Reconfigurable Arrays for Portable Ultrasound

Similar documents
Linear arrays used in ultrasonic evaluation

COMPUTER PHANTOMS FOR SIMULATING ULTRASOUND B-MODE AND CFM IMAGES

Ultrasound Bioinstrumentation. Topic 2 (lecture 3) Beamforming

Ultrasound Miniaturization

DYNAMIC ANALYSIS OF CMUTs IN DIFFERENT REGIMES OF OPERATION

An Overview Algorithm to Minimise Side Lobes for 2D Circular Phased Array

Ihor TROTS, Andrzej NOWICKI, Marcin LEWANDOWSKI

Transmitting Performance Evaluation of ASICs for CMUT-Based Portable Ultrasound Scanners

Ultrasound Beamforming and Image Formation. Jeremy J. Dahl

Capacitive Micromachined Ultrasonic Transducers (CMUTs) for Photoacoustic Imaging

FPGA-BASED CONTROL SYSTEM OF AN ULTRASONIC PHASED ARRAY

3-D Imaging using Row--Column-Addressed 2-D Arrays with a Diverging Lens

Integrated Reconfigurable High-Voltage Transmitting Circuit for CMUTs

Simulation of advanced ultrasound systems using Field II

Further development of synthetic aperture real-time 3D scanning with a rotating phased array

A Multichannel Pipeline Analog-to-Digital Converter for an Integrated 3-D Ultrasound Imaging System

A High-frequency Transimpedance Amplifier for CMOS Integrated 2D CMUT Array towards 3D Ultrasound Imaging

A Delta-Sigma beamformer with integrated apodization

3-D Imaging using Row Column-Addressed 2-D Arrays with a Diverging Lens: Phantom Study

Exercise 2: Simulation of ultrasound field using Field II

18th World Conference on Nondestructive Testing, April 2012, Durban, South Africa. Joanna X.Qiao 1, Matthias Jobst 2

FPGA-Based Control System of an Ultrasonic Phased Array Keywords: ultrasonic imaging, phased array, B-scan, FPGA

Operation and Sound Field of an Ultrasonic Biplane-Array

Investigation of PDMS as coating on CMUTs for imaging

Virtual ultrasound sources

Broadband Constant Beamwidth Beamforming MEMS Acoustical Sensors

High-frequency CMUT arrays for high-resolution medical imaging

INSTRUMENTATION DESIGN FOR ULTRASONIC IMAGING

3D synthetic aperture imaging using a virtual source element in the elevation plane

Broadband All-Optical Ultrasound Transducer

Topics 8/2/2017. Opportunities for ultrasonic imaging with software beamformation

Nuove tecnologie per ecografia ad ultrasuoni: da 2D a 4D

INSPECTION OF THERMAL BARRIERS OF PRIMARY PUMPS WITH PHASED ARRAY PROBE AND PIEZOCOMPOSITE TECHNOLOGY

AN EMAT ARRAY FOR THE RAPID INSPECTION OF LARGE STRUCTURES USING GUIDED WAVES. Paul Wilcox 1, Mike Lowe 2

Lesson 06: Pulse-echo Imaging and Display Modes. These lessons contain 26 slides plus 15 multiple-choice questions.

Multi-Element Synthetic Transmit Aperture Method in Medical Ultrasound Imaging Ihor Trots, Yuriy Tasinkevych, Andrzej Nowicki and Marcin Lewandowski

Lamb Wave Ultrasonic Stylus

SIMULATION OF B-SCAN IMAGES FROM TWO-DIMENSIONAL TRANSDUCER ARRAYS: PART II - COMPARISONS BETWEEN LINEAR AND TWO-DIMENSIONALPHASED ARRAYS

ECHO-CANCELLATION IN A SINGLE-TRANSDUCER ULTRASONIC IMAGING SYSTEM

ULTRASONIC IMAGING of COPPER MATERIAL USING HARMONIC COMPONENTS

VLSI Architecture for Ultrasound Array Signal Processor

Two-Dimensional Capacitive Micromachined Ultrasonic Transducer (CMUT) Arrays for a Miniature Integrated Volumetric Ultrasonic Imaging System

Three-dimensional investigation of buried structures with multi-transducer parametric sub-bottom profiler as part of hydrographical applications

Session: 2A NEW ULTRASOUND SYSTEMS Chair: H. Ermert University of Bochum 2A-1 10:30 a.m.

Parametric Beamformer for Synthetic Aperture Ultrasound Imaging

Comprehensive Ultrasound Research Platform

A Real-time Photoacoustic Imaging System with High Density Integrated Circuit

Medical Imaging (EL582/BE620/GA4426)

Development and Application of 500MSPS Digitizer for High Resolution Ultrasonic Measurements

Ultrasound physical principles in today s technology

Standard Guide for Evaluating Performance Characteristics of Phased-Array Ultrasonic Testing Instruments and Systems 1

A Modified Synthetic Aperture Focussing Technique Utilising the Spatial Impulse Response of the Ultrasound Transducer

A hand-held row-column addressed CMUT probe with integrated electronics for volumetric imaging

A TECHNIQUE TO EVALUATE THE IMPACT OF FLEX CABLE PHASE INSTABILITY ON mm-wave PLANAR NEAR-FIELD MEASUREMENT ACCURACIES

Ultrasound Nondestructive Evaluation (NDE) Imaging with Transducer Arrays and Adaptive Processing

! # % & () () +, & ). ) /) : 7 7

3-D Vector Flow Using a Row-Column Addressed CMUT Array

A COST-EFFECTIVE METHOD FOR ULTRASOUND VOLUMETRIC IMAGING

Ultrasonic Linear Array Medical Imaging System

30 th April 2008 Glasgow DSpace Service

Integrated 16-channel Transmit and Receive Beamforming ASIC for Ultrasound Imaging

Acoustical cross-talk in row column addressed 2-D transducer arrays for ultrasound imaging

12/26/2017. Alberto Ardon M.D.

REAL-TIME B-SCAN ULTRASONIC IMAGING USING A DIGITAL PHASED. Robert Dunki-Jacobs and Lewis Thomas General Electric Company Schenectady, New York, 12301

Piezoceramic Ultrasound Transducer Enabling Broadband Transmission for 3D Scene Analysis in Air

Random Thinning of Segmented Annular Arrrays

DACON INSPECTION SERVICES. Phased Array Ultrasonic Testing

Mathematical Modeling of Ultrasonic Phased Array for Obstacle Location for Visually Impaired

A miniature all-optical photoacoustic imaging probe

A SHEAR WAVE TRANSDUCER ARRAY FOR REAL-TIME IMAGING. R.L. Baer and G.S. Kino. Edward L. Ginzton Laboratory Stanford University Stanford, CA 94305

Ultrasound Physics. History: Ultrasound 2/13/2019. Ultrasound

BROADBAND CAPACITIVE MICROMACHINED ULTRASONIC TRANSDUCERS RANGING

RECENT ADVANCEMENTS IN THE APPLICATION OF EMATS TO NDE

IMAGING OF DEFECTS IN CONCRETE COMPONENTS WITH NON-CONTACT ULTRASONIC TESTING W. Hillger, DLR and Ing. Büro Dr. Hillger, Braunschweig, Germany

Extending Acoustic Microscopy for Comprehensive Failure Analysis Applications

Zero-Bias Resonant Sensor with an Oxide-Nitride Layer as Charge Trap

Architecture of Quality Imaging Mary K. Henne, MS, CNMT, RDMS, RVT Ultrasound Education Specialist GE Healthcare

Synthetic-aperture based photoacoustic re-beamforming (SPARE) approach using beamformed ultrasound data

Simulation of Algorithms for Pulse Timing in FPGAs

Generation Laser Scanning Method for Visualizing Ultrasonic Waves Propagating on a 3-D Object

DESIGN & VALIDATION OF A SEMI-FLEXIBLE PAUT PROBE FOR THE MANUFACTURING INSPECTIONS OF LARGE FORGED ROTORS

TESTING OF HEXAGONAL AND OTHER EDGE PROFILED MATERIALS BY USING ULTRASONIC PHASED ARRAY PROBES ARRANGED ANNULARLY

Sub-millimeter Wave Planar Near-field Antenna Testing

CHAPTER 1 INTRODUCTION

Advanced Ultrasonic Imaging for Automotive Spot Weld Quality Testing

A Comparison Between Conventional and Collapse-Mode Capacitive Micromachined Ultrasonic Transducers in 10-MHz 1-D Arrays

Evaluation of Transducers with Near-field Scanning of Their Sl~rfaces

Beamforming in ultrasound

MIL-STD-883H METHOD ULTRASONIC INSPECTION OF DIE ATTACH

Advancing MEMS R&D in materials, processes and devices to face major needs arising from the booming MEMS market

Tritech International Vehicle Sonar Developments

CIRCULAR DUAL-POLARISED WIDEBAND ARRAYS FOR DIRECTION FINDING

CMUT and PMUT: New Technology Platform for Medical Ultrasound Rob van Schaijk

Feasibility Study for ARL Inspection of Ceramic Plates Final Report - Revision: B

Easy Ultrasonic Phased Array Inspection of Corrosion - Resistant Alloys and Dissimilar Weld Materials

Ultrasound Brain Imaging System

DESIGN AND MANUFACTURE OF A HIGH-FREQUENCY ANNULAR ARRAY ULTRASOUND SYSTEM FOR MEDICAL IMAGING

Standard Guide for Evaluating Characteristics of Ultrasonic Search Units 1

Non-Ideal Quiet Zone Effects on Compact Range Measurements

Transcription:

Reconfigurable Arrays for Portable Ultrasound R. Fisher, K. Thomenius, R. Wodnicki, R. Thomas, S. Cogan, C. Hazard, W. Lee, D. Mills GE Global Research Niskayuna, NY-USA fisher@crd.ge.com B. Khuri-Yakub, A. Ergun, G. Yaralioglu E. L. Ginzton Laboratory Stanford University Stanford, CA-USA khuri-yakub@stanford.edu Abstract A collaborative effort is aimed at the development of reconfigurable array technology. The goal is to enable innovative medical ultrasound imagers ideally suited for portable ultrasound and applications such as remote emergency medicine and combat casualty care. Success depends on developing several technologies, the first of which is capacitive micromachined ultrasound transducers (cmuts). The monolithic nature of cmuts facilitates close connection with microelectronics. Thus a second technology under development is a switch matrix application specific integrated circuit (ASIC) that will enable the changing of interconnect between cmut cells. The reconfigurable array concept arises from this ability to dynamically combine cmut cells to form ideal apertures for a given imaging target (e.g. annular and phased apertures of various ring widths) and to move these apertures across the reconfigurable array plane [1-4]. Two central hypotheses are being tested: (1) the reconfigurable array can acquire acoustic pulse-echo data in a manner equivalent or superior to today s 1D piezoceramic arrays (2) reconfigurable array technology will enable highly portable ultrasound platforms. Keywords-reconfigurable; annular; array; cmut; ASIC; switch matrix; dynamic; phased; real-time I. INTRODUCTION Identification of internal bleeding in emergency patients or combat casualties is critical within the first hour after injury. To achieve speed in bleed localization a highly portable, high image quality, ultrasound system is desired. Laptop sized scanners are now readily available in the marketplace and promoted for niche applications especially where portability is important, such as general practice physician offices or remote area services, emergency rooms, and even ambulance care. But these devices generally have compromised beamformation because of choices for reduced channel count, use of synthetic aperture methods, lower frame rates, and lower number of A/D converter bits for the digitization process in order to reduce system size. It is of further concern that the clinical user base for these portable systems is less experienced than radiologists or sonographers in full-time practice, particularly since there is an art to achieving clinical utility in situations where image data acquisition is suboptimal. The proposed solution is to develop a reconfigurable array system that provides high-end image quality despite significantly reduced channel count. In the usual ultrasound imager design, each array element in the active aperture is associated with a system beamforming channel with its A/D converter and beamforming electronics. The reconfigurability concept in which array elements are formed by electronically combining smaller transducer subelements (see Fig. 1) removes this beamforming redundancy. cmuts are attractive for implementation of a reconfigurable array because they are formed using microelectronic processes and thus very apt to integration with front end electronics such as switching, pulser, and pre-amp circuitry. In addition, the high fractional bandwidth of cmuts can be used to achieve image quality improvements (spatial resolution and harmonic imaging). An appealing approach to reconfiguring such an array is to form an annular aperture. The acoustic fields generated by annular arrays are considered to be superior largely because of the symmetry. Thus axisymmetric dynamic focusing can be achieved without system changes. As a result, annular arrays have been considered to provide better image quality, albeit at the disadvantage of requiring a mechanical mover to acquire 2D image data. With the reconfigurable array, electronically scanned annular arrays can be attained for the first time. Figure 1. Reconfigurable Array Concept: Cells are hardwired together to form larger acoustic subelements. Thousands of subelements are connected or disconnected to each other and to system channels via an underlying switch matrix thus enabling electronically changing apertures. Some effort supported by US Army Medical Research Acquisition Activity DAMD17-2-181, 82 Chandler Street, Fort Detrick, MD 2172-514. The content of the information does not necessarily reflect the position or the policy of the Government and no official endorsement should be inferred. Some effort described was supported by Grant Number R1 EB2485 from NIBIB. Contents of this publication are solely the responsibility of the authors and do not necessarily represent the official views of NIH. -783-9383-X/5/$2. (c) 25 IEEE 495 25 IEEE Ultrasonics Symposium

II. RECONFIGURABLE ARRAY TECHNOLOGIES The first steps to prove that reconfigurable arrays can provide high-end image quality with significantly reduced channel count have been taken by demonstrating key technologies. The first was the development of a 2-ring static annular cmut array with similar performance to a conventional piezoceramic linear array. The second was the fabrication of a switch matrix ASIC that interconnects a 16 by 16 array of transducer elements to reconfigure the acoustic pattern in real-time with only 2 channels. A. cmut Annular Array In order to demonstrate that this new structure could perform in at least an equivalent fashion to a conventional array, several simulations have been run. The graphs of Fig. 2 show lateral beam profiles for different subelement sizes and how they compare with a linear array. For cases shown, the annular array beamshape is nearly equivalent to a linear array in the imaging plane. As expected due to the circular symmetry, the annular array is far superior in the elevation plane. An important factor in the quality of an ultrasound image is the sidelobe performance of the array. This aspect of the array design was tested against key array design parameters such as subelement size. Fig. 3 results show satisfactory performance for the 2um and 25um sizes, although some compromise is evident. Further results from these initial simulation studies are given in [2]. -2-4 -6 A) F=1mm -8 1 2 3 4 5-2 -4-6 B) F=4mm -8 1 2 3 4 5 3um 25um 2um Linear Lin Elev Figure 2. Effect of sub-element size on main beam for 2um, 25um, 3um sub-element sizes. Transmit focal points at 1 and 4 mm shown. Linear array radiation patterns included for comparison. -5 A) F=1mm -1 2 4 6-2 -4-6 -8 B) F=4mm -1 2 4 6 3um 25um 2um Linear Lin Elev Figure 3. Effect of subelement size on secondary lobes for 2um, 25um, 3um sub-element sizes. Transmit focal points at 1 and 4 mm shown. Linear array radiation patterns included for comparison. A new tool that produces increasingly realistic computer simulations of the acoustic fields generated by reconfigurable arrays has been developed. The main focus in this work has been the careful analysis of the configuration of the array subelements into elements, the influence of electrical loading, and the impact of resulting non-ideal acoustic fields. Field II from Jorgen Jensen at the Technical University of Denmark is being used as the acoustic calculation engine [5]. Fig. 4 shows two annular designs (2 rings and 32 rings). The tool generates these apertures using input variables such as number of rings, subelement pitch, subelement kerf, aperture width, steering angle, and focal depth. Next, switch states for the switch matrix ASIC are generated in order to configure the desired annular aperture on the reconfigurable array. Incorporated into the tool is the ability to analyze the role of missing subelements and possible delay errors associated with the circuitry. The acoustic performance of an aperture can then be simulated with Field II to generate 2D beam plots that can then be used to estimate the main lobe performance and secondary lobe levels. Simulated ultrasound images can also be produced to further assess the expected image quality from a given aperture. The beam profiles of Fig. 5 were generated using these tools. The images are one-way beam profiles assuming, 15um or 2um subelement size, 4um subelement kerf, 2mm aperture width, focal depth of 5mm and a 2-cycle, 5MHz pulse. The annular array with 32 rings has better sidelobe -783-9383-X/5/$2. (c) 25 IEEE 496 25 IEEE Ultrasonics Symposium

performance, especially in the nearfield. Since the aperture width is the same for both cases, the ring width for the 32-ring case is narrower and thus the elemental granularity with respect to delays is better. Using a finer subelement pitch of 15um shows further improvement, though the results are not as dramatic, since the ring width remains the same, however each ring s borders are smoother due to the smaller subelements. As is evident, this tool is ideal for the design of the reconfigurable apertures to meet application requirements and for assessing array hardware limitations. subelements sized at (28um) 2 that was integrated with a LOGIQ 7 scanner. Fig. 6 shows the array, which has a small wedge to bring out electrical connections from each ring. The image in fig. 7 was obtained by mechanically moving the array and obtaining M-mode scans while operating the array at 6.5MHz, 3cm focal depth, and 6dB dynamic range. This image shows similar performance of the cmut annular array as compared with a conventional PZT ultrasound probe, however, with 6 times lower channel count. Figure 6. 2-ring static annular cmut array; close-up of the electrical connect from the rings that removes a small wedge of acoustic area; close up of the array showing the center button. 4 mm 2 mm Figure 4. 2-ring and 32-ring apertures were generated by the reconfigurable array simulation tool assuming a 2mm aperture width, and subelement pitch of 2um, and equal width rings (~78 subelements) Based on the modeling results, array geometry such as number of rings, size of subelement, and width of rings were specified for a given application. Stanford then designed and fabricated a 2-ring static cmut annular array with Figure 7. This image was taken with the 2-ring static annular cmut. As can be seen, even the 2 mm spheres are clearly visible; this is only seen with conventional arrays at the elevation focus. B. Switch Matrix Electronics The first switch matrix ASIC contains a 16 by 16 array of unit cells with a total of 1,24 switches and their associated logic in a 5mm by 6.4mm chip. Figure 8 shows an 8 by 8 example of the core of the layout. Each row of cells shares high voltage analog signal, while each column of cells shares a a) b) c) Figure 5. One way beam profiles for 2 mm aperture width and a) 2 rings and 2 um subelement pitch, b) 32 rings and 2 um subelement pitch, c) 32 rings and 15 um subelement pitch -783-9383-X/5/$2. (c) 25 IEEE 497 25 IEEE Ultrasonics Symposium

digital control bus. Each cell also contains a signal pad for probing of the center point of the 4 switches and for connection to a transducer subelement. Digital circuitry is used to control the state of the switches. Power bus are routed down into the array from the pad ring at the chip edge. The analog data from these cells is read out along the analog data. 1 Column 1 Cell Digital data Matrix switch Signal pad Access switch Power Analog data Figure 8. Each cell of the switch matrix ASIC contains 3 matrix switches for connection between transducer subelements, 1 access switch for connection to a system channel, digital logic to set switch states, and a signal pad which will be connected to a transducer subelement. Individual cells are programmed and selected using digital data. This digital data is programmed into the array prior to transmit to reconfigure the switch matrix. There are three matrix switches that disconnect or connect one transducer subelement to another and one access switch to connect a given subelement to a system channel. Thus by setting the switches in a specific pattern one can electronically configure elements and apertures. and the measurement process repeated generating the set of measurements shown in fig. 11. This is a key demonstration of the ability to reconfigure an array on the fly. In the final version, the intent is to be able to change the aperture characteristics such as the number of annular array elements, aperture diameters, and aperture position on the 2D surface of the reconfigurable array. This enables acquisition of image data from a 3D volume immediately below the surface of the array. Figure 1. PZT 256 subelement transducer integrated on top of the switch matrix ASIC such that the subelements can be electronically interconnected to III. PROOF OF RECONFIGURABILITY CONCEPT The first prototype of a switch matrix ASIC has been built, integrated with a piezoceramic array, and tested for its ability to translate a 2D aperture on the array surface. Fig. 9 shows the switch matrix ASIC with its 16 by 16 matrix of switch cells and 1,24 total switches. Fig. 1 shows its integration with a 256-subelement piezoceramic array. form desired apertures Figure 11. Set of acoustic measurements showing an aperture walking from left to right on the 256 subelement PZT reconfigurable array Figure 9. A switch matrix ASIC with a 16 by 16 array of electronic subelements containing 1,24 switches To test out the functionality of the switch matrix/transducer assembly, the integrated device was placed in a water tank, an active aperture was formed by selection of a pattern of switches, and the resulting acoustic field was measured by a hydrophone. Next the pattern was translated to a new location IV. SUMMARY The key accomplishments achieved so far include investigating the behavior of reconfigurable arrays and using the information to design, build, and test a static annular array made out of cmut cells; designing a switch matrix ASIC to enable reconfiguring apertures and integrating the ASIC onto a 16 by 16 subelement PZT array; and finally integrating the reconfigurable array on an ultrasound system and demonstrating translation of an active aperture. The team is currently in the process of developing more complex assemblies of both the cmut array and the switch matrix and expects to image with these later in the year. In conclusion, good image quality using only 2 annular elements that is comparable to that of high-end transducers with 128 linear elements has been demonstrated. The result is intended to be a key step toward establishing the feasibility of highly portable systems with superior image quality. -783-9383-X/5/$2. (c) 25 IEEE 498 25 IEEE Ultrasonics Symposium

REFERENCES [1] K. Thomenius, R. Fisher, D. Mills, R. Wodnicki, C. Hazard, and L. Smith, Mosaic arrays using micromachined transducers, US Patent #6,865,14, March 8, 25. [2] C. Hazard, R. Fisher, D. Mills, L. Smith, K. Thomenius, and R.Wodnicki, Annular array beamforming for 2D arrays with reduced system channels, 23 IEEE ultrasonics symposium, October 23, vol. 2, pp. 1859-1862. [3] D. Bailey, J. Meyyappan, and G. Wade, A computer controlled transducer for real-time three-dimensional imaging, Acoustical Imaging, vol. 18, pp. 543-552. [4] R. Bele, Echography probe and apparatus incorporating such a probe, US Patent #4,641,66, February 1, 1987. [5] J.A. Jensen and N. B. Svendsen: Calculation of pressure fields from arbitrarily shaped, apodized, and excited ultrasound transducers, IEEE Trans. Ultrason., Ferroelec., Freq. Contr., 39, pp. 262-267, 1992. -783-9383-X/5/$2. (c) 25 IEEE 499 25 IEEE Ultrasonics Symposium