Self-navigation of STM tip toward a micron sized sample

Similar documents
A Project Report Submitted to the Faculty of the Graduate School of the University of Minnesota By

- Near Field Scanning Optical Microscopy - Electrostatic Force Microscopy - Magnetic Force Microscopy

PACS Nos v, Fc, Yd, Fs

Study of shear force as a distance regulation mechanism for scanning near-field optical microscopy

Scanning Tunneling Microscopy

A scanning tunneling microscopy based potentiometry technique and its application to the local sensing of the spin Hall effect

Radio-frequency scanning tunneling microscopy

INDIAN INSTITUTE OF TECHNOLOGY BOMBAY

Measurement of Microscopic Three-dimensional Profiles with High Accuracy and Simple Operation

Conductance switching in Ag 2 S devices fabricated by sulphurization

Prepare Sample 3.1. Place Sample in Stage. Replace Probe (optional) Align Laser 3.2. Probe Approach 3.3. Optimize Feedback 3.4. Scan Sample 3.

Electric polarization properties of single bacteria measured with electrostatic force microscopy

Proposal. Design of a Scanning Tunneling Microscope

Standard Operating Procedure of Atomic Force Microscope (Anasys afm+)

; A=4π(2m) 1/2 /h. exp (Fowler Nordheim Eq.) 2 const

Supporting Information. Vertical Graphene-Base Hot-Electron Transistor

Supporting Information. Atomic-scale Spectroscopy of Gated Monolayer MoS 2

Unit-25 Scanning Tunneling Microscope (STM)

Supplementary Figure 1 Schematic illustration of fabrication procedure of MoS2/h- BN/graphene heterostructures. a, c d Supplementary Figure 2

Akiyama-Probe (A-Probe) technical guide This technical guide presents: how to make a proper setup for operation of Akiyama-Probe.

Controller Design for Z Axis Movement of STM Using SPM Control Software

CONSTRUCTING A SCANNING TUNNELING MICROSCOPE FOR THE STUDY OF SUPERCONDUCTIVITY

attosnom I: Topography and Force Images NANOSCOPY APPLICATION NOTE M06 RELATED PRODUCTS G

SUPPLEMENTARY INFORMATION

Investigate in magnetic micro and nano structures by Magnetic Force Microscopy (MFM)

Near-field Optical Microscopy

IMAGING P-N JUNCTIONS BY SCANNING NEAR-FIELD OPTICAL, ATOMIC FORCE AND ELECTRICAL CONTRAST MICROSCOPY. G. Tallarida Laboratorio MDM-INFM

XYZ Stage. Surface Profile Image. Generator. Servo System. Driving Signal. Scanning Data. Contact Signal. Probe. Workpiece.

Giovanni P. Donati - MST-11 Daniel Some - MST-11 George Rodriguez - MST-11 Antoinette J. Taylor - MST-11

Lecture 20: Optical Tools for MEMS Imaging

Outline: Introduction: What is SPM, history STM AFM Image treatment Advanced SPM techniques Applications in semiconductor research and industry

photolithographic techniques (1). Molybdenum electrodes (50 nm thick) are deposited by

High resolution measurements The differential approach

Diamond X-ray Rocking Curve and Topograph Measurements at CHESS

Atomic Force Microscopy (Bruker MultiMode Nanoscope IIIA)

2. Pulsed Acoustic Microscopy and Picosecond Ultrasonics

INTRODUCTION: Basic operating principle of a MOSFET:

RHK Technology. Application Note: Kelvin Probe Force Microscopy with the RHK R9. ω mod allows to fully nullify any contact potential difference

BRIDGE VOLTAGE SOURCE

Recording EPR Spectra using ER 4102ST Resonator

:... resolution is about 1.4 μm, assumed an excitation wavelength of 633 nm and a numerical aperture of 0.65 at 633 nm.

Akiyama-Probe (A-Probe) guide

Fine structure of the inner electric field in semiconductor laser diodes studied by EFM.

Electrical Properties of Chicken Herpes Virus Based on Impedance Analysis using Atomic Force Microscopy

Akiyama-Probe (A-Probe) guide

attocfm I for Surface Quality Inspection NANOSCOPY APPLICATION NOTE M01 RELATED PRODUCTS G

Magnetic tunnel junction sensor development for industrial applications

Scanning eddy current dynamometer with 100 m resolution

Kit for building your own THz Time-Domain Spectrometer

Nanonics Systems are the Only SPMs that Allow for On-line Integration with Standard MicroRaman Geometries

Analysis of the process of anodization with AFM

Large Signal Displacement Measurement with an MTI Photonic Sensor Rev B

Resistive Switching Mechanisms on TaO x and SrRuO 3 Thin Film Surfaces Probed by Scanning Tunneling Microscopy

SPM The Industry s Performance Leader High Resolution Closed-loop System Fast, Easy Tip & Sample Exchange Versatility and Value Powerful Research

Design and characterization of 1.1 micron pixel image sensor with high near infrared quantum efficiency

Contents 1 Introduction 3 2 What is STM? 3 3 Scanning with 'easyscan' 4 4 Experiments Tip Preparation and Installation

Fabrication of large grating by monitoring the latent fringe pattern

Logic circuits based on carbon nanotubes

Operational Amplifiers

Advanced Nanoscale Metrology with AFM

MEMS for RF, Micro Optics and Scanning Probe Nanotechnology Applications

Weld gap position detection based on eddy current methods with mismatch compensation

in hbn encapsulated graphene devices

Supplementary Materials for

Spectrally Selective Photocapacitance Modulation in Plasmonic Nanochannels for Infrared Imaging

Park NX-Hivac: Phase-lock Loop for Frequency Modulation Non-Contact AFM

The Physics of Single Event Burnout (SEB)

3D simulations of the experimental signal measured in near-field optical microscopy

Optimal Preamp for Tuning Fork signal detection Scanning Force Microscopy. Kristen Fellows and C.L. Jahncke St. Lawrence University

LOW TEMPERATURE STM/AFM

Scanning Tunneling Microscopy

ATOMIC FORCE MICROSCOPY

State of the Art Room Temperature Scanning Hall Probe Microscopy using High Performance micro-hall Probes

Design, Fabrication and Characterization of Very Small Aperture Lasers

Lab Equipment EECS 311 Fall 2009

RECENTLY, using near-field scanning optical

SUPPLEMENTARY INFORMATION

Defense Technical Information Center Compilation Part Notice

The 34th International Physics Olympiad

PFM Experiments with High Voltage DC/AC Bias

In this lecture we will begin a new topic namely the Metal-Oxide-Semiconductor Field Effect Transistor.

Communication using Synchronization of Chaos in Semiconductor Lasers with optoelectronic feedback

Supplementary Figure 1 High-resolution transmission electron micrograph of the

DIODE lasers have some very unique qualities which have

An advanced impedance calibration method for nanoscale microwave imaging

Author(s) Issue Date Text Version author. DOI / /18/9/095501

GaAs polytype quantum dots

Active mechanical noise cancellation scanning tunneling microscope

Tip-induced band bending and its effect on local barrier height measurement studied by light-modulated scanning tunneling spectroscopy

Supplementary Information. The origin of discrete current fluctuations in a fresh single molecule junction

Impact of module parasitics on the performance of fastswitching

NanoFocus Inc. Next Generation Scanning Probe Technology. Tel : Fax:

High-speed wavefront control using MEMS micromirrors T. G. Bifano and J. B. Stewart, Boston University [ ] Introduction

Nano-structured superconducting single-photon detector

StarBright XLT Optical Coatings

A Prototype Wire Position Monitoring System

PowerAmp Design. PowerAmp Design PAD541 COMPACT POWER OP AMP

Large Signal Displacement Measurement with an Asylum SA Atomic Force Microscope Rev B

Application Note (A11)

Nanovie. Scanning Tunnelling Microscope

Transcription:

Self-navigation of STM tip toward a micron sized sample Guohong Li, Adina Luican, and Eva Y. Andrei Department of Physics & Astronomy, Rutgers University, Piscataway, New Jersey 08854, USA We demonstrate a simple capacitive based method to quickly and efficiently locate micron size conductive samples on insulating substrates in a scanning tunneling microscope (STM). By using edge recognition the method is designed to locate and identify small features when the STM tip is far above the surface allowing for crash-free search and navigation. The method can be implemented in any STM environment even at low temperatures and in strong magnetic field, with minimal or no hardware modifications. I. INTRODUCTION A scanning tunneling microscope 1,2 (STM) is a powerful tool to study materials with atomic resolution. In the STM topography mode a sharp metallic tip scans above a conductive sample surface while monitoring the tip-sample tunneling current which depends exponentially on the distance. Typically the measurement is carried out at a tip-sample distance of order 1 nm. The precise control of tip position (x, y, z) is usually realized by employing a piezo-electric tube scanner with pico-meter resolution 2. However, the high resolution of STM sets a limit to its field of view, the largest of which is usually about 1 m. In practice 2, one needs additional coarse positioning stages for tip navigation parallel to the sample surface (x, y) and for perpendicular (z) approach. These stages typically travel distances of a few mm with step size of 10-1000 nm. Although a reliable z-stage is a must for any STM, 1

the x-y navigation is not crucial for standard STM measurements on mm sized samples because the typical features appear everywhere. Recently there has been increased interest in studying micron sized samples, such as a graphene flake on a SiO 2 substrate. However, because the STM is intrinsically nearsighted it is quite challenging to locate a specific micron sized sample on a macroscopic substrate. Since STM images topography it is natural to try using topographic features 3,4 as guides. But this method is not practical because of the limitations inherent to the technique: the small field of view requires comparing many images frame by frame; the even smaller working distance does not favor high speed scanning due to the risk of tip crashing; the finite dynamic range of piezoelectric scanner in z direction, typically less than 300nm, usually cannot cover the height variation over distances of tens of m and therefore frequent tip retracting and approaching are necessary during the navigation. These factors render the STM navigation extremely slow. A further complication arises in the presence of insulating-conductive-boundaries, where navigating the STM tip will surely result in a crash into nearby insulating areas. It is possible to circumvent these complications by using additional setups, such as a scanning electron microscope 5 or a long range optical microscope 6, to visualize the STM tip during navigation. However these external aids are impractical at low temperatures and in high magnetic fields due to the harsh environment and limited space. Here we report a capacitance-based method to navigate the STM tip, which allows finding micron sized samples quickly and efficiently. The method consists of back-gate compensation 7,8, refocusing during navigation, and distinguishing edges of conducting electrodes and the sample. It requires no additional setup other than two independent bias voltages, one for the sample and 2

the other for a back-gate. The latter is simply a conductive plane below the sample separated by a thin insulating layer. II. CAPACITIVE PICKUP AND COMPENSATION The signal of interest in STM is the tunneling current: I GV (1) t s where V s is the sample bias voltage and G t the tunneling conductance. To measure scanning tunneling spectroscopy 2, one usually applies a small AC modulation, V ~ s, to the sample bias voltage so that there is an AC current, I ~, flowing through the STM tip, ~ ~ ~ I GV i CV (2) t s s There are two contributions to the AC current, one from tunneling (first term) and the other from capacitive pickup (second term) via the effective tip-sample capacitance C. Since the capacitive current is not limited to the tunneling regime, the two contributions are easily distinguishable. The pickup current is a function of the geometry through the tip-sample capacitance and can be used to roughly monitor tip approach starting from a few mm away from the sample surface. We will show that this signal can be used to resolve small structures. In principle, capacitive currents should depend on the relative lateral (x and y) position between tip and sample. As we observed experimentally, the effective capacitance change is ~30 af (10-18 F) when an STM tip with 3 m tip-sample distance scans across a 5 m conductive strip deposited on an insulating substrate. However, the background pickup is usually several orders of magnitude larger, ~ 6 ff (10-15 F). To measure the small change in capacitance, one needs very good background compensation, which could be accomplished with a capacitance bridge setup. We show that in most STM designs it is in fact possible, with only a minor modification to 3

use the sample as part of a built-in bridge circuit. For example, a graphene flake on SiO 2 is supported on a heavily doped silicon gate electrode. This gate can act as one arm of the capacitance bridge as illustrated in Fig. 1(b) and thus one can tune the out-of-phase voltage -V ~ gate to cancel the background current. In this circuit the variation of the tip current, d I ~, is directly ~ di dc proportional to dc t-s, the tip-sample capacitance and so are their spatial derivatives dx dx t s. It is worth noting that the presence of back gate also changes the electric field distribution around the sample. Without the gate as shown in Fig.1(c), the equipotential lines quickly lose the shape of the sample. A grounded back gate pushes up the equipotential lines [Fig.1(d)]. However, a dramatic change occurs when an opposite voltage is applied to the back gate. As seen in Fig.1(e), the equipotential lines develop strongly varying features above the edges which remain sharp and clearly distinguishable at distances comparable to the sample size. Thus one could resolve the thin bar with large tip sample distance. We next focus on the experimental results. Although the method reported here was successfully used in our recent study 7 of Landau levels in graphene on SiO 2 at 4.4K and 12T, the data reported here were taken at room temperature in order to have an independent check on tip navigation from an optical microscope. Typical AC voltages applied to the sample were 200 mv rms at 5 khz. The frequency is limited by the bandwidth of the current amplifier (10 9 V/A gain), which is also used to measure tunneling currents. 4

FIG.1. (Color online) Use of back gate in capacitance measurements. (a) Schematic experimental setup. (b) Equivalent circuit. The output voltage from the reference channel of a lock-in amplifier is split into two with 180 phase shift. One (+) is applied to the sample directly as V ~ ~, the other (-) is applied to the gate ( ) s through a pot resistor to adjust the amplitude. Capacitive pickup currents are measured with the same amplifier (hashed area) that is used for tunneling currents. (c) Electric field distribution near a conducting bar extending out of the paper. The strip is at 1V. Lines are equipotential contours and arrows show field directions. (d) and (e) Same as (c) but with a nearby back gate grounded and at -1V, respectively. Red arrows mark the steep potential lines near the sample edges. (c) to (e) were calculated with Field Precision TC (educational 7.0). V gate 5

III. EDGE DETECTION Fig.2 shows the variation of the capacitive currents as the STM tip moves above a 200 m wide Au film. The tip is 60 m away from the sample surface in Fig.2(a). As the tip moves across the sample, the current peaks near the center of the film. The asymmetry is due to the background, e.g. overall sample geometry and wiring. This background current could significantly distort the peak when the tip is 210 m away from the sample surface, as shown in Fig.2(b). FIG.2. (Color online) Variation of capacitive currents as STM tip moves above thin film. (a) Tip 60 m above a 200 m wide film. Thin line: pickup current. Thick line: spatial derivative di/dx. Insert: schematic of tip sample geometry. (b) Same as (a) but with tip-sample distance of 210 m. (c) spatial derivative di/dx for various tip-sample distances. Data are centered relative to the shaded area that shows the 200 m wide sample. Shaded area marks the film width. (d) Same as (c) but for a film with 5 m width. 6

Although the capacitive currents in Figs. 2(a) and (b) look quite different, their spatial derivatives, di/dx, appear to be similar and symmetric. The two turning points in di/dx correspond to the edges of the film. Fig. 2(c) plots a systematic study for various tip sample distances. The sharpness of the turning points decreases with increasing distance. Still the central part of the sample can be easily identified even if the tip-sample distance is comparable to the sample width. However, to identify the edges accurately, the distance has to be sufficiently small compared to the sample width so as to produce sharp turning points above the sample edges. For example, in Fig.2(d), the tip-sample distance has to be less than ~1.6 m to resolve the edges of a 5 m wide film. To check the reliability of this method, we repeated the measurements on films with different width shown in Fig. 3(b). As illustrated in Figs. 2(c) and 2(d), the di/dx signal exhibits sharp peaks at the position of the sample edges if the aspect ratio (tip height over sample width) is less than ~0.3. Therefore as the features become smaller it is important to approach the tip toward the sample surface in order to obtain the desired resolution. This is similar to the concomitant focus and magnification adjustment in an optical microscope. Fig. 3(a) compares the measured pad size obtained by this method with the actual size obtained from the optical image in Fig. 3(b). The measured sizes are very close to the actual ones for all film width down to the smallest measured feature provided the aspect ratio is kept below 0.3. For larger aspect ratios the measured widths are systematically larger than the actual widths. This error if not taken into account could lead to a fatal tip crash. Below we describe a navigation protocol which allows finding a micron-size sample reliably and safely. 7

FIG. 3. (Color on line) Sample widths measured by the capacitance method illustrated in Fig.2. (a) Measured size versus actual size. Tip sample distances adjusted for sharp edge resolution corresponding to aspect ratios of 0.3 or less. Red line is a guide to the eye through the values of the measured widths. (b) Optical image of sample. IV. PROTOCOL OF NAVIGATION Before demonstrating how to find a micron sized sample, we emphasize the key points in the navigation protocol. a) Identify the central region of a conductive sample from the spatial derivative of the capacitive currents. This step, which is done with the tip far away from the sample surface, (Fig.2) guarantees that the tip is targeting a conductive region. b) Use the STM mode to find the sample surface safely and retract the tip to a height corresponding to an aspect ratio of ~0.3. This step, similar to re-focusing in an optical microscope, enables sharper contrast. c) Identify the edges of a conductive sample from the spatial derivative of capacitive currents. 8

FIG.4. (Color online) Protocol of navigation. (a) Initial alignment between tip and sample at room temperature. (b) Possible drift of tip position after transferring and cooling down. (c) Tip position adjusted by centering and working in STM mode as discussed in text. (d) Tip near one edge ready to search for the lower boundary of the pad. (e) Derivative of capacitive currents in the first scan along the dashed line shown in (b). Arrow indentifies the center of the pad (f) Derivative of capacitive currents in along the dashed line in (c) with a smaller tip sample distance. Dotted line marks the choice of tip position in (d). (g) Derivative of capacitive currents in along the arrow in (d) to find the lower boundary. Units in (e)-(g) are arbitrary. Space directions are defined in the upper-right insert. Following these key points, navigation procedure will depend on the geometry of the metal pads contacting the micron sized sample of interest. Below we show how this procedure is applied to locate the sample in the left-lower corner of Fig.4 (a). 9

Starting at room temperature, we can easily position the tip ~0.2mm above the biggest pad using optical access [Fig.4(a)]. Once the STM is transferred into a magnet and is cooled down to 4K, the tip could drift away [Fig.4(b)] because of mechanical perturbations and thermal contractions. However, with a well-designed STM head, the drift is rarely larger than 100 m in any direction. Considering the weak tip height dependence of the overall structure in di/dx (Fig.2), one can always do an initial scan across the big pad to identify its center from di/dx in Fig.4(e) and position the tip accordingly. Once it is established that the STM tip is targeting a conductive surface, the surface is found using the STM tip-approach mechanism. During this stage the large AC modulation is turned off and the DC sample bias voltage, say 500mV, remains. When the surface is found, the tip is retracted to the appropriate distance [Fig.4.(c)]. This distance should be large enough to accommodate height variations of the sample during tip movements. We note that the height variation does not affect the identification of edges significantly because of the weak dependence shown in Fig. 2(c). With reduced tip-sample distance, the edges can be identified with better accuracy as shown in Fig. 4(f). Subsequently the tip is positioned near one of the edges [Fig.4(d)] and moved along the edge. The lower boundary of the big pad appears as a dip in di/dy [Fig.4(g)]. The procedure is then repeated on the smaller pad. For optimal imaging conditions the tip is retracted less after finding the surface of the smaller pad so as to maintain the aspect ratio within 0.3. Thus, as the tip approaches the targeted sample it also gets closer to the surface. V. CONCLUSIONS 10

We have demonstrated a capacitive pickup method and a navigation protocol to locate small samples on an insulating substrate using an STM tip. The method employs metallic guiding pads whose edges can be clearly identified by using back gate compensation and the spatial derivative of capacitive currents. This capacitance based method involves minimal modifications to the STM setup and can be applied more generally to other scanning probe microscopes equipped with coarse motor navigation. ACKNOWLEDGEMENTS This work was supported by the DOE under Grant No. DE-FG02-99ER45742 and partially supported by Grant No. NSF-DMR-0906711 and by Lucent. 1 G. Binnig, H. Rohrer, Ch. Gerber, and E. Weibel, Phys. Rev. Lett. 49, 57 (1982). 2 C. J. Chen, Introduction to Scanning Tunneling Microscopy (Oxford University Press, 2007). 3 E. Stolyarova, K. Rim, S. Ryu, J. Maultzsch, P. Kim, L. Brus, T. Heinz, M. Hybertsen, and G. Flynn, PNAS 104, 9209 (2007). 4 K. Xu, P. G. Cao, and J. R. Heath, Nano Lett. 9, 4446 (2009). 5 M. Ishigami, J. J. Chen, W. G. Cullen, M. S. Fuhrer, and E. D. Williams, Nano Lett. 7, 1643 (2007). 6 V. Geringer, M. Liebmann, T. Echtermeyer, S. Runte, M. Schmidt, R. Rückamp, M. C. Lemme, and M. Morgenstern, Phys. Rev. Lett. 102, 076102 (2009). 7 P. W. Kolb, R. S. Decca and H. D. Drew, Rev. Sci. Instrum. 69, 310, (1998). 8 S. B. Field and J. Barentine Rev. Sci. Instrum. 71, 2603 (2000). 9 A. Luican, G.Li, and E.Y. Andrei, Phys. Rev. B. 83, 041405 (2011). 11

12