Lecture 23 MNS 102: Techniques for Materials and Nano Sciences

Similar documents
Microscopy: Fundamental Principles and Practical Approaches

Introduction to Light Microscopy. (Image: T. Wittman, Scripps)

Imaging Introduction. September 24, 2010

VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES

Very short introduction to light microscopy and digital imaging

FLUORESCENCE MICROSCOPY. Matyas Molnar and Dirk Pacholsky

Resolution. Diffraction from apertures limits resolution. Rayleigh criterion θ Rayleigh = 1.22 λ/d 1 peak at 2 nd minimum. θ f D

Microscope anatomy, image formation and resolution

BASICS IN BIOIMAGING AND OPTICS PLATFORM EPFL SV PTBIOP LIGHT MICROSCOPY

Microscopy Training & Overview

Point Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy

Reflection! Reflection and Virtual Image!

ANSWER KEY Lab 2 (IGB): Bright Field and Fluorescence Optical Microscopy and Sectioning

PHYSICS. Chapter 35 Lecture FOR SCIENTISTS AND ENGINEERS A STRATEGIC APPROACH 4/E RANDALL D. KNIGHT

Microscopy. Matti Hotokka Department of Physical Chemistry Åbo Akademi University

Light microscopy BMB 173, Lecture 14, Feb. 21, 2018

The Nature of Light. Light and Energy

microscopy A great online resource Molecular Expressions, a Microscope Primer Partha Roy

SUBJECT: PHYSICS. Use and Succeed.

Heisenberg) relation applied to space and transverse wavevector

CCAM Microscope Objectives

Education in Microscopy and Digital Imaging

Applied Optics. , Physics Department (Room #36-401) , ,

INTRODUCTION THIN LENSES. Introduction. given by the paraxial refraction equation derived last lecture: Thin lenses (19.1) = 1. Double-lens systems


Basics of confocal imaging (part I)

Confocal Microscopy and Related Techniques

Motorized Axio Observer Start-up instructions

CCAM s Selection of. Zeiss Microscope Objectives

Script Bio 407 Applied Microscopy Light microscopy

Light Microscopy. Upon completion of this lecture, the student should be able to:

Katarina Logg, Kristofer Bodvard, Mikael Käll. Dept. of Applied Physics. 12 September Optical Microscopy. Supervisor s signature:...

OPELCO OPtical ELements COrporation LB Objective Series for Biological Use

Bio 407. Applied microscopy. Introduction into light microscopy. José María Mateos. Center for Microscopy and Image Analysis

Chapter 25. Optical Instruments

R.B.V.R.R. WOMEN S COLLEGE (AUTONOMOUS) Narayanaguda, Hyderabad.

LSM 510 META in Chang Gung University

Rates of excitation, emission, ISC

INTRODUCTION TO OPTICAL MICROSCOPY

Chapter Ray and Wave Optics

Final Exam, 150 points PMB 185: Techniques in Light Microscopy

Lecture 8. Lecture 8. r 1

BASICS IN LIGHT MICROSCOPY

Electron

Exercise 8: Interference and diffraction

Practical Flatness Tech Note

Advanced Optical Microscopy

Chapter 3 Op,cal Instrumenta,on

Chapter 16 Light Waves and Color

Systems Biology. Optical Train, Köhler Illumination

BEAM HALO OBSERVATION BY CORONAGRAPH

Basics of Light Microscopy and Metallography

Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE

Introduction to light microscopy

Applications of Optics

Observing Microorganisms through a Microscope LIGHT MICROSCOPY: This type of microscope uses visible light to observe specimens. Compound Light Micros

ELECTRON MICROSCOPY. 13:10 16:00, Oct. 6, 2008 Institute of Physics, Academia Sinica. Tung Hsu

BIOIMAGING AND OPTICS PLATFORM EPFL SV PTBIOP BASICS IN LIGHT MICROSCOPY

GIST OF THE UNIT BASED ON DIFFERENT CONCEPTS IN THE UNIT (BRIEFLY AS POINT WISE). RAY OPTICS

Tissue Preparation ORGANISM IMAGE TISSUE PREPARATION. 1) Fixation: halts cell metabolism, preserves cell/tissue structure

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy,

OPTICS GLOSSARY A BRIEF EXPLANATION OF CERTAIN OPTICAL TERMS PRE1305 OPTICAL MEASURING

Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal

Optical Components for Laser Applications. Günter Toesko - Laserseminar BLZ im Dezember

CHAPTER TWO METALLOGRAPHY & MICROSCOPY

Introduction to light microscopy

Lecture 2: Geometrical Optics. Geometrical Approximation. Lenses. Mirrors. Optical Systems. Images and Pupils. Aberrations.

The Compound Microscope. Brightfield: Köhler Illumination

a) How big will that physical image of the cells be your camera sensor?

Section A Conceptual and application type questions. 1 Which is more observable diffraction of light or sound? Justify. (1)

BASICS OF CONFOCAL IMAGING (PART I)

ELECTRON MICROSCOPY. 14:10 17:00, Apr. 3, 2007 Department of Physics, National Taiwan University. Tung Hsu

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

Chapter 3 Op+cal Instrumenta+on

General Physics II. Ray Optics

EE119 Introduction to Optical Engineering Spring 2003 Final Exam. Name:

DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS. GUI Simulation Diffraction: Focused Beams and Resolution for a lens system

Optical Design of. Microscopes. George H. Seward. Tutorial Texts in Optical Engineering Volume TT88. SPIE PRESS Bellingham, Washington USA

TSBB09 Image Sensors 2018-HT2. Image Formation Part 1

PHYSICS FOR THE IB DIPLOMA CAMBRIDGE UNIVERSITY PRESS

Will contain image distance after raytrace Will contain image height after raytrace

Characterization Methods. Prof.Dr.Figen KAYA A-209

Lecture 2: Geometrical Optics. Geometrical Approximation. Lenses. Mirrors. Optical Systems. Images and Pupils. Aberrations.

Supplementary Information for. Surface Waves. Angelo Angelini, Elsie Barakat, Peter Munzert, Luca Boarino, Natascia De Leo,

Prof. Enrico Gratton - Lecture 6 Fluorescence Microscopy

MICROSCOPE LAB. Resolving Power How well specimen detail is preserved during the magnifying process.

Development of a High-speed Super-resolution Confocal Scanner

ECEN 4606, UNDERGRADUATE OPTICS LAB

A broadband achromatic metalens for focusing and imaging in the visible

Microscopy. Lecture 2: Optical System of the Microscopy II Herbert Gross. Winter term

Invitation for a walk through microscopy. Sebastian Schuchmann Jörg Rösner

Physics 1C. Lecture 25B

Cardinal Points of an Optical System--and Other Basic Facts

Fundamentals of Light Microscopy II: Fluorescence, Deconvolution, Confocal, Multiphoton, Spectral microscopy. Integrated Microscopy Course

OPTICAL SYSTEMS OBJECTIVES

Flatness of Dichroic Beamsplitters Affects Focus and Image Quality

Microscopy Techniques that make it easy to see things this small.

Differential Interference Contrast (DIC) Verses Dark Field and Phase Contrast Microscopy. E. D. Salmon University of North Carolina at Chapel Hill

Mirrors and Lenses. Images can be formed by reflection from mirrors. Images can be formed by refraction through lenses.

PHYS 202 OUTLINE FOR PART III LIGHT & OPTICS

Transcription:

Lecture 23 MNS 102: Techniques for Materials and Nano Sciences Reference: #1 C. R. Brundle, C. A. Evans, S. Wilson, "Encyclopedia of Materials Characterization", Butterworth-Heinemann, Toronto (1992), Ch. 2.0, 2.1 Reference: Taewoo Lee et al. Optical Microscopy of Soft Matter Systems arxiv:1108.3287 - http://arxiv.org/ftp/arxiv/papers/1108/1108.3287.pdf Reference: D.B. Murphy, Fundamentals of Light Microscopy and Electronic Imaging", Wiley, New York (2001). http://www.biology.uoc.gr/courses/biol493/documents/book.pdf Light Microscopy - Overview Basic concepts: Numerical Aperture (NA), working distance (WD), resolution (R), magnification (M), depth of field or depth of focus (DOF) Types: Bright Field (BF), Dark Field (DF), Polarized Light, Differential Interference Contrast (DIC), Phase Contrast (PC). Fluorescence Microscopy Human eye can resolve ~ 0.1 mm or 100 microns 23-1

Wavelength sets Information Limit c =, where c = 3 x 10 10 cm/s E = h, where h = 6.62 x 10-27 erg s E = h c ( ) -1, where ( ) -1 is the wavenumber 23-2

Thin Lens Optics In thin lens optics Magnification (M) is defined by the ratio of the image distance to object distance, i.e., d i / d O d 0 d i Human eye ~ 0.1 mm Apparent resolvable distance by the eye for a magnified object is: R = 0.1 mm / M M R 1 100,000 nm 100 1,000 nm 1,000 100 nm 10,000 10 nm 100,000 1 nm 1,000,000 0.1 nm 23-3

Spherical aberration: Lens curvature deviates from the ideal shape, which causes different focal distances. Chromatic aberration: Lens focusses each wavelength at a different spot, but can be corrected by achromat, fluorite, or apochromat materials. Field of view flatness Immersion media Lens Imperfection 23-4

Optical Microscopes: History Saw single-cell organism in pond water Saw pores in cork cells 23-5

Basic Components > Illumination source: transmitted light vs reflected light > Condenser Lens > Specimen > Objective Lens > Viewer (eyepiece or CCD) 23-6

A Simple Light Microscope 23-7

Objective 23-8

Numerical Aperture & Working Distance Performance of an objective lens is defined by NA and WD. Numerical Aperture (NA) of an objective determines the range of angles over which the objective lens could accept light. NA = n sin θ, where n = refractive index of the medium between the objective lens and the sample (1 for air, 1.33 for water, 1.56 for oils) and = half-angle of the cone of light collected by the objective lens. NA < 1 for air objectives ; NA 1.5 for oil immersion objectives. Working Distance (WD) is the distance between the front lens of the objective to the specimen surface when the inspected area is in focus. This defines the max. specimen height. 23-9

Spatial Resolution set by Rayleigh Criterion Resolution = shortest distance between two points that can still be separated = k / NA where k = 0.61 (Rayleigh) or 0.5 (Abbe) NA = Numerical Aperture = n sin where n = refractive index of the medium, and = semi-angle (half-angle) of an objective lens; NA typically < 1 θ θ θ Effect of NA on the image of a point. Resolution achieved by separation of Airy disks, when max of 1 st disk overlaps first min of 2 nd disk (Rayleigh). 23-10

Resolution and PSF NA defines Point Spread Function (PSF), which is light intensity distribution in the image acquired by the microscope from a point source. PSF(r) = [ 2 J 1 (r a) / (r a) ] 2 where a = 2 (NA) / and J 1 is the Bessel function of the first kind. Airy diffraction pattern gives the first minimum at r = W/2. For two Airy discs to be resolved, they must be separated by R W/2. R is the lateral resolution and R = 0.61 / NA is the diffraction limit. Higher NA, better resolution but note R is also limited by the wavelength. Smaller WD > larger > larger NA > smaller R (i.e. better resolution). Conversely, large WD > poorer resolution. 23-11

Magnification Magnification is the enlargement of the incoming angular acceptance. M = tan / tan 0 / 0 d 0 d i For a typical stereo microscope with: (a) a 0.5X objective and a 1.0X objective, (b) a zoom range of 0.7 to 11.5X and (c) a 0.5X camera adapter, and the image is collected on a 2.2 megapixel camera (with a 1/1.8" or 0.55556 CCD chip) and projected onto a 24 monitor. Total Magnification = (Objective Mag) x (Zoom Mag) x (Camera Adapter Mag) x (Monitor Size/CCD Chip Size) Low end: Total Mag = 0.5 x 0.7 x 0.50 x (24/0.55556) = 7.6 High end: Total Mag = 1.0 x 11.5 x 0.5 x (24/0.5556) = 248.4 If we just have an eyepiece with a 10X mag, then we have 3.5 (low end) and 115 (high end). Olympus SZX16 Stereo microscope 23-12

Depth of Focus or Depth of Field (DOF) Depth of Focus refers to the image, while Depth of Field refers to the specimen. Depth of Field (or vertical resolution) measures the ability to produce a sharp image from a non-flat surface. DOF /NA for a set magnification, & can be increased by inserting an objective aperture. Light Microscope Electron Microscope Wavelength = 500 nm (150/V 0 ) = 0.0055 nm at 50 kv Refraction Index = n 1.5 (glass) 1.0 (vacuum) Half-angle = 70 deg 1 deg Resolution = 0.61 / NA where NA = n sin Depth of Field (DOF) = distance parallel to the optical axis that a feature on the specimen can be displaced without loss of resolution. 200 nm 0.16 nm DOF = λ n2 NA 2 + 250 NA 2 M 2 M = 10, DOF = 60 m M = 100, DOF = 8 m M = 1,000, DOF = 200 nm 0.1 mm DOF = M θ M = 10, DOF = 1,000 m M = 100, DOF = 100 m M = 1,000, DOF = 10 m M = 10,000, DOF = 1 m 23-13

R vs NA Higher NA > better (smaller) R For the same NA: Smaller > better R 23-14

DOF vs NA Higher NA > poorer (smaller) DOF For the same NA: Smaller > poorer DOF Ideally, we want small R (i.e. high resolution) and large DOF and this is mutually exclusive. 23-15

Homework 5A: Work through the following site: http://micro.magnet.fsu.edu/primer/anatomy/numaperture.html. When NA is increased, what happens to (a) the working distance, and (b) the brightness of the image in focus. To resolve two points separated by 1,000 nm (i.e. 1 micron) using light of 500 nm wavelength, what is the smallest NA required? Can one resolve these two points with light of 700 nm wavelength? (Use the Java app at the above site.) 23-16

Types of Light Microscopy Based on contrast modes and/or illumination methods: Bright Field a conventional light microscope Dark Field Polarized Light Phase contrast Fritz Zemike, Nobel Prize in Physics, 1953 Differential Interference Contrast (DIC) Georges Nomarski, 1955 Fluorescence Except for Fluorescence Microscopy, all five techniques involve transmission, absorption, refraction or scattering of light. 23-17

Bright Field vs Dark Field Microscopy Bright Field: Image comes from direct interaction of the specimen with unpolarized white light. Processes include absorption, refraction, scattering, and reflection. Pro: Easy to perform; good contrast of dark colours. Con: Bad contrast of light colours (biological samples); overillumination. Dark Field: Image is formed by collecting only light scattered by the specimen. Oblique illumination. Direct, unscattered light is blocked. Typically used for specimen with sparse scattering objects dispersed in a non-scattering medium. Pro: Effective on highly transmitting specimens (biological samples), less artifacts (no halos). Con: Intense illumination needed and this could cause specimen damage, blind for low scattering samples. 23-18

Köhler Illumination Kohler illumination is a technique used to provide extremely even specimen illumination and eliminate the image of the illumination source. This is achieved by making the image of light source perfectly defocussed in the sample and image planes. Homework 5B: In less than one page, discuss the difference between Kohler illumination and critical illumination. Explain how this can be achieved in an optical microscope. 23-19

Optics Setup 23-20

Polarization of Light and Wollaston Prisms 23-21

Polarized Light Microscopy A polarizer is used to create linearly polarized light for illuminating the specimen, and a second polarizer called the analyser to infer the polarization properties of the materials. When the polarizer and analyser is crossed, the transmitted light intensity is I perpendicular = I 0 sin 2 (2 ) sin 2 ( ) where is the angle between the polarization vector and the optical axis of the birefringent material, and is the phase (= 2 n d/ ) introduced by the thickness of the material d and n is the birefringence. When the polarizer and analyser is parallel, the transmitted light intensity is I parallel = I 0 [1 sin 2 (2 ) sin 2 ( )] Linearly polarized light is split by the birefringent specimen into two components (extraordinary and ordinary), propagating at different speeds and resulting in elliptical polarization. Use pattern to deduce spatial variations of optical axis orientation and the value of n. Birefringent materials are optically anisotropic materials that have a refractive index that depends on the polarization and propagation direction of light. E.g. SiC, many plastics, CaCO 3, ice, rutile TiO 2. 23-22

Differential Interference Contrast (DIC) & Phase Contrast (PC) Contrast is caused by difference in reflection, absorption, refraction, polarization, and other optical processes. DIC and PC are important for enhancing contrast in transparent specimen (e.g. cell). PC originates from the small phase shifts of light passing through a transparent specimen. DIC corresponds to interferometry due to slopes and valleys of the specimen, producing differences in the optical paths. In effect, the gradient of the refractive index within the specimen produces the DIC. BF (upper) vs DIC (lower) 23-23

Phase Contrast Change the phase by delaying or advancing a quarter wavelength. 23-24

PC Microscopy Small spatial variations in phase are transformed into corresponding changes in the intensity of transmitted light. Light passes through an annular ring before the condenser lens and is focussed onto the specimen. The light can either pass through the specimen undeviated (yellow) or diffracted with a changed phase (violet) depending on the composition of the specimen. Both undeviated and diffracted beams are collected by the objective and transmitted through a phase ring to the eyepiece. The phase ring introduces an additional phase shift to the undeviated light used as the reference. Pro: Phase contrast imaging is not sensitive to polarization and birefringence effects. Good for living cells and soft materials. 23-25

Differential Interference Contrast 23-26

DIC Microscopy White light is sent to the first polarizer. The polarized light is separated into two orthogonally polarized and spatially displaced (sheared) components by the first beam splitter (Nomarski-modified Wollaston prism). Each component propagates a different path, leading to different optical path difference after existing the sample. After the objective, the two components are recombined by a second beam splitter. The polarization of the resulting light is then analysed by the second polarizer (the analyser). The result is that the optical path length gradient along the shear direction will enhance contrast due to edges and interfaces of the specimen with different refractive indices. More gradient > more contrast. 23-27

Transmitted Light vs Reflected Light Illumination Thus far, we talk about transmitted light illumination, i.e. light from the illumination source passes through the specimen and is detected on the other side (the objective side). Here, light absorption is an important process, along with diffraction. We can also use reflected light illumination, i.e. light from the illumination source comes from the top of the specimen (not below the specimen). Here, light reflection (along with diffraction) is an important process b/c the objective is on the same side of the light illumination. Reflected light illumination is more popular in commercial microscopes. 23-28

BF vs DF Reflected Light Illumination Setup 23-29

PL & DIC Reflected Light Illumination Setup 23-30

Contrast Modes vs Image Quality BF DIC PL DF 23-31

Fluorescence Use of fluorescence in microscopy is very important to quantitative biology. Images show locations of the fluorescent markers. All non-stained tissue remains transparent. Amount of fluorescence can be measured accurately. This has led to recent advances in super-resolution microscopy, with resolution going below 60 nm. 23-32

Fluorescent Materials Small organic dye molecules Fluorescent proteins Green FP - M. Chalfie, O. Shimomura, R.Y. Tsien, Nobel Prize in Chemistry 2008 Quantum dots 23-33

Glow-stick & Chemiluminescence Oxidation of an diphenyl oxalate (top), decomposition of 1,2-dioxetanedione (middle), relaxation of dye (lower). Dye could be carcinogenic. 23-34

Fluorescence Microscopy Based on fluorescence phenomenon: (a) absorption of light by fluorescent dye molecules or fluorophores at a specific wavelength, thereby promoting to an excited state, followed by (b) emission of light at a longer wavelength. The lifetime of the excited state is usually in nanoseconds. The shift between the absorption peak and emission peak is called the Stokes shift and it is caused by nonradiative processes. This allows the use of optical filters to separate the excitation and emission signals. A filter cube contains an excitation filter, a dichroic mirror and an emission filter. Dichroic mirror is a very accurate mirror that reflects light of a small range of colours (wavelengths) while passing the other colours. 23-35

Epifluorescence Microscope Thick section of human skin: green = anti-basal lamina proten; red = neuronal processes; blue = collagen and elastin auto fluoresce Source: www.zeiss.com Pollen 23-36

Homework 5C: Discuss in one page or less, the difference between epifluorescence and transfluorescence. Identify the advantage of the former. 23-37