Wilcom MODEL T136BSBZW CIRCUIT TEST SET. Operating Instructions

Similar documents
Wilcom MODEL T136BGMZW CIRCUIT TEST SET. Operating Instructions

Wilcom MODEL T336B CIRCUIT TEST SET. Operating Instructions

Model 113 Scintillation Preamplifier Operating and Service Manual

Wilcom. T-304B Current Meter OPERATING INSTRUCTIONS

SX204. Headphone Amplifier. Symetrix. Owner s Manual

Model 7000 Low Noise Differential Preamplifier

DA208 & DA416 DISTRIBUTION AMPLIFIERS OPERATING AND MAINTENANCE MANUAL

Glass Electrode Meter

Far End Device II Model 1342

Model 9305 Fast Preamplifier Operating and Service Manual

SI-125 Power Amplifier Manual 6205 Kestrel Road; Mississauga, Ontario; Canada; L5T 2A1 November 2016, Rev 0.5

305 Stimulus Isolator. Description

Model 935A. Dual Tone Sender INSTRUCTION MANUAL

This manual should be read in its entirety prior to operation of the TS200 Test Set Page 2 of 9

INSTRUCTION MANUAL MODEL 2779 SUBCARRIER MODULATOR

Advanced Test Equipment Rentals ATEC (2832)

2001A. 200KHz Function Generator Instruction Manual. 99 Washington Street Melrose, MA Phone Toll Free

INSTRUCTION MANUAL MODEL 2455T SUBCARRIER MODULATOR

UNCONTROLLED WHEN PRINTED

Angular Rate Sensor. Owner's Manual

PCO-7114 Laser Diode Driver Module Operation Manual

Model 1792 VHF Radio. User s Manual. Rev. B FAA APPROVED NOT FAA APPROVED. ECP Dec 30

ADA416-XLR DISTRIBUTION AMPLIFIERS OPERATING AND MAINTENANCE MANUAL

Pen Multimeter. Model

SCREENING EFFICIENCY TESTER Model 271

DL102 Counter Loop Amplifier

AM-510 Commercial / Residential Multimeter. AM-510-EUR Digital Multimeter. Users Manual

LUDLUM MODEL MODEL AND MODEL GAMMA SCINTILLATORS. June 2017

HHMA2 DC / TRUE RMS AC NON-CONTACT MILLIAMMETER

Series 3000 Model R-107A

MaxLite LED Self-Driven LiteBars

AMP-12 OPERATOR S MANUAL

INSTRUCTION MANUAL. Model Autoranging DMM ProbeMeter TM. Measures voltage, resistance, frequency, capacitance, temperature, and duty cycle.

TIA-527 Balanced Optical/Electrical Converter. Operating Instructions

Synthesized Base Station Transmitter

Instruction Manual Model M Switch, DPDT, Manual Select

SPM-50 RF Spectrum Power Meter PC Software User Manual

TETRIS 1000 High Impedance Active Probe. Instruction Manual

40 Amp Digital Bidirectional PWM Motor Controller with Regenerative Braking BIDIR-340-DR

3M Dynatel Far End Device III

99 Washington Street Melrose, MA Fax TestEquipmentDepot.com # # AAC Clamp Meter. Instruction Manual

User s Guide ASSISTIVE LISTENING SYSTEMS

Amplified High Speed Photodetectors

AMP-13 OPERATOR S MANUAL

Classic Series Public Address Amplifiers C10 & C20 Models

INSTALLATION AND MAINTENANCE MANUAL FOR GROUND MONITOR GM-250 COPYRIGHT 1983 AMERICAN MINE RESEARCH, INC.

Pen Multimeter. Model

User s Guide. 400A AC/DC Clamp Meter. Model MA220

MICROTOOLS MICRONETBLINK KIT

Document Version 1.2

Model 1791 VHF Radio User's Manual

Model 863 Quad Timing Filter Amplifier Operating and Service Manual

V-136RTHF 36 ZONE TALKBACK INTERCOM/ PAGE CONTROL UNIT

34134A AC/DC DMM Current Probe. User s Guide. Publication number April 2009

TRANSDUCER IN-LINE AMPLIFIER

Model 416A Gate and Delay Generator Operating and Service Manual

Model 935A Current Source Operation Manual

TLS-3A. Telephone Line Simulator. User Manual , Rev. B Covers Model TLS-3A-01

CALIBRATED IMPULSE GENERATOR MODEL CIG khz 1 GHz

DM-46 Instruction Manual

User s Guide. MultiView Series Digital MultiMeters Models: MV110 MV120 MV130

Model 34A. 3Hz to 2MHz 2-Channel Butterworth/Bessel HP, LP, BP, BR Plug-In Filter Card for Model 3905/3916 Chassis.

F272xB Fiber optic 4-Wire Telephone/Audio Modem Technical Manual

Distribution Amplifiers 1

Model 229 Series 2 Leakage Current Tester OPERATOR S MANUAL

Model 3116 Double-Ridged Waveguide Horn

23A TRANSMISSION MEASURING SET (J94023A) (DESCRIPTION)

User Guide. Digital AC/DC Clamp Meter Model 38394

Model DB Disc Caliper Brake AIR CHAMP PRODUCTS. User Manual. (i) MTY (81)

200Amp AC Clamp Meter + NCV Model MA250

Warning, refer to accompanying documents.

TIA-500 Optical/Electrical Converter Operating Instructions

User s Guide. Digital AC/DC Clamp Meter Model 38394

12B/18 MultiMeter Instruction Sheet

201AP Charge Amplifier User Manual

Dawson DDM190. Digital Multimeter User s Manual

LUDLUM MODEL 182 RADON FLASK COUNTER. June 2011 Serial Number PR and Succeeding Serial Numbers

G7K. Latching Relay. Compact Mechanical Lock Latching Relays with Manual Buttons. Model Number Structure. Ordering Information

41P Portable Calibrator User Manual

80i-600A AC Current Probe

DA560D COMPACT SERIES. INSTALLATION / OWNER'S MANUAL Mobile Power Amplifiers

SIR, RADAN and UtilityScan are registered trademarks of Geophysical Survey Systems, Inc..

Installation & Operation Manual SAGA1-K Series Industrial Radio Remote Control

MBC Bipolar Microstep Driver. User s Guide E. Landon Drive Anaheim, CA

Series 500. Owner s Manual. Analog Transmitters by Data Industrial. Data Industrial. Data Industrial 2/95 PN 72806

User Manual. ProRF Encoder Transmitter & Receiver

Model 3101, 3102 and 3103 Conical Log-Spiral Antennas

TIA-3000 Optical/Electrical Converter. Operating Instructions

User s Manual. Ground Resistance Clamp On Tester MODEL

HTA125A/250A. Power Amplifiers. Installation & Use Manual

TETRIS User's Guide. High Impedance Active Probe DO177-1

SSR-150xxx-40VL-12P-xC-xxxCS

PR-216. High Performance Personal Receiver PR-216 OPERATOR S MANUAL

DA604D DA954D DA501D DA801D COMPACT SERIES. INSTALLATION / OWNER'S MANUAL Mobile Power Amplifiers

MLA High Performance Microstepping Driver. User s Guide E. Landon Drive Anaheim, CA

Instruction Manual Model Upconverter

FRYCORDER Power Cord Burn-in Generator

TRIAXIAL FLUXGATE MAGNETOMETER OWNER'S MANUAL

model 101 single channel microphone preamplifier owner s manual Rev C

Instruction Manual Model Upconverter

Transcription:

Wilcom MODEL T136BSBZW CIRCUIT TEST SET Operating Instructions

T136BSB Current Test Set Operating Instructions 811-230-007 February 2007 Copyright (c) 2007 Wilcom All Rights reserved Wilcom reserves the right to make changes to the material contained herein without notice and shall not be liable for errors contained herein or for incidental, consequential damage in connection with the furnishing, performance or use of this material. This document may not be copied or duplicated in part or in whole for any purpose without the express written permission of Wilcom. i

Table of Contents Page Introduction... 1 Specifications... 6 Circuit Description... 9 Input Circuit... 9 Measuring Circuit Input Section... 9 Circuit Measuring... 10 Batteries & PUSH to MEAS... 11 Operating the T136BSB... 12 Operating Procedure... 12 Applications... 14 Performance Limits... 14 Measurements at the Subscriber s Location... 14 Measurements at the C.O. Main Frame... 15 Balance Measurements... 16 Maintenance... 17 Battery Replacement... 17 Service and Repair... 18 Warranty... 19 Figures 1. T136BSB Circuit Test Set... 5 2. C-Message Weighting Characteristics... 8 3. Data Pad...15 ii

Introduction Model T136BSBZW is portable, compact combination level and noise measuring set designed to facilitate circuit and noise measurements by installers and repair persons without using elaborate or expensive equipment. It is designed to measure loop current, circuit loss, noise metalic and noise to ground, each over a range which exceeds normally acceptable limits. Meter scale colors indicate normally acceptable, marginal and unacceptable divisions of measurements. LOOP ma (DC loop current) measurement indicates the dc current drawn by a resistance of approximately 430 Ω connected across the line. Provisions have been made for dialing and holding a remote line. In the DIAL & BATT TEST position of the FUNCTION switch, the line connectors are switched through to a pair of square terminal posts for connecting a dial telephone hand set. In all other positions of the FUNCTION switch, a hold coil is connected across the line. The hold circuitry has a dc resistance of approximately 430Ω with ac impedance high enough to have no significant effect on measurements. The FUNCTION switch position called CKT LOSS provides terminated level measurement of a tone from -15 dbm to +1 dbm. Measurement response characteristics are flat from 300 Hz to 15 khz; 60 Hz is attenuated more than 25 db as input impedance is 600 Ω. The FUNCTION switch position designated CKT NOISE provides additional sensitivity and frequency weighting for measuring noise metallic on a C-Message basis as input impedance is 600 Ω. 1

Introduction The FUNCTION switch position designated PWR INFL (Power Influence) measures noise voltages between a circuit and ground on a C-Message weighted basis. The input circuit in this position has an impedance of 100,000 Ω between either input terminal and ground. In most conventional measuring sets, the reference level for noise-toground measurement is 40 db above the 600 Ω dbrn reference voltage. The PWR INFL scale of the T136BSBZW is calibrated to give a direct reading in dbrnc (600 Ω voltage equivalent) so that balance measurements can be made directly with no need for the conventional 40 db correction. 2

Introduction The T136BSBZW has four separate meter scales: LOOP ma Range: 0 to 100 ma Color Segments: Red Green below 23 ma 23 ma to 100 ma CKT LOSS Range: +1 dbm full scale, calibrated in 1 db increments to -15 dbm Color Segments: Red Yellow Green Yellow below -10 dbm -10 dbm to -8 dbm -8 dbm to 0 dbm 0 dbm to +1 dbm CKT NOISE (Noise Metallic) Range: +33 dbrnc full scale, calibrated in 1 db increments to 15 dbrnc Color Segments: Green Yellow Red below 20 dbrnc 20 dbrnc to 30 dbrnc 30 dbrnc to 33 dbrnc PWR INFL (Noise-to-Ground) Range: +93 dbrnc (600 Ω voltage equivalent) full scale; calibrated in 1 db increments to 75 dbrnc Color Segments: Green below 80 dbrnc Yellow 80 dbrnc to 90 dbrnc Red 90 dbrnc to 93 dbrnc 3

Introduction The 136BSBZW has two momentary push button RANGE EXTENDER switches, arranged so that the PUSH TO MEAS button and one RANGE EXTENDER button can be operated together with one finger. The RANGE EXTENDER button above the PUSH TO MEAS button on the panel extends the measuring range by 10 db. It is designated - 10 db because its use requires that 10 db be added to the meter scale reading. Similarly, use of the RANGE EXTENDER button below the PUSH TO MEAS button (+10 db) requires that 10 db be added to the meter reading. When the FUNCTION switch is in any measure position, the DIAL posts enable the tester to listen to the signals being measured without interfering with the measurements. Three 4 ft. long leads equipped with clips and colored green for TIP, red for RING and yellow for GROUND are permanently attached to the set. A non-locking push button switch is used to supply power to the measuring circuits from two small 9 volt batteries. No current is supplied by the batteries until the switch is depressed to take a reading. Consequently, the useful life of the batteries can approach shelf life because the ON period is usually very short. 4

Introduction Figure 1. The T136BSBZW Circuit Test Set 5

Specifications Measuring Ranges DC current (LOOP ma): Signal level (CKT LOSS): Noise Metallic (CKT NOISE): Noise-to-Ground (PWR INFL): Range Extension: Input Circuits Impedance CKT LOSS and CKT NOISE meas: PWR INFL measurement: Frequency Response: Hold Impedance DC Resistance: 0 to 100 ma, through 430 Ω and diode bridge. Measurements may be made with either polarity DC on the line. Scale with log graduation, -15 dbm to +1 dbm with 1 db divisions. Scale graduated from 15 to 33 dbrnc with 1 db divisions. Scale graduated from 75 dbrnc (600 Ω voltage equivalent) to 93 dbrnc with 1 db divisions. Add 10 db and 10 db push button switches. Range extension applies to all level and noise measurements. 600 Ω AC; 430 Ω DC. 100 kω tip/ring to ground. CKT LOSS measurement, range from 300 Hz to 15 khz. Attenuation at 60 Hz is more than 25 db. CKT NOISE and PWR INFL measurements, C- MSG weighting. Refer to Figure 2. 430 Ω over a range of 0-120 ma 6

Specifications Controls and Connections Switches: Connections: Multi-position FUNCTION switch. Push button PUSH TO MEAS switch. Push button +10 db range extender. Push button -10 db range extender. DIAL terminals: square clip-on type. LEADS: three 4 foot leads with test clips and colored: Green for Tip, Red for Ring, and Yellow for Ground Power Battery Type: Battery operation, two 9 volt batteries, Eveready 216 or equivalent NEDA No. 1604. Physical Height: Width: Length: Weight (with batteries): 5 7/8 in. (15.0 cm) 4 3/8 in. (11.2 cm) 6 3/8 in. (16.2 cm) 3 lbs. 4 oz. (1.47 kg) 7

Specifications Figure 2. C-message Weighting Characteristic 8

The Model T136BSBZW is a basic transmission measurement set providing loop testing in accordance with industry standards. The following subsections of this manual describe the T136BSBZW measurement operation. Input Circuit The input switching circuit connects the input leads to the selected measurement circuitry. a. In DIAL & BATT TEST, the T and R leads are connected to the square DIAL terminals. b. In LOOP ma, the T and R leads are connected through a combined impedance of 430 Ω. c. In both CKT LOSS and CKT NOISE the line is connected through dc blocking capacitors to the terminating resistance and input circuity, while the 430 Ω hold impedance is shunted across the line to draw dc holding current. d. In PWR INFL, each side of the line is connected through a series 100 kω resistor to a blocking capacitor to ground. This provides the longitudinal path with proper attenuation for noise-to-ground measurement. Measuring Circuit Input Section Circuit Description The function of the measuring circuit input section is to provide isolation and line impedance matching. The Measuring Circuit DC input consists of a series/parallel circuit 9

switched through additional poles of the FUNCTION switch to a shunt capacitor, series resistor and the meter itself. The PUSH TO MEAS switch need not be used for DC measurements. The CKT LOSS and CKT NOISE input is DC blocked for 100 volts. The DC shunt effect of the 430 Ω impedance prevents the voltage across the line from ever exceeding this value, even when loop extenders are used. Following the DC blocking, the line is connected to a 600 Ω precision impedance. Circuit Measurements Circuit Description All measurements taken using the T136BSBZW are referenced to 600 Ω, 1 mw. In the CKT LOSS mode this means a tone at.775 volts will result in a 0 dbm response. All other readings are scaled to this according to the following formula: db = 20 log Vmes/.775 where Vmes = the amplitude of the signal measured in volts. NOISE MEASUREMENTS are referenced to 600 Ω, -90 dbm which is equivalent to 0 dbrn and then weighted via the C-Message curve (see Figure 2). A signal of.775 mv will produce a meter reading of 30 dbrnc. All readings are referenced to the formula: dbm = 90-20 log Vmes/.775 POWER INFLUENCE levels are calculated in the same manner as CKT NOISE, except voltages between tip and ring to ground are measured. The preceding formula applies. 10

Circuit Description Batteries and PUSH TO MEAS Switch The measuring set is powered by a pair of 9-volt batteries. Since the batteries supply power only while the PUSH TO MEAS switch is held, battery life approaches shelf life. Battery voltage is connected to the positive and negative supply busses of the measuring set through transistors which are normally cut off. Pressing the PUSH TO MEAS switch turns on both transistors, supplying positive and negative voltages simultaneously. With the FUNCTION switch in BATT TEST position, pushing the PUSH TO MEAS switch supplies power to the measuring circuits to load the batteries. At the same time it connects positive and negative voltages through to the meter to test for satisfactory battery voltage. 11

Operation Connections to the circuit to be tested are made with the clip leads on the end of the red, green and yellow leads which represent ring, tip and ground, respectively. CAUTION: GROUND should be connected first. A lineman s dial telephone set is connected by clipping to the terminals marked DIAL. Operating Procedure Place the FUNCTION switch in the DIAL position. Dial tone should be heard in the lineman s handset. Dial the numbert of a milliwatt generator in the C.O. After the connection is established, the line current may be checked in LOOP ma. Turn the FUNCTION switch to the CKT LOSS position and press the PUSH TO MEAS button. This turns the set on; it must be held in place until the measurement has been made. Readings are made on the CKT LOSS scale of the meter. Noise measurements are made by placing the FUNCTION switch in the DIAL position (dial tone should be heard in the lineman s handset); and then dialing the number of the quiet (or balance) termination at the Central Office. When a connection has been established, turn the FUNCTION switch to the CKT NOISE position and press the PUSH TO MEAS button. The reading on the CKT NOISE scale on the meter is the noise metallic on the line. If the meter should read off scale, operate the RANGE EXTENDER switch to 12

Operation bring the reading on scale. If the RANGE EXTENDER switch is used, the appropriate correction must be added to the meter reading. PWR INFL, or noise to ground measurements, are made while connected to the quiet termination at the C.O; turn the FUNCTION switch to the PWR INFL position. The PWR INFL meter scale is calibrated to read the noise to ground directly in dbrnc (600 Ω voltage equivalent). If the meter should read off-scale, the range extender switch may be used as before. It is possible to listen to the signal by connecting an earphone to the DIAL posts. A battery test may be made by turning the FUNCTION switch to the BATT TEST position and pressing the PUSH TO MEAS button. If the meter does not read in the GOOD area, the batteries should be replaced in accordance with instructions in the Maintenance section. WARNING: This test is NOT designed to be used with voltage in excess of ringing voltage. 13

The following describes the applications for which the T136BSBZW Circuit Test Set is useful, including noise measurements at the subscriber and C.O. locations, and balance measurements. Performance Limits Applications The measurement ranges of the T136BSBZW has been designed to be within the normally acceptable limits of circuit loss, noise metallic and noise to ground. The colored segments of the meter scale indicate the various degrees of performance acceptability when direct readings are made on the meter (i.e., when the range extender switch is not used). The significance of each color is as follows: GREEN: YELLOW: RED: Acceptable performance. Marginal performance. Investigation of potential troubles should be initiated. Unacceptable performance. Immediate corrective action required. If reading is off-scale high, without using range extender switch, when measuring CKT NOISE or PWR INFL, unacceptable performance is indicated. An optional data pad (Wilcom part # 06800023) is available to record each measurement. See Figure 3. Measurements at the Subscriber s Location Measurements may be made from the terminal block or protection block at the customer s premises. It is not necessary to disconnect the customer s equipment, but the equipment must be in the ON HOOK condition when measurements are made. The leads on the 14

Applications Figure 3. T136BSB Data Pad 15

Applications T136BSBZW are connected to Tip, Ring and Ground as outlined in the Operation section. A good ground, preferably the ground at the protector block, must be available to make power influence measurements. Measurements at the C.O. Main Frame Noise measurements may be made at the main frame in the C.O. toward the subscriber or toward the office equipment. To Subscriber meaningful noise measurements can only be made toward the subscriber when there is a termination at the far end. The heat coils should be removed for the test. If a section of the subscriber s loop is to be measured, a termination should be connected at the far end of the section. Unterminated measurements will not provide useful data for analyzing noise problems; however, if a large number of unterminated loops aretested, useful statistical information can be compiled on the extent of noise problems on loops out of a C.O. A 900 Ω resistor in series with a 2 µf capacitor is a common termination used at the subscriber end. To C.O. Equipment it may be desirable to determine how much noise originates in the C.O. and the balance of the office equipment. Remove the heat coils and connect the Green and Red leads to the Tip and Ring terminals toward the office equipment connect the Yellow lead to C.O. ground bus. Dial the quiet termination and make CKT NOISE and PWR INFL measurements. 16

If the induced longitudinal voltage is low, it may not be possible to make power influence measurements. It is then necessary to introduce an external longitudinal voltage at a fixed frequency; any signal source between 500 Hz and 1000 Hz is satisfactory. The Wilcom Model T279 is useful for providing a balanced terminating circuit to which a longitudinal voltage source can be connected. Balance Measurements Applications Circuit balance can be determined from the CKT NOISE and PWR INFL measurements both made with the T136BSBZW. Balance (db) = PWR INFL - CKT NOISE The balance of a circuit will determine how susceptible it is to the induction of noise. The following Balance values provide a guide for determining acceptable balance: BALANCE CONDITION Over 60 db Excellent 50-60 db Good 40-50 db Fair Under 40 db Poor Usually any balance under 50 db requires improvement, particularly when the power influence is high (i.e., over 75 dbrnc as measured on the T136BSBZW). If power influence is low, a greater degree of unbalance can be tolerated; however, if the power influence should increase, the practice is to get as high a balance as possible. 17

The T136BSBZW should require little, if any, maintenance except for replacing the batteries. If any difficulties arise in the performance condition, please contact Wilcom Products, Inc. for further information. Battery Replacement The T136BSBZW only draws current when the PUSH TO MEAS switch is pressed; consequently, the batteries will have close to shelf life. The batteries should be checked periodically by placing the FUNCTION switch in the BATT TEST position and pressing the PUSH TO MEAS button. If the meter does not read in the GOOD segment, the batteries should be replaced. If a battery replacement is indicated, rotate the four fasteners on the bottom of the case 1/4 turn counter clockwise and lift the panel from the case. Disconnect the batteries, pull the batteries out, insert new batteries and reconnect the terminals. Use Eveready No. 216 or equivalent; the NEDA No. is 1604. Replace the case hatch, press down and turn the screw fasteners 1/4 turn clockwise Repair Maintenance Should the T136BSBZW require service, the unit may be sent to Wilcom at the address at the front of this manual, in accordance with the warranty instructions. Should technical assistance be necessary, contact our Applications Engineers at the following number: 1-800-222-1898. 18

Service and Repair 1. Call Wilcom Customer Service. Support personnel will determine if the equipment requires service, repair or calibration. 2. If the equipment must be returned to Wilcom for service, Wilcom Customer Service will issue a Return Material Authorization (RMA) number and the following address for return: Wilcom 73 Daniel Webster Highway Belmont, NH 03220 TEL (800) 222-1898 (USA only) or (603) 524-2622 FAX (603) 524-3735 www.wilcominc.com IMPORTANT Never send any equipment back to Wilcom without a Return Material Authorization (RMA) number. 3. Pack the equipment in its original shipping material. Be sure to include a statement or report fully detailing the defect and conditions under which it was observed. Also be sure to include a contact name and telephone number. 4. Return the equipment, prepaid, to the above address. Be sure to write the RMA on the shipping slip. Wilcom will refuse and return any package that does not bear the RMA. Ordering Information Orders for any of the Wilcom products and any of their optional accessories should be directed to the address shown above. 19

Warranty All products are warranted against defects in materials and workmanship. This warranty applies for a period of two (2) years from date of delivery, except for Fiber Optic instrumentation and equipment which have a one (1) year warranty on parts and two (2) years on labor. (The only exception in the digital testing equipment is the D550 Shark, which has a warranty of one (1) year from date of delivery.) Wilcom's obligation under this warranty is limited to servicing or adjusting each instrument returned to its factory within the warranty period, and to replace any components found to be defective. If determined that the defective condition is a result of misuse or abnormal operation, repairs will be billed. LIMITATION OF WARRANTY The foregoing warranties are the exclusive warranties provided by Wilcom. Wilcom will not be liable for any special, indirect, incidental or consequential damages whatsoever resulting from loss of use, loss of data or loss of profits arising out of or in connection with the use or performance of the product, even if Wilcom has been informed of the possibility of such damages in advance. All implied warranties, including without limitation warranties of merchantability and fitness for a particular purpose, as well as warranties arising from a course of dealing or usage of trade are expressly disclaimed. PROPRIETARY INFORMATION The information contained in this manual is the proprietary material of Wilcom, and may not be reproduced, used for manufacturing purposes, or disclosed to others for any use without written permission from Wilcom. 20