PXIe, 24-Channel, ±24 V, 50 ma Precision PXI Source Measure Unit

Similar documents
4-Channel, ±24 V, 150 ma PXI Source Measure Unit

4-Channel ±10 V, 100 ma, Precision PXI Source Measure Unit

4-Channel, ±24 V, 150 ma Precision PXI Source Measure Unit

4-Channel ±10 V, 100 ma, Precision PXI Source Measure Unit

4-Channel, ±6 V, 500 ma Precision PXI Source Measure Unit

PXI Contents. Definitions SPECIFICATIONS. ±100 V, 2 W Precision PXI Source Measure Unit. These specifications apply to the PXI-4132.

if it is used in a manner not described in the user documentation.

PXIe Contents SPECIFICATIONS. ±60 V, 3 A System PXI Source Measure Unit. These specifications apply to the PXIe-4138.

is used in a manner not described in the user documentation.

PXIe Contents SPECIFICATIONS. ±200 V, 1 A System PXI Source Measure Unit. These specifications apply to the PXIe-4136.

PXIe Contents SPECIFICATIONS. ±60 V, 3 A Precision System PXI Source Measure Unit. These specifications apply to the PXIe-4139.

PXIe Contents SPECIFICATIONS. ±200 V, 1 A Precision System PXI Source Measure Unit. These specifications apply to the PXIe-4137.

PXI Contents SPECIFICATIONS. ±20 V, 40 W PXI Source Measure Unit

Note Using the PXIe-5785 in a manner not described in this document might impair the protection the PXIe-5785 provides.

PXIe Contents SPECIFICATIONS. Precision, 10 fa Sensitivity Low-Current PXI Source Measure Unit. These specifications apply to the PXIe-4135.

NI PXIe Contents CALIBRATION PROCEDURE. Four Channel Precision Source-Measure Unit (SMU)

Current ON state current ma minimum, 6 ma typical, 10 ma maximum OFF state current ma maximum

PXIe Contents SPECIFICATIONS. 2-Channel, 60 V, 1 A PXI Programmable Power Supply. These specifications apply to the PXIe-4112.

NI PXIe Contents CALIBRATION PROCEDURE. Single-Channel Precision Source-Measure Unit (SMU)

NI PXIe Contents. Required Software CALIBRATION PROCEDURE. Dual-Output Programmable DC Power Supply

PXIe Contents SPECIFICATIONS. RF Amplifier

NI PXI-2521 Specifications 40-Channel DPST Relay Module

NI PXI-2520 Specifications 80-Channel SPST Relay Module

PXIe Contents SPECIFICATIONS. RF Preselector Module

PXIe Contents SPECIFICATIONS. 100 MVector/s PXI Digital Pattern Instrument

PXIe Contents SPECIFICATIONS. 3.3 GHz RF Analog Signal Generator

NI PXI Contents DEVICE SPECIFICATIONS. RF Attenuator

PXI Contents SPECIFICATIONS. 1.3 GHz RF Analog Signal Generator

PXIe Contents SPECIFICATIONS. 32-Channel Digital Pattern Instrument

PXIe Contents SPECIFICATIONS. 250 khz to 20 GHz RF Analog Signal Generator

PXI Contents SPECIFICATIONS. 6.6 GHz RF Analog Signal Generator

NI PXIe Contents. About These Specifications DEVICE SPECIFICATIONS. 3 GHz Ω Multiplexer (SP16T)

Input Characteristics. Measurement Range. SPECIFICATIONS NI PXIe-4357

PXIe Contents. Definitions SPECIFICATIONS. 8.5 GHz Vector Network Analyzer

PXIe Contents. Required Software CALIBRATION PROCEDURE

1- or 2-Port, High-Speed, Flexible Data Rate PXI CAN Interface Module

DEVICE SPECIFICATIONS NI PXIe-4302/ Ch, 24-bit, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module

These specifications apply to the PXIe-5113 with 64 MB and 512 MB of memory.

Power Amplifier. Input Characteristics. SPECIFICATIONS NI PXIe-4610 Power Amplifier. Signal Range

PXIe Contents. Definitions SPECIFICATIONS. 8.5 GHz Vector Network Analyzer

DEVICE SPECIFICATIONS NI PXIe Ch, 24-bit, 25.6 ks/s Simultaneous AC LVDT Input Module

Analog Characteristics

NI PXI-2797 Specifications 40 GHz 6 1 Terminated 50 Ω Multiplexer

NI PXI/PXIe-2540 Specifications

NI PXI/PXIe-2529 Specifications

Contents. Software Requirements

1- or 2-Port, Software-Selectable PXI CAN Interface Module

NI PXI/PXIe-2541 Specifications 300 MHz Ω Matrix

Analog Input. DEVICE SPECIFICATIONS NI PXIe-4300

NI PXIe-5630 Specifications

PXIe, 6½-Digit, ±300 V, Onboard 1.8 MS/s Isolated Digitizer,PXI Digital Multimeter

NI PXIe-5601 Specifications

NI PXI-2557 Specifications

NI PXI/PXIe Contents. About These Specifications DEVICE SPECIFICATIONS. Quad 2 1 Terminated Multiplexer

Refer to the NI Switches Help at ni.com/manuals for detailed topology information.

PXI Contents. PXI-2585 Specifications. Topology DEVICE SPECIFICATIONS. 10-Channel Multiplexer

NI 6143 Specifications

DEVICE SPECIFICATIONS NI PXIe-4304/ Ch, 24-bit, ±42 V, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module. ni.

NI 2865A 0.3 A Matrix Cards for NI SwitchBlock

SCXI MHz Quad 8x1 50 Ω Multiplexer

NI PXI Contents. About These Specifications DEVICE SPECIFICATIONS GHz Dual Transfer Switch 50 Ω Relay Module

CALIBRATION PROCEDURE PXIe-4302/4303 and TB-4302C 32 Ch, 24-bit, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module. ni.

PXIe, 7½-Digit, ±1,000 V, Onboard 1.8 MS/s Isolated Digitizer, PXI Digital Multimeter

6½-Digit, ±300 V, 1.8 MS/s Isolated Digitizer Included, PCI Digital Multimeter Device

NI PXI/PXIe-2527 Specifications

NI PXI-2555 Specifications

PCI Express, 6½-Digit, ±300 V Digital Multimeter Device

100 MHz Bandwidth, 12-Bit PXI Waveform Generator

NI PXI-2530 Specifications

Contents. CALIBRATION PROCEDURE NI PXIe-5698

This document lists the specifications for the NI PXI The PXI-5900 is a differential amplifier.

SPECIFICATIONS PXIe channel, 16-bit, 400 ks/s/ch, Ch-Ch Isolated Analog Input Module. ni.com/manuals

Specifications for the NI PXI/PCI-6552/6551

SPECIFICATIONS PXIe Ch (8 ADC), 2 MS/s, bit, Flexible Resolution PXI Analog Input Module

NI PXI-4461 Specifications

GETTING STARTED GUIDE NI khz to 100 MHz, 50 Ω, AC Coupled RF Receiver Module

NI PXI/PXIe-2532 Specifications

NI PXI/PXIe Contents DEVICE SPECIFICATIONS. 40 A Fault Insertion Unit

NI PXI/PXIe Contents DEVICE SPECIFICATIONS. 10 A Fault Insertion Unit

Contents. Software. Documentation. CALIBRATION PROCEDURE NI PXIe-6674T

CDA Contents. Definitions SPECIFICATIONS. Clock Distribution Device

PXI Contents. PXI-2584 Specifications DEVICE SPECIFICATIONS. 300 V CAT II, 600 V CAT I Multiplexers

NI PXI-2532 Specifications

USRP Contents. Definitions. Conditions SPECIFICATIONS. Software Defined Radio Device

USRP Contents. Definitions. Conditions SPECIFICATIONS. Software Defined Radio Device

80 MHz Bandwidth, 16-Bit PXI Waveform Generator. These specifications apply to the one-channel and two-channel PXIe-5433.

NI PXIe-5663 Specifications

GETTING STARTED GUIDE NI AI, ±5 V, 24 Bit, 51.2 ks/s/ch Simultaneous, AC/DC Coupling, IEPE AC Coupling

8-Channel, 12-Bit, 60 MHz PXI Express Oscilloscope

Terminology. Output Characteristics. DEVICE SPECIFICATIONS NI PXIe-4463 DSA Analog Output. ni.com/manuals

GETTING STARTED GUIDE NI AI, ±10 V, 24 Bit, 50 ks/s/ch Simultaneous

DEVICE SPECIFICATIONS Phase Matrix PXI GHz PXI Vector Signal Analyzer

USRP Contents. Definitions. Conditions SPECIFICATIONS. Software Defined Radio Reconfigurable Device

USRP Contents. Definitions SPECIFICATIONS. Software Defined Radio Reconfigurable Device

NI PXIe-5171R. Contents. Required Software CALIBRATION PROCEDURE

PXIe, 40 MHz Bandwidth, 16-Bit PXI Waveform Generator. These specifications apply to the one-channel and two-channel PXIe-5423.

High-Speed/Flexible Data-Rate CAN Characteristics 5 VDC

100-SPST (latching), 50-DPST. Refer to the NI Switches Help at ni.com/manuals for detailed topology information.

USRP Contents. Definitions SPECIFICATIONS. 50 MHz to 2.2 GHz Tunable RF Transceiver

PXI Contents SPECIFICATIONS. 6½-Digit, ±300 V, Onboard 1.8 MS/s Isolated Digitizer, L and C Measurement Support, PXI Digital Multimeter

Transcription:

SPECIFICATIONS PXIe-4163 PXIe, 24-Channel, ±24 V, 50 ma Precision PXI Source Measure Unit Contents Definitions...1 Conditions... 2 Instrument Capabilities... 2 SMU Specifications... 3 Voltage...3 Current... 4 DC Power...4 Additional Specifications...4 Supplemental Specifications... 6 Measurement and Update Timing...6 Triggers... 7 Calibration Interval... 8 Physical... 8 Environment...8 Operating Environment...8 Storage Environment...9 Shock and Vibration...9 Definitions Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty. The following characteristic specifications describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty. Typical specifications describe the performance met by a majority of models. Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing. Specifications are Nominal unless otherwise noted.

Conditions Specifications are valid under the following conditions unless otherwise noted. Ambient temperature 1 of 23 C ± 5 ºC Calibration interval of 1 year 30 minutes warm-up time Self-calibration performed within the last 24 hours nidcpower Aperture Time property or NIDCPOWER_ATTR_APERTURE_TIME attribute set to 2 power-line cycles (PLC) Fans set to the highest setting if the PXI Express chassis has multiple fan speed settings Instrument Capabilities The following table and figure illustrate the voltage and the current source and sink ranges of the PXIe-4163. Table 1. PXIe-4163 Current Source and Sink Ranges, Warranted Channels DC Voltage Ranges DC Current Source and Sink Ranges With a PXIe-1095 Chassis 0 through 23 ±24 V 10 μa 100 μa 1 ma 10 ma 50 ma With All Other Chassis 10 μa 100 μa 1 ma 10 ma 30 ma 1 The ambient temperature of a PXI system is defined as the temperature at the chassis fan inlet (air intake). 2 ni.com PXIe-4163 Specifications

Figure 1. PXIe-4163 Quadrant Diagram, Any Channel 50 40 1 ma/v Quadrant II Quadrant I 30 20 Current (ma) 10 0 10 20 30 1 ma/v 40 Quadrant III Quadrant IV 50 24 20 15 10 5 0 5 10 15 20 24 Voltage (V) Valid on any channel in a PXIe-1095 chassis. Valid on any channel in all other chassis. 1 1 Maximum 480 ma per module. SMU Specifications Voltage Table 2. Voltage Programming and Measurement Accuracy/Resolution, Warranted Range Resolution and Noise (0.1 Hz to 10 Hz) Accuracy (23 C ± 5 C) ± (% of Voltage + Offset) Tempco 2 ± (% of Voltage + Offset)/ C, 0 C to 55 C T cal ± 5 C 24 V 200 μv 0.05% + 5 mv 0.0005% + 1 µv 2 Temperature coefficient applies beyond 23 C ± 5 C within 5 C of Tcal. PXIe-4163 Specifications National Instruments 3

Current Table 3. Current Programming and Measurement Accuracy/Resolution, Warranted Range Resolution and Noise (0.1 Hz to 10 Hz) Accuracy (23 C ± 5 C) ± (% of Current + Offset) Tempco 3 ± (% of Current + Offset)/ C, 0 C to 55 C T cal ± 5 C 10 μa 100 pa 0.10% + 5 na 0.004% + 10 pa 100 μa 1 na 0.10% + 50 na 0.004% + 100 pa 1 ma 10 na 0.10% + 500 na 0.004% + 1 na 10 ma 100 na 0.10% + 5 μa 0.004% + 10 na 30 ma 500 na 0.10% + 25 μa 0.004% + 50 na 50 ma 4 500 na 0.10% + 25 μa 0.004% + 50 na DC Power With a PXIe-1095 chassis Per channel Total With all other chassis Per channel Total 1.2 W 28.8 W 0.7 W 11.5 W Additional Specifications Settling time 5 <500 μs, typical 6 Transient response 7 <100 μs, typical 8 3 Temperature coefficient applies beyond 23 C ± 5 C within 5 C of Tcal. 4 Only available when used with a PXIe-1095 chassis. 5 Current limit set to 1 ma and 10% of the selected current limit range. PXIe-4163 configured for fast transient response. 6 To settle to 0.1% of voltage step. 7 PXIe-4163 configured for fast transient response. 8 To recover within ±20 mv after a load current change from 10% to 90% of range. 4 ni.com PXIe-4163 Specifications

Wideband source noise 9 Remote sense Voltage Current Maximum lead drop Load regulation Voltage 10 Current Functional isolation voltage, any pin to earth ground Absolute maximum voltage to Output LO From Sense HI 11 15 mv RMS, typical <100 mv pk-pk, typical No additional error due to lead drop No additional error due to lead drop 1 V drop/lead 50 μv/ma, typical (30 pa + 20 ppm of range)/volt, typical 60 VDC When V Output HI > 0 V -0.5 V to (V Output HI + 0.5 V) When V Output HI 0 V From all other pins (V Output HI - 0.5 V) to 0.5 V ±25 V The following figures illustrate noise as a function of measurement aperture for the PXIe-4163. Figure 2. Voltage RMS Noise Versus Aperture Time 12, Typical 10 m 1 m Local Sense Remote Sense RMS Noise (V) 100 µ 10 µ 1 µ 10 µ 100 µ 1 m 10 m Time (s) 100 m 9 20 Hz to 20 MHz bandwidth. PXIe-4163 configured for normal transient response. Measured at the end of the 1 m SHDB62M-DB62M-LL cable. 10 At connector pins when using local sense. 11 Where V Output HI is the voltage at the Output HI pin in the same channel as a Sense HI pin. 12 All channels averaged. Channels 9 and 22 have degraded performance. PXIe-4163 Specifications National Instruments 5

Figure 3. Current RMS Noise Versus Aperture Time 13, Typical 10 m 1 m 100 µ 10 µ 30 ma and 50mA 10 ma 1 ma 100 µa 10 µa 1 µ RMS Noise (A) 100 n 10 n 1 n 100 p 10 p 1 p 100 f 10 µ 100 µ 1 m 10 m Time (s) 100 m Supplemental Specifications Measurement and Update Timing Available sample rates 14 (600 ks/s)/n where N = 6, 7, 8, 2 20 and S is samples. Sample rate accuracy Maximum measure rate to host 15 ±50 ppm 100,000 S/s per channel, continuous 13 All channels averaged. Channels 7, 9, and 11 have degraded performance. 14 When source-measuring, both the NI-DCPower Source Delay and Aperture Time properties affect the sampling rate. When taking a measure record, only the Aperture Time property affects the sampling rate. 15 Load dependent settling time is not included. Normal DC noise rejection is used. 6 ni.com PXIe-4163 Specifications

Maximum source update rate 16 Single channel All channels simultaneously Input trigger to Source event delay Source event jitter Measure event jitter Triggers Input triggers Types Sources (PXI trigger lines 0 to 7) 17 Polarity Minimum pulse width Destinations 18 (PXI trigger lines 0 to 7) 17 Polarity Minimum pulse width Output triggers (events) Types Destinations (PXI trigger lines 0 to 7) 17 Polarity Pulse width 100,000 updates/s 40,000 updates/s per channel 8.5 μs 1.7 μs 1.7 μs Start Source Sequence Advance Measure Active high (not configurable) 100 ns Active high (not configurable) >200 ns Source Complete Sequence Iteration Complete Sequence Engine Done Measure Complete Active high (not configurable) 230 ns 16 As the source delay is adjusted or if advanced sequencing is used, maximum source update rates may vary. 17 Pulse widths and logic levels are compliant with PXI Express Hardware Specification Revision 1.0 ECN 1. 18 Input triggers can come from any source (PXI trigger or software trigger) and be exported to any PXI trigger line. This allows for easier multi-board synchronization regardless of the trigger source. PXIe-4163 Specifications National Instruments 7

Calibration Interval Recommended calibration interval 1 year Physical Dimensions Weight Front panel connectors 3U, one-slot, PXI Express/CompactPCI Express module 2.0 cm 13.0 cm 21.6 cm (0.8 in. 5.1 in. 8.5 in.) 394 g (13.9 oz) Custom 62-position D-SUB, female Environment Maximum altitude Pollution Degree 2 2,000 m (800 mbar) (at 25 C ambient temperature) Indoor use only. Operating Environment Ambient temperature range With a PXIe-1095 chassis With all other chassis Relative humidity range 0 C to 55 C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2. Meets MIL-PRF-28800F Class 3 low temperature limit and MIL-PRF-28800F Class 2 high temperature limit.) 0 C to 45 C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2. Meets MIL-PRF-28800F Class 3 low temperature limit and MIL-PRF-28800F Class 2 high temperature limit.) 10% to 90%, noncondensing (Tested in accordance with IEC 60068-2-56.) 8 ni.com PXIe-4163 Specifications

Storage Environment Ambient temperature range Relative humidity range -40 C to 71 C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2. Meets MIL-PRF-28800F Class 3 limits.) 5% to 95%, noncondensing (Tested in accordance with IEC 60068-2-56.) Shock and Vibration Operating shock Random vibration Operating Nonoperating 30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Meets MIL-PRF-28800F Class 2 limits.) 5 Hz to 500 Hz, 0.3 g rms (Tested in accordance with IEC 60068-2-64.) 5 Hz to 500 Hz, 2.4 g rms (Tested in accordance with IEC 60068-2-64. Test profile exceeds the requirements of MIL-PRF-28800F, Class 3.) PXIe-4163 Specifications National Instruments 9

Information is subject to change without notice. Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on NI trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance for the NI global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015. 2018 National Instruments. All rights reserved. 376735A-01 January 19, 2018