Online data sheet GM35
C D E F Ordering information Type GM35 Part no. On request The exact device specifications and performance data of the product may deviate from the information provided here, and depend on the application in which the product is being used and the relevant customer specifications. Our regional sales organization will help you to select the optimum device configuration. Other models and accessories www.sick.com/gm35 Product description Simultaneous or as individual measurement: the GM35 gas analyzer measures CO 2, H 2 O and CO, or N 2 O, as well temperature and pressure and it does this reliably, quickly, and economically. s an in-situ gas analyzer, the GM35 detects these gas components directly in the gas stream without gas sampling. Reliability, precision, and short response time of the GM35 offer a key advantage for efficient control loops in all CO and CO 2 -generating processes. H I J K L M N O P Q R S T 15267 14181 certified certified t a glance Dynamic humidity correction Fast in-situ measurement directly in the process Simultaneous determination of up to three gas components, temperature, and pressure No gas sampling and conditioning Gas testable version of measuring probe available Integrated self-test and control functions Your benefits Dynamic humidity measurement directly in the process Provision of real humidity reference values Unbiased measuring values due to in-situ measurement Fast or short-term fluctuations in the process are being detected Representative measurement by selection of an appropriate probe or cross-duct type 2 GS NLYZERS SICK Online data sheet 2016-07-25 17:26:28 Subject to change without notice
Fields of application Emission monitoring according to EN 14181, e. g., in power or cement plants Control of the water injection upstream of the electrostatic precipitator Monitoring of the combustion efficiency in hazardous waste incineration Process monitoring in urea production Tube rupture detection in heat exchangers Detailed technical data GM35 system Measured values Performance-tested measurands Measurement principles Measuring ranges Certified measuring ranges CO CO 2 H 2 O N 2 O CO CO 2 H 2 O CO, CO 2, H 2 O, N 2 O CO, CO 2, H 2 O Gas filter correlation, Interference filter correlation 0... 180 ppm / 0... 20,000 ppm 0... 22.5 Vol.-% / 0... 100 Vol.-% 0... 25 Vol.-% / 0... 100 Vol.-% 0... 60 ppm / 0... 2,500 ppm Measuring ranges refer to 1 m measuring path Measuring ranges depend on application and device version 0... 75 mg/m³ / 0... 2,000 mg/m³ 0... 15 Vol.-% 0... 25 Vol.-% Cross-duct and GMP version are type approved Zero point drift Reference point drift ± 2 % relative to measuring range end value ± 2 % within the maintenance interval, relative to measuring range full scale mbient temperature 40 C... +55 C Temperature change maximum ±10 C/h Storage temperature 40 C... +55 C mbient humidity 96 % Relative humidity; bedewing of optical surfaces not permitted Conformities Electrical safety Enclosure rating Operation Model Test functions pproved for plants requiring approval 2001/80/EC (13. ImSchV) 2000/76/EC (17. ImSchV) 27.ImSchV T-Luft (Prevention of ir Pollution) EN 15267 EN 14181 MCERTS CE IP 65 / NEM 4x Menu-driven operation via separate control unit Cross-duct version Measuring probe version utomatic check cycle for zero and span point 2016-07-25 17:26:28 Online data sheet Subject to change without notice GS NLYZERS SICK 3
C D E F H I J K L M N O P Q R S T Sender/receiver unit Enclosure rating Power supply Voltage Frequency Power consumption Open measuring probe (GMP) nalyzer unit of the measuring system IP 65 / NEM 4x 291 mm x 527 mm x 529 mm 29 kg 115 V C / 230 V C 48... 62 Hz 350 W Measuring probe in open design with integrated purge air control system Process temperature +430 C With N 2 O measurement max. 180 C Process pressure Process gas velocity Dust load 120 hpa Depending on purge air supply 40 m/s 3 g/m³ See dimensional drawings 25 kg Material in contact with media Stainless steel 1.4571, stainless steel 1.4539 uxiliary gas connections Integrated components Gas-testable measuring probe (GPP) Purge air Pressure sensor PT1000 temperature sensor Process temperature +430 C With N 2 O measurement max. 180 C Process pressure Process gas velocity Dust load Material in contact with media Power supply uxiliary gas connections Integrated components Reflector unit Voltage Frequency Power consumption Measuring probe with gas permeable filter element for adjustment with test gas 250 hpa Depending on purge air supply 40 m/s 30 g/m³ See dimensional drawings 45 kg Stainless steel 1.4571, stainless steel 1.4539, ceramics, PTFE 115 V C / 230 V C 50 Hz / 60 Hz 150 W Test gas Purge air Pressure sensor PT1000 temperature sensor Reflector unit with hollow triple reflector 4 GS NLYZERS SICK Online data sheet 2016-07-25 17:26:28 Subject to change without notice
Process temperature +500 C With N 2 O measurement max. 180 C Process pressure Depending on purge air supply 291 mm x 280 mm x 161 mm 1.5 kg GM35 evaluation unit; steel sheet enclosure Enclosure rating nalog outputs nalog inputs Digital outputs Digital inputs Interfaces and bus protocols RS-232 RS-232 Indication Input Operation Model Power supply Voltage Frequency Power consumption The evaluation unit serves as the user interface and is responsible for data processing and output as well as control and monitoring functions. IP 65 / NEM 4x 3 outputs: 0/4... 20 m, 500 Ω Electrically isolated 1 input: 0... 20 m, 100 Ω 3 relay contacts: 48 V C, 1, 60 W / 48 V DC, 1, 30 W Preset for failure, maintenance and functional control 3 inputs: 24 V SOPS ET Proprietary service interface LC display Status LEDs: "Power", "Maintenance" and "Failure" rrow keys Functional keys Menu-driven operation via LC-display and membrane keyboard Steel sheet enclosure 200 mm x 346 mm x 97.5 mm 3 kg 115 V / 230 V 50 Hz / 60 Hz 50 W GM35 WE evaluation unit; cast metal enclosure Enclosure rating IP 67 nalog outputs nalog inputs Digital outputs Digital inputs The evaluation unit serves as the user interface and is responsible for data processing and output as well as control and monitoring functions. 3 outputs: 0/4... 20 m, 500 Ω Electrically isolated 1 input: 0... 20 m, 100 Ω 3 relay contacts: 48 V C, 1, 60 W / 48 V DC, 1, 30 W Preset for failure, maintenance and functional control 3 inputs: 24 V 2016-07-25 17:26:28 Online data sheet Subject to change without notice GS NLYZERS SICK 5
C D E F GM35 Interfaces and bus protocols Indication Input Operation Model Power supply RS-232 RS-232 Voltage Frequency Power consumption SOPS ET Proprietary service interface LC display Status LEDs: "Power", "Maintenance" and "Failure" rrow keys Functional keys Menu-driven operation via LC-display and membrane keyboard Cast metal enclosure 289 mm x 370 mm x 138 mm 3 kg 115 V / 230 V 50 Hz / 60 Hz 50 W H I J K L M N O P Q R S T Connection unit Purge air fixture; sender/receiver unit uxiliary gas connections Integrated components Purge air fixture; reflector unit uxiliary gas connections SLV4-2 purge air unit, 2H1300, 3-ph Gas flow rate To lengthen the internal CN-us connection with cable provided by the customer 125 mm x 103 mm x 57 mm 3 kg Fixture to flanges with connections for purge air and external cabelling 320.9 mm x 360 mm x 220 mm 7 kg Purge air PT1000 temperature sensor Pressure sensor Fixture to flanges with connections for purge air and external cabelling 320.9 mm x 360 mm x 220 mm 7 kg Purge air mbient temperature 20 C... +40 C Enclosure rating IP 54 Power supply Three-phase current Unit to provide dust-free air for flushing of optical surfaces 38 m³/h... 63 m³/h t 30 hpa counter pressure, depending on low pressure inside the filter 550 mm x 550 mm x 258 mm (for details see dimensional drawings) 18 kg 3-phase, Δ: 200... 240 V, 50 Hz, 2.6, 350 W 3-phase, Δ: 220... 275 V, 60 Hz, 2.3, 450 W 3-phase, Y: 345... 415 V, 50 Hz, 1.5, 350 W 3-phase, Y: 380... 480 V, 60 Hz, 1.3, 450 W 6 GS NLYZERS SICK Online data sheet 2016-07-25 17:26:28 Subject to change without notice
uxiliary gas connections Test functions Integrated components Purge air: 40 mm Pressure switch (switching point 35 hpa) 2-step air filter, type Europiclon, dust capacity 200 g 2016-07-25 17:26:28 Online data sheet Subject to change without notice GS NLYZERS SICK 7
C D E F Dimensional drawings (Dimensions in mm (inch)) GM35 sender/receiver unit 27 (1.06) 110 (4.33) 390 (15.35) 291 (11.46) 135 (5.31) 156 (6.14) 200 (7.87) 531,7 (20.93) 6,8 (0.27) H I J K L M N O P Q R S T > 106 (4.17) 643 (25.31) 8 GS NLYZERS SICK Online data sheet 2016-07-25 17:26:28 > 95 99,7 (3,93) Subject to change without notice
Open measuring probe (GMP) 227 (8.94) L2 L1 L3 Ø 114.3 (4.50) 10 (0.39) 76.4 (3.01) Ø 254 (10.00) Ø 14 (0.55) Ø 200 (7.87) Ø 200 (7.87) Ø 18 (0.71) Ø 40 (1.57) 360 (14.17) 233.2 (9.18) 55 (2.17) 144 (5.67) 176.9 (6.98) Ø 215.8 (8.50) GMP measuring probes Measuring gap L3 (active measuring path) Probe length, nominal L1 250 500 750 1,000 1,250 1,500 1,750 L2 900 935 296 46 --- --- --- --- --- 1,500 1,644 1,004.5 754.5 504.5 254.5 --- --- --- 2,000 2,128 1,489 1,239 989 739 489 239 --- 2,500 2,628 1,988 1,738 1,488 1,238 988 738 488 ll dimensions in mm pplication-specific lengths available on request 2016-07-25 17:26:28 Online data sheet Subject to change without notice GS NLYZERS SICK 9
C D E F H I J K L M N O P Q R S T Gas-testable measuring probe (GPP) 227 (8.94) 10 (0.39) Ø 254 (10.00) 360 (14.17) GPP measuring probes Probe length, nominal 233 (9.17) L1 L2 238 (9.37) L1 Ø 254 (10.00) Ø 215.8 (8.50) Ø 200 (7.87) L3 Ø 18 (0.71) Ø 22.4 (0.88) Measuring gap L3 (active measuring path) Ø 114.3 (4.50) 227 477 727 977 1,000 904 353 103 --- --- 1,500 1,614 1,063 813 563 313 2,000 2,098 1,547 1,297 1,047 797 2,500 2,598 2,047 1,797 1,547 1,297 ll dimensions in mm pplication-specific lengths available on request 10 GS NLYZERS SICK Online data sheet 2016-07-25 17:26:28 L2 Subject to change without notice
GM35 reflector unit 135 (5.31) 291 (11.46) 144 (5.67) Ø 200 (7.87) 280 (11.02) R = 295 (11.61) 146 (5.75) 161 (6.34) 35.5 (1.40) max. 120 2016-07-25 17:26:28 Online data sheet Subject to change without notice GS NLYZERS SICK 11
C D E F Evaluation unit; steel sheet enclosure 200 (7.87) 160 (6.30) Ø 8 (0.31) 322 (12.68) 300 (11.81) 97.5 (3.84) 90 (3.54) 82 (3.22) H I J K L M N O P Q R S T Evaluation unit; cast metal enclosure 260 (10.24) 289 (11.38) 370 (14.57) 25 (0.98) 23 (0.91) 235 (9.25) Ø 7 (0.28) 330 (12.99) 85 (3.35) R = 273 (10.75) 12 GS NLYZERS SICK Online data sheet 2016-07-25 17:26:28 138 (5.43) Subject to change without notice
GM35 connection unit 57 (2.24) PG11 PG13,5 125 (4.92) 6 (0.24) 113 (4.45) Ø 5 (0.20) 23 (0.91) 14 (0.55) 80 (3.15) 52 (2.05) SLV4-2 purge air unit, 2H1300 Ø 10 (0.39) 550 (21.65) 470 (18.50) 40 (1.57) 40 470 (18.50) 128 (5.04) (1.57) 550 (21.65) 258 (10.16) 2016-07-25 17:26:28 Online data sheet Subject to change without notice GS NLYZERS SICK 13
Online data sheet SICK at a glance SICK is one of the leading manufacturers of intelligent sensors and sensor solutions for industrial applications. unique range of products and services creates the perfect basis for controlling processes securely and efficiently, protecting individuals from accidents and preventing damage to the environment. We have extensive experience in a wide range of industries and understand their processes and requirements. With intelligent sensors, we can deliver exactly what our customers need. In application centers in Europe, sia and North merica, system solutions are tested and optimized in accordance with customer specifications. ll this makes us a reliable supplier and development partner. Comprehensive services complete our offering: SICK LifeTime Services provide support throughout the machine life cycle and ensure safety and productivity. For us, that is Sensor Intelligence. Worldwide presence: Contacts and other locations www.sick.com SICK G Waldkirch Germany www.sick.com