An oscillator circuit to produce a radio-frequency discharge and application to metastable helium saturated absorption spectroscopy

Similar documents
arxiv: v1 [physics.atom-ph] 17 Feb 2012

A New Approach to Modeling the Impact of EMI on MOSFET DC Behavior

A high PSRR Class-D audio amplifier IC based on a self-adjusting voltage reference

Wireless Energy Transfer Using Zero Bias Schottky Diodes Rectenna Structures

Power- Supply Network Modeling

Resonance Cones in Magnetized Plasma

Neel Effect Toroidal Current Sensor

A 100MHz voltage to frequency converter

Direct optical measurement of the RF electrical field for MRI

BANDWIDTH WIDENING TECHNIQUES FOR DIRECTIVE ANTENNAS BASED ON PARTIALLY REFLECTING SURFACES

High finesse Fabry-Perot cavity for a pulsed laser

Optical component modelling and circuit simulation

MODELING OF BUNDLE WITH RADIATED LOSSES FOR BCI TESTING

Enhanced spectral compression in nonlinear optical

INVESTIGATION ON EMI EFFECTS IN BANDGAP VOLTAGE REFERENCES

Computational models of an inductive power transfer system for electric vehicle battery charge

Analysis of the Frequency Locking Region of Coupled Oscillators Applied to 1-D Antenna Arrays

Design of Cascode-Based Transconductance Amplifiers with Low-Gain PVT Variability and Gain Enhancement Using a Body-Biasing Technique

Gate and Substrate Currents in Deep Submicron MOSFETs

Ironless Loudspeakers with Ferrofluid Seals

A design methodology for electrically small superdirective antenna arrays

Small Array Design Using Parasitic Superdirective Antennas

Floating Body and Hot Carrier Effects in Ultra-Thin Film SOI MOSFETs

L-band compact printed quadrifilar helix antenna with Iso-Flux radiating pattern for stratospheric balloons telemetry

Dynamic Platform for Virtual Reality Applications

SUBJECTIVE QUALITY OF SVC-CODED VIDEOS WITH DIFFERENT ERROR-PATTERNS CONCEALED USING SPATIAL SCALABILITY

Superconducting magnetometers with sensitivities approaching gauss

70km external cavity DWDM sources based on O-band Self Seeded RSOAs for transmissions at 2.5Gbit/s

Reconfigurable antennas radiations using plasma Faraday cage

Indoor Channel Measurements and Communications System Design at 60 GHz

COTS-Based Modules for Far-Field Radio Frequency Energy Harvesting at 900MHz and 2.4GHz

High-power diode-pumped Q-switched Er3+:YAG single-crystal fiber laser

MODAL BISTABILITY IN A GaAlAs LEAKY WAVEGUIDE

Concepts for teaching optoelectronic circuits and systems

E H Mode Transition of an Inductively Coupled Plasma Torch at Atmospheric Pressure

X-Ray Beam Position Monitor Based on a Single Crystal Diamond Performing Bunch by Bunch Detection

RFID-BASED Prepaid Power Meter

A Switched-Capacitor Band-Pass Biquad Filter Using a Simple Quasi-unity Gain Amplifier

High acquisition rate infrared spectrometers for plume measurement

Design of the photomultiplier bases for the surface detectors of the Pierre Auger Observatory

PMF the front end electronic for the ALFA detector

Compound quantitative ultrasonic tomography of long bones using wavelets analysis

Long reach Quantum Dash based Transceivers using Dispersion induced by Passive Optical Filters

3-axis high Q MEMS accelerometer with simultaneous damping control

Time and frequency metrology accredited laboratories in Besançon

STUDY OF RECONFIGURABLE MOSTLY DIGITAL RADIO FOR MANET

MAROC: Multi-Anode ReadOut Chip for MaPMTs

Design of an Efficient Rectifier Circuit for RF Energy Harvesting System

Electrical model of an NMOS body biased structure in triple-well technology under photoelectric laser stimulation

A STUDY ON THE RELATION BETWEEN LEAKAGE CURRENT AND SPECIFIC CREEPAGE DISTANCE

A. Mandelis, R. Bleiss. To cite this version: HAL Id: jpa

Coupled optoelectronic oscillators: design and performance comparison at 10 GHz and 30 GHz

Gis-Based Monitoring Systems.

The Galaxian Project : A 3D Interaction-Based Animation Engine

Complementary MOS structures for common mode EMI reduction

PICOSECOND AND FEMTOSECOND Ti:SAPPHIRE LASERS

On the role of the N-N+ junction doping profile of a PIN diode on its turn-off transient behavior

Study on a welfare robotic-type exoskeleton system for aged people s transportation.

analysis of noise origin in ultra stable resonators: Preliminary Results on Measurement bench

A Comparison of Phase-Shift Self- Oscillating and Carrier-based PWM Modulation for Embedded Audio Amplifiers

S-Parameter Measurements of High-Temperature Superconducting and Normal Conducting Microwave Circuits at Cryogenic Temperatures

Electronic sensor for ph measurements in nanoliters

Intracavity testing of KTP crystals for second harmonic generation at 532 nm

Influence of ground reflections and loudspeaker directivity on measurements of in-situ sound absorption

Development of an On-Chip Sensor for Substrate Coupling Study in Smart Power Mixed ICs

Enhancement of Directivity of an OAM Antenna by Using Fabry-Perot Cavity

Embedded Multi-Tone Ultrasonic Excitation and Continuous-Scanning Laser Doppler Vibrometry for Rapid and Remote Imaging of Structural Defects

Design and fabrication of an asymmetric twin-core fiber directional coupler for gain-flattened EDFA

Estimation of the uncertainty for a phase noise optoelectronic metrology system

Reconfigurable Patch Antenna Radiations Using Plasma Faraday Shield Effect

Sub-Threshold Startup Charge Pump using Depletion MOSFET for a low-voltage Harvesting Application

PANEL MEASUREMENTS AT LOW FREQUENCIES ( 2000 Hz) IN WATER TANK

100 µw Coreless Flyback Converter for microbial fuel cells energy harvesting

Susceptibility Analysis of an Operational Amplifier Using On-Chip Measurement

New Structure for a Six-Port Reflectometer in Monolithic Microwave Integrated-Circuit Technology

Arcing test on an aged grouted solar cell coupon with a realistic flashover simulator

Design of induction heating lines using ELTA program

UV Light Shower Simulator for Fluorescence and Cerenkov Radiation Studies

A 180 tunable analog phase shifter based on a single all-pass unit cell

WIRELESS CHIPLESS PASSIVE MICROFLUIDIC TEMPERATURE SENSOR

Exploring Geometric Shapes with Touch

VR4D: An Immersive and Collaborative Experience to Improve the Interior Design Process

Application of CPLD in Pulse Power for EDM

QPSK-OFDM Carrier Aggregation using a single transmission chain

A low power 12-bit and 25-MS/s pipelined ADC for the ILC/Ecal integrated readout

Refraction of TM01 radially polarized mode from a chemically etched fiber

A technology shift for a fireworks controller

Thermal Behavior of High Power GaAs-Based Laser Diodes in Vacuum Environment

Characterization of Few Mode Fibers by OLCI Technique

Pushing away the silicon limits of ESD protection structures: exploration of crystallographic orientation

A Low-cost Through Via Interconnection for ISM WLP

Linear MMSE detection technique for MC-CDMA

Low temperature CMOS-compatible JFET s

Measures and influence of a BAW filter on Digital Radio-Communications Signals

Design and Realization of Autonomous Power CMOS Single Phase Inverter and Rectifier for Low Power Conditioning Applications

A sub-pixel resolution enhancement model for multiple-resolution multispectral images

A Novel Piezoelectric Microtransformer for Autonmous Sensors Applications

A new inductorless DC-DC piezoelectric flyback converter

3D MIMO Scheme for Broadcasting Future Digital TV in Single Frequency Networks

NOVEL BICONICAL ANTENNA CONFIGURATION WITH DIRECTIVE RADIATION

Transcription:

An oscillator circuit to produce a radio-frequency discharge and application to metastable helium saturated absorption spectroscopy Frédéric Moron, Lynn Hoendervanger, Marie Bonneau, Quentin Bouton, Alain Aspect, Denis Boiron, David Clément, Christoph I Westbrook To cite this version: Frédéric Moron, Lynn Hoendervanger, Marie Bonneau, Quentin Bouton, Alain Aspect, et al.. An oscillator circuit to produce a radio-frequency discharge and application to metastable helium saturated absorption spectroscopy. Review of Scientific Instruments, American Institute of Physics, 2012, 83, pp.044705. <10.1063/1.4705999>. <hal-00672488> HAL Id: hal-00672488 https://hal.archives-ouvertes.fr/hal-00672488 Submitted on 19 Nov 2015 HAL is a multi-disciplinary open access archive for the deposit and dissemination of scientific research documents, whether they are published or not. The documents may come from teaching and research institutions in France or abroad, or from public or private research centers. L archive ouverte pluridisciplinaire HAL, est destinée au dépôt et à la diffusion de documents scientifiques de niveau recherche, publiés ou non, émanant des établissements d enseignement et de recherche français ou étrangers, des laboratoires publics ou privés.

An oscillator circuit to produce a radio-frequency discharge and application to metastable helium saturated absorption spectroscopy F. Moron, A. L. Hoendervanger, M. Bonneau, Q. Bouton, A. Aspect, D. Boiron, D. Clément, and C. I. Westbrook Citation: Review of Scientific Instruments 83, 044705 (2012); doi: 10.1063/1.4705999 View online: http://dx.doi.org/10.1063/1.4705999 View Table of Contents: http://scitation.aip.org/content/aip/journal/rsi/83/4?ver=pdfcov Published by the AIP Publishing Articles you may be interested in Influence of oxygen traces on an atmospheric-pressure radio-frequency capacitive argon plasma discharge Phys. Plasmas 18, 103502 (2011); 10.1063/1.3643224 Measurements of helium metastable density in an atmospheric pressure glow discharge Appl. Phys. Lett. 85, 1487 (2004); 10.1063/1.1784514 Calculated characteristics of radio-frequency plasma display panel cells including the influence of xenon metastables J. Appl. Phys. 92, 6990 (2002); 10.1063/1.1521258 A rf discharge cell for saturated absorption spectroscopy of metastable argon Rev. Sci. Instrum. 73, 493 (2002); 10.1063/1.1433953 Microfabrication and characterization of superconducting radio-frequency oscillators Rev. Sci. Instrum. 72, 4302 (2001); 10.1063/1.1412258

REVIEW OF SCIENTIFIC INSTRUMENTS 83, 044705 (2012) An oscillator circuit to produce a radio-frequency discharge and application to metastable helium saturated absorption spectroscopy F. Moron, A. L. Hoendervanger, M. Bonneau, Q. Bouton, A. Aspect, D. Boiron, D. Clément, a) and C. I. Westbrook Laboratoire Charles Fabry, Institut d Optique, CNRS, Univ Paris Sud, 2 Avenue Augustin Fresnel, 91127 Palaiseau Cedex, France (Received 16 February 2012; accepted 7 April 2012; published online 26 April 2012) We present a rf gas discharge apparatus which provides an atomic frequency reference for laser manipulation of metastable helium. We discuss the biasing and operation of a Colpitts oscillator in which the discharge coil is part of the oscillator circuit. Radiofrequency radiation is reduced by placing the entire oscillator in a metal enclosure. 2012 American Institute of Physics. [http://dx.doi.org/10.1063/1.4705999] INTRODUCTION At present, all laser cooling and trapping of noble gas atoms is performed using atoms in excited, metastable states. 1 An important part of such manipulation experiments is an absolute optical frequency reference to maintain the laser frequency at appropriate values. Such a reference is typically provided by an auxiliary low pressure gas cell in which a plasma discharge produces metastable atoms and in which saturated absorption spectroscopy can provide a spectroscopic signal to lock a laser on a desired atomic transition. Such discharge cells have employed both direct current) 2, 3 and radiofrequency (rf) (Refs. 4 7) techniques. Radiofrequency discharges are appealing because they lend themselves to the use of sealed glass cells 5 7 with no internal metal parts. On the other hand, they have the drawback of radiating rf power into a laboratory in which many other sensitive electrical measurements are being performed, rf evaporation in a magnetic trap, for example. 8 In addition, when rf power is generated by an external oscillator and amplifier, impedance matching to the discharge coil is important and mismatch can result in substantial losses. 5 In this work, we describe an efficient, low cost apparatus which provides good spectroscopic signals in a rf-discharge cell of helium. To avoid impedance matching problems, we use a Colpitts oscillator design in which the discharge coil is included as part of the oscillator circuit. 9 We describe a biasing method which allows us to easily vary the rf amplitude during operation. This feature is important because the voltage necessary to strike the discharge is much higher than that necessary to maintain it or that which optimizes the saturated absorption signal. The entire apparatus can operate in a metal enclosure, thus limiting rf interference in the laboratory. RADIO-FREQUENCY COLPITTS OSCILLATOR The oscillator we use to generate the rf signal is a LC oscillator circuit in the Colpitts configuration 9 (see Fig. 1). The resonant frequency f 0 is close to 21 MHz before the plasma turns on in the cell. Our approach shares features with the work of Ref. 10 where the LC oscillator circuit is in the Hartley configuration. In both circuits, the resonant frequency is set by the inductance of the coil wrapped around the glass cell (as well as by the capacitors C4 and C5 offig.1), therefore avoiding any problem of impedance matching between the oscillator and the load. 5 Below we discuss our circuit in more detail and emphasize the differences with the work of May and May, 10 especially with regard to the possibility to start the plasma without the need for an additional Tesla coil. D9 1N4148 D10 1N4148 26V R4 4k7 AJ2 10k V+ 270nF/600V C3' 100pF C6 1 C3 100nF 2 3 Coil around He cell Q2 MJF18004 R2 68R/3W C4 2p2 C5 4p7 a) Author to whom correspondence should be addressed. Electronic mail: david.clement@institutoptique.fr. FIG. 1. Colpitts oscillator generating a rf signal in the frequency range of 12 21 MHz with varying amplitude V + (160 360 V) in the coil which is wrapped around a sealed glass cell containing about 1 mbar of He gas. 0034-6748/2012/83(4)/044705/4/$30.00 83, 044705-1 2012 American Institute of Physics

044705-2 Moron et al. Rev. Sci. Instrum. 83, 044705 (2012) TR1 IEC FILTERED 315mA 9V AC 26V 26V CASE GND FILTER 9V AC 2x115V 2x9V 30VA PD1 1.5A 2200µF C17 BOOST CONVERTER 26V -- 360V L11 V+ 100µH 530k Transorb1 1.5KE400A RF OSCILLATOR L on He cell 6 turns R6 2k2 AJ1 50k FIG. 2. Schematic of the electronic circuit located inside the box (see details in the Appendix). The potentiometer AJ1 controls the voltage V + and can be seen on the side of the box in the picture of Fig. 3. Starting a plasma discharge requires much more power than maintaining one. For this reason, in addition to the supply which generates the steady state power, a Tesla coil is often used to strike the discharge. 10 To avoid the Tesla coil, we supply the Colpitts circuit with a varying voltage V + allowing a high enough power to start the plasma (up to 30 W with V + = 360 V). V + is given by a boost converter the details of which can be found later in the text (see Fig. 2). Once the plasma is on, we reduce V + to a value which optimizes our spectroscopic observations (see text below and Fig. 4). The need to tune V + made us choose the scheme described in Fig. 1 where the base bias of the transistor Q2 is connected to a constant positive voltage of 26 V, unlike common base Colpitts oscillators. This solution has two advantages. First, it allows us to solve the problem of adjusting the circuit to obtain an oscillation. Indeed, oscillation of the Colpitts circuit is critical 9 and requires adjustment of the potentiometer AJ2 (see Fig. 1) to adapt the transistor base current. In the common base Colpitts configuration, the tuning of V + would require adjusting AJ2 every time V + is modified. Having a constant base bias voltage (and thus a constant transistor base current) allows a single adjustment of AJ2. Second, electrical consumption is much reduced with the constant 26 V voltage compared to that using V +. Because the inductance of the circuit depends on the state of the plasma, the resonant frequency of the circuit changes with the applied voltage V +. In our circuit, the resonant frequency with the plasma on varies from 12 MHz at V + = 140 V to 20 MHz at V + = 360 V. In practice, we start the plasma discharge at 21 MHz with full rf power ( 30 W with V + = 360 V) and then we reduce the rf amplitude to a working point where the resonant frequency happens to be 14 MHz ( 7 W with V + = 170 V). and its compactness. The converter provides enough power to start the plasma inside the cell and it allows us to later diminish the electrical power (through the potentiometer AJ1) in order to control the metastable atom density in the cell and thus the amount of absorption of the laser light propagating through the cell (see Fig. 4). Two frequency filtering elements have been placed in the electrical circuit. An inductance of 100μH (inductance L11 in Fig. 2) decouples the Colpitts oscillator from the boost converter. As the box is directly plugged onto the electrical network at 50 Hz, a line filter prohibits any leak rf signal from propagating in the electrical network. All components shown in Fig. 2 are placed in an aluminum box of thickness 3 mm to reduce the rf signal radiated into the laboratory. For such a box, the skin depth for a 10 MHz rf signal is 35 μm, corresponding to an enormous attenuation over 3 mm. However, the holes (of diameter 5 mm) ELECTRONICS INSIDE THE CLOSED BOX The oscillator is supplied by a boost converter as depicted in Fig. 2 (see Fig. 5 for a more detailed diagram). The use of a boost converter is motivated by its efficient power production FIG. 3. View inside the box with the rf-discharge cell. The box has dimensions 160 mm 160 mm 90 mm. A metal shield separates the transformer and the electronic circuit from the glass cell and the coils. Points A and B are the positions where we measure the current induced by the radiated magnetic field in a 5 cm-loop. Point A lies in front of the 5 mm hole to allow the laser beam to enter the cell.

044705-3 Moron et al. Rev. Sci. Instrum. 83, 044705 (2012) Contrast of the satured absorption peak 0.20 0.16 0.12 0.08 0.04 0 100 150 200 250 Voltage V (V) 300 350 FIG. 4. Contrast of the saturated absorption peak (red circles, left scale) and optical transmission of the laser through the glass cell (blue squares, right scale) as a function of the voltage V +. The minimum voltage to maintain the plasma is V + = 130V. Data points below this value are taken without plasma on. Lines are guides to the eyes. Inset: picture of the saturated absorption spectrum of metastable helium. The ratio of voltages V 1 /V 2 defines the contrast of the saturated absorption peak of the 2 3 S 1 2 3 P 2 transition. The other absorption peak visible corresponds to the transition 2 3 S 1 2 3 P 1. required for letting the laser beam propagate through the cell will limit the attenuation of the radiated rf signal in the laboratory. The picture in Fig. 3 shows a view inside the box. The electrical circuit of Fig. 2 is on the left-hand side and the He glass cell wrapped with the coil is on the right-hand side of the picture. Having rf signals confined inside the box avoids both problems of rf radiation from the coil wrapped around the cell as well as rf radiation through cables that would connect an external rf generator. 5 Ignition of the plasma is more difficult with the box closed, presumably because of parasitic capacitive coupling to the box walls. Indeed, the initial version of the boost converter did not provide enough power (maximum value V + 280 V instead of 360 V) to start the discharge with the box closed. Using a more powerful boost converter as described here, permits ignition of the plasma with the box closed in a few seconds. We emphasize that the conditions for ignition of the plasma are related to the properties of the coil, the box, and of the sealed glass cell (e.g., the gas pressure, about 1 mbar in our case). V 2 + V 1 1.0 0.8 0.6 0.4 0.2 0.0 Optical transmission through the cell front of the glass cell) and at point B, on the side of the glass cell, both with the box closed and opened. The attenuation of the magnetic field amplitude at point A is 100 while it is 300 at point B. This difference is expected as point A lies in front of the small hole made in the box for the laser beam to propagate. In this respect point B is more representative of the attenuation on the radiated rf magnetic field in the lab room. ABSORPTION SPECTROSCOPY SIGNAL We have used our device to lock a laser to the atomic transition 2 3 S 1 2 3 P 2 at 1083nm of metastable He. Our laser source is a fiber-laser with 2 W output power. A small part of the laser power is double-passed in an acousto-optic modulator to produce a 10 khz frequency modulation (FM) of the laser frequency. The FM laser beam is sent through the cell in a saturated absorption spectroscopy configuration. When the plasma is on, optical transmission goes to zero with increasing V + as the density of metastable helium atoms increases. In Fig. 4, we plot a measurement of the optical transmission through the glass cell as a function of the voltage V +. The transmission is unity for voltages below the minimum for maintaining the plasma (130 V). To choose the working point we monitor the amplitude of the saturated absorption peak. The contrast of this peak, plotted in Fig. 4, is defined as the ratio of the voltage amplitude of the saturated absorption peak to that of the total absorption (as sketch in the inset of Fig. 4). The optimum value for V + corresponds to the deepest saturated absorption feature. It happens to be V + 170 V, well below the ignition point of the plasma and just above the minimum voltage to maintain the plasma. Finally, an error signal is obtained from the demodulated absorption signal. 11 We easily obtain stable locking with a few mw of laser power delivered to the cell. Thanks to the use of a toroidal transformer (TR1 in Fig. 2), we observe no 50 Hz oscillations in the absorption signal. CONCLUSION With the exception of the glass cell, all the components of our device are inexpensive and readily available. The attenuation provided by the metal box seems to be adequate for continuous use in our laboratory. The design should be easily adaptable to other metastable species. MEASUREMENT OF rf ATTENUATION To test the attenuation of the radiated rf signal when the box is closed, we have measured the voltage amplitude of the signal picked up by a loop of copper wire with a diameter of 5 cm similar to that of the glass cell. The induced current in the loop is proportional to the rf magnetic field and the measured voltage amplitude is therefore proportional to the amplitude of the radiated rf field. We have repeated this measurement at point A, in front of the hole made in the box (just in ACKNOWLEDGMENTS We acknowledge support from the Triangle de la Physique Contract No. 2010-062T, the IFRAF Institute, the ANR and the ERC Grant No. 267 775 Quantatop. APPENDIX See Fig. 5.

044705-4 Moron et al. Rev. Sci. Instrum. 83, 044705 (2012) FIG. 5. Complete schematic diagram of the electronics. 1 W. Vassen et al., Rev. Mod. Phys. 84, 175 (2012). 2 W. Lu, D. Milic, M. D. Hoogerland, M. Jacka, K. Baldmwin, and S. J. Buckman, Rev. Sci. Instrum. 67, 3003 (1996). 3 J. E. Lawler, A. Siegel, B. Couillaud, and T. W. Hansch, J. Appl. Phys. 52, 4375 (1981). 4 C. Y. Chen, K. Bailey, Y. M. Li, T. P. O Connor, Z.-T. Lu, X. Du, L. Young, and G. Winkler, Rev. Sci. Instrum. 72, 271 (2001). 5 C. I. Sukenik and H. C. Busch, Rev. Sci. Instrum. 73, 493 (2002). 6 D. I. Hoult and C. M. Preston, Rev. Sci. Instrum. 63, 1927 (1992). 7 J. C. J. Koelemeij, W. Hogervorst, and W. Vassen, Rev. Sci. Instrum. 76, 033104 (2005). 8 A. Browaeys, A. Robert, O. Sirjean, J. Poupard, S. Nowak, D. Boiron, C. I. Westbrook, and A. Aspect, Phys.Rev.A64, 034703 (2001). 9 A. E. Clifford and A. H. Wing, Electronic Circuits and Tubes (McGraw- Hill, New York, 1947). 10 R. D. May and P. H. May, Rev. Sci. Instrum. 57, 2242 (1986). 11 N. Vansteenkiste, C. Gerz, R. Kaiser, C. Salomon, and A. Aspect, J. Phys. II 1, 1407 1428 (1991).