These Specifications DEVICE SPECIFICATIONS. 3 GHz Ω Multiplexer (SP16T)

Terminology. Output Characteristics. DEVICE SPECIFICATIONS NI PXIe-4463 DSA Analog Output. ni.com/manuals

SCXI HIGH-VOLTAGE ATTENUATOR TERMINAL BLOCK

R Series for USB Multifunction RIO with Kintex-7 70T FPGA

100 MHz Bandwidth, 12-Bit PXI Waveform Generator

Contents. CALIBRATION PROCEDURE NI PXIe GHz and 14 GHz RF Vector Signal Analyzer

NI PXIe-5667 (3.6 GHz)

Contents. USER MANUAL NI ISM-7400 Integrated Stepper

Reconfigurable 6 GHz RF Vector Signal Transceiver with 1 GHz Bandwidth

PXIe Contents. Definitions SPECIFICATIONS. 8.5 GHz Vector Network Analyzer

NI PXIe-5601 Specifications

Analog Input. DEVICE SPECIFICATIONS NI PXIe-4300

Specifications for the NI PXI/PCI-6552/6551

SCXI TERMINAL BLOCK

PXIe Contents SPECIFICATIONS. 250 khz to 20 GHz RF Analog Signal Generator

NI PXI/PXIe-2529 Specifications

R Series for USB Multifunction RIO with Kintex-7 70T FPGA. may impair the protection the NI USB-7855R provides.

GETTING STARTED GUIDE NI Port DeviceNet Module

NI PXI-2557 Specifications

NI Introduction. Contents CALIBRATION PROCEDURE

PCI Contents SPECIFICATIONS. 8-Channel, 12-Bit, 60 MHz PCI Oscilloscope Device

PXIe Contents. Definitions SPECIFICATIONS. 8.5 GHz Vector Network Analyzer

Transcription:

CALIBRATION PROCEDURE NI PXIe-4113 Dual-Output Programmable DC Power Supply This document contains the verification and adjustment procedures for the NI PXIe-4113 (NI 4113). Refer to ni.com/calibration for more information about calibration solutions. Contents Required Software...1 Related Documentation...2 Password... 2 Calibration Interval... 2 Test Equipment...2 Test Conditions...3 Calibration Overview...4 Verification...5 Voltage Output and Measurement...5 Current Output and Measurement...7 Load Regulation...9 Remote Sense...12 Adjustment... 13 Adjusted Specifications...13 Initiating the Adjustment Session... 14 Connecting and Configuring Equipment for Voltage Adjustment...14 Adjusting Voltage Output and Measurement... 14 Connecting and Configuring Equipment for Current Adjustment...15 Adjusting Current Output and Measurement... 16 Closing the Adjustment Session... 17 Reverification...17 Worldwide Support and Services... 17 Required Software Calibrating the NI 4113 requires you to install the following software on the calibration system: NI-DCPower 15.1 or later. Supported application development environment (ADE) LabVIEW or LabWindows /CVI. Supported operating system Windows.

You can download all required software from ni.com/downloads. When you install NI-DCPower, you need to install support only for the application software that you intend to use. Access calibration support in the locations shown in the following table. ADE Calibration Support Location LabVIEW LabWindows/CVI NI-DCPower Calibration palette NI-DCPower function panel (nidcpower.fp) Related Documentation You might find the following documents helpful as you perform the calibration procedure: NI PXIe-4112/4113 Getting Started Guide NI DC Power Supplies and SMUs Help NI PXIe-4113 Specifications NI-DCPower Readme LabVIEW Help Visit ni.com/manuals for the latest versions of these documents. Password The default password for password-protected operations is NI. Calibration Interval Recommended calibration interval 2 years Test Equipment The following table lists the equipment NI recommends for the performance verification and adjustment procedures. If the recommended equipment is not available, select a substitute using the minimum requirements listed in the table. 2 ni.com NI PXIe-4113 Calibration Procedure

Table 1. Required Equipment for Calibration Required Recommended Parameter Minimum Equipment l(s) Measured Specifications One digital multimeter (DMM), optional second DMM 1 NI PXI-4071 All parameters Voltage: <±200 ppm accuracy and <30 µv resolution Programmable electronic load Agilent N3302A or Sorensen SLM-60-30-150 Load regulation Constant current and constant voltage modes: sink at least 6 A at 10 V 100 mω precision current shunt Guildline 9230 A-30 or Ohm Labs CS10 Current output and measurement, voltage and current load regulation ±100 ppm tolerance, ±50 ppm stability, ±5 ppm/c temperature coefficient. Minimum current 10 A. Two 50 Ω resistors Vishay PTF5650R000BZEK Remote sense output 0.1% tolerance, 1/8 W 1 kω resistor Vishay PTF651K0000BYEK Remote sense output 0.1% tolerance, 1/4 W Test Conditions The following setup and environmental conditions are required to ensure the NI 4113 meets published specifications: Keep cabling as short as possible. Long cables act as antennas, picking up extra noise that can affect measurements. Verify that all connections to the device, including front panel connections and screws, are secure. Ensure that the PXI chassis fan speed is set to HIGH, that the fan filters (if present) are clean, and that the empty slots contain slot blockers and filler panels. For more information about cooling, refer to the Maintain Forced-Air Cooling Note to Users document available at ni.com/manuals. 1 If you make measurements using only one DMM, sequentially set up the DMM connections as specified by the procedure steps. NI PXIe-4113 Calibration Procedure National Instruments 3

Allow a warm-up time of at least 30 minutes after the chassis is powered on and NI-DCPower is loaded and recognizes the NI 4113. The warm-up time ensures that the NI 4113 and test instrumentation are at a stable operating temperature. Use shielded copper wire for all cable connections to the device. Use twisted-pair wire to eliminate noise and thermal offsets. To ensure the system has had adequate time to settle, wait one second after requesting a new current or voltage or after changing a load before taking a measurement. Keep relative humidity between 10% and 70%, noncondensing. Test limits in this document are based on the December 2014 edition of the NI PXIe-4113 Specifications. Set the nidcpower Aperture Time property or NIDCPOWER_ATTR_APERTURE_TIME attribute to 1 power-line cycle (PLC). Set the nidcpower Aperture Time Units property or NIDCPOWER_ATTR_APERTURE_TIME_UNITS to power line cycles. Do not use the NI-DCPower Soft Front Panel (SFP) to request test points for any adjustment functions because you cannot set aperture time using the SFP. Ensure that properties or attributes for the device that are not specified in calibration procedures are set to their default values. Plug the chassis and the instrument standard into the same power strip to avoid ground loops. When making measurements, configure any specified digital multimeters (DMMs) with the best available levels and limits for each specified test point. An external power supply capable of providing 48 V and greater than 3.5 A should be connected to the AUX Power front panel input at all times. Temperature conditions specific to verification procedures: Maintain an ambient temperature of 23 C ± 5 C. Temperature conditions specific to adjustment procedures: Maintain an ambient temperature of 23 C ± 1 C. The NI 4113 temperature will be greater than the ambient temperature. Calibration Overview Calibration includes the steps shown in the following figure. 4 ni.com NI PXIe-4113 Calibration Procedure

Figure 1. Calibration Overview Verify Document Pre-Adjustment Results Adjust (Calibration Constants, Dates, and Temperatures Updated) Verify Document Post-Adjustment Results Yes Meets Calibration Test Limits? No Calibration/ Verification Complete Review Verification/ Adjustment Procedure or Return Device 1. Initial Setup Install the NI 4113 and configure it in Measurement & Automation Explorer (MAX). 2. Verification Verify the existing operation of the NI 4113. This step confirms whether the device is operating within the published specifications prior to adjustment. 3. Adjustment Adjust the calibration constants of the NI 4113. 4. Reverification Repeat the Verification procedure to ensure that the device is operating within the published specifications after adjustment. Verification The performance verification procedures assume that adequate traceable uncertainties are available for the calibration references. Voltage Output and Measurement Connecting and Configuring Equipment for Voltage Verification 1. Make the necessary connections for this procedure, as shown in the following figure. NI PXIe-4113 Calibration Procedure National Instruments 5

Figure 2. Voltage Verification or Adjustment Connection Diagram NI-DCPower Device Sense Sense DMM Voltage 2. Set the nidcpower Output Function property or NIDCPOWER_OUTPUT_FUNCTION attribute to DC Voltage for the NI 4113. Verifying Voltage Output Compare a set of voltages measured by an external DMM to the voltage test points requested by the NI 4113. Refer to the following table as you complete the following steps. Table 2. Voltage Output Verification Level Range Limit Range and Limit Test Point As-Found Output Test Limit (% of Voltage Offset) 10 V 6 A 0.1 V 0.1% 15 mv 10 V 1. Set the first specified level range, limit range, and limit on the NI 4113. 2. Set the level on the NI 4113 to the first specified test point. 3. Compare a DMM voltage measurement to the voltage output test limits. a) Take a voltage measurement using the DMM. b) Calculate the lower and upper voltage output test limits using the following formula: Voltage Output Test Limits = Test Point ± ( Test Point % of Voltage Offset) c) Verify the DMM measurement falls within the test limits. 4. Repeat the previous two steps for each test point specified in the level range. 5. If more than one level range is specified, repeat the previous steps using the values specified in each level range. Verifying Voltage Measurement Compare a set of measured voltages reported by the NI 4113 to the voltages measured by an external DMM. Refer to the following table as you complete the following steps. 6 ni.com NI PXIe-4113 Calibration Procedure

Table 3. Voltage Measurement Verification Level Range Limit Range and Limit Test Point As-Found Measurement Test Limit (% of Voltage Offset) 10 V 6 A 0.1 V 0.1% 15 mv 10 V 1. Set the first specified level range, limit range, and limit on the NI 4113. 2. Set the level on the NI 4113 to the first specified test point. 3. Compare a NI 4113 voltage measurement to the voltage measurement test limits. a) Take a voltage measurement using the DMM, and take a voltage measurement using the NI 4113. b) Calculate the lower and upper voltage measurement test limits using the following formula: Voltage Measurement Test Limits = DMM Measured Voltage ± ( DMM Measured Voltage % of Voltage Offset) c) Verify the NI 4113 measurement falls within the test limits. 4. Repeat the previous two steps for each test point specified in the level range. 5. If more than one level range is specified, repeat the previous steps using the values specified in each level range. Current Output and Measurement Connecting and Configuring Equipment for Current Verification 1. Make the necessary connections for this procedure, as shown in the following figure. Figure 3. Current Verification or Adjustment Connection Diagram NI-DCPower Device Sense Sense Precision Shunt DMM Voltage 2. Set the nidcpower Output Function property or NIDCPOWER_OUTPUT_FUNCTION attribute to DC Current for the NI 4113. NI PXIe-4113 Calibration Procedure National Instruments 7

Verifying Current Output Compare a set of currents measured by an external DMM to the current test points requested by the NI 4113. Refer to the following table as you complete the following steps. Table 4. Current Output Verification Level Range Limit Range and Limit Test Point As-Found Output Test Limit (% of Current Offset) 6 A 10 V 0.1 A 0.15% 20 ma 6 A 1. Set the first specified level range, limit range, and limit on the NI 4113. 2. Set the level on the NI 4113 to the first specified test point. 3. Calculate the current through the shunt by completing the following steps. a) Take a voltage measurement across the shunt using the DMM. b) Divide the voltage measurement by the calibrated value of the shunt. c) Record the calculated value as DMM Measured Current. 4. Calculate the lower and upper current output test limits using the following formula: Current Output Test Limits = Test Point ± ( Test Point % of Current Offset) 5. Verify that the recorded DMM value falls within the test limits. 6. Repeat the previous four steps for each test point specified in the level range. 7. If more than one level range is specified, repeat the previous steps using the values specified in each level range. Verifying Current Measurement Compare a set of measured currents reported by the NI 4113 to the currents measured by an external DMM. Refer to the following table as you complete the following steps. Table 5. Current Measurement Verification Level Range Limit Range and Test Point As-Found Measurement Test Limit Limit (% of Current Offset) 6 A 10 V 0.1 A 0.15% 20 ma 6 A 1. Set the first specified level range, limit range, and limit on the NI 4113. 2. Set the level on the NI 4113 to the first specified test point. 8 ni.com NI PXIe-4113 Calibration Procedure

3. Calculate the current through the shunt by completing the following steps. a) Take a voltage measurement across the shunt using the DMM. b) Divide the voltage measurement by the calibrated value of the shunt. c) Record the calculated value as DMM Measured Current. 4. Calculate the lower and upper current measurement test limits using the following formula: Current Measurement Test Limits = DMM Measured Current ± ( DMM Measured Current % of Current Offset) 5. Take a current measurement using the NI 4113. 6. Record the value from the previous step. 7. Verify that the recorded NI 4113 value falls within the test limits. 8. Repeat the previous seven steps for each test point specified in the level range. 9. If more than one level range is specified, repeat the previous steps using the values specified in each level range. Load Regulation Connecting and Configuring Equipment for Voltage Load Regulation Verification 1. Make the necessary connections for this procedure, as shown in the following figure. NI-DCPower Device Figure 4. Voltage Load Connection Diagram 23 Electronic Load Sense Sense DMM1 Voltage Precision Shunt Sense Sense Constant Current Remote Sense Ensure cable resistance <0.1 Ohm DMM2 Voltage 2 Configure the electronic load for remote sense mode, and ensure that cables not corrected by remote sense have a total resistance of <0.1 Ω. 3 If you make measurements using only one DMM, sequentially set up the DMM connections as specified by the procedure steps. NI PXIe-4113 Calibration Procedure National Instruments 9

2. Connect the DMM to the NI 4113 connector pins with dedicated pairs of wires to avoid including lead drop in the DMM measurement. 3. Set the nidcpower Output Function property or NIDCPOWER_OUTPUT_FUNCTION attribute to DC Voltage for the NI 4113. Verifying Voltage Load Regulation Use the NI 4113 in constant voltage mode to confirm the output voltage change falls within calculated limits while varying the load current. Run this test only after successfully verifying voltage output and measurement. Refer to the following table as you complete the following steps. Table 6. Voltage Load Regulation Verification Level Range Limit Range and Limit Test Point Load 1 Load 2 10 V 6 A 10 V 0.6 A 5.4 A 1. Set the programmable electronic load to the first specified value (Load1). 2. Set the first specified level range, limit range, and limit on the NI 4113. 3. Set the level on the NI 4113 to the first specified test point. 4. Take a voltage measurement across the output terminals of the NI 4113 using DMM1. 5. Record the voltage from the previous step as V1. 6. Calculate the current through the shunt by completing the following steps. a) Take a voltage measurement across the shunt using DMM2. b) Divide the voltage measurement by the calibrated value of the shunt. 7. Record the current from the previous step as I1. 8. Change the load from Load1 to Load2. Repeat the previous four steps. This time, record the values as V2 and I2. 9. Calculate the upper and lower voltage load regulation test limits using the following formula, and then record the test limits. Voltage Load Regulation Test Limit = ± (I1 - I2) 0.0005 V/A 10. Calculate the voltage change using the following formula, and record the value. V1 - V2. 11. Verify the voltage change falls within the voltage load regulation test limits. 12. If more than one level range is specified, repeat the previous steps using the values specified in each level range. Connecting and Configuring Equipment for Current Load Verification 1. Make the necessary connections for this procedure, as shown in the following figure. 10 ni.com NI PXIe-4113 Calibration Procedure

NI-DCPower Device Figure 5. Current Load Connection Diagram 45 Electronic Load Sense Sense DMM1 Voltage Precision Shunt Sense Sense Constant Voltage Remote Sense Ensure cable resistance <0.1 Ohm DMM2 Voltage 2. Connect the DMM to the NI 4113 connector pins with dedicated pairs of wires to avoid including lead drop in the DMM measurement. 3. Set the nidcpower Output Function property or NIDCPOWER_OUTPUT_FUNCTION attribute to DC Current for the NI 4113. Verifying Current Load Regulation Use the NI 4113 in constant current mode to confirm the output current change falls within calculated limits while varying the load voltage. Run this test only after successfully verifying current output and measurement. Refer to the following table as you complete the following steps. Table 7. Current Load Regulation Verification Level Range Limit Range and Limit Test Point Load 1 Load 2 6 A 10 V 6 A 1 V 8.5 V 1. Set the programmable electronic load to the first specified value (Load1). 2. Set the first specified level range, limit range, and limit on the NI 4113. 3. Set the level on the NI 4113 to the first specified test point. 4. Take a voltage measurement across the output terminals of the NI 4113 using DMM1. 5. Record the voltage from the previous step as V1. 4 Configure the electronic load for remote sense mode, and ensure that cables not corrected by remote sense have a total resistance of <0.1 Ω. 5 If you make measurements using only one DMM, sequentially set up the DMM connections as specified by the procedure steps. NI PXIe-4113 Calibration Procedure National Instruments 11

6. Calculate the current through the shunt by completing the following steps. a) Take a voltage measurement across the shunt using DMM2. b) Divide the voltage measurement by the calibrated value of the shunt. 7. Record the current from the previous step as I1. 8. Change the load from Load1 to Load2. Repeat the previous four steps. This time, record the values as V2 and I2. 9. Calculate the upper and lower current load regulation test limits using the following formula, and then record the test limits. Current Load Regulation Test Limit = (V1 - V22) 0.00025 A/V 10. Calculate the current change using the following formula: I1 - I2 11. Verify the current change falls within the current load regulation test limits. 12. If more than one level range is specified, repeat the previous steps using the values specified in each level range. Remote Sense Connecting and Configuring Equipment for Voltage Remote Sense Accuracy Verification 1. Make the necessary connections for this procedure, as shown in the following figure. Figure 6. Voltage Remote Sense Output Connection Diagram 6 NI-DCPower Device Load1 Sense Sense Load 2 DMM2 Voltage DMM Load Voltage Measurements Load3 DMM1 Voltage DMM LO Lead Drop Measurements 2. Set the nidcpower Sense property or NIDCPOWER_ATTR_SENSE attribute to Remote. 3. Set the nidcpower Output Function property or NIDCPOWER_OUTPUT_FUNCTION attribute to DC Voltage for the NI 4113. 6 If you make measurements using only one DMM, sequentially set up the DMM connections as specified by the procedure steps. 12 ni.com NI PXIe-4113 Calibration Procedure

Verifying Voltage Remote Sense Accuracy Use the NI 4113 in constant voltage mode with a test circuit to simulate the voltage drop between the device and a load. Complete this procedure only after successfully completing all previous Verification procedures. Refer to the following table as you complete the following steps. Table 8. Voltage Remote Sense Output Verification Level Limit Test Load 1 Load 2 Load 3 Voltage Minimum Range Range and Point Remote Lead Drop Limit Sense Test Limit 10 V 6 A 10 V 50 Ω 1 kω 50 Ω 0.1% 15 mv 0.4 V 1. Set the first specified level range, limit range, and limit on the NI 4113. 2. Set the level on the NI 4113 to the first specified test point. 3. Measure the LO lead drop with DMM1 from the negative terminal of the device to the negative side of Load2. Record the measurement as Lead Drop. 4. Verify the Lead Drop measurement is greater than or equal to the minimum lead drop. 5. Measure the load voltage with DMM2 across Load2 where the sense leads connect. Record the measurement as Load Voltage. 6. Verify the Load Voltage measurement falls within the voltage remote sense test limits. 7. If more than one level range is specified, repeat the previous steps using the values specified in each level range. Adjustment This section describes the steps needed to adjust the NI 4113 to meet published specifications. Adjusted Specifications Adjustment corrects the following specifications for the device: Voltage programming accuracy Current programming accuracy Voltage measurement accuracy Current measurement accuracy NI PXIe-4113 Calibration Procedure National Instruments 13

Following the adjustment procedure automatically updates the calibration date and temperature on the device. Note You do not need to separately adjust both measurement and output. The architecture of the NI 4113 ensures that if measurement is accurate, then output is as well, and vice versa. Initiating the Adjustment Session Initiate an external calibration session (a special type of NI-DCPower session) by calling the nidcpower Initialize External Calibration VI or nidcpower_initextcal function. (Optional) You can close the session and commit the new constants to hardware after you complete each adjustment procedure. Related Information Closing the Adjustment Session on page 17 Connecting and Configuring Equipment for Voltage Adjustment 1. Make the necessary connections for this procedure, as shown in the following figure. Figure 7. Voltage Verification or Adjustment Connection Diagram NI-DCPower Device Sense Sense DMM Voltage 2. Set the nidcpower Output Function property or NIDCPOWER_OUTPUT_FUNCTION attribute to DC Voltage for the NI 4113. Adjusting Voltage Output and Measurement Compare a set of voltages measured by the external DMM to both a set of voltage test points requested by the NI 4113 and to the measured voltages reported by the NI 4113. Refer to the following table as you complete the following steps. 14 ni.com NI PXIe-4113 Calibration Procedure

Table 9. Voltage Output and Measurement Adjustment Level Range Limit Range and Limit Test Point 10 V 6 A 0.1 V 1. Set the first specified level range, limit range, and limit on the NI 4113. 2. Set the level on the NI 4113 to the first specified test point. 10 V 3. Take a voltage measurement using the DMM, and take a voltage measurement using the NI 4113. 4. Store the values from the previous step as inputs for the nidcpower Cal Adjust VI or function called in the following steps. 5. Repeat the previous three steps for each test point specified in the level range. 6. Update the measurement calibration constants by configuring and calling the nidcpower Cal Adjust Voltage Measurement VI or nidcpower_caladjustvoltagemeasurement function. a) Input the DMM measurements as the measured outputs. b) Input the NI 4113 measurements as the reported outputs. c) Input the specified level range as the range. 7. Update the output calibration constants by configuring and calling the nidcpower Cal Adjust Voltage Level VI or nidcpower_caladjustvoltagelevel function. a) Input the DMM measurements as the measured outputs. b) Input the test points as the requested outputs. c) Input the specified level range as the range. 8. If more than one level range is specified, repeat the previous steps using the values specified in each level range. Related Information For information about VIs or functions, refer to the programming reference of the NI DC Power Supplies and SMUs Help. Connecting and Configuring Equipment for Current Adjustment 1. Make the necessary connections for this procedure, as shown in the following figure. NI PXIe-4113 Calibration Procedure National Instruments 15

Figure 8. Current Verification or Adjustment Connection Diagram NI-DCPower Device Sense Sense Precision Shunt DMM Voltage 2. Set the nidcpower Output Function property or NIDCPOWER_OUTPUT_FUNCTION attribute to DC Current for the NI 4113. Adjusting Current Output and Measurement Compare a set of currents measured by the external DMM to both a set of current test points requested by the NI 4113 and to the measured currents reported by the NI 4113. Refer to the following table as you complete the following steps. Table 10. Current Output and Measurement Adjustment Level Range Limit Range and Limit Test Point 6 A 10 V 0.1 A 1. Set the first specified level range, limit range, and limit on the NI 4113. 2. Set the level on the NI 4113 to the first specified test point. 3. Calculate the current through the shunt by completing the following steps. a) Take a voltage measurement across the shunt using the DMM. b) Divide the voltage measurement by the calibrated value of the shunt. 4. Store the value from the previous step as input for the nidcpower Cal Adjust VI or function called in the following steps. 5. Take a current measurement using the NI 4113. 6. Store the value from the previous step as input for the nidcpower Cal Adjust VI or function called in the following steps. 7. Repeat the previous five steps for each test point specified in the level range. 8. Update the measurement calibration constants by configuring and calling the nidcpower Cal Adjust Current Measurement VI or nidcpower_caladjustcurrentmeasurement function. a) Input the DMM measurements as the measured outputs. 6 A 16 ni.com NI PXIe-4113 Calibration Procedure

b) Input the NI 4113 measurements as the reported outputs. c) Input the specified level range as the range. 9. Update the output calibration constants by configuring and calling the nidcpower Cal Adjust Current Limit VI or nidcpower_caladjustcurrentlimit function. a) Input the calculated shunt current measurements as the measured outputs. b) Input the test points as the requested outputs. c) Input the specified level range as the range. 10. If more than one level range is specified, repeat the previous steps using the values specified in each level range. Related Information For information about VIs or functions, refer to the programming reference of the NI DC Power Supplies and SMUs Help. Closing the Adjustment Session Close the session and commit the new constants to hardware by calling the nidcpower Close External Calibration VI or nidcpower_closeextcal function and specifying Commit as the calibration close action. Reverification Repeat the Verification section to determine the as-left status of the device. Note If any test fails reverification after performing an adjustment, verify that you have met the Test Conditions before returning your device to NI. Refer to the Worldwide Support and Services section for information about support resources or service requests. Related Information Test Conditions on page 3 Worldwide Support and Services The NI website is your complete resource for technical support. At ni.com/support, you have access to everything from troubleshooting and application development self-help resources to email and phone assistance from NI Application Engineers. Visit ni.com/services for NI Factory Installation Services, repairs, extended warranty, and other services. Visit ni.com/register to register your NI product. Product registration facilitates technical support and ensures that you receive important information updates from NI. A Declaration of Conformity (DoC) is our claim of compliance with the Council of the European Communities using the manufacturer s declaration of conformity. This system affords the user protection for electromagnetic compatibility (EMC) and product safety. You NI PXIe-4113 Calibration Procedure National Instruments 17

can obtain the DoC for your product by visiting ni.com/certification. If your product supports calibration, you can obtain the calibration certificate for your product at ni.com/calibration. NI corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas, 78759-3504. NI also has offices located around the world. For telephone support in the United States, create your service request at ni.com/support or dial 1 866 ASK MYNI (275 6964). For telephone support outside the United States, visit the Worldwide Offices section of ni.com/ niglobal to access the branch office websites, which provide up-to-date contact information, support phone numbers, email addresses, and current events. Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on NI trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/ legal/export-compliance for the NI global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015. 2014 2015 National Instruments. All rights reserved. 374458B-01 Dec15