CALIBRATION PROCEDURE PXIe-4302/4303 and TB-4302C 32 Ch, 24-bit, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module. ni.

Similar documents
NI PXIe Contents CALIBRATION PROCEDURE. Four Channel Precision Source-Measure Unit (SMU)

NI PXIe Contents. Required Software CALIBRATION PROCEDURE. Dual-Output Programmable DC Power Supply

NI PXIe Contents CALIBRATION PROCEDURE. Single-Channel Precision Source-Measure Unit (SMU)

PXIe Contents. Required Software CALIBRATION PROCEDURE

CALIBRATION PROCEDURE NI PXIe-4330/4331. Contents

Contents. CALIBRATION PROCEDURE NI PXIe-4463 DSA Analog Output

Contents. CALIBRATION PROCEDURE NI PXIe-5698

PXIe, 7½-Digit, ±1,000 V, Onboard 1.8 MS/s Isolated Digitizer, PXI Digital Multimeter

Contents. Software Requirements

Calibrating the NI 5653 requires you to install one of the following packages on the calibration system. NI-RFSA 2.4 or later NI-RFSG 1.

Contents. Software Requirements. Documentation Requirements. Calibration Interval CALIBRATION PROCEDURE NI TB-4300B

Contents. Software. Documentation. CALIBRATION PROCEDURE NI PXIe-6674T

NI PXIe-5171R. Contents. Required Software CALIBRATION PROCEDURE

NI 6624 Calibration Procedure

GETTING STARTED GUIDE NI AI, ±5 V, 24 Bit, 51.2 ks/s/ch Simultaneous, AC/DC Coupling, IEPE AC Coupling

GETTING STARTED GUIDE NI khz to 100 MHz, 50 Ω, AC Coupled RF Receiver Module

Required Software. Related Documentation. Password. Calibration Interval

PXIe Contents CALIBRATION PROCEDURE. 10 GHz or 20 GHz RF Analog Signal Generator

Contents. Software Requirements. CALIBRATION PROCEDURE NI PXIe-5663E

Contents. Software Requirements CALIBRATION PROCEDURE NI PXI-5663

GETTING STARTED GUIDE NI AI, ±10 V, 24 Bit, 50 ks/s/ch Simultaneous

Contents. CALIBRATION PROCEDURE PXIe-5673 Vector Signal Generator

Power Amplifier. Input Characteristics. SPECIFICATIONS NI PXIe-4610 Power Amplifier. Signal Range

if it is used in a manner not described in the user documentation.

NI sbrio-9632/9642 Verification Procedure

PXIe, 24-Channel, ±24 V, 50 ma Precision PXI Source Measure Unit

is used in a manner not described in the user documentation.

GETTING STARTED GUIDE NI AI, ±60 V, 24 Bit, 1 ks/s/ch Simultaneous

GETTING STARTED GUIDE NI AI Differential/32 AI Single-Ended, ±200 mv to ±10 V, 16 Bit, 250 ks/s Aggregate, Fuel-Cell Measurements

These specifications apply to the PXIe-5113 with 64 MB and 512 MB of memory.

DEVICE SPECIFICATIONS NI PXIe Ch, 24-bit, 25.6 ks/s Simultaneous AC LVDT Input Module

PXIe Contents. Required Software CALIBRATION PROCEDURE

DEVICE SPECIFICATIONS NI PXIe-4302/ Ch, 24-bit, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module

Input Characteristics. Measurement Range. SPECIFICATIONS NI PXIe-4357

DEVICE SPECIFICATIONS NI PXIe-4304/ Ch, 24-bit, ±42 V, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module. ni.

PXIe Contents SPECIFICATIONS. 32-Channel Digital Pattern Instrument

Contents. CALIBRATION PROCEDURE NI PXIe-6555/6556. ni.com/manuals

Note Using the PXIe-5785 in a manner not described in this document might impair the protection the PXIe-5785 provides.

PXIe Contents SPECIFICATIONS. 100 MVector/s PXI Digital Pattern Instrument

Terminology. Output Characteristics. DEVICE SPECIFICATIONS NI PXIe-4463 DSA Analog Output. ni.com/manuals

SCXI Terminal Block

Contents. CALIBRATION PROCEDURE NI PXIe-5668R 14 GHz and 26.5 GHz Signal Analyzer

SPECIFICATIONS PXIe Ch (8 ADC), 2 MS/s, bit, Flexible Resolution PXI Analog Input Module

SPECIFICATIONS PXIe channel, 16-bit, 400 ks/s/ch, Ch-Ch Isolated Analog Input Module. ni.com/manuals

Analog Characteristics

Contents. CALIBRATION PROCEDURE NI PXIe GHz and 14 GHz RF Vector Signal Analyzer

Current ON state current ma minimum, 6 ma typical, 10 ma maximum OFF state current ma maximum

SCXI TERMINAL BLOCK

Contents. USER MANUAL NI ISM-7400 Integrated Stepper

Analog Input. DEVICE SPECIFICATIONS NI PXIe-4300

NI PXI Contents DEVICE SPECIFICATIONS. RF Attenuator

SCXI HIGH-VOLTAGE ATTENUATOR TERMINAL BLOCK

OPERATING INSTRUCTIONS NI Port, Low-Speed/Fault Tolerant CAN Module

Reconfigurable 6 GHz RF Vector Signal Transceiver with 1 GHz Bandwidth

SPECIFICATIONS PXIe ks/s, 119 db, 6 Gains, AC/DC-Coupled, 4-Input PXI Sound and Vibration Module

USRP-2940/2942/2943/2944/2945

NI 6143 Specifications

This document lists the specifications for the NI PXI The PXI-5900 is a differential amplifier.

PXIe Contents SPECIFICATIONS. RF Amplifier

NI PXIe-5667 (3.6 GHz)

SA1500X. Operating Basics. Cautions DEVICE SPECIFICATIONS. High Impedance Active Probe

100 MHz Bandwidth, 12-Bit PXI Waveform Generator

Contents. CALIBRATION PROCEDURE NI 5421/ MS/s Arbitrary Waveform Generator

NI PXIe-5630 Specifications

OPERATING INSTRUCTIONS AND SPECIFICATIONS. NI 9211E 4-Channel Thermocouple Input Module

DEVICE SPECIFICATIONS Phase Matrix PXI GHz PXI Vector Signal Analyzer

NI PXI/PXIe-2541 Specifications 300 MHz Ω Matrix

NI PXI/PXIe-2540 Specifications

PXIe Contents CALIBRATION PROCEDURE. Reconfigurable 6 GHz RF Vector Signal Transceiver with 200 MHz Bandwidth

NI PXI-2521 Specifications 40-Channel DPST Relay Module

SCXI MHz Quad 8x1 50 Ω Multiplexer

PXIe Contents SPECIFICATIONS. RF Preselector Module

NI PXI-2797 Specifications 40 GHz 6 1 Terminated 50 Ω Multiplexer

NI PXI/PCI-5412 Specifications 14-Bit 100 MS/s Arbitrary Waveform Generator

NI PXI-2520 Specifications 80-Channel SPST Relay Module

NI PXIe Contents. About These Specifications DEVICE SPECIFICATIONS. 3 GHz Ω Multiplexer (SP16T)

NI REM Remote I/O Overview DATASHEET. Temperature Input Module for Remote I/O

NI PXI-4110 Calibration Procedure Addendum

NI PXI/PXIe-2529 Specifications

NI PXI-5422 Specifications 16-Bit 200 MS/s Arbitrary Waveform Generator

PXIe, 6½-Digit, ±300 V, Onboard 1.8 MS/s Isolated Digitizer,PXI Digital Multimeter

4-Channel, ±24 V, 150 ma Precision PXI Source Measure Unit

Contents CALIBRATION PROCEDURE NI PXI-5422

6 1 2-Digit Digital Multimeter, 1.8 MS/s Isolated Digitizer, and LCR Meter

4-Channel, ±24 V, 150 ma PXI Source Measure Unit

CDA Contents. Definitions SPECIFICATIONS. Clock Distribution Device

NI DAQPad -6020E Family Specifications

4-Channel, ±6 V, 500 ma Precision PXI Source Measure Unit

4-Channel ±10 V, 100 ma, Precision PXI Source Measure Unit

4-Channel ±10 V, 100 ma, Precision PXI Source Measure Unit

NI PXI/PXIe-2527 Specifications

OPERATING INSTRUCTIONS AND SPECIFICATIONS. NI 9207E 16-Channel, ±20 ma/±10 V, 24-Bit Analog Input Module

R Series for USB Multifunction RIO with Kintex-7 70T FPGA

PXIe Contents SPECIFICATIONS. 250 khz to 20 GHz RF Analog Signal Generator

PXIe Contents. Definitions SPECIFICATIONS. 8.5 GHz Vector Network Analyzer

NI 6023E/6024E/6025E Family Specifications

NI 2865A 0.3 A Matrix Cards for NI SwitchBlock

PXIe Contents. Definitions SPECIFICATIONS. 8.5 GHz Vector Network Analyzer

SC Express. PXIe-4309 User Manual. 32 Ch (8 ADC), 2 MS/s, bit, Flexible Resolution PXI Analog Input Module. PXIe-4309 User Manual

PXI Contents. Definitions SPECIFICATIONS. ±100 V, 2 W Precision PXI Source Measure Unit. These specifications apply to the PXI-4132.

Transcription:

CALIBRATION PROCEDURE PXIe-4302/4303 and TB-4302C 32 Ch, 24-bit, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module Français Deutsch ni.com/manuals This document contains the verification and adjustment procedure for the National Instruments PXIe-4302/4303 module and the verification procedure for the National Instruments TB-4302C terminal block. Contents Software... 1 Documentation... 2 PXIe-4302/4303 Verification and Adjustment... 3 Test Equipment... 3 Connecting the TB-4302... 4 Test Conditions...5 Initial Setup... 6 Accuracy Verification... 6 Adjustment... 8 EEPROM Update... 10 Reverification... 10 TB-4302C Verification... 10 Test Equipment... 10 Accuracy Verification... 11 Specifications... 14 World Wide Support and Services... 14 Software Calibrating the PXIe-4302/4303 requires the installation of NI-DAQmx on the calibration system. Driver support for calibrating the PXIe-4302/4303 was first available in NI-DAQmx 15.1. For the list of devices supported by a specific release, refer to the NI-DAQmx Readme, available on the version-specific download page or installation media.

You can download NI-DAQmx from ni.com/downloads. NI-DAQmx supports LabVIEW, LabWindows /CVI, C/C++, C#, and Visual Basic.NET. When you install NI-DAQmx, you only need to install support for the application software that you intend to use. No other software is required to verify the operation of the TB-4302C. Documentation Consult the following documents for information about the PXIe-4302/4303, NI-DAQmx, and your application software. All documents are available on ni.com, and help files install with the software. NI PXIe-4302/4303 and TB-4302/4302C User Guide and Terminal Block Specifications NI-DAQmx driver software installation and hardware setup. NI PXIe-4302/4303 User Manual PXIe-4302/4303 usage and reference information. NI PXIe-4302/4303 Specifications PXIe-4302/4303 specifications and calibration interval. NI-DAQmx Readme Operating system and application software support in NI-DAQmx. NI-DAQmx Help Information about creating applications that use the NI-DAQmx driver. LabVIEW Help LabVIEW programming concepts and reference information about NI-DAQmx VIs and functions. NI-DAQmx C Reference Help Reference information for NI-DAQmx C functions and NI-DAQmx C properties. NI-DAQmx.NET Help Support for Visual Studio Reference information for NI-DAQmx.NET methods and NI-DAQmx.NET properties, key concepts, and a C enum to.net enum mapping table. 2 ni.com PXIe-4302/4303 and TB-4302C Calibration Procedure

PXIe-4302/4303 Verification and Adjustment This section provides information for verifying and adjusting the PXIe-4302/4303. Test Equipment Table 1 lists the equipment recommended for the performance verification and adjustment procedures of the PXIe-4302/4303. If the recommended equipment is not available, select a substitute using the requirements listed in Table 1. Table 1. Recommended Equipment for PXIe-4302/4303 Verification and Adjustment Equipment Recommended Model Requirements DMM PXI-4071 Use a DMM that has an accuracy of 13 ppm or better when measuring the 10 V range, an accuracy of 30 ppm or better when measuring the 100 mv range, and an offset error of 0.8 V or better at 0 V. PXI Express Chassis PXIe-1062Q If this chassis is unavailable, use another PXI Express chassis, such as the PXIe-1082 or PXIe-1078. Connection Accessory TB-4302 SMU PXIe-4139 Noise (0.1 Hz to 10 Hz, peak to peak) is 60 V or better at 10 V. Noise (0.1 Hz to 10 Hz, peak to peak) is 2 V or better at 100 mv. PXIe-4302/4303 and TB-4302C Calibration Procedure National Instruments 3

Connecting the TB-4302 The TB-4302 provides connections for the PXIe-4302/4303. Figure 1 shows the pin assignments of the TB-4302. Figure 1. TB-4302 Circuit Board Parts Locator Diagram R1 R5R6 R4 CR2 C4 R3 R8 C6 CR1 U4 S/N NI TB-4302 COPYRIGHT 2015 C2 U2 R2 U3 R7 C1 U1 C3 C5 UL94V-0 1 J2 8 1 J4 8 158016B-01L AI5+ AI2+ AI1+ AI0+ AIGND J5 AIGND J3 J9 J7 AIGND AIGND J6 J8 AI15+ AI14+ AI13+ AI12+ AI10+ AI9+ AI8+ 8 1 8 1 U6 U5 U7 8 J10 J12 1 1 8 CR3 AI23+ AI22+ AI21+ AI20+ AI18+ AI7+ AI6+ FOR PATENTS: NI.COM/PATENTS AIGND AIGND J11 J13 J18 J17 J15 DGND PFI0 AIGND AIGND J14 J16 AI4- AI4+ AI3- AI3+ AI2+ AI5- AI7- AI7+ AI6- AI6+ AI0- AI1- AI9- AI13- AI15- AI14- AI11+ AI10- AI11- AI8- AI12- AI7- AI21- AI23- AI22- AI19+ AI18- AI19- AI6- AI20- AI31- AI31+ AI30- AI30+ AI29- AI29+ AI28- AI28+ AI17- AI17+ AI16- AI16+ AI5- AI5+ AI4- AI4+ 8 1 8 1 Each channel consists of two terminal connections specific to that channel as shown in Table 2. You can verify or adjust the accuracy for any or all channels depending on the desired test coverage. Refer to Figure 2 and only connect the input channels required for verification or adjustment in parallel. Refer to Table 2 for the analog signal names of the TB-4302. Table 2. TB-4302 Analog Signal Names Signal Name AI+ AI- AIGND Signal Description Positive input voltage terminal Negative input voltage terminal Analog ground input Complete the following steps to connect the TB-4302. 1. Install the PXIe-4302/4303 and the TB-4302 in the PXI Express chassis according to the instructions in the NI PXIe-4302/4303 and TB-4302/4302C User Guide and Terminal Block Specifications. 2. Configure the PXIe-4139 to voltage output mode and enable remote sensing. Connect the PXIe-4139 output to the TB-4302 as shown in Figure 2. 4 ni.com PXIe-4302/4303 and TB-4302C Calibration Procedure

3. Use two 10 k resistors with 1% or better tolerance to build a voltage divider to bias the PXIe-4139 output and set the common-mode input of the PXIe-4302/4303 to zero volts. Connect one resistor between AI+ and AIGND and the other between AI- and AIGND as shown in Figure 2. 4. Connect the PXI-4071 to measure the differential voltage across the TB-4302 AI+ and AIterminals. A detailed wiring diagram is shown in Figure 2. Figure 2. Connecting the TB-4302 HI SENSE HI AI0+ 4139 LO SENSE LO AI0 AI1+ AI1 HI AI31+ 4071 LO AI31 TB-4302 10 kω 10 kω Test Conditions The following setup and environmental conditions are required to ensure the PXIe-4302/4303 meets published specifications. Keep connections to the PXIe-4302/4303 as short as possible. Long cables and wires act as antennas, picking up extra noise that can affect measurements. Verify that all connections to the TB-4302 are secure. Use shielded copper wire for all cable connections to the TB-4302. Use twisted-pair wire to eliminate noise and thermal offsets. Maintain an ambient temperature of 23 C ±5 C. The PXIe-4302/4303 temperature will be greater than the ambient temperature. Keep relative humidity below 80%. Allow a warm-up time of at least 15 minutes to ensure that the PXIe-4302/4303 measurement circuitry is at a stable operating temperature. Ensure that the PXI/PXI Express chassis fan speed is set to HIGH, that the fan filters are clean, and that the empty slots contain filler panels. For more information, refer to the Maintain Forced-Air Cooling Note to Users document available at ni.com/manuals. PXIe-4302/4303 and TB-4302C Calibration Procedure National Instruments 5

Initial Setup Refer to the NI PXIe-4302/4303 and TB-4302/4302C Installation Guide and Terminal Block Specifications for information about how to install the software and hardware and how to configure the device in Measurement & Automation Explorer (MAX). Note When a device is configured in MAX, it is assigned a device identifier. Each function call uses this identifier to determine which DAQ device to verify or to verify and adjust. This document uses Dev1 to refer to the device name. In the following procedures, use the device name as it appears in MAX. Accuracy Verification The following performance verification procedures describe the sequence of operations and provides the test points required to verify the PXIe-4302/4303. The verification procedures assume that adequate traceable uncertainties are available for the calibration references. The PXIe-4302/4303 has 32 independent analog input channels. The input range of each channel can be set to 10 V or 100 mv. You can verify the accuracy of either range for any or all of the channels depending upon your desired test coverage. Complete the following steps to verify the voltage mode accuracy of the PXIe-4302/4303. 1. Set the PXIe-4139 voltage output to zero volts. 2. Connect the PXIe-4139 and PXI-4071 to the TB-4302 as shown in Figure 2. 3. Use Table 3 to configure the PXIe-4139 to output a Test Point value for the appropriate range shown in Table 6, beginning with the values in the first row. Table 3. PXIe-4139 Voltage Output Setup Configuration Value Function Sense Range Voltage output Remote 600 mv range for test points less than 100 mv 60 V range for all other test points Current Limit Current Limit Range 20 ma 200 ma 6 ni.com PXIe-4302/4303 and TB-4302C Calibration Procedure

4. Refer to Table 4 to configure the PXI-4071 and acquire a voltage measurement. Table 4. PXI-4071 Voltage Measurement Setup Configuration Value Function Range DC measurement 1 V range for test points less than 100 mv. 10 V range for all other test points. Digital Resolution Aperture Time Autozero ADC Calibration Input Impedance DC Noise Rejection 7.5 digits 100 ms On On > 10 G High Order Number of Averages 1 Power Line Frequency Dependent upon the local power line characteristics. 5. Acquire a voltage measurement with the PXIe-4302/4303. a. Create a DAQmx task. b. Create and configure the AI channel according to the values shown in Table 5. Table 5. AI Voltage Mode Setup Configuration Channel Name Task Sample Mode Value Dev1/aix, where x refers to the channel number AI voltage Finite samples Sample Clock Rate 5000 Samples per Channel 5000 Maximum Value Appropriate maximum range value from Table 6 Minimum Value Appropriate minimum range value from Table 6 Units Volts c. Start the task. d. Average the readings that you acquired. PXIe-4302/4303 and TB-4302C Calibration Procedure National Instruments 7

e. Clear the task. f. Compare the resulting average to the Lower Limit and Upper Limit values in Table 6. If the result is between these values, the device passes the test. Table 6. Voltage Measurement Accuracy Limits Minimum Range (V) Maximum Test Point (V) Lower Limit (V) Upper Limit (V) -0.1 0.1-0.095 DMM Reading - 0.0007 V DMM Reading + 0.0007 V -0.1 0.1 0 DMM Reading - 0.000029 V DMM Reading + 0.000029 V -0.1 0.1 0.095 DMM Reading - 0.0007 V DMM Reading + 0.0007 V -10 10-9.5 DMM Reading - 0.004207 V DMM Reading + 0.004207 V -10 10 0 DMM Reading - 0.001262 V DMM Reading + 0.001262 V -10 10 9.5 DMM Reading - 0.004207 V DMM Reading + 0.004207 V 6. For each value in Table 6, repeat steps 3 through 5 for all channels. 7. Set the PXIe-4139 output to be zero volts. 8. Disconnect the PXIe-4139 and PXI-4071 from the TB-4302. Adjustment The following performance adjustment procedure describes the sequence of operations required to adjust the PXIe-4302/4203. Complete the following steps to adjust the accuracy of the PXIe-4302/4203. 1. Set the PXIe-4139 output to be zero volts. 2. Connect the PXIe-4139 and PXI-4071 to the TB-4302 as shown in Figure 2. 3. Call the DAQmx Initialize External Calibration function with the following parameters: Device In: Dev1 Password: NI 1 4. Call the 4302/4303 instance of the DAQmx Setup SC Express Calibration function with the following parameters: calhandle in: calhandle output from DAQmx Initialize External Calibration rangemax: Appropriate Range Max starting with the value in the first row of Table 7 rangemin: Appropriate Range Min starting with the value in the first row of Table 7 physical channels: dev1/ai0:31 1 The default password is NI, which can be changed. Refer to the NI-DAQmx Help for details. 8 ni.com PXIe-4302/4303 and TB-4302C Calibration Procedure

Table 7. Voltage Mode Adjustment Test Points Range (V) Max Min Test Points (V) 0.1-0.1-0.09-0.06-0.03 10-10 -9 5. Refer to Table 3 to configure the PXIe-4139. Set the PXIe-4139 output equal to the first Test Point for the corresponding range in Table 7 that was configured in step 4. 6. Enable the PXIe-4139 output. 7. Refer to Table 4 to configure the PXI-4071 and acquire a voltage measurement. 8. Call the 4302/4303 instance of the DAQmx Adjust SC Express Calibration function with the following parameters: calhandle in: calhandle output from DAQmx Initialize External Calibration reference voltage: DMM measurement value from step 7 9. Repeat steps 5 through 8 for the remaining Test Point values from Table 7 for the corresponding range that was configured in step 4. 10. Repeat steps 4 through 9 for the remaining ranges from Table 7. 11. Call the 4302/4303 instance of the DAQmx Adjust SC Express Calibration function with the following parameters: calhandle in: calhandle output from DAQmx Initialize External Calibration action: commit 0 0.03 0.06 0.09-6 -3 0 3 6 9 PXIe-4302/4303 and TB-4302C Calibration Procedure National Instruments 9

EEPROM Update When an adjustment procedure is completed, the PXIe-4302/4303 internal calibration memory (EEPROM) is immediately updated. If you do not want to perform an adjustment, you can update the calibration date without making any adjustments by initializing an external calibration and closing the external calibration. Reverification Repeat the Accuracy Verification section to determine the As-Left status of the device. Note If any test fails Reverification after performing an adjustment, verify that you have met the Test Conditions before returning your device to NI. Refer to World Wide Support and Services for assistance in returning the device to NI. TB-4302C Verification This section provides information for verifying the performance of the TB-4302C. Test Equipment Table 8 lists the equipment recommended for verifying the shunt value of the TB-4302C. If the recommended equipment is not available, select a substitute using the requirements listed in Table 8. Table 8. Recommended Equipment for PXIe-4302/4303 Verification and Adjustment Equipment Recommended Model Requirements DMM PXI-4071 Use a DMM with an accuracy of 136 ppm or better when measuring 5 Ω in 4-wire mode. 10 ni.com PXIe-4302/4303 and TB-4302C Calibration Procedure

Accuracy Verification The TB-4302C has a total of 32, 5 shunt resistors, one for each channel. The reference designators of the shunt resistors range from R10 to R41 as shown in Figure 3. Figure 3. TB-4302C Circuit Board Shunt Resistor Locator Diagram S/N NI TB-4302C C2 R1 R5 R6 R4 U2 R2 U3 R3 8 R7 1 1 8 1 8 8 1 1 C6 8 C1 B1 A1 J1 COPYRIGHT 2014 C4 CR2 R8 U4 CR1 C1 U1 C3 C5 1 J2 8 1 J4 8 158141A-01L AI7 AI6 AI5 AI4 AI3 AI2 AI1 AI0 RT5 RT6 RT7 RT8 R14 R15 R16 R17 RT2 RT3 RT4 R11 R12 R13 RT1 R10 J3 J7 J5 J9 R24 R23 R22 RT15 RT14 RT13 R21 R20 R19 R18 R25 RT10 RT9 RT16 RT12 RT11 AI15 AI4 AI3 AI2 AI1 AI0 A9 AI8 AIGND J8 J12 J6 J21 CR3 U9 U11 U10 J19 J10 AI23 AI22 AI21 AI20 AI19 AI18 AI17 AI16 RT21 RT22 RT23 RT24 RT19 RT20 RT18 RT17 R26 R30 R31 R32 R33 J13 J18 R27 R28 R29 J15 J11 J17 PFI0 GND R41 R40 R39 R38 RT32 RT31 RT30 RT29 R36 R35 R34 RT27 RT26 RT25 R37 RT28 AI31 AI30 AI29 AI28 AI27 AI26 A25 AI24 FOR PATENTS: NI.COM/PATENTS 8 J14 1 1 2 3 4 1 R10, R11, R12, R13, R14, R15, R16, R17 (Bottom to Top) 2 R21, R20, R19, R18, R25, R24, R23, R22 (Bottom to Top) 3 R26, R27, R28, R29, R30, R31, R32, R33 (Bottom to Top) 4 R37, R36, R35, R34, R41, R40, R39, R38 (Bottom to Top) PXIe-4302/4303 and TB-4302C Calibration Procedure National Instruments 11

Table 9 shows the correlation between the AI channels and the shunt reference designators. Table 9. Channel to Shunt Reference Designator Correlation Channel CH0 CH1 CH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10 CH11 CH12 CH13 CH14 CH15 CH16 CH17 CH18 CH19 CH20 CH21 CH22 CH23 CH24 CH25 CH26 CH27 CH28 CH29 CH30 CH31 Shunt Reference Designator R10 R11 R12 R13 R14 R15 R16 R17 R21 R20 R19 R18 R25 R24 R23 R22 R26 R27 R28 R29 R30 R31 R32 R33 R37 R36 R35 R34 R41 R40 R39 R38 12 ni.com PXIe-4302/4303 and TB-4302C Calibration Procedure

The following performance verification procedure describes the sequence of operations to verify the shunt values of the TB-4302C. 1. Open the TB-4302C enclosure. 2. Configure the PXI-4071 for 4-wire resistance measurement mode as shown in Table 10. Table 10. PXI-4071 Voltage Measurement Setup Function Range Configuration Value 4-wire resistance measurement 100 Digital Resolution 7.5 Aperture Time Autozero ADC Calibration Input Impedance DC Noise Rejection 100 ms On On > 10 G High order Number of Averages 1 Power Line Frequency Offset Compensated Ohms Dependent upon the local power line characteristics. On 3. Locate R10 on the TB-4302C. Refer to Figure 3. 4. Hold the HI and HI_SENSE probes of the PXI-4071 to one pad of R10 and hold the LO and LO_SENSE probes to the other pad of R10. 5. Acquire a resistance measurement with the PXI-4071. 6. Compare the results to the Lower Limit and Upper Limit values in Table 11. If the results are between these values, the device passes the test. Table 11. 5 Shunt Accuracy Limit Nominal Upper Limit Lower Limit 5 5.025 4.975 7. Repeat steps 3 through 6 for all other 5 shunt resistors. Note If the TB-4302C fails verification, refer to World Wide Support and Services for assistance in returning the terminal block to NI. PXIe-4302/4303 and TB-4302C Calibration Procedure National Instruments 13

Specifications Refer to the NI PXIe-4302/4303 Specifications document for detailed PXIe-4302/4303 specification information. Refer to the NI PXIe-4302/4303 and TB-4302/4302C User Guide and Terminal Block Specifications document for detailed TB-4302C specification information. World Wide Support and Services The National Instruments website is your complete resource for technical support. At ni.com/ support you have access to everything from troubleshooting and application development self-help resources to email and phone assistance from NI Application Engineers. Visit ni.com/services for NI Factory Installation Services, repairs, extended warranty, and other services. Visit ni.com/register to register your National Instruments product. Product registration facilitates technical support and ensures that you receive important information updates from NI. National Instruments corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas, 78759-3504. National Instruments also has offices located around the world. For telephone support in the United States, create your service request at ni.com/support or dial 1 866 ASK MYNI (275 6964). For telephone support outside the United States, visit the Worldwide Offices section of ni.com/niglobal to access the branch office websites, which provide up-to-date contact information, support phone numbers, email addresses, and current events. Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for more information on National Instruments trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patents Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015. 2015 National Instruments. All rights reserved. 377005A-01 Sep15