RADIOMETRIC AND PHOTOMETRIC MEASUREMENTS AT THE LNE-INM/CNAM

Similar documents
Current and Future Realizations NRC Photometric and Spectroradiometric Calibration Chains

NIST MEASUREMENT SERVICES:

Photometry for Traffic Engineers...

Photometry for Traffic Engineers...

Part 1: New spectral stuff going on at NIST. Part 2: TSI Traceability of TRF to NIST

Measurement Services Optics Laboratory

Lighting Terminologies Introduction

NFMS THEORY LIGHT AND COLOR MEASUREMENTS AND THE CCD-BASED GONIOPHOTOMETER. Presented by: January, 2015 S E E T H E D I F F E R E N C E

New automated laser facility for detector calibrations

Intercomparison of radiation temperature measurements over the temperature range from 1600 K to 3300 K

Technical Notes. Integrating Sphere Measurement Part II: Calibration. Introduction. Calibration

High Illuminance Calibration Facility and Procedures

Application Note (A16)

Development of 2 Total Spectral Radiant Flux Standards at NIST

Report No.: HZ j. Total Luminous Flux (Lumens) Stabilization Time (Light & Power) Table 1: Executive Data Summary CRI

Report No.: HZ f. Total Luminous Flux (Lumens) Stabilization Time (Light & Power) Table 1: Executive Data Summary CRI

Radiometric and Photometric Measurements with TAOS PhotoSensors

Developement of the method for realization of spectral irradiance scale featuring system of

\Ç à{x ÇtÅx Éy ALLAH à{x `xüv yâä

We bring quality to light. MAS 40 Mini-Array Spectrometer

Report No.: HZ h/R1. Total Luminous Flux (Lumens) Stabilization Time (Light & Power) Table 1: Executive Data Summary CRI

SPECTRORADIOMETRY METHODS: A GUIDE TO PHOTOMETRY AND VISIBLE SPECTRORADIOMETRY

Company synopsis. Regulations and Standards

Report No.: HZ h. Total Luminous Flux (Lumens) Stabilization Time (Light & Power) Table 1: Executive Data Summary CRI

Quality Manual of Luminous Intensity Laboratory

IESNA LM-79: Measurement and Test Report for Halco Lighting Technologies ATL 2940-A PACIFIC DRIVE. Dec 08, 2014

07-Lighting Concepts. EE570 Energy Utilization & Conservation Professor Henry Louie

Radiometry vs. Photometry. Radiometric and photometric units

Of straying photons, shiny apertures and inconstant solar constants Advances in TSI radiometery

Report No.: HZ b. Total Luminous Flux (Lumens) Stabilization Time (Light & Power) Table 1: Executive Data Summary CRI

Instruction Manual of Luminance and Spectral Radiance Calibrations

LED14.5WPAR30S/FL/827-DIM

Keysight Technologies Optical Power Meter Head Special Calibrations. Brochure

Originally presented as a Keynote to the CIE Conference, Paris, April Reproduced with the kind permission of Jean Bastie.

LM Test Report for OSRAM SYLVANIA INC. 100 Endicott Street Danvers, MA 01923

(Lumens) Stabilization Time CRI (K) (Light & Power) Table 1 Executive Data Summary

IESNA LM-79: Measurement and Test Report for Jiawei Technology (USA) Limited Lincoln Ave.Hayward, CA USA.

IESNA LM-79: Measurement and Test Report for EiKO Global, LLC. Sep 09, 2013 LED8WMR16/38/830-DIM. David Zhang BTR

Report No.: HZ b. Total Luminous Flux (Lumens) Stabilization Time (Light & Power) Table 1: Executive Data Summary CRI

(Lumens) Stabilization Time CRI (K) (Light & Power) Table 1 Executive Data Summary

Photometric Calibrations

The Art of Light Measurement. Avantes BV Apeldoorn, The Netherlands

Report No.: HZ a/R2. Total Luminous Flux (Lumens) Stabilization Time (Light & Power) Table 1: Executive Data Summary CRI

Report No.: HZ c/R1. Total Luminous Flux (Lumens) Stabilization Time (Light & Power) Table 1: Executive Data Summary CRI

Report No.: HZ a/R2. Total Luminous Flux (Lumens) Stabilization Time (Light & Power) Table 1: Executive Data Summary CRI

The Physikalisch-Technische Bundesanstalt,

Total solar irradiance measurements with PREMOS/PICARD

Detector-Based Traceability Chain for Spectral Irradiance using Tunable Laser-Based Facility at PTB

TSBB09 Image Sensors 2018-HT2. Image Formation Part 1

LM Test Report. for Elec-Tech International Co.,Ltd. LED PAR30 Spot light Model: XX(XX:41-50)

UV-VIS-IR Spectral Responsivity Measurement System for Solar Cells

TEST REPORT. Job No SHA Date: Apr 12,2017 REPORT NO SHA-006. TEST OF ONE LED Luminaire MODEL NO.

INDEPENDENT TEST LABORATORY REPORT No

CCT/10-13 MeP-K direct methods. Section 3.2: Absolute (spectral-band) radiometry (radiation thermometry)

LED Tester BTS256-LED

IESNA LM-79: Measurement and Test Report for Green Creative Ltd. Room , New Victory House, Wing Lok Street, Central, HONG KONG

Capabilities of NIST SIRCUS for Calibrations of SSI Vis-IR Instruments

LM Test Report. for Elec-Tech International Co.,Ltd. LED CEILING LIGHT Model: XX(XX: 01-10)

LM Test Report. for Maxlite SK America Inc. 2*4 Retrofit Kits Model: RKT4514U5550DV

LM Test Report. for Elec-Tech International Co.,Ltd. LED CEILING LIGHT Model: XX(XX: 61-70)

CU-LASP Test Facilities! and Instrument Calibration Capabilities"

LM Test Report. for Elec-Tech International Co.,Ltd. LED CEILING LIGHT Model: XX(XX: 01-10)

INDEPENDENT TEST LABORATORY REPORT No

INDEPENDENT TEST LABORATORY REPORT No

LM Test Report. for Elec-Tech International Co., Ltd. BR40 LAMP Model: XX(XX:01-10), WBR40L27S2A

LM Test Report. for. Philips (China) Investment Co., Ltd. InstantFit LEDtube Model:

BGS400 Lamp & Luminaire Goniophotometer (C-γ, B-ß Plane)

NIST Agency Report May 2012 OUTLINE. The case for traceability NMI capabilities A view to the future the HIP Current/recent NIST activities

NEW MEASUREMENT STANDARDS AND METHODS FOR PHOTOMETRY AND RADIOMETRY

Diogen Lighting,Inc 14 Inverness Drive East,Ste.E-128 Englewood, CO T8L W BEST Test Service Shenzhen Co., Ltd.

Optical and Quantum Electronics (2005) 37: Ó Springer 2005 DOI /s

LM Test Report. for EiKO Global, LLC. 25W Floodlight Model: FLM-2C-N-U

Model S471-LED Optometer

Observational Astronomy

Luminaire photometry has been performed traditionally using a standard

CIE 220:2016 Characterization and Calibration Method of UV Radiometers

Report No.: HZ w. Stabilization Time (Light & Power) CRI. Table 1: Executive Data Summary. Figure 1- Overview of the sample

Novel Approach for LED Luminous Intensity Measurement

Electrical Illumination and Design

LM Test Report. for Maxlite SK America Inc. WALLPACK Model: MLSWP30LED50DS

NIST EUVL Metrology Programs

Improved Radiometry for LED Arrays

Irradiance Calibration Using a Cryogenic Radiometer and a Broadband Light Source

LINEARPYROMETER LP4. Technical Documentation KE November TN

Lighting Sciences, Inc.

NIRCam optical calibration sources

Measurements of Infrared Sources with the Missile Defense Transfer Radiometer

Transport Canada Standard for LED Signal Modules at Highway/Railway Grade Crossings. TC E-14 (October 10, 2003)

INDEPENDENT TEST LABORATORY REPORT No

INDEPENDENT TEST LABORATORY REPORT No

Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, 77. Table of Contents 1

Advantages of white LED lamps and new detector technology in photometry

LM Test Report. for Maxlite Inc. LED Canopy Model: MLCAN20LED50

Relation between the illuminance responsivity of a photometer and the spectral power distribution of a source

Lighting Sciences

OL Series 426 Low Light Level lntegrating Sphere Calibration Standards

SPECTRAL IRRADIANCE DATA

Photometric Indoor Test Report

LTE. Tester of laser range finders. Integrator Target slider. Transmitter channel. Receiver channel. Target slider Attenuator 2

NVLAP LAB CODE:

Transcription:

RADIOMETRIC AND PHOTOMETRIC MEASUREMENTS AT THE LNE-INM/CNAM Jean Bastie, Laura Patricia Gonzalez Galvan. Institut National de Métrologie Conservatoire National des Arts et Métiers 61 rue du Landy 93210 La Plaine Saint Denis - France Telephone : +33 1 40 27 20 25, Fax : +33 1 58 80 89 00, e-mail : bastie@cnam.fr Abstract : Optical radiation can be measured according two ways. One way is radiometry which deals with the physical aspect of optical radiation and the other way is photometry which is concerned by the action of optical radiation on the eye of an observer. The results of radiometric measurements are expressed in the usual physical SI units and photometric quantities have their own SI units. This paper describes the use of a cryogenic radiometer for realizing radiometric references. It gives also the links existing between radiometric and photometric units and explains how to realize the candela which is the photometric base unit in the SI system of luminous intensity and the lumen which is the photometric unit for flux, starting from the cryogenic radiometer as reference. 1. INTRODUCTION The various quantities defined for radiometric measurements are used for characterizing the power transferred by optical radiation and use physical units. The photometric measurements are used to quantify the visual effect of radiation on the human eye. But, as eye is a very complicated organ and not the same for everybody, it is necessary to define a standard observer which is mainly described by the luminous efficiency function V() [1]. They use special SI units, the photometric units. At present time radiometric and photometric measurements are closely linked through the present definition of the candela. In this paper, the present radiometric reference of the LNE-INM/CNAM is described. Then the use of this radiometric reference for realizing the photometric units is presented.the two major photometric quantities for which the laboratory is providing standards are the luminous intensity and the luminous flux. 2. RADIOMETRIC REFERENCES 2.1 Operating principle Cryogenic radiometers are recognised as by far the most accurate radiometric standards. [2] The principle of this apparatus is very simple and more than 100 years old. It is based on the electrical substitution. In a first step, the optical radiation to be measured is put into an absorbing cavity giving a rise in temperature of this cavity. In a second step, the optical radiation is replaced by an electrical heating in order to get the same rise in temperature. If the radiometer would be perfect, the optical power will be equal to the electrical power. In practice, it is necessary to do several corrections to get the accurate results. Working at very low temperature allows to increase the responsivity of the device and to reduce dramatically the correction factors and their associated uncertainties. The cryogenic radiometer used for the present measurement is the LaseRad system from Cambridge Research Instrumentation. Customized specifications provide an operating power range around 1 mw, and enable measurements to be carried out between 250 nm 2000 nm. [3] Evacuation port Cavity assembly Liquid Helium reservoir liquid Helium port Electrical connectors Apertures : 5 mm 6 mm 7.5 mm Brewster angle window assembly Figure 1- Schematic drawing of the cryogenic radiometer. 1

2.2 Description of the system The design of the cryogenic radiometer is shown in figure 1. Its basic element is the highly absorptive cavity which heats up when it is either irradiated or electrically heated. The cavity is mounted horizontally and is thermally linked to the helium liquid reservoir. Both are placed in a high-vacuum enclosure, which is closed by a quartz window. To minimize reflective losses, this window can be conveniently adjusted to the Brewster angle. Considering the shape and the location of the cavity, the radiometer can only be used with collimated laser beam of diameter up to 2 mm. 2.3 Correction factors To know accurately the radiant power in the laser beams, which are used in transfer detector calibration, several corrections must be applied to the cryogenic radiometer laser power measurement. These corrections are listed in table 1 which gives an example of the value of these different correction factors for the wavelength of 543 nm as well as the uncertainty on their determination. Table 1 Correction factors and uncertainty budget of the cryogenic radiometer Measurement Correction Relative uncertainty (1σ) Cavity absorptance 0.99988 1x10-5 Window transmission 0.99974 3x10-5 Heating non-equivalence 1.00000 1x10-5 Electrical power measurement 1.00000 3x10-5 Global correction 0.99960 5x10-5 2.4 Spectral responsivity detector calibration The calibration of a detector is carried out at some laser wavelengths by direct comparison to the cryogenic radiometer. The experimental set-up used for these measurements is shown in figure 2. The light emitted by the laser is power stabilized by a feedback photodiode and a liquid crystal modulator. The achieved stability is in the range of few parts in 10 5 during the time needed for a comparison, typically 15 minutes. The spatial filter adjust the beam size to the right diameter and removed the stray light. The test detector is compared to the cryogenic radiometer by putting, successively the two detectors into the laser beam using a translation stage. detector to calibrate Shutter aperture cryogenic radiometer spatial filter feedback photodiode liquid crystal modulator Polariser He-Ne laser He-Ne laser Argon laser Figure 2 Experimental set-up for calibrating detectors against the cryogenic radiometer The uncertainty on the results of these calibrations depends on the stability of the detector under test. At present time, the best transfer detectors are Si photodiode trap detectors. They are calibrated with a global standard uncertainty comprised between 1 or 2 parts in 10 4, in the visible range [4]. To extend the spectral responsivity calibration of the detector all over the spectral range, in a second step we use a relative spectral responsivity measurement set-up. The measurement of the relative spectral responsivity is carried out by comparison of the response of the test detector to that of a non selective cavity shape pyroelectric detector, when they are irradiated by the same monochromatic flux. In this calibration, the major cause of uncertainty is coming from the low responsivity of the cavity shape pyroelectric detector giving a signal to noise ratio in the range of 10 3 and a standard uncertainty of about 2 10-3 in the visible range used for detector calibration. 3. REALISATION OF THE CANDELA 3.1 Principle of measurement The present definition of the candela is : The candela is the luminous intensity, in a given direction, of a source that emits a monochromatic radiation of frequency 540.10 12 hertz and that has a radiant intensity in that direction of 1/683 watt per steradian.[5] It can be also written in a numerical form according to the following equation. Iv = Km I e, ( ). V ( ). d This definition gives only the numerical relationship between the luminous quantities and the radiometric quantities. The luminous intensity I v is linked to the spectral radiant intensity distribution I e, (), weighted by the V() function and multiplied by K m. V() is the spectral luminous efficiency of standard observer defined by the CIE in 1924 and K m is the 2

maximum luminous efficiency which fixes the relationship between luminous and radiant quantities. By definition K m is equal to 683 lm/w. The principle used to realize the candela is directly linked to the definition. The source to be calibrated in luminous intensity I v, irradiates through a filter, a radiometer able to measure the irradiance it receives, E e in an absolute way, in Wm -2. The transmittance of the filter τ() matches as closely as possible the V() function. In these conditions the luminous irradiance is given by In this equation the term S e, F = S Ev = ( Km. Ee ). F e, ( ). V ( ). d ( ). τ ( ). d With : S e, () relative spectral distribution of the source to be calibrated, τ() transmission of the real filter, V() spectral luminous efficiency function. is the spectral matching factor which takes into account for the discrepancy between the perfect V() filter and the realized filter. 3.2 Standard photometers For realizing the luminous intensity unit according to the principle described previously, it is necessary to built standard photometers directly calibrated by comparison to the cryogenic radiometer.[6] These standard photometers are made with silicon trap detectors, a set of colored glass filters which adjust the spectral responsivity of the detector to match the V() function, and a calibrated aperture (figure 3). Trap detector Detector 1 Det. 2 Det. 3 Connector Temperature control Oven Filter Thermistor Peltier cooler Aperture Cap Figure 3 Schematic drawing of a standard photometer. The aperture and the V() filter are put in a temperature controlled housing. For characterizing these photometers we have had to realize experimental set-up for measuring the spectral responsivity of detectors (described previously), the spectral transmittance of filters and the area of apertures. 3.3. Characterization of the V() filters Using the relative spectral responsivity of the trap detector selected for realizing the photometers and the V() function, we have determined the theoretical filter to match V(). From glass filter catalogue from Schott, we have selected some glasses which seem appropriate for approximating the ideal filter. Then, we have determined the number of suitable filters and calculated their thickness to approximate the ideal filter. Three filters were calculated, realized and measured on our experimental set-up for filter transmittance measurement.[7] The results of the study of the V() filter is shown in figure 4 which displays the deviation of the realized filters from the ideal filters. In the visible spectral range the relative standard uncertainties are usually less than 10-3. 0,15 0,10 0,05 0 350 400 450 550 600 650 700 750 800-0,05-0,10-0,15 Filtre 1 Filtre 2 Filtre 3 Wavelength [nm] Figure 4 Transmittance of the realized V() filter. 3.4. Aperture measurements To measure the area of each aperture we have used a non contact method developed in our laboratory.[8] The relative uncertainties achievable for aperture from 6 to 10 mm in diameter are in the range from 3.10-4 to 5.10-5. 3.5. Photometer calibration The components previously studied have been put together to realize five photometers. The complete photometers were calibrated directly, in absolute spectral responsivity by comparison to a transfer detector calibrated against the cryogenic radiometer on an experimental set-up using a double 3

monochromator. The luminous responsivity and the spectral matching factor (SMF) were calculated for each photometer using the absolute spectral responsivity. The results of this study are given in the table 3. Table 3 Responsivity of the standard photometers Photometer PH-04-A PH-04-B PH-04-C PH-04-D PH-04-E Responsivity A/lx 2.5115.10-8 1.4848.10-8 2.3335.10-8 1.6304.10-8 9.6814.10-9 SMF 1.0022 1.0063 1.0038 1.0151 1.0369 performances of this apparatus must be as high as possible in order to keep the added uncertainties lower or at least of the same order of magnitude as the uncertainties achieved in realizing the candela. 4.2 The goniophotometer : To realize the luminous flux unit, the lumen, a large size goniophotometer (7 m diameter) has been built (figure 5).[9] 3.6. Luminous intensity measurement The standard photometers are designed to measure illuminance in the plane of its entrance aperture. So to calibrate incandescent standard lamps in luminous intensity we have to measure accurately the distance between the filament of the lamp and the aperture of the photometer. This is done on a photometric bench with a very good quality ruler and precision indexes. 3.7. Uncertainty budget The components of the present uncertainty budget are detailed in table 4. The first component is the reproducibility of measurements (type A). The second one is related to the geometrical parameters. The uncertainties on electrical measurements for the photometer as well as for the lamp are components three and four. The uncertainty on the luminous responsivity which includes mainly the uncertainty on the spectral responsivity of the photometer is the last one and the highest. The present total relative standard uncertainty is 0.22 (k=1). Table 4 Uncertainty budget for the realization of the candela (relative values). Reproducibility 0.01 Solid angle 0.07 Photometer current 0.06 Lamp intensity 0.06 Luminous responsivity 0.19 Global uncertainty (1σ) 0.22 4. LUMINOUS FLUX MEASUREMENT 4.1 Principle of measurement Traditionally a luminous flux standard lamp is a lamp for which the luminous flux is measured in the space all around it. The luminous intensity distribution can be measured with a goniophotometer. The Figure 5 Schematic drawing of the goniophotometer. A standard lamp of luminous flux must be operated in a prescribed burning position, generally vertical, cap up. For this reason the goniophotometer realizes the spatial measurement of the luminous intensity according to the following method : the lamp is rotated around its vertical axis over a full circle (360 ). The photometer is rotated in a vertical plane containing the axis of the lamp. Its rotation is only a half of a circle (180 ). The lamp is put at the center of the circle described by the photometer. These two rotations allow measuring the luminous intensity distribution of the lamp all over the space around it. The main characteristics of the apparatus are : - Distance between source and detector : 3.4 m - Photometer V() corrected, diameter : 60 mm - Speed of motion of the detector : 4 /s - Uncertainty on the angular setting : 0.02 4

Before and after each set of luminous flux measurement, the photometer is calibrated using luminous intensity standard lamps. A set of measurement lasts about one month. The total flux emitted by the lamp is calculated by integrating the luminous intensity distribution over 4π steradians. 4.3 Incandescent standard lamps The standard lamps usually used for maintaining the luminous flux unit, the lumen, are incandescent lamps specially manufactures in order to be stable and reproducible, with a smooth intensity distribution. They are supplied by DC current. For this type of lamps the temporal aspects of the on the fly measurements is relatively easy to take into account for.[10] The achievable uncertainty is 4.10-3. In this uncertainty, 2.2.10-3 are coming from the realization of the luminous intensity unit. 5 CONCLUSION It is a long way between the primary standard, the cryogenic radiometer, and the lamps used for transferring the luminous flux unit to industry. It involves a lot of specific measurements in almost all the fields of activities of the radiometry, spectroradiometry, photometry and spectro-photometry. Due to this large quantity of extra measurements needed to realize luminous flux standard lamps, the traceability is not very easy to determine and need to be carefully check at each step. Another consequence of that is also the deterioration of the uncertainty along the chain. Starting with an uncertainty of approximately 0.01% at the top of the calibration chain we have only 0.4% at the end for the realization of luminous flux standard lamps. Nevertheless, in spite the need of numerous additional measurements for characterizing the photometers and the goniophotometer, the achievable uncertainty is perfectly acceptable for most of the common use. REFERENCES [1] Photometry The CIE system of physical photometry, Standard CIE S 010/E : 2004 [2] J.E. MARTIN, N.P. FOX and P.J KEY., A Cryogenic Radiometer for Absolute Radiometric Measurements, Metrologia, 21, 147-155, 1985. [3] C.C. HOYT and P.V. FOUKAL, Cryogenic Radiometers and their application to Metrology, Metrologia, 28, 163-167, 1991 [4] J-M COUTIN, O TOUAYAR and J. BASTIE, The using conditions of the BNM-INM cryogenic radiometer as the basis for the French optical radiation measurement scales, Proceedings of the 24 th CIE Session1999, Warsaw (Poland), pages 222-224. [5] Conférence Générale des Poids et Mesures (CGPM) : Comptes Rendus de la 16 ème Séance, 1979. [6] L. P. GONZALEZ GALVAN, Matérialisation de la candela (unité d intensité lumineuse dans le système international), à l aide de photomètres dont la chaîne de raccordement au radiomètre cryogénique primaire est parfaitement établie, Ph.D. Thesis 2005 [7] P. LECOLLINET et J. BASTIE, Mesure des facteurs de transmission de filtres optiques au BNM- INM, Actes des conférences Métrologie 1997, Besançon, France, pages 343-347. [8] A. RAZET, Etalonnage de diamètres de diaphragmes pour des mesures radiométriques, Bulletin du BNM n 120 volume 2001-2, pages 27-33. [9] J. BASTIE, Luminous flux measurement at the BNM-INM photometry laboratory, Proceedings of Light & lighting 2000, Bucharest (Romania). [10] J. BASTIE, B. ANDASSE and R. FOUCART, Luminous flux measurements with a goniophotometer; study of time effects on data collection. Proceedings of the 22 nd CIE Session 1991, Melbourne (Australia), pages 44-47. 5