41P Portable Calibrator User Manual

Similar documents
45EMD Portable Calibrator User Manual

201AP Charge Amplifier User Manual

200B Clipper Module User Manual

DC200A Displacement Clipper User Manual

INSTRUCTION MANUAL LKG 601 Electrical Safety Analyzer

Model 7000 Low Noise Differential Preamplifier

34134A AC/DC DMM Current Probe. User s Guide. Publication number April 2009

2001A. 200KHz Function Generator Instruction Manual. 99 Washington Street Melrose, MA Phone Toll Free

Glass Electrode Meter

OPERATION & SERVICE MANUAL FOR FC 110 AC POWER SOURCE

Radio Remote(s) (Installation Manual)

Signal Conditioning Amplifier

Model MV106J/MV116J. ±10nVdc to ±11Vdc Precision DC Voltage Standard Source. Operating Manual

Signal Conditioning Amplifier

INSTRUMENTS, INC. Model 2960AX Disciplined Quartz Frequency Standard 2960AX. Section Page Contents

Series 500. Owner s Manual. Analog Transmitters by Data Industrial. Data Industrial. Data Industrial 2/95 PN 72806

INSTRUCTION MANUAL LKG

BC145 SIGNAL ISOLATOR BOARD

INSTRUCTION MANUAL FOR CELL SIMULATOR MODEL 2410

Model 9305 Fast Preamplifier Operating and Service Manual

Model 9302 Amplifier-Discriminator Operating and Service Manual

A53106 SERIES DC-TO-DC CONVERTER

ELECTROSURGICAL UNIT ANALYZER

Model 113 Scintillation Preamplifier Operating and Service Manual

TRANSDUCER IN-LINE AMPLIFIER

Safety. This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages may be present.

Advanced Test Equipment Rentals ATEC (2832)

Model 863 Quad Timing Filter Amplifier Operating and Service Manual

SI-125 Power Amplifier Manual 6205 Kestrel Road; Mississauga, Ontario; Canada; L5T 2A1 November 2016, Rev 0.5

Model Hz to 10MHz Precision Phasemeter. Operating Manual

Model 1791 VHF Radio User's Manual

MODEL 3810/2 Line Impedance Stabilization Network

ECG PATIENT SIMULATOR FIVE LEAD

Advanced Test Equipment Rentals ATEC (2832)

VT1586A Rack Mount Terminal Panel Installation and User s Manual

Model 8140VT VERSATAP INSTALLATION AND OPERATION MANUAL

SC200 Series. Signal Conditioner MNX10020, REV H 10/17/ Connection Technology Center, Inc Rae Boulevard Victor, NY (585)

80i-110s AC/DC CURRENT PROBE

CEC VIBRATION TRANSMITTER

PARALLEL MULTI-AMP KIT for 7200 Series AMPLIFIERS INSTRUCTION SHEET

SX204. Headphone Amplifier. Symetrix. Owner s Manual

Model 5100F. Advanced Test Equipment Rentals ATEC (2832) OWNER S MANUAL RF POWER AMPLIFIER

TRANSDUCER IN-LINE AMPLIFIER

15B+/17B+/18B+ Calibration Manual. Digital Multimeter

Wilcom MODEL T136BSBZW CIRCUIT TEST SET. Operating Instructions

Wilcom MODEL T136BGMZW CIRCUIT TEST SET. Operating Instructions

Model 34A. 3Hz to 2MHz 2-Channel Butterworth/Bessel HP, LP, BP, BR Plug-In Filter Card for Model 3905/3916 Chassis.

PULSE DISTRIBUTION AMPLIFIER OPERATING MANUAL

A Channel Amplifier

AMP-12 OPERATOR S MANUAL

Mobile Activation Case

User s Guide. MultiView Series Digital MultiMeters Models: MV110 MV120 MV130

FRYCORDER Power Cord Burn-in Generator

An American Control Electronics Brand PCM4 SERIES USER MANUAL PCM4.

HP 86290B RF PLUG-IN GHz HEWLETT PACKARD

VIM Series. 90 & 60 W, Efficient, CV Class 2 LED Drivers ORDERING INFORMATION

Model 3210C. 100 Ampere AC Current Standard. Operating Manual

R-X SERIES. Decade Resistor

PXIe-1209 Dual 100 MHz Pulse Generator

ECG PATIENT SIMULATOR SIX LEAD

200Amp AC Clamp Meter + NCV Model MA250

Operating Manual. Model Channel. DC Signal Conditioner Amplifier

VGM Series. 100 & 60 W, Efficient, CV Class 2 LED Drivers for Signage Applications ORDERING INFORMATION

IRRIGATION 810-T PLUS TRANSMITTER GUIDE

TEGAM, INC. SINGLE/DUAL CHANNEL HIGH VOLTAGE AMPLIFIER MODEL 2340/2350. Instruction Manual PN# CD Publication Date: June 2006 REV.

9 PIECE TUNGSTEN CARBIDE HOLE SAW KIT. Model 90721

LINCO MEASUREMENT MODEL CP-2B MASTER METER PROVER COUNTER INSTRUCTION MANUAL

Classic Series Public Address Amplifiers C10 & C20 Models

Demo kit by Cedrat Technologies PRODUCT AND WARRANTY INFORMATION

AC/DC Current Oscilloscope Probe Model SL261

Signal Conditioning Amplifier

AL8805EV1 EVALUATION BOARD USER GUIDE

Introduction. Specifications. Features. Controls. Model 103

P5100A & P5150 High Voltage Probes Performance Verification and Adjustments

Type N Coaxial Calibration Kit

PIEZO FILM LAB AMPLIFIER

DL102 Counter Loop Amplifier

1510A PRECISION SIGNAL SIMULATOR

INSTRUCTIONS PS2500 POWER SUPPLY

CSM-S USER S MANUAL TRIGGER DISTRIBUTION MODULE Release April 7, VXI Technology, Inc.

P5100A & P5150 High Voltage Probes Performance Verification and Adjustments

ASX SERIES PACIFIC OPERATION MANUAL AC POWER SOURCE POWER SOURCE

MODEL W Power Amplifier

HTA125A/250A. Power Amplifiers. Installation & Use Manual

PI-150 Broadband Power Indicator

POWER SUPPLY MODEL XP-720. Instruction Manual ELENCO

Shiba Drive USER MANUAL

Series APD4059. For Load Cells/Pressure Transducer Transmitters. Field Rangeable, with Calibration Resistor. Standard Features.

Applications AP7350 GND

HydroLynx Systems, Inc.

Copyright 2014 by Minarik Drives

TK 100 Probe Calibrator. Instruction Manual

Smart-UPS On-Line Isolation and Step-Down Transformer SRT5KTF

Pen Multimeter. Model

AM-510 Commercial / Residential Multimeter. AM-510-EUR Digital Multimeter. Users Manual

WRM-10 TM TRANSFORMER WINDING RESISTANCE METER

Bi-Directional P-Channel MOSFET/Power Switch

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE. ISO-STIM 01D STIMULUS ISOLATION UNIT ±100 V / ±10 ma, bipolar output

RHS SERIES RH Set Point Controller/Probe

Compact Autoranging Clamp Meters. Models (400 A AC), (400 A AC/DC)

Transcription:

Trig-Tek 41P Portable Calibrator User Manual Publication No. 980961 Rev. A Astronics Test Systems Inc. 4 Goodyear, Irvine, CA 92618 Tel: (800) 722-2528, (949) 859-8999; Fax: (949) 859-7139 atsinfo@astronics.com atssales@astronics.com atshelpdesk@astronics.com http://www.astronicstestsystems.com Copyright 2011 by Astronics Test Systems Inc. Printed in the United States of America. All rights reserved. This book or parts thereof may not be reproduced in any form without written permission of the publisher.

THANK YOU FOR PURCHASING THIS ASTRONICS TEST SYSTEMS PRODUCT For this product, or any other Astronics Test Systems product that incorporates software drivers, you may access our web site to verify and/or download the latest driver versions. The web address for driver downloads is: http://www.astronicstestsystems.com/support/downloads If you have any questions about software driver downloads or our privacy policy, please contact us at: atsinfo@astronics.com WARRANTY STATEMENT All Astronics Test Systems products are designed to exacting standards and manufactured in full compliance to our AS9100 Quality Management System processes. This warranty does not apply to defects resulting from any modification(s) of any product or part without Astronics Test Systems express written consent, or misuse of any product or part. The warranty also does not apply to fuses, software, non-rechargeable batteries, damage from battery leakage, or problems arising from normal wear, such as mechanical relay life, or failure to follow instructions. This warranty is in lieu of all other warranties, expressed or implied, including any implied warranty of merchantability or fitness for a particular use. The remedies provided herein are buyer s sole and exclusive remedies. For the specific terms of your standard warranty, contact Customer Support. Please have the following information available to facilitate service. 1. Product serial number 2. Product model number 3. Your company and contact information You may contact Customer Support by: E-Mail: atshelpdesk@astronics.com Telephone: +1 800 722 3262 (USA) Fax: +1 949 859 7139 (USA)

RETURN OF PRODUCT Authorization is required from Astronics Test Systems before you send us your product or sub-assembly for service or calibration. Call or contact Customer Support at 1-800-722-3262 or 1-949-859-8999 or via fax at 1-949-859-7139. We can also be reached at: atshelpdesk@astronics.com. If the original packing material is unavailable, ship the product or sub-assembly in an ESD shielding bag and use appropriate packing materials to surround and protect the product. PROPRIETARY NOTICE This document and the technical data herein disclosed, are proprietary to Astronics Test Systems, and shall not, without express written permission of Astronics Test Systems, be used in whole or in part to solicit quotations from a competitive source or used for manufacture by anyone other than Astronics Test Systems. The information herein has been developed at private expense, and may only be used for operation and maintenance reference purposes or for purposes of engineering evaluation and incorporation into technical specifications and other documents which specify procurement of products from Astronics Test Systems. TRADEMARKS AND SERVICE MARKS All trademarks and service marks used in this document are the property of their respective owners. Racal Instruments, Talon Instruments, Trig-Tek, ActivATE, Adapt-A-Switch, N-GEN, and PAWS are trademarks of Astronics Test Systems in the United States. DISCLAIMER Buyer acknowledges and agrees that it is responsible for the operation of the goods purchased and should ensure that they are used properly and in accordance with this document and any other instructions provided by Seller. Astronics Test Systems products are not specifically designed, manufactured or intended to be used as parts, assemblies or components in planning, construction, maintenance or operation of a nuclear facility, or in life support or safety critical applications in which the failure of the Astronics Test Systems product could create a situation where personal injury or death could occur. Should Buyer purchase Astronics Test Systems product for such unintended application, Buyer shall indemnify and hold Astronics Test Systems, its officers, employees, subsidiaries, affiliates and distributors harmless against all claims arising out of a claim for personal injury or death associated with such unintended use.

FOR YOUR SAFETY Before undertaking any troubleshooting, maintenance or exploratory procedure, read carefully the WARNINGS and CAUTION notices. This equipment contains voltage hazardous to human life and safety, and is capable of inflicting personal injury. If this instrument is to be powered from the AC line (mains) through an autotransformer, ensure the common connector is connected to the neutral (earth pole) of the power supply. Before operating the unit, ensure the conductor (green wire) is connected to the ground (earth) conductor of the power outlet. Do not use a two-conductor extension cord or a three-prong/two-prong adapter. This will defeat the protective feature of the third conductor in the power cord. Maintenance and calibration procedures sometimes call for operation of the unit with power applied and protective covers removed. Read the procedures and heed warnings to avoid live circuit points. Before operating this instrument: 1. Ensure the proper fuse is in place for the power source to operate. 2. Ensure all other devices connected to or in proximity to this instrument are properly grounded or connected to the protective third-wire earth ground. If the instrument: - fails to operate satisfactorily - shows visible damage - has been stored under unfavorable conditions - has sustained stress Do not operate until performance is checked by qualified personnel.

Publication No. 980961 Rev. A 41P User Manual Table of Contents Chapter 1... 1-1 Introduction... 1-1 Features... 1-2 Specifications... 1-2 Outputs (mv and pc)... 1-2 Controls... 1-2 Indicators... 1-3 Power... 1-3 Dimensions... 1-3 Chapter 2... 2-1 Operations... 2-1 FREQUENCY HZ POWER Switch... 2-1 UNITS mv pc Switch... 2-1 AMPLITUDE and X1-X10 Switches... 2-2 Rechargeable Battery... 2-2 Chapter 3... 3-1 Performance Test... 3-1 Test Equipment... 3-1 Switch Settings... 3-1 Performance Test... 3-1 Chapter 4... 4-1 Calibration Procedure... 4-1 Test Equipment... 4-1 Adjustments... 4-1 Calibration Procedures... 4-1 Astronics Test Systems i

41P User Manual Publication No. 980961 Rev. A List of Figures Figure 1-1, 41P Portable Calibrator...1-1 Figure 4-1: Potentiometer Adjustment Location Diagram...4-3 ii Astronics Test Systems

Publication No. 980961 Rev. A 41P User Manual DOCUMENT CHANGE HISTORY Revision Date Description of Change A 1/6/2011 Document Control release Astronics Test Systems iii

41P User Manual Publication No. 980961 Rev. A This page was left intentionally blank. iv Astronics Test Systems

Publication No. 980961 Rev. A 41P User Manual Chapter 1 Introduction The Trig-Tek 41P Portable Calibrator is a hand held signal generator with accurate frequency and level of mv and pc (millivolts and picocoulomb outputs). Primarily used in Environmental Test Laboratories and Engine Test cells, the 41P is ideal as a sensor simulator for remote calibration of monitoring instrumentation. Figure 1-1, 41P Portable Calibrator A crystal controlled oscillator provides five stable selectable frequencies: 100 and 1000 Hz, 61.4 Hz, 140 Hz and 318 Hz. Since the 41P has both mv and pc outputs it is possible to calibrate piezoelectric or charge mode amplifiers. A four-digit thumb switch combined with an X10 switch allows level selection from 0.1 to 10000 mv and PC, available in DC, Peak, Pk-Pk, and RMS units. An offset feature allows 0 to +12 Volts DC to be added to the AC signals. The rechargeable battery provides 8 hours of continuous service and has a low battery indicator on the front panel. Astronics Test Systems Introduction 1-1

41P User Manual Publication No. 980961 Rev. A Features Features include: Millivolts and picocoulomb outputs 0.1% frequency stability AC or DC output DC, Pk, Pk-Pk, RMS 0.1 to 10000 mv-pc High Accuracy DC Offset Specifications Outputs (mv and pc) Impedance Level Accuracy Frequencies Frequency Accuracy Units mv output < 50 Ohms 0.5% of reading ±0.1% of fullscale ± 3 LSD at 25 C ( ± 0.015% / C) 61.4, 140, 318, 100 and 1000 Hz 0.5% @ 25 C, 0.015% / C DC, RMS, Pk, and Pk-Pk Controls Frequency, Power Amplitude Selector Units DC Offset X1-X10 Switch Selects desired frequency or power ON/OFF Four-digit thumb switch allows level selection of 0.1 to 999.9 Selects DC, Peak, RMS, or Pk-Pk Variable from 1 to 12 VDC (positive or negative depending on position of UNITS switch In X10 position, output level is multiplied by 10 Introduction 1-2 Astronics Test Systems

Publication No. 980961 Rev. A 41P User Manual Indicators CHRG ON BATT LOW LED illuminates when charging. LED illuminates when low battery condition exists. Power 110 VAC with supplied charger 6 VDC rechargeable batteries Dimensions 2.5 high x 6 wide x 7 long Astronics Test Systems Introduction 1-3

41P User Manual Publication No. 980961 Rev. A This page was left intentionally blank.. Introduction 1-4 Astronics Test Systems

Publication No. 980961 Rev. A 41P User Manual Chapter 2 Operations The Trig-Tek 41P Portable Calibrator is contained in a high impact plastic case. It is lightweight and very durable. It provides settings of 000.1 to 999.9 mv or pc output levels with an X10 to provide 0001 to 9999 mv or pc at the two separate output jacks. The internal battery will provide more than 8 hours of operation before recharge. FREQUENCY HZ POWER Switch This switch has six positions, OFF, 61.4 Hz, 100 Hz, 140 Hz, 318 Hz, and 1000 Hz. The 100 Hz and 1000 Hz are frequencies widely used for checking various equipment. The other three frequencies are unique: 61.4 Hz is where Acceleration Peak is equal to Velocity Peak; 140 Hz is where Acceleration Peak is equal to Displacement Pk-Pk; and 318 Hz is where Velocity Peak is equal to Displacement Pk-Pk. The OFF position de-energizes the circuitry. UNITS mv pc Switch This switch has five positions. The first is DC. When the switch is in this position, only DC voltages appear at the mv (millivolt) output. The output level of the DC is determined by the AMPLITUDE switch, and can provide outputs of 000.1 to 999.9 millivolts or 0001 to 9999 millivolts DC in two ranges by placing the X1-X10 Switch to X10. The next position of the switch is PEAK. When in this position the AMPLITUDE is calibrated for millivolts peak at the mv output. The output at the BNC connector labeled pc (picocoulomb) is a charge output equivalent to the applied voltage through a precision capacitor. The precision 0.3% capacitor is 1000 pf. Hence, if 100 mv PEAK is set on the front panel switch, a voltage of 100 mv is applied through the 1000 pf capacitor providing an equivalent charge of 100 pc PK at the BNC PC output. The third position of the switch is PK-PK (minus). In this position the mv and pc outputs are calibrated in PK-PK units. While in the PK-PK (minus) position, the mv output can be offset 0 to -12 VDC adjustable by the DC OFFSET control to the right of the UNITS mv-pc switch. Note: When no DC offset is needed, be sure the DC OFFSET control is turned completely counterclockwise to 0. The fourth position of the switch is PK-PK+ (plus). In this position the mv and pc outputs are calibrated in PK-PK units. While in the PK-PK+ position, the mv output Astronics Test Systems Operations 2-1

41P User Manual Publication No. 980961 Rev. A can be offset using the DC SET control for the 0 to +12 VDC. The last switch position is RMS+ (plus). In this position the mv and pc outputs are calibrated in RMS units and the output can be offset 0 to +12 VDC using the OFFSET control. AMPLITUDE and X1-X10 Switches The AMPLITUDE four-digit thumbwheel switch providing settings from 000.0 to 999.9 mv or pc amplitude for the units selected by the UNITS mv-pc switch. The 000.1 to 999.9 amplitude setting can be multiplied by the X1-X10 switch above the AMPLITUDE switch. When switched to X1 the amplitude is 000.0 to 999.9 mv and pc. In the X10 position, the amplitudes of 0000 to 9999 mv and pc can be set. Rechargeable Battery The 41P Portable Calibrator uses a six-volt rechargeable battery pack. The battery supplies power to the 41P for up to 8 hours when fully charged. In order to recharge the battery, it is necessary to plug the charge adapter into 115 VAC power and then connect it to the 41P. To extend the life of the battery, the calibrator should be completely discharged before recharging. The battery requires ten hours to fully charge. The 41P provides circuitry such that the charger may be left plugged-in indefinitely without damaging the battery. When the low battery (BATT LOW) LED illuminates, it does not necessarily mean the battery is fully discharged but that the battery s voltage has dropped below what is required to maintain the specifications for the 41P. The 41P will probably continue to operate, but the output voltages may not be within specifications. The unit can be operated with the charger plugged-in. Operations 2-2 Astronics Test Systems

Publication No. 980961 Rev. A 41P User Manual Chapter 3 Performance Test The following performance test procedure is a method of testing the 41P Portable Calibrator unit to verify that it is performing within the manufacturer's specifications. The unit uses integrated circuits and very stable parts, so should not require calibration more often than every twelve months. In the event that a reading is out of tolerance, refer to Chapter 4 of this manual. Test Equipment Note: Equivalent equipment can be used. 1. AC-DC Digital Multimeter Keithley Model 191 2. Frequency Counter Leader Model LDC-822 Switch Settings Set the FREQUENCY switch to 318 Hz, the AMPLITUDE switch to indicate 9999, and the UNITS switch to DC. Set the X1-X10 switch for X10. Turn the OFFSET knob to the full counterclockwise position ( 0 ). Performance Test 1. Connect the Digital Multimeter to the mv OUTPUT (BNC) jack. 2. Observe an indication of 9999 ±60 mvdc on the multimeter. 3. Set the X1-X10 switch to X1. 4. Set the AMPLITUDE switch to 0000. 5. Observe an indication of 0 ±5.0 mvdc. 6. Set the AMPLITUDE switch to 100.0. 7. Observe an indication of 100 ±10.5 mvdc. 8. Set the AMPLITUDE Switch to 010.0. 9. Observe an indication of 10.0 ±1.5 mvdc. 10. Set the AMPLITUDE switch to 900.0, and the UNITS switch to RMS+. 11. Observe an indication of 900 ±14.5 mvac RMS. 12. Set the UNITS switch to PEAK. Astronics Test Systems Performance Test 3-1

41P User Manual Publication No. 980961 Rev. A 13. Observe an indication of 636 ±3.2 mvac RMS. 14. Set the UNITS switch to PK-PK+. 15. Observe an indication of 318 ±1.6 mvac RMS. 16. Set the FREQUENCY switch to 61.4 Hz. 17. Connect the Frequency Counter to the mv OUTPUT (BNC) jack. 18. Observe an indication of 61.44 ±0.31 Hz on the Frequency Counter. 19. Set the FREQUENCY switch to 140 Hz. 20. Observe an indication of 139.9 ±0.696 Hz. 21. Set the FREQUENCY switch to 318 Hz. 22. Observe an indication of 318 ±1.59 Hz. 23. Set the FREQUENCY switch to 100.0 Hz. 24. Observe an indication of 100.0 ±0.5 Hz. 25. Set the FREQ switch to 1000 Hz. 26. Observe an indication of 1000 ±5 Hz. Performance Test 3-2 Astronics Test Systems

Publication No. 980961 Rev. A 41P User Manual Chapter 4 Calibration Procedure The unit has been carefully calibrated using frequency counters and multimeter that are traceable to thenational Bureau of Standards. Do not make any adjustments unless properly calibrated equipment is available. Make sure the battery is fully charged. Test Equipment Note: Equivalent equipment can be used. 1. AC-DC Digital Multimeter Keithley Model 191 2. Frequency Counter Leader Model LDC-822 Adjustments There are four separate adjustments inside the 41P Portable Calibrator. Refer to Figure 4-1 for potentiometer locations on the circuit board inside the case. 1. Remove the back cover of the 41P by taking off the six screws found on the back of the calibrator. 2. Pull the cover back off. Place the unit in front of you with the front panel down. 3. Looking down into the calibrator circuit board, you will see the four potentiometers which provide the following adjustments: R1 RMS Level R2 PEAK Level R3 DC Zero R4 DC Level Calibration Procedures 1. Connect a digital multimeter to the mv (BNC) OUTPUT jack. 2. Set FREQUENCY switch to 318 Hz. 3. Place the UNITS switch to DC. Set the DC OFFSET Control completely counterclockwise to 0. 4. Set the AMPLITUDE switch to 0000. Astronics Test Systems Calibration Procedure 4-1

41P User Manual Publication No. 980961 Rev. A 5. Adjust the DC ZERO Adjustment R3 for a 0 ±1.0 mvdc indication on the multimeter. 6. Set the AMPLITUDE switch to 900. Set X1-X10 switch to X1. 7. Set the DC LEVEL Adjustment R4 for a 900 ±1.0 mv indication on the multimeter. 8. Place the UNITS switch to RMS. Make sure the DC OFFSET Control is in the full counterclockwise ( 0 ) position. 9. Turn the FREQ Switch to 318 Hz. 10. Set the RMS Level Adjustment R1 for an indication of 900 ±1.0 mv RMS on the multimeter. 11. Place the UNITS switch to PEAK. 12. Set the PEAK Level Adjustment R2 for a 636.3 k ±1.0 mv RMS indication on the multimeter. Note: A high grade 0.3% precision 1000 pf capacitor is in series with the BNC output connector. The above calibrations have automatically adjusted the pc (picocoulomb) output, as these calibrated voltages are used via the capacitor to the pc (picocoulomb) output. Calibration Procedure 4-2 Astronics Test Systems

Publication No. 980961 Rev. A 41P User Manual Figure 4-1: Potentiometer Adjustment Location Diagram Astronics Test Systems Calibration Procedure 4-3

41P User Manual Publication No. 980961 Rev. A This page was left intentionally blank. Calibration Procedure 4-4 Astronics Test Systems