Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT

Similar documents
Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM

Robert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center. Thursday September 8, 2011

GS Introduction to Medical Physics IV Laboratory 5 Gamma Camera Characteristics

NM Module Section 2 6 th Edition Christian, Ch. 3

Parameters Affecting on Intrinsic Uniformity Test For MEDISO

CHAPTER 8 GENERIC PERFORMANCE MEASURES

Acceptance Testing and Annual Physics Survey Recommendations for Gamma Camera, SPECT, and SPECT/CT Systems

Initial Certification

Intrinsic and Tomographic Evaluation of Siemens e.cam SPECT System at the Korle-Bu Teaching Hospital (Ghana)

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

International Journal of Scientific & Engineering Research, Volume 4, Issue 9, September ISSN

Quality control in dual head γ- cameras: Comparison between methods and softwares used for image analysis

Investigation of Multiple Head Registration / Center of Rotation for SPECT Gamma Cameras

QUALITY CONTROL OF IMAGING DEVICES. P.S. Soni

CHAPTER 15 DEVICES FOR EVALUATING IMAGING SYSTEMS

Nuclear Associates , &

A new operative gamma camera for Sentinel Lymph Node procedure

How Gamma Camera s Head-Tilts Affect Image Quality of a Nuclear Scintigram?

Photomultiplier Tube

ACR Update in Nuclear Medicine Accreditation

QC Testing for Computed Tomography (CT) Scanner

J of Nuclear Medicine Technology, first published online July 27, 2011 as doi: /jnmt

Factors Affecting the resolution of SPECT Imaging. h.

T h e P h a n t o m L a b o r a t o r y

Y11-DR Digital Radiography (DR) Image Quality

CZT Technology: Fundamentals and Applications

Evaluation of XRI-UNO CdTe detector for nuclear medical imaging

2/14/2019. Nuclear Medicine Artifacts. Symmetric energy windows

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

Results of the Measurement of the Collimator Hole Angulation for Different Collimators of SPECT with Adaptive Quality Control Phantom

Nuclear Associates

PHYSICS ADVANCED LABORATORY I COMPTON SCATTERING Spring 2002

NON-UNIFORM ATTENUATION CORRECTION USING SIMULTANEOUS TRANSMISSION AND EMISSION CONVERGING TOMOGRAPHY

CyberKnife Iris Beam QA using Fluence Divergence

SPECT Reconstruction & Filtering

Digital Radiography : Flat Panel

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

Performance evaluation of two CZT gamma ray imaging systems

Chiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard.

... In vivo imaging in Nuclear Medicine. 1957: Anger camera (X;Y) X Y

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Pinhole collimator design for nuclear survey system

Deadtime correction for two multihead Anger cameras in 131 I dual-energywindow-acquisition

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

Primer on molecular imaging technology

Test Equipment for Radiology and CT Quality Control Contents

Ergo TM Imaging System

Application of the HICAM Camera for Imaging of Prompt Gamma Rays in Measurements of Proton Beam Range

Nuclear medicine is critically dependent on the accurate,

Features and Weaknesses of Phantoms for CR/DR System Testing

Contrast Evaluation for a Dual Head SPECT System with Different Energy Peaking- Drift

Exprerimental Evaluation of a Dedicated Pinhole SPECT System for Small Animal Imaging and Scintimammography

Electronic Instrumentation for Radiation Detection Systems

Appendix I: Artefacts and Trouble-shooting

A high energy gamma camera using a multiple hole collimator

3D Diode Array Commissioning: Building Confidence in 3D QA Technology

10/3/2012. Study Harder

Master of Science Thesis. SIMIND Based Pinhole Imaging

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline

10/26/2015. Study Harder

Reconstruction Filtering in Industrial gamma-ray CT Application

Industry Breakthrough

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

AUTOMATED AND QUANTITATIVE METHOD FOR QUALITY ASSURANCE OF DIGITAL RADIOGRAPHY IMAGING SYSTEMS

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA

QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING

First Applications of the YAPPET Small Animal Scanner

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

Ph 3324 The Scintillation Detector and Gamma Ray Spectroscopy

K-edge subtraction X-ray imaging with a pixellated spectroscopic detector

IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY

A Fast Monolithic System for Proton Imaging. Fritz DeJongh ProtonVDA Inc October 2017

Acceptance Testing of a Digital Breast Tomosynthesis Unit

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist

Nuclear Associates , , CT Head and Body Dose Phantom

Design Optimization of a Small-animal SPECT System Using LGSO Continuous Crystals and Micro Parallel-hole Collimators

Industry Breakthrough

Amorphous Selenium Direct Radiography for Industrial Imaging

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC

Hardware for High Energy Applications 30 October 2009

Maximizing clinical outcomes

Backgrounds in DMTPC. Thomas Caldwell. Massachusetts Institute of Technology DMTPC Collaboration

LaBr 3 :Ce, the latest crystal for nuclear medicine

DALLA LUCE VISIBILE AI RAGGI X: NUOVI RIVELATORI DI IMMAGINI PER RAGGI X A DISCRIMINAZIONE IN ENERGIA ED APPLICAZIONI

ipix Gamma Imager Product Introduction Steve Laskos Product Management Director

SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES.

Prompt-Gamma Based Range Verification in Particle Therapy: New prospects (also) for 4D?

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

Development of an innovative LSO-SiPM detector module for high-performance Positron Emission Tomography

Changing the Shape of Nuclear Medicine

BASICS OF FLUOROSCOPY

PET/CT Instrumentation Basics

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY

IBEX MATERIALS DETECTION TECHNOLOGY

FRCR Nuclear Medicine

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Design Optimization of a Small-animal SPECT System Using LGSO Continuous. Crystal and a Micro Parallel-hole Collimator

PET Detectors. William W. Moses Lawrence Berkeley National Laboratory March 26, 2002

Transcription:

Quality control of Gamma Camera By Dr/ Ibrahim Elsayed Saad 242 NMT

WHAT IS QUALITY? The quality of a practice is to fulfill the expectations and demands from: Patient Clinicain Your self

Quality assurance and quality control The concept of quality in the term quality assurance expresses the closeness with which the outcome of a given procedure approaches some ideal, free from all errors and artifacts (Whole procedure). The term quality control is used in reference to the specific measures taken to ensure that one particular aspect of the procedure is satisfactory (Single step in the procedure).

NUCLEAR MEDICINE SERVICE facilities patient care waiting time staff reporting competence experience optimisation radiopharmaceuticals methods examination technique instrumentation etc etc Primary service Secondary service Nuclear medicine examination or treatment

QA-PROGRAMME OBJECTIVES * Improvement in the quality of the diagnostic information. * Use of minimum amount of radionuclide activity to ensure the production of the desired diagnostic information. * Effective use of available resources

QA Medical exposure Choice of examination Determination of technical parameters Optimization of administered activity Methods of reducing the absorbed dose Quality control of equipment and radiopharmaceutical Quality assurance of methods Safe routines to avoid misadministration

Gamma Camera Used to measure the spatial and temporal distribution of a radiopharmaceutical

Test Schedule for Gamma Camera System Test Frequency in routine testing Acceptance Referenc e once per Quarter Week Half year 1-Physical Inspection A 2-Test of Intrinsic Flood-field Uniformity 3- Test of Intrinsic Spatial Resolution 4- Test of Intrinsic Count-rate Performance 5- Test of Spatial Linearity and Spatial Resolution A R Q W A R H A R H A 6- Test of Basic Computer Timing A R H 7- Test of Computer Timing in Dynamic Acquisition R A R H 8- Test of camera Sensitivity A R W

SOURCES FOR QC OF GAMMA- CAMERAS <1 mm Point source Collimated line source Line source Flood source Tc99m, Co57, Ga67

Phantoms for QC of gamma-cameras Bar phantom Slit phantom Orthogonal hole phantom Total performance phantom

Phantoms for QC of gamma-cameras

Physical Inspection The Purpose of this test is to inspect a cameracomputer system for shipping damage and production flaws. Procedure: 1-Image Display Devices 2-Image Recording Devices 3-Electrical Connections, Fuses and Cables 4-Operation and Service Manuals 5-Check for controls that are difficult to operate or are noisy and switches that do not throw securely. Inspect keyboards for damage.

Special definitions for quality control System: a term used to refer to the performance of the camera detector, as it would be used in a clinical environment including components such as a collimator and supporting gantry. Energy window: A range of gamma and X-ray energies, which are to be accepted and processed. The window is expressed as a range of energies or as a percentage of expected peak energy. When expressed as percentage the peak energy must always be specified, and the window is always symmetrical about the peak energy value. Photopeak: the characteristic energy distribution resulting from the collection of total photon energy absorbed by the detector.

Differential uniformity: The amount of count density change per defined unit distance when the detector s incident gamma radiation is a homogenous flux over the field of measurement. Integral uniformity: A measure of the maximum count density variation over a defined large area of the scintillation detector for a uniform input gamma flux to the Useful Field of View of the camera. Pixel: a picture element used to store a value in a digital memory. It represents an area defined by dimensions in X and Y directions at a known position defined by X and Y coordinates.

Scatter: photons that have lost part of their energy due to interaction with a medium, such as water, plastic, or tissue. Sensitivity: The observed count rate per unit of radioactivity at a specified distance. Spectrum: a plot of the number of detected gamma rays versus the measured energy of the gamma rays.

Energy resolution: A term used to characterize the ability of a gamma camera to distinguish between photons of different energies Spatial resolution: A term that characterizes the scintillation camera s ability to accurately resolve spatially separated radioactive sources.

Full Width at Half Maximum (FWHM): The measure of the spread of a point or line spread function measured between locations 50 %down on each side from the peak amplitude. Full Width at Tenth Maximum (FWTM): The measure of the spread of a point or line response function measured between locations 90 % down on each side from the peak amplitude.

Differential linearity: The amount of positional distortion or displacement over a limited distance Useful Field of View (UFOV): The area of the detector, which is used for imaging gamma rays and x- rays. It is defined by a dimensioned figure supplied by the manufacturer. Central field of view (CFOV): The area defined by scaling all linear dimensions of the useful field of view (UFOV) by a factor of 75%.

Intrinsic: A term used to describe performance characteristics of a scintillation camera that exclude external variables, which affect these specifications such as collimators or display devices. Extrinsic: A term that is used to describe the total system performance including photons attenuated throw passing into the collimator, so that it includes all external factors such as collimator.

EnergyWindow Peaking 15 % window 25 % window 35 % window

140 KeV 137 KeV Wrong setting of energy window

1) Test of Intrinsic Flood-field Uniformity The purpose of this test is to test the combined intrinsic response of a cameracomputer system to a spatially uniform flux of incident gamma radiation over the camera field-of-view. Materials used are: Point source consisting of 10-20 MBq (0.3-0.5 mci) Tc-99m, giving a count rate not greater than 30 000-50 000 c /s with a 20% PHA window. Source mounting for point source.

Procedure 1. Remove the collimator from the detector head. Align the head and the source mounting. 2. Mount the source in the source mounting with distance equal to 5 times the average detector diameter, which is (220 cm). 3. Centre a 20% PHA window on the Photopeak (Check of Energy Calibration of PHA). 4. Acquire an image on the cameras display device at a preset count of 15 x 10 6 using a 512X512 matrix. 6. Record the collection time for the camera. 7. Remove the source and mount the collimator.

Data Analysis: Determine the maximum (Max) and minimum (Min) counts in the pixels lying within the UFOV and the CFOV. The integral uniformity, IU, is then given by: IU = 100 * Max - Min Max + Min Determine for each row or column of pixels in the X and Y directions within the UFOV and the CFOV, the maximum count difference in any 6 contiguous pixels. Determine the highest value of this maximum count difference in the sets of rows and columns. The differential uniformity, DU, is then given by: DU =100* Hi ' Low Hi + Low

2) Test of Intrinsic Spatial Resolution The purpose of this test is to test the combined intrinsic spatial resolution of a camera- computer system in terms of the full width at half-maximum (FWHM) and full width at tenth-maximum (FWTM). Procedures: 1-The collimator is removed from the detector face. 2- Position the intrinsic-resolution phantom ( slit phantom) and then adjust the source on the source holder to be in front of the detector face with distance equal to 5 times the average detector diameter, which is (220 cm). 3-The acquisition parameters for this test is adjusted as follows: a)15000 K. counts. b)512x512x16 acquisition matrix. c)20 % PHA window on the photopeak which is 140 KeV for Tc-99m.

4-Acquire an image for the camera device at the preset count of 15 x 10 6 5-By using the software provided on the processing computer connected to the camera, by which we can draw a line on a single slit of the slits in the phantom image, this will give us a profile which indicate to the peak from which we can calculate the FWHM in Pixels. And then by knowing the matrix size used while acquiring the image we can calculate the FWHM in millimeters.

6-Then the estimation of the intrinsic spatial resolutions is done in terms of FWHM of the line spread function. counts Pixels

At acceptance testing, the calculated values of FWHM in the X and Y directions should be compared with the manufacturer's worst-case values. At routine testing, the calculated values should be compared with the reference values.

3) Test of Intrinsic Spatial Linearity The same procedures for testing the intrinsic resolution is prepared. The resulted peaks for the lines separating the lead sheets is plotted on a graph and then the spatial distortion (linearity) is measured.

Test of Intrinsic Spatial Linearity

Test of Intrinsic Spatial Linearity Count pixels

Then the calculation of the distance between each adjacent peaks is done, and calculating the difference between the maximum distance and the minimum distance this will indicate the shifting between the parallel slits, this is directly the spatial linearity.