Applications Note. Successfully using the Model 8000 NERO dental x-ray machine timer performance. max to evaluate

Similar documents
Nuclear Associates

Fluke199XRAY. Users Manual. Medical ScopeMeter

IX Feb Operation Guide. Sequence Creation and Control Software SD011-PCR-LE. Wavy for PCR-LE. Ver. 5.5x

Ansur TNT Users Manual. Plug-In

Nuclear Associates

35080M kvp Divider. Operators Manual

Nuclear Associates

Nuclear Associates

Nuclear Associates

How to measure mas using the current clamp with TNT 12000

Progeny Imaging Veterinary

Nuclear Associates EZ CR-DIN Phantoms

APPLICATION NOTE. RTI Electronics AB, Sweden Revision A, July kvp and time measurements with the Barracuda or Piranha on a typical dental unit

ACCU-GOLD QUICK START MANUAL

Victoreen 4000M+ X-Ray Test Device

Nuclear Associates

EKA Laboratory Muon Lifetime Experiment Instructions. October 2006

Nuclear Associates

EA SPORTS PGA TOUR Golf Team Challenge Upgrade Instructions

Progeny Imaging. User Guide V x and Higher. Part Number: ECN: P1808 REV. F

Alignment of the camera

My Blogs: To Add New Blog Post: o Click on the My Learn360 link. You will then see eight different tabs (below).

Exercise 8. Troubleshooting a Radar Target Tracker EXERCISE OBJECTIVE

1.1. Log on to the TUMI system (you cannot proceed further until this is done).

GE AMX 4+ Portable X-Ray

SKF Static Motor Analyzer Baker DX Series

PANalytical X pert Pro High Resolution Specular and Rocking Curve Scans User Manual (Version: )

SOP: EDAX Eagle III Microspot XRF

TNT DoseMate Dosimeter Detector

Nuclear Associates , , CT Head and Body Dose Phantom

Nuclear Associates

Operating Procedures for MICROCT1 Nikon XTH 225 ST

Nuclear Associates

The ideal K-12 science microscope solution. User Guide. for use with the Nova5000

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

facewho? Requirements Analysis

Nuclear Associates , &

Table 1. Placing the Sensor in the Sensor Cradle. Step Instruction Illustration

R&S RTO-K92 emmc Compliance Test Test Procedures

SYNGUIDER USER'S MANUAL

EULAMBIA ADVANCED TECHNOLOGIES LTD. User Manual EAT-EOM-CTL-2. Alexandros Fragkos

Conversational CAM Manual

Calibration check of dosimeters measuring whole body vibrations. Calibration check bench user manual

OPERATOR'S INSTRUCTIONS

SARG: The Graphical User Interface Manual

Introduction to Oscilloscopes Instructor s Guide

Veterinary Digital X-Ray System Quick Start Guide

User Operation of JEOL 1200 EX II

Figure 1 The Raith 150 TWO

Nuclear Associates

Digital Dental Quality Assurance Manual Initial Baseline and Longitudinal Assessment

PX8000 Precision Power Scope with Features of High-accuracy Power Meter and Waveform Measuring Instrument

TEP Principles of Digital X-ray Imaging

How to remove Jammed Paper

Introduction to Circular Pattern Flower Pot

expertmeter High Performance Analyzer PM180 Fault Locator Application Note BB0165 Rev. A2

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: )

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

PROGRAMMABLE INTRACRANIAL SELF- STIMULATION (ICSS) CURRENT STIMULATOR

UCE-DSO212 DIGITAL OSCILLOSCOPE USER MANUAL. UCORE ELECTRONICS

Using the R-Axis Spider to run thin films

Reflection and absorption of sound (Item No.: P )

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS

Nikon Firmware Update for Coolpix 950 Version 1.3

ScopeMeter Test Tool CSV files available for FlukeView software and Microsoft Excel Application Note

Operating Instructions Pocket Pictor For use with Pocket Pc s

Exercise 2: Hodgkin and Huxley model

Ph 3455 The Franck-Hertz Experiment

1. Preliminary sample preparation

Quick Start Guide for the PULSE PROFILING APPLICATION

Unit. Drawing Accurately OVERVIEW OBJECTIVES INTRODUCTION 8-1

UCE-DSO210 DIGITAL OSCILLOSCOPE USER MANUAL. FATIH GENÇ UCORE ELECTRONICS REV1

Agilent 5DX Series 2L, 3, and 5000 X-ray Safety Test Report

Experiment 2: Electronic Enhancement of S/N and Boxcar Filtering

ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY or local phone

ThermaViz. Operating Manual. The Innovative Two-Wavelength Imaging Pyrometer

POLYMOBIL III. Function description X038I. Wiring index E from Serial no. 3000

Nuclear Associates

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

SPECTRALIS Training Guide

MEASURE Evaluation. Global Positioning System Toolkit

Direct waveform transfer from a DS1000Z scope to a DG4000 generator

DUKANE. Frequency Tracking in iq Systems APPLICATION NOTE AN513. Intelligent Assembly Solutions

Notes on DYNALYZER HVU Usage

Time-Varying Signals

RADIOACTIVITY INDICATOR RADIASCAN-501(701) OPERATIONS MANUAL

Version 1.0. TechnicVR. Student Guide

iworx Sample Lab Experiment AN-2: Compound Action Potentials

ArbStudio Triggers. Using Both Input & Output Trigger With ArbStudio APPLICATION BRIEF LAB912

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual

MAXYM Mortiser Operating Manual

Recovering highlight detail in over exposed NEF images

KODAK RVG 6100 and 5100 Digital Radiography Systems

Lesson 4 Extrusions OBJECTIVES. Extrusions

Exercise 6. Range and Angle Tracking Performance (Radar-Dependent Errors) EXERCISE OBJECTIVE

ArbExpress AXW100 Waveform Creation and Editing Tool for Tektronix AWG/AFG Version

Standard Operating Procedure

How to remove Jammed Paper

CNC Chucker Lathe P/N 6600, 6610, and 6620

RaySafe X2. Effortless measurements of X-ray

Transcription:

Successfully using the Model 8000 NERO dental x-ray machine timer performance max to evaluate Information in this article applies to Model 8000 NERO max firmware release 2.3 or higher. Introduction The more that is known about the characteristics of the x-ray machine under test, the easier it is to achieve reasonable results in a timely manner. For instance, if it is known that the x-ray machine does not utilize a filament preheat, the use of a measurement delay is probably unnecessary. However, if the x-ray machine is known to have a filament preheat, a measurement delay is most likely needed. If the x-ray machine's timer is set in pulses, using the NERO max in its pulse counting mode (%kv set to 1φPULSE) will yield exposure time results in pulses. If the x-ray machine's timer is set in fractions of seconds or milliseconds, a %kv setting of ZERO, 75%, 80% or 90% may be applicable (depending upon manufacturers specifications or other criteria). Typically, for single-phase self-rectified dental units a %kv setting of 1φPULSE or ZERO is used. For dental x-ray units with three phase or high frequency generators, the %kv should be set to 75%, 80% or 90%.. Making a measurement 1. Insert the W/Al filter card into the detector and set it to the 50-100 kv range. 2. Set the detector on a flat, stable surface and position the x-ray tube so that the cone just touches the detector's top surface. Note: The NERO max detector's ion chamber is located at the detector's top surface. Do not depress the top of the ion chamber. This can cause incorrect exposure measurements. Make sure that the detector is aligned along the axis of the x-ray tube and that the tube is perpendicular to the detector's top surface and is centered over the detector crosshairs. 3. Turn on the NERO max and select RADIO mode. 4. Select LOW sensitivity. 5. Set the measurement delay to 0 milliseconds. 6. If the x-ray machine's timer is set in pulses, set the %kv to 1φPULSE. If the x-ray machine's timer is set in seconds, fractions of seconds or milliseconds, set %kv to ZERO, 75%, 80% or 90% (depending upon manufacturers specifications or other criteria). 7. Press the ENTER button and wait for the MAKE EXPOSURE prompt. 8. Make an exposure. 9. If the measured exposure time or number of pulses is within the allowed tolerance, the x-ray machine probably does not have a filament preheat or a malfunctioning timer and the results of the exposure are probably correct. If the measured exposure time or number of pulses is significantly less than the set exposure time or pulses, the x-ray machine may have a filament preheat or a malfunctioning timer. When this occurs, a measurement delay can be used to determine whether a filament preheat is used or if the timer is malfunctioning. For a solution, see "What to do if the measured exposure time is too short". If the measured exposure time or number of pulses is significantly longer than the set exposure time or pulses, the x-ray machine may have a filament preheat or its timer is malfunctioning. For a solution, see "What to do if the measured exposure time is too long". 8000AN1 rev 1 31 oct 05 Page 1 of 6

What to do if the measured exposure time is too short Short exposure times may be caused by a timer that is malfunctioning or filament preheat characteristics. To determine whether a short measured exposure time is due to a malfunctioning timer or filament preheat characteristics: 1. If the NERO max is still displaying the results of the last exposure, press the MODE key to return to the menu screen. 2. Set the %kv to ZERO. 3. Press the ENTER button and wait for the MAKE EXPOSURE prompt. 4. Make an exposure. 5. Record the displayed exposure time. 6. Press the MODE key to return to the menu screen. 7. Enter a NEGATIVE measurement delay greater than the exposure time measured with no delay. For instance, if the measured exposure time of the previous exposure was 18 milliseconds, enter a delay of -20 milliseconds. Generally, rounding the delay up to the nearest ten milliseconds is adequate. A negative delay instructs the NERO max not to include the delay time in the measured exposure time. 8. Error! No sequence specified.set the %kv to that used for the first exposure, prior to setting the %kv to ZERO. 9. Press the ENTER button and wait for the MAKE EXPOSURE prompt. 10. Make an exposure. 11. If the measured exposure time or number of pulses is within the allowed tolerance, this x-ray machine probably has a filament preheat and the measurement delay is sufficient to skip over the effects of the preheat. If the NERO max displays a "DELAY TOO LONG" error message, the x-ray machine's timer may be malfunctioning. This error message is displayed if no x-rays are detected within one second after the delay time has elapsed. When a measurement delay is used, the NERO max waits for up to one second after the delay time has elapsed for an exposure to resume. If no x-rays are present after this time, the "DELAY TOO LONG" error message is displayed. What to do if the measured exposure time is too long Long exposure times may be caused by a timer that is malfunctioning or filament preheat characteristics. To accurately determine the cause of long measured exposure times requires examination of the machine's output radiation and/or kv waveforms. If excessive measured exposure times are due to filament preheat effects as shown in the first waveform of the following example, the NERO max Excel Add-In may be used to determine the appropriate measurement delay to use. To determine the correct measurement delay to use: 1. If the NERO max is still displaying the results of the last exposure, press the MODE key to return to the menu screen. 2. Connect a computer to the NERO max and start the NERO max Excel Add-In. Information on installation and use of the NERO max Excel Add-In may be found in the NERO max Excel Add-In manual. 3. Open a new Excel workbook by clicking on File then New and selecting Workbook from the "New" menu box or click on the "New" toolbar icon. 4. Click on worksheet cell A1. 5. Press the ENTER button on the NERO max and wait for the MAKE EXPOSURE prompt. 6. Make an exposure. 7. After the exposure results are displayed in the first row of the Excel worksheet, click on NERO max and select Retrieve Rad Waveform. 8. After the radiation waveform is displayed, examine the waveform for preheat characteristics such as those shown in the first waveform of the following example. If no preheat effects are present; the x-ray machine's timer may be malfunctioning. 8000AN1 rev 1 31 oct 05 Page 2 of 6

9. Determine the measurement delay time that is adequate to skip over the preheat portion of the output waveform. The delay should terminate after the last pulse of the filament preheat. 10. If the NERO max is still displaying the results of the last exposure, press the MODE key to return to the menu screen. 11. Enter the measurement delay needed to skip over the filament preheat effects. Enter a NEGATIVE delay if the delay time is not to be included in the measured exposure time. 12. Press the ENTER button and wait for the MAKE EXPOSURE prompt. 13. Make an exposure. 14. If the measured exposure time or number of pulses is not within the allowed tolerance, the x-ray machine's timer may be malfunctioning and further examination of the output waveform can be used to confirm the results. A digital storage oscilloscope may also be connected to the scope output on the rear panel of the NERO max readout to view radiation output waveforms. Information on using the NERO max scope output may be found in section 3.2 of the NERO max instruction manual. Examples The radiation waveform below is from a single-phase dental x-ray machine (Gendex GX-770) that utilizes a filament preheat. The exposure time on this generator set in pulses (instead of milliseconds) but does not include the filament preheat pulses. For instance, when the exposure time control is set for 10 pulses, the generator actually outputs 24 pulses, 14 of which are filament preheat pulses. Radiation Waveform 30-60 kv range 30000 25000 20000 counts 15000 Channel A Channel B 10000 5000 Filament Preheat Pulses 0 0 50 100 150 200 250 300 350 400 450 msec 8000AN1 rev 1 31 oct 05 Page 3 of 6

Below is an example of what the NERO max sees when making a measurement without a delay on this type of generator. Note that the NERO max only sees two pulses during the filament preheat period. This is because the intensity of the remaining filament preheat pulses is so low that they fall below the detectability limit of the NERO max. This causes the NERO max to stop acquiring data, ending the measurement after the first two pulses resulting in a measured exposure time of 18.36 milliseconds. Radiation Waveform 500 450 400 350 counts 300 250 200 Channel A Channel B 150 100 50 0 0 2 4 6 8 10 12 14 16 18 20 msec 8000AN1 rev 1 31 oct 05 Page 4 of 6

To accurately assess this generator s performance, a measurement delay should be used to skip over the filament preheat pulses. The radiation waveform below shows the effect of using a - 20 millisecond delay. Again, the x-ray generator was set for 14 pulses but the measurement delay caused the NERO max to skip the first 2 preheat pulses and only record the desired portion of the generator s output. Radiation Waveform -20 Millisecond Delay 6000 5000 4000 counts 3000 Channel A Channel B 2000 1000 0 0 20 40 60 80 100 120 140 160 180 msec The NERO max Excel Add-In may be used to view radiation and kv waveforms to determine the optimum measurement delay to use. Information on using the NERO max Excel Add-In to view waveforms may be found in the NERO max Excel Add-In manual. In addition, a digital storage oscilloscope may be connected to the scope output on the rear panel of the NERO max readout to view radiation output waveforms. Information on using the NERO max scope output may be found in section 3.2 of the NERO max instruction manual. Technical assistance For technical assistance, please contact Fluke Biomedical, 6045 Cochran Road, Cleveland OH 44139, USA, tel 440.248.9300, fax 440.349.2307, email rmsinfo@flukebiomedical.com, or internet www.flukebiomedical.com/rms. 8000AN1 rev 1 31 oct 05 Page 5 of 6

FLOWCHART: HOW TO USE THE MODEL 8000 NERO max TO EVALUATE DENTAL X-RAY MACHINE TIMER PERFORMANCE Select Radio Mode Low Sensitivity Set %kv to the appropriate value and press ENTER Seconds Is timer set in pulses or seconds? Pulses Select Radio Mode Low Sensitivity Set %kv to 1φ PULSE and press ENTER Too Short Is exposure time within tolerance, too short, too long? Too Long Select Radio Mode Low Sensitivity Set %kv to ZERO and press ENTER Within Tolerance DONE Examine the output waveform with the NERO max Excel Add-In or digital oscilloscope for filament preheat effects. NOTE: Exposure time cannot exceed 480mS when using the NERO max Excel Add-In to view waveforms. Record the measured exposure time Examine the output waveform with the NERO max Excel Add-In or digital oscilloscope to determine exposure time. NOTE: Exposure time cannot exceed 480mS when using NERO max Excel Add-In to view waveforms. NO Is there a filament preheat? Press the MODE key to exit to the menu screen. ENTER A NEGATIVE DELAY SLIGHTLY LONGER THAN THE MEASURED EXPOSURE TIME Set %kv to the appropriate value YES Press the MODE key to exit to the menu screen. ENTER A NEGATIVE DELAY SUFFICIENT TO SKIP OVER THE FILAMENT PREHEAT PORTION OF THE OUTPUT WAVEFORM and press the ENTER key Is exposure time within tolerance? NO Examine the output waveform with the NERO max Excel Add-In or digital oscilloscope to determine exposure time. NOTE: Exposure time cannot exceed 480mS when using NERO max Excel Add-In to view waveforms. YES DONE 8000AN1 rev 1 31 oct 05 Page 6 of 6