Performance Evaluation of SiPM Detectors for PET Imaging in the Presence of Magnetic Fields

Similar documents
An innovative detector concept for hybrid 4D-PET/MRI Imaging

Investigation of Solid-State Photomultipliers for Positron Emission Tomography Scanners

Development of PET using 4 4 Array of Large Size Geiger-mode Avalanche Photodiode

Study of Silicon Photomultipliers for Positron Emission Tomography (PET) Application

Design and development of compact readout electronics with silicon photomultiplier array for a compact imaging detector *

PUBLISHED BY IOP PUBLISHING FOR SISSA MEDIALAB. Imaging performance of silicon photomultipliers coupled to BGO and CsI:Na arrays

Effects of Dark Counts on Digital Silicon Photomultipliers Performance

Tutors Dominik Dannheim, Thibault Frisson (CERN, Geneva, Switzerland)

Design and development of compact readout electronics with silicon photomultiplier array for a compact imaging detector

Characterisation of SiPM Index :

Development of an innovative LSO-SiPM detector module for high-performance Positron Emission Tomography

Initial results on Sipm array based on a symmetric resistive voltage division readout

Investigation of low noise, low cost readout electronics for high sensitivity PET systems based on Avalanche Photodiode arrays

Timing Resolution Performance Comparison for Fast and Standard Outputs of SensL SiPM

The digital Silicon Photomultiplier A novel Sensor for the Detection of Scintillation Light

POSITRON emission tomography (PET) scanners, which

Highlights of Poster Session I: SiPMs

Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET

PoS(PhotoDet 2012)022

P ILC A. Calcaterra (Resp.), L. Daniello (Tecn.), R. de Sangro, G. Finocchiaro, P. Patteri, M. Piccolo, M. Rama

764 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 51, NO. 3, JUNE 2004

Performance measurements of a depth-encoding PET detector module based on positionsensitive

Recent Development and Study of Silicon Solid State Photomultiplier (MRS Avalanche Photodetector)

Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud

Thomas Frach, Member, IEEE, Walter Ruetten, Member, IEEE, Klaus Fiedler, Gunnar Maehlum, Member, IEEE, Torsten Solf, and Andreas Thon

Silicon Photomultiplier

Cross-Strip Multiplexed Electro-Optical Coupled Scintillation Detector for Integrated PET/MRI

PoS(PhotoDet 2012)016

Design of a Novel Front-End Readout ASIC for PET Imaging System *

Characterization of a 64 Channel PET Detector Using Photodiodes for Crystal Identification *

Nuclear Instruments and Methods in Physics Research A

TM-xx-xx-xx / Seite 2

Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud

Characterization of Silicon Photomultipliers and their Application to Positron Emission Tomography. Zhiwei Yang. Abstract

Solid-State Photomultiplier in CMOS Technology for Gamma-Ray Detection and Imaging Applications

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

60 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 1, FEBRUARY /$ IEEE

Arrays of digital Silicon Photomultipliers Intrinsic performance and Application to Scintillator Readout

Characterization of a prototype matrix of Silicon PhotoMultipliers (SiPM s)

2594 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 5, OCTOBER /$ IEEE

A test of silicon photomultipliers as readout for PET

Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud

Review of Solidstate Photomultiplier. Developments by CPTA & Photonique SA

LaBr 3 :Ce, the latest crystal for nuclear medicine

A comparative study of the time performance between NINO and FlexToT ASICs

Application of Silicon Photomultipliers to Positron Emission Tomography

Performance Assessment of Pixelated LaBr 3 Detector Modules for TOF PET

Design Studies of A High-Performance Onboard Positron Emission Tomography For Integrated Small Animal PET/CT/RT Radiation Research Systems

Simulation of Algorithms for Pulse Timing in FPGAs

Silicon Photo Multiplier SiPM. Lecture 13

Development of a simplified readout for a compact gamma camera based on 2 2 H8500 multi-anode PSPMT array

A Study of Silicon Photomultiplier Sensor Prototypes for Readout of a Scintillating Fiber / Lead Sheet Barrel Calorimeter

Nuclear Instruments and Methods in Physics Research A

Small Prototype Gamma Spectrometer Using CsI(Tl) Scintillator Coupled to a Solid-State Photomultiplier

SiPMs for solar neutrino detector? J. Kaspar, 6/10/14

EndoTOFPET-US: an endoscopic Positron Emission Tomography detector for a novel multimodal medical imaging tool

RECENTLY, the Silicon Photomultiplier (SiPM) gained

Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM

Novel scintillation detectors. A. Stoykov R. Scheuermann

White Paper. Zecotek MAPD (Multi- pixel Avalanche Photo Diode) Enabling the future of imaging and detection

Characterization of Silicon Photomultiplier Readout Designs for Use in Positron. Emission Tomography Systems. Chen-Yi Liu. the University of Manitoba

MPPC versus MRS APD in two-phase Cryogenic Avalanche Detectors

SPMMicro. SPMMicro. Low Cost High Gain APD. Low Cost High Gain APD. Page 1

Index terms: PET, Silicon Photo-multiplier, Small Animal PET insert for MRI scanner. Size: 1.2x1.2 mm, Pitch: mm Thickness: 4 mm

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

SensL B-Series Silicon Photomultipliers for TOF- PET. NDIP2014 Kevin O Neill 4 th July, 2014

Low Dark Count UV-SiPM: Development and Performance Measurements P. Bérard, M. Couture, P. Deschamps, F. Laforce H. Dautet and A.

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare

Advanced Materials Research Vol

LaBr 3 :Ce scintillation gamma camera prototype for X and gamma ray imaging

Electronic Instrumentation for Radiation Detection Systems

Moderne Teilchendetektoren - Theorie und Praxis 2. Dr. Bernhard Ketzer Technische Universität München SS 2013

A Survey of Power Supply Techniques for Silicon Photo-Multiplier Biasing

Direct Measurement of Optical Cross-talk in Silicon Photomultipliers Using Light Emission Microscopy

Direct Measurement of Optical Cross-talk in Silicon Photomultipliers Using Light Emission Microscopy

A Measurement of the Photon Detection Efficiency of Silicon Photomultipliers

High granularity scintillating fiber trackers based on Silicon Photomultiplier

arxiv: v2 [physics.ins-det] 10 Jan 2014

Concept and status of the LED calibration system

PRELIMINARY RESULTS OF PLASTIC SCINTILLATORS DETECTOR READOUT WITH SILICON PHOTOMULTIPLIERS FOR COSMIC RAYS STUDIES *

Status of the LED calibration system

Positron Emission Tomography

Total Absorption Dual Readout Calorimetry R&D

Future directions in Nuclear Medicine Instrumentation

Instructions for gg Coincidence with 22 Na. Overview of the Experiment

Digital coincidence acquisition applied to portable β liquid scintillation counting device

Silicon Photomultiplier Evaluation Kit. Quick Start Guide. Eval Kit SiPM. KETEK GmbH. Hofer Str Munich Germany.

Journal of Radiation Protection and Research

arxiv: v3 [astro-ph.im] 17 Jan 2017

Characterization of New-Generation Silicon Photomultipliers for Nuclear Security Applications

Development of a High-Resolution and Depth-of- Interaction Capable Detector for Time-of-Flight PET

A high energy gamma camera using a multiple hole collimator

A PET detector module using FPGA-only MVT digitizers

Application Notes: Discrete Amplification Photon Detector 5x5 Array Including Pre- Amplifiers Board

Development of the LBNL Positron Emission Mammography Camera

MICRO PIXEL AVALANCHE PHOTODIODE AS ALTERNATIVE TO VACUUM PHOTOMULTIPLIER TUBES

PoS(PhotoDet 2012)058

Design and Simulation of a Silicon Photomultiplier Array for Space Experiments

Silicon Photomultipliers. Dieter Renker

Overview Full Featured Silicon Photomultiplier Module for OEM and Research Applications The is a solid state alternative to the Photomultiplier Tube (

Transcription:

2008 IEEE Nuclear Science Symposium Conference Record M02-4 Performance Evaluation of SiPM Detectors for PET Imaging in the Presence of Magnetic Fields Samuel España, Student Member, IEEE, Gustavo Tapias, Luis M. Fraile, Joaquín L. Herraiz, Student Member, IEEE, Esther Vicente, Student Member, IEEE, Jose Udias, Member, IEEE, Manuel Desco, Juan J. Vaquero, Senior Member, IEEE Abstract The multi-pixel photon counter (MPPC) or silicon photo-multiplier (SiPM), recently introduced as a solid-state photodetector, consists of an array of Geiger-mode photodiodes (microcells). It is a promising device for PET thanks to its potential for high photon detection efficiency (PDE) and immunity to high magnetic fields. It is also very easy to use, with simple electronic read-out, high gain and small size. In this work we evaluate the performance of three 1 1 mm 2 and one 6 6 mm 2 (2 2 array) SiPMs offered by Hamamatsu for their use in PET. We examine the dependence of the energy resolution and the gain of these devices on the thermal and reverse bias when coupled to LYSO scintillator crystals. We find that the 400 and 1600 microcells models and the 2 2 array are suitable for small size crystals, like those employed in high resolution small animal scanners. The good performance of these devices up to 7 Tesla has also been confirmed. I. INTRODUCTION OSITRON emission tomography (PET) scanners have been Pused to investigate biochemical and pathological phenomena, to diagnose disease, and to determine prognosis after treatment. On the other hand, magnetic resonance imaging (MRI) provides unsurpassed soft-tissue contrast and does not require the use of ionizing radiation. Therefore, it is expected that combined PET/MR scanners represent the future for biomedical imaging [1], and that these scanners will either supplement or compete with PET/CT for basic research and clinical applications. However, one major obstacle of the combined PET/MR system with photomultiplier tubes (PMT) is that these tubes are extremely sensitive to magnetic fields. PET/MR using PMTs involves the use of long optical fibers to transfer the light signals from the scintillation crystals located inside the MR scanner [2]. This method produces a degradation of energy resolution due to light losses during the optical transfer. The use of semiconductor photo-detectors, such as avalanche photodiodes (APD) instead of PMTs offers an alternative development option which is currently being Manuscript received November 14, 2008. This work was supported in part by the MEC (FPA2007-07393), CDTEAM (CENIT-Ingenio 2010) Ministerio de Industria, Spain, UCM (Grupos UCM; 910059), CPAN (Consolider- Ingenio 2010) CSPD-2007-00042 projects, and the RECAVA-RETIC network. S. España, L. M. Fraile, J. L. Herraiz, E. Vicente and J. M. Udías are with the Grupo de Física Nuclear, Dpto. Física Atómica, Molecular y Nuclear, Universidad Complutense de Madrid, Spain (S. España email: samuel@nuclear.fis.ucm.es). G. Tapias, M. Desco and J. J. Vaquero are with the Unidad de Medicina y Cirugía Experimental, Hospital General Universitario Gregorio Marañón, Madrid, Spain (J. J. Vaquero email: juanjo@mce.hggm.es). pursued by several research groups [3]. APDs are compact and insensitive to magnetic fields as compared with PMTs, and these properties are desirable for PET/MR. One disadvantage of presently available APDs is that they have low gains of an order of a few hundreds, and thus require sophisticated preamplifiers. New alternatives for PET/MR scanners are Silicon Photomultiplier detectors (SiPM). These devices are solidstate photodetectors consisting of an array of photodiodes (microcells) operated in Geiger mode [4]. They exhibit good photon detection efficiency (PDE) and they should not be affected by external magnetic fields. They are easy to use, require simple electronics, do not need high voltage power supplies and provide a high gain, all in a compact size. Their insensitivity to magnetic field makes them suitable for development of PET scanners for simultaneous PET/MR studies. Several groups are studying the SiPM performance for PET purpose showing promising results [5, 6]. II. MATERIALS AND METHODS In this work we evaluate three 1 1 mm 2 SiPM models with 100, 400 and 1600 microcells and a 6 6 mm 2 (2 2 array) SiPM with 3600 microcells per element, all of them from Hamamatsu, for their use in PET and PET/MR. They were coupled to 1.5 1.5 12 mm 3 LYSO crystals. We have studied the performance of these devices and their immunity to magnetic fields of up to 7 Tesla with superconducting magnet from Bruker. TABLE I SIPM USED ON THE PRESENT STUDY Series 11-100P 11-050P 11-025P MPPC-33-2x2-50 5900 Active area (mm2) Number of microcells Pixel size (μm) Fill factor P.D.E. (%) 1 1 100 100 78.5 75 1 1 400 50 61.5 50 1 1 1600 25 30.8 25 6 6 (2 2 array) 3600/ element 50 61.5 50 Data acquisition was performed using an Agilent 6000 series oscilloscope (2 GSa/s). The oscilloscope was placed in the MRI controller room using 5 m long cables for the output 978-1-4244-2715-4/08/$25.00 2008 IEEE 3591

signals. The power supply for the SiPM and the amplifier circuit was placed in the MR scanning room behind the 5 Gauss line and were connected by 2 m long cables. A. Single SiPM Fig. 1 shows the experimental setup employed for the 1 1 mm 2 SiPMs. Each SiPM is coupled with optical grease to a LYSO crystal wrapped with Teflon. A Delrin case was used to prevent external light from reaching the SiPM. Fig. 2. Picture (top) and scheme (middle) of the readout electronics built for the 1 1 mm 2 SiPMs. Pulses obtained from the single SiPMs with different pixel size (bottom). We have measured the energy resolution of these devices, and the dependence of the gain with reverse bias and temperature with a 22Na radioactive source in the presence of magnetic fields of 0 and 7 T. Fig. 1. Experimental setup used for 1 1 mm 2 SiPM. fast shaper, providing pulses with 100 ns width, has been designed and constructed for the readout of the SiPMs (see Fig. 2). This allows for very high counting rates to be handled. B. SiPM array A 4 4 LYSO crystal matrix was coupled to the 2 2 SiPM array. Each crystal element, individually wrapped in Teflon, has a size of 1.5 1.5 12 mm 3 (see Fig. 3), wrapped individually with Teflon. Fig. 3. Experimental setup used for SiPM array. Readout electronics, specifically tailored to this application was built in order to acquire the signals from the SiPM channels as well as an additional signal with the sum of all channels, which is used as external trigger. Fig. 4. Scheme of the electronics readout used for the SiPM array. 3592

The energy resolution and flood histogram were measured for both 0 and 7 Tesla. A 18 F radioactive source was used for these measurements. III. RESULTS In Fig. 7, the relative gains for the same measurements shown in Fig. 6 are graphed. The slope is smaller for the SiPM with 1600 microcells followed by those with 400 and 100 microcells. The slopes are comparable for the 0 and 7 Tesla measurements. A. Single SiPM Fig. 5 shows the energy spectra obtained for the three 1 1 mm 2 SiPMs using a 22 Na radioactive source. They must be linearized since the SiPMs saturate when most of the pixels are triggered. The 511 kev peak is clearly resolved in all cases but the 1.275 MeV peak is not so well resolved due to saturation. Obviously the higher the number of microcells the less the saturation and therefore the 1275 kev peak is better observed for the SiPM with 1600 microcells. Fig. 7. Relative gain as a function of voltage. The energy resolution at 511 kev was also measured for a fixed voltage for each SiPM as a function of temperature. The SiPMs with 1600 and 400 microcells exhibit again the most stable behavior compared with the 100 microcells SiPM (see Fig. 8). Fig. 5. Energy spectra obtained for the three single SiPMs using a 22 Na radioactive source. The energy resolution at 511 kev is shown in Fig. 6 as the percentage FWHM of the peak, as a function of applied voltage. The best FWHM was obtained for he SiPMs with 1600 and 400 microcells. Furthermore, the 1600 microcells FWHM does not change with voltage. Red lines are the results at 0 Tesla measurement and green lines are for those at 7 Tesla. Both measurements were performed at the same temperature of 20ºC and yield very similar results. Fig. 8. Energy resolution as a function of temperature. The variation of the relative gain as a function of the temperature shows once again more uniform behavior of the 40 and 1600 microcells SiPMs (see Fig. 9). Fig. 6. Energy resolution as a function of voltage. Fig. 9. Relative gain as a function of temperature. 3593

ΔGain ΔVoltage 1 Gain Variation = 100 (% V ) (1) Gain Table II shows the gain variation (see eq. 1) with voltage and with temperature as a summary of the results obtained with the single SiPMs. The variation is smaller for the 400 and 1600 microcells SiPMs. TABLE II SUMMARY OF RESULTS FOR SINGLE SIPM BY MEANS OF GAIN VARIATION Gain Variation 100 microcells 400 microcells 1600 microcells 0 Tesla 7 Tesla Constant T 230 % V-1 235 % V-1 110 % V-1 110 % V-1 70 % V-1 75 % V-1 Constant V -2.7 % ºC-1-0.4 % ºC-1-1.1 % ºC-1 B. SiPM array Fig. 10 shows the four channel signals of an event. The total width of the pulses is about 200 ns and the height is a few hundreds mv. These pulses are integrated and Anger logic is employed to compute the centroid of the interactions. Fig. 11. Flood histogram (top) and horizontal line profile (bottom). Energy spectra for every crystal element measured at 0 and 7 Tesla are shown in Fig. 12. The differences for the results with and without magnetic field are negligible. An additional acquisition was performed using a 22 Na in order to test the linearity of the energy spectrum. Fig. 12 (bottom) shows the good linearity for the energy spectrum obtained with a single crystal placed at the center of the matrix. Fig. 10. Pulses of four channels for one event detected with the SiPM array. The flood histogram shown in Fig. 11 was obtained in an acquisition of 5 minutes with a 18 F source, accumulating 5 10 5 counts. All the crystals of the matrix are clearly resolved. On the same figure, the horizontal profile of counts inside the red rectangle of the flood histogram is shown. Red and green lines of the figure represent the results obtained for 0 and 7 Tesla measurements respectively showing the same results. A fixed value of the bias voltage of 68.2 V was applied, while the temperature was kept at 20ºC. A peak to valley ratio of 10:1 is achieved. 3594

[4] Renker, D., Geiger-mode avalanche photodiodes, history, properties and problems. Nucl. Inst. Meth. In Phy. Res. A, 2006. 567: p. 48-56. [5] Otte, A.N., et al., A test of Silicon Photomultipliers as readout for PET. Nucl. Inst. Meth. In Phy. Res. A, 2005. 545(3): p. 705-15. [6] Llosá, G., et al. Silicon Photomultipliers and SiPM matrices as Photodetectors in Nuclear Medicine. in IEEE NSS-MIC. 2007. Fig. 12. Energy spectra obtained for each crystal element using 18 F (top) and for one crystal at the center of the matrix using 22 Na. The energy resolution at 511 kev for sample crystals placed at different location inside the matrix is shown in table III. TABLE III ENERGY RESOLUTION FOR CRYSTAL ELEMENT AT DIFFERENT POSITIONS INSIDE THE MATRIX FWHM (%) 0 Tesla 7 Tesla Center 11.2 11.9 Center edge row 14.1 14.4 Corner 21.7 21.4 IV. CONCLUSIONS Single SiPMs with an active area of 1 1 mm2 have been tested at 0 and 7 Tesla. We found no significant magnetic field effect. The 400 and 1600 microcells SiPMs seem more suitable for PET purposes exhibiting better and more stable energy resolution. A 2 2 SiPM array has also been examined in conjunction with a 4 4 LYSO crystal matrix. All the 4 4 crystals of 1.5 mm pitch size were perfectly resolved by the 2 2 SiPM array, yielding a 10:1 peak to valley ratio in the counts profile at both 0 and 7 Tesla. Our measurements show the strong potential for PET/MR scanners with scintillator crystal sizes of 1 mm. ACKNOWLEDGMENT Part of the computations of this work were done at the High Capacity Cluster for Physical Techniques'' of the Faculty for Physical Sciences of the UCM, funded in part by the UE under the FEDER program and in part by the UCM. E. Vicente acknowledges support from CSIC-JAE predoctoral fellowships respectively. REFERENCES [1] Gaa, J., E.J. Rummeny, and M.D. Seemann, Whole-body imaging with PET/MRI. Eur. J. Med. Res., 2004. 9: p. 309-12. [2] Shao, Y., et al., Simultaneous PET and MR imaging. Phys Med Biol, 1997. 42: p. 1965-70. [3] Pichler, B., et al. Performance Test of a LSO-APD PET Module in a 9.4 Tesla Magnet. in IEEE NSS-MIC. 1998. 3595