Analog Voltage Detector 0-40kV Operating Instructions VDA040CTM

Similar documents
Digital Voltage Detector 0-999kV Operating Instructions VD1000PTM

Analog Voltage Detector 1-450kV Operating Instructions VDAO450TM. Patent No. 6,275,022

PHASE ROTATION METER. Operating and Instruction Manual. a n d A C C E S S O R I E S

TAG5000 WIRELESS PHASER. Instruction Manual HD ELECTRIC COMPANY 1475 LAKESIDE DRIVE WAUKEGAN, ILLINOIS U.S.A.

AT Advanced Wire Tracer. Users Manual

OPERATOR S MANUAL Model 160 Volt-Ohm-Milliammeter

200Amp AC Clamp Meter + NCV Model MA250

INSTRUCTION MANUAL UTL260

2011 / Circuit Tracer

99 Washington Street Melrose, MA Fax TestEquipmentDepot.com # # AAC Clamp Meter. Instruction Manual

CHANCE Proximity Voltage Indicator

Safety. This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages may be present.

Horizontal Mezzanine Gate

S&C Scada-Mate Switching Systems Outdoor Distribution Interrupter Replacement 14.4 kv through 34.5 kv

GT-1050A 2 GHz to 50 GHz Microwave Power Amplifier

INSTRUCTION MANUAL. Model Autoranging DMM ProbeMeter TM. Measures voltage, resistance, frequency, capacitance, temperature, and duty cycle.

Instruction Sheet SNE SERIES. Cable Chase

Current Probes. User Manual

RIGOL. User s Guide. RP5600 Passive Probe. July 2010 RIGOL Technologies, Inc.

Broadband Current Probe Series Operation Manual

Pen Multimeter. Model

Grid Radar Installation Manual

User Guide. Digital AC/DC Clamp Meter Model 38394

User s Guide. Digital AC/DC Clamp Meter Model 38394

2015 RIGOL TECHNOLOGIES, INC.

Installation Guide. Model VLS-TM1540. Universal Tilt/Pivot Wall Mount for use with Flat Panel Displays

TA MHz oscilloscope probe TA MHz oscilloscope probe

CAP ANALOG 410 INSTALLATION & OPERATION MANUAL

Electrical Multimeter

MODEL 3810/2 Line Impedance Stabilization Network

HIGH VOLTAGE PHASE COMPARATORS

R-X SERIES. Decade Resistor

600A Clamp Meters w/tightsight Display

User s Guide. 400A AC/DC Clamp Meter. Model MA220

PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT. HRRS 5kV AND 10kV SERIES. High Resistance, 5kV and 10kV Decade Substituter User and Service Manual

POWER TRACE TM TASCO, INC.

14 Piece Slide Hammer and Puller Set

PBM-UNI Installation Guide

INSTRUCTION SHEET STRUT DRIVER RETROFIT KIT PN: 22-79SDRF

Models Tension/Compression Pancake Load Cell Model 3174 with tension base DESCRIPTION FEATURES

INSTRUCTION MANUAL INF Fax: (503)

Copyright 2014 by Minarik Drives

TK 100 Probe Calibrator. Instruction Manual

UNIVERSAL TOOL TABLE. Owner s Manual

COOPER POWER SERIES. External application of a motor start capacitor on a single-phase voltage regulator. Voltage Regulators MN225063EN

DIGITAL DUAL DISPLAY AC/DC CLAMP METER MODEL-860A OPERATION MANUAL

Operation Instructions for. Energized Insulator Testers. Insulators No. 2, 3 & 4 DAMAGED

30 Bending Brake. Model Assembly and Operating Instructions. Distributed exclusively by Harbor Freight Tools.

Pen Multimeter. Model

Access Hatch. Installation Instructions and Operators Manual. PS DOORS Contact Information. Model AH-710 Standard High Neck Radius Cut

OPERATOR S MANUAL Model 77E Pneumatic Cable Stripper

Angular Rate Sensor. Owner's Manual

Instructions for Installation. S&C ELECTRIC COMPANY Specialists in Electric Power Switching and Protection TABLE OF CONTENTS INTRODUCTION INSTALLATION

Broadband Current Probe Series Operation Manual

Model 229 Series 2 Leakage Current Tester OPERATOR S MANUAL

CT238A. AC/DC Current Probe. User Manual. For detailed specifications and ordering info go to

HUSTLER 7' & 8' POOL TABLE ASSEMBLY INSTRUCTIONS

Warning, refer to accompanying documents.

Operator s Manual. PP016 Passive Probe

Model DB Disc Caliper Brake AIR CHAMP PRODUCTS. User Manual. (i) MTY (81)

Owner s Manual & Safety Instructions

INSTALLATION INSTRUCTIONS

NOVA-EXT Versatile Projector Mount Model: NOVA-EXT

PHV RO High Voltage Passive Probe. Instruction Manual

OPERATION & SERVICE MANUAL FOR FC 110 AC POWER SOURCE

Model 228 Current Leakage Tester INSTRUCTION MANUAL

User Manual. Smart-UPS On-Line. Uninterruptible Power Supply. Isolation Transformer Models: SURT001 and SURT002

Rojek Woodworking Machinery Interchangeable Shaper Spindle Installation and User s Guide

P5100A & P5150 High Voltage Probes Performance Verification and Adjustments

INSTRUCTION MANUAL LKG 601 Electrical Safety Analyzer

Installation Instructions. Monitor Roll Stand

RIGOL. User s Guide. RP1000D Series High Voltage Differential Probe. Feb RIGOL Technologies, Inc

Model Tension/Compression Pancake Load Cell DESCRIPTION FEATURES

Autoranging Multimeter Extech EX503

This manual should be read in its entirety prior to operation of the TS200 Test Set Page 2 of 9

STAFF User Manual. Manual Part #

Operator s Manual. PP022 Passive Probe

MANUAL METAL SHRINKER/STRETCHER

9 PIECE TUNGSTEN CARBIDE HOLE SAW KIT. Model 90721

Instruction Manual For Baldor Polishing Lathes

312, 316, 318. Clamp Meter. Users Manual

User s Guide. MultiView Series Digital MultiMeters Models: MV110 MV120 MV130

DM-45 Digital Multimeter

Glass Electrode Meter

User s Manual. MiniTec TM Series. Model MN26 (Model MN26T includes temperature probe) Mini Autoranging MultiMeter

LOGIC MODULE 3 INPUT AND/OR INSTRUCTION MANUAL MODEL# POWER INPUT: +12 volts DC LOGIC INPUT: +12 volts DC or Ground 3 YEAR WARRANTY

FLOW SWITCH 600 Series Velocity Flow Sensor. Instruction Manual

PRC-LA Installation Guide

Agilent 5DX System. X-ray Safety Test Procedure. for use with Software Release 8.20 and later

PRODUCT SAFETY SOLUTIONS MODEL HWI-01. Model HWI-01 Serial Number:

P5100A & P5150 High Voltage Probes Performance Verification and Adjustments

Model 113 Scintillation Preamplifier Operating and Service Manual

CM605. User Manual AC/DC LOW CURRENT CLAMP-ON METER ENGLISH

Model 72D Twin Blade Rotary Wire Stripper

MODEL SETUP & OPERATION MANUAL DOVETAIL JIG FEATURES SPECIFICATIONS

CT-2 and CT-3 Channel Taggers OPERATION MANUAL

User s Guide. Model MA A AC Mini Clamp-on Meter

PHV 1000-RO High Voltage Passive Probe. Instruction Manual

LUDLUM MODEL 182 RADON FLASK COUNTER. June 2011 Serial Number PR and Succeeding Serial Numbers

netzerotools.com 374/375/376 Users Manual Clamp Meter

Transcription:

Analog Voltage Detector 0-40kV Operating Instructions VDA040CTM

CONTENTS Limitation of Warranty and Liability 2 Product Safety Information 3 Design and Function 4 Voltage Detection in the C Position 5-6 Voltage Detection in the C Position Secondary Voltage & Capacitive Test Points 7 Voltage Detection in the URD Position 8 Voltage Detection in the OH Position 9 Voltage Detection in the x 4 Position 10 Parts & Accessories 11 1

Limitation of Warranty and Liability Bierer & Associates Inc. warrants this product to be free from defects in workmanship and material, under normal use and service conditions for a period of one year from date of shipment. Due to continuous product improvement and development, Bierer & Associates Inc. reserves the right to modify product designs and specifications without notice. It is impossible to eliminate all risks associated with the use of high voltage electrical devices including this device. Risks of serious injury or death are inherent in working around energized electrical systems. Such risks include but are not limited to variations of electrical systems and equipment, manner of use or applications, weather and environmental conditions, operator mentality, and other unknown factors that are beyond the control of Bierer & Associates Inc. Bierer & Associates Inc. do not express or imply to be an insurer of these risks, and by purchasing or using this product you AGREE TO ACCEPT THESE RISKS. IN NO EVENT SHALL Bierer & Associates Inc. BE LIABLE FOR ANY INCIDENTAL, CONSEQUENTIAL OR SPECIAL DAMAGES RESULTING FROM THE USE OR HANDLING OF THIS PRODUCT. SAFETY MESSAGE DEFINITIONS per ANSI Z535 These instructions contain important safety messages to alert the user to potentially hazardous situations, how to avoid the hazard, and the consequences of failure to follow the instruction. The safety alert symbol identifies a safety message. The signal word following the symbol indicates: DANGER A hazardous situation which, if not avoided, will result in death or serious injury and equipment damage. A hazardous situation which, if not avoided, could 2

result in death or serious injury and equipment damage. CAUTION A hazardous situation which, if not avoided, could result in minor or moderate injury and equipment damage. NOTICE Important safety message relating to equipment damage only. PRODUCT SAFETY INFORMATION 1. Meter assembly and live line tool adapters shall be considered non-insulating. Do not let live line tool fittings come in contact with energized or grounded conductors. 2. Use appropriate length live line tools for voltage being worked and maintain minimum approach distances as outlined in OSHA 1910.269, Table R-6. 3. All Phasing Meters and Voltage Detectors manufactured during and after 2007 will have a limit mark engraved on the high voltage probe(s) 2.5 inches from the tip to indicate to the user the physical limit that should not be exceeded when approaching and contacting an electrical conductor or other electrical test points. Zero Ohm insulated adapter (81280IE) should be used if limit mark will be exceeded. 4. This equipment should be used only by qualified employees, trained in and familiar with the safety-related work practices, safety rules and other safety requirements associated with the use of this type of equipment. 5. These instructions are not intended as a substitute for adequate training, nor do they cover all details or situations which could be encountered when operating this type of equipment. 6. Before operating this equipment, read, understand and follow all instructions contained in this manual. Keep instructions with equipment. INSPECTION & MAINTENANCE BEFORE USE 1. Prior to using any high voltage test equipment a careful inspection should be made to ensure the unit is free from any contaminants such as dirt, grease, etc. and that there are no apparent physical damages. 3

2. High voltage probe assemblies shall be wiped clean prior to each usecdesign and FUNCTION Limit Mark -see Product Safety Information, item 3, page 3. The VDA040C is a combination capacitive/direct contact, capacitive type voltage detector for use on voltages up to 40 kv Line-to-Ground (69 kv Phase-to-Phase), and is over-voltage protected to 66 kv Line-to- Ground (115 kv Phase-to-Phase). NOTE: Meter scale indicates Line-to-Ground values. Not calibrated to read Phase-to-Phase. Each unit has a five position switch for the following functions: C URD OH Capacitive Voltage Detection Voltage Detection, URD Equipment Voltage Detection, Overhead Lines X4 Meter Scale Times X 4 T Meter Test Meter should deflect full scale in the T position. Deflection below full scale indicates low battery. Replaceable 9V battery is located behind the live line tool attachment threaded into the meter housing. As with all voltage detectors, readings can be affected by a variety of field conditions. For example, if the live line tool attachment is close to another phase, ground or voltage source, the readings may be high. If the live line tool attachment is close to the same phase, readings may be low. If there is any doubt about the meter reading under any circumstances, the line or equipment shall be considered energized and appropriate safety precautions taken, i.e., confirm visual open gaps, tag outs, hold orders and sources of induced voltage. 4

Voltage Detection in the C Position In the C position, the VDA040C can be used as a non-contact proximity voltage detector on bare overhead primary conductors. It can also be used in direct contact with secondary voltages and capacitance test points to indicate the presence of voltage. 1 2 Test Voltage detector for proper operation by turning selector switch to the T position. Meter should deflect full scale. (See Warning page 4). Turn the selector switch to the C position. 3 4 5 From the ground, hold the VDA040C by the live line tool attachment and point the unit toward the energized line. Any meter deflection indicates the presence of voltage. (See statement page 4). If voltage is not present, meter should read zero volts. (See statement page 4). Re-test voltage detector by turning the selector switch to the T position. NOTE: Use of the VDA040C in the C position on primary voltages with an insulated live line tool or from an insulated aerial device, is NOT recommended since the readings may be diminished and could result in false de energized readings. DO NOT make contact with any energized conductors or equipment when holding unit by hand. If there is any doubt about the meter reading in the C position, the line or equipment shall be considered energized, and 5

appropriate safety precautions taken. To confirm the presence of nominal voltages, induced voltage prior to installing ground, re-test the line or equipment using the direct contact methods described on the following pages. Unit shall be tested before and after each use on a known voltage source. Failure to do so could result in false negative indications. Battery Replacement A standard 9 volt battery is located behind the 5/8 x 11 threaded live line tool fitting on the meter housing. Two flat edges are furnished for use with a wrench or slip joint pliers to remove and install the fitting from the probe housing. 6

Voltage Detection in the C Position Secondary Voltages and Capacitance Test Points 1 2 Test Voltage detector for proper operation by turning selector switch to the T position. Meter should deflect full scale. (See Warning page 4). Turn the selector switch to the C position. 3 4 5 Attach voltage detector to appropriate length live line tool for voltage being tested. Make direct contact with line or equipment under test. Any meter deflection indicates the presence of voltage. If no voltage is present, meter should read zero volts. 6 Re-test voltage detector by turning the selector switch to the T position. DO NOT make contact with any energized conductors or equipment when holding unit by hand. Readings should take into account proximity to other phases and grounded surfaces and be consistent with previous experience on the same voltage and circuit configuration with this voltage detector. If there is any doubt about the meter reading in the C position, the line or equipment shall be considered energized and appropriate safety precautions taken. Unit shall be tested before and after each use on a known voltage source. Failure to do so could result in false negative indications. 7

Voltage Detection in the URD Position probe. a 1 2 3 Test Voltage detector for proper operation by turning selector switch to the T position. Meter should deflect full scale. (See Warning page 4). Thread appropriate adapter, bushing or elbow into meter Attach voltage detector to appropriate length live line tool for voltage being tested. 4 Turn the selector switch to the URD position. Make direct contact with URD equipment under test. If 5 a equipment is energized, meter should read approximate Line-to-Ground voltage. For voltages above 10 kv, Line-to- Ground meter will read full scale. 6 7 If URD equipment is de-energized, meter should read zero volts. Re-test voltage detector by turning the selector switch to the T position. DO NOT make contact with any energized conductors or equipment when holding unit by hand. Readings should take into account proximity to other phases and grounded surfaces and be consistent with previous experience on the same voltage and circuit configuration with this voltage detector. If there is any doubt about the meter reading in the URD position, the line or equipment shall be considered energized and appropriate safety precautions taken. Unit shall be tested before and after each use on a known voltage source. Failure to do so could result in false negative indications. 8

Voltage Detection in the OH Position 1 2 Test Voltage detector for proper operation by turning selector switch to the T position. Meter should deflect full scale. (See Warning page 4). Turn the selector switch to the OH position. 3 4 5 Attach voltage detector to appropriate length live line tool for voltage being tested. Make direct contact with line or equipment under test. If equipment is energized, meter should read approximate Line-to-Ground voltage. For voltages above 10 kv Lineto-Ground, use the X4 position. If no voltage is present, meter should read zero volts. Readings other than zero volts may indicate the presence of induced voltage. 6 Re-test voltage detector by turning the selector switch to the T position. DO NOT make contact with any energized conductors or equipment when holding unit by hand. Readings should take into account proximity to other phases and grounded surfaces and be consistent with previous experience on the same voltage and circuit configuration with this voltage detector. If there is any doubt about the meter reading in the OH position, the line or equipment shall be considered energized and appropriate safety precautions taken. Unit shall be tested before and after each use on a known voltage source. Failure to do so could result in false negative indications. 9

Voltage Detection in the X4 Position 1 2 Test Voltage detector for proper operation by turning selector switch to the T position. Meter should deflect full scale. (See Warning page 4). Turn the selector switch to the X4 position. 3 4 5 Attach voltage detector to appropriate length live line tool for voltage being tested. Make direct contact with line or equipment under test. If equipment is energized, meter should read approximate Line-to-Ground voltage. (Actual reading times (X) 4 = voltage present). If line or equipment is de-energized, meter should read zero volts. Readings other than zero volts may indicate the presence of induced voltage. (Actual reading times (X) 4 = induced voltage present). 6 Re-test voltage detector by turning the selector switch to the T position. DO NOT make contact with any energized conductors or equipment when holding unit by hand. Readings should take into account proximity to other phases and grounded surfaces and be consistent with previous experience on the same voltage and circuit configuration with this voltage detector. If there is any doubt about the meter reading in the X4 position, the line or equipment shall be considered energized and appropriate safety precautions taken. Unit shall be tested before and after each use on a known voltage source. Failure to do so could result in false negative indications. 10

PARTS & ACCESSORIES PART NO. 8128TBALB 8128LHM 8128LPM 81280IE PA165UGA 10022CHL 10022HHSL DESCRIPTION 15-25kV Bushing Adapter Hook Probe Adapter Straight Probe Adapter Insulated Extension Adapter QC to Universal/Grip All combo Adapter 1 Handle w/threaded Ferrule and Cap 1 Handle w/threaded Ferrule and Stud *Nominal one inch in diameter and two feet in length; handle assemblies may be two, four or six feet in length. Limit Mark All Phasing Meters and Voltage Detectors manufactured after 2007 will have a limit mark engraved on the high voltage probe(s) 2.5 inches from the tip to indicate to the user the physical limit that should not be passed when approaching and contacting an electrical conductor or other electrical test points. Technical & Service Bierer & Associates Inc. Manufacturing & Repair 10730 Farrow Rd. Blythewood SC 29016 Tel: (803) 786-4839 Fax: (803) 786-5457 www.bierermeters.com 11