Improving a commercially available heterodyne laser interferometer to sub-nm uncertainty

Similar documents
A Fast Phase meter for Interferometric Applications with an Accuracy in the Picometer Regime

Absolute distance interferometer in LaserTracer geometry

A Multiwavelength Interferometer for Geodetic Lengths

Stability of a Fiber-Fed Heterodyne Interferometer

Real-time periodic error correction: experiment and data analysis

Jones matrix analysis of high-precision displacement measuring interferometers

Fiber characterization for application in heterodyne laser interferometry with nanometer uncertainty, part I: polarization state measurements

Department of Mechanical Engineering and Automation, Harbin Institute of Technology Shenzhen Graduate School, Shenzhen, , China

Bench-top setup for validation of real time, digital periodic error correction

INTERFEROMETRIC VIBRATION DISPLACEMENT MEASUREMENT

CHAPTER 5 FINE-TUNING OF AN ECDL WITH AN INTRACAVITY LIQUID CRYSTAL ELEMENT

Module 5: Experimental Modal Analysis for SHM Lecture 36: Laser doppler vibrometry. The Lecture Contains: Laser Doppler Vibrometry

LASER VIBROMETER CALIBRATION AT HIGH FREQUENCIES USING CONVENTIONAL CALIBRATION EQUIPMENT

ADALAM Sensor based adaptive laser micromachining using ultrashort pulse lasers for zero-failure manufacturing D2.2. Ger Folkersma (Demcon)

High stability multiplexed fibre interferometer and its application on absolute displacement measurement and on-line surface metrology

Agilent 5527A/B-2 Achieving Maximum Accuracy and Repeatability

Sub-millimeter Wave Planar Near-field Antenna Testing

CALIBRATION OF LASER VIBROMETER STANDARDS ACCORDING TO ISO

NEW LASER ULTRASONIC INTERFEROMETER FOR INDUSTRIAL APPLICATIONS B.Pouet and S.Breugnot Bossa Nova Technologies; Venice, CA, USA

Polarization Experiments Using Jones Calculus

A precise and wide-dynamic-range displacement-measuring homodyne quadrature laser interferometer

Periodic Error Correction in Heterodyne Interferometry

University of Huddersfield Repository

Fabry Perot Resonator (CA-1140)

Fiber Optic Sensing Applications Based on Optical Propagation Mode Time Delay Measurement

PREPARED BY: I. Miller DATE: 2004 May 23 CO-OWNERS REVISED DATE OF ISSUE/CHANGED PAGES

a 1550nm telemeter for outdoor application based on off-the-shelf components

Installation and Characterization of the Advanced LIGO 200 Watt PSL

ULTRASONIC TRANSDUCER PEAK-TO-PEAK OPTICAL MEASUREMENT

Results from the Stanford 10 m Sagnac interferometer

Compensation for the Variable Cyclic Error in Homodyne Laser Interferometers

la. Smith and C.P. Burger Department of Mechanical Engineering Texas A&M University College Station Tx

Stabilizing an Interferometric Delay with PI Control

Development of C-Mod FIR Polarimeter*

EVALUATION OF A FOCUSED LASER SPOT DIAMETER FOR AN OPTICAL ANGLE SENSOR. Yuki Shimizu, Taiji Maruyama, Shota Nakagawa, Yuan-Liu Chen, Wei Gao

Multiply Resonant EOM for the LIGO 40-meter Interferometer

Agilent 10774A Short Range Straightness Optics and Agilent 10775A Long Range Straightness Optics

Experiment O11e Optical Polarisation

Swept Wavelength Testing:

Nanometer-level repeatable metrology using the Nanoruler

High Sensitivity Interferometric Detection of Partial Discharges for High Power Transformer Applications

ECE 185 ELECTRO-OPTIC MODULATION OF LIGHT

Dynamic Phase-Shifting Electronic Speckle Pattern Interferometer

Agilent 10705A Single Beam Interferometer and Agilent 10704A Retroreflector

9. Microwaves. 9.1 Introduction. Safety consideration

LOS 1 LASER OPTICS SET

Verification of large CMMs - Artefacts and Methods - Eugen Trapet, Spain (Trapet Precision and ISM3D)

Initial Results from the C-Mod Prototype Polarimeter/Interferometer

Laser interferometric measuring system for positioning in nanometrology

Absolute distance interferometer in LaserTracer geometry

echo-based range sensing L06Ua echo-based range sensing 1

Development of innovative fringe locking strategies for vibration-resistant white light vertical scanning interferometry (VSI)

SUPPLEMENTARY INFORMATION DOI: /NPHOTON

Polarization Sagnac interferometer with a common-path local oscillator for heterodyne detection

Pre-Lab 10. Which plan or plans would work? Explain. Which plan is most efficient in regard to light power with the correct polarization? Explain.

Designing Optical Layouts for AEI s 10 meter Prototype. Stephanie Wiele August 5, 2008

COATS: compact optical 5DoF attitude sensor for space applications

Spectrally resolved frequency comb interferometry for long distance measurement

3.0 Alignment Equipment and Diagnostic Tools:

A TECHNIQUE TO EVALUATE THE IMPACT OF FLEX CABLE PHASE INSTABILITY ON mm-wave PLANAR NEAR-FIELD MEASUREMENT ACCURACIES

SPRAY DROPLET SIZE MEASUREMENT

Frequency stabilized three mode HeNe laser using nonlinear optical phenomena

Status of the BIPM Watt Balance

Wavelength Control and Locking with Sub-MHz Precision

Frequency stability and reproducibility of iodine stabilised He-Ne laser at 633 nm

How to Build a Gravitational Wave Detector. Sean Leavey

Active Vibration Control in Ultrasonic Wire Bonding Improving Bondability on Demanding Surfaces

Interferometric OSEM Sensor Development

attosnom I: Topography and Force Images NANOSCOPY APPLICATION NOTE M06 RELATED PRODUCTS G

Optics and Laser Heads for Laser-Interferometer Positioning Systems Product Overview

Optical generation of frequency stable mm-wave radiation using diode laser pumped Nd:YAG lasers

1.6 Beam Wander vs. Image Jitter

Keysight Technologies Optics and Laser Heads for Laser-Interferometer Positioning Systems

Lab 5 - Electro-Optic Modulation

Bias errors in PIV: the pixel locking effect revisited.

Synthesis of projection lithography for low k1 via interferometry

Advances in laboratory modeling of wave propagation

NCSL International 2995 Wilderness Place, Suite 107 Boulder, Colorado Office: (303) Fax: (303)

REAL TIME THICKNESS MEASUREMENT OF A MOVING WIRE

UPGRADE OF THE MEDIUM AND HIGH FREQUENCY VIBRATION CALIBRATION REFERENCE EQUIPMENT AND EXTENSION TO LOW FREQUENCIES

D.C. Emmony, M.W. Godfrey and R.G. White

A refractivity-compensated absolute distance interferometer as prospective novel primary standard for baseline calibrations

Figure 4.1 Vector representation of magnetic field.

Two-Mode Frequency Stabilization of an Internal-Mirror 612 nm He-Ne Laser

Single Photon Interference Laboratory

CALIBRATION PROCEDURE FOR STABILIZED LASERS USING THE METHOD OF OPTICAL BEATS MEASUREMENT UNCERTAINTY

Investigation of an optical sensor for small angle detection

CO2 laser heating system for thermal compensation of test masses in high power optical cavities. Submitted by: SHUBHAM KUMAR to Prof.

visibility values: 1) V1=0.5 2) V2=0.9 3) V3=0.99 b) In the three cases considered, what are the values of FSR (Free Spectral Range) and

Experimental Competition

Development of a Low Cost 3x3 Coupler. Mach-Zehnder Interferometric Optical Fibre Vibration. Sensor

Precision displacement interferometry with stabilization of wavelength on air

PREPARED BY: I. Miller DATE: 2004 May 23 CO-OWNERS REVISED DATE OF ISSUE/CHANGED PAGES

6 Experiment II: Law of Reflection

International Conference on Space Optics ICSO 2000 Toulouse Labège, France 5 7 December 2000

7 WAVEMETER PROJECT #6 MODEL OEK-100. Measure the Wavelength of An Unknown laser Using 633nm and 543 nm HeNe lasers

3B SCIENTIFIC PHYSICS

Heterodyne interferometric technique for displacement control at the nanometric scale

MICROMACHINED INTERFEROMETER FOR MEMS METROLOGY

Amplitude independent RF instantaneous frequency measurement system using photonic Hilbert transform

Transcription:

Improving a commercially available heterodyne laser interferometer to sub-nm uncertainty H. Haitjema, S.J.A.G. Cosijns, N.J.J. Roset and M.J.Jansen Eindhoven University of Technology, PO Box 513, 56 MB Eindhoven, The Netherlands ABSTRACT Laser interferometer systems are known for their high resolution, and especially for their high range/resolution ratio. In dimensional metrology laboratories, laser interferometers are popular workhorses for the calibration of displacements. The uncertainty is usually limited to about 1 nm due to polarization- and frequency mixing. For demanding applications however nanometer uncertainty is desired. We adapted a commercially available heterodyne laser interferometer by feeding the measurement signal into a fast lock-in amplifier and use the laser interferometer reference signal as a reference. By measuring both the in-phase and quadrature component an uncorrected phase can be directly measured. By recording both components while the phase changes between andπ a typical ellipse is recorded from which the first and second harmonics of periodic deviations can be derived. These can be corrected independent of their origin. s using a Babinet-Soleil compensator show that this method can reduce significant non-linearities (16 nm top-bottom) to a standard deviation down to about.1 nm. This is confirmed in the evaluation of some standard optical-set-ups. With this method, also optical set-ups can be analysed to predict the non-linearities when a non-compensated standard interferometer is used. Keywords: Laser interferometry, heterodyne, sub-nm, non-linearity, correction 1. INTRODUCTION Laser interferometer systems are known for their high resolution, and especially for their high range/resolution ratio. In dimensional metrology laboratories, laser interferomters are popular workhorses for the calibration of displacements. Their uncertainty is limited to the refractive index of air for long range. For short range calibration the limitations are in the photonic noise and residual nonlinearities repeating each fraction of a wavelength. The nonlinearities can be present in the phase measurement system of the interferometer itself, 1 but they can also result from polarization states of the laser and the optics used: beam splitters, retardation plates used with plane mirrors and corner cubes can influence the polarization state in interferometers using polarizing optics, or influence the contrast in interferometers with non-polarizing optics. Compensation of nonlinearities becomes important within the nanometer region. For homodyne laser interferometers the compensation method of Heydemann is used widely. 3 A quadrature phase measurement method in combination with a lock-in amplifier was used previously to compensate a heterodyne interferometer with a split frequency of 16 khz. 4 We are using a combination of both methods for a commercial heterodyne laser interferometer with a split frequency of.7 MHz and compare the results with calculations using a theoretical model.. PRINCIPLE OF HETERODYNE LASER INTERFEROMETRY The laser shown in figure 1 consists of a laser source which emits two orthogonal polarized beams with a different frequency (f 1 and f ). These two beams can be represented as: Further author information: (Send correspondence to H.Haitjema) S.J.A.G.Cosijns: E-mail: s.j.a.g.cosijns@tue.nl, Phone: +31 ()4 47 6 H.Haitjema: E-mail: h.haitjema@tue.nl E 1 = E sin(πf 1 t + φ 1 ) (1) E = E sin(πf t + φ ). ()

Reference signal f R = f 1 f Reference arm Polarisers Laser 5 5 f arm f 1 Triple mirror f 1 + f 1 signal f M = f 1 f + f 1 Polarizing beam splitter L Figure 1. Schematical representation of the principle of a heterodyne laser interferometer. A non-polarizing beam splitter divides the beam in two parts: one reference part and one measurement part. The reference part passes a combining polarizer and a detector resulting in an interference signal of which only the alternating current is measured using a band-pass filter, eliminating both the DC component and optical frequencies. This signal can be represented by: I ref = 1 E [cos (π (f 1 f ) t +(φ 1 φ ))]. (3) The measurement part is divided by a polarizing beam splitter in a reference arm and in a measurement arm. The signal in the measurement arm receives a Doppler shift as a result of the moving corner cube, resulting in a phase shift of the measurement signal compared to the reference signal 5 : φ 1 =π f 1 t = 4πn l λ 1. (4) with λ 1 representing the wavelength of the light reflected by the mirror and n the refractive index of air in the measurement path. With a polarizer the signal from the reference arm and the signal from the measurement arm are combined and result in an interference pattern. From this interference pattern the measurement signal is constructed containing the phase information and therefore containing the displacement of the corner cube. In an ideal laser interferometer E 1 enters the measurement arm and E enters the reference arm and the detected signal can be represented by: I meas = 1 E [cos (π (f 1 f ) t +(φ 1 φ )+ φ 1 )]. (5) Where f 1 and f are the frequency of measurement and reference arm, φ 1 and φ are the initial phase of both frequencies and φ 1 is the phase difference as a result of the target movement. The measurement with a laser interferometer is based on measuring the phase difference between the measurement signal (5) and the reference signal (3). In a common laser interferometer the separation between both arms and therefore both polarization states is not ideal, part of the measurement frequency enters the reference arm and part of the reference frequency enters the measurement arm and gets a Doppler shift also. This results in an extra phase term in the measurement signal which now can be represented as: I measnl = 1 E [cos(π (f 1 f ) t +(φ 1 φ )+ φ 1 + φ nonlin )]. (6) The calculation in the interferometer however uses only one frequency resulting in non-linearities in the measurement results. The extra phase term therefore represents the non-linearity of the system.

3. PRINCIPLE OF QUADRATURE DETECTION WITH A LOCK-IN AMPLIFIER A lock-in amplifier enables the phase detection between two signals with the same carrier frequency, due to a method called Phase Sensitive Detection. This method enables the filtering of a certain frequency. Noise signals with a different frequency are filtered out. For this method a reference frequency and phase is needed. In the heterodyne laser interferometer signals are generated: I ref = A sin(π(f 1 f )t + φ ref ) (7) I meas = B sin(π(f 1 f )t + φ meas + φ nonlin ). (8) Where I ref is the reference signal and I meas stands for the measurement signal. These signals can well be analysed using a lock-in amplifier. With use of a mixer these signals are multiplied and passed through a band pass filter: S = 1 AB cos(φ ref φ meas φ nonlin ). (9) To enable a phase measurement a second reference signal, retarded π measurement signal resulting in a sine function: in phase, is multiplied with the S 9 = 1 AB sin(φ ref φ meas φ nonlin ). (1) From simulations it is known that many harmonics of nonlinearity exist, however the first and second order are dominant. 6 Therefore the above mentioned signals can be rewritten as follows: S = x + R cos(φ) (11) S 9 = y + R( 1 )sin(φ α). (1) r Where the offset x and y gives a first order non-linearity and the ellipticity characterized by r and α gives the second order non-linearity, R represents the amplitude of the signals. 4. MEASUREMENT RESULTS With use of a Stanford Research SRS844 lock-in amplifier the phase quadrature measurement is carried out. Both output signals are captured with a PC. After an initiating measurement the ellipse parameters are calculated using a least square fit on the parameters of (11) and (1) using the method by Heydemann. These parameters are used to compensate the interferometer system. The schematical setup is shown in figure. Note that Picotto 7 and Sacconi 8 used a similar setup, just he uses a phase detector instead of a lock-in amplifier. In order to verify the compensation method three kinds of measurements were done. First a measurement of the nonlinearity resulting from a common path setup with use of a Babinet Soleil Compensator replacing the interferometer optics is carried out. The second is a measurement of the entire laser interferometer including the interferometer optics, both linear optics as well as high stable flat mirror optics. The final measurement was a comparison of flat mirror laser interferometer and an inductive probe.

S S 9 Output Lock-in amplifier.7 MHz Reference signal Input signal Heterodyne laser Detector Non-polarizing splitter Displacement Figure. Schematical representation of the setup. 4.1. Common path verification The Babinet Soleil Compensator provides a common path for both polarization components, and therefore cancels out any influence of the refractive index of air. It also provides a convenient way to apply an optical path difference of parts of a wavelength. This optical path difference is produced due to a difference in refractive index for two orthogonal polarization states. The phase difference introduced between both polarization axes can be represented as 9 : φ =π n n λ d. (13) Where λ and λ are the vacuum wavelengths of both polarization states, n and n are the refractive indices for both polarization arms and d represents the thickness of the Babinet Soleil Compensator. By varying the thickness of the compensator (through movement of the wedges compared to each other, see figure 3) the optical path difference between both polarization states can be varied. A rotation of the optics (left in figure 3) around the laser beam axis can be simulated by rotating the Babinet Soleil compensator about the laser beam axis. λ Laser + receiver Laser Pitch Receiver Yaw Roll Figure 3. Schematical representation of the simulation of the interferometer optics with a Babinet Soleil Compensator. s were caried out as following: in the experimental setup the Babinet Soleil Compensator is shifted in small steps over a bit more than two wavelength path-difference. Both signals S and S 9 are recorded and a least-squares fit to equations 11 and 1 is made. From these data the corrected phase (φ) is derived and compared to the uncorected phase φ =arctan 1 ( S9 S ) as it is measured in commercially available systems. A measurement was taken with a rotation of 1 of the Babinet Soleil Compensator. Figure 4 gives the corrected and uncorrected results where on the right the S 9 signal is plotted against the S signal and on the left the

phase deviation is re-calculated to a length deviation when linear optics would be used. The deviations from the fitted curve indicate the remaining deviations of the system after correction. Assuming that noise in the recorded intensities can be phase noise as well as amplitude noise, de deviations in amplitude can be considered as a possible phase error 1 : φ = R (14) R Note that this is a worst-case estimation: if the laser intensity fluctuates this will also give an amplitude variation but not a phase variation. The non-linearity as it is measured in figure 4 without correction can be characterised by a standard deviation of 4.4 nm; after correction this is reduced to a standard deviation of.8 nm. From model calculations 6 and measurements of the laser polarization state 11 this type of deviation is known; it is also known that the deviation turns into a second-order deviation once the receiver is given the same rotation as the optics (in this case the Babinet Soleil Compensator). The results of this experiment are given in figure 5. Comparing the right hand figures 4 and 5 one can observe that in figure 4 the measured circle is shifted downward relative to the undisterbed corrected circle and becomes somewhat elliptical. In figure 5 the centre point is shifted back and only the ellipticity remains. This is reflected in the fitted parameters: r and α are similar in both figures, but in figure 5 the bias in y-direction is almost vanished. This is reflected in a smaller linearity deviation for the uncorrected case at the left-hand of figure 5. In this case the standard deviation of the measurement reduces from 1.17 nm for the uncompensated measurement to below.1nm for the corrected measurement. Non-linearity in nm 8 6 4-4 -6 Fit -8 Compensated -1 1 3 4 5 6 (x,y )=(.,-.435), r=1.678, R=.3686, α =.8331.3..1 S 9 -.1 -. -.3 (a) results -.4 -...4 S Calculated non-linearity in nm 8 6 4-4 -6-8 1 3 4 5 6 7 (b) Simulated results Figure 4. Results of a measurement with a rotated Babinet Soleil Compensator 1. Non-linearity in nm 1.5 1.5 -.5-1 -1.5 Fit Compensated -1 1 3 4 5 6 (x,y )=(5.5147e-5,-3.139e-5), r=1.68, R=.39556, α =.961.3..1 S 9 -.1 -. -.3 (a) results -.4 -...4 S Calculated non-linearity in nm 1.5 1.5 -.5-1 -1.5 1 3 4 5 6 7 (b) Simulated results Figure 5. Results of a measurement with a rotated Babinet Soleil Compensator and a rotated receiver polarizer (both 1 ).

4.. Verification of basic interferometer set-ups To verify the effect of the optics as predicted in 1 also two setups were built with interferometer optics. One with linear optics and a second one with flat mirror optics. To generate a stable but slowly varying displacement an aluminium distance tube was placed between the polarizing beamsplitter and the measurement optics. The displacement was generated by warming up the tube by hand and then letting it cool down slowly to room temperature. The result for the linear optics is given in figure 6. This figure shows that already without compensation the non-linearity is within 1 nm. The compensation reduces the standard deviation of the nonlinearity to.1 nm. Internal receiver receiver Retro reflector Distance tube Retro reflector Polarizing beam splitter (a) Linear optics Non-linearity in nm.6.5 Fit Compensated.4.3..1 -.1 -. -.3 -.4-6-5-4-3-1 1 (x,y )=(8.386e-5,5.36e-5), r=1.98, R=.198, α =.44. S 9.1 -.1 (b) results -. -. -.1.1. S Figure 6. Results of a measurement with linear optics. The flat mirror optics gives a more peculiar behaviour, as depicted in figure 7. The same kind of behaviour 13 was noticed by Číp. By modelling the optics and laser head polarization errors it was tried to explain the deviation. This can be done if the assumption is made that the optics were misaligned by 4.9. The right hand sight of figure 7 gives the result of the model calculation. The model was based on a Jones matix calculation. The polarization errors of the optics were measured by ellipsometry 1 and the polarization error of the laserhead was measured as explained in. 11 The ellipticity of the laserhead was 1 : 148 for f 1 and 1 : 168 for f in E-field one left hand elliptical and one right hand elliptical. The non-orthogonality between both laser polarizations was.1. The compensation reduces the standard deviation of the non-linearity from.35 nm down to. nm. 4.3. Comparison probe and flat mirror optics In the laboratory of Precision Engineering at Eindhoven University of Technology a calibration system for probes of Coordinate Measuring Machines was developed based on a flat mirror interferometer (see figure 8). In earlier probe calibrations it appeared to show non-linearities. 14 Therefore this system was tested also. The results are given in figure 9. On the left hand side both uncompensated and compensated data are shown of the setup as shown in figure 8. Due to the probe properties the improvement remains limited. On the right hand of figure 9 the results are shown when the ellipticity of the laser head is increased by means of a rotated λ 4 -plate. In this figure the reduction of the second order non-linearity is clearly visible. However a first order non-linearity with the same order of magnitude as for the left-hand of the figure remains. For the future we plan to test the setup on a traceable calibration setup in which no non-linearities are inherent. 15 5. CONCLUSIONS Phase measurement by lock-in technique was succesfully applied to a commercially available heterodyne laser interferometer system operating at a frequency difference of.7 MHz. Applying the Heydemann correction to

Internal receiver Flat mirror Distance tube λ 4 -plate Flat mirror Polarizing receiver beam splitter Retro reflector (a) Flat mirror optics Non-linearity in nm.5.4 (b1) Calculated non-linearity in nm.3..1 -.1 -. Fit -.3 Compensated -6-5 -4-3 -1 1.5.4 (b) results (b).3..1 -.1 -. -.3-6 -5-4-3-1 1 Figure 7. Results of a measurement with flat mirror optics (b1) and results of the simuated non-linearity from these optics (b). Probe Reference mirror mirror Reference mirror Bending mirror Polarizing λ beam splitter 4 -plate Bending mirror Retro reflector (a) Optical setup (b) Practical setup Figure 8. setup. the measured signals reduces the periodic errors from several nanometers to less than.1 nm. The shape of periodic errors found was consistent with model calculations for the used setups. This method can also be used for measuring periodic errors of existing non-compensating setups. Acknowledgement This work was financially supported by the Dutch ministry of economic affairs in the framework of the program IOP Precisie Technologie. REFERENCES 1. P. H. N.M. Oldham, J.A. Kramar and E. Teague, Electronic limitations in phase meter for heterodyne interferometry, Precision Engineering 15, pp. 173 179, July 1993.. P. L. Heydemann, Determination and correction of quadrature fringe measurement errors in interferometers, Applied optics, pp. 338 3384, October 1981. 3. M.J.Downs and W. Rowley, A proposed design for a polarisation insensitive optical interferometer system with sub-nanometric capability, Precision Engineering 15, pp. 81 86, October 1993.

1 Compensated laser Laser data 15 Compensated laser Laser data 8 6 1 Difference in nm 4-4 Difference in nm 5-6 -5-8 -1-7 -6-5 -4-3 -1-1 -7-6 -5-4 -3-1 Figure 9. Difference between laser and probe. 4. T. C. TaeBong Eom et al., A simple method for the compensation of the nonlinearity in the heterodyne interferometer, Science and Technology 13, pp. 5, January. 5. Jens Flügge, Vergleichende untersuchungen zur messtechnischen leistungsfhigkeit von laserinterferometern und inkrementellen masstabmesssystemen. PhD thesis, Physikalisch Technische Bundesanstalt, February 1996. 6. S. Cosijns, H. Haitjema, and P. Schellekens, Modelling and verifying non-linearities in heterodyne displacement interferometry, Precision Engineering 6, pp. 448 455,. 7. M. G.B.Picotto, A sample scanning system with nanometric accuracy for quantitative spm measurements, Ultramicroscopy 86, pp. 47 54, 1. 8. G. P. A. Sacconi and W. Pasin, Interferometric calibration of piezo-capacitive displacement actuators, Proceedings of 4th IMEKO symposium on Laser metrology, 1996. 9. R. Azzam and N. Bashara, Ellipsometry and polarized light, Elsevier Science, Amsterdam, 1999. 1. J. L. Chien-ming Wu and R. D. Deslattes, Heterodyne interferometer with subatomic periodic nonlinearity, Applied optics 38, pp. 489 494, July 1999. 11. D.Lorier, B.Knarren, S.Cosijns, H.Haitjema, and P.Schellekens, Polarisation state analysis of heterodyne laser beams, CIRP Annals 5, August 3. 1. S. Cosijns, H. Haitjema, and P. Schellekens, The influence of polarization states on non-linearities in laser interferometry, Proceedings of the 3 rd Euspen International Conference, Eindhoven, The Netherlands, pp. 593 596, May. 13. O. Cip and F. Petru, A scale-linearization method for precise laser interferometry, Meas. Sci. Technol. 11, pp. 133 141,. 14. W. Pril, Development of high precision mechanical probes for coordinate measuring machines. PhDthesis, TUE, December. 15. H. Haitjema, P. Schellekens, and S. Wetzels, Calibration of displacement sensors up to 3 µm with nanometre accuracy and direct traceability to a primary standard of length, Metrologia 37(1), pp. 5 33,.