Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.

Similar documents
Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400.

JEOL 6500 User Manual

JEOL 6700 User Manual 05/18/2009

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014

Instructions for Tecnai a brief start up manual

Using the Hitachi 3400-N VP-SEM

SEM Training Notebook

FE-SEM SU-8020 Operating manual (Preliminary version)

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual

SEM OPERATION IN LOW VACUUM MODE

FEI Tecnai G 2 F20 Operating Procedures

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier).

User Operation of JEOL 1200 EX II

SEM Training Notebook

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below.

Dickinson College Department of Geology

2. Raise HT to 200kVby following the procedure explained in 1.6.

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

Basic Users Manual for Tecnai-F20 TEM

2 How to operate the microscope/obtain an image

Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016

SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right

MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope

Check that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course)

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack).

STEM alignment procedures

1. Specimen Holder Removal, Loading, and Insertion

RAITH e-line OPERATING INSTRUCTIONS

1.1. Log on to the TUMI system (you cannot proceed further until this is done).

ZEISS EVO SOP. May 2017 ELECTRON OPTICS

Cressington 108 Auto/SE Sputter Coater Standard Operating Procedures (S.O.P)

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope

1.1. In regular TEM imaging mode, find a region of interest and set it at eucentric height.

Protective Equipment Nitrile gloves for handling sample holder and safety glasses for filling liquid nitrogen dewar.

05/20/14 1. Philips CM200T. Standby Condition

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire

General information. If you see the instrument turned off, notify MIC personnel. MIC personnel will help you insert your samples into the instrument.

1. Preliminary sample preparation

Model SU3500 Scanning Electron Microscope

Things to check before start-up.

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

JEOL JEM-1400 Transmission Electron Microscope Operating Instructions

Zeiss AxioImager.Z2 Brightfield Protocol

Brightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer

CNC Using the FlexiCam CNC and HMI Software. Guldbergsgade 29N, P0 E: T:

Bruker Dimension Icon AFM Quick User s Guide

STANDARD OPERATING PROCEDURE: JEOL TEM-2100

SAMUEL ROBERTS NOBLE ELECTRON MICROSCOPY LABORATORY. Operating Procedures for the Zeiss 9 S-2. Transmission Electron Microscope

The University of Texas at Austin Electron Paramagnetic Resonance Facility. Bruker EMX+ CW EPR Spectrometer

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

Recording EPR spectra using the Loop Gap Resonator (LGR)

FEI Quanta 200 ESEM Basic instructions

How to use the Jeol 1010 TEM of GI (Liesbeth own GI version)

Installation and Operating Instructions Desk top Laser

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

Standard Operating Procedure Hitachi UHR CFE SU8230 SEM

Zeiss LSM 880 Protocol

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface.

Nanoscale Fabrication & Characterization Facility. Raith e-line EBL Users Guide (updated:aug 2 nd, 2017)

Recording EPR Spectra using ER 4102ST Resonator

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

FEI Titan Image Corrected STEM

JEOL 2010 FasTEM & DigitalMicrograph User's Guide

Guide to Confocal 5. Starting session

The user should already be familiar with operation of the instrument in STEM mode, use of the Microscope Control interface, and TIA.

JEOL 5800LV SCANNING ELECTRON MICROSCOPE OPERATOR'S MANUAL

Jeol JEM Responsible personell: Endy ( ) Online booking is compulsory!

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Bruker Dimension Icon AFM Quick User s Guide

Learning to use your easypod. [somatropin (rdna origin) for injection]

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

Standard Operating Manual

Zeiss LSM 780 Protocol

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

WITec Alpha 300R Quick Operation Summary October 2018

COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual

Strata DB235 FESEM FIB

UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY

Materials Polishing Manual. By Thomas Perry Daniel Webster College Version 1 August 10, 2007

ILFORD SPORTSVIEW PROJECTOR INSTRUCTION BOOK

Nikon SIM-E & A1-R System

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

LSM 800 Confocal Microscope Standard Operation Protocol

JEOL JEM 2010 TRAINING TRANSMISSION ELECTRON MICROSCOPE USER MANUAL

CM20 USER GUIDE. Duncan Alexander, CIME 2010

Print Head Installation Guide

AutoSeal FD 2006IL / FE 2006IL

Zeiss Axio Imager.A1 manual

The ideal K-12 science microscope solution. User Guide. for use with the Nova5000

Standard Operating Procedure Hitachi UHR CFE SU8230 SEM

INTRODUCTION CHAPTER 1

Transcription:

Smith College August 2005 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check, 1 Specimen Insertion, 1 Startup, 2 Filament Saturation, 2 Beam Alignment, 3 Check Alignment, 3 Viewing, 4 Screen Text, 5 Photography, 6 Instrument setup, 6 Film insertion, 6 Film processing, 7 Digital Image Capture, 8 Filament De-saturation, 9 Specimen Removal, 9 Shut Down, 10 ATTACHMENT 1: Critical Point Dryer, 11 ATTACHMENT 2: Sputter Coater, 12

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400 Pre-Check - Check logbook - Check vacuum (10-7 range) Specimen Insertion - Check stage controls: WD (working distance) 39 mm X (direction) 25 mm Y (direction) 35 mm T (tilt) 0 R (rotation) 0 - Insert specimen into holder with forceps and tighten screws - Attach specimen holder to insertion rod DO NOT TOUCH THE METAL INSERTION ROD - Place rod with holder in port and press the Red Button on top - Wait for the red light to go out before opening the Air Lock door To open the door: Turn flat on the knob 90 towards you Pull out till it stops - Push rod in and secure holder on dovetail slide - Unscrew rod CAREFULLY and retract fully - Close airlock door by pushing in and turning the knob 90 to the ceiling - Push Red Button while supporting the rod with your hand - Remove the rod once the airlock has vented 1

Startup - Press FAST scan on the keyboard - Turn up CRT1 (left) and CRT2 (right) using wheels located under screens - Press ESC on keyboard to show cursor on CRT2 (ESC toggles cursor back and forth between the screens) - Press BREAK on the keyboard to show text (image settings) on CRT2. To remove text (image settings), press BREAK again - To set operating parameters: Press PF-5 to show conditions Use arrows to highlight START UP Press L (load) Press PF-2 followed by 1to show EOS-1 screen Filament Saturation - Press MODE button in panel 1to show scan line on CRT1 - Press the ACCEL VOLTAGE button (light should go on) - SLOWLY turn up the EMISSION knob (clock wise) to saturate the filament. The line will move up, drop down, move up and plateau (saturation). - During saturation use BRIGHTness knob to center the scan line on the screen 2

Beam Alignment - Check gun alignment TILT / SHIFT button, light should be off - Turn PROBE CURRENT up to 13 and use TILT control knobs (X and Y) to find the point of maximal deflection - Press TILT / SHIFT button, light should go on - Drop PROBE CURRENT to 1 and obtain maximal deflection using SHIFT control knobs (X and Y) - Return PROBE CURRENT to 13 (SHIFT light turned off) and again maximize signal using TILT control knobs - Set PROBE CURRENT back to 7 (or otherwise if desired) - Use FOCUS knob to adjust the scan line for maximum detail - Use CONTrast knob to adjust peaks and valleys (2-3 cm amplitude) - Use BRIGHTness knob to position the scan line mid-screen - Press PIC button on panel 1 to show image Check Alignment (optional) - Recommended for high magnification and high resolution work: - Press MODE on panel 1 one or more times to get cross-hairs - Press ESC (cursor on CRT1) followed by BREAK to show image settings including a scale marker - Turn MAGNIFICATION up to 5000X - Run PROBE CURRENT from 13 to 1, shift should be less than one micron else repeat 'Beam Alignment' - When finished press BREAK to remove image settings from CRT1 - Press ESC to return to CRT2 - Press PF-2 followed by 1 to show parameters (EOS-1 screen) if not already there - Press MODE one or more times to get scan line - Press PIC button to show image 3

Viewing - Several adjustments to the image on CRT1 can be made: - Use FOCUS knob to focus image, press FINE button for fine focussing ability (light should go on) - Use STIGMATOR knobs (X and Y) to correct for astigmatism - Use knobs at the front of the specimen chamber to move the specimen: X and Y (direction) R (rotation) T (tilt) - Adjust magnification using MAGNIFICATION knob - Press WFM to properly adjust contrast and brightness. (Use CONTrast and BRIGHTness knobs to move peaks up to 2 nd line from top and valleys about halfway between 2 nd and 3 rd line from bottom. Turn off WFM) - Use PROBE CURRENT knob to adjust condenser lens aperture - To change working distance turn big knob on top of the specimen chamber - To change accelerating voltage follow the instructions on 'Filament De-saturation', change accelerating voltage using the knob right of ACCEL VOLTAGE button and re-saturate the filament (see 'Filament Saturation') - To change specimen follow the instructions on 'Filament De-saturation', 'Specimen Removal', 'Specimen Insertion' and 'Filament Saturation ' - Notice OPTimum APERture on EOS-1 screen. DO NOT CHANGE! 4

Screen Text - Press ESC on the keyboard to get cursor on CRT1 (ESC toggles cursor back and forth between screens) - Press BREAK to show text (image settings) on CRT1. To remove text (image settings), press BREAK again - To change existing text (image settings): a) Use arrows to place cursor on JEOL b) Press INS on keyboard c) Type text up to 6 characters (CLS will remove INS effect) d) Press ENTER - To delete text (image settings): a) Place cursor on block of text b) Press ENTER - To return text (image settings): a) Place cursor on empty block b) Press ENTER - To move whole text (image settings), use arrow-keys on keyboard - To write different information on CRT1: a) Press BREAK and cursor will appear b) Move cursor with arrow-keys to desired location c) Type text (CLS will erase text) d) Press ENTER - To add text (image settings) to written information, press BREAK. To remove text (image settings), press BREAK again 5

Photography Instrument setup - Find specimen and focus (see 'Viewing') - Press SCAN PIC button on panel 1 one or more times to focus on a brighter and smaller part of your image: Press IMAGE / POSITION button (lights on) to move frame over image using knobs located left and right Press IMAGE / POSITION button (lights off) to move image in the frame using knobs located left and right - Press D-MAG button on panel 1 to show actual size of the print. Turn off D-MAG - Press WFM to adjust control and brightness (see 'Viewing' for detail). Turn off WFM - To remove information temporary from CRT2: a) Press ESC to get cursor or highlighted block on CRT2 (if not already there) b) Press PF-1 followed by BREAK c) To get information back, press PF-2 followed by 1 Film insertion - Two types of Polaroid film (4 X 5 inches) can be used: a) Type 52: delivers a positive print only b) Type 55: delivers a positive print and a negative - Push lever on film holder (Model 545) to the right (L) position - Slide film all the way in (camera side down) until a "click" is heard - Pull cardboard cover back as far as it will go - Be sure f-stop setting is at f16 for type 52 film or f5.6 for type 55 film (dark red knob right of film holder) - Press LEFT button on panel 1 to begin the super slow scan for taking a picture. When finished CRT goes back to rapid scan mode - Push cardboard sleeve all the way back into film holder - Pull lever on film holder to the left and pull the film smoothly out of the film holder 6

Photography - continue: Film processing Type 52 - Develop film for 15 seconds (start as film leaves the film holder) - Peel off paper covers - Use special Coater to coat the print as soon as possible. DO THIS ON PAPER TOWEL, NOT ON CONSOLE OF INSTRUMENT! - Dispose film pack and used up Coaters properly Type 55 - Develop film for 20 seconds (start as film leaves the film holder) - Open pack and separate print from negative quickly and carefully. BE CAREFULL NOT TO TOUCH THE SURFACE OF THE NEGATIVE - In warm water, fold ends of negative and tear them away. Discard the tabs - Swirl negative in warm water to cause the developing layer to drop off - Rinse for at least 5 minutes in warm water - Dip negative in wetting agent solution and hang up to dry - Use special Coater to coat the print as soon as possible. DO THIS ON PAPER TOWEL, NOT ON CONSOLE OF INSTRUMENT! - Dispose film-pack and used up Coaters properly 7

Digital Image Capture - If computer is not ON, turn ON computer. Turn ON monitor and printer - Load ISIS by clicking on Link ISIS icon or select Link ISIS from the Start menu - After a message about Resetting Hardware, a Welcome Window will appear - Choose an operator, which will ordinarily be you or the name of a course (SEM interterm) - Click on the blue Labbook button to get the Labbook window - To obtain an image with correct scale-bar: - Select the X-Ray Analysis button (first button from left) in the Labbook window - Select the Microscope button (bottom right) in the X-Ray Analysis window - Click on the magnifying glass button, enter the working magnification (see SEM monitor) and select OK. Every time you change the working magnification on the SEM you need to adjust the magnification in the ISIS program before acquiring an image - Go back to the Labbook window by selecting Labbook from the menu bar - Select the Autobeam button (second from left) in the Labbook window - Go to view in the Autobeam window and select Scale Marker Bar from the dropdown menu. This should be checked off - To obtain an image without scale-bar: - Select the Autobeam button (second from left) in the Labbook window - Go to view in the Autobeam window and make sure the Scale Marker Bar from the dropdown menu is NOT checked off - To acquire an image, click on the round Start button on the upper left of this window. This digital image will improve by averaging over time - To stop acquiring an image, click the square Stop button next to the Start button - To change image size and detail (# of pixels) as well as the speed of acquisition, select Image setup from the Edit menu - To adjust contrast and brightness, click on the third button from left in the Autobeam window - To save an image select Export as a TIFF or Export as a BMP from File menu. Drive d is for ZIP disk - To print an image, select Print from File menu - When finished close all windows (including Labbook window) to log OFF 8

Filament De-saturation - SLOWLY turn EMISSION knob all the way down (counter clock wise) - Press ACCEL VOLTAGE button (light should go out) Specimen Removal - Check stage controls: WD (working distance) 39 mm X (direction) 25 mm Y (direction) 35 mm T (tilt) 0 R (rotation) 0 - DO NOT TOUCH THE METAL INSERTION ROD - Place rod in port and press the Red Button on top - Wait for the red light to go out before opening the Air Lock door To open the door: Turn flat on the knob 90 towards you Pull out till it stops - Push rod in and contact specimen holder - Screw rod in and pull rod and holder out - Close airlock door by pushing in and turning the knob 90 to the ceiling - Push Red Button while supporting the rod with your hand - Remove the rod once the airlock has vented - Remove specimen holder from insertion rod - Loosen screws and remove specimen from holder with forceps 9

Shut Down - After 'Filament De-saturation' and 'Specimen Removal': - Turn electronics of the microscope to the proper shut-down position: Press PF-5 to show conditions Use arrows to highlight SHUT DOWN Press L - If image settings show, press PF-1 followed by BREAK to remove data on CRT2 - Press ESC to move cursor from CRT2 to CRT1 - If there is information on CRT1, press BREAK (only cursor in left corner is shown) - Turn down CRT1 and CRT2 - To reduce use of scan coils, press SLOW scan on the keyboard twice. A scan line will appear - Use BRIGHTness knob to reduce the brightness of the scan line - Sign out Logbook - Leave room CLEAN and TIDY! 10

ATTACHMENT 1: Critical Point Dryer 1. Turn power switch ON and make sure the hood fan is ON 2. Ensure that all three valves are closed 3. Fill chamber with 100% EtOH and transfer specimen(s) into chamber 4. Put the lid on the chamber, tightening all three screws evenly 5. Open tank valve, allowing CO2 into pipes 6. Open the cool valve, allowing the temperature to drop to about 5 ºC 7. Close cool valve 8. Open the inlet valve 9. Allow chamber to fill for about 3 minutes (set timer) 10. Open the exhaust valve slightly, purging the chamber for 30 seconds 11. Close exhaust valve 12. Repeat purging procedure (step 9-11) five times. Use pennies to keep track 13. Open the exhaust valve slightly and use a Kimwipe to confirm (smell) that there is no EtOH left 14. Close the exhaust valve, close the inlet valve, close the tank valve, and open the cool valve to drain the line. Close the cool valve 15. Turn on the heat switch, allowing the temperature to rise to about 40 ºC, at which the indicator light will go out. During this time, if the pressure in the chamber rises above 1400 psi, open the exhaust valve very slightly to minimally decrease the pressure to below 1400 psi 16. When the heat indicator light goes out, vent the chamber by opening the exhaust valve very slightly. Use a container with water to check the exhaust speed (separate air bubbles). The pressure must be decreased to zero slowly! 17. Remove the camber lid and your specimen(s) 18. Make sure that all switches are turned OFF and that the lid is re-secured before you leave 19. Sign out Critical Point Dryer Logbook 11

ATTACHMENT 2: Sputter Coater Sample insertion 1. Lift the lid off the vacuum chamber 2. Carefully remove the glass cylinder and lay it on its side where it won t roll off the counter. This prevents messing up the vacuum grease on the top and bottom rims 3. Place your specimens on the central area of the pedestal, not the raised ring around the periphery 4. Replace the glass cylinder and close the lid Control vacuum 1. CLOSE the vent toggle valve (down) 2. CLOSE the increase/decrease needle valve (clockwise). Be sure the high voltage control switch is OFF and the high voltage dial is at O 3. TURN the main power switch ON 4. Wait 15 20 min. to create a vacuum in the vacuum chamber 5. OPEN the increase/decrease needle valve (3 full turns counter clockwise) to increase the air flow in the vacuum chamber Coating 1. High voltage switch should be OFF; set high voltage dial to 9 2. Press the high voltage switch ON 3. Allow about 10 seconds for the instrument to stabilize and the vacuum chamber to outgas 4. Use the needle valve to ADJUST the current so that the current meter reads 10 milliamps: - if the current is to high, turn the needle valve toward decrease (clockwise) - if the current is to low, turn the needle valve toward increase (counter clockwise) 5. Coat specimens for 2 to 3 minutes 6. While coating, touch up the current flow if necessary, to keep the meter at 10 milliamps 7. After coating, turn high voltage switch OFF and set dial back to 0 Sample removal 1. TURN main power OFF 2. Immediately OPEN the air inlet toggle valve (lift) 3. Open the lid 4. Remove the glass cylinder (lay it on its side where it won t roll) 5. Remove your specimen to a safe storage container Shut down 1. CLEAN THE CYLINDER inside with a paper towel wet with absolute EtOH or acetone 2. Check that the vacuum seals are clean and greased and reassemble the machine 3. Do not pump down, leave it under atmospheric pressure, close increase/decrease needle valve (clockwise) 4. Sign out Sputter Coater Logbook 12

13