Model 532 ma/v Loop Calibrator with Loop Diagnostic Operating Instructions

Similar documents
PIECAL Model 541 Frequency Calibrator with Totalizer Operating Instructions

Dual T/C & RTD Calibrator With Auto Stepping

Meriam Model M ma Loop Calibrator Operating Instructions

PIECAL 311 Automated Universal RTD Calibrator Operating Instructions. Product Description. Practical Instrument Electronics

PIECAL 211 Automated Universal RTD Calibrator Operating Instructions. Product Description. Practical Instrument Electronics

PIECAL 211 Automated RTD Calibrator Operating Instructions. Product Description. Practical Instrument Electronics

PIECAL 311 Automated Universal RTD Calibrator Operating Instructions. Product Description

Milliamp Calibrator. Model 434. General description. Calibrate with laboratory accuracy. All 4 to 20 ma loop functions

Model 422 Diagnostic Thermocouple & Milliamp Calibrator Operating Instructions

PIECAL 820 Multifunction Process Calibrator

Process Calibrator. TechChek 820

PIE 525. Diagnostic Thermocouple RTD & Milliamp Calibrator Operating Instructions. Practical Instrument Electronics

PIE 525B. Automated Thermocouple & RTD Calibrator Operating Instructions. Practical Instrument Electronics

Practical Instrument Electronics

POWER SERIES Plus. Volt / Amp / Hertz. Digital Switchboard Meter. User s Manual IM2493VAF-3

Practical Instrument Electronics. PIE 311 Diagnostic RTD & Milliamp Calibrator. Best RTD & Milliamp Calibrator on the market!

Practical Instrument Electronics

User s Guide CL535. Frequency Calibrator with Totalizer. Shop online at. omega.com For latest product manuals: omegamanual.

User s Guide CL309A Milliamp Loop Calibrator. Shop online at omega.com SM

Practical Instrument Electronics PIECAL 820

ProScale. Compact LCD Readout. Installation & Operation

ProScale Compact LCD Readout. Operation. Firmware version C 2.xxx and higher

DIGITAL SOUND LEVEL DSM8930 METER USER S MANUAL. Please read this manual carefully and thoroughly before using this product.

Frequency Calibrator with Totalizer

User Manual January Opticom Infrared System RC790 Remote Coding Unit

LBI-38976A. Mobile Communications MDX/ORION DESK TOP STATION. Operator's Manual

Microwave Meter. Instruction Manual

Frequency Calibrator with Totalizer

Process Calibrator. TechChek 830-KP

BC145 SIGNAL ISOLATOR BOARD

Series 3000 Model R-107A

PIE mA/10-50mA Process Calibrator. Practical Instrument Electronics

Firmware Version d & higher Installation & Operation

INSTALLATION INSTRUCTIONS

Firmware Version d & higher Installation & Operation

PIE 830. Process Calibrator. Practical Instrument Electronics

INSTALLATION INSTRUCTIONS

INSTALLATION AND SETUP GUIDE

Passive Component LCR Meter

SA-027HQ 365-Day Annual Timer

Operator Manual 1.4 FRACSIM MINI

Automated Thermocouple & RTD Calibrator

Digi-Stop. Installation & Operation

99 Washington Street Melrose, MA Fax TestEquipmentDepot.com # # AAC Clamp Meter. Instruction Manual

EX Multi-Point Process Switch KING-GAGE ma Process Control. Single-Channel Analog. Start-Up/Operation Instructions.

12B/18 MultiMeter Instruction Sheet

600A Clamp Meters w/tightsight Display

CCE Image may differ from the actual product By Martin Labbé, eng., Jasmin Goupil & Louis Perreault

2006 MFJ ENTERPRISES, INC.

Infrared Gun. Perfect For. Easy To Use. Features. Get Professional Results Every Time! Battery Installation. Model: IN1022

Independent Technology Service Inc Independence Ave. Chatsworth, California Toll Free:

MBC Bipolar Microstep Driver. User s Guide E. Landon Drive Anaheim, CA

THE PROBE, 5 METER, 2WIRE, LOOP POWERED SINGLE POINT LEVEL MEASUREMENT SYSTEM (ULTRASONIC)

PIN/PINLESS DEEP SENSING MOISTURE METER WITH SPHERICAL SENSOR AND REMOTE PROBE

Passive Component LCR Meter

LBI Mobile Communications MDX DESK TOP STATION. Operator's Manual

Digi-Fence. User Manual (all models) For All Models with Digital Readout Firmware version d & Higher

Digi-Stop. User Manual for: Digi-Stop. Digital Readout Firmware version d & Higher

Wilcom MODEL T136BSBZW CIRCUIT TEST SET. Operating Instructions

OPERATING AND MAINTENANCE INSTRUCTIONS FOR CONSTANT CURRENT RECTIFIERS USING SATURABLE CORE REACTORS

USER MANUAL MODEL Parallel to Serial/ Serial to Parallel Interface Converter

PIE 820-ELITE Multifunction Diagnostic Process Calibrator

9320 Manual Portable Battery Powered Indicator

CDT. Service and Installation Manual. Manual Revision Oct 2014

Copyright 2014 by Minarik Drives

Easy-Link Plus Version 2.2

8003 1/8 DIN Temperature Controller

Digi-Fence. User Manual for: Digi-Fence (all models) Digital Readout Firmware version d & Higher

SSR-150xxx-40VL-12P-xC-xxxCS

COMMERCIAL TRANSMITTER INSTRUCTIONS

SMH250D/SMU250K User s Guide

This manual should be read in its entirety prior to operation of the TS200 Test Set Page 2 of 9

USER S MANUAL PROCESS CONTROL MODULE PCM3

Kotron Sentinel Model 804 RF Level Transmitter

Far End Device II Model 1342

HHMA2 DC / TRUE RMS AC NON-CONTACT MILLIAMMETER

Instruction Notes for 108A L Sensor Input

UNCONTROLLED WHEN PRINTED

OPERATOR S MANUAL Model 160 Volt-Ohm-Milliammeter

Micro-Ohmmeter Model UM200

CHAPTER MAINTENANCE AND TROUBLESHOOTING. In This Chapter... Maintenance and Inspection Troubleshooting...6 3

Blue Point Engineering

IXIAN TM class V 1.5. EC Transmitter This is an evolving document. EC Transmitter IXIAN TM. Features

Property of Ramsey Electronics, Inc. Do not reproduce or distribute.

Owner s. Manual. Expand Your Factory Radio. Honda/Acura. Media GateWay PXAMG. HD Radio Operation. isimple Connect

4590 Tank Side Monitor. Service Manual. Mark/Space Communication Protocol. Software Version v2.03 SRM009FVAE0808

SMART TWEEZERS Model ST-1 User s Manual Version 1.0

Wilcom MODEL T136BGMZW CIRCUIT TEST SET. Operating Instructions

Programmable K-Factor Scaler B and Programming Software Kit B

Operating Instructions

User s Guide Instructions for Installation and Operation

f i r e - p a r t s. c o m

Owner s Manual. Model FR-1400 Two Way Family Radio A 1 of 20. Customer Service Manufacturer will reduce to 75 per cent.

MC133C/133Ci. Electric Power/Energy Calibrator. Operation manual

INSTRUCTION MANUAL INF Fax: (503)

CBCT TM CIRCUIT BREAKER COIL TESTER

Pen Multimeter. Model

PCO-7114 Laser Diode Driver Module Operation Manual

Note: There are two versions of the PIECAL 211 and 311

LPU-2127 User Manual

Transcription:

Practical Instrument Electronics Model 532 ma/v Loop Calibrator with Loop Diagnostic Operating Instructions A. Basic Keypad Operations j EZ-Check Switch/EZ-Step Pushbutton Slide the switch to select the user stored values for calibration points. Press the button to adjust the output by the user defined step size. Press and hold the button to activate the auto step/ramp mode. m UNITS/STEP TIME Button Press and release UNITS/STEP TIME to change how current is displayed: either in milliamperes or % of 4-20 ma. Voltage is only displayed in volts. Press and hold UNITS/STEP TIME to change step size. Refer to section H. READ/SOAK TIME Button Press and release READ/SOAK TIME to change read modes. These are: Read Milliamps Power and Measure 2-Wire Transmitter Read Volts Press and hold READ/SOAK TIME to change soak time. Refer to section H. STORE/CLEAR/DIAGNOSTIC Button In any source mode: Press STORE/CLEAR/DIAGNOSTIC to save the current reading in the EZ-Check HI or LO position. The EZ-Check switch must be set to HI or LO. The display will flash STORED to confirm. k SOURCE/STEP SIZE Button Press and release SOURCE/STEP SIZE to change source modes. These are: Source Milliamps 2-Wire Transmitter Simulate Source Volts Press and hold SOURCE/STEP SIZE to change step size. Refer to section H. In any read mode: Press STORE/CLEAR/DIAGNOSTIC to clear the values saved in the EZ-Check HI and LO positions. The display will flash CLEARED to confirm. Press and hold STORE/CLEAR/DIAGNOSTIC to activate diagnostic mode. See section I. EZ-Dial Knob Turn the EZ-Dial knob to adjust the output level. Press and turn to adjust 100X faster. ON/OFF Button Press ON/OFF to turn the Model 532 on or off. Practical Instrument Electronics, Inc. Copyright ª 2011. All rights reserved. 532-9002 Rev C 7/13/11 1-8

B. Model 532 Configuration Press the EZ-Dial Knob while turning the Model 532 on to access the configuration mode. Turn the EZ-Dial Knob to select configuration items. Press the EZ-Dial Knob to change configuration items. Turn the unit off or just wait approximately 8 seconds to exit the configuration mode. Auto Off - ON (default)/off Auto Off is ON, by default, to save battery life by turning the unit off after 30 minutes of inactivity. Turn Auto Off to OFF to prevent automatic shutdown. This is typically useful for manual loading or continuous use. EZ-Step - ON/OFF (default) If EZ-Step is ON manual and automatic stepping/ramping is available. If EZ-Step is OFF the EZ-Step pushbutton will be disabled and the step direction indicator will not be displayed. HART Compatibility Mode - ON/OFF (default) The Model 532 has a HART compatibility mode. This mode is useful when the devices being powered communicate using the HART protocol. In this mode the Model 532 connects a 250 Ω load resistor in series with the output in both Source and Power Measure 2-Wire transmitter modes. This eliminates the requirement of an external 250 Ω load resistor. This resistor is typically shown in connection diagrams and manuals for HART devices. If HART Compatibility Mode is ON, a 250 Ω load resistor is automatically switched in series with the output in Source and Power Measure 2-Wire Transmitter modes. The output compliance with HART Compatibility Mode ON is 950 Ω at 20 ma. If HART Compatibility Mode is OFF there is no 250 Ω load resistor in series with the output. This will increase the output compliance voltage to drive 1200 Ω at 20 ma. EZ-Check HI/LO Readings ON/OFF (default) If the EZ-Check HI/LO Readings option is ON, the highest and lowest readings will automatically be saved in the HI and LO EZ-Check positions. If this option is OFF the HI and LO positions will show the current reading. Loop Diagnostic ON (default)/off Loop Diagnostic may be turned off to prevent entry into diagnostic mode. This may make the Model 532 simpler to operate by eliminating accidental entry into diagnostic mode. Factory Reset ON/OFF (default) If Factory Reset is ON, the unit will restore all factory defaults when the Model 532 is turned OFF and back ON. This will reset any changes made in the Model 532 Configuration options, returning the unit to its simplest factory configuration. Practical Instrument Electronics, Inc. Copyright ª 2011. All rights reserved. 532-9002 Rev C 7/13/11 2-8

C. EZ-Dial Knob Adjust the output up and down with the EZ-Dial knob. The increment is 0.001 ma (or 0.01 % if display units are % of 4-20 ma.) Press while turning to adjust 100X faster 0.100 ma (or 1.00 %.) D. EZ-Check Switch The EZ-Check switch has three positions -- high, set, and low. Its position is shown at the left edge of the display with HI and LO indicators. Use of the EZ-Check switch depends on mode: Source Modes: Slide the EZ-Check switch to the HI and LO positions to recall the settings stored in those positions. While in the HI and LO positions, dial the EZ-Dial knob to change the display. Press STORE/CLEAR to save new settings in the HI and LO positions. The display will flash STORED to confirm. Hint: For faster calibrations, the position of the switch can be felt. This feature allows continuous monitoring of the device being calibrated without looking back at the Model 532 display. This is also useful in poor lighting or under difficult operating conditions. Read Modes: In read modes, the Model 532 calibrator records the maximum and minimum readings observed in each mode. Slide the EZ-Check switch to the HI and LO positions to display the readings. Press STORE/CLEAR to clear the readings. The display will flash CLEARED to confirm. By default, the Model 532 has EZ-Check HI/LO Readings OFF. Refer to Model 532 Configuration, section B. E. EZ-Step Pushbutton/ Manual Step The EZ-Step pushbutton is a feature only in source modes. Press and hold the EZ-Step pushbutton for less than one second to cause the output to step up or down by the EZ-Step size. The EZ-Step direction is indicated on the display ( or ). Press the EZ-Dial knob to change the step direction. Stepping and auto step/ramp limits are defined by the EZ-Check HI and LO settings. The step direction changes when a limit is reached. The step size is computed as the difference between the EZ-Check HI and the EZ-Check LO divided by the number of steps. See section H. By default, the Model 532 has EZ-Step OFF. Refer to Model 532 Configuration, section B. F. Auto Step/Ramp Press the EZ-Step pushbutton for more than one second to activate auto step/ramp mode. The Model 532 will automatically step by the selected EZ-Step size and time. Press the EZ-Step pushbutton again to deactivate auto step/ramp mode. The EZ-Step direction is indicated on the display ( or ). Press the EZ-Dial knob to change the step direction. The step direction can be changed while automatically stepping/ramping. Practical Instrument Electronics, Inc. Copyright ª 2011. All rights reserved. 532-9002 Rev C 7/13/11 3-8

Stepping and auto step/ramp limits are defined by the EZ-Check HI and LO settings. The step direction changes when a limit is reached. Figure 1 shows how the Step/Ramp Parameters are used to configure automatic stepping/ramping. Figure 1 24 20 ramp time step time = ramp time # of steps 16 soak time ma 12 8 step size = (EZ-Step HI EZ-Step LO) # of steps 4 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 time Note: The Model 532 s ability to detect overload/undervoltage conditions may be limited by the rate of change in the output when using automatic stepping/ramping. Turn auto step/ramp off while connecting or disconnecting the Model 532. G. Quick Reference Bar Graph The Quick Reference Bar Graph indicates the input and output level on the Model 532 in % of 4-20 ma with 1% resolution. Practical Instrument Electronics, Inc. Copyright ª 2011. All rights reserved. 532-9002 Rev C 7/13/11 4-8

H. Manual Step and Auto Step/Ramp Parameter To Change the EZ-Step Size: 1. Press and hold the SOURCE/STEP SIZE button for more than ¾ of a second. 2. The display will flash EZ-STEP SIZE. 3. Turn the EZ-Dial knob to select from 2 to 16 steps between the EZ-Check limits. 4. Turn the EZ-Dial clockwise past 16 steps to select continuous ramp mode. 5. Press the SOURCE/STEP SIZE button again to return to the normal display. Note: If the EZ-Step option is turned off, the display will flash EZ-STEP OFF. Refer to Model 532 Configuration, section B. To Change the EZ-Step Time: 1. Press and hold the UNITS/STEP TIME button for more than ¾ of a second. 2. The display will flash EZ-STEP TIME. 3. Turn the EZ-Dial knob to select from 5 to 900 second ramp time. The time per step is calculated based on the selected EZ-Step size. 4. Press the SOURCE/STEP SIZE button to return to the normal display. To Change the Soak Time: 1. Press and hold the READ/SOAK TIME button for more than ¾ of a second. 2. The display will flash SOAK TIME. 3. Turn the EZ-Dial knob to select from 0 to 900 second soak time. Note: A soak time of 0 defeats the soak period. The step time will be used instead. 4. Press the READ/SOAK TIME button again to return to the normal display. To Change the EZ-Step TM Direction: 1. Press and release the EZ-Dial knob without turning. 2. The display will change to show the EZ-Step direction selected ( or ). Practical Instrument Electronics, Inc. Copyright ª 2011. All rights reserved. 532-9002 Rev C 7/13/11 5-8

I. Loop Diagnostic Press and hold the STORE/CLEAR/DIAGNOSTIC button for more than ¾ of a second to activate loop diagnostic mode. This is available in each of the Model 532 s operating modes. Read Modes: Read Milliamps Loop current is displayed. Power and Measure 2-Wire Transmitter Loop current, voltage, and resistance is displayed. Read Volts Voltage is displayed. Source Modes: The EZ-Dial knob, EZ-Check switch, and EZ-Step pushbutton function normally when the loop diagnostic is activated. Source Milliamps Loop current, voltage, and resistance is displayed. The resistance is highlighted if it exceeds the Model 532 s output capability (1200 W at 20 ma with HART Compatibility disabled.) 2-Wire Transmitter Simulate Loop current is displayed. The Model 532 automatically performs a test every 7 seconds to compute the loop power supply voltage and loop resistance. The Model 532 requires approximately 2 volts across its terminals to operate in 2 Wire Simulate mode. The voltage display is highlighted if there is less than 2 volts present. Source Volts Loop current, voltage, and resistance is displayed. The resistance is highlighted if it exceeds the Model 532 s output capability (20 ma into 1200 W with HART Compatibility disabled.) Practical Instrument Electronics, Inc. Copyright ª 2011. All rights reserved. 532-9002 Rev C 7/13/11 6-8

In all loop diagnostic modes, AC voltage is displayed. If more than 2 VAC is present on the Model 532 s terminals, the display is highlighted to alert the user of a potential problem. In all loop diagnostic modes, HART activity is indicated with ' HART DETECTED '. The ' symbols flash on and off with the actual HART messages. J. Modes of Operation Read Milliamp Connect the Model 532 in series with the process loop to monitor current. Observe correct polarity. Current limiting above 24.000 ma is indicated by a flashing CURRENT LIMITED display. Power and Measure 2 Wire Transmitter The Model 532 provides power to the process loop while displaying output current. Use this mode to test a transmitter s ability to control loop current. Current limiting above 24.000 ma is indicated by a flashing CURRENT LIMITED display. Read Volts The Model 532 measures +/- 30 VDC with 4X overrange ability. The display flashes OVERRANGE when the 30 volt limit is exceeded. 0.001 volt resolution below 24.00 volts provides the ability to calibrate 1-5 volt instrumentation. Practical Instrument Electronics, Inc. Copyright ª 2011. All rights reserved. 532-9002 Rev C 7/13/11 7-8

Source Milliamp Connect the Model 532 directly to 4-20 ma receiver equipment, alarms, panel meters, etc. Use the EZ-Dial Knob and EZ-Check Switch to adjust loop current. The display flashes HIGH Ω when the loop resistance is too high or the leads are open. 2 Wire Transmitter Simulate Substitute the Model532 for a 2 wire transmitter. Use the EZ-Dial Knob and EZ- Check Switch to adjust loop current. At least 2 volts of loop power is required, else the display flashes CHECK LOOP SUPPLY. Source Volts The Model 532 sources 0.000-24.000 volts. This is useful for powering transmitters and receiver equipment. Use the EZ-Dial Knob and EZ-Check Switch to adjust output voltage. The display flashes LOW W when the output is overloaded. K. Warranty Our equipment is guaranteed against defective material and workmanship (excluding batteries) for a period of three years from the date of shipment. Claims under guarantee can be made by returning the equipment prepaid to our factory. The equipment will be repaired, replaced or adjusted at our option. The liability of Practical Instrument Electronics (PIE) is restricted to that given under our guarantee. No responsibility is accepted for damage, loss or other expense incurred through sale or use of our equipment. Under no condition shall Practical Instrument Electronics, Inc. be liable for any special, incidental or consequential damage. Practical Instrument Electronics, Inc. Copyright ª 2011. All rights reserved. 532-9002 Rev C 7/13/11 8-8