An 8-Channel Parallel Multispectral TCSPC FLIM System

Similar documents
Non-Descanned FLIM Detection in Multiphoton Microscopes

Megapixel FLIM with bh TCSPC Modules

TCSPC at Wavelengths from 900 nm to 1700 nm

Instrument response function. Left linear scale, right logarithmic scale. FWHM is 120 ps.

DCS-120. Confocal Scanning FLIM Systems. Based on bh s Multidimensional Megapixel FLIM Technology

DCS-120. Confocal Scanning FLIM Systems. Based on bh s Multidimensional Megapixel FLIM Technology

The DCS-120 Confocal Scanning FLIM System

PZ-FLIM-110. Piezo Scanning FLIM System. Based on bh s Megapixel FLIM Technology. Complete FLIM Microscopes FLIM Upgrades for Existing Microscopes

Multiphoton FLIM with the Leica HyD RLD Detectors

IR Antibunching Measurements with id201 InGaAs Gated SPAD Detectors

Wide-Field TCSPC FLIM with bh SPC-150 N TCSPC System and Photek FGN Detector

Multi-wavelength TCSPC lifetime imaging Wolfgang Becker a, Axel Bergmann a, Christoph Biskup b, Thomas Zimmer b, Nikolaj Klöcker c, Klaus Benndorf b

Becker & Hickl GmbH DCS-120. Confocal Scanning FLIM Systems. An Overview

High Performance Photon Counting. User Manual PML-16-C. 16 Channel Detector Head for Time-Correlated Single Photon Counting. Becker & Hickl GmbH

BDS-SM Family Picosecond Diode Lasers

TCSPC for FLIM and FRET in Microscopy

Practical work no. 3: Confocal Live Cell Microscopy

Time-Correlated Single Photon Counting

Becker & Hickl GmbH. Technology Leader in Photon Counting

BDS-MM Family Picosecond Diode Lasers

BDS-SM Family Picosecond Diode Lasers

Ultraviolet and Blue Picosecond Diode Laser Modules

Multifluorescence The Crosstalk Problem and Its Solution

3D light microscopy techniques

NUV and Blue ps Diode Lasers

Time-Correlated Single Photon Counting Systems

Boston Electronics Corporation 91 Boylston Street, Brookline MA USA (800) or (617) fax (617)

Time Correlated Single Photon Counting Systems

Picosecond Light Sources

TCSPC measurements with the InGaAs/InP Single- photon counter

Simultaneous Phosphorescence and Fluorescence Lifetime Imaging by Multi-Dimensional TCSPC and Multi-Pulse Excitation

PML Channel Detector Head for Time-Correlated Single Photon Counting

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005

Confocal Microscopy. Kristin Jensen

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

PCS-150 / PCI-200 High Speed Boxcar Modules

WHITE PAPER FAST PROTEIN INTERACTION BINDING CURVES WITH INO S F-HS CONFOCAL MICROSCOPE

Confocal, hyperspectral, spinning disk

3D light microscopy techniques

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope

Solea. Supercontinuum Laser. Applications

Time-Correlated Single Photon Counting

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Detectors for High-Speed Photon Counting

FLIM Protocol. 1. Install IRF sample on the microscope using the stage insert. IRF sample options include urea crystals or BBO crystal.

SHM-180 Eight Channel Sample & Hold Module

DeltaMyc. Fluorescence Lifetime Mapping Microscope. Affordable Fluorescence Lifetime Imaging Microscopy (FLIM)

Confocal Microscopy. (Increasing contrast and resolu6on using op6cal sec6oning) Lecture 7. November 2017

Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement

Boston Electronics Corporation 91 Boylston Street, Brookline MA USA (800) or (617) fax (617)

Bio 407. Applied microscopy. Introduction into light microscopy. José María Mateos. Center for Microscopy and Image Analysis

User Handbook. DPC Channel Photon Correlator

Shreyash Tandon M.S. III Year

ADVANCED METHODS FOR CONFOCAL MICROSCOPY II. Jean-Yves Chatton Sept. 2006

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST

Maria Smedh, Centre for Cellular Imaging. Maria Smedh, Centre for Cellular Imaging

ScanArray Overview. Principle of Operation. Instrument Components

Supplemental Information

Fluorescence lifetime imaging for the two-photon microscope: time-domain and frequency-domain methods

Simple setup for nano-second time-resolved spectroscopic measurements by a digital storage oscilloscope

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

MULTIPHOTON MICROSCOPY. Matyas Molnar Dirk Pacholsky

Dynamic Confocal Imaging of Living Brain. Advantages and risks of multiphoton microscopy in physiology

Resolution. Diffraction from apertures limits resolution. Rayleigh criterion θ Rayleigh = 1.22 λ/d 1 peak at 2 nd minimum. θ f D

FLIM on a wide field fluorescence microscope

INTRODUCTION TO MICROSCOPY. Urs Ziegler THE PROBLEM

DETECTORS Important characteristics: 1) Wavelength response 2) Quantum response how light is detected 3) Sensitivity 4) Frequency of response

BIOIMAGING AND OPTICS PLATFORM EPFL SV PTBIOP LASER SCANNING CONFOCAL MICROSCOPY PRACTICAL CONSIDERATIONS

Operating Instructions for Zeiss LSM 510

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

The Zeiss AiryScan System, Confocal Four.

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy

Supplementary Figure S1: Schematic view of the confocal laser scanning STED microscope used for STED-RICS. For a detailed description of our

LSM 710 Confocal Microscope Standard Operation Protocol

Point Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy

Technology Note ZEISS LSM 880 with Airyscan

Microscopic Structures

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background

Imaging Retreat - UMASS Customized real-time confocal and 2-photon imaging

Components of confocal and two-photon microscopes

Reflecting optical system to increase signal intensity. in confocal microscopy

Zeiss 780 Training Notes

LSM 510 META in Chang Gung University

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy,

LABORATÓRIUMI GYAKORLAT SILLABUSZ SYLLABUS OF A PRACTICAL DEMONSTRATION. financed by the program

1 Co Localization and Working flow with the lsm700

Fiber Laser Chirped Pulse Amplifier

Bi Imaging. Multicolor Imaging: The Important Question of Co-Localization. Anna Smallcombe Bio-Rad Laboratories, Hemel Hempstead, UK

Light Microscopy for Biomedical Research

A New Single-Photon Avalanche Diode in 90nm Standard CMOS Technology

Fundamentals of Light Microscopy II: Fluorescence, Deconvolution, Confocal, Multiphoton, Spectral microscopy. Integrated Microscopy Course

Cavity QED with quantum dots in semiconductor microcavities

Confocal Microscopy and Related Techniques

Alba v5 Laser Scanning Microscope

長庚大學共軛焦顯微鏡課程 長庚大學共軛焦顯微鏡課程. Spot light 長庚大學

Application Note. The New 2D Superresolution Mode for ZEISS Airyscan 120 nm Lateral Resolution without Acquiring a Z-stack

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide

Real-time Fluorescence Lifetime Imaging Microscopy Implementation by Analog Mean-Delay Method through Parallel Data Processing

LSM 780 Confocal Microscope Standard Operation Protocol

Transcription:

An 8-Channel Parallel Multispectral TCSPC FLIM System Abstract. We describe a TCSPC FLIM system that uses 8 parallel TCSPC channels to record FLIM data at a peak count rate on the order of 50 10 6 s -1. By using a polychromator for spectral dispersion and a multi-channel PMT for detection we obtain multi-spectral FLIM data at acquisition times on the order of one second. We demonstrate the system for recording transient lifetime effects in the chloroplasts in live plants. Count Rates in TCSPC FLIM Experiments TCSPC FLIM [3, 6] delivers single- and multi-wavelength fluorescence lifetime images at an accuracy essentially limited by the photon statistics [1, 11], and a time resolution limited by the transit-time-spread of the detector [4, 5, 12]. This is a clear advantage in the typical FLIM applications, such as FRET measurements, investigation of membrane proteins, or autofluorescence imaging. All these applications have in common that the fluorophore concentration is low. Moreover, the excitation power has to be kept low to avoid cell damage and lifetime changes by photobleaching. The count rates available from the samples are therefore on the order of only 10 4 s -1 to a few 10 5 s -1. This is one to two orders of magnitude lower than the counting capability of a single TCSPC FLIM device [5]. Consequently, the acquisition time required to achieve a given lifetime accuracy is exclusively given by the photostability of the sample, not by the counting capability of the TCSPC module. Technical efforts to increase the counting capability of the TCSPC FLIM electronics therefore do not result in shorter acquisition times in these applications. There are, however, certain classes of FLIM experiments that can be run at very high count rate. One of these is imaging of chlorophyll in plant tissue. Chlorophyll is highly photostable and emits strong fluorescence around 700 nm. The fluorescence lifetime of chlorophyll in live plants changes with the time of exposure [9]. Recording these changes requires fast FLIM [10]. Another potential application is, surprisingly, autofluorescence. Autofluorescence signals are usually weak, but exceptions do exist. Fig. 1 shows autofluorescence FLIM images of live yeast cells. The bright cells (probably apoptotic ones) are about 50 times brighter than the dim ones (visible only in Fig. 1, right). Fig. 1: FLIM images of yeast cells. Left: Intensity scale 0 to 3000 counts per pixel. Right: Intensity scale 0 to 300 counts per pixel. The bright cells are about 50 times brighter than the dim ones. Becker & Hickl DCS-120 confocal scanning FLIM system, excitation at 407 nm. eight-channel-flim-apn01.doc Oct. 2009 1

Recording lifetimes for both types of cells in the same image requires the count rate of the bright cells to be kept below about 4 MHz. The count rate in the dim cells is then less than 100 khz. The high contrast thus results in an unnecessarily long acquisition time. The Way to Higher Counting Capability The count rate of TCSPC FLIM is limited by counting loss due to the dead time of the TCSPC electronics, by pile-up effects, and saturation in the detector [4, 5, 10]. Counting loss could, in principle, be reduced by decreasing the dead time of the electronics. Pile-up-effects could be reduced by recording more than one photon per signal period. Unfortunately, both approaches have detrimental effects on the accuracy of the FLIM data. Reducing the dead time increases the background caused by the afterpulsing of the detector [4]. Recording several photons per laser period leads to an overlap of subsequent single-photon pulses and thus causes count-rate-dependent distortions of the decay curves [10]. In any case, detector nonlinearity and detector overload preclude the use of a single detector at count rates much higher than 10 7 s -1. A better way to increase the counting capability of TCSPC FLIM is therefore to use several parallel TCSPC channels. Standard FLIM systems for the bh DCS-120 confocal scanner [7] or the Zeiss LSM 710 NLO [8] system therefore have two parallel TCSPC channels. Four-channel FLIM systems are available as well [5] and have been shown to record FLIM at an average count rate of 13 MHz [2]. 8-Channel Parallel FLIM To further increase the counting capability of TCSPC FLIM we built an eight-channel parallel FLIM system. The architecture of the system is described below. We use the Becker & Hickl DC-120 confocal scanning FLIM system [7], see Fig. 2, left. An LOT MS 125 polychromator is attached to one confocal output of the DCS-120 scanner. The polychromator is used without an input slit; the light focused into the slit plane by a lens. from confocal pinhole Slit plane Mirror Mirror Lens LOT MS 125 Grating R5900-L16 Detector Fig. 2: Eight-channel FLIM system. Left: DCS-120 scanner. Middle: Principle of spectral dispersion. Right: FLIM electronics. The spectrum at the polychromator output is detected by a Hamamatsu R5900-L16-01 multichannel PMT. Only eight of the 16 channels of the R5900 are used; the remaining ones are terminated via 50 Ω into ground. Alternatively, connecting every two PMT channels in parallel is possible, but not recommended. The PMT channels have slightly different signal transit times. Connecting channels in parallel thus impairs the time resolution. Moreover, combining channels would also combine the dark counts and thus result in an unwanted increase of the thermal background. The electronics of the 8-channel FLIM system is shown in Fig. 2, right. We use two Simple- Tau 154 bus extension boxes holding four bh SPC-150 TCSPC modules each. The boxes are connected to a standard Pentium PC via bus extension cables. The PC also contains a DCC-100 detector controller card [5] and the scan controller card of the DCS-120 system [7]. 2 eight-channel-flim-apn01.doc Oct. 2009

Fig. 3 shows autofluorescence FLIM of a Drosophila eye. The excitation wavelength was 407 nm. The images were scanned at a resolution of 256 x 256 pixels and a rate of 500 ms per frame. The intensities were normalised on the brightest pixels. Fig. 3: Multi-wavelength FLIM of a Drosophila eye. Autofluorescence, excitation at 407 nm. Double-exponential decay model. Left: Amplitude weighted mean lifetime. Right: Relative amplitude of fast decay component. The total count rate (summed over all TCSPC channels and averaged over the image) obtained from the Drosophila eye was about 4 10 6 s -1. That means the rate in the brightest pixels was more than 10 10 6 s -1. A single-channel FLIM system would still deliver reasonable lifetimes under these conditions but inevitably develop visible intensity nonlinearity. A time-series recording with the 8-channel system is shown in Fig. 4. A moss leaf was scanned with 407 nm excitation wavelength, and at a scan rate of 4 frames per second. The data format of the TCSPC data was 256 x 256 pixels x 256 time channels. The lifetime shown is an amplitudeweighted average of the two lifetime components of a double-exponential fit. The intensity scale of all images was normalised on the intensity of the brightest pixel. Fig. 4: Multi-wavelength time-series FLIM obtained from a moss leaf. Each image is 256 x 256 pixels, 256 time channels. Left to right: Wavelength, from 580 nm to 720 nm. Top to bottom: First four steps of the time series, acquisition time 5 seconds per step. eight-channel-flim-apn01.doc Oct. 2009 3

In the wavelength intervals from 580 nm to 640 nm the fluorescence comes from the cell membranes. At 660 nm fluorescence of the chloroplasts starts to show up. From 680 nm to 720 nm the images are totally dominated by chlorophyll fluorescence. The lifetime of the chloroplasts decreases over the time of exposure, due to induction of a non-photochemical quenching transient. The lifetime in the cell membranes remains constant. The variation of the lifetime can be seen more clearly in Fig. 5. The figure shows lifetime images in the 700 nm interval for the first four steps of the time series. The lifetime scale has been stretched to 200 ps to 400 ps. Decay curves in a selected spot of the images are shown in the second row. Fig. 5: First four steps of the time series, 700 nm interval. Upper row: Lifetime images. Lower row: Decay curve in pixel position X=51, Y=147. Summary The results presented show that an eight-channel parallel TCSPC FLIM system is feasible and operable. The advantage over a single-channel multispectral FLIM system [5, 6] is that the applicable count rate is 8 times higher. Moreover, possible saturation in a single spectral channel does not cause artefacts in other channels. The parallel-channel architecture allows relatively large FLIM data formats to be used. However, large pixel and time channel numbers in combination with the eight channels result in large data file size. Every single recording step of the time-series shown in Fig. 4 produces 260 Mbytes of data. The rate at which a time series can be recorded is thus limited by the speed at which the computer can save the data. The rate scales with the image format; for images of 128 x 128 pixels x 256 time channels a speed of about 1.3 seconds per step is reached. It should be noted that the 8-channel system is built from standard components of the bh modular FLIM systems. Moreover, the application of the system is not restricted to bh DCS-120 confocal scanning system. It can be used with any confocal or multiphoton laser scanning microscope that gives the user access to the fluorescence light via an optical port. 4 eight-channel-flim-apn01.doc Oct. 2009

References 1. R.M. Ballew, J.N. Demas, An error analysis of the rapid lifetime determination method for the evaluation of single exponential decays, Anal. Chem. 61, 30 (1989) 2. W. Becker, A. Bergmann, G. Biscotti, K. Koenig, I. Riemann, L. Kelbauskas, C. Biskup, High-speed FLIM data acquisition by time-correlated single photon counting, Proc. SPIE 5323, 27-35 (2004) 3. W. Becker, A. Bergmann, M.A. Hink, K. König, K. Benndorf, C. Biskup, Fluorescence lifetime imaging by timecorrelated single photon counting, Micr. Res. Techn. 63, 58-66 (2004) 4. W. Becker, Advanced time-correlated single-photon counting techniques. Springer, Berlin, Heidelberg, New York, 2005 5. W. Becker, The bh TCSPC handbook, third edition. Becker & Hickl GmbH (2006), available on www.beckerhickl.com 6. W. Becker, A. Bergmann, C. Biskup, Multi-Spectral Fluorescence Lifetime Imaging by TCSPC, Micr. Res. Tech. 70, 403-409 (2007) 7. Becker & Hickl GmbH, DCS-120 confocal scanning FLIM systems, user handbook. Available on www.beckerhickl.com 8. Becker & Hickl GmbH, Modular FLIM systems for Zeiss LSM 510 and LSM 710 laser scanning microscopes. User handbook. Available on www.becker-hickl.com 9. Govindjee, Sixty-three Years Since Kautsky: Chlorophyll Fluorescence, Aust. J. Plant Physiol. 22, 131-160 (1995) 10. V. Katsoulidou, A. Bergmann, W. Becker, How fast can TCSPC FLIM be made? Proc. SPIE 6771, 67710B-1 to 67710B-7 (2007) 11. M. Köllner, J. Wolfrum, How many photons are necessary for fluorescence-lifetime measurements?, Phys. Chem. Lett. 200, 199-204 (1992) 12. D.V. O Connor, D. Phillips, Time-correlated single photon counting, Academic Press, London (1984) eight-channel-flim-apn01.doc Oct. 2009 5