Development of a Low Cost ECG Data Acquisition Module

Similar documents
Biomedical Instrumentation (BME420 ) Chapter 6: Biopotential Amplifiers John G. Webster 4 th Edition

BME 405 BIOMEDICAL ENGINEERING SENIOR DESIGN 1 Fall 2005 BME Design Mini-Project Project Title

Indigenous Design of Electronic Circuit for Electrocardiograph

Portable, Low Cost, Low Power Cardiac Interpreter

Bio-Potential Amplifiers

Development of Electrocardiograph Monitoring System

AC-Coupled Front-End for Biopotential Measurements

ECE 480 Design Team 6 Electrocardiography and Design

INTERFERENCE REDUCTION IN ECG RECORDINGS BY USING A DUAL GROUND ELECTRODE

IMPROVEMENTS IN ELECTROCARDIOGRAPHY SMOOTHENING AND AMPLIFICATION

Lecture 4 Biopotential Amplifiers

Data acquisition and instrumentation. Data acquisition

EE 230 Experiment 10 ECG Measurements Spring 2010

REAL-TIME WIRELESS ECG AND ITS SIGNAL DISPLAY ON LABVIEW

Electrocardiogram (EKG) Data Acquisition and Wireless Transmission

Design and Implementation of Digital Stethoscope using TFT Module and Matlab Visualisation Tool

STM32 microcontroller core ECG acquisition Conditioning System. LIU Jia-ming, LI Zhi

TRANSDUCER INTERFACE APPLICATIONS

Wireless Transmission of Real Time Electrocardiogram (ECG) Signals through Radio Frequency (RF) Waves

COMPARISON OF VARIOUS FILTERING TECHNIQUES USED FOR REMOVING HIGH FREQUENCY NOISE IN ECG SIGNAL

BIOMEDICAL INSTRUMENTATION PROBLEM SHEET 1

Biomedical. Measurement and Design ELEC4623. Lectures 9 and 10 Practical biopotential amplifier design and multilead ECG systems

ELR 4202C Project: Finger Pulse Display Module

Electrocardiogram (ECG)

AD8232 EVALUATION BOARD DOCUMENTATION

Design on Electrocardiosignal Detection Sensor

Nizamuddin M., International Journal of Advance Research, Ideas and Innovations in Technology.

CHAPTER 7 HARDWARE IMPLEMENTATION

Changing the sampling rate

CHAPTER 3. Instrumentation Amplifier (IA) Background. 3.1 Introduction. 3.2 Instrumentation Amplifier Architecture and Configurations

A Body Area Network through Wireless Technology

Kanchan S. Shrikhande. Department of Instrumentation Engineering, Vivekanand Education Society s Institute of.

A Comprehensive Model for Power Line Interference in Biopotential Measurements

GENERATION OF SIGNALS USING LABVIEW FOR MAGNETIC COILS WITH POWER AMPLIFIERS

Implementation of wireless ECG measurement system in ubiquitous health-care environment

Wireless ECG System Based on ARM LPC 2103 Processor

Operational Amplifier BME 360 Lecture Notes Ying Sun

Florida Atlantic University Biomedical Signal Processing Lab Experiment 2 Signal Transduction: Building an analog Electrocardiogram (ECG)

Ballistocardiograph 1

NOISE REDUCTION TECHNIQUES IN ECG USING DIFFERENT METHODS Prof. Kunal Patil 1, Prof. Rajendra Desale 2, Prof. Yogesh Ravandle 3

EDL Group #3 Final Report - Surface Electromyograph System

A SMART METHOD FOR AUTOMATIC TEMPERATURE CONTROL

6.101 Introductory Analog Electronics Laboratory

Filtration Of Artifacts In ECG Signal Using Rectangular Window-Based Digital Filters

Name Kyla Jackson, Todd Germeroth, Jake Spooler Date May 5, 2010 Lab 3E Group 3 Experiment Title Project Deliverable 3

In-depth Analysis of Cardiac Signals Using Novel Equipment and Software

Electromagnetic Compatibility to Bio-Medical Signals Using Shielding Methods

INTERNATIONAL JOURNAL OF ENGINEERING SCIENCES & RESEARCH TECHNOLOGY

EMG Electrodes. Fig. 1. System for measuring an electromyogram.

A Reliable Non-Contact ECG Measurement System with Minimal Power Line Disturbance

Fault Diagnosis in H-Bridge Multilevel Inverter Drive Using Wavelet Transforms

ELG3336 Design of Mechatronics System

Biomedical Sensor Systems Laboratory. Institute for Neural Engineering Graz University of Technology

About the Tutorial. Audience. Prerequisites. Copyright & Disclaimer. Linear Integrated Circuits Applications

Massachusetts Institute of Technology MIT

Deepali Shukla 1 (Asst.Professor), Vandana Pandya 2 (Asst.Professor) Medicaps Institute of Technology & Management, Indore (M.P.

International Journal of Electronics and Communication Engineering & Technology (IJECET), ISSN COMMUNICATION ENGINEERING & TECHNOLOGY (IJECET)

DESIGNING A VIRTUAL MACHINE FOR IDENTIFICATION OF CARDIAC ARRHYTHMIAS USING LAB VIEW

Design of CMOS Instrumentation Amplifier

A PC Based Cost Effective Advanced Cardio Signals Monitoring System

A Low-Noise AC coupled Instrumentation Amplifier for Recording Bio Signals

InstrumentationTools.com

DESIGN OF A LOW COST EMG AMPLIFIER WITH DISCREET OP-AMPS FOR MACHINE CONTROL

A Lower Transition Width FIR Filter & its Noise Removal Performance on an ECG Signal

ANALYSIS AND DESIGN OF HIGH CMRR INSTRUMENTATION AMPLIFIER FOR ECG SIGNAL ACQUISITION SYSTEM USING 180nm CMOS TECHNOLOGY

EECE Circuits and Signals: Biomedical Applications. Lab ECG I The Instrumentation Amplifier

Keywords: Data Acquisition, ECG, LabVIEW, Virtual instrumentation

A Design Of Simple And Low Cost Heart Rate Monitor

WIRELESS ELECTRONIC STETHOSCOPE USING ZIGBEE

Designing and Implementation of Digital Filter for Power line Interference Suppression

Laboratory Project 1B: Electromyogram Circuit

LOW VOLTAGE / LOW POWER RAIL-TO-RAIL CMOS OPERATIONAL AMPLIFIER FOR PORTABLE ECG

Instrumentation amplifier

III Lead ECG Pulse Measurement Sensor

IMPLEMENTATION OF DIGITAL FILTER ON FPGA FOR ECG SIGNAL PROCESSING

Applied Electronics II

CHAPTER 7 INTERFERENCE CANCELLATION IN EMG SIGNAL

Brain-computer Interface Based on Steady-state Visual Evoked Potentials

DESIGN OF OTA-C FILTER FOR BIOMEDICAL APPLICATIONS

LabVIEW Based Biomedical Signal Acquisition and Processing

Noise Reduction Technique for ECG Signals Using Adaptive Filters

Texas Components - Data Sheet. The TX53G1 is an extremely rugged, low distortion, wide dynamic range sensor. suspending Fluid.

BME 701 Lecture 1. Measurement and Instrumentation

EMG click PID: MIKROE-2621

PORTABLE ECG MONITORING APPLICATION USING LOW POWER MIXED SIGNAL SOC ANURADHA JAKKEPALLI 1, K. SUDHAKAR 2

UNIT I. Operational Amplifiers

ELEC4623 / ELEC9734 BIOMEDICAL ENGINEERING LABORATORY 3: DESIGN, TESTING AND ANALYSIS OF A HIGH QUALITY ISOLATED BIOPOTENTIAL AMPLIFIERS

C H A P T E R 02. Operational Amplifiers

Analysis of Instrumentation Amplifier at 180nm technology

EE 300W 001 Lab 2: Optical Theremin. Cole Fenton Matthew Toporcer Michael Wilson

IMPLEMENTATION OF REAL TIME BRAINWAVE VISUALISATION AND CHARACTERISATION

Laboratory Project 1: Design of a Myogram Circuit

Spring 2014 EE 445S Real-Time Digital Signal Processing Laboratory Prof. Evans. Homework #2. Filter Analysis, Simulation, and Design

Lecture 14 Interface Electronics (Part 2) ECE 5900/6900 Fundamentals of Sensor Design

PHYS225 Lecture 15. Electronic Circuits

Ideal Op Amps. The Two Golden Rules for circuits with ideal op-amps*

Bioelectric Signal Analog Front-End Module Electrocardiograph

Two stage Cascade BJT Amplifier

A Novel Low Noise High Gain CMOS Instrumentation Amplifier for Biomedical Applications

Chapter 4 4. Optoelectronic Acquisition System Design

Transcription:

Development of a Low Cost ECG Data Acquisition Module Deboleena Sadhukhan 1, Rohit Mitra 2, Avik Kundu 2, Madhuchhanda Mitra 3 Research Scholar, Department of Applied Physics, University of Calcutta, Kolkata, West Bengal, India 1 M.Tech Student, Department of Applied Physics, University of Calcutta, Kolkata, West Bengal, India 2 Associate Professor, Department of Applied Physics, University of Calcutta, Kolkata, West Bengal, India 3 ABSTRACT: The paper describes the development of a low cost and simple amplifier circuit for ECG acquisition from a single lead. The acquisition circuit uses clip-type flat metal plate limb electrodes to sense the heart signals and a basic amplifier circuit is designed using JFET OP-AMP IC LF-353 with the required gain to suitably amplify the signal. The amplified data fed into a computer using USB-69 is then denoised, processed and displayed using LabVIEW software. The developed ECG acquisition module is evaluated by visual comparison of simultaneously recorded data acquired by the module with and by the MP-15 amplifier system from BIOPAC Systems Inc. Tests have been performed in the laboratory on several volunteers in the age group of 28-6 and the results were quiet satisfactory. KEYWORDS: ECG acquisition, instrumentation amplifier, LabVIEW software, filtering. I.INTRODUCTION The electrocardiogram (ECG) is the recording of the electrical activity of the myocardium of the heart during one cardiac cycle and is characterized by a recurrent sequence of P, QRS, T and a conditional U wave [1]. The electrical impulses within the heart act as a source of voltage, which generates a current flow in the torso and corresponding potentials on the skin. ECG is recorded by placing electrodes on the body surface and a standard 12 lead system is used to get an overall view of the heart s activity [2]. The most prevalent and significant among these is Lead II for diagnosing rhythm problems. Signals from Lead II measure the variations in potential between the right arm and the left leg, with the electrode of the left arm acting as the ground. Cardiovascular diseases are one of the predominant causes of death, all over the world. ECG is used as an important diagnostic tool for various cardiac diseases due to its ability to correlate the different ECG wave signatures with the actual operation of the heart and its ease of recording in a non-invasive manner. Hence development of EKG acquisition hardware has impacted the progression of research in electrophysiology. Design of different amplifier circuits suitable for ECG acquisition has been proposed in literature [3-6]. Several researches have been done for development of wireless ECG acquisition module also [7, 8]. The main advantage of the wireless measurement technology is the increased patients mobility, degree of freedom and convenience since it is not restricted by lead wires. But most of them are not cost effective and involve complex circuitry. Presently, different ECG acquisition devices are commercially available, but most of them are highly priced due to the use of inbuilt isolation and filter circuits. But in India a vast majority of the country belongs to the rural areas. Hence there is an immense need for development of a low-cost indigenous ECG acquisition system. In this paper we propose a low cost and simple acquisition circuitry for ECG data acquisition from a single lead as shown in Fig. 1. Signal from the clip-type flat metal plate limb electrodes is suitably amplified using a simple instrumentation amplifier circuit designed with high input impedance JFET OP-AMP ICS. No separate isolation and hardware filtering has been used to avoid circuit complexity and signal distortion due to use of additional components. The amplified signal is then fed into a PC using USB-69. The raw signal is then processed and displayed on the monitor using LabVIEW software and also stored for future analysis. Copyright to IJIRSET www.ijirset.com 148

Patient ECG ELECTRODES AMPLIFIER CIRCUIT NI USB- 69 LabVIEW PROGRAMMING (data processing) ECG display Fig. 1 Block diagram of the proposed ECG acquisition module II.HARDWARE DESCRIPTION The hardware part of the module consists of the ECG sensors or the electrodes, the amplifier circuit with suitable characteristics, power supply for the amplifier and an USB module for data transmission to the computer. For recording of ECG signal, the standard 12-Lead electrode configuration is commonly used which measures the potential at the frontal plane and the transverse plane of the body. The standard limb leads comprising of Lead-1, Lead-2 and Lead-3 [2] are the basic measuring leads. The leads measure potentials at one part of the body with respect to another part and hence do not require any reference electrode (Wilson Central Terminal), as needed in Augment Unipolar Leads or Unipolar Chest Leads. A. ECG Electrodes In order to measure and record potentials, it is necessary to provide some interface between the body and the electronic measuring apparatus. Bio-potential electrodes serve as this interface which acts as a transduction section thus converting the ionic currents in the body to electronic currents at the metal-electrolyte interface. These transducers, known as bare-metal or recessed electrodes, generally consist of a metal such as silver or stainless steel, with a jelly electrolyte that contains chloride and other ions. A clip-attached system flat metal plate electrode is used instead of disposable foam-pad electrodes due to their cost effectiveness and adaptability for continuous usage over longer periods. Fig. 2. Clip type flat metal plate electrodes B. Amplifier circuit The amplifier circuit for ECG measurement must be able to deal with the extremely weak nature of the signal of the order of few microvolts and also its low drive. The requirements for a typical ECG amplifier [2] include the following- Capability to sense low amplitude signals in the range of.5 1 mv Very high input impedance, > 5 Mega-ohms Very low input leakage current, < 1 micro-amp Flat frequency response of.5 15 Hz Copyright to IJIRSET www.ijirset.com 149

High Common Mode Rejection Ratio To meet all the specifications JFET OP-AMP IC LF-353 was chosen. Additionally, it is extremely low priced as compared to the other commonly used biomedical instrumentation amplifier ICs and its low input bias current range provides the basic isolation between patient s body and the electronic components and thus eliminates the need of separate isolation for the limb leads. High common mode rejection ratio (CMRR) and low-input voltage noise are two of the most desirable features for biopotential amplifiers in order to, respectively, reject external interference and enable high-resolution measurements. High common mode input impedance is also desirable in order to achieve a high CMRR despite the electrode s impedance mismatch [4]. All these objectives are easily achieved by using a high-gain instrumentation amplifier circuit as shown in Fig. 3. Fig. 3. Circuit diagram of the amplifier circuit The inputs are buffered or isolated using the voltage followers. The second stage is the differential amplifier providing the necessary gain. A final non-inverting amplifier stage is added to provide the required additional gain. The common mode voltage available at the point A is further minimized by using an Op-AMP A3. The D.C. components in the common- mode signal are blocked by the capacitor while the A.C. components are grounded using the negative feedback concept. The common mode point A can also be used to implement a Right Leg Drive circuit for augmentation in the removal of common mode signals and also ensuring patient safety. A non-inverting amplifier is also used after op-amp A4 for further enhancement of the amplified signal derived from the Instrumentation Amplifier block. Gain of the Instrumentation Amplifier block only is given by:- Av1 = V1 / (V1 V2) = (1 + 2 R2 / R1) (R5 / R4). Gain of the Non-Inverting Amplifier (N.I.) is given by Av2 = (1 + R7 / R6) Therefore, total gain of the amplifier circuit is given as:- Av = Av1 * Av2 = (1 + 2 R2 / R1) (R5 / R4) (1 + R7 / R6) We choose a gain of 9 for Instrumentation Amplifier part comprising of op-amps A1, A2, A3 and A4 and a value of 1 kilo ohm (standard value) for R2. Keeping R5=R4=R2 for good control over C.M.R.R., the value of R1 can be Copyright to IJIRSET www.ijirset.com 15

derived as 224.71ohm but unfortunately this is not a standard value. We use R1 = 22ohm (standard resistor value with +/- 1% tolerance) and this gives a gain of approximately Av1 = 92. For Non-Inverting Amplifier R6 and R7 are chosen as 1K and 2K respectively and this gives a gain of Av2 =3. Therefore, total gain of the amplifier circuit is Av=92 * 3 =276. The gain is sufficient to suitably amplify the ECG signals for different heart abnormalities like Hyperkalemia which has high T-wave amplitudes and small or indiscernible P-waves used by excess potassium in the body. No separate isolation and hardware filtering has been used to avoid circuit complexity and signal distortion due to use of additional components The circuit is fabricated on a bread board and powered from a standard +/- 12V regulated power supply as shown in Fig.4. The amplified ECG signal from the amplifier circuit is then fed into a PC using USB-69 made by National Instruments. Fig.4. Fabricated hardware III. SOFTWARE DESCRIPTION FOR ECG PROCESSING & DISPLAY LabVIEW, shortform for Laboratory Virtual Instrument Engineering Workbench, is a programming environment in which programs can be created using a graphical notation (connecting functional nodes via wires through which data flows). A program is made for preliminary level processing of the acquired raw data and displaying it on the computer screen and also storing the records for future analysis. The block diagram and the front panel are shown in Fig. 7 and Fig. 8, respectively. A. Programming for ECG Data Acquisition The configuration of the data acquisition from USB-69 module is done by using the DAQ ASSISTANT function which is a graphical interface that is used to configure the measurement tasks (configured through MAX) and channels on the DAQ card. For a NI USB 69 having a 14- bit resolution ADC, we select a sampling frequency much more than the Nyquist sampling frequency. Since the ECG signal is a bipolar signal, the input voltage is selected to be +/-1 V that gives a total range of 2 Volts. This is done because the maximum amplitude in the ECG signal is that of R-peak with a value of 1.6mV and the smallest detectable change in the input voltage called the code width (Vcw) is then given as Vcw = I/P Voltage Range / 2 RESOLUTION = 2 / 2 14 = 1.2mV i.e. sufficient to detect R-Peak. The terminal configuration parameter is selected to be differential since an ideal differential measurement system reads only the potential difference between its two terminals- the (+) and the ( ) inputs. It completely rejects any voltage present at the instrumentation amplifier inputs with respect to the amplifier ground. In other words, an ideal differential measurement system completely rejects the common mode voltage. The acquisition time can also be fixed according to the user requirement. Copyright to IJIRSET www.ijirset.com 151

B. Programming for ECG Filtering ECG signals are routinely contaminated by noise due to motion artefacts, power line noise, and electrode contact noises, all of which decreases the accuracy of ECG interpretation. The goal of ECG filtering is to separate the valid cardiac components from the background noises so as to obtain a signal that is qualified for reliable interpretation. Different kinds of digital filters have been used in literature for ECG filtering [9,1]. The main source of artefact in the acquired ECG signal displayed on the LabVIEW front panel was Power Line Interference and slow base- line wandering and some high frequency components. In India, the power line frequency is 5 Hz. So an IIR band stop filter in the frequency range of 48Hz to 52Hz with a much higher order and Butterworth topology is used for power line removal as shown in the Fig. 5. Fig. 5. Bandstop filter configuration for power line noise removal The high frequency components were removed by a Moving Average (Smoothing) Filter. It is probably the most optimal smoothing filter in the time domain but an exceptionally bad low-pass filter in the frequency domain with the roll-off very slow and stop-band attenuation ghastly. Clearly, the moving average filter cannot separate one band of frequencies from another but for signals buried in random noise, the smoothing action of moving average filter is very good but it reduces the sharpness of the edges as well. Multiple-pass moving average filters involve passing the input signal through a moving average filter two or more times. Two passes are equivalent to using a triangular filter kernel which is better than a rectangular filter kernel as a triangular filter kernel is a rectangular filter kernel convolved with itself, therefore the same is used in our case too. Fig- 6 shows the configuration settings for the moving average filter. The following Fig. 7 and Fig. 8 shows the complete block diagram programming and the corresponding ECG display on the front panel of the LabVIEW. The filtered ECG data is also stored in measurement files for future analysis and display. The storage file can be specified by the user. Copyright to IJIRSET www.ijirset.com 152

Fig. 6. Smoothing filter configuration for high frequency removal Fig. 7. Overall block diagram for ECG filtering and display Copyright to IJIRSET www.ijirset.com 153

Fig. 8. ECG display on the LabVIEW front panel IV. EXPERIMENTAL RESULTS The developed ECG acquisition module is tested in the laboratory to acquired leadii ECG data from several volunteers in the age group of 28-6, both male and female. Fig 9 shows the complete experimental set up. Performance evaluation of the developed module is performed by visual comparison of simultaneously recorded data acquired by the module and by the MP-15 amplifier system from BIOPAC Systems Inc. A few of the visual comparison test results plotted in Matlab 7.1 are shown in Fig 1. Fig.9. Overall experimental setup Copyright to IJIRSET www.ijirset.com 154

-1 N M V ORIGINAL DATA DENOISED DATA -1.5 1 2 3 4 5 6 7 8 9 1 1.5 5, 1, m V Ampliteve in mv PID 1/S1 LEAD 2 ORIGINAL DATA.5 input SNR = 13.64 db A M P L I T U D E -1-1.5 1 2 3 4 5 6 7 8 9 1 1 a m p l i t u d e.5 output SNR = 32.89 db 5, 1, -1-1 1 2 3 4 5 6 7 8 9 1 1 2 3 4 5 6 7 8 9 1 Time in msec..5 a m p l i t u d e i n m DENOISED DATA (a) (b) Fig. 1. Visual comparison test results a) Data recorded by MP-15 system by Biopac b) Simultaneous data recorded by the developed acquisition module The test results reveal quiet a satisfactory performance of the developed module. As shown in the fig. 1, the different ECG waveforms were well acquired by the module. The wave signatures of the ECG beats were well correlated in both the data sets. However, due to the use of filtering in the developed module, the amount of noise present in the data varied. Also, acquisition of chest lead ECG has been purposely avoided to ensure patient safety. IV. CONCLUSION The performance of the developed module was quiet satisfactory. The acquisition circuitry does not involve any costly components and hence is very cheap and simple as compared to the other commercial ECG acquisition modules. However, as the acquisition circuitry does not use any separate isolation, it is not suitable to use it for acquisition of chest lead ECG, too ensure patient safety. Some modifications still need to be incorporated for performance enhancement of the circuit. Like, making the circuit battery powered will make the circuit indigenous and also will allow more patient safety. Also, much functionality could be added to the PC program, to perform direct ECG analysis and disease identification. REFERENCES 1. L. Schamroth, An Introduction to Electrocardiography, 7th ed. Oxford: Blackwell Science ltd; 199. 2. J. D. Bronzio, The Biomedical Engineering Handbook, IEEE Press, 2. 3. M.S.J.Steyaert, W.C.Sansen, C.Zhongyuan, A Micro-power low noise Monolithic Instrumentation Amplifier for Medical purposes, IEEE Journal of Solid state circuits, 1987, SC-22(6), pp 1163-1168 4. E.M. Spinelli, N. Martínez, M. A. Mayosky, R.P. Areny, Novel Fully Differential Biopotential Amplifier With DC Suppression, IEEE Transactions on BME, Vol. 51, no. 8, 24 5. M.J. Burke, D.T. Gleeson, A micropower dry electrode ECG Pre-amplifier IEEE Transaction on Biomedical Engineering, 2, vol.- 47 (2) 155-162 Copyright to IJIRSET www.ijirset.com 155

6. C.J. Yen, W.Y. Chung, M.C. Chi, Micro-power low offset Instrumentation Amplifier IC design for biomedical system applications, IEEE Transactions on Circuits and Systems, 24, vol- 51(4), pp. 691-699. 7. M. Chaitanya Suman and K. Prathyusha, Wireless ECG System Based on ARM LPC 213 Processor, IJECT Vol. 3, Issue 1, Jan.-March 212 8. E.S. Valchinov and N.E. Pallikarakis, Wearable Wireless Biopotential Electrode for ECG Monitoring, IFMBE Proceedings 16, pp. 373 376, 27. 9. J. A Van Alste,,, T. S. Schilder, Removal of Baseline Wander and Power-Line Interference from the ECG by an Efficient FIR Filter with a Reduced Number of Taps, IEEE Transactions on Biomedical Engineering, BME-32, issue 12, 1985, pp. 152-16. 1. Y. Sun, K.L. Chan, and S.M. Krishnan, ECG signal conditioning by morphological filtering. Computers in biology and medicine, 32(6): 465-479, 2. BIOGRAPHY Deboleena Sadhukhan is currently pusuing her Ph.D. from the Department of Applied Physics, University of Calcutta. After graduating with Physics Hons., she did her B.Tech and M.tech in Instrumentation Engineering from the Department of Applied Physics, University of Calcutta and stood First class First from the university in both. She also has a one year teaching experience in a reputed engineering college and also served as Guest Lecturer in the Department of Applied Physics, University of Calcutta. Her research interest is Biomedical signal processing. [1] She has currently been awarded the DST-INSPIRE fellowship. Rohit Mitra is pursuing M.Tech in Instrumentation and Control Engineering in the Department of Applied Physics, Calcutta University and have appeared for the 5th Semester examination. He have passed B.Tech under West Bengal University of Technology (WBUT) in the year 29 in Electronics and Instrumentation Engineering. Avik Kundu Mitra is pursuing M.Tech in Instrumentation and Control Engineering in the Department of Applied Physics, Calcutta University and have appeared for the 5th Semester examination. He completed B. Tech in Electronics and Instrumentation Engineering from Future Institute of Engineering and Management in 211. Copyright to IJIRSET www.ijirset.com 156