80i-110s. AC/DC Current Probe Instructions. Introduction

Similar documents
80i-110s AC/DC CURRENT PROBE

323/324/325 Clamp Meter

The table below lists the symbols used on the Clamp and/or in this manual. Important Information. See manual.

374 FC/375 FC/376 FC Clamp Meter

374/375/376 Clamp Meter

374/375/376 Clamp Meter

15B+/17B+/18B+ Calibration Manual. Digital Multimeter

Electrical Multimeter

netzerotools.com 374/375/376 Users Manual Clamp Meter

374/375/376. Calibration Manual. Clamp Meter

902 FC. HVAC True-rms Clamp Meter

312, 316, 318. Clamp Meter. Users Manual

T+ and T+ PRO Electrical Tester

36 AC/DC True RMS. Clamp Meter. P Read First: Safety Information

15B & 17B. Multimeters. Users Manual

712C. Users Manual. RTD Calibrator

T6-1000/T Service Information. Electrical Tester

CT238A. AC/DC Current Probe. User Manual. For detailed specifications and ordering info go to

374/375/376 Clamp Meter

12B/18 MultiMeter Instruction Sheet

Model 77 Series IV. Digital Multimeter. Users Manual

1630-2/ FC. Calibration Manual. Earth Ground Clamp

712B. Users Manual. RTD Calibrator. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

34134A AC/DC DMM Current Probe. User s Guide. Publication number April 2009

80i-600A AC Current Probe

Model 77 Series IV. Digital Multimeter. Users Manual

Circuit Breaker Finder

AM-510 Commercial / Residential Multimeter. AM-510-EUR Digital Multimeter. Users Manual

Agilent 1146A AC/DC Oscilloscope Current Probe

51-54 Series II. Thermometer

AM-500 Autoranging Mulitmeter. AM-500-EUR Digital Multimeter. Users Manual

AC/DC Current Probe GCP-100 QUICK START GUIDE ISO-9001 CERTIFIED MANUFACTURER

SureGrip TM. XWWarning

175, 177, 179. True-rms Multimeters. Users Manual

PT-2000series AC/DCcurrentprobe PT-2710 PT-2720 PT-2740 PT-2770 MANUAL INSTRUCTION

99 Washington Street Melrose, MA Fax TestEquipmentDepot.com # # AAC Clamp Meter. Instruction Manual

15B & 17B. Users Manual. Multimeters

Instruction Manual ICM 135R Electrical Clamp Multimeter EN FR IT DE ES

T5-600/T Electrical Tester

For detailed specifications and ordering info go to

712B/714B. Calibration Manual. RTD/Thermocouple Calibrator

374/375/376 Clamp Meter. Users Manual

CM605. User Manual AC/DC LOW CURRENT CLAMP-ON METER ENGLISH

Digital Clamp Meter Model: &

AC FLEXIBLE CURRENT PROBE CA 3000

AC/DC CLAMP METER USER S MANUAL

ICM 136R Operating Instruction Electrical Clamp-Multimeter

USER MANUAL 600A AC Clamp Meter + NCV Model MA610

AC/DC Current Oscilloscope Probe Model SL261

HART Mode. Users Guide. July Fluke Corporation. All rights reserved. All product names are trademarks of their respective companies.

Operator s Manual. PP016 Passive Probe

79/26 Series III Multimeter

200Amp AC Clamp Meter + NCV Model MA250

AC / DC STROOMTANG MULTIMETER NI 30 AC / DC CLAMP MULTIMETER NI 30. Handleiding / Manual

Fluke 1587 FC/1577 Insulation Multimeters

DM-46 Instruction Manual

USER'S MANUAL ACDC-100 TRMS ACDC-100. Versatile AC/DC Clamp-on Multimeter Series

INSTRUCTION MANUAL UTL260

NOTE: Fully read and understand this manual before using this Digital Multimeter.

AMM-1022 Digital Multimeter USER`S MANUAL

MR461 MR561 AC/DC OSCILLOSCOPE CURRENT PROBES. User Manual. AC/DC CURRENT PROBE Model MR 561. AC/DC CURRENT PROBE Model MR V CAT III - 600A

16 Multimeter with Temperature

DIGITAL DUAL DISPLAY AC/DC CLAMP METER MODEL-860A OPERATION MANUAL

LH240 LH1040 LH2040 LH2015

Models 110, 111 & 112

User Guide. Digital AC/DC Clamp Meter Model 38394

16 Multimeter with Temperature

1535/1537. Calibration Manual. Insulation Tester

600A Clamp Meters w/tightsight Display

AC/DC Current Probes Model MR411 Model MR521

User s Guide. Digital AC/DC Clamp Meter Model 38394

TA MHz oscilloscope probe TA MHz oscilloscope probe

Model: PR-55. High Voltage Oscilloscope Probe 10 kv, 40 MHz USER MANUAL

AM-520 AM-530. Users Manual. HVAC Multimeter. True-rms Electrical Contractor Multimeter

BENNING. CM7 Electrical Tester. Service Information

Dawson DDM190. Digital Multimeter User s Manual

1507/1503. Insulation Testers. Users Manual

3000 FC. Users Manual. Wireless Multimeter

USER'S MANUAL. ACD-6 Pro. ACD-6 TRMS Pro. Versatile Clamp-on Multimeter Series

User Manual Digital Clamp Multimeter. model no.: MSR-C600

5730A. Service Manual. Multifunction Calibrator

AT Advanced Wire Tracer. Users Manual

AC/DC DIGITAL CLAMP METER OPERATION MANUAL

742A Series Resistance Standards

AMP-25 AMP-25-EUR. Mini-Clamp TRMS AC / AMP-25. ZERO 2 Sec

RIGOL. User s Guide. RP1000D Series High Voltage Differential Probe. Feb RIGOL Technologies, Inc

DM-46 Instruction Manual

MiniFlex MF MF FLEXIBLE AC CURRENT SENSOR. User Manual

ET-3888 True RMS Clamp Meter. User Manual

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

EM420A/420B DIGITAL MULTIMETER OWNERS MANUAL Read this owners manual thoroughly before use

PKT 512A-RO High Impedance Passive Cable Divider

Operator s Manual. PP022 Passive Probe

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL:

User s Manual. MiniTec TM Series. Model MN26 (Model MN26T includes temperature probe) Mini Autoranging MultiMeter

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: MODEL: Multimeter Service Information

Fluke 370 FC Series True-rms Wireless AC/DC Clamp Meters

PHV RO High Voltage Passive Probe. Instruction Manual

1. SAFETY INFORMATION.1 2. DESCRIPTION SPECIFICATIONS.6 4. OPERATING INSTRUCTION Voltage measurement Current measurement 10

KMD-S04 Multímetro de bolsillo

Transcription:

80i-110s.book Page 1 Monday, December 5, 2016 12:12 PM 80i-110s AC/DC Current Probe Instructions Introduction The Fluke 80i-110s (the Probe or Product) is a clamp-on AC/DC Current Probe that reproduces current waveforms found in commercial and industrial power distribution systems. The Probe performance is optimized for accurate reproduction of currents at line frequency and up to the 50th harmonic waveform. The 80i-110s is also compatible with any instrument capable of millivolt measurements. The Probe provides these benefits: Accurate AC, DC, and AC+DC current measurements for Electrical, Electronic, and Automotive applications. Shielded for high-noise immunity around electronic motor drives and ignition systems. Wide measurement range from 50 ma to 100 A, useful to 10 ma. Jaw shaped for easy access to cramped spaces. Tactile barrier (see Figure 1, item ). Safety-designed 600 V insuiated BNC - compatible with Fluke ScopeMeter test tools, Power Harmonic analyzers, and oscilloscopes. Selectable output of 10 millivolts/1 amp for the 100 A range, and 100 mv/1 A for the 10 A range. 1 Figure 1. 80i-110s AC/DC Current Probe PN 4862926 October 2005, Rev. 1, 1/17 2005-2017 Fluke Corporation. All rights reserved. Specifications are subject to change without notification. All product names are trademarks of their respective companies.

80i-110s.book Page 2 Monday, December 5, 2016 12:12 PM How to Contact Fluke To contact Fluke, call one of the following telephone numbers: Technical Support USA: 1-800-44-FLUKE (1-800-443-5853) Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853) Canada: 1-800-36-FLUKE (1-800-363-5853) Europe: +31 402-675-200 Japan: +81-3-6714-3114 Singapore: +65-6799-5566 Anywhere in the world: +1-425-446-5500 To view, print, or download the latest manual supplement, visit http://us.fluke.com/usen/support/manuals. Go to www.fluke.com to register your product and find more information. Safety A Warning identifies conditions and procedures that are dangerous to the user. Warning To prevent personal injury and for safe operation of the Product: Read all safety information before you use the Product. Carefully read all instructions. Do not use the Product if it is altered or damaged. Do not alter the Product and use only as specified, or the protection supplied by the Product can be compromised. Comply with local and national safety codes. Use personal protective equipment (approved rubber gloves, face protection, and flame-resistant clothes) to prevent shock and arc blast injury where hazardous live conductors are exposed. Hold the Product behind the tactile barrier. Before each use, examine the Product. Look for cracks or missing pieces of the clamp housing or output cable insulation. Also look for loose or weakened components. Carefully examine the insulation around the jaws. Do not touch voltages >30 V ac rms, 42 V ac peak, or 60 V dc. Use the Clamp only on insulated conductors. Use caution around bare conductors or bus bars. To prevent electrical shock, do not touch the conductor.

80i-110s.book Page 3 Monday, December 5, 2016 12:12 PM Symbols Symbol Definition WARNING - RISK OF DANGER. WARNING. HAZARDOUS VOLTAGE. Risk of electric shock. Consult user documentation. Double Insulated Earth Battery Application around and removal from uninsulated hazardous live conductors is permitted. Conforms to European Union directives. Conforms to relevant Australian Safety and EMC standards. Measurement Category III is applicable to test and measuring circuits connected to the distribution part of the building s lowvoltage MAINS installation. This product complies with the WEEE Directive marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste. Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation" product. Do not dispose of this product as unsorted municipal waste. Before You Start These items are included in the shipment box: AC/DC Current Probe, 80i-110s Instructions (this document) 9 volt Battery, type IEC 6LR61 Check the contents of the box for completeness. If something in the box is damaged or missing, contact your distributor or the nearest sales or service office.

80i-110s.book Page 4 Monday, December 5, 2016 12:12 PM Battery Warning To prevent electrical shock, unclamp the Probe from any conductor, and disconnect the Scopemeter test tool or any other measurement device before you install or replace the battery. Before the first use, install the battery. See Figure 2. To install the battery: 1. Unclamp the Probe from any conductor and disconnect the ScopeMeter test tool or any other measurement device. 2. Make sure that the Probe is in the OFF position. 3. Locate the battery cover on the handle and loosen the screw with a flatblade screwdriver. 4. Slide the battery cover away from the Probe. 5. Install the battery (IEC 6LR61). Arrange the battery leads so that they are not pinched between the handle bottom and the battery cover. 6. Reinstall the battery cover and secure the screw. + Figure 2. Battery Installation Warning To prevent possible electrical shock, fire, or personal injury, replace the batteries when the low battery indicator shows to prevent incorrect measurements.

80i-110s.book Page 5 Monday, December 5, 2016 12:12 PM Compatibility The Product is compatible with any Fluke ScopeMeter test tool, Power Harmonics Analyzer, Oscilloscope, Multimeter, or other voltage measurement device with these features: BNC input connector (PN PM9081/001 BNC-to-banana adapter) for use with standard inputs on a digital multimeter (DMM). Input accuracy of 2 % or better to take full advantage of the accuracy of the Probe. Input impedance of 1 MΩ in parallel with a maximum of 100 pf. A passband of more than four times the frequency of the waveform to be measured. Instructions To use the Probe: 1. Connect the Probe to the input on the measuring instrument. When the ScopeMeter test tool or an oscilloscope is used, it must have DC-coupled input. When you are using a digital multimeter, use the BNC-to- banana adapter (PM9081/001) to connect the Probe to the input. 2. On the Probe, select the least sensitive range (10 mv/a). Make sure that the green ON-indicator lights. See Figure 4 for selector switch and green On indicator locations. 3. On the Probe, rotate the ZERO thumbwheel to adjust the reading to zero. See Figure 4 for the ZERO rotary knob location. 4. Select the appropriate probe sensitivity on your ScopeMeter test tool or oscilloscope. 5. Clamp the Probe around the conductor. Make sure that the arrow mark on the jaw of the Probe points toward the correct orientation. See Figure 3. 6. Observe the current value or waveform on your display or the current value readout on the multimeter. 7. On the ScopeMeter test tool, adjust the vertical range knob and time division knob for the best display. 8. If the red OL-indicator lights, the Probe is in overload mode. Figure 3. Current Probe Orientation

80i-110s.book Page 6 Monday, December 5, 2016 12:12 PM 9. At completion, set the range selection switch to OFF again. A measurement setup with the Probe and a ScopeMeter test tool is shown in Figure 4. X. XX A Figure 4. Measurement Setup Measurement Guidelines Use these guidelines to position the jaws: Center the conductor inside the jaws. Position the Probe perpendicular to the conductor. Make sure that the arrow marked points toward the correct direction. on the jaw of the Probe Use these guidelines to take measurements: If possible, avoid measurements close to other current-carrying conductors. On the Probe, the 100 mv/a range offers the best accuracy. Maintenance Before each use, inspect the Probe. Look for cracks or missing portions of the housing and output cable insulating cover and for loose or weakened components. Pay particular attention to the insulation surrounding the jaws. If a Probe fails this inspection, tape it shut to prevent unintended operation. To check Probe performance, complete the Performance Verification procedure. Warning To avoid electric shock: Do not perform any service procedures unless you are qualified to do so. Read the "Safety" information at the beginning of this instruction sheet before proceeding.

80i-110s.book Page 7 Monday, December 5, 2016 12:12 PM If the Probe does not perform properly, use these steps to help isolate the problem: 1. Test the battery. Be sure that the green ON-indicator lights when you select the 10 mv/a range or the 100 mv/a range. 2. Inspect the jaw mating surface for cleanliness. If any foreign material is present, the jaws will not close properly and errors will result. 3. Verify that the Probe is zeroed properly. For zeroing be sure that the ScopeMeter test tool or oscilloscope is DC-coupled. 4. Verify that the function selection on the ScopeMeter test tool or oscilloscope is correct, for example, the display vertical resolution is not too low or too high. Repairs or service not covered in this document must be performed only at a Fluke Service Center. A Probe under warranty will be repaired or replaced (at Fluke's discretion) and returned at no charge. Cleaning and Storage Periodically wipe the case with a damp cloth and detergent. Do not use abrasives or solvents. Open the jaws and wipe the magnetic pole pieces with a lightly oiled cloth. Do not allow rust or corrosion to form on the magnetic core ends. If the Probe is not used for long time periods (>60 days) the battery should be removed and stored separately. Accessories An adapter accessory is required to use the Probe with a digital multimeter. The BNC-to-Banana Adapter, Fluke Part PM9081/001, is available to order. Specifications Maximum Voltage between any Terminal and Earth Ground... 600 V Dimensions... 67 mm x 231 mm x 36 mm (2.6 in x 9.1 in x 1.4 in) Weight... 330 g (11.6 oz), battery included Cable... 1.6 m (63 in) Maximum Conductor Size... 11.8 mm (0.46 in) Maximum Jaw Opening... 12.5 mm (0.49 in) Temperature Operating... 0 C to 50 C (32 F to 122 F) Storage... -30 C to +70 C (-22 F to +158 F) Relative Humidity (Operating) 0 % to 85 %... 0 C to 35 C (32 F to 95 F) 0 % to 45 %... 35 C to 50 C (95 F to 122 F) Altitude Operating... <2000 m (<6560 ft) Storage... <12 000 m (<40 000 ft) Demagnetize Probe... Open and close Probe jaws several times

80i-110s.book Page 8 Monday, December 5, 2016 12:12 PM Battery Type... 9 volt, IEC 6LR61 Consumption... 8.6 ma typical 12 ma maximum Service life with Alkaline IEC 6LR61 battery Typical... 55 hours Minimum... 40 hours Battery indicator (ON)... Green LED dims when battery voltage is <6.5 V Safety General... IEC 61010-1: Pollution Degree 2 Measurement... IEC 61010-2-032: CAT III 600 V Electromagnetic Compatibility (EMC) International... IEC 61326-1: Portable Electromagnetic Environment CISPR 11: Group 1, Class A Group 1: Equipment has intentionally generated and/or use conductively coupled radio-frequency energy which is necessary for the internal functioning of the equipment itself. Class A: Equipment is suitable for use in all establishments other than domestic and those directly connected to a low voltage power supply network which supplies buildings used for domestic purposes. There may be potential difficulties in ensuring electromagnetic compatibility in other environments, due to conducted and radiated disturbances. Caution: This equipment is not intended for use in residential environments and may not provide adequate protection to radio reception in such environments. Emissions that exceed the levels required by CISPR 11 can occur when the equipment is connected to a test object. USA (FCC)... 47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103. Electrical Specifications Valid temperature... 23 C ±3 C (73 F ±5 F). Current Ranges... 0 A dc to 10 A dc or ac peak 0 A dc to 100 A dc or ac peak Signals... 10 A range: 100 mv/a 100 A range: 10 mv/a Basic Accuracy (DC to 1kHz): Input Current (DC or AC peak) 0 A to 10 A Error (after zero check) Range 100 mv/a <3 % of reading +50 ma Range 10 mv/a - 0 A to 40 A - <4 % of reading +50 ma 40 A to 80 A - <12 % of reading +50 ma 80 A to 100 A - <15 % of reading

80i-110s.book Page 9 Monday, December 5, 2016 12:12 PM Extended Accuracy: For other frequencies, refer to the appropriate input current range and add the error listed below to the "Basic Accuracy" error. Additional Error Frequency Range 100 mv/a Range 10 mv/a 1 khz to 5 khz 3% 3% 5 khz to 20 khz 12 % 12 % >20 khz not specified not specified Input Load Impedance (of host instrument)... >1 MΩ in parallel with up to 100 pf Useful Bandwidth (-3 db)... 0 khz to 100 khz Rise or Fall Time... <4 μsec noise level Range 10 mv/a typical... 480 μv pk-pk Range 100 mv/a typical... 3 mv pk-pk Max. nondestructive current... 0 khz to 2 khz 140 A peak 2 khz to 10 khz 110 A peak 10 khz to 20 khz 70 A peak 20 khz to 50 khz 30 A peak 50 khz to 100 khz 20 A peak Temperature coefficient... 2000 ppm/ C max. for temperature from 0 C to 50 C (32 F to 132 F) Performance Verification Verify Probe accuracy with the test setup shown in Figure 5. Required test equipment is defined in Table 1. Toroid coil construction is illustrated in Table 2. To verify the Probe accuracy: 1. Set up connection. 2. Make the checks called for in Table 3 (100 mv/a). 3. Make the checks called for in Table 4 (10 mv/a). Table 1. Required Test Equipment Required Recommended AD/DC Calibrator Fluke 5520A Digital Multimeter (DMM) Fluke 45 Small insulated screwdriver Spectrol Banana-to-BNC Adapter Fluke PM9081/001 10-turn Toroid Coil See Figure 6

80i-110s.book Page 10 Monday, December 5, 2016 12:12 PM 5520A DMM Figure 5. Performance Test and Calibration Setup Table 2. Toroid Coil Construction 10.5 cm (4 in) D Item Description 10 turns wound on cylindrical form using 10-gauge magnet wire. Remove form and tape coil together making sure 10 wires are taped. Tape D Fan out coil to approximately 270 with spacing 5 cm (2 in) using tape to maintain coil shape.

80i-110s.book Page 11 Monday, December 5, 2016 12:12 PM Table 3. Performance Test Points: Current Range 0 A to 10 A (100 mv/a) DC Measurement: 5520A Settings DC Amps Measured Low Limit High Limit 0.1 A 1A 92 mv 108 mv 0.5 A 5A 480 mv 520 mv 0.9 A 9A 868 mv 932 mv AC Measurement: 5520A Settings RMS Amps Measured Low Limit High Limit 0.1 A, 60 Hz 1A 92 mv 108 mv 0.3 A, 400 Hz 3A 286 mv 314 mv 0.5 A, 2 khz 5A 465 mv 535 mv 0.6 A, 4 khz 6A 559 mv 641 mv Table 4. Performance Test Points: Current Range 0 A to 100 A (10 mv/a) DC Measurement: 5520A Settings DC Amps Measured Low Limit High Limit 1A 10 A 95.5 mv 104.5 mv 3A 30 A 287.5 mv 312.5 mv 5A 50 A 439.5 mv 560.5 mv 7A 70 A 615.5 mv 784.5 mv 9A 90 A 765.0 mv 1035.0 mv AC Measurement: 5520A Settings RMS Amps Measured Low Limit High Limit 1 A, 5 khz 10 A 92.5 mv 107.5 mv 2 A, 1 khz 20 A 185.5 mv 214.5 mv 3 A, 400 Hz 30 A 287.5 mv 312.5 mv 5 A, 400 Hz 50 A 439.5 mv 560.5 mv 7 A, 60 Hz 70 A 595.0 mv 805.0 mv

80i-110s.book Page 12 Monday, December 5, 2016 12:12 PM LIMITED WARRANTY AND LIMITATION OF LIABILITY This Fluke product will be free from defects in material and workmanship for one year from the date of purchase. This warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Fluke s behalf. To obtain service during the warranty period, contact your nearest Fluke authorized service center to obtain return authorization information, then send the product to that Service Center with a description of the problem. THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you. Fluke Corporation P.O. Box 9090 Everett, WA 98206-9090 U.S.A. 11/99 Fluke Europe B.V. P.O. Box 1186 5602 BD Eindhoven The Netherlands