SPECIFICATIONS NI PXIe-4322 This document lists specifications for the NI PXIe-4322 module. These specifications are typical for the range of 0 C to 55 C unless otherwise stated. The system must be allowed to warm up for 15 minutes to achieve the rated accuracy. All specifications are subject to change without notice. Visit ni.com/manuals for the most current specifications and product documentation. Note Keep the filler panels on all unused slots in your chassis to maintain forced air cooling. Analog Characteristics Number of channels... 8 analog output channels DAC resolution... 16 bits Type of DAC... R-2R Monotonicity... 16 bits DNL... ±1 LSB max INL (best fit)... ±4 LSBs max Power-on output state 1 Voltage mode... 0 V Current mode... 0 ma Note You can program the power-on output states. Refer to your software documentation for information about programming the power-on output states using NI-DAQmx with LabVIEW or other National Instruments application development environments (ADEs). Power-off output state... High impedance 1 When the output stage powers on, a current glitch occurs for 1 ms peaking at 5 μa. When the output stage powers off, a current glitch occurs for 3 ms peaking at 5 μa.
Protection Overvoltage...±120 VDC Short circuit...indefinitely Update rate Maximum...250 ks/s per channel Minimum...No minimum Timing accuracy...50 ppm of sample rate Timing resolution...10 ns Data transfers...dma (scatter-gather), programmed I/O Output FIFO size...8,191 samples shared among channels used AO waveform modes Nonperiodic waveform Periodic waveform regeneration mode from onboard FIFO Periodic waveform regeneration from host buffer including dynamic update Voltage Mode Output voltage range 1 Nominal...±16 V Minimum...±16.57 V Typical...±16.70 V Maximum...±16.83 V Current drive...±20 ma per channel max Output impedance...25 mω Noise (rms) 10 Hz to 1 khz bandwidth...30 μv rms 10 Hz to 300 khz bandwidth...250 μv rms 10 Hz to 20 MHz bandwidth...500 μv rms Slew rate...±10 V/μs Crosstalk Channel-to-channel @ 10 khz...-100 db Common-mode voltage @ 60 Hz...-120 db 1 Refer to the Increasing Output Voltage Range in Voltage Mode section in the NI PXIe-4322 User Manual for information about how to increase the nominal output voltage range by connecting multiple voltage channels in series. 2 ni.com NI PXIe-4322 Specifications
Settling time 1000 pf load, to 1 LSB 20 V step... 20 μs 1 V step... 12 μs 0.1 V step... 10 μs 500 Ω 100 pf, to 1 LSB 20 V step... 20 μs Capacitive drive... 4500 pf Accuracy 1 Measurement Conditions Percent of Output (Gain Error) Percent of Range* (Offset Error) Calibrated, max (0 C to 55 C) 0.076% 0.018% Calibrated, max (0 C to 40 C) 0.054% 0.014% Calibrated, max (23 C ±5 C) 0.014% 0.007% Calibrated, typ (23 C ±5 C) 0.010% 0.003% * Range equals 16 V. Stability Gain drift... 7 ppm/ C Offset drift... 25 μv/ C Absolute Voltage Output Accuracy Equation AbsoluteVoltageAccuracy = Output * (GainError) + Range * (OffsetError) Absolute Voltage Output Accuracy Example For a 10 V voltage output, the absolute output accuracy for an external temperature range of 18 C to 28 C is as follows: GainError = 0.014% OffsetError = 0.007% AbsoluteAccuracy = 10 V * (GainError) + 16 V * (OffsetError) = 2.52 mv 1 Accuracies listed are warranted for the conditions described in the table for up to one year from the module external calibration. NI PXIe-4322 Specifications National Instruments 3
Current Mode Output current range 1 Nominal...±20 ma Minimum...±20.6 ma Typical...±20.9 ma Maximum...±21.1 ma Compliance voltage...±16 V per channel max Output impedance...100 MΩ Noise (rms) 10 Hz to 1 khz bandwidth...50 na 10 Hz to 300 khz bandwidth...600 na Slew rate...±20 ma/μs Crosstalk Channel-to-channel @ 1 khz...-100 db Common-mode voltage @ 60 Hz...50 na/v Settling time 100 Ω load Full-scale step to 2 LSB...20 μs 800 Ω load Full-scale step to 2 LSB...25 μs 2 ma step to 1 LSB...15 μs Inductive drive...10 μh 1 Refer to the Increasing Output Current Range in Current Mode section of the NI PXIe-4322 User Manual for information about how to increase the nominal output current range by connecting current channels in parallel. 4 ni.com NI PXIe-4322 Specifications
Accuracy 1 Measurement Conditions Percent of Output (Gain Error) Percent of Range* (Offset Error) Calibrated, max (0 C to 55 C) 0.12% 0.05% Calibrated, max (0 C to 40 C) 0.09% 0.035% Calibrated, max (23 C ±5 C) 0.033% 0.019% Calibrated, typ (23 C ±5 C) 0.028% 0.004% *Range equals 20 ma. Stability Gain drift... ±15 ppm/ C Offset drift... ±75 na/ C Absolute Current Output Accuracy Equation AbsoluteCurrentAccuracy = Output * (GainError) + Range * (OffsetError) Absolute Current Output Accuracy Example For a 10 ma current output, the absolute output accuracy for an external temperature range of 18 C to 28 C is as follows: GainError = 0.033% OffsetError = 0.019% AbsoluteAccuracy = 10 ma * (GainError) + 20 ma* (OffsetError) = 7.1 μa Synchronization Reference clock source... PXIe_DSTAR<A.. B>, PXI_STAR, PXIe_Clk100, PXI_TRIG<0..7> Digital Triggers Source... PXI_TRIG<0..7>, PXI_STAR, PXIe_DSTAR<A..B> PFI<0..1> Purpose... Start Trigger, Pause Trigger 1 Accuracies listed are warranted for the conditions described in the table for up to one year from the module external calibration. NI PXIe-4322 Specifications National Instruments 5
Polarity...Software-selectable Debounce filter settings...disable, 90 ns, 5.12 μs, 2.56 ms, custom interval Clocking Source...Onboard Clock, PXI_TRIG<0..7>, PXI_STAR, PXIe_DSTAR<A..B>, PXIe_Clk100 (RefClk only) Destination...Sample Clock, Sample Clock Timebase, Reference Clock Polarity...Software-selectable (except Reference Clock) Debounce filter settings (Sample Clock only)...disable, 90 ns, 5.12 μs, 2.56 ms, custom interval Reference clock locking frequencies Locking Input Frequency (MHz) Reference Signal 10 20 100 PXIe_DSTAR<A..B> PXI_STAR PXIe_Clk100 PXI_TRIG<0..7> Note National Instruments does not recommend locking to non-selected frequencies. Output Timing Signals Source...Start Trigger, Pause Trigger, Sample Clock, various derived timebases and clocks Destination...PXI_TRIG<0..7> PXIe_DSTARC Polarity...Software-selectable 6 ni.com NI PXIe-4322 Specifications
Bus Interface Form factor... x1 PXI Express peripheral module, specification rev 1.0 compliant Slot compatibility... x1 and x4 PXI Express or PXI Express hybrid slots DMA channels... 1 analog output NI PXIe-4322 modules may be installed in PXI Express slots or PXI Express hybrid slots. Calibration Recommended warm-up time... 15 minutes Calibration interval... 1 year Power Requirements +3.3 V... 800 ma +12 V... 700 ma Physical Requirements Dimensions... Standard 3U PXIe, 16 10 cm (6.3 3.9 in.) Weight... 161 g (5.7 oz) I/O connector... 96-pin male DIN 41612/IEC 60603-2 connector Caution Clean the hardware with a soft, nonmetallic brush. Make sure that the hardware is completely dry and free from contaminants before returning it to service. Environmental Specifications Maximum altitude... 2,000 m (800 mbar), at 25 C ambient temperature Pollution Degree... 2 Indoor use only NI PXIe-4322 Specifications National Instruments 7
Operating Environment Ambient temperature range...0 C to 55 C (Tested in accordance with IEC-60068-2-1 and IEC-60068-2-2. Meets MIL-PRF-28800F Class 3 low temperature limit and MIL-PRF-28800F Class 2 high temperature limit.) Relative humidity range...10% to 90%, noncondensing (Tested in accordance with IEC-60068-2-56.) Storage Environment Ambient temperature range...-40 C to 71 C (Tested in accordance with IEC-60068-2-1 and IEC-60068-2-2. Meets MIL-PRF-28800F Class 3 limits.) Relative humidity range...5% to 95% noncondensing (Tested in accordance with IEC-60068-2-56.) Shock and Vibration Operating shock...30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC-60068-2-27. Meets MIL-PRF-28800F Class 2 limits.) Random vibration Operating...5 Hz to 500 Hz, 0.3 g rms Non-operating...5 Hz to 500 Hz, 2.4 g rms (Tested in accordance with IEC-60068-2-64. Nonoperating test profile exceeds the requirements of MIL-PRF-28800F, Class 3.) Safety Voltage Connect only voltages that are within the following limits: Between any two terminals of an isolated channel...±120 VDC Isolation Channel to channel Continuous...300 V rms, Measurement Category II (Basic) Withstand...1,500 V rms, verified by 5 s dielectric withstand test 8 ni.com NI PXIe-4322 Specifications
Channel to earth ground Continuous... 300 V rms, Measurement Category II Withstand... 3,000 V rms, verified by a 5 s dielectric withstand test Measurement Category II is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level electrical distribution, such as that provided by a standard wall outlet, for example, 115 V for U.S. or 230 V for Europe. Caution Do not use for measurements within Measurement Categories III or IV. Safety Caution The protection provided by the NI PXIe-4322 can be impaired if it is used in a manner not described in this document. Caution When hazardous voltages (>30 V rms /42.4 V pk /60 VDC) are present on any terminal, safety low-voltage ( 30 V rms /42.4 V pk /60 VDC) cannot be connected to any other terminal. Caution Do not supply hazardous voltages (>30 V rms /42.4 V pk /60 VDC) to the terminal block without the terminal block being connected to the NI PXIe-4322. This product meets the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use: IEC 61010-1, EN 61010-1 UL 61010-1, CSA 61010-1 Note For UL and other safety certifications, refer to the product label or the Online Product Certification section. Electromagnetic Compatibility This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use: EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity EN 55011 (CISPR 11): Group 1, Class A emissions EN 55022 (CISPR 22): Class A emissions EN 55024 (CISPR 24): Immunity AS/NZS CISPR 11: Group 1, Class A emissions AS/NZS CISPR 22: Class A emissions FCC 47 CFR Part 15B: Class A emissions ICES-001: Class A emissions NI PXIe-4322 Specifications National Instruments 9
Note In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia and New Zealand (per CISPR 11) Class A equipment is intended for use only in heavy-industrial locations. Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes. Note For EMC declarations and certifications, and additional information, refer to the Online Product Certification section. CE Compliance This product meets the essential requirements of applicable European Directives as follows: 2006/95/EC; Low-Voltage Directive (safety) 2004/108/EC; Electromagnetic Compatibility Directive (EMC) Online Product Certification Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/ certification, search by model number or product line, and click the appropriate link in the Certification column. Environmental Management NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers. For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document. 10 ni.com NI PXIe-4322 Specifications
Waste Electrical and Electronic Equipment (WEEE) EU Customers At the end of the product life cycle, all products must be sent to a WEEE recycling center. For more information about WEEE recycling centers, National Instruments WEEE initiatives, and compliance with WEEE Directive 2002/96/EC on Waste and Electronic Equipment, visit ni.com/environment/ weee. RoHS National Instruments (RoHS) National Instruments RoHS ni.com/ environment/rohs_china (For information about China RoHS compliance, go to ni.com/environment/rohs_china.) Where to Go for Support The National Instruments Web site is your complete resource for technical support. At ni.com/ support you have access to everything from troubleshooting and application development self-help resources to email and phone assistance from NI Application Engineers. National Instruments corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas, 78759-3504. National Instruments also has offices located around the world to help address your support needs. For telephone support in the United States, create your service request at ni.com/support and follow the calling instructions or dial 512 795 8248. For telephone support outside the United States, visit the Worldwide Offices section of ni.com/ niglobal to access the branch office Web sites, which provide up-to-date contact information, support phone numbers, email addresses, and current events. NI PXIe-4322 Specifications National Instruments 11
Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for more information on National Instruments trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patents Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. 2013 National Instruments. All rights reserved. 375956B-01 May13