Calibration of High-Voltage Test Equipment

Similar documents
PRODUCT PORTFOLIO TEST AND MEASURING SYSTEMS

IMPORTANCE OF ACCURATE MEASUREMENTS DURING THE LIFE CYCLE OF UTILITIES

High-Voltage Capacitive Compensation Unit for Transformer and Shunt Reactor Testing, Type HVCC

SYSTEMS AND COMPONENTS FOR CABLE TESTING

HV Module Systems for Testing, Training and Research

Error vs. Uncertainty Historical Perspective

Field Measurement of Transmission Cable Dissipation Factor

Calibration Laboratory Assessment Service CLAS Certificate Number Page 1 of 10

Technical Questionnaire 9.101/6 Transformer Test System. Personal Data. Application. Test of. Delivery scope

Power Quality Measurements the Importance of Traceable Calibration

LOVAG GENERAL INSTRUCTION G2 MEASUREMENT UNCERTAINTY. This instruction of a general nature and does not relate to specific standards.

Mitigating Murphy s Law While Test. Frédéric Dollinger

Peter Dack, Vice President Sales & Marketing Radian Research, Inc.

Prepared by Mick Maytum

Checklist for electromechanical and electronic active energy meters in accordance with the MID Directive (2004/22/EC)

Lotus Automation (Ireland) Ltd. Trading As. LotusWorks. Building 3, Finisklin Business Park, Sligo. Calibration Laboratory Registration number: 277C

Table of Contents. Compendium SPEKTRA. Calibration Systems CS18. Vibration and Shock Exciters. Vibration Control Systems. Services

A New Method for the Calibration of the mv Ranges of an AC Measurement Standard

Operating Instructions / Voltage Tester per IEC with Phase and Polarity Tester and Phase Sequence Indicator

34134A AC/DC DMM Current Probe. User s Guide. Publication number April 2009

Lotus Automation (Ireland) Ltd. Trading As. LotusWorks. Building 3, Finisklin Business Park, Sligo. Calibration Laboratory Registration number: 277C

INTERNATIONAL STANDARD

IEC/CIGRE UHV Symposium Beijing Paper 4.2. Challenges on the measuring and testing techniques for UHV AC and DC equipment

MEASUREMENT AND STANDARDS

PHV RO. High impedance passive probe. Features: CeramCore TM Hybrid Probe. Modular Construction. Coaxial Design

SYSTEMS AND COMPONENTS FOR TRANSFORMER TESTING

Supplied in carry case with additional accessories The PHV 1000-RO is a 400 MHz, standard sized, 100:1 passive probe designed for instruments

Test Report CESI A6/ Approved Page 1. MF Trasformatori s.r.l. - Calcinato (BS) Italy

EU-TYPE EXAMINATION CERTIFICATE

PHV 1000-RO. High impedance passive probe. Features: CeramCore TM hybrid probe. Modular construction. Coaxial design

Microphone calibration service for airborne ultrasound

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

CALIBRATION OF LASER VIBROMETER STANDARDS ACCORDING TO ISO

Moving Test - MT3000

Resistive capacitive voltage divider

Lecture 36 Measurements of High Voltages (cont) (Refer Slide Time: 00:14)

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

Bilateral Comparison EURAMET.AUV.V-K1.2. (vibration acceleration) Swiss Federal Office of Metrology METAS Christian Hof

POWER TRANSFORMER SPECIFICATION, DESIGN, QUALITY CONTROL AND TESTING 18 MARCH 2009

RAP Ripple Control Coupling for parallel injection of RC signals in Medium and High Voltage Networks

Power Engineering II. High Voltage Testing

For the National Voluntary Laboratory Accreditation Program

TT 0710/03. TYPE TEST OF 3 Phase Solid Earthing Bonding Link Box, Type: LB.W.3SA.3.1

TR CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF ELECTRICAL DCLF METROLOGY. Approved By: Senior Manager: Mpho Phaloane Revised By:

Amendment to Test Report

Traceability for Oscilloscopes and Oscilloscope Calibrators

HV AC TESTING OF SUPER-LONG CABLES

Modular and Upgradable Fully Automated High Resistance Measurement Systems

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Signature 1100H+ Performance Verification Manual Version July 11, 2013

Reconfiguration of 3 MV Marx Generator into a Modern High Efficiency System

Keysight Technologies Optical Power Meter Head Special Calibrations. Brochure

EFFECT OF INTEGRATION ERROR ON PARTIAL DISCHARGE MEASUREMENTS ON CAST RESIN TRANSFORMERS. C. Ceretta, R. Gobbo, G. Pesavento

Capacitive voltage transformers

CNY17 Series. Optocoupler with Phototransistor Output. Description. Applications. Order Nos. and Classification table is on sheet 2.

IEC Electrical fast transient / Burst immunity test

Coherence and time-frequency analysis of impulse voltage and current measurements

High Voltage Engineering

TCET1100/ TCET1100G. Pb Pb-free. Optocoupler, Phototransistor Output, High Temperature. Vishay Semiconductors

Certificate of Calibration

INTERNATIONAL. High-voltage test techniques Partial discharge measurements

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: MODEL: Multimeter Service Information

About the power standards calibration in the National Institute of Metrology, Romania

VARIABLE FREQUENCY RESONANT TEST SYSTEM

PXIe Contents. Required Software CALIBRATION PROCEDURE

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

INTERNATIONAL STANDARD

MEASUREMENT OF THE INTERNAL INDUCTANCE OF IMPULSE VOLTAGE GENERATORS AND THE LIMITS OF LI FRONT TIMES

2-3 Calibration of Standard Voltage and Current Generator

Temperature References for Highest Accuracy Industrial Thermocouple Measurements

Modeling and Practical Suggestions to Improve ESD Immunity Test Repeatability

OP735. Benchtop Optical Power Meter Instruction Manual

Organisation Internationale de Métrologie Légale

CERTIFICATE OF ACCREDITATION

Baro Transmitter PTB x1.073

Electric Stresses on Surge Arrester Insulation under Standard and

Borderline between CIPM MRA and testing activities

Digital temperature transmitter for resistance sensors Model T15.H, head mounting version Model T15.R, rail mounting version

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

Calibration Specialists Ltd

TEST REPORT. Table Of Contents

WaveLink High Bandwidth Differential Probing System (16 GHz 25 GHz)

Resistive capacitive voltage divider

TECHNICAL SPECIFICATION

DM-45 Digital Multimeter

High-Voltage Test and

MAVOWATT 4 Multifunction Power Meter

ILAC input to CIPM MRA Review Workshop October 2015

SITRANS F flowmeters. SITRANS F M System information MAGFLO electromagnetic flowmeters. 4/18 Siemens FI

OIML R 130 RECOMMENDATION. Edition 2001 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Octave-band and one-third-octave-band filters

Commissioning Manual Distribution Transformers - Three-phase Oil Filled

CNY75(G) Series. Optocoupler with Phototransistor Output. Description. Applications. Order Nos. and Classification table is on sheet 2.

CIPM and CCPR What are these organizations and how do they affect my testing results. Maria Nadal Photometry, Surface Color and Appearance NIST

NEMA Standards Publication ICS Adjustable Speed Electrical Power Drive Systems

3.2 Measurement of high voltages

INTERNATIONAL STANDARD

For the National Voluntary Laboratory Accreditation Program

HiAS TM 744 HIGHEST RESOLUTION IMPULSE ANALYZING SYSTEM

TECHNICAL SPECIFICATION

Transcription:

Workshop 2000, Alexandria, Virginia, 13 & 14 September 2000 paper No.: 6 of High-Voltage Test Equipment Uwe Clauss, Stefan Maucksch, HIGHVOLT Prüftechnik Dresden GmbH, Dresden, Germany 1. Abstract The main key for accurate measurements of voltages and currents in high voltage applications is the performance of the measuring system. It requires a precise traceable to Standards. International Standards, e.g. IEC 60060-2 or IEEE Standard no.4-1995 define the configuration of measuring systems (MS) and the demands on the performance as well as the principal methods. With respect to the acquisition of valid data during HV tests it must be in the interest of the operator to assure the correct performance of the HV measuring system; i.e. accurate. The scientific background of the of HV measuring equipment has been considered in many publications. This paper is written especially for the users of MS as an introduction to some theoretical and practical aspects of s. 2. Introduction to HV The user of measuring equipment in general, and in high voltage applications in particular, is confronted with the requirements of technical and quality standards, that require the use of calibrated equipment. In addition the must be traceable to Standards /1-3/. The term traceability is the property of a result of a measurement whereby it can be related to appropriate standards, generally national or international standards, through an unbroken chain of comparisons. The result is a hierarchy, as shown in Figure 1. A measurement result is complete only when accompanied by a quantitative statement of its uncertainty. The uncertainty is required in order to decide if the result is adequate for its intended purpose and to ascertain if it is consistent with other similar results. Typical uncertainty levels for high voltage quantities are for MS of test equipment /1/, 1 % for Reference Measuring Systems (RMS) /1/, 0.5 % for Reference Standards, < 0.5 % for Standard (e.g. Germany PTB: DC 0.001 %, AC 0.02 %, LI/SI 0.). Figure 1: hierarchy /4/ Illustrated in table 1 is how different levels of the hierarchy interact with the metrological infrastructure. Copyright HV Technologies, Inc. - 2000 6-1

For the MS used in the HV test field of a company at least the first must be performed by an laboratory (see IEC 60060-2, annex A). If a company operates a large number of MS, the periodical re of the MS may be performed by an inhouse section, that operates RMS calibrated by a Institute or an laboratory. Normally, this company's section is granted the right to calibrate the companies own systems after it is inspected by an accreditation body. The in-house section must have a close cooperation with an laboratory to ensure the performance of the RMS. Accredited laboratories perform most of the routine work for HV test fields of companies and in-house sections. They operate reference standards which are industrial manufacture standards with the best possible characteristic. Institute Accredited laboratories In-house section HV test department Tasks To maintain and disseminate the Standards To safeguard the metrological infrastructure of a country Supervision of measuring equipment for in-house purposes Measurements and tests as part of quality assurance measures Basis for the or measurement Statutory duty to represent SI units and ensure international compatibility Institute or another laboratory Institute or an laboratory laboratory or Factory certificate, mark or the like Documentation of the or measurement certificate for reference standard certificate for working standard or factory standard Factory certificate, mark or the like for test equipment Test mark or like Table 1: system /4/ Accredited laboratories issue certificates with a predefined structure ( object, method, measuring conditions, measuring results, uncertainty of measurement). The mutual recognition of these certificates is ensured by multilateral agreements of the accreditation bodies. Institutes operate very special standards to achieve the lowest possible uncertainty. They are responsible for representing the SI units and have to assure the international compatibility. 3. Equipment for in-house sections The selection of the equipment for an in-house section depends on the type of voltage to be measured and on the range of testing voltages applied for quality assurance measures. In general, the should meet, if technically possible, the range of testing voltages. Because of the difficulties to meet the uncertainty level of 1 % for RMS at higher voltages, the standard IEC 60060-2 has made the compromise to allow s at a lower voltage level. The voltage level can be as low as 20 % of the rated testing voltage (MS for 630 kv AC can be tested with a minimum of 126 kv). The questions are: is there only one type or are there more types of voltage to be measured and what kind of RMS should be chosen? The following recommendations are intended to answer some of those questions: RMS for single types of voltages: 300 kv-dc RMS with Precision Divider and Peak Voltmeter (MR 0.5/300, MU 15) For the HV arm of the divider highly stable, non-inductive HV resistors are used. The low voltage arm is designed as a tap of the high voltage part. Therefore the high voltage and the low voltage arms have the same temperature coefficient of voltage of 0.005 % and consequently the scale factor may be considered as temperature stabilized. The divider is arranged inside an insulating glass fiber reinforced tube. Creepage currents over the tube are flowing directly to earth and have no effect on the scale factor of the divider. Creepage currents over the surface of the divider, e.g. in case of high ambient humidity, are avoided by the self heating of the divider (temperature on the surface during operation, approx. 60 C is much higher as the ambient temperature). HV Technologies, Inc. - Workshop 2000, 13 & 14 September, Alexandria, Virginia 6-2

200 kv-ac RMS with special Measuring Capacitor and Peak Voltmeter (MCF 600/200, MU 15) Voltage dependent non-linearity is low due to the use of an advanced capacitor technology CDCT Low temperature coefficient because of compensation of the temperature coefficient of high voltage arm and low voltage arm The proximity effect may be neglected because the stray capacitances are much lower than the capacitance of the measuring capacitor. 700 kv-li RMS with Controlled Resistive Lightning Impulse Voltage Divider and Digital Recorder (SMR 10/700, TR-AS 100-10) The proximity effect is compensated by the capacitive grading of the voltage distribution over the divider. The construction and the materials of the divider ensure the compensation of the temperature influence on the scale factor and avoid aging of the divider resistors. As digital transient recorder the type TR-AS 100-10 (100 Ms/s, 10 bit, 2 channels) is used. RMS for more than one type of voltages Universal RMS based on a Damped Capacitive Divider with parallel resistive path and Digital Recorder (MCR 600/500spec, TR-AS 100-10) Adapted system for 500 kv LI and SI, 200 kv AC and 200 kv DC Only one instrument to measure the relevant voltage quantities at LI, SI, AC and DC Calibrators In addition to the RMS, calibrators are required for s and adjustments of the instruments. For AC/ DC, precision Fluke calibrators (uncertainty < 0.05 %) and for LI/SI, the impulse calibrator Dr. Strauss KAL 1000 (uncertainty 0.7 %) are recommended. Software and know-how package CALLAB The package contains the software for calculation of measuring deviations, scale factors, uncertainty of measurement and the working instructions to perform s. 4. procedure For the user of a measuring system the question arises; what does he have to know about the procedure? The answer depends on the kind of system. New manufactured MS The first of a new manufactured MS may be performed by an laboratory at the factory (manufacturer), because in most of the cases the conditions at the factory are similar to the conditions at the user's location. Only in cases where the MS will be built into existing test equipment (e.g. AC cascade transformers) the has to be performed on-site at the user's location. Existing MS With the exception of small transportable MS, the has to be performed on-site at the users location. Preparing the, first it is necessary to check if the MS is ready for. It needs to be checked if the configuration of the MS meets requirements of IEC 60060-2 and if the instrument is functioning properly. For support and help you might contact a laboratory. To ensure a successful the voltage generators must fulfil the demands on the voltage shape according to IEC 60060-1. The HV test field must have sufficient space for the arrangement of the RMS. Ambient temperature and humidity on-site must be within a range for which the metrological behavior of the RMS is known. Especially high humidity (> 80 %) should be avoided. If all conditions are fulfilled you can proceed with the. Average duration times for the are: systems up to 1 MV one day, systems > 1 MV two days. After the the user gets a certificate which contains a conformity declaration as described in IEC 60060-2. If the conformity was proved by the measurements the MS is considered as an Approved Measuring System (AMS) for the applied conditions. The certificate is valid for the moment of. No laboratory can foresee the long term stability. Therefore the user is responsible for the determination of a suitable HV Technologies, Inc. - Workshop 2000, 13 & 14 September, Alexandria, Virginia 6-3

re- period. The maximum re- period is 5 years according to IEC 60060-2, if performance checks demonstrate that the long term stability is within the range of ± 1%. 5. Performance check of MS For most of the users there are no additional AMS or RMS available to proceed with the performance check by using the comparison measurement method. Therefore the performance check has to be performed by measurement of the output voltage of the test system under well defined conditions. The results of this measurement have to be compared with the results of the first performance check. Conditions Test system: without load or with defined load, defined input voltage Measuring system: defined measuring range and setting Proceeding Observation and recording of the ambient temperature and humidity and of the output voltage of the test system (10 observations) Calculation of the mean value of the output voltage for the 10 observations Calculation of the relative deviation U URef δu = x100% URef U Ref voltage determined in the first performance check U mean value of the 10 observations Evaluation The relative deviation should be within the range of ±1 %. If the condition is not fulfilled the reason has to be investigated. 6. Record of Performance All documents of the MS (principal circuit, technical data, routine and type test reports, certificates and the reports of the performance checks) are collected in the Record of Performance. It is in the responsibility of the operator of the measuring system to establish and maintain the record of performance. All new measuring systems delivered by HIGHVOLT are accompanied by a Record of Performance. It has the required structure and includes all existing documents. 7. Experiences of an laboratory (DKD-K-24501) /5/ New MS delivered by HIGHVOLT to HVT Type of MS Quantity Uncertainty Limit IEC 60060-2 300 kv-li 800 kv-li 1000 kv-li 1.2 % T 1, T 2 5.1 % 1.4 % T 1, T 2 6.0 % 1.0 % T 1, T 2 7.6 % 100 kv-ac / 2, U r.m.s. 0.8 % 100 kv-ac / 2, U r.m.s. 1.5 % Table 2: Uncertainties of MS calibrated by DKD-K-24501 The uncertainty of the measurements (Table 2) is well inside the limits. Existing MS If the existing MS is based on modern design there are no problems to meet the required limits of uncertainty. For older MS the weak points are the instruments and impulse voltage dividers without damping resistors. In some cases a bad characteristic of an instrument can be compensated by corrections. But in cases of analog oscilloscopes and instruments there are no chances to fulfill the required uncertainty. An older MS might be improved by a upgrade with new low voltage arms and instruments. 8. Conclusions The of MS must be considered as an element of the system of the country under consideration. regulations have to be taken into account. At HIGHVOLT MS of test equipment, RMS and service are available to fulfil the demands of IEC 60060-2. HV Technologies, Inc. - Workshop 2000, 13 & 14 September, Alexandria, Virginia 6-4

References /1/ International Electrotechnical Commission, High-voltage test Techniques IEC 60060-2, 1994 /2/ IEEE Power Engineering Society, IEEE Standard Techniques for High-Voltage Testing IEEE Std 4-1995, Oct 12, 1995 /3/ International Standard Organization ISO 9000 series publications about Quality Management Systems /4/ European cooperation for Accreditation of Laboratories (EAL) Traceability of Measuring and Test Equipment to Standards EAL-G12, edition 1, November 1995 /5/ of High Voltage Measuring Systems according to IEC 60060-2 Information sheet 10.3/2e HIGHVOLT Prüftechnik Dresden GmbH HV Technologies, Inc. - Workshop 2000, 13 & 14 September, Alexandria, Virginia 6-5