Meeting Measurement Challenges For Low-Power, Pulsed, Or Modulated Light Sources

Similar documents
Optical Power Meter Basics

Goals of the Lab: Photodetectors and Noise (Part 2) Department of Physics. Slide 1. PHYSICS6770 Laboratory 4

NON-AMPLIFIED PHOTODETECTOR USER S GUIDE

GFT1504 4/8/10 channel Delay Generator

Pulse Timing and Latency Measurements Using Wideband Video Detectors

Model 310H Fast 800V Pulse Generator

LB480A Pulse Profiling USB PowerSensor+ Data Sheet

Pyroelectric, Photodiode and RP Heads for Repetitive Energy Measurements

LB680A Pulse Profiling USB PowerSensor+ Data Sheet

SIGNAL RECOVERY: Sensors, Signals, Noise and Information Recovery

Optical to Electrical Converter

Data Sheet. Peak, CW & Average. Power Sensors. Taking performance to a new peak

Novel laser power sensor improves process control

2014 PRODUCT SELECTION GUIDE

DESIGN OF AN ANALOG FIBER OPTIC TRANSMISSION SYSTEM

Modular Controller. Key Features. Model P hotonics. Comprehensive laser diode protection features. Advanced 16-bit control technology

HFD3029. Schmitt Input, Non-Inverting TTL Output Receiver

NON-AMPLIFIED HIGH SPEED PHOTODETECTOR USER S GUIDE

ModBox-FE-NIR Near-Infra Red Front-End Laser Source

771 Series LASER SPECTRUM ANALYZER. The Power of Precision in Spectral Analysis. It's Our Business to be Exact! bristol-inst.com

LB480A Pulse Profiling USB PowerSensor+ Data Sheet

RF Generators. Requirements:

3.003 Lab 3 Part A. Measurement of Speed of Light

Class Room Experiments on Laser Physics. Alika Khare

RTD), LM335 Voltage, or AD592 current. c. Constant TE Current (8350 only)

HFD /XXX. h 479. Schmitt Input, Non-Inverting TTL Output Receiver

High-Power Highly Linear Photodiodes for High Dynamic Range LADARs

pulsecheck The Modular Autocorrelator

HIGH SPEED FIBER PHOTODETECTOR USER S GUIDE

APE Autocorrelator Product Family

Agilent 83711B and 83712B Synthesized CW Generators

R. J. Jones College of Optical Sciences OPTI 511L Fall 2017

Arbitrary/Function Waveform Generators 4075B Series

POLYTECHNIC UNIVERSITY Electrical Engineering Department. EE SOPHOMORE LABORATORY Experiment 3 The Oscilloscope

2520 Pulsed Laser Diode Test System

New automated laser facility for detector calibrations

It s Our Business to be EXACT

Wavelength Meter Sensitive and compact wavemeter with a large spectral range for high speed measurements of pulsed and continuous lasers.

Amplified Photodetectors

INSTRUMENTATION SELECTION GUIDE

The ASTRI SST-2M Illuminator

Fiber Optic Communication Link Design

R a) Explain the operation of RC high-pass circuit when exponential input is applied.

GFT Channel Digital Delay Generator

DIGITAL UTILITY SUB- SYSTEMS

Physics 262. Lab #1: Lock-In Amplifier. John Yamrick

Information & Instructions

WORLD LEADING PRODUCTS FOR LASER SCIENTISTS AND ENGINEERS

ASTABLE MULTIVIBRATOR

Measure the roll-off frequency of an acousto-optic modulator

Sensitivity of Series Direction Finders

DETECTING THE RATIO OF I AC

the reactance of the capacitor, 1/2πfC, is equal to the resistance at a frequency of 4 to 5 khz.

Experiment No.15 DC-DC Converters

Detecting the Ratio of I ac. /I ave. photoelastic modulators

Traceability for Oscilloscopes and Oscilloscope Calibrators

CHAPTER 5 FINE-TUNING OF AN ECDL WITH AN INTRACAVITY LIQUID CRYSTAL ELEMENT

EE 3101 ELECTRONICS I LABORATORY EXPERIMENT 9 LAB MANUAL APPLICATIONS OF IC BUILDING BLOCKS

EE 330 Laboratory 8 Discrete Semiconductor Amplifiers

Gentec-EO USA. T-RAD-USB Users Manual. T-Rad-USB Operating Instructions /15/2010 Page 1 of 24

Multiple Instrument Station Module

Homework Set 3.5 Sensitive optoelectronic detectors: seeing single photons

ECEN 4606, UNDERGRADUATE OPTICS LAB

11Beamage-3. CMOS Beam Profiling Cameras

EXPERIMENT 2.2 NON-LINEAR OP-AMP CIRCUITS

Lab 0: Orientation. 1 Introduction: Oscilloscope. Refer to Appendix E for photos of the apparatus

WORLD LEADING PRODUCTS FOR LASER SCIENTISTS AND ENGINEERS

Keysight Technologies Optical Power Meter Head Special Calibrations. Brochure

UNIVERSITY OF NORTH CAROLINA AT CHARLOTTE Department of Electrical and Computer Engineering

The Sampling Theorem:

Class #9: Experiment Diodes Part II: LEDs

OFC SYSTEM: Design Considerations. BC Choudhary, Professor NITTTR, Sector 26, Chandigarh.

Electronic Circuits I Laboratory 03 Rectifiers

U1604A Handheld Oscilloscopes, 40 MHz

Optimizing Averaging for Better Power Measurements

Cost-Effective Traceability for Oscilloscope Calibration. Author: Peter B. Crisp Head of Metrology Fluke Precision Instruments, Norwich, UK

University of Washington Department of Electrical Engineering EE 351: Introduction to Energy Devices and Systems. Lab 1: Power Electronic Converters

Status of Primex Beam Position Monitor July 29 th, 2010

MEASURING HUM MODULATION USING MATRIX MODEL HD-500 HUM DEMODULATOR

PLL Synchronizer User s Manual / Version 1.0.6

Exp 3 COLCULATE THE RESPONSE TIME FOR THE SILICON DETECTOR

Lab 1: Basic Lab Equipment and Measurements

Mach 5 100,000 PPS Energy Meter Operating Instructions

Supplementary Information

671 Series LASER WAVELENGTH METER. The Power of Precision in Wavelength Measurement. It's Our Business to be Exact! bristol-inst.

quantiflash Calibration Light Source for Cytometry

Keysight Technologies N9320B RF Spectrum Analyzer

Beam Diagnostics, Low Level RF and Feedback for Room Temperature FELs. Josef Frisch Pohang, March 14, 2011

Agilent N9320B RF Spectrum Analyzer

BEAMAGE-3.0 KEY FEATURES BEAM DIAGNOSTICS AVAILABLE MODELS MAIN FUNCTIONS SEE ALSO ACCESSORIES. CMOS Beam Profiling Cameras

Draw in the space below a possible arrangement for the resistor and capacitor. encapsulated components

Directly Chirped Laser Source for Chirped Pulse Amplification

Operational Amplifiers

ERS KEY FEATURES BEAM DIAGNOSTICS MAIN FUNCTIONS AVAILABLE MODEL. CMOS Beam Profiling Camera. 1 USB 3.0 for the Fastest Transfer Rates

Experiment 1.A. Working with Lab Equipment. ECEN 2270 Electronics Design Laboratory 1

LadyBug Technologies, LLC LB5918L True-RMS Power Sensor

Real-Time Phase Noise Analyzer

ME 365 EXPERIMENT 1 FAMILIARIZATION WITH COMMONLY USED INSTRUMENTATION

Silicon Photodiodes - SXUV Series with Platinum Silicide Front Entrance Windows

Cornerstone Electronics Technology and Robotics Week 21 Electricity & Electronics Section 10.5, Oscilloscope

Transcription:

Meeting Measurement Challenges For Low-Power, Pulsed, Or Modulated Light Sources By Denise Ullery, Sylvia Tan, and Jay Jeong, Newport Corporation (www.newport.com) Traditionally, power meters have been used for continuous-wave (CW) light sources in conjunction with photodiode or thermopile detectors for average power measurements, while energy meters and pyroelectric detectors have been used for pulsed laser beam applications. Recent advancements in photonics technologies have broadened the applications horizon into many different areas. An increasingly larger number of complex and diverse forms of light sources are designed into various systems and experimental setups. These complex light sources drive the need for more sensitive and versatile power measurement capabilities with unprecedented accuracy. It is particularly challenging to measure a low-average-power pulsed light source, such as an LED or a low-power laser diode, especially when the electrical designers do not know the detailed optical characteristics of the light source. In addition to average optical power, many other parameters about the light source are important to obtain. For instance, what does the optical signal look like? Key information such as shape, frequency, and power level is needed to determine the system power budget and the overall system performance, as well as to validate the effectiveness of lower cost components. It is also important to identify which signal characteristics the detector or the detection system will measure. Researchers and engineers are interested in the pulse width and pulse shape of the signal. Typically, one would use an optical power meter to measure the pulse energy or the average power by using a meter and a National Institute of Standards and Technology (NIST)-calibrated detector. A biased detector, connected to an oscilloscope, would be used to determine the pulse shape and to measure frequency and other temporal information.

Recent developments in power meter technology have now enabled decent temporal signal measurements, eliminating the need for multiple instruments. One of the most advanced power meters in the market has been designed by Newport Corporation, Model 2936-C. This state-of-the-art, dualchannel optical power meter has several sophisticated features that allow the user to achieve various measurements, including frequency measurements of pulses up to 250 khz (see figure 1). Figure 1: Newport Corporation s optical power meter Model 2936-C with the 918D Series Photodiodes allows various measurements for pulsed or modulated light input as well as continuous wave. Analog And Digital Filtering The analog filter is an electronic, low-pass filter implemented on an optical power meter while the digital filter averages the most recent 10, 100, 1,000, or 10,000 samples. The 2936-C optical power meter has selectable 5 Hz, 1 khz, 12.5 khz, and 250 khz low-pass filters. An optical power meter is essentially an amp meter with very low noise and high accuracy. Therefore, the proper filter setting would make a huge impact on the signal response of the optical meter. For the best CW measurement or average power measurement of a modulated/pulsed light source, the heaviest filtering would be utilized, i.e., the 10,000 point digital filter with the 5 Hz analog filter. For a modulated or pulsed signal, an appropriate analog filter must be selected to obtain the best waveform, while the digital filter must not be used at all. Moreover, applying the digital and the analog filters on the pulsed or modulated input signal effectively averages out the signal.

Statistics And Time Plot Robust statistical measurement capability, such as average, minimum, maximum, and standard deviation is a standard feature in many optical power meters nowadays. With a photodiode detector, Newport s 2936-C power meter can log a data point every 0.1 msec, up to 250,000 data samples per channel, which can be easily stored on a USB thumb drive for further analysis using a spreadsheet program like Excel. In addition, the collected data can be viewed in a scope-like time plot directly from the meter display, as shown in figure 2. One can easily see whether the input signal is the expected form or if it is distorted due to overly heavy analog filtering, excessive noise, or any other external influences. Figure 2: The time plot of the data stored by the internal sample storage. Peak-To-Peak And RMS Measurement With the fast 250 khz sampling rate of the 2936-C, modulated and pulsed signals can be measured in the Peak-to-Peak Mode. The power meter will determine the difference between the maximum and the minimum value within the given time window, which can be defined by the user. For different measurement ranges of the power meter, there are trade-offs between bandwidth and noise level. Therefore, when using large bandwidth ranges, it is important that the signal being measured has large enough power to ensure a good signal-to-noise ratio. The RMS (root mean square) measurement is another measurement mode available in the 2936-C optical power meter.

Triggering Triggering is an especially useful feature when synchronizing the measurement with an external event. Typically, an external event generates a pulsed light output, whose value can be measured in the CW or Peak-to-Peak mode. In the Newport 2936-C power meter, it is possible to set the trigger to be responsive to an external transistor-transistor logic (TTL) signal, the press of a soft key trigger by the user, or to a computer command. Testing And Results To illustrate some of the features and capabilities available with the Newport optical power meter, a series of tests were run with a 50% duty cycle to compare the performance of a Newport 818-BB-21 fast-biased detector to a Newport 818-SL/DB semiconductor detector. The repetition rate for all tests was 1 khz with a pulse width of 500 μs. We connected the 818-BB-21 detector directly to channel 1 of the oscilloscope. The 818-SL/DB detector was connected to the 2936-C optical power meter. We then connected the analog output of the 2936-C to channel 2 of the oscilloscope. With a 50% duty cycle we noticed nearly identical waveforms when comparing the output of the 818- BB-21 to that of the analog output of the 2936-C (see figure 3). To measure CW power, we used the 2936-C digital filter set to 10,000 samples and the analog filter set to 5 Hz. The 2936-C measured a CW power of 67.76 μw and peak-to-peak power of 134.5 μw. The peak power as measured by the 2936-C (no filters, manual range mode) is in close agreement with the theoretical peak power.

Figure 3: Measurement comparison of a pulsed optical signal, 1 khz repetition rate with a 50% duty cycle, as measured by a Newport Model 818-BB-21 biased detector (top waveform) and an 818-SL/DB (bottom waveform) silicon photodiode detector. In conclusion, recent improvements in optical power meter technology enable more complex measurements, such as power measurement of low-power pulsed or modulated light sources. We constructed a simple pulsed light source measurement using a Newport optical power meter and a fastbiased detector. Comparable pulse shapes and the resultant NIST-traceable average optical power in the CW measurement mode and the amplitude in the Peak-to-Peak mode were clearly demonstrated.