WWW.VTIINSTRUMENTS.COM RX1032 PRECISION TEMPERATURE MEASUREMENTS FOR RUGGED ENVIRONMENTS FEATURES 32-Channel Thermocouple Measurement System Advanced CJC Implementation 1000 Sample/sec/channel Scan Rate Precise Distributed Measurement Synchronization with IEEE-1588 Rugged Design Broad Operating Temperature Range IP65 Environmental Rating Ethernet Connectivity via LXI Open Industry Standard IVI-COM, IVI-C, LabVIEW Driver Support 1 2016 VTI Instruments Corporation. All rights reserved.
RX1032 OVERVIEW The RX1032 thermocouple measurement instrument is a ruggedized version of the popular EX1000 series, the most advanced, full-featured data acquisition solutions available on the market today. The RX1032 unit provides superior measurement accuracy and repeatability by implementing fully integrated signal conditioning and advanced cold junction compensation (CJC). The RX1032 series uses military grade connectors for power and connectivity while thermocouple and sensor inputs are routed through a protected entry to a screw terminal panel for quick and easy connectivity. The rugged construction of the enclosure enables the measurement system to be mounted on test pylons, in test cells and in other harsh environments. SCALABLE FOR HIGH- SPEED SYNCHRONIZED DATA ACQUISITION The RX10XX family supports easy integration and synchronization of multiple devices through an industry standard IEEE-1588 implementation. Multiple RX1032 instruments can easily be distributed and placed extremely close to the measurement points of interest, reducing the length of analog cable, minimizing errors induced by noisy environments. All measurement data is returned with IEEE-1588 timestamp codes with typical accuracies of <200µS ensuring that the acquired data is tightly correlated across the entire test article. APPLICATIONS Large Scale Engine Test Harsh Environment Health Monitoring HALT / HASS Rocket Motor Reliability Wind Tunnel Test COLD JUNCTION COMPENSATION The heart of any truly accurate thermocouple measurement system is the CJC implementation. The RX1032 combines multiple precision thermistors, a significant thermal mass, and careful parts placement to provide world-class measurement performance. 2
SPECIFICATIONS CHANNELS CHANNELS 32 differential inputs CHANNEL TYPES Thermocouple inputs: J, K, T, E, S, R, B, N SAMPLING RATE 1000 Sa/sec/ch maximum TEMPERATURE RESOLUTION <0.1 C TEMPERATURE ACCURACY See Thermocouple Accuracy table TEMPERATURE NOISE, PEAK-TO-PEAK 0.08 Cpp typical (J, K, T, E) PROGRAMMABLE FILTERS BESSEL (2 POLE) BUTTERWORTH (1 POLE) COMMON MODE REJECTION RATIO 4 HZ FILTER 1 KHZ FILTER 4 Hz, 15 Hz, 40 Hz, 100 Hz, 500 Hz (-3 db cutoff frequency) 1000 Hz (-3 db cutoff frequency) DC: 100 db MINIMUM; (50/60) HZ: 140 db TYPICAL, 120 db MINIMUM DC: 100 db MINIMUM; (50/60) HZ: 100 db TYPICAL, 80 db MINIMUM 3
LXI SPECIFICATIONS CLOCK SPECIFICATIONS CLOCK OSCILLATOR ACCURACY SYNCHRONIZATION ACCURACY TIMESTAMP ACCURACY RESOLUTION ±50 PPM REPORTS SYNCHRONIZED WHEN < ±200 μs OF THE 1588 MASTER CLOCK AS GOOD AS TIME SYNCHRONIZATION DOWN TO 50 ns 25 ns IEEE 1588- BASED TRIGGER TIMING ALARM TRIGGER TIME ACCURACY RECEIVE LAN [0-7] EVENT TRIGGER TIME ACCURACY FUTURE TIMESTAMP PAST/ZERO TIMESTAMP AS GOOD AS TIME SYNCHRONIZATION DOWN TO 50 μs 50 μs AS GOOD AS TIME SYNCHRONIZATION DOWN TO 50 μs 50 μs TYPICAL 1 ms MAXIMUM HARDWARE TRIGGER TIMING LXI TRIGGER BUS DIO BUS 55 μs TYPICAL DIO BUS 57 μs TYPICAL ENVIRONMENTAL SPECIFICATIONS TEMPERATURE -40 C TO +71 C (REDUCED ACCURACY), 0 C TO +50 C (FULL OPERATING ACCURACY) HUMIDITY IP 65 RATED ENCLOSURE, 5% TO 95% ALTITUDE SHOCK AND VIBRATION UP TO 4600 M CONFORMS TO MIL-PRF-28800F RANDOM VIBRATION 30 MIN PER AXIS, 10-500Hz MIL-PRF-28800F CLASS 2 5 TO 55HZ RESONANCE SEARCH PER MIL-PRF-28800F CLASS 2, SINUSOIDAL EACH AXIS SHOCK 30G/AXIS, 11MS HALF SINE PULSE PER MIL-PRF-28800F CLASS 2 GENERAL INPUT PROTECTION TRANSDUCER SCREW TERMINAL INSTRUMENT CONNECTION POWER INPUT 15V ZENER PROTECTION M3 X 6MM (PAN HEAD PHILLIPS) 10/100 BASE-T INPUT CONNECTOR SCREW TERMINAL INPUT VOLTAGE DC, 18-36V DC DIMENSIONS (H x W x D) 125 mm x 238 mm x 249 mm; 4.92 x 9.05 x 9.45 WEIGHT 5.3 kg; 11 lbs. 4
THERMOCOUPLE ACCURACY Operating Temperature +15 C to +25 C -40 C to +50 C +50 C to +70 C Type E, J, K, N, and T Typical ±0.12 C ±0.22 C ±0.53 C Max ±0.20 C ±0.35 C ±0.93 C Type R, S Typical ±0.48 C ±0.88 C ±2.12 C Max ±0.80 C ±1.40 C ±3.72 C Type B Typical ±0.96 C ±1.76 C ±4.24 C Max ±1.60 C ±2.80 C ±7.44 C 1) After 1-hour warmup period and temperature stable for 1 hour after any temperature change 2) Exclusive of thermocouple, wiring, noise, common mode errors 3) Temperature variation less than 0.1 C/min 4) Typical is based on 2σ and Max is based on 4σ Specifications contained within this document are subject to change without notice. Revision 20161021. 5
ORDERING INFORMATION Model RX1032 Configuration Ruggedized 32 Channel Thermocouple Instrument BLOCK DIAGRAM RX-10XX CHASSIS MOUNTING DIMENSIONS (IN MM) VIBRATION/SHOCK ISOLATOR MOUNTING DIMENSIONS (IN MM) CONTACT +1 949-955-1894 sales@vtiinstruments.com Corporate Headquarters 2031 Main St. Irvine, CA 92614 6