Basic Components of Spectroscopic. Instrumentation

Similar documents
CHAPTER 7. Components of Optical Instruments

Spectrophotometer. An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer :

Spectroscopy in the UV and Visible: Instrumentation. Spectroscopy in the UV and Visible: Instrumentation

Chemistry Instrumental Analysis Lecture 7. Chem 4631

The equipment used share any common features regardless of the! being measured. Electronic detection was not always available.

Components of Optical Instruments 1

SCCH 4: 211: 2015 SCCH

Components of Optical Instruments. Chapter 7_III UV, Visible and IR Instruments

Ultraviolet Visible Infrared Instrumentation

Components of Optical Instruments

Components of Optical Instruments

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region

Chemistry Instrumental Analysis Lecture 10. Chem 4631

DETECTORS Important characteristics: 1) Wavelength response 2) Quantum response how light is detected 3) Sensitivity 4) Frequency of response

The FTNIR Myths... Misinformation or Truth

University of Wisconsin Chemistry 524 Spectroscopic Components *

Chemistry 524--"Hour Exam"--Keiderling Mar. 19, pm SES

Infrared Detectors an overview

Observational Astronomy

CONFIGURING. Your Spectroscopy System For PEAK PERFORMANCE. A guide to selecting the best Spectrometers, Sources, and Detectors for your application

NIR SPECTROSCOPY Instruments

Period 3 Solutions: Electromagnetic Waves Radiant Energy II

CHAPTER 4. Optical Sources,Transducers,and Measurement Systems

THROUGHPUT OF AN OPTICAL INSTRUMENT CHEM 314

Radiation transducer. ** Modern electronic detectors: Taking the dark current into account, S = kp + bkgnd over the dynamic range.

Modern Instrumental Methods of Analysis Prof. Dr. J.R. Mudakavi Department of Chemical Engineering Indian Institute of Science, Bangalore

COMPONENTS OF OPTICAL INSTRUMENTS. Chapter 7 UV, Visible and IR Instruments

COMPONENTS OF OPTICAL INSTRUMENTS. Topics

OPTI510R: Photonics. Khanh Kieu College of Optical Sciences, University of Arizona Meinel building R.626

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

Guide to SPEX Optical Spectrometer

Radiation detectors Photographic detection

Thermo Scientific icap 7000 Plus Series ICP-OES: Innovative ICP-OES optical design

Ground-based optical auroral measurements

Section 1: SPECTRAL PRODUCTS

LlIGHT REVIEW PART 2 DOWNLOAD, PRINT and submit for 100 points

SPECTRAL IRRADIANCE DATA

GLOSSARY OF TERMS. Terminology Used for Ultraviolet (UV) Curing Process Design and Measurement

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26

UV-VIS-IR Spectral Responsivity Measurement System for Solar Cells

Page 1. Ground-based optical auroral measurements. Background. CCD All-sky Camera with filterwheel. Image intensifier

Lecture 18: Photodetectors

Spark Spectral Sensor Offers Advantages

08-2 EE 4770 Lecture Transparency. Formatted 16:41, 12 February 1998 from lsli Steradian. Example

Engineering Medical Optics BME136/251 Winter 2018

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

Nd: YAG Laser Energy Levels 4 level laser Optical transitions from Ground to many upper levels Strong absorber in the yellow range None radiative to

6 Experiment II: Law of Reflection

AP Chemistry Cell Phone Spectroscopy Lab Adopted from Alexander Scheeline Department of Chemistry University of Illinois at Urbana-Champaign

Maya2000 Pro Spectrometer

Lithography. 3 rd. lecture: introduction. Prof. Yosi Shacham-Diamand. Fall 2004


The Photoelectric Effect

Photoelectric effect

Part I. Experimental Investigation

A Software Implementation of Data Acquisition Control and Management for Czerny Turner Monochromator

Exercise 8: Interference and diffraction

Instrumentation. 1) Sources 2) Wavelength Selectors 3) Detectors Recommended Reading: Applied Spectroscopy

Measurement of the Speed of Light in Air

Absentee layer. A layer of dielectric material, transparent in the transmission region of

David Jameson 4/4/2018

Doppler-Free Spetroscopy of Rubidium

Improving the Collection Efficiency of Raman Scattering

A Novel Multipass Optical System Oleg Matveev University of Florida, Department of Chemistry, Gainesville, Fl

UVS-2800 Spectro UV-VIS Split Beam (PC) is a precise scanning

RANDY W. ALKIRE, GEROLD ROSENBAUM AND GWYNDAF EVANS

Experimental Analysis of Luminescence in Printed Materials

NanoSpective, Inc Progress Drive Suite 137 Orlando, Florida

Systems & Accessories

Physics 308 Laboratory Experiment F: Grating Spectrometer

LAB ASSIGNMENT No. 1 Characteristics of IC Temperature Sensor (LM 335)

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science

Fundamentals of Laser

combustion diagnostics

Terahertz spectroscopy measurements

Introduction to the operating principles of the HyperFine spectrometer

PHYSICS. Speed of Sound. Mr R Gopie

Lecture 8 Optical Sensing. ECE 5900/6900 Fundamentals of Sensor Design

Operating Manual. Model 721N. Visible Spectrophotometer

Laser Telemetric System (Metrology)

PART XI: DETECTION OF RADIATION (IUPAC Recommendations 1995) (Originally authored by K. Laqua, B. Schrader, D. S. Moore, and T.

Chapter 3 OPTICAL SOURCES AND DETECTORS

Ph 3455 The Photoelectric Effect

Absorption: in an OF, the loss of Optical power, resulting from conversion of that power into heat.

TIME-PRESERVING MONOCHROMATORS FOR ULTRASHORT EXTREME-ULTRAVIOLET PULSES

Light has some interesting properties, many of which are used in medicine:

EXPERIMENT 3 THE PHOTOELECTRIC EFFECT

Section 18.3 Behavior of Light

DIODE LASER SPECTROSCOPY (160309)

Photons and solid state detection

Uses of Electromagnetic Waves

LE/ESSE Payload Design

Section 1: Sound. Sound and Light Section 1

Test 1: Example #2. Paul Avery PHY 3400 Feb. 15, Note: * indicates the correct answer.

Principles of Fluorescence Techniques 2014 Urbana-Champaign, Illinois April 7-10, 2014

David Jameson 3/23/2015

The Nature of Light. Light and Energy

HR2000+ Spectrometer. User-Configured for Flexibility. now with. Spectrometers

SCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS

LIGHT AND LIGHTING FUNDAMENTALS. Prepared by Engr. John Paul Timola

Transcription:

Basic Components of Spectroscopic Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11451 Saudi Arabia Building: 05, Office: AA53 Tel. 014674198, Fax: 014675992 Web site: http://fac.ksu.edu.sa/aifseisi E-mail: ahmad3qel@yahoo.com aifseisi@ksu.edu.sa Instrumentation

Basic components of spectroscopic instruments - A source of energy that can be input to the sample. - A means for isolating a narrow range of wavelengths. - Sample container. - A detector for measuring the signal. - A signal processor to display the signal in a form convenient for the analyst.

Sources of Energy The sources of energy could be in different forms: (1) Sources of electromagnetic radiation (photons) A source of electromagnetic radiation must provide an output that is both intense and stable in the desired region of the electromagnetic spectrum. The energy levels have well-defined values (i.e., they are quantized). Absorption only occurs when the photon s energy matches the difference in energy, ΔE, between two energy levels. Simplified energy level diagram showing absorption of a photon ΔE = E 1 - E 0

Sources of electromagnetic radiation are classified as: - Continuum source: a source that emits radiation over a wide range of wavelengths. - Line source: a source that emits radiation at only select wavelengths (narrow wavelength ranges). A list of the most common sources of electromagnetic radiation. Source Wavelength region Useful for H 2 and D 2 lamp continuum source from 160 380 nm UV molecular absorption Tungsten (W) lamp continuum source from 320 2400 nm Vis molecular absorption Xe arc lamp continuum source from 200 1000 nm molecular fluorescence Nernst glower continuum source from 0.4 20 µm IR molecular absorption Globar continuum source from 1 40 µm IR molecular absorption Nichrome wire continuum source from 0.75 20 µm IR molecular absorption Hollow cathode lamp line source in UV/Vis atomic absorption Hg vapor lamp line source in UV/Vis molecular fluorescence Laser line source in UV/Vis atomic and molecular absorption, fluorescence and scattering

(2) Sources of thermal energy The most common sources of thermal energy are flames and plasmas. -Flame sources use the combustion of a fuel and an oxidant such as acetylene and air, to achieve temperatures of 2000 3400 K. -Plasmas, which are hot, ionized gases, provide temperatures of 6000 10000 K. (3) Chemical sources of energy Exothermic reactions also may serve as a source of energy. In chemiluminescence the analyte is raised to a higher-energy state by means of a chemical reaction, emitting characteristic radiation when it returns to a lowerenergy state. When the chemical reaction results from a biological or enzymatic reaction, the emission of radiation is called bioluminescence. Commercially available light sticks, glow and the flash of light from a firefly are examples of chemiluminescence and bioluminescence, respectively.

Wavelength Selection Wavelength selector is a component used to select and isolate the required wavelengths or range of wavelengths where the analyte is the only absorbing species (to obtain a certain wavelength or a narrow band of wavelengths). The ideal wavelength selector has a high throughput of radiation and a narrow effective bandwidth. A high throughput is desirable because more photons pass through the wavelength selector, giving a stronger signal with less background noise. A narrow effective bandwidth provides a higher resolution, with spectral features separated by more than twice the effective bandwidth being resolved.

Unfortunately, we cannot isolate a single wavelength of radiation from a continuum source. Instead, a wavelength selector passes a narrow band of radiation characterized by a nominal wavelength, an effective bandwidth, and a maximum throughput of radiation. The effective bandwidth is defined as the width of the radiation at half the maximum throughput. Nominal wavelength: the wavelength which a wavelength selector is set to pass. Effective bandwidth: the width of the band of radiation passing through a wavelength selector measured at half the band s height. Generally these two features of a wavelength selector are in opposition. Conditions favoring a higher throughput of radiation usually provide less resolution. Decreasing the effective bandwidth improves resolution, but at the cost of a noisier signal. For a qualitative analysis, resolution is generally more important than the throughput of radiation; thus, smaller effective bandwidths are desirable. In a quantitative analysis a higher throughput of radiation is usually desirable.

There are two types of wavelength selector: 1. Filters 2. Monochromators Filters Filters are the simplest devices, consisting of only a material that selectively transmits the desired wavelengths and absorbs all other wavelengths. Filters use either an absorption or interference of radiation. -Absorption filters work by selectively absorbing radiation from a narrow region of the electromagnetic spectrum. A simple example of an absorption filter is a piece of colored glass. -Interference filters use constructive (in phase) and destructive (out of phase) interference to isolate a narrow range of wavelengths. Interference filters are more expensive than absorption filters, but have narrower effective bandwidths

Monochromators If measurements need to be made at two different wavelengths, then the filter must be changed in between measurements. absorption or interference filter do not allow for a continuous selection of wavelength. Monochromators can give a much narrower range of wavelengths and are easily adjustable over a wide spectral range (scan spectrum, vary continuously). The dispersing element may be a prism or a grating. The radiation is collected by a collimating mirror, which reflects a parallel beam of radiation to a diffraction grating. The diffraction grating is an optically reflecting surface with a large number of parallel grooves. Diffraction by the grating disperses the radiation in space, where a second mirror focuses the radiation onto a planar surface containing an exit slit. As a result, a polychromatic source of radiation at the entrance slit is converted at the exit slit to a monochromatic source of finite effective bandwidth. The choice of which wavelength exits the monochromator is determined by rotating the diffraction grating. A narrower exit slit provides a smaller effective bandwidth and better resolution, but allows a smaller throughput of radiation.

Types of monochromators: Collimating mirror Focusing mirror Grating monochromator Prism monochromator Light source

Interferometers An interferometer provides an alternative approach for wavelength selection. Instead of filtering or dispersing the electromagnetic radiation, an interferometer simultaneously allows source radiation of all wavelengths to reach the detector. Michelson interferometer (beam splitter) One fixed mirror and the other mirror is moving sample

Radiation from the source is focused on a beam splitter that transmits half of the radiation to a fixed mirror, while reflecting the other half to a movable mirror. The radiation recombines at the beam splitter, where constructive and destructive interference determines, for each wavelength, the intensity of light reaching the detector. As the moving mirror changes position, the wavelengths of light experiencing maximum constructive interference and maximum destructive interference also changes. The signal at the detector shows intensity as a function of the moving mirror s position, expressed in units of distance or time. The result is called an interferogram, or a time domain spectrum. The time domain spectrum is converted mathematically, by a process called a Fourier transform, to the normal spectrum (also called a frequency domain spectrum) of intensity as a function of the radiation s energy. Since it does not use slits and has fewer optical components from which radiation can be scattered and lost, interferometers provide higher throughput of source radiation than monochromators. Furthermore, since all frequencies are monitored simultaneously, an entire spectrum can be recorded at significantly lower time in compared with monochromators.

Sample Containers (Holders) The cells or cuvettes that hold the samples must be made of material that is transparent to radiation in the spectral region of interest. Sample containers should also in compatible with the sample type, size and state (solid, solution and gas). Quartz or fused silica is required for work in the ultraviolet region (below 350 nm), both of these substances are transparent in the visible region. Silicate glasses can be employed in the region between 350 and 2000 nm. Plastic containers can be used in the visible region. Crystalline NaCl is the most common cell windows in the IR region.

Detectors The first detector for optical spectroscopy was the human eye (limited to Visible region, limited accuracy and sensitivity to electromagnetic radiation). Modern detectors use a sensitive transducer to convert a signal consisting of photons into an easily measured electrical signal. Transducer: a device that converts a chemical or physical property, such as ph or photon intensity, to an easily measured electrical signal, such as a voltage or current. Device converts a signal in one form of energy to another form of energy.

The photoelectric effect is the observation that many metals (such as alkali metals or semiconductor material such as gallium arsenide) emit electrons when light shines upon them. Electrons emitted in this manner can be called photoelectrons. Incoming light Ejected electrons Metal

Photon transducer Phototubes and photomultipliers contain a photosensitive surface that absorbs radiation in the ultraviolet, visible, and near infrared (IR), producing an electric current proportional to the number of photons reaching the transducer. Other photon detectors use a semiconductor as the photosensitive surface. When the semiconductor absorbs photons, valence electrons move to the semiconductor s conduction band, producing a measurable current. Thermal transducer Infrared radiation generally does not have sufficient energy to produce a measurable current when using a photon transducer. A thermal transducer, therefore, is used for infrared spectroscopy. A thermal detector consists of a tiny blackened surface that absorbs infrared radiation and increases in temperature as a result. The temperature rise is converted to an electrical signal that is amplified and measured.

Phototube Vacuum phototube, which consists of a semicylindrical cathode and a wire anode sealed inside an evacuated transparent envelope. The concave surface of the electrode supports a layer of photoemissive material that tends to emit electrons when it is irradiated. When a voltage is applied across the electrodes, the emitted electrons flow to the wire anode generating a photocurrent that is generally about one tenth as great as that associated with a photovoltaic cell for a given radiant intensity. dynodes Photomultiplier tubes A photon of radiation entering the tube strikes the cathode, causing the emission of several electrons. These electrons are accelerated towards the first dynode (which is 90V more positive than the cathode). The electrons strike the first dynode, causing the emission of several electrons for each incident electron. These electrons are then accelerated towards the second dynode, to produce more electrons which are accelerated towards dynode three and so on. Eventually, the electrons are collected at the anode. Each original photon produces about 10 6-10 7 electrons.

Photodiode array: a linear array of photodiodes providing the ability to detect simultaneously radiation at several wavelengths. A special feature of diode array detectors is the ability to perform spectroscopic scanning and precise absorbance readings at a variety of wavelengths while the peak is passing though the flow cell. UV detector Diode array detector Photodiode array allows for the best wavelength(s) to be selected for actual analysis. This is particularly important when no information is available on molar absorptivities at different wavelengths. The second major advantage is related to the problem of peak purity, absorbance rationing at several wavelengths is particularly helpful in deciding whether the peak represents a single compound or, is in fact, a composite peak.

Detector Wavelength range Output signal Photon detectors Phototube 200 1000 nm current Photomultiplier 110 1000 nm current Si photodiode 250 1100 nm current Photoconductor 750 6000 nm change in resistance Photovoltaic cell 400 5000 nm current or voltage Thermal detectors Thermocouple 0.8 40 µm voltage Thermistor 0.8 40 µm change in resistance Pneumatic 0.8 1000 µm membrane displacement Pyroelectric 0.3 1000 µm current

Signal Processors and Readout A device that displays the signal from the transducer in a form that is easily interpreted by the analyst. The signal processor also may be used to control various instrumental parameters, to calibrate the detector s response, to amplify the signal from the detector, to remove noise by filtering, or to mathematically transform the signal. Examples of signal processors include analog or digital meters, recorders and computers equipped with digital acquisition boards.

Basic Components of Spectroscopic Instrumentation Source of light Wavelength selector Sample Detector Signal processor