Spectroscopy of Ruby Fluorescence Physics Advanced Physics Lab - Summer 2018 Don Heiman, Northeastern University, 1/12/2018

Similar documents
SECOND HARMONIC GENERATION AND Q-SWITCHING

Microscopy. The dichroic mirror is an important component of the fluorescent scope: it reflects blue light while transmitting green light.

Electronics - PHYS 2371/2

Instytut Fizyki Doświadczalnej Wydział Matematyki, Fizyki i Informatyki UNIWERSYTET GDAŃSKI

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region

Instructions for the Experiment

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

R. J. Jones Optical Sciences OPTI 511L Fall 2017

A Narrow-Band Tunable Diode Laser System with Grating Feedback

Spectroscopy Lab 2. Reading Your text books. Look under spectra, spectrometer, diffraction.

Nd: YAG Laser Energy Levels 4 level laser Optical transitions from Ground to many upper levels Strong absorber in the yellow range None radiative to

LEP Optical pumping

Optical Gain Experiment Manual

Chemistry 524--"Hour Exam"--Keiderling Mar. 19, pm SES

Experimental Physics. Experiment C & D: Pulsed Laser & Dye Laser. Course: FY12. Project: The Pulsed Laser. Done by: Wael Al-Assadi & Irvin Mangwiza

Ph 77 ADVANCED PHYSICS LABORATORY ATOMIC AND OPTICAL PHYSICS

FPPO 1000 Fiber Laser Pumped Optical Parametric Oscillator: FPPO 1000 Product Manual

ECE 340 Lecture 29 : LEDs and Lasers Class Outline:

Key Questions. What is an LED and how does it work? How does a laser work? How does a semiconductor laser work? ECE 340 Lecture 29 : LEDs and Lasers

Measuring Kinetics of Luminescence with TDS 744 oscilloscope

CHAPTER 5 FINE-TUNING OF AN ECDL WITH AN INTRACAVITY LIQUID CRYSTAL ELEMENT

COMPONENTS OF OPTICAL INSTRUMENTS. Chapter 7 UV, Visible and IR Instruments

COMPONENTS OF OPTICAL INSTRUMENTS. Topics

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

QE65000 Spectrometer. Scientific-Grade Spectroscopy in a Small Footprint. now with. Spectrometers

Examination Optoelectronic Communication Technology. April 11, Name: Student ID number: OCT1 1: OCT 2: OCT 3: OCT 4: Total: Grade:

EE119 Introduction to Optical Engineering Spring 2003 Final Exam. Name:

Luminous Equivalent of Radiation

University of Wisconsin Chemistry 524 Spectroscopic Components *

Chapter 14. Tunable Dye Lasers. Presented by. Mokter Mahmud Chowdhury ID no.:

Exp. No. 13 Measuring the runtime of light in the fiber

PHY385H1F Introductory Optics Term Test 2 November 6, 2012 Duration: 50 minutes. NAME: Student Number:.

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science

How is Light Absorbed and Transmitted?

EE119 Introduction to Optical Engineering Fall 2009 Final Exam. Name:

Chapter 16 Light Waves and Color

Observational Astronomy

Unit 8: Light and Optics

Experiment 1: Fraunhofer Diffraction of Light by a Single Slit

Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120)

FIBER OPTICS. Prof. R.K. Shevgaonkar. Department of Electrical Engineering. Indian Institute of Technology, Bombay. Lecture: 18.

PHY 431 Homework Set #5 Due Nov. 20 at the start of class

Cavity QED with quantum dots in semiconductor microcavities

HR2000+ Spectrometer. User-Configured for Flexibility. now with. Spectrometers

Instruction manual and data sheet ipca h

Vertical External Cavity Surface Emitting Laser

Doppler-Free Spetroscopy of Rubidium

Make Your Own Digital Spectrometer With Diffraction Grating

880 Quantum Electronics Optional Lab Construct A Pulsed Dye Laser

AP Chemistry Cell Phone Spectroscopy Lab Adopted from Alexander Scheeline Department of Chemistry University of Illinois at Urbana-Champaign

AgilOptics mirrors increase coupling efficiency into a 4 µm diameter fiber by 750%.

Exam 4. Name: Class: Date: Multiple Choice Identify the choice that best completes the statement or answers the question.

DIODE LASER SPECTROSCOPY (160309)

Basic concepts. Optical Sources (b) Optical Sources (a) Requirements for light sources (b) Requirements for light sources (a)

Solid-State Laser Engineering

Spectrometer using a tunable diode laser

You won t be able to measure the incident power precisely. The readout of the power would be lower than the real incident power.

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION

Kit for building your own THz Time-Domain Spectrometer

Renishaw InVia Raman microscope

EE119 Introduction to Optical Engineering Spring 2002 Final Exam. Name:

Photonics and Optical Communication Spring 2005

Components of Optical Instruments. Chapter 7_III UV, Visible and IR Instruments

Diffraction. Interference with more than 2 beams. Diffraction gratings. Diffraction by an aperture. Diffraction of a laser beam

Chemistry 61: Instrumental Analysis Building and Operating Your Own Diode Array Spectrometer

Educational Spectrophotometer Accessory Kit and System OS-8537 and OS-8539

Chapter Ray and Wave Optics

Lecture 6 Fiber Optical Communication Lecture 6, Slide 1

combustion diagnostics

Absorption: in an OF, the loss of Optical power, resulting from conversion of that power into heat.

IBIL setup operation manual for SynerJY software version

Optodevice Data Book ODE I. Rev.9 Mar Opnext Japan, Inc.

Spectroscopy in the UV and Visible: Instrumentation. Spectroscopy in the UV and Visible: Instrumentation

Maya2000 Pro Spectrometer

Introduction to the operating principles of the HyperFine spectrometer

LlIGHT REVIEW PART 2 DOWNLOAD, PRINT and submit for 100 points

Experiments in Photonics

LASERS. & Protective Glasses. Your guide to Lasers and the Glasses you need to wear for protection.

Be aware that there is no universal notation for the various quantities.

Test 1: Example #2. Paul Avery PHY 3400 Feb. 15, Note: * indicates the correct answer.

EDUCATIONAL SPECTROPHOTOMETER ACCESSORY KIT AND EDUCATIONAL SPECTROPHOTOMETER SYSTEM

3D light microscopy techniques

Physics 431 Final Exam Examples (3:00-5:00 pm 12/16/2009) TIME ALLOTTED: 120 MINUTES Name: Signature:

R.B.V.R.R. WOMEN S COLLEGE (AUTONOMOUS) Narayanaguda, Hyderabad.

DEVELOPMENT OF CW AND Q-SWITCHED DIODE PUMPED ND: YVO 4 LASER

Optical behavior. Reading assignment. Topic 10

Thesis. Submitted to. The School of Engineering of the UNIVERSITY OF DAYTON. In Partial Fulfillment of the Requirements for.

SA210-Series Scanning Fabry Perot Interferometer

Vision. The eye. Image formation. Eye defects & corrective lenses. Visual acuity. Colour vision. Lecture 3.5

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background

plasmonic nanoblock pair

Nd:YSO resonator array Transmission spectrum (a. u.) Supplementary Figure 1. An array of nano-beam resonators fabricated in Nd:YSO.

Tunable KiloArc. Tunable Broadband Light Source.

SUPPLEMENTARY INFORMATION

Period 3 Solutions: Electromagnetic Waves Radiant Energy II

Lasers PH 645/ OSE 645/ EE 613 Summer 2010 Section 1: T/Th 2:45-4:45 PM Engineering Building 240

Will contain image distance after raytrace Will contain image height after raytrace

CHAPTER 7. Components of Optical Instruments

PH 481/581 Physical Optics Winter 2014

AP B Webreview ch 24 diffraction and interference

Transcription:

1 Spectroscopy of Ruby Fluorescence Physics 3600 - Advanced Physics Lab - Summer 2018 Don Heiman, Northeastern University, 1/12/2018 I. INTRODUCTION The laser was invented in May 1960 by Theodor Maiman. This first laser was constructed of a cylindrical ruby crystal surrounded by a photographic flash lamp, all contained in a polished aluminum cylinder (on left). The flash lamp was used to excite the chromium ions in the sapphire host crystal. As the excited Cr(3+) ions de-excite they emit light as individual photons. Then as these photons travel back and forth in the optical cavity between the mirror-coated ends on the crystal, they induce other excited Cr ions to de-excite causing "stimulated emission." Rapidly, all of the ions become de-excited and generate a lasing light pulse. The light beam is coherent in the sense that the photons all travel in the same direction and have the same phase. LASER is an acronym for Light Amplification by Stimulated Emission of Radiation. This type of laser requires three energy levels, as shown in the diagram on the right. Absorbed pump light excites the Cr 3+ ion into excited states. The lifetime of these levels is short (50 ns), so that the excited ion quickly relaxes by making a transition to the long-lived metastable state. The energy which is lost in this process is nonradiative and goes into heating the crystal by generating phonons or vibrational excitations of the crystal atoms. The metastable energy level must have a lifetime which is long enough to enable the Cr ions to remain excited until a photon having the precise energy comes along to de-excite it. This lab experiment does not produce lasing in ruby, but investigates the excited states via the absorption spectrum and spontaneous fluorescence. The lasing line in ruby is the so-called "R-line" having a wavelength of ~0.7 μm. The fluorescence lifetime of the R-line is several ms. Fluorescence of the R-line can be excited by light in any of three absorption bands, at 250, 410, and 550 nm. A green laser used here for exciting the R-line fluorescence.

2 CAUTION Never look directly into any laser beam. Also, make sure there are no reflections that exit the bench top. Place unnecessary optical components to block stray beams from exiting the bench top. This green laser has a power of approximately 5 mw. Compare this to the 1 mw of sunlight that would enter your 1 mm diameter pupil if you looked directly at the sun. This experiment also introduces the widely used tool the spectrometer for optical spectroscopy. A spectrometer is a instrument used for measuring the intensity of light as a function of wavelength. Spectrometers usually contain a diffraction grating (or prism) to disperse the light, thereby spreading out the light of differing wavelengths into different positions. The spectrometer unit used here has an internal CCD (charged coupled device) silicon detector, essentially a digital camera detector, to measure the light intensity at various positions along its length. In the figure, light from an optical fiber enters at position #1; reflects off the collimating mirror #4; is diffracted by the grating #5; reflects from focusing mirror #6; then finally is read out by the CCD detector #9. For more information on Ocean Optics spectrometers see their website, http://www.oceanoptics.com/.. II. APPARATUS aluminum breadboard with 1/4-20 tapped holes green laser diode, ~5 mw @ 532 nm USB2000-FLG or USB4000FL Ocean Optics spectrometer, OceanView software, USB cable lens, 25 mm focal length, 25 mm diameter, for collecting and focusing fluorescence lens, 200 mm focal length, 12.5 mm diameter, for focusing the laser beam mirror, in x-y adjustable mount photodiode (PD) detector neutral-density (ND) optical filter (black) long-pass optical filter (red-orange) optical fibers, 50 and 600 micron core white light illuminator ruby crystal, Al 2 O 3 :Cr, approximately 0.05% Cr, length L=2.00±0.05 mm

3 III. PROCEDURE A. Room Light Spectrum 1. Start the OceanView software. Note the following: adjust Integration Time to keep the peaks below saturation; check the Electric Dark box to remove the background; if the data looks noise then increase the Scans to Average; use Copy to Clipboard then save the clipboard. 2. Insert one end of the 50 μm core fiber optic (FO) into the spectrometer input. Adjust the collection (integration) time so that the highest spectral peak is near but not above saturation. Store data in tab-delimited format after you click on the storage disk icon. Calibrate the spectrometer use the green laser to determine the shift in the wavelength. If the shift is >1 nm from the expected value, correct the wavelength of all subsequent spectra. 3. Now, point the open end of the FO at the wall or ceiling to collect stray room light. You should see many spectral peaks from the fluorescent room lights. Store the spectrum and include in report. What is the characteristic linewidth (full-width at half-maximum) of the narrowest spectral line? List in table form the wavelengths and intensities of the 5 or 6 strongest lines in counts/sec. Why are there separate spectral peaks? What are the colors of the strongest peaks. B. Absorption Spectrum The light intensity after passing through a material of length L is given by I = I o (1-R) 2 exp(-αl), where I o is the initial light intensity (see Appendix). The factor (1-R) 2 corrects for reflection losses from the two surfaces, where R = (n-1) 2 /(n+1) 2 is the reflectivity. Here, the absorption spectrum α(λ) of the ruby crystal will be measured. Make a new setup where a white light illuminates one end of the 600 μm core optical fiber, but leave a space large enough (10-20 cm) to slide in the ruby crystal just in front of the FO. Cover apparatus with black cloth to shield the room light. As before, adjust the collection time and the lamp/fo such that the spectrometer intensity is just below saturation. 1. Collect and store spectra I o (λ) without the ruby crystal, then collect I (λ) with the ruby crystal directly in front of the input end of the FO. 2. Roughly compute the transmission T=I / I o at a wavelength of λ~700 nm. Retake the measurements until T is in the range 80 % to 90 %. Assuming that α=0 at λ~700 nm and the accepted value for n, what is the expected T at λ~700 nm. Discuss. Plot I(λ) and I o (λ) on the same graph (after you subtract a constant background from each). Compute and plot the transmission spectrum, T(λ) = I(λ) / I o (λ).

4 Plot the spectrum of the absorption coefficient as a function of wavelength, α(λ). For the absorption maxima, what is the absorption length, 1/α, and uncertainty. Smooth out the noise and discuss the absorption peaks. What is wavelength and width of the main absorption peak? C. Optical Fluorescence Setup Mount the optical components on the breadboard approximately as shown in the diagram and photograph. Before aligning the beam, you must place the black neutral-density (ND) optical filter in front of the laser to reduce the beam intensity. D. Ruby Fluorescence Spectrum Collect laser-excited fluorescence from the ruby crystal and plot its spectrum. Optical Setup Before beginning, slide the black neutral-density (ND) filter in front of the laser to reduce the beam intensity. First, set all of the optical components to equal heights above the plate. Now the collection optics must be finely aligned. Insert one end of the 600 μm FO into the mount that points at the collection lens, and shine a white light source into the other end. Using the thin lens formula, compute the image and object distances for the f=25 mm focal length collection lens. While holding a white piece of paper at the front face of the ruby crystal, adjust the position of the lens and FO so that the white spot image is focused at the front face of the ruby crystal. Now adjust the mirror to overlap the green laser beam and the white spot image on the ruby crystal. Measurement Insert one end of the FO into the holder pointing at the collection lens and the other end into the spectrometer. Look for the R-line fluorescence in the spectrum. Maximize the intensity of the R-line emission by turning the adjusting screws on the mirror mount. Cover the breadboard with the black cloth. Store the spectrum and include in the report. Record the wavelength of the R-line. Discuss.

5 E. Fluorescence Lifetime of Ruby R-line This section describes measurement of the lifetime of the ruby R-line fluorescence. 1. Laser setup Insert the 200 mm focusing lens in the laser beam, in order to produce a smaller laser spot on the ruby crystal. Turn the diode laser power switch OFF (down). Connect a BNC cable from the TTL output of the function generator to the diode laser power supply. Connect another BNC cable from the Output of the function generator to channel-1 of the oscilloscope. Adjust the function generator to output a square wave having a period of 30 to 50 ms (select the 10 Hz pushbutton). Trigger the scope on the edge of the square wave. 2. Collection setup Use the FO to collect the fluorescence from the collection lens, and attach the other end of the FO to the photodiode (PD). Connect the BNC output of the PD to channel-2 of the scope. Turn the PD switch on (up). Next, remove the ND filter from the laser beam. Place the long-pass (red-orange) filter between the collection lens and the FO. This blocks (absorbs) the green laser light, while transmitting the red R-line. On the oscilloscope you should see a signal from the PD which is synchronous with the function generator. Capture one complete period of the PD output and the function generator for your report. Curve fit the decay of the PD signal to an exponential function. In the region of interest, plot the data as points and the curve fit as a solid line. What is the exponential decay time of the fluorescence? What is the uncertainty? IV. SUMMARY Discuss how you used the thin lens formula to optimize the collection of the fluorescence. Compute the photon energies (E) of the green laser line, the absorption peak, and the R-line. [E(eV)=1239.513/λ(nm)]. Make a TABLE listing the important values (with uncertainties) and reference the expected values. Always discuss the origin of the uncertainties. V. OPTIONAL Measure the index of refraction of the ruby crystal using a reflection from the surface.

6 V. APPENDIX A. Thin Lens Formula The thin lens formula is used here to determine the parameters for focusing the image of the ruby fluorescence into the optical fiber. Convex focusing lens are useful for projecting an image of an object from one side of a lens onto the other side. See the diagrams. The distance of an image from the lens is related to the distance of the object from the lens by the thin lens formula, 1/i + 1/o = 1/f, O i where i is the image distance, o the object distance, and f the focal length of the lens. The magnification of the lens is M=i/o. A special case of this formula is the focusing of a collimated laser beam. There the object distance is infinite, so the laser beam is focused to a small diameter spot at a distance i=f from the lens, shown in the lower diagram. f B. Light Absorption Materials which absorb light, such as the ruby crystal used here, have an absorption coefficient α which characterizes the depth to which light penetrates into the crystal. The light power P (or intensity I ) decreases as a function of increasing distance x into the material given by P(x) = P o exp(-αx), where P o is the initial power at x=0. The absorption length, 1/α, characterizes the penetration depth of the light. Light energy absorbed in the ruby crystal is converted into energy stored in the excited Cr ions (plus some energy lost which goes into heating the crystal). Note that light intensity is the light power divided by the cross sectional area of the light beam, I = P / A. C. Optical Excitation Each photon of green light absorbed by the ruby crystal raises the energy of a Cr ion to an excited state. This state rapidly relaxes to the long-lived metastable state having a lifetime t. Before the light is turned on, all the Cr ions are in the ground state. After turning on an excitation source, the number of excited ions increases. After time t of weak excitation, the number of excited ions is given by N (t) = N o [1- exp(- t / τ)], where N o is the number of ions which are excited after the crystal has been weakly illuminated for a long time t >> τ. Neglecting stimulated emission, the excitation process comes to equilibrium when the rate of excitation equals the rate of decay from the metastable state. The rate at which Cr ions are excited is P / hν, where hν = hc/λ is the photon energy. The rate of decay is 1/τ. At equilibrium, the number of excited Cr ions per unit volume is given by n o = α I τ/hν. After the exciting light is removed, the number of excited Cr ions decreases exponentially for increasing time as N (t) = N o exp (- t/τ).

7 D. Chromaticity Diagram All the possible colors of light which the human eye can distinguish can be made by adding relative amounts of the three additive primary colors: red, green and blue. These are related to the three wavelength-dependent photoreceptors in the human eye. These three colors are also used in TV and computer color monitors. In the mid-19 th century, J.C. Maxwell first described a diagram, the Maxwell triangle, to quantitatively represent all possible colors using the three primary colors. Spectra of the three particular primary colors are shown on the left. The Maxwell triangle has been updated into the universally accepted CIE chromaticity diagram, shown on the right. In this diagram, the x- and y-axis are the relative amounts of red and green light, respectively, and the amount of blue is 1-x-y. The outer rim of the tongue shape represents pure or saturated colors (hues), and is light of a single wavelength. Going inward from the edge towards the center is equivalent to adding white light, referred to as changing the tint. Pure white light is composed of an equal mixture of the three primary colors, x=y=z=0.33. Note that the light from the green laser used in the experiment (532 nm) lies on the upper, outer edge at x=0.21 and y=0.77. It is interesting that after dark you can see many variations of white lights that are distinguishable. http://www.efg2.com/lab/graphics/colors/chromatici ty.htm Spectra of the three colors (left), CIE 1931 Color Matching Functions (right).