Minnesota Rules, Chapter 4732 X-ray Revision PROPOSED NON-MEDICAL CABINET X-RAY SYSTEMS, 1.0

Similar documents
Minnesota Rules, Chapter 4732 X-ray Revision PROPOSED NON-MEDICAL ANALYTICAL X-RAY SYSTEMS, 1.0

Minnesota Rules, Chapter 4732 X-ray Revision

Minnesota Rules, Chapter 4732 X-ray Revision

Minnesota Rules, Chapter 4732 X-ray Revision

Minnesota Rules, Chapter 4732 X-ray Revision

Minnesota Rules, Chapter 4732 X-ray Revision

RADIATION SAFETY REQUIREMENTS FOR ANALYTICAL X-RAY EQUIPMENT

TABLE OF CONTENTS. References

PART H RADIATION SAFETY REQUIREMENTS FOR NON-HEALING ARTS RADIATION GENERATING DEVICES (RGD)

(a) This subchapter applies to installations using analytical x-ray equipment and establishes requirements for their use.

APPENDIX K UNF RADIATION GENERATING DEVICE SAFETY PROCEDURES

Minnesota Rules, Chapter 4732 X-ray Revision

By: Georgia Institute of Technology

1. Oklahoma Administrative Code (OAC) 252:410, titled Radiation Management

Ch. 223 VETERINARY MEDICINE CHAPTER 223. VETERINARY MEDICINE GENERAL PROVISIONS X-RAYS RADIOACTIVE MATERIAL. Authority

LASER RADIATION REQUIREMENTS

VACUUM INTERRUPTER APPLICATION NOTES Filename: VIAN X-Rays and Vacuum Interrupters Revision: 0 PAGE 1

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS

RULES OF TENNESSEE DEPARTMENT OF ENVIRONMENT AND CONSERVATION DIVISION OF RADIOLOGICAL HEALTH CHAPTER USE OF X-RAY APPARATUS

Nuclear Associates , , CT Head and Body Dose Phantom

ANALYTICAL X-RAY EQUIPMENT RADIATION SAFETY MANUAL

2014 NEC Changes Part 1

LOCKOUT / TAGOUT PROGRAM

GE AMX 4+ Portable X-Ray

Verizon NEBS Compliance and Quality Assurance (NQA) Labeling Requirements for Light Emitting Products

05/20/14 1. Philips CM200T. Standby Condition

Agilent 5DX System. X-ray Safety Test Procedure. for use with Software Release 8.20 and later

RadView Manual Collimator

Agilent 5DX Series 2L, 3, and 5000 X-ray Safety Test Report

SECTION AUTOMATIC TRANSFER SWITCH

NJDEP Medical Physicist s Radiographic QC Survey Registration Number:

Product Guide Verizon Delaware LLC. Section 31 Delaware LLC Original Sheet 1. Connection With Certain Facilities of Others

CDRH 21 CFR Part 1040 Informative Laser Report

Keene State College Center for Environmental BioGeoChemistry Radiation Protection Program Revised January

DENTAL X-RAY OPERATOR'S INSTRUCTIONS. (for USA) Wall Mount Type...WK

QXR-W XRF SAFETY MANUAL

UNSIGNED HARDCOPY NOT CONTROLLED. Reference Designation Marking and Miscellaneous Coding

SAFETY AND HEALTH STANDARD ELECTRICAL GROUNDING Effective Date: 07/17/10 Standard: Document Number: KUCSH0039 Rev: 4

CDR PanElite. j~áåíéå~ååé=fåëíêìåíáçåë==== English

Category: ELECTRICITY Requirement: EL-ENG Page: 1 of 13. Document(s): S-E-01, S-E-04 Issue Date: Effective Date:

Laser User Manual. Therapy Unit INTRODUCTION

Power-Sure 700. Technical Data TD158001EN. Contents

CITY AND COUNTY OF DENVER POLICY DENVER FIRE DEPARTMENT. Emergency Responder Radio Enhancement Coverage System (RES)

SECTION EMERGENCY RESPONDER RADIO COVERAGE SYSTEMS

Series NRX with PXR - Instructions for Neutral Current Sensor Type NF

1/4in. SHEET PALM SANDER

AN EXAMPLE OF A STANDARD ARC FLASH PPE LABELING STRATEGY

SUGGESTED SPECIFICATION for Series 300 Automatic Transfer Switches

Siemens ST950 & ST900 Statement of Compliance Against TR 2513 Issue A, September 2005 Performance Specification for Wig Wag Signal Control Equipment

NATIONAL RULES FOR ELECTRICAL INSTALLATIONS. Fourth Edition. Extracts from Amendment No.1: June 2011

Wireless Power Transfer Devices (Wireless Chargers)

Instruction Manual. Self-Leveling Cross-Line Laser Level Model No and

Operating instructions Diffuse reflection sensor with background suppression. OJ51xx laser / / 2010

Wireless Power Transfer Devices (Wireless Chargers)

Radiological Safety Analysis Document for the CLAS12 Engineering and the first physics run of Run Group A

L25 Series Operator Manual

Nemko-CCL, Inc West Alexander Street Salt Lake City, UT

X-band Magnetron. Cooling (note 5) Water Output coupling (note 6) UG51/U Magnet (note 7) Integral, Permanent

I-01 NAVIGATIONAL WARNING RECEIVERS

800 MHz BIDIRECTIONAL AMPLIFIER Technical manual

REQUIREMENTS FOR USE OF NONMEDICAL X;RAY GENERATORS AT SLAC*

LASER ENHANCED REVOLVER GRIP OWNER S MANUAL RED LASER GREEN LASER

C-ARM FLUOROSCOPIC AND SPOT-FILM SYSTEMS

SWITCHING SAFETY & RELIABILITY CONFERENCE

Manual Unihan UPWL6024

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

MARICOPA COUNTY ZONING ORDINANCE Chapter 11 General Regulations

PREFACE ********************************************************** IT IS NOT INTENDED THAT THESE STANDARDS BE COPIED AND USED AS A SPECIFICATION!

Issue Date: Effective Date: Supersedes: S-E-06 (rev. 6)

X band Magnetron. Water: Anode cavity Forced-air: Input ceramics and terminals Output coupling (note 6) UG51/U Magnet (note 7) Integral, Permanent

PORTROD WALL MOUNTED HEIGHT ROD

FEI Titan Image Corrected STEM

Pipe Laser Model No Instruction Manual

Provläsningsexemplar / Preview

TABLE of CONTENTS ############################# Revised Rule Revised Rule 91.3-B 14. Revised Rule 92.1-F(2). 15

XRC X-Ray Calibration System

Digital Electronic Thermostat With RF

Laser Displacement Sensor 3Z4M

PRODUCTIVELY SAFER LOCK-OUT TAG-OUT PROCEDURE

P10 LASER DISTANCE MEASURE. IMPORTANT: Read before Using

Heritage MedCall. Sentry E-Call Model HM-527 Resident Host Panel

IChapter Number,ChapterTiiie I0400: _OJ TJse of X-Ray Apparatus! Rule Number Rule Title

AT&T INDIANA GUIDEBOOK. PART 2 - General Terms and Conditions 1st Revised Sheet 1 SECTION 9 - Connections

Installation and user s guide H A. NCi-5 non-contact tool setting interface

Section 16621A - AUTOMATIC TRANSFER SWITCH. Part 1 General

Wiring Methods, Components, and Equipment for General Use -- OSHA Standard

CLARK COUNTY FIRE CODE AMENDMENTS

Product Safety and RF Energy Exposure Booklet for Unication Two-Way Portable Radios

ISO INTERNATIONAL STANDARD. Safety of machinery Basic concepts, general principles for design Part 1: Basic terminology, methodology

INTELLIMETER REGISTER

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film

Indoor Contact Input Photosensor

Installation/User Manual

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing of fusionwelded

THE KENYA POWER AND LIGHTING CO. LTD. SPECIFICATIONS for IMPULSE VOLTAGE TEST EQUIPMENT

Centurion Laser Shower

Alternating-Current High-Voltage Power Vacuum Interrupters Safety Requirements for X-Radiation Limits

AT Advanced Wire Tracer. Users Manual

OPERATORS MANUAL WEEKENDER STEEL LADDER RACK

Notice of Rulemaking Hearing

Transcription:

Minnesota Rules, Chapter 4732 X-ray Revision PROPOSED NON-MEDICAL CABINET X-RAY SYSTEMS, 1.0 4732.#### NON-MEDICAL CABINET X-RAY SYSTEMS Subpart 1. Applicability. A registrant s non-medical cabinet x-ray system must comply with this subpart. Certified non-medical cabinet x-ray system must also comply with applicable requirements of Code of Federal Regulations, title 21, section 1020.40, or successor requirements. X-RAY SYSTEMS Subp. 2. Safety Device. A registrant is responsible for the safety device requirements for cabinet x-ray systems. A cabinet x-ray system safety device must: Commented [BB(1]: CFR; Ports and apertures 1020.40(c)(3) Ohio; 3701:1-68-06, A (2,3) A. prevent insertion of any part of the human body through: Commented [BB(2]: CA; 30337 (1) any port into the primary beam; and (2) any aperture. Subp. 3. Warning Lights. A. A visible warning light labeled with the words "X-RAY ON", or other visible warning indicator that indicates the cabinet x-ray system is producing Commented [BB(3]: Ohio; page 1 3701:1-68-04, (A) (2) ANSI N43.2-2001; 6.2.2.1.3 AK, FL, IN, ME, NM, AL (420-3-26-.11): same and includes fail-safe CA; 30337, NJ,NC ionizing radiation, must be located near a switch that energizes an x-ray tube and illuminated only when the tube is energized. Page 1

B. Warning lights must be labeled so that the purpose is easily identified. On equipment installed after the effective date of this part, a warning device must have a fail-safe design. Commented [BB(4]: Iowa; Ch 45, page 24, Section 45.5(3) (b) (3) AK, IN, KY, LA,NM - similar, effective date different ME- similar, has installed after effective date of these regulations, must have. Subp. 4. Beam ports. Unused beam ports on radiation producing housings must be secured in the closed position to prevent opening. Subp. 5. Controls and indicators. All non-medical cabinet x-ray systems must provide: A. A key-actuated [safety switch] control to prevent x-ray generation with the Commented [BB(5]: SSRCR; page H13, Section H.8 (d) ANSI N43.2-2001; 6.2.2.2.2 FL, GA, IL, IN, IA, LA, ME, NM; similar to MN TX(289.228), AL (420-3-26-.11) Commented [BB(6]: Ohio; 3701:1-68-06, A (6) a,b,c,d CA; 30337 key removed. B. A key-actuated [safety switch] control to initiate and terminate the generation of x-rays other than by functioning of a safety interlock or the main power control. C. Two independent means which indicate when x-rays are being generated, unless the x-ray generation period is less than one-half second, in which case Commented [TP(7]: This CFR provision is wordy and unclear. Looking for Focus Group input. the indicators shall be activated for one-half second, and which are discernible from any point at which initiation of x-ray generation is possible. Failure of a single component of the cabinet x-ray system shall not cause failure of both indicators to perform their intended function. One, but not both, of the indicators required by this subdivision may be a milliammeter labeled to indicate x-ray tube current. All other indicators shall be legibly labeled "X-RAY ON". Page 2

D. Additional means other than milliammeters which indicate when and only when x-rays are being generated, unless the x-ray generation period is less Commented [TP(8]: This CFR provision is wordy and unclear. Looking for Focus Group input. than one-half second in which case the indicators shall be activated for onehalf second, as needed to insure that at least one indicator is visible from each door, access panel, and port, and is legibly labeled "X-RAY ON". Subp. 6. Safety Interlocks. A. Each door of a cabinet x-ray system must have a minimum of two safety interlocks. One, but not both are required interlocks shall be such that door Commented [JC(9]: CFR; 1020.40, (c)(4), OH has, AK has this Commented [BB(10]: Ohio; 3701:1-68-06, A(4)a,b,c,d Same: NJ opening results in physical disconnection of the energy supply circuit on the high-voltage generator, and such disconnection shall not be dependent upon any moving part other than the door. B. Each access panel must have at least one safety interlock. C. Following interruption of x-ray generation by the functioning of any safety Commented [BB(11]: ANSI; N43.3, 7.5.4 interlock, use of a control panel provided in accordance with paragraph (c)(6)(ii) of this section shall be necessary for resumption of x-ray generation. D. Failure of any single component of the cabinet x-ray system must not cause Commented [JC(12]: Need to coordinate this CFR reference with MDH rule. failure of more than one required safety interlock. Subp. 7. Floors. A cabinet x-ray system must have a permanent floor. Any support Commented [JC(13]: CFR; 1020.40, (C)(2), NJ surface to which a cabinet x-ray system is permanently affixed may be deemed the floor of system. Page 3

Subp. 8. Ground fault. An accidental electrical grounding of an electrical conductor must not result in the generation of x-rays. Subp. 9. Labeling. A registrant is responsible for labeling cabinet x-ray systems according to this subpart. A. All cabinet x-ray systems must be labeled near any switch that energizes an Commented [JC(14]: CFR; 1020.40, (c)(5) Ohio; 3701:1-68-06, A(5)-similar CA; 30337 Commented [JC(15]: CFR; 1020.40, (c)(8) Ohio; 3701:1-68-06, 6(e) CA; 30337, NJ Commented [JC(16]: ANSI N43.2 2001; 6.2.2.1.5 x-ray tube with a readily visible and discernible sign bearing the radiation symbol and the words CAUTION RADIATION THIS EQUIPMENT PRODUCES IONIZING RADIATION WHEN ENERGIZED. B. All cabinet x-ray systems must be placed adjacent to each [beam] port and labeled with "CAUTION DO NOT INSERT ANY PART OF THE BODY WHEN Commented [BB(17]: CA; 30337 SYSTEM IS ENERGIZED - X-RAY HAZARD". Subp. 10. Safety device evaluation. An operator must evaluate a cabinet x-ray safety device upon [initial] installation and every six months (180 days). A safety device evaluation Commented [JC(18]: SSRCR; page H6, Section H.6 (j) Ohio; page 3, 3701:1-68-04, Section (C) (3), D ANSI; N43.3, 8.7 must include the interlocks, warning lights, and required emergency shut-off switches. A. The evaluation must verify that: (1) all cabinet x-ray safety devices are functioning as designed; and (2) all tags and labels are legible and visible. B. If a cabinet x-ray safety device is not functioning as designed, then it must be: (1) labeled immediately as defective; and Page 4

(2) removed from service until the safety device is corrected. C. Registrant must maintain a safety device evaluation record. The record must include: (1) date of evaluations; (2) a list of (all) the safety devices evaluated; (3) results of the evaluation; (4) survey instrument model and serial numbers; (5) survey instrument current calibration date; (6) the individual performing the evaluation; and (7) corrective actions recommended and performed for any safety device that fails the required evaluation. D. A cabinet x-ray system that is locked out and tagged DO NOT USE by the RSO is exempt from this part. E. A cabinet x-ray system that is returned to service after being locked-out and tagged must be evaluated before use if the date of the last safety device evaluation exceeds the six-month interval. Subp. 11. Radiation emission limit. Radiation emitted from the cabinet x-ray system must not exceed an exposure of 0.5 milliroentgen (0.005 msv) in one hour at any point 5 centimeters outside the external surface when: Page 5 Commented [JC(19]: CFR, 1020.40m (c)(1) Ohio; 3701:1-68-06, A(1) CA; 30337, NJ Commented [JC(20]: Should the radiation emission limit provision be consistent (ie less detailed) with closed-beam analytical x-ray systems? The following language is from the Non-medical analytical x-ray systems rule draft: Closed-beam analytical x-ray systems must not exceed a dose rate of 0.5 mrem (0.005 msv) in one hour at a distance of five centimeters measured outside at the nearest accessible surface.

A. Measurements are averaged over a cross-sectional area of ten square centimeters with no linear dimension greater than 5 centimeters; B. The cabinet x-ray system is operated at combinations of x-ray tube potential, current, beam orientation, and conditions of scatter radiation which produce the maximum x-ray exposure at the external surface; and C. The door(s) and access panel(s) are fully closed as well as fixed at any other position(s) which will allow the generation of x-ray radiation. Subp. 12. Additional controls and indicators for cabinet x-ray systems designed to admit humans. Cabinet x-ray systems designed to admit humans must: Commented [JC(21]: CFR; 1020.40, (c)(7), AK same as this Ohio; 3701:1-68-06, B, (1,2,3,4,5,6,7) A. Provide a control within the cabinet for preventing and terminating x-ray generation, which cannot be reset, overridden or bypassed from the outside of the cabinet. B. Not allow the initiation of x-ray exposure from within the cabinet. C. Audible and visible warning signals within the cabinet which are actuated for at least 10 seconds immediately prior to the first initiation of x-ray Commented [BB(22]: Do we need to include this in subpart 9? Section F, Ohio Commented [BB(23]: ANSI; N43.3, 7.5.2 generation after closing any door designed to admit humans. Failure of any single component of the cabinet x-ray system must not cause failure of both the audible and visible warning signals. D. A visible warning signal within the cabinet must remain actuated when and only when x-rays are being generated, unless the x-ray generation period is Page 6

less than one-half second in which case the indicators shall be activated for one-half second. E. Signs must indicate the meaning of the warning signals required according to this subpart and: (1) Must contain instructions for the use of the control provided pursuant to subpart 10 A. (2) Must be legible, visible, and illuminated when the main power control is in the "on" position. F. A means for a person within the enclosure to be able to egress at all times. G. In addition to subpart 9, the safety systems in this subpart must be tested for proper operation: (1) The audible and visible warning signals must be checked daily; (2) The control for x-ray prevention and termination must be checked monthly; (3) If the safety system(s) in this subpart do not function as designed, then the safety system(s) must be: a. Labeled immediately as defective; and b. Removed from service until the safety device is corrected. Page 7

(4) This subpart does not apply to a cabinet system that is lock-out/tag-out by a registrant and tagged DO NOT USE. Subp. 13. Additional requirements for x-ray cabinet inspection systems. Cabinet x-ray systems designed for the inspection of baggage and [industrial quality control] must: A. At the control area, an operator must be immediate attendance at all times Commented [JC(24]: CFR; 1020.40 (c)(10) AK same as this Ohio; 3701:1-68-06, C(1,2,3) Commented [JC(25]: For inspecting food products on similar conveyor belt systems. and must be able to view the ports and doors during exposure. B. Enable the operator to terminate the exposure or preset succession of exposures at any time for exposures of one-half second or greater duration. C. Allow for completion of the exposure or preset succession of exposures of less than one-half second duration and must enable the operator to prevent additional exposures. Subp. 14. Radiation survey. A registrant is responsible for performing a survey of cabinet x-ray systems that comply with the radiation emission requirements under subpart 12. Commented [JC(26]: SSRCR; page H7, Section H.6. (e) Ohio; 3701:1-68-06, E TX(289.228) AL (420-3-26-.11),NJ A radiation survey must be performed: A. upon installation of the equipment, and at an interval not to exceed 12 Commented [BB(27]: ANSI; N43.3, 8.1, 8.6 months; B. after any change to the local components in the system, including the initial arrangement, number, or type; Page 8

C. after any maintenance that requires the disassembly, removal, or repair of a Commented [JC(28]: ANSI; N43.3, 8.2 local component in the system; D. during maintenance, calibration, and other procedure that requires the presence of a primary x-ray beam while any local component in the system is disassembled or removed; E. after a bypass of a safety device or interlock; F. when a visual inspection of the analytical x-ray systems an abnormal condition, according to subpart 12; and G. radiation survey instruments must be used according 4732.####. CONDITIONS OF OPERATIONS Subp. 15. Safety Procedures. A registrant must develop and comply with operating procedures for cabinet x-ray systems that include step-by-step instructions to accomplish the Commented [JC(29]: SSRCR; page H4, in definitions Louisiana; page 233, in definitions FL, IN, IA; procedures not listed task. A. Operating procedures may be maintained in electronic or written form and must include: (1) sample insertion and manipulation; (2) (2) equipment alignment; (3) (3) routine maintenance by the registrant; Commented [JC(30]: Are subitems (1) and (2) needed? (4) (4) bypassing a safety device; (5) (5) Lock out/tag out; and Page 9

(6) (6) Record retention procedures. B. No individual may operate a cabinet x-ray system in any manner other than Commented [BB(31]: NJ, that specified in the operating procedures unless the individual has obtained written approval from the radiation safety officer (RSO). C. Operating procedures must be available to all operators of cabinet x-ray systems. this subpart. Subp. 16. Bypassing a safety system. A registrant is responsible for the requirements of Commented [JC(32]: SSRCR; page H8, Section H.6. (H) (2) TX(289.228), AL (420-3-26-.11); A. A registrant must follow operating procedures under subpart 15 for bypassing a safety system. B. An operator is prohibited from bypassing a safety device, interlock, or removing shielding unless an operator obtains approval from the RSO. Commented [JC(33]: Is this needed? C. An approval for bypassing a safety device under item A may be in electronic or written form and must be: (1) authorized or signed; and dated by the RSO; and (2) for a specified and limited period of time. D. When a safety device or interlock is bypassed, a sign bearing the words "SAFETY SYSTEM NOT WORKING," must be placed: (1) on the cabinet x-ray system; and (2) at the control switch. Page 10

E. A registrant must maintain utilization data when bypassing a safety device or interlock. Utilization data may be maintained in electronic or written form and must include: (1) date(s) of the alteration; (2) type of alteration; (3) name of the individual who made the alteration; (4) length of time the unit remained in the altered condition; (5) name of the individual who restored the unit to normal operation; and (6) post bypass survey that is authorized or signed by the RSO. Subp. 17. Repair or modification. A qualified service provider must install, repair, or Commented [JC(34]: ANSI; N43.3, 9.2.3 make modifications to a registrant s cabinet x-ray system. A. The x-ray source power switch must be locked out and tagged for routine shutdown before repair or modification to a cabinet x-ray system. B. A qualified service provider must verify that the x-ray source is off, and will remain off, before an operation that involves removing the covers, shielding materials, x-ray source housings, modifications to collimators or beam stops. Subp. 18. Records. Commented [JC(35]: There will be one records provision applicable to all registrants. Records of safety device tests, check dates, findings and corrective actions must be available for inspection and maintained. SSRCR; page H9, Section H.6 (j) Page 11