Very short introduction to light microscopy and digital imaging

Similar documents
BASICS IN BIOIMAGING AND OPTICS PLATFORM EPFL SV PTBIOP LIGHT MICROSCOPY

Imaging Introduction. September 24, 2010

microscopy A great online resource Molecular Expressions, a Microscope Primer Partha Roy

Systems Biology. Optical Train, Köhler Illumination

Education in Microscopy and Digital Imaging

Microscopy: Fundamental Principles and Practical Approaches

Microscope anatomy, image formation and resolution

Introduction to Light Microscopy. (Image: T. Wittman, Scripps)


VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES

INTRODUCTION TO OPTICAL MICROSCOPY

Katarina Logg, Kristofer Bodvard, Mikael Käll. Dept. of Applied Physics. 12 September Optical Microscopy. Supervisor s signature:...

Light Microscopy. Upon completion of this lecture, the student should be able to:

EE119 Introduction to Optical Engineering Spring 2003 Final Exam. Name:

Basics of Light Microscopy and Metallography

BASICS IN LIGHT MICROSCOPY

Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE

Properties of optical instruments. Visual optical systems part 2: focal visual instruments (microscope type)

Optics Day 3 Kohler Illumination (Philbert Tsai July 2004) Goal : To build an bright-field microscope with a Kohler illumination pathway

Microscopy Techniques that make it easy to see things this small.

Microscopy. Matti Hotokka Department of Physical Chemistry Åbo Akademi University

BIOIMAGING AND OPTICS PLATFORM EPFL SV PTBIOP BASICS IN LIGHT MICROSCOPY

Chapter 3 Op,cal Instrumenta,on

Nikon E800 Operating Instructions.

2/4/15. Brightfield Microscopy! It s all about Magnification..! or is it?!

Lecture 23 MNS 102: Techniques for Materials and Nano Sciences

MICROSCOPE LAB. Resolving Power How well specimen detail is preserved during the magnifying process.

Microscopy Training & Overview

Bio 407. Applied microscopy. Introduction into light microscopy. José María Mateos. Center for Microscopy and Image Analysis

Observing Microorganisms through a Microscope LIGHT MICROSCOPY: This type of microscope uses visible light to observe specimens. Compound Light Micros

Light microscopy BMB 173, Lecture 14, Feb. 21, 2018

The Compound Microscope. Brightfield: Köhler Illumination

Optical Design of. Microscopes. George H. Seward. Tutorial Texts in Optical Engineering Volume TT88. SPIE PRESS Bellingham, Washington USA

Chapter 3 Op+cal Instrumenta+on

ECEN 4606, UNDERGRADUATE OPTICS LAB

ANSWER KEY Lab 2 (IGB): Bright Field and Fluorescence Optical Microscopy and Sectioning

The microscope is useful in making observations and collecting data in scientific experiments. Microscopy involves three basic concepts:

FLUORESCENCE MICROSCOPY. Matyas Molnar and Dirk Pacholsky

Reflection! Reflection and Virtual Image!

Figure 3.4 Approximate size of various types of cells. ~10 um. Red Blood Cells = mm 1500 um. Width of penny Pearson Education, Inc.

TOPICS Recap of PHYS110-1 lecture Physical Optics - 4 lectures EM spectrum and colour Light sources Interference and diffraction Polarization

Point Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy

Characterization Microscope Nikon LV150

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy,

Marine Invertebrate Zoology Microscope Introduction

INSTRUCTIONS FOR COURSE WORK 4 (AxioVert) Instructor: Anne Vaahtokari (MIU) 1. Purpose of the work

Invitation for a walk through microscopy. Sebastian Schuchmann Jörg Rösner

Observing Microorganisms through a Microscope

Prof. Enrico Gratton - Lecture 6 Fluorescence Microscopy

CHAPTER TWO METALLOGRAPHY & MICROSCOPY

EPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY

Nikon Instruments Europe

Light Microscopy for Biomedical Research

Physics 431 Final Exam Examples (3:00-5:00 pm 12/16/2009) TIME ALLOTTED: 120 MINUTES Name: Signature:

Motorized Axio Observer Start-up instructions

Rates of excitation, emission, ISC

Köhler Illumination: A simple interpretation

PHYSICS. Chapter 35 Lecture FOR SCIENTISTS AND ENGINEERS A STRATEGIC APPROACH 4/E RANDALL D. KNIGHT

OPTICAL PRINCIPLES OF MICROSCOPY. Interuniversity Course 28 December 2003 Aryeh M. Weiss Bar Ilan University

Tissue Preparation ORGANISM IMAGE TISSUE PREPARATION. 1) Fixation: halts cell metabolism, preserves cell/tissue structure

A BRIEF INTRODUCTION TO MICROSCOPY The two key properties of a microscope that allow you to see microbes are resolution and magnification.

Properties of optical instruments. Projection optical systems

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005

EE119 Introduction to Optical Engineering Fall 2009 Final Exam. Name:

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

Week IV: FIRST EXPERIMENTS WITH THE ADVANCED OPTICS SET

INTRODUCTION THIN LENSES. Introduction. given by the paraxial refraction equation derived last lecture: Thin lenses (19.1) = 1. Double-lens systems

Laboratory Introduction

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION

Opti 415/515. Introduction to Optical Systems. Copyright 2009, William P. Kuhn

2013 LMIC Imaging Workshop. Sidney L. Shaw Technical Director. - Light and the Image - Detectors - Signal and Noise

Differential Interference Contrast (DIC) Verses Dark Field and Phase Contrast Microscopy. E. D. Salmon University of North Carolina at Chapel Hill

Optical System Design

SUBJECT: PHYSICS. Use and Succeed.

Nikon E800 Microscope. Operating Instructions

Chapter 25. Optical Instruments

Nature Protocols: doi: /nprot Supplementary Figure 1. Schematic diagram of Kőhler illumination.

CCAM Microscope Objectives

Phys 531 Lecture 9 30 September 2004 Ray Optics II. + 1 s i. = 1 f

Transmission electron Microscopy

Using the Nikon TE2000 Inverted Microscope

Basics of confocal imaging (part I)

Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal

Resolution. Diffraction from apertures limits resolution. Rayleigh criterion θ Rayleigh = 1.22 λ/d 1 peak at 2 nd minimum. θ f D

Using Microscopes. Life Science: Molecular

Phy Ph s y 102 Lecture Lectur 21 Optical instruments 1

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands

Nikon E800 Operating Instructions.

The Microscope. Packet #2. 10/17/2016 9:12:02 PM Ryan Barrow 2012

High resolution extended depth of field microscopy using wavefront coding

Shaping light in microscopy:

LEOK-3 Optics Experiment kit

Leica DB LB Research microscope and Studo Lite Imaging software

Criteria for Optical Systems: Optical Path Difference How do we determine the quality of a lens system? Several criteria used in optical design

Manual for BMS E1 eplan series, compound microscope

Opto Engineering S.r.l.

OPTICAL SYSTEMS OBJECTIVES

Phys 102 Lecture 21 Optical instruments

OPTICS DIVISION B. School/#: Names:

EE119 Introduction to Optical Engineering Spring 2002 Final Exam. Name:

Transcription:

Very short introduction to light microscopy and digital imaging Hernan G. Garcia August 1, 2005 1 Light Microscopy Basics In this section we will briefly describe the basic principles of operation and alignment of a microscope. These notes are based on [1], [4] and the microscopy introductory lectures by Rudi Rottenfusser at the 2005 MBL Physiology Course. In figure 1 we present the basic elements required for magnification. The specimen lies on the focal plane of the objective lens, which focuses it to infinity. The tube lens then focuses the parallel rays on the intermediate image plane. The obtained real image is now a magnified version of the sample. A second round of magnification, which gets multiplied to the previously described one, is obtained by focusing this real image to infinity once again using the eyepiece. This creates a new image focused to infinity, a condition to which the eye is adapted for viewing. eye or camera eyepiece intermediate image plane { tube lens objective back focal plane objective lens specimen Figure 1: Basic diagram of the imaging light path [4] One of the keys to obtaining sharp images is correct alignment. In the imaging path we are able to move the objective around to make the plane of the specimen we are interested in coincide with the objective focal plane. However, this is not the only required alignment: the transmitted illumination has also to be set up correctly to obtain a homogeneous specimen illumination. In figure 2 we present both light paths. In the illumination path the light coming out from the lamp is focused on the condenser diaphragm by the collector lense. Previously to arriving at that plane 1

it can be filtered spatially (its edges can be cut) by the field diaphragm, the reason for this will become clear when we describe the alignment protocol. The condenser diaphragm lies on the focal plane of the condenser lens, which makes sure that the light coming from that plane is collimated and goes through the specimen as parallel beams, which ensures the most possible homogeneous illumination. Finally, we see the illumination is focused once more on the objective back focal plane, the conjugate plane to the specimen plane. imaging light paths illuminating light paths eye or camera eyepiece intermediate image plane { tube lens objective back focal plane objective lens specimen condenser lens condenser diaphragm views field diaphragm collector lens lamp Figure 2: Imaging and illuminating light paths [4] The key parameter in all optical elements of the microscope is the numerical aperture (NA). It is defined as n sin(α), where n is the index of refraction of the medium and α is the maximum angle of light compatible with the optical element (fig. 3). Since light rays corresponding to bigger angles give the highest spatial resolution (they are the ones corresponding to higher wave numbers in fourier space), this magnitude is just telling us what the maximum achievable resolution is. We see that if the objective is in air (n 1) the maximum NA will be of order one. Therefore, by immersing it into a different medium like oil or water a higher resolution can be achieved because more light rays can be collected, as can be seen in figure 4. Finally, in figure 5 we present the different parameters of an objective. Each detail might have relevance when it comes down to not only resolution and magnification power, but also to spherical or color aberrations. 2

Figure 3: Numerical apertures of different objectives [2] Figure 4: Light rays coming from the specimen and collected by the objective in the case of a) a dry (or air) objective and b) an oil immersion objective [2] 1.1 Aligning a Light Microscope: Köhler Illumination Obtaining a homogeneous illumination is key to ensure that the imaging is being taken at the highest possible resolution supported by the microscope and to optimize the contrast. In the following paragraph we describe the protocol to obtain Köhler illumination. 1. Open field and condenser diaphragms. 2. Focus on specimen. This ensures that the optimization will be done for that particular position of the objective along the optical axis. 3. Correct for proper color temperature. A yellowish color might be present if the direct output of the lamp is used without any neutral density filters. 4. Close field diaphragm. 3

Figure 5: Specifications of an objective. [2] 5. Focus field diaphragm by moving the condenser up and down. 6. Center the field diaphragm. This will center the illumination and focus it on the sample. 7. Open the field diaphragm to fill the view of the observer. If the diaphragm is opened beyond the field of view, that light will just contribute to the background due to scattering with the sample outside the field. 8. Look at the objective back focal plane. This can be done either by using a Bertrand lens or by removing the ocular and looking at the tube. 9. Center an in-focus image of the lamp and mirror images at the objective back focal plane using the focus and adjustment screws on the lamp housing. Close the condenser aperture to 0.3 0.9 the NA of the objective. This is a compromise between having a higher NA, which increases the resolution, and not illuminating on a bigger sample area, which will increase the background. 10. Done! 1.2 Different techniques The site www.microscopyu.com hosted by Nikon is a great source of information for different imaging techniques, alignment protocols and various other information on microscopy. 1.2.1 Darkfield The basic concept of this microscopy is to illuminate the sample with high spatial frequencies, filtering the direct light. This can be done by shinning a hollow cone of light on the sample. This way a black background is obtained where outlines of cells, for example, can be easily seen. This technique, however, has some disadvantages. The resolution is usually not very high, thick samples are hard to image and the depth of focus is low. Additionally, internal cellular structure cannot be easily interpreted because of the lack of information given by the direct beams of light. More information on darkfield can be obtained from http://www.microscopyu.com/articles/stereomicroscopy/stereodarkfield.html. 4

1.2.2 Phase Contrast A microscope in phase contrast mode detects slight differences in the light s phase induced by the sample. As can be seen in figure 6, the illumination goes through a condenser annulus, where only certain spatial frequencies are selected. When the phase in the illumination light gets changed by the different features of the sample it will get diffracted and, therefore bent. By filtering out the original components using the phase plate we just get an image of the sample s effect on the light. More information on phase contrast microscopy can be found at http://www.microscopyu.com/articles/phasecontrast/phasemicrosco Figure 6: Diagram of a phase contrast microscope [2] 1.2.3 DIC In DIC microscopy the illumination is polarized and then divided into two rays with different polarization, the ordinary and extraordinary beams, by the Condenser Nomarski Prism. The polarized rays will also be separated spatially by an amount smaller than the optical resolution limit. These two beams will go through the specimen and will accumulate different phases, which upon recombination by the Objective Nomarski Prism, will interfere to give a signal related to the local properties of the sample. Before acquisition the original light beam is filtered out by another polarizer, the analyzer, with its axis oriented perpendicularly to the axis of the original polarizer. This way, even though the linear component of the light does not get imaged, the resulting elliptical and circular components will still be detected. A diagram of a DIC microscope can be found in figure 7. A more detailed description of the DIC microscope can be found at http://www.microscopyu.com/articles/dic/desenarmontdicintro.html. 2 Digital Imaging This section is based on the microscopy lectures by Jennifer Waters at the 2005 MBL Physiology course. Light can be detected and quantified in a variety of ways. In microscopy systems the most commonly used detection device is the charged-coupled device (CCD), a solid-stated based detector. As we will see CCDs allow for light detection with spatial discrimination. On the other hand, if one 5

Figure 7: Diagram of a DIC microscope [2] is interested in obtaining light from only one source, point detectors with no spatial discrimination such as photomultiplier tubes (PMTs) might be more appropriate. Charged-coupled devices consist of a dense matrix of photodiodes, each of them with a charge storage region. As photons reach each pixel electrons will be accumulated in the well. This process will go on for as long as the integration or exposure time lasts or until the well is saturated, which happens when a density of 1000electrons/µm 2 is reached. There are many parameters that are key to a camera s performance. The compromise between all of them will usually depend on the type of experiment. Ideally, the CCD spatial resolution has to be matched to the resolution of the microscope. The pixel size should be at least 1/2-1/3 of the Airy disk size. In figure 8 we show the relation between pixel size and spatial resolution. Image Pixels Digital Image Figure 8: Resolution limitations of a CCD camera (Jennifer Waters, Nikon Imaging Center, Harvard Medical). There are different noise sources in these systems. First, we can divide them into signal, optical 6

and camera noise. Any fluctuations in the object we are imaging falls into the category of signal noise. The fluctuations of photon counting statistics are included here. Optical noise consists of any photons detected by the camera that come from sources other than the area of interest in our field of view. By camera noise we understand all sources of change in the detector output that are not produced by light photons. The camera noise has two different components: dark and readout noise. Dark (or thermal) noise is caused by electrons that accumulate in the chip due to thermal fluctuations. This type of noise accumulates with exposure time and can be decreased by cooling the chip. Readout noise is related to errors in the digitalization of the signal. This noise is independent of the exposure time and can be reduced, for example, by binning the CCD. Binning consists in averaging the signal over a certain number of pixels. This will obviously lead to a loss in resolution, but since the signal is now digitalized only once for that set of pixels, the readout noise is decreased. Now we can turn to the analysis and visualization of acquired digital images. The first lesson here is that we should never judge an image by the way it looks on the screen. As an example, in figure 9 we present the same picture with different gray levels. The lesson to be learned here is that the differences are very subtle to the naked eye, which means that our eyes should not always be trusted when processing digital images. 256 Gray levels 64 Gray levels 16 Gray levels Figure 9: Same picture displayed with different gray levels (Butch Moomaw, Hamamatsu) Since a computer can only display 2 8 = 256 different gray levels, and most cameras can generate images of higher bit depth, images will have to be rescaled for proper viewing on a computer screen. Finally, we might want to perform some basic image processing on our images. The most commonly used parameters are brightness, contrast and gamma. All of these transformations consist of changing the mapping between image and displayed intensity. In figure 11, for example, we see how the brightness shifts the mapping up and down, saturating some of the pixels. The slope of this mapping can be changed by adjusting the contrast (fig. 12). Finally, it can also be useful not to use a linear mapping. We can obtain log-scale-like mapping by adjusting the image s gamma value. This allows us, for example, to give pixels with lower intensity a comparable importance with those of a higher intensity, as can be seen in figure 13. 7

Actual gray values 198-1265 Displaying 0-4095 Actual gray values 198-1265 Displaying 198-1265 Auto-scaled Actual gray values 198-1265 Displaying 236-546 Figure 10: Same image displayed with different rescalings (Jennifer Waters, Nikon Imaging Center, Harvard Medical) Raw Brightness JC Waters, 2005 Figure 11: Adjusting the brightness of a digital image (Jennifer Waters, Nikon Imaging Center, Harvard Medical). 3 Calibrating a Miscrocope Using a Resolution Target One of the first and easiest ways one can start being quantitative when approaching biology problems is to include scale bars in every image taken. As we will see throughout the first weeks of the course this will allow us to make all kinds of measurements and estimations not only about numbers and sizes, but also about composition and rates (once we include time in the picture, of course). In the lab we will use a standard resolution target [3], which can be used for both testing the quality of an optical system or calibrating it. In fig. 14 we present a picture of the resolution target and a table that shows how to read it. The target is separated into groups and each group is divided into different elements. For example, in the lower right corner you have group 0, element 1. Group 0 continues in the upper left corner with element 2 and so on. The table gives the density of lines in lines per millimeter, therefore the periodicity of the lines in each element is 1/density. 8

Raw Contrast JC Waters, 2005 Figure 12: Adjusting the contrast of a digital image (Jennifer Waters, Nikon Imaging Center, Harvard Medical). Raw Gamma JC Waters, 2005 Figure 13: Adjusting the gamma value of a digital image (Jennifer Waters, Nikon Imaging Center, Harvard Medical). When used for testing the quality of an optical system the idea is that diffraction (light getting diffracted from the feature s edges) and aberrations (mainly due to imperfections in the mirrors and lenses that are part of the system) are the ultimate limitations. The presence of simple periodic features makes it easy to determine when the pattern is starting to be affected by these effects. This resolution target has a range from 1mm to 4 µm, which is going to be reasonable for most applications in this lab. When using regular objectives one has to make sure that is the focus is not made through the target s glass, the side with the features should be facing the objective in order to reduce possible aberrations. When using an immersion oil objective a large cover slip should be put in between making sure that no oil gets on the target. Each magnification will have a suitable group to look at. A good check is to test the linearity of the microscope: is there a linear change in magnification when one changes objectives? Additionally, one might want to play with the focus and take different 9

Line Pairs per millimeter Element Group Number Number 0 1 2 3 4 5 6 7 1 1.00 2.00 4.00 8.00 16.00 32.00 64.00 128.00 2 1.12 2.24 4.49 8.98 17.96 35.92 71.84 143.70 3 1.26 2.52 5.04 10.08 20.16 40.32 80.63 161.30 4 1.41 2.83 5.66 11.31 22.63 45.25 90.51 181.00 5 1.59 3.17 6.35 12.70 25.40 50.80 101.60 203.20 6 1.78 3.56 7.13 14.25 28.51 57.02 114.00 228.10 Line pairs/mm = LP Line width (mm) = 1/(2LP) Space width (mm) = 1/(2LP) Line length = 5(line width) Figure 14: Picture of the resolution target and table that shows how to read it [3] pictures for different focus settings which are close to what appears to be the right one. In that way, when looking at the pictures later with a program like Photoshop, one can have an idea of the sensitivity of the system to small displacements of the focal plane. References [1] Eugene Hecht. Optics. Addison-Wesley, Reading, Mass., 4th edition, 2002. Eugene Hecht. ill.; 24 cm. [2] MicroscopyU. Nikon microscopyu (www.microscopyu.com). [3] Newport. Usaf-1951 test targets (www.newport.com). [4] E. D. Salmon and J. C. Canman. Proper aligment and adjustment of the light microscope. In Current Protocols in Immunology, pages 21.1.1 21.1.26. John Wiley & Sons, Inc., 2002. 10