Combining Images for SNR improvement. Richard Crisp 04 February 2014

Similar documents
Everything you always wanted to know about flat-fielding but were afraid to ask*

Everything you always wanted to know about flat-fielding but were afraid to ask*

Residual Bulk Image Characterization using Photon Transfer Techniques

Simple RBI Shot Noise Measurement/Interpretation. Richard Crisp February 13, 2014

ASTROPHOTOGRAPHY (What is all the noise about?) Chris Woodhouse ARPS FRAS

Struggling with the SNR

Camera Selection Criteria. Richard Crisp May 25, 2011

WEBCAMS UNDER THE SPOTLIGHT

Lecture 30: Image Sensors (Cont) Computer Graphics and Imaging UC Berkeley CS184/284A

Cross-Talk in the ACS WFC Detectors. II: Using GAIN=2 to Minimize the Effect

What an Observational Astronomer needs to know!

BASLER A601f / A602f

Your Complete Astro Photography Solution

Amplifier Luminescence and RBI. Richard Crisp May 21,

Astrophotography. An intro to night sky photography

Hewett 1 Imaged by Amateur

A Test of non-standard Gain Settings for the NICMOS Detectors

Camera Test Protocol. Introduction TABLE OF CONTENTS. Camera Test Protocol Technical Note Technical Note

Astronomy 341 Fall 2012 Observational Astronomy Haverford College. CCD Terminology

Errata to First Printing 1 2nd Edition of of The Handbook of Astronomical Image Processing

Image Processing for Comets

Figure 1 HDR image fusion example

Shooting Sky Flats in the Daytime. Richard Crisp 12/24/2009

Temperature Dependent Dark Reference Files: Linear Dark and Amplifier Glow Components

Image Processing Tutorial Basic Concepts

CCD reductions techniques

The DSI for Autostar Suite

Step 5) Split the red data using the Multi Scale Decomposition tool into a detail and residual background image.

Noise Analysis of AHR Spectrometer Author: Andrew Xiang

OPTOLONG L Pro pollution filter testing

Pixel Response Effects on CCD Camera Gain Calibration

Observing*Checklist:*A3ernoon*

CCD Characteristics Lab

Astrophotography. Playing with your digital SLR camera in the dark

MAOP-702. CCD 47 Characterization

AstraLux SNR and DR considerations

The Noise about Noise

Scientific Image Processing System Photometry tool

CCD vs CMOS for Video Astronomy by Jim Thompson, P.Eng Test Report November 20 th, 2017

Victoria RASCals Star Party 2003 David Lee

This particular case study is an experimental trial

Name: Date: Period: Activity 4.6.2: Point-Slope Form of an Equation. 0, 4 and moving to another point on the line using the slope.

Basler aca km. Camera Specification. Measurement protocol using the EMVA Standard 1288 Document Number: BD Version: 03

Setting GAIN and OFFSET on cold CMOS camera for deep sky astrophotography

Frame Calibration* CCD, Video & DSLR. * Also known as reduction

Basler aca gm. Camera Specification. Measurement protocol using the EMVA Standard 1288 Document Number: BD Version: 01

Basler aca640-90gm. Camera Specification. Measurement protocol using the EMVA Standard 1288 Document Number: BD Version: 02

I was not able to test the configuration of UHC + MC IR cut with the UHC on the camera side of the IR as my UHC is 2 and the MC IR is 1.25.

Basler ral km. Camera Specification. Measurement protocol using the EMVA Standard 1288 Document Number: BD Version: 01

Chasing Faint Objects

Noise and ISO. CS 178, Spring Marc Levoy Computer Science Department Stanford University

Getting The Most From Your Imaging Equipment. John Smith Advanced Imaging Conference October 28, 2012

An Inherently Calibrated Exposure Control Method for Digital Cameras

You, too, can make useful and beautiful astronomical images at Mees: Lesson 1

6. Very low level processing (radiometric calibration)

Introduction to Astrophotography

INCREASING LINEAR DYNAMIC RANGE OF COMMERCIAL DIGITAL PHOTOCAMERA USED IN IMAGING SYSTEMS WITH OPTICAL CODING arxiv: v1 [cs.

Atik Infinity; StarlightXpress Ultrastar; and Mallincam StarVision.

The 0.84 m Telescope OAN/SPM - BC, Mexico

ObsAstro Documentation

CCD Image Calibration Using AIP4WIN

Temperature Reductions to Mitigate the WF4 Anomaly

ObsAstro Documentation

Abstract. Preface. Acknowledgments

High Dynamic Range Imaging

SBIG ASTRONOMICAL INSTRUMENTS

AIC Narrowband Imaging Things That Make a Difference Saturday, October 27, 2007 Neil Fleming. (

A Kalman-Filtering Approach to High Dynamic Range Imaging for Measurement Applications

SEAMS DUE TO MULTIPLE OUTPUT CCDS

High Contrast Imaging using WFC3/IR

MY ASTROPHOTOGRAPHY WORKFLOW Scott J. Davis June 21, 2012

Note: These sample pages are from Chapter 1. The Zone System

Photometry. Variable Star Photometry

Dark current behavior in DSLR cameras

Camera Calibration Certificate No: DMC III 27542

E19 PTC and 4T APS. Cristiano Rocco Marra 20/12/2017

Ron Brecher. AstroCATS May 3-4, 2014

Padova and Asiago Observatories

On spatial resolution

Single Photon Interference Katelynn Sharma and Garrett West University of Rochester, Institute of Optics, 275 Hutchison Rd. Rochester, NY 14627

THE CCD RIDDLE REVISTED: SIGNAL VERSUS TIME LINEAR SIGNAL VERSUS VARIANCE NON-LINEAR

Orion StarShoot G3 Deep Space Imaging Cameras

A 120dB dynamic range image sensor with single readout using in pixel HDR

Response Curve Programming of HDR Image Sensors based on Discretized Information Transfer and Scene Information

Photometry of the variable stars using CCD detectors

Residual bulk image quantification and management for a full frame charge coupled device image sensor. Richard Crisp

F-number sequence. a change of f-number to the next in the sequence corresponds to a factor of 2 change in light intensity,

How to capture the best HDR shots.

WFC3 TV2 Testing: UVIS Shutter Stability and Accuracy

Flat Fields. S. Eikenberry Obs Tech

New Bad Pixel Mask Reference Files for the Post-NCS Era

Tabletop Lighting Variations. February 9, 2008 By Carol McGonegal Colby Munger

SNR IMPROVEMENT FOR MONOCHROME DETECTOR USING BINNING

On the Bench: QHY-10 Craig Stark

Use of the Shutter Blade Side A for UVIS Short Exposures

It Takes Two to Tango

INTRODUCTION TO CCD IMAGING

Astrophotography Basics

STIS CCD Saturation Effects

Total Comet Magnitudes from CCD- and DSLR-Photometry

Transcription:

Combining Images for SNR improvement Richard Crisp 04 February 2014 rdcrisp@earthlink.net

Improving SNR by Combining Multiple Frames The typical Astro Image is made by combining many sub-exposures (frames) into a single exposure using averaging, median combines etc. The SNR is improved by this combining of data, but: Is there a limit to how many frames should be combined?

Signal To Noise Ratio Theory Signal to Noise ratio is defined: = In the general case there are many noise components: the most commonly encountered significant ones are included in the equation below = + _ + + + (1) (2) Many of these noise components can be made negligible by picking proper operating temperature and performing simple image calibration Fixed Patten Noise is eliminated by proper Flat-Fielding Dark Shot Noise: cooling can render this parameter insignificant Dark Fixed Pattern Noise: dark-subtraction ( despiking ) eliminates what remains of this term that cooling doesn t directly eliminate Applying these techniques the noise equation reduces to: = + _ (3)

Signal To Noise Ratio Theory Since = (4) Substituting (4) into (3) the noise equation becomes: = + _ (5) Substituting (5) into (1) we get = + _ (6) If the read noise is small compared to the signal, it further reduces to = (7)

Doubling the SNR: Implications Neglecting read noise, FPN and dark noise components, to double the SNR, we need 4x as much signal = 2 = 2 = 4 = : = 4* While doubling the number of exposures does not have any limit, there are practical considerations given to the total amount of time available for taking the image data One situation where a very large number (> 200) of exposures may be beneficial is when capturing the faint nebulosity surrounding a very bright object that you wish to avoid excessively saturating. Example: Trapezium/M42 from camera with ABG and 20,500 e- saturation signal: up to 233 exposures of 90 sec.

Example HDR Image taken with large # of exposures M42/Trapezium Region LRGB image KAI29050 sensor: saturation signal is 20,500 e- 4 hours of R, G, B total (1H 20M/channel), 120 second exposures 7.25 hours of Luminance, 90 second exposures

Region of interest Dark features in this region are buried in the read noise with 90 sec exposures Example HDR Image

ABG Sensor (KAI29050) with 20,500 e- saturation signal on HDR object (M42) 5frames 10frames 15frames 30frames 60frames 101frames 233frames Very faint parts of image are buried in read noise: use large numbers of frames to improve SNR (ie 90 second exposures in Orion Nebula)

Examination of Noise using Bias Frames Since a Bias frame is a zero length exposure it is largely a measure of the system read noise It does take a finite time to readout/download from the sensor so dark signal is accumulated during that time. Cooling can make the dark signal less important Examining stacks of cooled bias frames therefore is a way to estimate the impact of read noise contribution to the overall noise in a stack of short exposures such as in the M42 example Theory says that a quadrupling of # of frames is needed for a 2x reduction of noise. Let s confirm by experiment

Experimental verification of Theory Procedure: Take 1024 identical bias frames Average 2, 4, 8, 16, 32, 64, 128, 256, 512, 1024 frames together Crop a 10,000 pixel box in a clean area Measure Std Dev in this box and record Plot Std Dev vs # frames on Logarithmic axes

Bias Frames Analyzed Standard Deviation of the entire cropped image is measured and recorded for each image

Noise Reduction by Stacking 100 Bias: Random Noise (DN) vs # Frames Averaged 10 Slope = -1/2 Means its describing equation is of the form Noise ~ 1/Sqrt(# frames stacked)) Noise (DN) 1 1 10 100 1000 10000 Noise (DN) # frames stacked 0.1

Practical Observations In this example with 1 image, the effective read noise contributed by the camera is 14.33 DN. Stacking 20 frames it is reduced to about 3 DN. To reduce below 1 DN requires around 250 frames For images of very faint objects it may be necessary to take a large number of exposures, particularly if the exposure time is limited out of dynamic range considerations. For M42, the Trapezium stars limit the exposure time so many exposures are needed to get the faint nebulosity with acceptable SNR 100 Bias: Random Noise (DN) vs # Frames Stacked 10 Noise (DN) 1 3 250 20 1 10 100 1000 10000 Noise (DN) # frames stacked 0.1