Instant super-resolution imaging in live cells and embryos via analog image processing

Similar documents
Nature Methods: doi: /nmeth Supplementary Figure 1. Schematic of 2P-ISIM AO optical setup.

Point Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005

BASICS OF CONFOCAL IMAGING (PART I)

Development of a High-speed Super-resolution Confocal Scanner

3D light microscopy techniques

Multicolor 4D Fluorescence Microscopy using Ultrathin Bessel Light sheets

Supplementary Information. Stochastic Optical Reconstruction Microscopy Imaging of Microtubule Arrays in Intact Arabidopsis thaliana Seedling Roots

Basics of confocal imaging (part I)

Technology Note ZEISS LSM 880 with Airyscan

Opterra II Multipoint Scanning Confocal Microscope. Innovation with Integrity

High resolution extended depth of field microscopy using wavefront coding

EUV microscopy - a user s perspective Dimitri Scholz EUV,

Rapid Non linear Image Scanning Microscopy, Supplementary Notes

3D light microscopy techniques

Spectral Imaging with the Opterra Multipoint Scanning Confocal

Application Note. The New 2D Superresolution Mode for ZEISS Airyscan 120 nm Lateral Resolution without Acquiring a Z-stack

Nikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON

Applications of Optics

ANSWER KEY Lab 2 (IGB): Bright Field and Fluorescence Optical Microscopy and Sectioning

1 Co Localization and Working flow with the lsm700

Microscopy from Carl Zeiss

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Practical work no. 3: Confocal Live Cell Microscopy

Multifluorescence The Crosstalk Problem and Its Solution

Travel to New Dimensions- LSM 880. The Resolution of a Microscope is limited. The Resolution of a Microscope is limited. Image. Image. Object.

High-resolution, low light-dose lightsheet microscope LATTICE LIGHTSHEET

Zeiss 780 Training Notes

Rates of excitation, emission, ISC

Practical Flatness Tech Note

Rapid Adaptive Optical Recovery of Optimal Resolution over Large Volumes

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST

Light Microscopy. Upon completion of this lecture, the student should be able to:

Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal

Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement

Leica TCS SP8 Quick Start Guide

Welcome to: LMBR Imaging Workshop. Imaging Fundamentals Mike Meade, Photometrics

Rapid three-dimensional isotropic imaging of living cells using Bessel beam plane illumination

The DCS-120 Confocal Scanning FLIM System

Confocal Microscopy. (Increasing contrast and resolu6on using op6cal sec6oning) Lecture 7. November 2017

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

FLUORESCENCE MICROSCOPY. Matyas Molnar and Dirk Pacholsky

Nature Neuroscience: doi: /nn Supplementary Figure 1. Optimized Bessel foci for in vivo volume imaging.

Nikon Instruments Europe

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy,

Things to check before start-up.

The Zeiss AiryScan System, Confocal Four.

Leica TCS SP8 Quick Start Guide

SUPPLEMENTARY INFORMATION

Shreyash Tandon M.S. III Year

Supplementary Materials

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Chapter 23 Study Questions Name: Class:

Reflecting optical system to increase signal intensity. in confocal microscopy

High-speed 1-frame ms scanning confocal microscope with a microlens and Nipkow disks

Imaging Introduction. September 24, 2010

User manual for Olympus SD-OSR spinning disk confocal microscope

Zeiss 880 Training Notes Zen 2.3

ADVANCED METHODS FOR CONFOCAL MICROSCOPY II. Jean-Yves Chatton Sept. 2006

LEICA TCS SP5 AOBS TANDEM USER MANUAL

Point Spread Function Estimation Tool, Alpha Version. A Plugin for ImageJ

We attempted to separate the two dyes by acquiring images using a single excitation wavelength and just two emission wavelengths.

3. are adherent cells (ie. cells in suspension are too far away from the coverslip)

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

Aberrations and adaptive optics for biomedical microscopes

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI)

Single-shot three-dimensional imaging of dilute atomic clouds

LSM 710 Confocal Microscope Standard Operation Protocol

Opterra. Multipoint Scanning Confocal Microscope. Innovation with Integrity. Cell-Friendly, High-Speed, High-Resolution Imaging

Confocal and 2-photon Imaging. October 15, 2010

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

SPINNING DISK CSU-X1 USER MANUAL

Imaging Retreat - UMASS Customized real-time confocal and 2-photon imaging

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

The Optics of Mirrors

Inside the LSM 880 NLO + Airyscan

Εισαγωγική στην Οπτική Απεικόνιση

Confocal, hyperspectral, spinning disk

Advanced Optical Microscopy lecture. 03. December 2012 Kai Wicker

a) How big will that physical image of the cells be your camera sensor?

Instructions for the Experiment

Flatness of Dichroic Beamsplitters Affects Focus and Image Quality

LSM 510 Meta Training Notes

Zeiss LSM 880 Protocol

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Be aware that there is no universal notation for the various quantities.

OPTICS DIVISION B. School/#: Names:

CCAM s Selection of. Zeiss Microscope Objectives

Introduction to light microscopy

More fancy SPIM, Even fancier SPIM

Final Exam, 150 points PMB 185: Techniques in Light Microscopy

ECEN 4606, UNDERGRADUATE OPTICS LAB

Swept-Field User Guide

ECEN 4606, UNDERGRADUATE OPTICS LAB

Multi-channel imaging cytometry with a single detector

Multidimensional Imaging with the Opterra Multipoint Scanning Confocal System

Transcription:

Nature Methods Instant super-resolution imaging in live cells and embryos via analog image processing Andrew G. York, Panagiotis Chandris, Damian Dalle Nogare, Jeffrey Head, Peter Wawrzusin, Robert S. Fischer, Ajay Chitnis, Hari Shroff

Supplementary Fig. 1, Instant SIM Optical Setup. 488 nm and 561 nm lasers are used to excite fluorescence from the sample. Lasers are combined with dichroic DC and passed through an AOTF (providing fast shuttering, and power control in conjunction with half-wave plates placed in front of each laser), beam expander, converging microlens array (to form a multiplicity of excitation foci, inset 1), compensator plate (to minimize astigmatism), thick dichroic TDC, scan lens telescope (Scan Lens 1 and 2) with galvanometric mirror (Galvo) placed between lenses (to scan the multifocal excitation array through the sample, thus covering the imaging field, inset 2), tube lens and objective. Multifocal fluorescence from sample is collected with the same objective and passed through tube lens, Scan lenses 1 and 2, Galvo, and reflected from TDC. In the descan path (post Galvo), fluorescence is stationary, so a pinhole array placed at the focal plane of Scan lens 1 serves to reject out of focus fluorescence (inset 3). Relaying the emission to a second microlens array is achieved by relay lenses; the second microlens array serves to locally contract each focus 2x, while maintaining the foci orientation (inset 4 and Supplementary Fig. 3). The locally scaled foci are rescanned and imaged onto a scmos camera via Scan lens 3, the Galvo, and Scan lens 4; excitation light is removed by emission filters inside a Filter wheel placed immediately before the camera.

Supplementary Fig. 2, Brightfield imaging with instant SIM. A lamp is used to illuminate the sample from above. Transmitted light is collected via the objective, passed through the Tube lens and Scan lens 2, and directed through Scan lens 4 via removable mirrors before imaging onto the camera.

Supplementary Fig. 3, Forming an upright image half the size of an object. A virtual object (O) placed at the right-hand focal point (f) produces an upright image (I) one half the object size, located on the same side of the lens. Three rays were traced from the virtual object (red dotted lines, for clarity shown only on the right side of the lens): (1) a ray parallel to the optic axis is refracted through the lens and passes through the right focal point; (2) a ray en route to the left focal point emerges horizontal to the optic axis; (3) a ray that passes through the center point of the lens continues undeflected. The intersection of these rays at f/2 indicates the image position. Tracing the rays back through the lens (on its left side) indicates the real rays (black) that would have formed the object (if the lens had not been placed in the optical path); these real rays are refracted by the lens to form the image.

Supplementary Fig. 4, Example control voltages for 5-timepoint, 2-color timelapse 2D acquisition. See online methods for accompanying description.

Supplementary Fig. 5, Example control voltages for a 4-slice, 2-color Z-stack. See online methods for accompanying description. See also higher magnification view in Supplementary Fig. 6.

Supplementary Fig. 6, Vibration control scheme for rapid Z axis piezo stage movement. Higher magnification view of voltages in Supplementary Fig. 5, illustrating a triple step voltage routine that enables ~10 ms piezo movement with minimal vibration. See online methods for further details.

Supplementary Fig. 7, Immunolabeled microtubules viewed by instant SIM (a) Maximum intensity projection of Alexa Fluor 488 antibody labeled microtubules in a U2OS cell, after analog processing shown in Fig. 1 and deconvolution. Scalebar: 10 m. (b) Top panel: Higher magnification view of the yellow boxed region in (a); Middle and bottom panels indicate apparent microtubule width and separation, corresponding to the red and blue lines marked in top panel. Scalebar: 500 nm.

Supplementary Fig. 8, Cross-sectional lineouts of 100 nm yellow green beads, as viewed in line-scanning, spinning-disk confocal, and instant SIM systems. Lateral (top) and axial (right) lineouts through 5 yellow green 100 nm fluorescent beads, pre-(left column) and post-(right column) deconvolution. Beads always appear sharpest in instant SIM. See also Table 1, Fig. 2, and Supplementary Fig. 9.

Supplementary Fig. 9, Apparent size of subdiffractive beads as viewed in line-scanning, spinning-disk confocal, and instant SIM systems. XZ cross-sectional views of 100 nm yellow-green fluorescent beads before (left column) and after (right column) deconvolution. See also Supplementary Fig. 8 and Table 1. Scalebars: 200 nm.

Supplementary Fig. 10, Voids in mitochondria, as revealed by instant SIM. Higher magnification view of a subregion of the Tom20-mCherry image in Fig. 2. Red arrows mark the inner mitochondrial space (voids). Scalebar: 1 m.

Supplementary Fig. 11, Mitochondria visualized in line-scanning, spinning disk confocal, and instant SIM systems, without deconvolution. Live MRL-TR transformed human lung fibroblasts expressing TFAM-GFP (green) and Tom20- mcherry (red) were imaged in instant SIM (a-c), a spinning disk confocal microscope (d-f), and a line scanning confocal microscope (g-h). Data and annotations are identical to those in Fig. 2, but data has not been deconvolved. Instant SIM achieves better resolution than either spinning disk or line-scanning confocal microscopy.

Supplementary Fig. 12, Deconvolution improves spinning disk and line-scanning confocal resolution, but not to the extent offered by instant SIM. Tom20-mCherry images were acquired with a line-scanning confocal microscope (left), a spinning disk confocal microscope (right), and instant SIM (bottom). Unlike instant SIM, submitochondrial features are not resolved in either line-scanning or spinning disk microscopes, even after deconvolution (right column). Scalebars: 1 m.

Supplementary Fig. 13, Axial sectioning of different microscopes Point-scanning (LSM 510), line-scanning, spinning disk, and instant SIM systems were used to image a thick fluorescent lake. Top: axial response as a function of defocus, bottom: central difference quotient (approximate derivative) of the same data. Instant SIM gives similar sectioning to spinning disk confocal, and worse sectioning than line- or point-scanning confocal techniques.

Supplementary Table 1, Primer sequences Name Sequence SEC61A Forw SEC61A Rev Tom20 Forw Tom20 Rev TFAM Forw TFAM Rev ACAGTCGACATGGCAATCAAATTTCTG ATAGGTACCTTAGAAGAGCAGGGCCCCC CACAGATCTATGGTGGGTCGGAACAGCGC ATAGAGCTCTTCCACATCATCTTCAGCC CACAGATCTATGGCGCTGTTCCGGGGAAT ATAGAGCTCATGCTCAGAGATGTCTCCGG See online methods, Molecular biology and cloning procedures for accompanying context.

Supplementary Note 1, Explanation of resolution-doubling in instant SIM. To understand the mechanism of resolution doubling in instant SIM, it is helpful to first understand the mechanism of resolution doubling in a standard point-scanning confocal microscope. A point-scanning confocal microscope with a pinhole closed almost completely produces a point-spread function (PSF) that is 1.4x narrower in the lateral direction, and doubles the frequency content of the resulting image 1. To illustrate the mechanism in more detail, we examine simulated images of test objects. Fig. S1 shows a cross section of three idealized test objects. The x-direction points leftright, the y-direction points into and out of the page, and the z-direction points up-down along the page. Test objects are invariant in the y-direction. The left test object is a thin line, like a microtubule. The middle test object is a cylindrical shell with two septa dividing it into quadrants. The right test object is a set of parallel sheets with variable spacing, oriented in the xy and xz planes. Fig. S1, Idealized test objects used in simulations Fig. S2 shows these test objects viewed with a confocal microscope. The test object is blurred with a calculated point spread function (PSF) using three different pinhole sizes. PSFs were calculated using the PSF Generator ImageJ plugin, available at http://bigwww.epfl.ch/algorithms/psfgenerator/#bw, with NA 1.4, index 1.5, wavelength 500, and pixel size 25 nm.

Fig. S2, Test objects viewed through a confocal microscope The top panel of Fig. S2 shows an image taken with a fully-open pinhole. In this case, the microscope's PSF is determined simply by the excitation PSF, the shape of the focused excitation laser beam. Lateral features (shown by the width of the line-like test object in the inset) are reasonably resolved, but axial structure of the sheets oriented in the xy-plane is almost completely obscured. Axial features are poorly imaged because the excitation laser beam does not die off in the axial direction, it merely spreads out.

The middle panel of Fig. S2 shows an image taken with a partially-closed pinhole (225 nm diameter in the image plane). In this case, the microscope's PSF is the product of its excitation and emission PSFs, and the emission PSF is the convolution of the widefield emission PSF with the partially open pinhole. Convolution with the partially-open pinhole further blurs the widefield emission PSF, but the product of this blurred emission PSF with the excitation PSF is still narrower than the excitation PSF alone. Additionally, the product of the excitation and emission PSFs dies off in the axial direction, so thin sheets oriented in the xy plane are now visible. Because the transverse resolution is improved in this configuration due to the combination of excitation and emission PSFs, confocal microscopy with a partially closed pinhole can be regarded as a form of structured illumination microscopy, albeit without fully doubled resolution compared to widefield microscopy. The bottom panel of Fig. S2 shows an image taken with a pinhole that is almost completely closed (25 nm diameter in the image plane). As in the middle panel, the microscope's PSF is the product of the excitation and emission PSFs, but now the emission PSF is not blurred by convolution with a large pinhole. Since the excitation and emission PSFs are nearly equal for most fluorescence microscopes, the resulting PSF is effectively the square of the widefield PSF, improving resolution by 2. Since multiplication in the spatial domain is equivalent to convolution in the spectral domain, it is clear that the frequency content of this microscope's image is doubled, implying that deconvolution can improve spatial resolution a full factor of two compared to widefield microscopy. However, this doubled resolution is usually not useful in practice. Fig. S3 shows simulated relative noise levels for the same images shown in Fig. S2. Assuming an ideal detector with no noise, measured signals still contain Poisson noise, and the signal-to-noise ratio (SNR) scales as the square root of the signal. The top panel of Fig. S3 shows a very bright widefield signal, difficult to distinguish from the noise-free case. In the middle panel, the test object is the same brightness as the top panel, but the partially-closed confocal pinhole rejects a substantial portion of the emission light. Noise is noticeable in this image, but the resolution is still much better than in the widefield case. In the bottom panel, the fully-closed pinhole rejects far too much of the emission light; any improvement in resolution compared to the middle panel is masked by the severe degradation in SNR.

Fig. S3, The influence of noise when imaging with a pinhole Ref. 1 suggests a method for improving resolution without sacrificing signal: close the pinhole nearly all the way, but use multiple nearly-closed pinholes to collect and combine the majority of the emission light. However, a naive summing of images will not work, as lateral misalignment alters the collected image. The top panel of Fig. S4 shows one such image, collected with a laterally displaced pinhole (150 nm effective displacement to the left in the image plane). Superficial features of the image are similar to the bottom panel of Fig. S2, but the

image is even dimmer, and more importantly, the image is laterally shifted. A second image collected with a second pinhole shifted in the opposite direction (150 nm effective displacement to the right in the image plane) produces a similar image, but shifted in the opposite direction. Simply adding these two images, as shown in the bottom panel of Fig. S4, produces a 'doubled' image, with substantially worse resolution than the individual images. Fig. S4, The result of laterally shifting the pinholes

Combining additional nearly-closed pinholes does not solve the problem; the top panel of Fig. S5 shows the result of adding the images collected by several small pinholes. The doubleimage is eliminated, but resolution is not improved compared to the partially-closed pinhole shown in the middle panel of Fig. S2. However, there is an alternative method for eliminating the double-image: simply shift the individual images appropriately before summing them! The bottom panel of Fig. S5 is a sum of the same images used to construct the top panel, but each image has been shifted before summing. As expected 1,2, the appropriate shift amount is half the amount that the pinhole is displaced, and the appropriate shift direction is opposite to the direction that the pinhole was displaced. For example, the 150 nm pinhole displacements used in Fig. S4 collect images that must be shifted 75 nm before summing with an image from an undisplaced pinhole. Fig. S5, Summing, then shifting pinholed images improves resolution without sacrificing signal The key principle which allows resolution doubling in a confocal point-scanning microscope without sacrificing signal is thus to use multiple small pinholes, but shift the images appropriately in order to compensate for the shift induced by pinhole misalignment before summing their signals. In previous digital implementations such as image scanning microscopy

(ISM) 2 or multifocal structured illumination microscopy (MSIM) 3, shifting is accomplished computationally. In instant SIM, shifting is accomplished optically. As shown in the top panel of Fig. S6, this splits the job of the confocal pinhole between two separate types of pinhole, "micropinholes", and the "macropinhole". Figure S6, Macro- and micropinholes in instant SIM The macropinhole plays the same role as the partially-closed pinhole used for the middle panel of Fig. S2-S3: it rejects out-of-focus light to provide optical sectioning, but is not small enough to double transverse resolution, or to badly degrade signal levels. In MSIM, the macropinhole was applied with digital postprocessing; in ISM and instant SIM, the macropinhole is implemented optically with a physical pinhole. The micropinholes play the role of the nearly-closed pinhole, allowing doubled transverse resolution. As shown in the top panel of Fig. S6, the micropinholes are not centered, so their individual images must be corrected. In ISM, MSIM, and instant SIM, the pixels of the camera detector are used as micropinholes. In ISM and MSIM, the correction is implemented by collecting multiple exposures with different illumination positions, and digitally reassigning the collected signal to different image bins using interpolation. The reassignment is simple: since a pinhole displaced by X collects an image displaced by X/2, shrink the image of each pinhole by a

factor of 2 towards the center of the pinhole. In instant SIM, this correction is implemented by using a microlens to 'shrink' pinholed images by a factor of 2 before they reach the detection camera; no interpolation is required, due to the analog nature of reality. Note that these two approaches, digital and analog, are entirely equivalent, except that the analog approach is much faster. With this understanding of how a single illumination point at a particular sample position contributes signal to the image, we construct the full super-resolution image by scanning the illumination over the entire sample. In ISM, a single excitation point is used, providing excellent sectioning but slow speed. The camera is placed in the descanned emission path, so micropinholes are stationary with respect to the macropinhole and the required shift, applied digitally, is independent of scan position. In MSIM, multiple excitation points are used to increase acquisition speed, albeit at the cost of pinhole crosstalk. In MSIM the emission is not descanned, so micropinholes are not stationary with respect to the macropinhole. The required shift, implemented digitally, changes with scan position. Instant SIM, like MSIM, uses multiple excitation points to speed acquisition. Like ISM, the emission is descanned, so the required shift is independent of scan position, but the shifting is performed optically (with a microlens array), rather than digitally. Following the shifting, the beam is rescanned to produce a stationary image of the sample on the camera, as in MSIM. In all implementations, as long as the sample is painted evenly, the resulting image appears smooth. We note that our implementation of instant SIM has several dead microlenses, resulting in a striping artifact in portions of our images. We anticipate that better quality control would solve this problem. Noise levels, of course, are crucial to image quality and resolution. ISM uses a moderate number of optics in the emission path, but a very small scan step size. This provides reasonable transmission efficiency (preventing excess Poisson noise), but requires many exposures (and thus an increased level of read noise). MSIM uses fewer emission optics and a larger step size than ISM, improving both Poisson and read noise, although presumably provides worse sectioning than ISM due to pinhole crosstalk. Instant SIM has more emission optics than ISM, but uses only a single camera exposure, thus increasing Poisson noise but minimizing read noise. Finally, we note that digital macropinholes compared to analog macropinholes give approximately equivalent levels of background rejection; emission light that falls outside the macropinhole is ignored in either case. We note that if excitation and emission PSFs are not so nearly identical (as in two-photon microscopy or stimulated emission depletion microscopy, for example), the appropriate amount to shrink the image of the pinhole is no longer a factor of 2; smaller excitation PSFs demand more shrinking, and larger excitation PSFs demand less shrinking, assuming the emission PSF stays the same size. The appropriate shift can be estimated precisely by calculating the position of the maximum of the product of the excitation PSF with the shifted emission PSF.

1 Sheppard, C. J. R. Super-resolution in Confocal Imaging. Optik 80, 53-54 (1988). 2 Muller, C. B. & Enderlein, J. Image Scanning Microscopy. Physical Review Letters 104, 198101 (2010). 3 York, A. G. et al. Resolution Doubling in Live, Multicellular Organisms via Multifocal Structured Illumination Microscopy. Nature Methods 9, 749-754 (2012).