1000TR. Instructions

Similar documents
Model. Instruction Manual. Digital ph Pen. reedinstruments. REED Instruments

Instruction manual SMART 96-2 / WG-2 ph

Microprocessor Transmitter/Controller for ph - Type for redox - Type B Operating Instructions 9.

W100/W600/W900 Controller ph Electrode Troubleshooting Guide

ph or ORP Transmitter

Troubleshooting Guide for Aquatrac Smart AS and Flex

UNICONT. PMG-400 Universal controller and display unit USER'S AND PROGRAMMING MANUAL 1. pmg4111a0600p_01 1 / 24. ST edition

COMPACT-LINE. KOBO-RP Transmitter for ORP Model ARM-Z

PH AUTOCLEAN ph + ORP CONTROLLER MICROPROCESSOR BASED

DIGITAL ph / ORP / mv CONTROLLER PMH-2 USERS GUIDE. PMH-2 Instruction manual PAGE 1

COMPACT-LINE. KOBO-pH Transmitter for ph-value Model APM-Z

GREISINGER electronic GmbH D Regenstauf, Hans-Sachs-Straße 26 phone: / , fax: / ,

PH METER OPERATION MANUAL. Model : PH-208

4-20mA, 0-10V and general process measuring bargraph displays

INSTRUCTION MANUAL Redox sensor M 1322 C

Measuring and Controlling Unit Multronic OC

Datasheet Platinum Resistance Pt100 In-head (Push Button) Temperature Transmitter TX203P mA output, default range supplied C

OPTISENS PH 8390 Technical Datasheet

INSTRUCTION MANUAL DISSOLVED OXYGEN-METER MODEL CDO-01

SMARTPAT PH 1590 Technical Datasheet

Conductivity +/ 2% 1 or 2 point remotely through PLC or directly on board. Any two lead Conductivity probe (K 0.01, K 0.1, K 1.

DIGITAL proportional ph CONTROLLER PPH-1 USERS GUIDE. PPH-1 Instruction manual PAGE 1

Signet 2751 DryLoc ph/orp Smart Sensor Electronics

C 2030 חפר מיכשור ומערכות בע"מ טכנולוגיות לטיפול במים ושפכים מיכשור אנליטי למדידה ובקרת תהליכים OPERATOR'S MANUAL. 2- Wire E.C transmitter Din Rail

SPECIFICATION SHEET. 2-WIRE TYPE ph ANALYZER/TRANSMITTER SYSTEM CONFIGURATION

Lines, plugs and sockets for ph, redox, conductivity and temperature sensors

ABB ph/orp analyzer and probe package

USER MANUAL. EPP Intelligent Positioner Control Unit 1/22.

Process Calibrator. TechChek 820

Thermo Scientific Orion Star A111 ph Benchtop Meter

OVEN INDUSTRIES, INC. Model 5C7-362

COMPACT-LINE. Transmitter for ph-value

OPTISENS PH 8500 Handbook

Measuring and Controlling Unit Multronic

Operating instructions Electronic level sensor LK / / 2008

IXIAN TM class V 1.5. EC Transmitter This is an evolving document. EC Transmitter IXIAN TM. Features

DUAL OUTPUT AC CURRENT/VOLTAGE TRANSDUCER

MHPS MHPS. Modular pressure transmitter. Technical documentation. Table of content. Characteristics - applications - technical data

Please enter the identity code of your device here!

Installation and Operating Instructions. Power IT Power Factor Controller RVC

R7420B. MicroniK 100 HONEYWELL TEMPERATURE CONTROLLER PRODUCT DATA GENERAL FEATURES

Remote Controlled FOUNDATION Fieldbus Two-Wire ph/orp Transmitter

Endress Hauser. SmarTec M CLD 133 Conductivity Measurement

Conductivity Transmitter N-LF V AC

CX105 Conductivity/Resistivity Transmitter

M1 Series. Humidity - Temperature Transmitter INSTRUCTION MANUAL

Please enter the identity code of your device here!

OPTISENS ORP 8590 Technical Datasheet

DISSOLVED OXYGEN TRANSMITTER

high accuracy tester-calibrator PJ 6301

8000 SERIES PRECISION MULTIMETER VERIFICATION AND ADJUSTMENT GUIDE

Dual Channel Conductivity Controller EC Operational Manual CH1 CH ATC 25.0 MTC 25.0

EV-06 with option AAG...

USER'S GUIDE. Vaisala HUMICAP Humidity and Temperature Transmitters HMD70U/Y U303EN11

HBPH-MK2 Brine/NH3 Leakage Sensor

Please completely read through operating instructions! Do not discard! The warranty shall be invalidated by damage caused by operating errors!

CONDUCTIVITY/ TDS METER + PH meter. Model : DCT2001

EXPERT-LINE. KOBO-pH. Transmitter for ph-value and ORP Model APM-X

Universal Alarm Indicator

Dissolved Oxygen Transmitter

Operating Instructions

Please enter the identity code of your device here!

SEM104 SERIES. SEM104P Pt100 Temperature Transmitter. SEM104TC Thermocouple Temperature Transmitter INDEX SECTION CONTENTS PAGE NO.

GE Fanuc IC695ALG600. Rx3i PacSystem

LS3300 AC Power Calibrator Simple Adjustment Procedure for Voltage, Current, and Frequency

Roline L1 Series. Humidity - Temperature Transmitters INSTRUCTION MANUAL

FC-33, DC SELECTABLE SIGNAL CONDITIONER

MEAS CAL ALARM CONF APT2000 TC

DIGITAL MULTIMETER CONTENTS DIGITAL MULTIMETER CONTENTS

PIECAL 311 Automated Universal RTD Calibrator Operating Instructions. Product Description. Practical Instrument Electronics

User s Manual. Model MXS Universal Computing Unit (1-input, Isolated 2-output Type) MODEL AND SUFFIX CODES

just below the screen. Data collection will begin, and a graph will show your data being plotted in real time.

Commissioning Instructions Rev. 03

Additel 875 Series Dry Well Calibrators

BU: EPBP GPG: DIN Rail Products Devices for the permanent control of insulation on supply lines for medical locations ISOLTESTER-DIG-RZ/RS/PLUS

Type PLC Fitting. Valve for Continuous control. Diaphragm valve. Output

SET Installation and Operating Instructions. Level switch for two sensors

- HD HD HD pag. 338 Portable ph meters. - HD HD pag. 342 Portable ph and conductivity meters

Sensor Troubleshooting Application Note

1800 Series Attachable Loop-Powered Digital Indicators Operating Manual

ph / ORP Transmitter / Controller PH2000/ PH3000/ PH5000/ PH ph / ORP Operation Manual

TxBlock-USB Transmitter

B850 Boiler House Energy Monitor

AMM-1022 Digital Multimeter USER`S MANUAL

Computer-14d - xx - 144a

CL900. True RMS 1000V 2000A 60MΩ ENGLISH. INSTRUCTION MANUAL 2000A Digital Clamp Meter. Measurement Technology

2-Wire Toroidal Conductivity Transmitters Series APT2000 Specifications

Bedienungsanleitung Operating manual Mode d emploi Instrucciones de operación

RMO500 M I C R O O H M M E T E R

SC200 Series. Signal Conditioner MNX10020, REV H 10/17/ Connection Technology Center, Inc Rae Boulevard Victor, NY (585)

Please enter the identity code of your device here!

phix Compact 4.10 PH, TEMPERATURE AND REDOX MEASUREMENT DATASHEET EN 4.10 PHIX DATASHEET 1812

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL:

Earth Leakage Monitoring System IsoBase IsoHub IsoOut

269 Troubleshooting Guide 11/4/98 1

Operating Instructions CAMILLE BAUER METRAWATT METRAOHM /9.97. Low impedance resistance meas. instrument

R24 DISPLAY MODULE WITH ACTIVE OUTPUT AND RELAY

Paperless Recorder PR20

DVM98. True RMS Digital Multimeter. 1 Safety information. 1.1 Preliminary. 1.2 During use

CMM System - Encapsulated Modules

Transcription:

1000TR ph Instructions

CONTENTS 1. INTRODUCTION... 2 1.1 COMMON INTRODUCTION... 2 1.2 PARTS & ACCESSORIES... 2 2. INSTALLATION... 3 2.1 CASING... 3 2.2 MOUNTING... 3 2.3 ELECTRICAL INSTALLATION... 3 2.3.1 Connection of the power supply... 4 2.3.2 Connection of the electrode... 4 2.3.3 Connection of a temperature probe ( RTD Pt100 )... 4 2.3.4 Connection of a recorder... 4 2.4 CHECK THE CONNECTIONS... 4 3. FUNCTIONS... 5 3.1 COMMON FUNCTIONS... 5 3.2... 5 3.2.1 MEASURE... 5 3.2.2 CAL. 1... 5 3.2.3 CAL. 2... 5 3.2.4 CAL. 3... 5 3.2.5 TEMP... 5 3.2.6 OUTPUT... 5 3.2.7 ALARM... 6 3.3 KEY FUNCTIONS... 6 4. PARAMETER SETTINGS... 7 4.1 DEFAULT SETTINGS... 7 4.2 PARAMETER SETTINGS... 7 4.2.1 Setting of the temperature compensation... 7 4.2.2 Set the output signal... 7 4.3 CALIBRATION... 8 4.4 OPERATION... 10 4.5 CHANGING THE PARAMETER SETTINGS... 10 4.5.1 Change the temperature compensation... 10 4.5.2 Change the output range... 11 5. MAINTENANCE... 12 5.1 HARDWARE CHECK... 12 5.2 PRE-CALIBRATION... 12 5.3 RESET OF DEFAULT SETTINGS... 13 6. TROUBLE SHOOTING... 14 7. SPECIFICATIONS... 15-1 -

1. INTRODUCTION 1.1 COMMON INTRODUCTION 1000TR is a new CE-approved series of transmitters from Elmacron AB. In the series, there are instruments for ph, Redox%, Redox mv and temperature measurement. 1000TR is easy to program, calibrate and use. The layout on the instrument is clear; a large display with four LED-segments, diode indication for the menu and four large function keys. All settings are made through the function keys on the front panel. When connecting an electrode to 1000TR-pH the display shows the actual ph-value in the solution measured. Manual / Automatic temperature compensation is built-in to give the highest possible accuracy over the whole range. 1000TR-pH is provided with an isolated output 0/4-20 ma, proportional to the ph-value, for connection to, for example, a computer, printer or other recording equipment. When measuring is interrupted, the output stays frozen at the last measured value. 1.2 PARTS & ACCESSORIES 1000TR is delivered without connection cables and electrodes. Article Function Code number Power supply cord, standard Power supply 20-R037-000 Low noise coax cable with BNC-connection Connection of electrode/probe 60-T810-0xx* ( xx* = 01, 02, 03, 05, 07 or 10 meter ) Pt100-probe, glass. Type 2771001-16T2V12 Automatic temperature compensation 60-K118-000 ph-buffer 7.00, 500 ml Calibration of ph-electrode 90-T514-000 ph-buffer 4.00, 500 ml Calibration of ph-electrode 90-T513-000 ph-buffer 10.00, 500 ml Calibration of ph-electrode 90-T516-000 - 2 -

2. INSTALLATION 2.1 CASING The casing is made of die-cast polystyrol with a transparent front cover that is closed with a snaplock. The connection terminals are located in a separate partition at the lower part of the transmitter. At the bottom of the instrument there are four cable glands ( two 15,2 and two 18,6 ) for the electrical connections. The protection class is IP65. 2.2 MOUNTING The instrument is designed for wall mounting. Be sure that the instrument is mounted at a nonvibrating place. 1000TR is mounted vertically with two fastening screws ( 7) through the mounting brackets on the instrument. 2.3 ELECTRICAL INSTALLATION It is recommended that each instrument is provided with a separate power switch. The electrode coax cable must be protected by a screen and may not be installed near power cables. Avoid extension of cables. The connection terminals are located in a separate space at the lower part of the transmitter. - 3 -

2.3.1 Connection of the power supply Connect the power supply to terminal 1 (earth), terminal 2 (zero) and terminal 3 (live). 2.3.2 Connection of the electrode Electrode cabling is a critical part of the whole system. Use a low noise coax cable between the electrode and the input terminal on the controller. Low noise cables has, in general, a black semi-conductive sheath between the centrewire insulation and the shield. Remove this sheath. Keep the cable away from power wires. Max recommended length of cable is 10 meters. Connect the shield ( reference ) of the coax cable to terminal 4 ( EL - ). Connect the centre ( glass electrode ) of the coax cable to terminal 5 ( EL + ). 2.3.3 Connection of a temperature probe ( RTD Pt100 ) When automatic temperature compensation is configured, a Pt100-probe is connected to terminal 6, 7, 8 and 9 ( 1A, 1B, 1C and 1D ). 2.3.3.1 Connection of 4-wire Pt100 2.3.3.2 Connection of 2-wire Pt100 The temperature sensor has to be installed in the same solution being measured, close to the electrode either in-line or in the tank. A pre-set calibration has to be done when installing a new Pt100- sensor ( see chapter 5.2 ). When manual temperature compensation is configured a 107 Ω-resistor is connected to the temperature input and a jumper is set between terminals as shown in the figure below. 2.3.4 Connection of a recorder Connect the recorder to terminal 10 ( REC + ) and terminal 11 ( REC - ). 2.4 CHECK THE CONNECTIONS Before the power supply is turned on; check that the connections are mechanical and electrical correct. - 4 -

3. FUNCTIONS 3.1 COMMON FUNCTIONS All settings are made through the function keys on the front panel. The keys + and are accelerating in three steps when they are kept pressed. Chosen function is indicated by a green diode in the menu, at alarm a red diode is activated at the same time as the diode at the actual function activates. When pressing M you step down in the menu. When calibration or setting mode is entered the output signal stays "frozen" at the latest value, until = is pressed. 3.2 MEASURE CAL. 1 CAL. 2 CAL. 3 TEMP. Offset High-buffer Low-buffer Aut./Man. 0/4-20 ma OUTPUT ALARM 3.2.1 MEASURE When power supply is connected the measuring starts.. 3.2.2 CAL. 1 Calibration point 1. Pre-set value: ph 7.00, other values can be chosen in the calibration mode. 3.2.3 CAL. 2 Calibration point 2. Pre-set value: ph 10.00, other values > ph 7.00 can be chosen in the calibration mode. 3.2.4 CAL. 3 Calibration point 3. Pre-set value: ph 4.00, other values < ph 7.00 can be chosen in the calibration mode. 3.2.5 TEMP. The instrument is provided with a choice between automatic and manual temperature compensation. The values shown when automatic temperature compensation is activated, the value is corresponding to the temperature 20 C. In manual compensation mode, the temperature is set by the operator. 3.2.6 OUTPUT Can be set to 0-20 ma or 4-20 ma. The function is scaleable. - 5 -

3.2.7 ALARM When an error occurs in the measuring mode, the alarm diode activates at the same time as the error code is shown on the display. At serious errors, the error code is flashing on the display. 3.3 KEY FUNCTIONS KEY FUNCTION 1 ( at parameter setting mode ) FUNCTION 2 ( in measuring mode ) Steps down the menu Accelerating key. Increase the value in setting mode. Used when making a choice according to the display. Interrupt measuring together with Accelerating key. Decrease the value in setting mode. Interrupt measuring together with Confirm performed settings and choices. Starts measuring. - 6 -

4. PARAMETER SETTINGS 4.1 DEFAULT SETTINGS When delivered, 1000TR have the following default settings. Output 0-20 ma Temperature compensation Manual, 20 C Offset ph = 7.00 Low-buffer ph = 4.00 High-buffer ph = 10.00 Slope 100% ( 20 ) 4.2 PARAMETER SETTINGS All settings has to be confirmed with = to be saved. To leave the setting mode without saving the settings: press M. Before starting up the transmitter the following parameters has to be set: Temperature compensation and output range. Turn the power on. 4.2.1 Setting of the temperature compensation 4.2.1a) Manual temperature compensation For automatic temperature compensation, see 4.2.1.b). 1. Interrupt the measuring by pressing + and at the same time until the measuring is interrupted and the diode at CAL.1 is activated. 2. Press M until the diode at TEMP. is activated. At the display: In--. Press = to confirm. 3. Set the correct temperature with + and -. Press = to confirm. The diode at OUTPUT activates and the display shows 0-20. 4.2.1b) Automatic temperature compensation In automatic temperature compensation a Pt100 sensor has to be installed ( see chapter 2.3 for connection ). NOTE! Save the 107Ω resistor. A pre-calibration according to 5.2 has to be performed when a new Pt100 sensor is installed. 1. Interrupt the measuring by pressing + and at the same time until the measuring is interrupted and the diode at CAL.1 is activated. 2. Press M until the diode at TEMP. is activated. At the display: In--. 3. Press + until the display shows Au--. Press = to confirm. The diode at OUTPUT activates and the display shows 0-20. 4.2.2 Set the output signal 1. Choose between 0-20 and 4-20 ma with +. 2. Press = to confirm. 3. The display shows SPAn. Press = to confirm. 4. The display shows 0000. Set the ph-value corresponding to 0 (alt. 4) ma. 5. Press = to confirm. The display shows 0014. - 7 -

6. Set the ph-value corresponding to 20 ma. Press = to confirm. 7. The diode at MEASURE is activated and the display shows Hold. 8. Press = to start the measurement. NOTE! The ph-values have to be set so that the 0/4 ma-level is corresponding to a lower value than the 20 ma value. In other cases the output signal stays at 0 ma. 4.3 CALIBRATION When delivered the transmitter is calibrated according to standard: offset ( zero point ) at ph 7.00, and slope 100% at 20 C. The transmitter should be calibrated on a regular basis according to ph, to reach a high security in the measurement. Calibration can be performed at 1, 2 or 3 ( recommended ) calibration points. The transmitter is preset for calibration at ph 7.00 ( offset ), ph 4.00 (low buffer) and ph 10.00 (high buffer), other values can be chosen. If calibration at two points is desired, it is recommended that the calibration points are chosen at the "zero-point" and at a point close to the expected ph. During the calibration, the latest output signal stays "frozen". When a new electrode is installed or at the first start-up of the instrument, we advise you to do as follows: Check that the electrode glass membrane is stored wet. If the protective cap is empty and the electrode is dry, immerse the electrode in a buffer solution or tap water ( do not use destilled water ) for three hours before calibration ( and using ). If the transmitter is set for manual temperature compensation, the temperature of the buffer solution has to be set. ( see chapter 5.1.1 ). If the transmitter is set for automatic temperature compensation, the Pt100-sensor has to be immersed in the buffer solution together with the electrode. 1. Interrupt the measuring by pressing + and at the same time. 2. The instrument automatically steps down to CAL. 1 ( calibration point 1). Pre-set zeropoint is at ph 7.00, if another offset value is desired it can be set with + and -. 3. Rinse the electrode membrane carefully with destilled water. 4. Put the electrode in buffer 1. Press = to start. Wait until the value at the display has stabilised. Use + and to set the correct value. Press = to save the calibration. If the measured value differs more than ± 1.5 ph-units* from the pre-set buffer value, calibration can not be performed. When pressing = a message is shown, ( see chapter 6 ). Locate and correct the error and try again. Press M to interrupt the calibration. 5. The instrument automatically steps to CAL. 2 (calibration point 2). At calibration point 2, the instrument is calibrated at a high-buffer value ( ph > 7 ). If no calibration at point 2 is desired, the point can be ignored by pressing M until the diode at CAL.3 is activated. 6. The pre-set value on buffer 2 is ph 10.00, if another offset value is desired it can be set with + and -. 7. Rinse the electrode membrane carefully with destilled water. - 8 -

8. Put the electrode in buffer 2. Press = to start. Wait until the value at the display has stabilised. Use + and to set the correct value. Press = to save the calibration. If the phslope differs more than +10%, -20%* the calibration is not accepted and the value can not be adjusted. If the ph-slope differs more than +10%, -20%* the alarm diode is activated and the calibration can not be done, when pressing = an error message is shown, (see chapter 6). Accept the error by pressing =. Locate and correct the error and try again. Press M to interrupt the calibration. 9. The instrument automatically steps to CAL. 3 (calibration point 3). At calibration point 3 the instrument is calibrated at a low-buffer value ( ph < 7 ). If no calibration at point 3 is desired, the point can be ignored by pressing M. 10. The pre-set value on buffer 3 is ph 4.00, if another offset value is desired it can be set with + and -. 11. Rinse the electrode membrane carefully with destilled water. 12. Put the electrode in buffer 3. Press = to start. Wait until the value at the display has stabilised. Use + and to set the correct value. Press = to save the calibration. If the phslope differs more than +10%, -20%* the calibration is not accepted and the value can not be adjusted. If the ph-slope differs more than +10%, -20%* the alarm diode is activated and the calibration can not be done, when pressing = an error message is shown, (see chapter 6). Accept the error by pressing =. Locate and correct the error and try again. Press M to interrupt the calibration. 13. Step to MEASURE with M. 14. Rinse the electrode membrane carefully with destilled water. 15. Do not forget to reset to the process temperature when manual temperature compensation is chosen. 16. The instrument is now ready for measuring. Press M to start. The accepted range is calculated in relation to the distance from the offset value. Example: Offset ( zero ) ph 7.00 Calibration point 1 ph 10.00 Calibration point 2 ph 4.00 Accepted ph-slope is +10%, -20% from ideal slope. The deviation is calculated according to: ph 10.00 = 3, which gives max deviation = +10 % of 3 = 0,3 ph-units, -20 % of 3 = -0,6 ph-units. ph 4.00 = -3, which gives max deviation = +10 % of -3 = -0,3 ph-units, -20 % of -3 = 0,6 ph-units. The accepted intervals in the example: calibration point 1 = ph 9,4 - ph 10,3 calibration point 2 = ph 3,7 - ph 4,6 Accepted limits for the offset value is ph 5,5 - ph 8,5-9 -

ph 11 10 9 8 1,5 ph 11 10 9 8 0,3 0,6 7 6 5 4 3 1,5 0,6 0,3 7 6 5 4 3 4.4 OPERATION Measuring is started when the power supply is turned on. After an interruption off the measuring it can be restarted by pressing = when the MEASURE diode is activated. The display shows the actual ph-value in the solution measured at the same time as the corresponding ma-signal is sent to the output. During the operation, the diode at MEASURE is activated. 4.5 CHANGING THE PARAMETER SETTINGS All settings has to be confirmed with = to be saved. To leave the setting mode without saving the settings: press M. 4.5.1 Change the temperature compensation Manual temperature compensation 1. When manual temperature compensation is configured a 107 Ω-resistor is connected to the temperature input and a jumper is set between terminals as shown in figure below. 2. Interrupt the measuring mode by pressing + and at the same time until the measurement is interrupted and the diode at CAL.1 is activated. 3. Press M until the diode at TEMP. is activated. 4. Press + until the display shows In-- Press = to confirm. 5. Set the correct temperature with + and -. Press = to confirm. The diode at OUTPUT is activated and the display shows 0-20. 6. Step to MEASURE with M. Hold is shown on display, start measuring by pressing =. Automatic temperature compensation - 10 -

1. Connect a Pt100-sensor to the instrument ( see chapter 2.3 for connection instructions). Make a pre-calibration according to chapter 5.2. 2. Interrupt the measuring mode by pressing + and at the same time until the measurement is interrupted and the diode at CAL.1 is activated. 3. Press M until the diode at TEMP. is activated. 4. Press + until the display shows Au--. Press = to confirm. 5. The diode at OUTPUT is activated and the display shows 0-20. 6. Step to MEASURE with M. Hold is shown on display, start measuring by pressing =. 4.5.2 Change the output range 1. Interrupt the measuring mode by pressing + and at the same time until the measurement is interrupted and the diode at CAL.1 is activated. 2. Step to OUTPUT with M. 3. Choose between 0-20 and 4-20 ma with +. 4. Press = to confirm. 5. The display shows SPAn. Press = to confirm. 6. The display shows 0000. Set the ph-value corresponding to 0 (alt. 4) ma. 7. Press = to confirm. The display shows 0014. 8. Set the ph-value corresponding to 20 ma. Press = to confirm. 9. The diode at MEASURE is activated and the display shows Hold. 10. Press = to start the measurement. NOTE! The ph-values have to be set so that the 0/4 ma-level is corresponding to a lower value than the 20 ma value. In other cases the output signal stays at 0 ma. - 11 -

5. MAINTENANCE 5.1 HARDWARE CHECK A check of the hardware is necessary only if you suspect that there is something wrong with your transmitter. 1. Disconnect the power supply. 2. Short-circuit the electrode input. 3. Connect a ma-meter to the REC-output. 4. Press M and connect the power supply, keep M pressed during 5 seconds until the four upper diodes are activated. 5. The transmitter performs a self-test according to table 5.1.1 5.1.1 Hardware check CH01 Test of LED on display All 4 LED-segments are activated CH02 Test of function keys Press the keys one by one and the corresponding sign is shown on the display CH03 Test of output signal Press =. 20.00 is shown on the display at the same time as the corresponding signal is sent to the output. Press =. 04.00 is shown on the display at the same time as the corresponding signal is sent to the output. Press = CH04 test of diodes All the menu diodes are activated The version number is shown, then the instrument starts to measure. 5.2 PRE-CALIBRATION The transmitter is always pre-calibrated when delivered. Pre-calibration is only necessary when the instrument has been shut of for a longer period or when a new Pt100-sensor is connected. 1. Turn the power off. Short-circuit the electrode input. Connect a 100 Ω-resistor to the Pt100-input terminal clamps. Make a connection between the terminals according to the figure below or, when a new Pt100-sensor is connected, put the sensor in icy water ( 0 C ). - 12 -

2. Press and turn on the power supply. Keep the key pressed until the display shows CAL1. (10 seconds). Every second diode is activated. CAL1. 3. Press = to continue. 4. On the display E0.00 during 5 s. 5. On the display CAL2. 6. On the display P0.0 during 5 s. 7. On the display CAL3. 8. On the display 0.01A CAL2.. Press = to continue. CAL3. Press = to continue. 0.01A. Press + or - to receive 10.0 ma on the ma-instrument. NOTE! You have to press + or - repeatedly to get a change in the signal, 9. Press = to continue. The display shows the version number during 5 sec. 10. The instrument returns to measuring mode, the measuring starts automatically. 11. If manual temperature compensation is chosen the 100 Ω-resistor is replaced with a 107 Ω-resistor. 5.3 RESET OF DEFAULT SETTINGS The settings are reset according to chapter 4.1. 1. Turn off the power supply. 2. Press + and turn on the power supply. Keep the key pressed until the display shows C l r O (10 seconds). The four lower diodes are activated. 3. Press = to continue. 4. The display shows the version number and then the measuring starts up. - 13 -

6. TROUBLE SHOOTING CODE INDICATES PROBABLE CAUSE CORRECTION The temperature in the The temperature in the measured Correct the temperature E-01* measured solution is to low. solution is below 0 C Broken Pt100-sensor Replace the sensor E-02* The temperature in the measured solution is too high. Damage at the connections of the Pt100-sensor The temperature in the measured solution is above 100 C Check the connections Correct the temperature Broken Pt100-sensor Replace the sensor E-03 Buffer 1 measured value > + 1.5 ph the measured value exceeds the upper interval limit E-04 Buffer 1 measured value < - 1.5 ph or Or the measured value exceeds the lower interval limit Damage at the connections of the Pt100-sensor Faulty buffer solution Old/bad electrode Faulty buffer solution Old/bad electrode Check the connections Check the buffer solution Replace the electrode Check the buffer solution Replace the electrode * If the error message appears during normal operation it is probably the Pt100-sensor that is broken, the fault has to be corrected before the measuring can be continued. To continue the measurement with manual temperature compensation until a new Pt100-sensor is connected: Press = to accept the error message. Set the temperature compensation to the manual mode ( see chapter 4.2.3 A ). - 14 -

7. SPECIFICATIONS Instrument Version Dimensions 186x131x103 Weight Ca 1,0 kg Max length, electrode cable 10 meters Display Four 7-segment LED Keyboard 4 push buttons Connections Terminals Power supply 230 VAC, 50 Hz Backup > 10 years Range, ph 0.00-14.00 ph Accuracy, ph ± 0.01 ph Resolution, ph ± 0.01 ph Offset Range ± 1.5 ph, steps 0.01 ph Slope +10%, -20% Calibration 1, 2 or 3 points Temp. range 0-100 C Temperature compensation manual automatic with a connected 4-wire Pt100-sensor Input ph-inp z = 10 12 Ω Outputs Output 0-20 ma / 4-20 ma Max load, output 700 Ω - 15 -