Ion optics innovations for increased sensitivity in hybrid MS systems

Similar documents
Agilent 6400 Series Triple Quadrupole LC/MS System

Agilent 6400 Series Triple Quadrupole LC/MS System

Introducing the Agilent 7000A Triple Quadrupole GC/MS

Ultrasensitive LC MS/MS: Agilent 6470 and 6495 LC-QQQ

LC/MS/MS. Page Header. triple quadrupole mass spectrometer.

Mass Spectrometry and the Modern Digitizer

Ultra High Definition LC/MS

Collision cell: Fragmentation efficiency > 90 % Duty cycle: Variable between 1 and 50 % (user selectable trapping time/scan time ratio) Transmission:

The Extrel MAX Systems are UHV-compatible flange mounted Quadrupole Mass

QUANTOF. High-resolution, accurate mass, quantitative time-of-flight MS technology

Welcome! Denver, CO Houston, TX February 2009

AccuTOF LC Training Course

830 Vacuum Quality Monitor

Characterization of Common Electron Multipliers in Harsh Environments

835 Vacuum Quality Monitor (VQM TM ) 2012 Brooks Automation, Inc. Proprietary Information

Spectroscopy in the UV and Visible: Instrumentation. Spectroscopy in the UV and Visible: Instrumentation

Supplementary Information

The Agilent OneNMR Probe

The development of the low-pressure linear ion

The Best Way to Get the Right Answers... is in a Flash

Dust Measurements With The DIII-D Thomson system

Fragmentation of Ions in a Low Pressure Linear Ion Trap

Technical Specifications for the procurement of Inductively Coupled Plasma Mass Spectrometer (ICP-MS)

Historical. McPherson 15 Mount

Performance Characterization Of A Simultaneous Positive and Negative Ion Detector For Mass Spectrometry Applications

A New Conjoined RF Ion Guide for Enhanced Ion Transmission

Advantages of Multi-Turn-TOF Technology

Electron Multipliers. Introduction to ETP Electron Multipliers Selection by Technique: LC-MS 242

Picosecond Time Analyzer Applications in...

Performance characteristics of a new wide range, fast settling electrometer design for a residual gas analysis mass spectrometer

MIMS Workshop F. Hillion. MIMS Workshop

SCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS

Cylindrical Toroidal Ion Trap Mass Spectrometer. Daniel Austin and Nick Taylor Brigham Young University, Provo, Utah

Agilent 10774A Short Range Straightness Optics and Agilent 10775A Long Range Straightness Optics

Residual Gas Analyzers

HIQUAD. New high-end mass spectrometer! Fast, flexible and easy to operate.

The 5975C Series MSDs: Normalized Instrument Tuning. Technical Overview. Introduction

Thermo Scientific icap 7000 Plus Series ICP-OES: Innovative ICP-OES optical design

SCCH 4: 211: 2015 SCCH

Multistage Transversal Modulation Ion Mobility Spectrometry: Reducing the Voltage Required for High Resolution IMS for pre-existing mass spectrometers

PUV3402 LED multiwave photometer A new approach to online process photometry

Spectrophotometer. An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer :

CHAPTER 11 HPD (Hybrid Photo-Detector)

Residual Gas Analyzers RGA100/200/ amu, 200 amu and 300 amu RGAs

Major Fabrication Steps in MOS Process Flow

TRITON Plus Thermal Ionization MS. Another step ahead. in TIMS. Earth and Planetary Sciences Nuclear Sciences

True simultaneous ICP-OES for unmatched speed and performance

Unpolarized Cluster, Jet and Pellet Targets

Observational Astronomy

J.A. Casey and J.H. Irby. M.I.T. Plasma Fusion Center

The Development of Miniature Electron Multipliers for Use In Portable Mass Spectrometers

Autoresonant Trap Mass Spectrometry

SCIEX Presentation. Name of Speaker, Identification. QTRAP Functionalities. Alexandre Paccou Name of Speaker, Identification. SCIEX Presentation.

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

Agilent 8700 LDIR Chemical Imaging System. Bringing Clarity and Unprecedented Speed to Chemical Imaging.

ICP-MS. plasma 3. Multi-Collector ICP-MS.

Superior ICP-OES optical design for unmatched speed and performance

StarBright XLT Optical Coatings

High specification CCD-based spectrometry for metals analysis

Gamma Spectrometer Initial Project Proposal

ToF-AMS DAQ. Joel Kimmel Univ. of Colorado at Boulder & Aerodyne Research, Inc AMS Users Meeting 2007

Fast Methods for Small Molecules

Practical Quadrupole Theory: Graphical Theory

LC/MS Purification System TRILUTION LC v3.0 User s Guide

MRI SYSTEM COMPONENTS Module One

Contemporary TOF instruments with orthogonal

BioInstrumentation Laboratory

Capillary Flow Technology: GCxGC Flow Modulator GET A SECOND DIMENSION OF INFORMATION ON COMPLEX MIXTURES

8.2 Common Forms of Noise

AutoMax Fast, automated method optimization

Automated Orthogonal Control System for Electrospray Ionization Mass Spectrometry

SECONDARY ELECTRON DETECTION

NEPTUNE Plus Multicollector ICPMS. Another step ahead. in MC-ICP-MS. Earth and Planetary Sciences Nuclear Sciences

Three-dimensional (3-D) ion traps can benefit from

Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors

R. J. Jones Optical Sciences OPTI 511L Fall 2017

Grant No. DE-FG-05-85ER Progress Report One. R. H. Atalla. The Institute of Paper Chemistry. April, Prepared for

Density and temperature maxima at specific? and B

Improving Signal-to-Noise by Identifying Sources of Noise in Mass Spectrometer Systems

The Development of a Novel Electron Multiplier with an Onboard Integral High Voltage Power Supply for use in Mass Spectrometers

Damage-free failure/defect analysis in electronics and semiconductor industries using micro-atr FTIR imaging

Scintillation Counters

Design of ESS-Bilbao RFQ Linear Accelerator

A New Class of Robust Sub-nanosecond TOF Detectors with High Dynamic Range

CHAPTER 7. Components of Optical Instruments

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background

Assembly and Experimental Characterization of Fiber Collimators for Low Loss Coupling

880 Quantum Electronics Optional Lab Construct A Pulsed Dye Laser

N-Methyl-1-(6-methylpyridin-2-yl)propan-2-amine

Agilent Cary 610/620 FTIR microscopes and imaging systems RESOLUTION FOR EVERY APPLICATION

Sub-mm Linear Ion Trap Mass Spectrometer Made Using Lithographically Patterned Ceramic Plates

Subminiature Photoionization VOC Sensor Boris Dolgov, Baseline-MOCON, Inc.

event physics experiments

Non-Destructive Ion Trap Mass Analysis at High Pressure. Supporting information

Abstract No. 32. Arne Bengtson and Tania Irebo. Swerea KIMAB AB, Isafjordsgatan 28A, SE Kista, Sweden

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

Determination of the N-Nitrosamine Content in Rubber Articles Using the Agilent 7000A Triple Quadrupole GC/MS System

High resolution ion mobility-mass spectrometry for separation and identification of isomeric lipids

HIGH-FIELD ASYMMETRIC WAVEFORM ION MOBILITY SPECTROMETRY: EVALUATION OF NOVEL CELL GEOMETRIES AND TECHNIQUES

INTERNATIONAL STANDARD

Transcription:

Ion optics innovations for increased sensitivity in hybrid MS systems Abstract Sensitivity, mass resolution, scan speed, spectral fidelity, mass accuracy, and mass range are just a few of the parameters that are used to gauge the performance of a mass spectrometer. While exactly which of these parameters is most important to the instrument user can vary by application, sensitivity is almost always of great interest since sensitivity determines the limit of detection. This overview describes shared ion optics design features that are key to achieving outstanding sensitivity in the Agilent Technologies 6400 Series Triple Quadrupole and 6500 Series Quadrupole Time-of-Flight LC/MS systems. It also describes unique design features that contribute to sensitivity of the individual systems.

Ion optics innovations for increased sensitivity in hybrid MS systems Introduction Maximum sensitivity is not achieved by the performance of any single mass spectrometer component, but results from the interaction of many components. It requires: Maximizing ion generation while simultaneously minimizing the generation of noise Efficient ion transmission and mass filtering with as little ion loss as possible Efficient ion detection that simultaneously reduces or eliminates signal from non-ion sources (either chemical or electronic) The ion optics of Agilent triple quadrupole and quadrupole time-of-flight (Q-TOF) systems are shown in Figure 1. The major differences between the two instruments are in those components that come after the collision cell and are associated with the second mass filter and detector. Figure 1. Q-TOF (top) and triple quadrupole (bottom) share most ion optics components up through the collision cell. This reduces the number of variables in multi-instrument research. They differ in the components associated with the second mass filter and detector. 2

Electrospray ionization source The first critical component in determining the sensitivity of the triple quad and Q-TOF is the electrospray ionization (ESI) source. The primary design goal of the ion source is to maximize the number of ions produced by the ESI technique while minimizing the generation of neutral particles and high-energy ion clusters that cause noise. Figure 2 shows the orthogonal design of the patented Agilent ESI source. The source incorporates many design features to improve the ion transmission while minimizing the noise that is generally associated with atmospheric pressure ionization. Nebulizing gas Eluent spray Entrance to dielectric capillary + + + + + + + Ions Heated nitrogen drying gas + + + + + + + + + + Dielectric capillary Figure 2. Electrospray ion source is designed to maximize ion generation while minimizing the amount of chemical noise generated and transmitted into the vacuum system. Generation of small, uniform droplets The patented mechanical centering design of the ESI nebulizer uses a sheath of nitrogen flowing concentric to the LC effluent to produce droplets. A converging nozzle mechanism accelerates the nebulizing gas. The resulting high velocity and uniform flow of nitrogen produces smaller, more uniform droplets. Smaller, more uniform droplets desolvate in a more consistent manner, ultimately delivering more ions into the instrument and increasing the signal strength. Orthogonal nebulization The ESI nebulizer is oriented orthogonally to axis of the dielectric sampling capillary. The orthogonal orientation greatly reduces the number of neutral particles entering the mass spectrometer, thus eliminating noise that would otherwise be generated by those particles. This improves the ultimate signal-to-noise ratio of the instrument. Heated drying gas A counter current of heated nitrogen drying gas effectively desolvates the ions formed in the ESI spray. By desolvating the ions in the ion source before they enter the sampling capillary, the drying gas significantly reduces the amount of solvent vapor entering the capillary. Ultimately, this increase signal strength and reduces the number of solventadduct ions. www.agilent.com/chem/ms 3

Ion optics innovations for increased sensitivity in hybrid MS systems Dielectric sampling capillary The 6400 and 6500 Series LC/MS systems use a dielectric sampling capillary to convey ions from atmospheric pressure in the ESI source to the vacuum region inside the mass spectrometer. Because a substantial pressure drop (~3X) occurs gradually over the entire length of the capillary, expansive cooling of the ions upon their exit from the capillary is reduced. This significantly reduces the formation of adduct ions. By contrast, in mass spectrometers that use an orifice or series of orifices instead of a long capillary, pressure reduction often occurs more abruptly with greater expansive cooling. This can result in formation of solvent clusters (sovlent adducts) and generation of matrix ions significantly above the molecular weight of the analytes. Heated, counter-current drying gas in the ion source also makes it possible for the 6400 and 6500 Series LC/MS systems to use a sampling capillary that is only indirectly heated. LC/MS systems that do not feature heated drying gas often use directly heated, high-temperature capillaries to accomplish desolvation. High temperature capillaries are prone to contamination, and without prior reduction of solvent load by the drying gas, they also tend to produce more solvent adducts. The Agilent dielectric sampling capillary carries a high voltage at its entrance and a low voltage at its exit. This allows the ESI nebulizer to be grounded, eliminating any chance of high-voltage discharge from the nebulizer to personnel using the mass spectrometer. Skimmer Ions exit the capillary with substantial momentum. Beyond the exit of the capillary is a skimmer (Figure 3) that helps maximize ion transmission while minimizing the transmission of neutral gas molecules. Heavier sample ions pass through the skimmer while lighter, neutral gas molecules are pumped away by the vacuum system. The skimmer angle, hole size, position, and distance from the exit of the capillary have been carefully balanced and optimized to achieve maximum ion transmission. Figure 3. The skimmer angle, hole size, position, and distance from the exit of the capillary have been carefully balanced and optimized to achieve maximum ion transmission. 4

Octopole ion guide An octopole ion guide (Figure 4) helps maintain a high degree of ion transmission as pressure is reduced from ~1 Torr to ~10 4 Torr by a split-flow turbomolecular pump. Four-, six- or eight-rod ion guide assemblies can be used to perform this function. An eightrod ion guide was chosen because it requires lower operating voltages than a four- or six-rod ion guide, and because it is more efficient at transmitting ions over a broad range of masses. A 5 MHz RF drive frequency applied to the octopole rods contributes to the broad mass-transmission range. Lenses 1 and 2 The electrical fields generated by lenses 1 and 2 (Figure 5) optimize transmission of ions from the octopole ion guide into the first mass filter (Q1). Lens 1, carrying a DC voltage, shields the octopole ion guide from effects of the DC and RF voltages on lens 2. The DC voltage on lens 2 accelerates the ions. The RF voltage applied to lens 2 matches the phase angle and amplitude of the RF voltage on the quadrupole to improve acceptance of ions into the quadrupole. The voltages applied to lens 2 increase ion transmission by approximately 2.5 times for ions with m/z greater than 600 and 5 times for ions with m/z greater than 1500 (Figure 6) compared to the same system without these voltages. Figure 4. The eight-rod ion guide requires lower operating voltages than a four- or six-rod ion guide and more efficiently transmits ions over a broad range of masses. Figure 5. Lenses 1 and 2 optimize transmission of ions from the octopole ion guide to the first mass filter (Q1). www.agilent.com/chem/ms 5

Ion optics innovations for increased sensitivity in hybrid MS systems % Relative improvement 1000 900 800 700 600 500 400 300 200 100 0 Isolation width narrow medium wide 118 322 622 922 1222 1522 1822 2122 2422 2722 m/z Quadrupole mass filters The 6400 Series triple quadrupole LC/MS systems feature two quadrupole mass filters. The 6500 Series Q-TOF LC/MS systems use a single quadrupole mass filter in combination with a timeof-flight mass filter. bolic rods made from 440C stainless steel alloy. During manufacturing, the rods are positioned to a precision of better than 1 micron and then locked into alignment. Ion throughput in a quadrupole mass filter (Figure 7) is greatly affected by Figure 6. Compared to the same system without the lens 2 RF, a system with it can increase ion transmission by 2.5 times for ions with m/z greater than 600 and by 5 times for ions with m/z greater than 1500. Quadrupole design requires the balancing multiple parameters to optimize ion throughput, mass filtering and, in some instruments, ion polarity switching. The quadrupole mass filters used in the 6400 Series and 6500 Series LC/MS systems are composed of four eight-inch hyperspacing between opposite rods. This spacing is known as the 2R0 (two R zero) spacing. Wider 2R0 spacing improves ion acceptance and transmission. However, the wider the 2R0 spacing, the greater the voltages that must be applied to the rods to achieve the proper mass filtering. Among other consequences, higher voltages make it harder to quickly switch ion polarity. The quadrupole mass filters in the 6400 Series and 6500 Series LC/MS system have a 2R0 of 8.89 mm. The hyperbolic shape of the quadrupole rods extends the hyperbolic area of the electrical fields the area in which the ions are transmitted all the way to the rods. This results in a larger effective 2R0 without the voltage and polarity switching disadvantages of a physically larger 2R0 spacing. Figure 7. The spacing between opposite quadrupole rods called 2R0 has a significant influence on ion throughput. The hyperbolic shape of the rods in the Agilent quadrupoles creates a larger effective 2R0 without incurring the disadvantages of a physically larger 2R0. 6

Post-Q1 filter The post-q1 filter (Figure 8), also called the beam shaper, is a radio-frequency (RF) ion guide that is capacitively coupled to Q1. It focuses the ion beam and guides the ions into the highpressure collision cell, ensuring that as many ions as possible enter the collision cell. The filter helps counter the effects of the flow of collision gas that is exiting the collision cell. Collision cell The collision cell (Figure 9) is a hexapole assembly. Nitrogen or argon collision gas (user s choice) is gently admitted into the cell at approximately 0.5 ml/min to avoid turbulence that could affect ion transmission. The collision cell operates at a relatively high pressure, which helps eliminate noise by reducing the number of excited neutrals and clusters that are transmitted through the cell by collisions that occur in the cell. Triple Quadrupole In early mass spectrometers that employed two quadrupole mass filters, the collision cell between them was frequently also a quadrupole. Thus, this type of hyphenated mass spectrometer was called a triple quadrupole mass spectrometer. The name became somewhat traditional and is frequently used for any mass spectrometer that uses a quadrupole mass filter collision cell quadrupole mass filter configuration, even if the collision cell is not a quadrupole. Figure 8. The post-q1 filter (beam shaper) improves transmission of ions from the quadrupole mass filter to the collision cell. Figure 9. Linear axial acceleration in the octopole collision cell facilitates the use of higher collision gas pressure and reduces carryover that could cause cross-experiment memory effects. www.agilent.com/chem/ms 7

Ion optics innovations for increased sensitivity in hybrid MS systems To avoid ion stall out, a negative consequence of a high collision gas pressure, a voltage potential is applied along the length of the resistively coated hexapole rods. This creates linear axial acceleration that maintains ion momentum. This allows more MRM experiments to be performed in a given time, and reduces cross-experiment memory effects due to ion latency. The energy of the ions entering the collision cell, the collision gas pressure, and the linear voltage gradient are balanced so that ions of all masses exit the collision cell with a negligible (almost 0 ev), but nearly uniform energy. In the case of the Q-TOF instruments, this allows the same mass calibration factors to be applied to MS and MS/MS ions. The result is better than 5-ppm mass accuracy for MS/MS ions. Components exclusive to triple quadrupole Ion detection system The triple quadrupole LC/MS systems employ a high-energy-dynode electron multiplier for ion detection (Figure 11). Ions first pass through an iris lens whose voltage is ramped according to the mass to improve focusing of ions into the detector. The detector incorporates two dynodes that improve the ion-to-electron yield. In positive mode, the conversion (primary) dynode operates at 10 KV. Positive ions impact the conversion dynode and generate secondary electrons. These electrons are then attracted to the 1.5 KV matching (secondary) dynode. Pre-Q2 filter Similar to the post-q1 filter, the pre-q2 filter (Figure 10) is a radio-frequency (RF) ion guide that is capacitively coupled to the second quadrupole mass filter (Q2). Its transmit ions from the collision cell to Q2 with minimum loss while allowing the vacuum system to pump away neutral collision gas molecules exiting the collision cell. Quadrupole mass filter 2 In the 6400 Series triple quadrupole LC/MS systems, the second quadrupole mass filter (Q2) is identical to the first (Figure 10). It uses hyperbolic rods to increase the effective 2R0 and maximize ion throughput, and hence sensitivity, without incurring the voltage and polarity switching disadvantages of a physically larger 2R0 spacing. Figure 10. In the triple quadrupole, the second quadrupole mass filter is identical to the first. It uses hyperbolic rods to maximize ion throughput, and hence sensitivity. Figure 11. The triple quadrupole detector is a continuous dynode electron multiplier. Since the conversion dynode, matching dynode, and electron multiplier horn are all off axis, neutral molecules miss the detector, eliminating neutral noise. 8

Additional electrons are generated and cascade into the electron multiplier. Since the high-energy conversion dynode, matching dynode, and electron multiplier are all off the main axis of the ion path, neutral molecules miss the detector, eliminating neutral noise. This allows the detector to operate at higher gains for a greater signal without simultaneously increasing noise. A typical gain for the entire detection system is 10 6 to 10 7. In negative ion mode, the conversion dynode operates at +10 kv. Negative ions impact the conversion dynode, producing secondary positive ions. These positive ions are attracted to the matching dynode at 1.5 kv. They impact the matching dynode and produce secondary electrons. These electrons are then directed to the electron multiplier. Gain characteristics are similar to those for positive ions. Detection electronics The electron multiplier current is sampled using an analog-to-digital converter with a logarithmic (log) amplifier. The log amplifier and the rest of the electronics are specifically designed to minimize electronic noise. filter requires a lower pressure for optimum operation than does a quadrupole mass filter. After the collision cell, the Q-TOF vacuum system differs significantly from the vacuum system in the triple quadrupole. Its higher pumping capacity maintains the lower pressure required by a TOF mass analyzer. The Q-TOF vacuum system uses two split-flow turbomolecular pumps to obtain the 10 7 Torr pressure required for TOF operation. Ion focus lens The ion focus lens (Figure 12) is a round, flat lens with a DC voltage applied to it. This voltage reaccelerates the ions, whose velocities are minimal after exiting the collision cell. Quadrupole ion guide The quadrupole ion guide (Figure 12) is a 20-mm section of quadrupole. Different voltages are applied to the two opposing rod pairs. The resulting electromagnetic fields reduce the velocity in the vertical axis and flatten the ion beam to better match it to the shape of the entrance slit in the slicer. Slicer The slicer (Figure 13) is a series of horizontal slits that prevent ions with excess velocity in the vertical axis from reaching the time-of-flight mass analyzer. By eliminating ions with excess vertical velocity, the quadrupole ion guide and slicer improve mass accuracy and especially resolving power (better than 20,000 at m/z 1,500). Time-of-flight mass analyzer The 6500 Series Q-TOF LC/MS systems employ a time-of-flight (TOF) mass analyzer as the second mass analyzer. A TOF mass analyzer delivers a higher percentage of ions over its complete mass range to its detector than most other types of mass analyzers. This translates to better sensitivity when full scan spectra are needed. The TOF mass analyzer in the 6500 Series uses orthogonal acceleration and a two-stage ion mirror (reflectron). Components exclusive to Q-TOF Octopole ion guide 2 Instead of the pre-q2 filter in triple quadrupole instruments, the Q-TOF instruments have a second octopole ion guide almost identical to the first (Figure 12). This second octopole ion guide maintains maximum ion transmission while neutral collision gas escaping the collision cell is pumped away by the vacuum system. This is essential since the time-of-flight mass Figure 12. In the Q-TOF instruments, the post-collision-cell ion optics help with removal of the collision gas, shape the ion beam, and eliminate ions with non-uniform vertical velocity to improve the mass accuracy and mass resolution of the time-of-flight mass filter. www.agilent.com/chem/ms 9

Ion optics innovations for increased sensitivity in hybrid MS systems Pulser Pulsed acceleration orthogonal to the axis of the ion source and ion optics (Figure 14) is particularly advantageous for systems that generate a continuous beam of ions. It provides a very precise start to the flight time and, with the quadrupole ion guide and slicer, a much narrower range of initial energies in the ions. The result is better mass resolution and mass accuracy. 6500 Series Q-TOFs uses pulsed orthogonal acceleration at default frequency of approximately 14 KHz (mass range dependent) to deliver as many ions to the detector as possible. Figure 13. The slicer prevents ions with non-uniform velocity in the vertical axis from reaching the time-of-flight mass filter. This improves mass accuracy and resolving power. Ion mirror Ion mirrors improve mass resolution while allowing a smaller instrument due to the folded flight path. The 6500 Series uses a two-stage ion mirror, also called a reflectron, with a harp design that improves resolution and sensitivity. The ion mirror self-compensates for thermal changes. Low-expansion flight tube The flight tube is constructed from a metal alloy with an extremely low coefficient of thermal expansion in the operating temperature range of the Q-TOF. This minimizes changes in flight distance due to changes in laboratory temperature and helps maintain the instrument s superlative mass accuracy. Figure 14. When combined with the quadrupole ion guide and slicer, pulsed acceleration orthogonal to the major axis of the ion optics results in better mass resolution and mass accuracy. Detector The QTOF detection system uses a microchannel-plate detector (Figure 15). The ion beam strikes the microchannel plate and generates a stream of electrons. The electrons strike a scintillator, Figure 15. The Q-TOF uses a microchannel-plate detector that provides high gain and low noise for better sensitivity. The Q-TOF detection system also features a wide dynamic range. 10

producing photos. The photons strike a photomultiplier tube, generating an amplified signal proportional to ion flux. The photomultiplier features high gain with low noise. Because it is isolated from the ions, it also has a long lifespan The dynamic range of the detection system is greater than 10 4. Detector electronics Similar to the triple quadrupole systems, the Q-TOF uses an analog-to-digitalconversion (ADC) -based detection system. An analog sampling system avoids many of the ion current saturation and mass assignment inaccuracies associated with a time-to-digital (TDC) detector systems. 1 The detector signal is sampled at a 4 GHz frequency (32 Gbit/sec). This high sampling rate improves the resolution, mass accuracy and accurate abundance measurement of the TOF, especially at lower masses since the low-mass ions are traveling at greater velocities and require more frequent sampling for accuracy. Conclusion Maximizing sensitivity in a mass spectrometer requires optimizing the generation, transmission, and detection of ions while simultaneously minimizing the generation, transmission, and detection of chemical and electronic noise. In Agilent 6400 Series Triple Quadrupole LC/MS systems and Agilent 6500 Series Q-TOF LC/MS systems, the entire ion path has been designed with these requirements in mind from ion generation in the ion source, through ion fragmentation and mass filtering, to ion detection and signal processing. The result is instruments that achieve superlative sensitivity performance without sacrificing other aspects of MS performance (Figures 16 and 17). x10 1 4 MRM 271.2 -> 145.1 Noise (PeakToPeak) = 4.00; SNR (0.874min) = 9.5 0.874 147 17β-Estradiol Limit of Detection: 500 fg on column, negative ion mode with 11% RSD 3 2 1 0 x10 1 4 MRM 271.2 -> 145.1 Noise (PeakToPeak) = 5.00; SNR (0.847min) = 7.8 0.847 152 3 2 1 0 x10 1 5 4 MRM 271.2 -> 145.1 Noise (PeakToPeak) = 4.00; SNR (0.876min) = 10.5 0.876 180 3 2 1 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1 1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8 1.9 2 Abundance vs. Acquisition Time (min) Figure 16. Analysis of 17β-estradiol using the 6410 Triple Quadrupole LC/MS demonstrates outstanding sensitivity and repeatability, which are due in large part to unique design features of the ion optics. www.agilent.com/chem/ms 11

Abundance x10 2 1.8 1.6 1.4 1.2 1 0.8 0.6 0.4 0.2 0 86.09630 72.08054 185.16380 218.14626 244.09567 297.17728 327.20248 357.17580 388.30220 431.1809 459.21972 494.26657 527.24317 595.29565 MS/MS spectrum m/z 582.31920 -> ** 500 amol BSA on column 100 200 300 400 500 600 700 800 900 1000 1100 1200 Mass-to-charge (m/z) 708.39349 770.34811 837.44537 951.47825 1063.50869 Figure 17. Analysis of 500 amol of BSA injected on column demonstrates the superb sensitivity of the 6510 Quadrupole Timeof-Flight LC/MS. Protein database searching on this data provided positive identification. References 1. John Fjeldsted, "Time-of-Flight Mass Spectrometry," Agilent Technical Overview, Publication Number 5989-0373EN, 2003. About Agilent Technologies Agilent Technologies is a leading supplier of life science research systems that enable scientists to understand complex biological processes, determine disease mechanisms, and speed drug discovery. Engineered for sensitivity, reproducibility, and workflow productivity, Agilent's life science solutions include instrumentation, microfluidics, software, microarrays, consumables, and services for genomics, proteomics, and metabolomics applications. Learn more: www.agilent.com/chem/ms Buy online: www.agilent.com/chem/store Find an Agilent customer center in your country: www.agilent.com/chem/contactus U.S. and Canada 1-800-227-9770 agilent_inquiries@agilent.com Europe info_agilent@agilent.com Asia Pacific adinquiry_aplsca@agilent.com Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance or use of this material. Agilent Technologies, Inc. 2007 Printed in the U.S.A. 5989-7408EN