A High-Speed Imaging Colorimeter LumiCol 1900 for Display Measurements

Similar documents
Imaging Photometer and Colorimeter

Ideal for display mura (nonuniformity) evaluation and inspection on smartphones and tablet PCs.

Achieving 100,000 : 1 contrast measurement

Color Measurement with the LSS-100P

Ideal for display mura (nonuniformity) evaluation and inspection on smartphones and tablet PCs.

Miniaturized Spectroradiometer

Cvision 2. António J. R. Neves João Paulo Silva Cunha. Bernardo Cunha. IEETA / Universidade de Aveiro

Multispectral. imaging device. ADVANCED LIGHT ANALYSIS by. Most accurate homogeneity MeasureMent of spectral radiance. UMasterMS1 & UMasterMS2

Spectroradiometer CS-2000/2000A. The world's top-level capability spectroradiometers make further advances with addition of second model to lineup.

High-Accuracy Luminance & Chromaticity Measurement Comparable to Many Spectroradiometers

Spectroradiometer CS-2000/2000A. The world's top-level capability spectroradiometers make further advances with addition of second model to lineup.

WP640 Imaging Colorimeter. Backlit Graphics Panel Analysis

CS-2000/2000A. Spectroradiometer NEW

We bring quality to light. LumiCam 1300 Imaging Photometer/Colorimeter

Image acquisition. In both cases, the digital sensing element is one of the following: Line array Area array. Single sensor

An imaging device for multispectral analysis in the visible range. P. Fiorentin, E. Pedrotti, A. Scroccaro

DISPLAY metrology measurement

1. Introduction 2. Objectivation of Inspections via Measurements 3. Practical Application of Micro Sensor Systems 4. Summary & Conclusions

POTENTIAL OF MULTISPECTRAL TECHNIQUES FOR MEASURING COLOR IN THE AUTOMOTIVE SECTOR

Ultimate and Versatile 2D Spectroradiometer is finally released!

MEASURING HEAD-UP DISPLAYS FROM 2D TO AR: SYSTEM BENEFITS & DEMONSTRATION Presented By Matt Scholz November 28, 2018

CS6640 Computational Photography. 6. Color science for digital photography Steve Marschner

Acquisition. Some slides from: Yung-Yu Chuang (DigiVfx) Jan Neumann, Pat Hanrahan, Alexei Efros

RGB Laser Meter TM6102, RGB Laser Luminance Meter TM6103, Optical Power Meter TM6104

NFMS THEORY LIGHT AND COLOR MEASUREMENTS AND THE CCD-BASED GONIOPHOTOMETER. Presented by: January, 2015 S E E T H E D I F F E R E N C E

There are many types of light sources, ranging from natural sunlight to man-made artificial lightings.

APPLICATIONS OF HIGH RESOLUTION MEASUREMENT

Gamma Scientific Display Measurement Systems

To discuss. Color Science Color Models in image. Computer Graphics 2

Novel Approach for LED Luminous Intensity Measurement

Light. intensity wavelength. Light is electromagnetic waves Laser is light that contains only a narrow spectrum of frequencies

Contents. Contents. Part 4. Specification. Part 1. Feature. Feature Merit Configuration. Specification Dimension Diagram

NEAR EYE DISPLAY (NED) SPECTRORADIOMETER SYSTEMS DATASHEET

Robust, Highly-visible and Facile Bioconjugation Colloidal Crystal Beads for Bioassay

Multiplex Image Projection using Multi-Band Projectors

WHITE PAPER. Guide to CCD-Based Imaging Colorimeters

Color , , Computational Photography Fall 2017, Lecture 11

DEFINING A SPARKLE MEASUREMENT STANDARD FOR QUALITY CONTROL OF ANTI-GLARE DISPLAYS Presented By Matt Scholz April 3, 2018

TOP 200. Telescopic Optical Probe for Radiance and Luminance Measurements. Two Global Leaders. One Complete Solution.

Photometry for Traffic Engineers...

Understand brightness, intensity, eye characteristics, and gamma correction, halftone technology, Understand general usage of color

Color images C1 C2 C3

Color , , Computational Photography Fall 2018, Lecture 7

Color Restoration of RGBN Multispectral Filter Array Sensor Images Based on Spectral Decomposition

ISS-30-VA. Product tags: Integrating Sphere Source. Gigahertz-Optik GmbH 1/5

LED Tester BTS256-LED

Digital Image Processing

A Spectral Database of Commonly Used Cine Lighting Andreas Karge, Jan Fröhlich, Bernd Eberhardt Stuttgart Media University

External Optical Probes. EOP-xxx and ISP 40. Two Global Leaders. One Complete Solution.

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy,

Color Science. CS 4620 Lecture 15

Broadbandradiometer specbos 1211

Assignment: Light, Cameras, and Image Formation

PERCEIVING COLOR. Functions of Color Vision

BTS2048-UV. Product tags: UV, Spectral Data, LED Binning, Industrial Applications, LED.

True energy-efficient lighting: the fundamentals of lighting, lamps and energy-efficient lighting

Instructions for the Experiment

Technical Notes. Integrating Sphere Measurement Part II: Calibration. Introduction. Calibration

COLOR. and the human response to light

IMPROVING AUTOMOTIVE INSPECTION WITH LIGHT & COLOR MEASUREMENT SYSTEMS

In sum the named factors cause differences for multicolor LEDs visible with the human eye, which can be compensated with color sensors.

Modifications of a sinarback 54 digital camera for spectral and high-accuracy colorimetric imaging: simulations and experiments

Lumen lm 1 lm= 1cd 1sr The luminous flux emitted into unit solid angle (1 sr) by an isotropic point source having a luminous intensity of 1 candela

Colorimetry evaluation supporting the design of LED projectors for paintings lighting: a case study

Photometry and Light Measurement

Spectroradiometer series NEW

INNOVATIVE CAMERA CHARACTERIZATION BASED ON LED LIGHT SOURCE

University of British Columbia CPSC 414 Computer Graphics

Color Reproduction. Chapter 6

Chroma Meter CL-200A. Even measures new light sources such as LED lighting and organic EL lighting

Report No.: HZ f. Total Luminous Flux (Lumens) Stabilization Time (Light & Power) Table 1: Executive Data Summary CRI

Announcements. The appearance of colors

12/02/2017. From light to colour spaces. Electromagnetic spectrum. Colour. Correlated colour temperature. Black body radiation.

BTS2048-VL-CP-ILED-B-IS-1.0-HL

Printing Beyond srgb Color Gamut by. Mimicking Silicon Nanostructures in Free-Space

Variable microinspection system. system125

Camera Requirements For Precision Agriculture

Color Science. What light is. Measuring light. CS 4620 Lecture 15. Salient property is the spectral power distribution (SPD)

Leica RCD30 Calibration Certificate

VIDEO-COLORIMETRY MEASUREMENT OF CIE 1931 XYZ BY DIGITAL CAMERA

Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters

Measuring colour in a world of light

Technical Explanation for Displacement Sensors and Measurement Sensors

MUSKY: Multispectral UV Sky camera. Valentina Caricato, Andrea Egidi, Marco Pisani and Massimo Zucco, INRIM

SPECTRAL SCANNER. Recycling

The Art of Light Measurement. Avantes BV Apeldoorn, The Netherlands

Understanding Imaging System Specifications for Pixel-Level Measurement of Displays

Color Digital Imaging: Cameras, Scanners and Monitors

Photometry for Traffic Engineers...

Visual Perception. Overview. The Eye. Information Processing by Human Observer

The Science Seeing of process Digital Media. The Science of Digital Media Introduction

Company synopsis. MSU series

COLOR. Elements of color. Visible spectrum. The Fovea. Lecture 3 October 30, Ingela Nyström 1. There are three types of cones, S, M and L

Image Formation: Camera Model

Technological Advances in General Lighting. New Lightmeter for Solid State Lighting. State-of-the-Art LED Illuminance Meter

Sunderland, NE England

High spatial resolution imaging colorimeter and gloss-meter for measurements of small parts

CRISATEL High Resolution Multispectral System

Whiteness indices and UV standards

Camera Requirements For Precision Agriculture

Transcription:

A High-Speed Imaging Colorimeter LumiCol 19 for Display Measurements Shigeto OMORI, Yutaka MAEDA, Takehiro YASHIRO, Jürgen NEUMEIER, Christof THALHAMMER, Martin WOLF Abstract We present a novel high-speed imaging colorimeter especially designed for test applications and requirements typically found in display measurement. It combines the benefits of a red-green-blue (RGB) complementary metal oxide semiconductor sensor (CMOS) camera and a high accuracy spot colorimeter. Using the colorimeter as reference for the camera enables precise characterizations of the whole display area with respect to luminance and chromaticity. This device has potential for substantial time- and costsavings, because it measures faster than conventional imaging colorimeter. It combines the functionality of two measurement devices into a single body and may reduce the number of test stations in production lines. 1 Background As display technologies improve, comprehensive characterizations of all optical display properties like contrast, gamma-curve, flicker, uniformity of luminance and color become increasingly important. This is not only true for R&D in laboratory, but in particular also for quality control, two dimensional inspections and adjustments in production lines. A number of different measurement devices like spectrometers, imaging colorimeters and conventional charge coupled device image sensor (CCD) cameras are used in order to realize the versatile measurements and test applications. Slowing down the tact times should be avoided because it leads to cost increase in production lines. Fig. 1 shows an example of uniformity measurement of an organic light-emitting diode (OLED) display in accordance with the Information Display Measurements Standard (IDMS) 1). The display shown in Fig. 1 has 9 windows located on black background. Spectra were measured with the Konica Minolta CS- spectroradiometer. Fig. shows the spectra of the three windows (T/R, M/C and B/L). In Fig. 1, from B/L to T/R, it is observed that the colors of the areas gradually shift toward blue. Also as shown in Fig., the Le values in 53 nm and 6 nm are decreasing respectively. From these figures, we can clearly observe a non-uniformity. Therefore, correcting the non-uniformity of displays is an important application in production lines. By using LumiCol 19, non-uniformity can be measured by a single measurement, and the tact time of non-uniformity adjustment can be significantly reduced. * R&D Division Sensing Business Unit, KONICA MINOLTA, INC. **Instrument Systems GmbH KONICA MINOLTA TECHNOLOGY REPORT VOL. 14 (17) 113

L v.95 u.184 v.461 L v 1.118 u.183 v.466 L v 1.37 u.179 v.471 L v.81 u.185 v.457 L v.953 u.183 v.464 L v 1.15 u.181 v.467 L v.761 u.183 v.456 L v.856 u.183 v.459 L v 1. u.181 v.463 Relative tristimulus values x(λ) y(λ) z(λ) Fig. 1 Example of uniformity inspection. View of OLED displays. Each display image is at RGB output signal level [19/19/19]. Where Lv denotes luminance value, and u and v denote color coordinates. Wavelength (nm) Fig. 3 CIE1931 XYZ color matching functions. Normalized Le (λ) T/R M/C L/B Relative responsibity Sb(λ) Sg(λ) Sr(λ) Wavelength (nm) Fig. Spectra of an OLED display at RGB output signal level [19/19/19] at different display positions. B/L marks the bottom left of the nine windows in Fig. 1. M/C refers to the middle center and T/R to the top right measurement spot, respectively. Le denotes light emission quantity. Prior Technology Conventional imaging colorimeters using a tristimulus filter wheel combine several sequential exposures. Exposures with all filters, which are typically four to six, cause a significant delay due to the mechanical movement of the wheel. The other method is to use an RGB camera. RGB cameras using a Bayer pattern can measure all color values at once. However, there are large differences in spectral transmittance between RGB filters and tristimulus XYZ filters. Fig. 3 shows the CIE1931 color matching functions x (λ), y (λ), and z (λ) which are equivalent to human eye s sensitivity. Fig. 4 shows the RGB spectral responsivities of the camera. To compensate the difference, RGB values obtained from the camera are calibrated to match the tristimulus values from a reference spectrometer, by using a display with typical spectrum as a calibration source and measuring different colors on the display screen. When a display with the different spectrum is measured, however, the difference of the spectrum of displays results in a measurement error. Wavelength (nm) Fig. 4 RGB camera spectral responsivity. 3 Measuring Principle We solved these problems by a real-time correction method. The real-time correction is to convert RGB values into tristimulus values by using simultaneous measurement of a built-in colorimeter. LumiCol 19 combines a fast and precise XYZ sensor with an RGB CMOS camera. We embed a modified spot-colorimeter ) as the reference XYZ sensor. The built-in XYZ sensor acquires X, Y and Z values while the camera using a Bayer pattern measures the RGB values, simultaneously. Thus, typical production requirements, such as high speed, high accuracy, low maintenance effort, and cost efficiency can be fulfilled. The special design and calibration procedure of LumiCol 19 thereby transfers high accuracy measurement values of the spot colorimeter to the D image of the camera. Thanks to the combination of a camera and a spot colorimeter, many different measurement tasks typically found in a production environment can be covered with the single device. At the same time, the measurement speed of LumiCol 19 is extremely high. In contrast, a conventional filter-wheel imaging colorimeter such as Instrument Systems LumiCam 114 KONICA MINOLTA TECHNOLOGY REPORT VOL. 14 (17)

13 3) requires several sequential exposures, one for every color filter, causes a significant delay causedue to by the change of the filters. Table 1 compares measurement speeds of LumiCol 19 4) and LumiCam 13. Luminance 1cd/m 1cd/m Table 1 Performance of measurement speed. LumiCol 19 Spot XYZ + RGB camera.5 sec. 1.5 sec. LumiCam 13 XYZ Filter Wheel + B/W Camera 6 sec. not described Fig. 5 shows a schematic of LumiCol 19. A beam splitter allows for simultaneous measurements of camera and colorimeter. Fig. 5 (b) shows a camera s field of view by the blue rectangle area and the measurement area of the colorimeter by the green circular area. The colorimeter serves as reference for realtime correction during each measurement. The acquisition of the XYZ values by LumiCol 19 consists of two steps: transformation from RGB to XYZ and real-time correction. In the first step, RGB data acquired by the camera is transformed into tristimulus value (X, Y, Z ) by the following equation: X m m 1 m R Y = m 1 m 11 m 1 G (1) Z m m 1 m B where m, m 1,, m are coefficients of the transformation matrix. The transformation matrix is calculated during the calibration at the factory, using a spectrometer and a typical display as a calibration source with different colors on the display screen. When color number i is displayed, CIE tristimulus values XYZ measured by the spectrometer are given by the following equations: X i = x (λ) Le i (λ) dλ R i = Sr (λ) Le i (λ) dλ G i = Sg (λ) Le i (λ) dλ (3) B i = Sb (λ) Le i (λ) dλ Given a set of (X i, Y i, Z i ) and (R i, G i, B i ), the transformation matrix is found in a way that a deviation between (X i, Y i, Z i ) and (X i, Y i, Z i ) is minimized in the XYZ color space. As mentioned, a certain amount of error may still remain because the spectrum of customer s deviceunder-test (DUT) is likely to be different from the spectrum of the display used for the factory calibration. In the real-time correction, the XYZ sensor is used to correct the tristimulus values after application of the transformation matrix, i.e. in the CIE 1931 color space. This is performed by comparing the measurement results of the XYZ sensor and the approximated tristimulus values over the camera pixels within the area of the colorimeter spot (the green area in Fig. 5 (b)). The three independent parameters (scaling factors) of the diagonal 3 3 matrix can thus simply be found by dividing each of the second tristimulus values by the values from the XYZ sensor. Finally, this correction is applied to all pixels in the entire camera s field, as explained above, so as to obtain the corrected tristimulus values. (a) Design X α X Y = β Y (4) Z γ Z XYZ colorimeter D-RGBcamera Y i = ȳ (λ) Le i (λ) dλ () Beam splitter Lens Z i = z (λ) Le i (λ) dλ (b) Field of view where X i, Y i, Z i are tristimulus values of color number i. These values are precisely measured with the spectrometer. On the other hand, the RGB values of the camera are given by the following equations: Camera CA-31 Fig. 5 Design and field of view of LumiCol 19. KONICA MINOLTA TECHNOLOGY REPORT VOL. 14 (17) 115

As a result, a high accuracy of the colorimeter measurement is achieved over camera s entire field of view. This real-time correction for the camera measurements works best with displays showing uniform color pattern, and can be used with any uniform DUT in principle. ΔL V 1.% (a) Luminance 1 cd/m 1 cd/m 1 cd/m (LLM) 4 Measurement Performance The measurement performance resulting from the above-described principle can be demonstrated by the accuracy of the colorimeter with respect to a reference instrument and by means of the response nonuniformity (RNU). Moreover, the high measurement speed and time-saving potential are briefly highlighted. Accuracy of luminance measurements Fig. 6 shows the accuracy of LumiCol 19 colorimeter measurements with respect to luminance L v and the color coordinates u and v, respectively, determined for four different devices (with 1 measurements on average). The measurements were performed on a display showing white images with two different luminance levels, 1 and 1 cd/m. The measurements of the built-in XYZ sensor were compared to the measurement of a high-end spectroradiometer. In addition, the results of measurements using an optional low-luminance mode (LLM) are shown. In the LLM, averaged dc-calibrations are performed prior to each measurement. The sum of the mean value and the threefold standard deviation (ave + 3σ) of the luminance measurement (Fig. 6 (a)) of the colorimeter with respect to a spectroradiometer are.9 % and.77 % for L v =1 cd/m (circles) and L v =1 cd/m (triangles), respectively. For the color coordinates u (Fig. 6 (b)) and v (Fig. 6 (c)), the deviations are given as absolute values. At a luminance level of L v = 1 cd/m (circles), the comparison between built-in colorimeter and spectroradiometer reveals mean deviations of.4 for u and.18 for v. For L v = 1 cd/m (triangles), the mean deviations are. for u and.54 for v. The low luminance mode, represented by diamonds, reduces the deviations for measurements at L v = 1 cd/ m to.7 for u and. for v. Accuracy of color measurements Fig. 7 shows the accuracy, again using a high-end spectroradiometer as reference instrument, for measurements on a DUT showing colored images (red [55//], green [/55/], and blue [//55] represented by Δu x1-3 Δv x1-3 3 1 3 1 (b) u 1 cd/m 1 cd/m 1 cd/m (LLM) (c) v 1 cd/m 1 cd/m 1 cd/m (LLM) Fig. 6 Measurement accuracy of LumiCol 19 (for white). The deviations from spectroradiometer measurements with respect to luminance and chromaticity were determined on a display showing white images of different luminance levels (circles: 1 cd/m ; triangles: 1 cd/m ). In addition, the deviations for measurements in the LLM are shown (diamonds: 1 cd/m ). colored circles, squares and triangles). Three graphs show the deviations of luminance L v and color coordinates u and v (c), respectively. For luminance, the ave + 3σ values are below.6 % for red and green (circles and squares) and below.9 % for blue (triangles). For the color coordinate u, the ave + 3σ values are.3,.4, and.8 for the colors red, green and blue, respectively (same symbols as in Fig. 7 (a)). In Fig. 7 (c), the accuracy for the color coordinate v is shown. Here, the ave + 3σ values are.9,.8, and.1 for red, green, and blue, respectively. In sum, it was shown that the accuracy of the novel measurement device for luminance and color measurements on a display of arbitrary uniform content is within the range of state-of-the-art imaging colorimeters based on the filter wheel technology. 116 KONICA MINOLTA TECHNOLOGY REPORT VOL. 14 (17)

(a) Luminance Red Green Blue 1.% (a) Luminance RNU Averaged RNU + 3σ ΔL V RNU L V Δu x1-3 (b) u Red Green Blue RNU u x1-3 3 1 (b) u RNU Averaged RNU + 3σ Δv x1-3 (c) v Red Green Blue RNU v x1-3 8 6 4 (c) v RNU Averaged RNU + 3σ Fig. 7 Measurement accuracy of LumiCol 19 (for color). The deviations from spectroradiometer measurements with respect to luminance and chromaticity were determined on a display showing colored images (red, green, and blue represented by circles, squares and triangles respectively). Fig. 8 Response non-uniformity of four different LumiCol 19 devices determined from measurements on a uniform light source with a luminance level of 1 cd/m (represented by symbols). The RNUs were determined for the quantities luminance (a) and color coordinates u and v. The dashed lines mark the ave + 3σ values. Response non-uniformity (RNU) The response non-uniformity is a measure for the accuracy of the pixel-wise two-dimensional measurement. The results of the evaluation below show how a conventional RGB CMOS camera can be utilized for high-accuracy color measurements in the CIE 1931 XYZ color space. This is achieved by the specific calibration procedure applied to, i.e. flat field correction, color calibration, and color fine-correction to the reference sensor. The response non-uniformity in Fig. 8 was determined from measurements of four different LumiCol 19 devices on a uniform light source (L v = 1 cd/ m ). For the evaluation of the RNU, an 8 8 binning was applied to the XYZ images, resulting in a resolution of 4x15 pixels. After binning, the XYZ-image was transformed into (L v, u, v ). For each channel (L v, u, v ), the mean value was calculated. For chromaticity (u, v ), the response non-uniformity is defined as the largest deviation from the mean value within the binned image. For luminance L v, the RNU is defined as the largest deviation from the mean value divided by the mean value of the binned image. Figs. 8 (a)-(c) show the results. For luminance measurements, the RNUs, determined for four different devices, are all below.8 % (symbols in Fig. 8 (a)). Adding the standard deviation 3σ to the mean value sums up to around.95 % (dashed line). Figs.8 (b) and (c) show the corresponding values for the color coordinates u and v. The maximum value of Δu and Δv is.65. The ave + 3σ values are. for u and.73 for v. KONICA MINOLTA TECHNOLOGY REPORT VOL. 14 (17) 117

5 Conclusion This report presents a novel approach to imaging colorimetry focusing on the typical needs and requirements of testing and adjusting in display production environments. The basic principle for LumiCol 19 is the usage of a spot colorimeter as reference for D measurement. On the one hand, the built-in XYZ sensor enables to determine a transformation matrix during the calibration procedure, converting the RGB values of the camera pixels into the CIE 1931 XYZ tristimulus values; the colorimeter is used for a realtime correction during each measurement to improve the accuracy of the camera measurement. The resulting accuracy of the LumiCol 19 for luminance and color measurements on displays is within the accuracy range of state-of-the-art imaging colorimeters based on the filter-wheel technology. It was shown that this is also true for uniformity of the RGB CMOS camera by RNU. Moreover, LumiCol 19 measures significantly faster because all color values are captured with a single exposure. The chromaticity response of non-uniformity by the pixel-wise two-dimensional measurement is below.65. The imaging colorimeter measures ten times faster than the conventional filter-wheel imaging colorimeter. LumiCol19 has been developed in a short period of time by a complementary combination of Instrument Systems imaging colorimeter technology and Konica Minolta s XYZ sensor technology. 6 References 1) Information Display Measurements Standard V1.3aa. chapter 8.1. Sampled Vantage Point Uniformity ) Konica Minolta Holdings, Inc. CA-31 catalogue http://www.konicaminolta.com/instruments/ download/catalog/display/pdf/ca31_catalog_eng. pdf 3) Instrument Systems GmbH, LumiCam13 data sheet http://www.instrumentsystems.jp/fileadmin/ editors/downloads/products/lumicam13_e.pdf 4) Instrument Systems GmbH, LumiCol 19 data sheet http://www.instrumentsystems.com/fileadmin/ editors/downloads/products/lumicol19_en.pdf This paper is written based on Publication M. Wolf, J. Neumeier. A High-Speed -in-1 Imaging Colorimeter for Display Production Applications. 7- SID 16 Digest: 974-977. 118 KONICA MINOLTA TECHNOLOGY REPORT VOL. 14 (17)