Calibration of 100 MΩ Hamon resistor using current-sensing Wheatstone bridge. Ivan Leniček 1, Roman Malarić 2, Alan Šala 3

Similar documents
Application of Digital Sampling Method for Voltage Transformer Test Set Calibrations. Hüseyin Çaycı

Realization and traceability of AC power standard at frequency of 50 Hz

Low Cost, General Purpose High Speed JFET Amplifier AD825

Investigation on the realization of an automated and guarded Hamon GΩ network

TRANSDUCER INTERFACE APPLICATIONS

Resistance Measurements Systems w/sub PPM Accuracy - 1uΩ to 1GΩ. Duane Brown, Measurements International

Model 6600A Dual Source High Resistance Bridge

Modular and Upgradable Fully Automated High Resistance Measurement Systems

8558A 8 1/2 Digit Multimeter

Low Cost Instrumentation Amplifier AD622

Standard Resistors Precision DC Shunts Digital Thermometers Non-inductive AC Shunts Intelligent Teraohmmeter Precision Decade Standards Precision Air

8000 SERIES PRECISION MULTIMETER VERIFICATION AND ADJUSTMENT GUIDE

UNIVERSITI MALAYSIA PERLIS

UNIT- IV ELECTRONICS

EXAM Amplifiers and Instrumentation (EE1C31)

High Current, High Power OPERATIONAL AMPLIFIER

Comparison of RMS Value Measurement Algorithms of Non-coherent Sampled Signals

Proposal for instrumentation to calibrate DCCT s up to 24 ka

CA-550 Series / CA-650 Series

HP 34401A Specifications 8

Quad Picoampere Input Current Bipolar Op Amp AD704

Chapter 9: Operational Amplifiers

EE431 Lab 1 Operational Amplifiers

Precision in Practice Achieving the best results with precision Digital Multimeter measurements

A New Generation of Hamon Resistor Standards

5075 Precision Digital Multimeter Time Electronics Calibration, Test & Measurement

Infrared Communications Lab

Publication II by authors

Ultrastable Low-Noise Current Amplifiers With Extended Range and Improved Accuracy

2.996/6.971 Biomedical Devices Design Laboratory Lecture 7: OpAmps

CONDUCTIVITY sensors are required in many application

Variable-Gain High Speed Current Amplifier

Lecture 14 Interface Electronics (Part 2) ECE 5900/6900 Fundamentals of Sensor Design

Calibration Techniques for the Home Lab

Model 6517B Electrometer / High Resistance Meter Specifications

Investigation of Two Different Techniques for Accurate Measurements of Sinusoidal Signals

Introduction to Analog Interfacing. ECE/CS 5780/6780: Embedded System Design. Various Op Amps. Ideal Op Amps

THERMAL NOISE. Advanced Laboratory, Physics 407, University of Wisconsin. Madison, Wisconsin 53706

ADVANCED THERAPY DOSIMETER MODEL 35040

DUAL ULTRA MICROPOWER RAIL-TO-RAIL CMOS OPERATIONAL AMPLIFIER

Variable Gain Sub Femto Ampere Current Amplifier

8588A Reference Multimeter

Chapter 9: Operational Amplifiers

Differential Amplifier : input. resistance. Differential amplifiers are widely used in engineering instrumentation

AC Resistance Thermometry Bridges and their Advantages By Peter Andrews


Experiment 1A. Integrated Laboratory Experiment CHEMICAL INSTRUMENTATION PART A. ELECTRONIC INSTRUMENTATION

CERTIFICATE OF ACCREDITATION

INTEGRATED CIRCUITS. AN109 Microprocessor-compatible DACs Dec

Lab 2: Optical Theremin Team 2 Flyback By Brian Pugh, Andrew Baker, and Michael Betts

Performance Assessment of Resistance Ratio Bridges used for the Calibration of SPRTs

Precision amplifier for bridge circuits AM467 PRINCIPLE FUNCTION

QUAD 5V RAIL-TO-RAIL PRECISION OPERATIONAL AMPLIFIER

EE LINEAR INTEGRATED CIRCUITS & APPLICATIONS

LOW NOISE AMPLIFIER SA SERIES

Single Supply, Rail to Rail Low Power FET-Input Op Amp AD820

Testing & Calibration Lab, 204, Diamond Industrial Estate No. 2, Ketki Pada Road, (Near Dahisar Toll Naka), Dahisar (East), Mumbai, Maharashtra

Module 2. Measurement Systems. Version 2 EE IIT, Kharagpur 1

APPLICATION NOTE. Wide Range of Resistance Measurement Solutions from μω to PΩ

Maximizing your reference multimeter, minimizing measurement uncertainties

IC Preamplifier Challenges Choppers on Drift

Dual, Current Feedback Low Power Op Amp AD812

R-X SERIES. Decade Resistor

Linear IC s and applications

Quad Picoampere Input Current Bipolar Op Amp AD704

Comparison of the NIST and NRC Josephson Voltage Standards (SIM.EM.BIPM-K10.b)

Fluke 83V and 87V Digital Multimeters

Discover. Blue Box. the. Difference. High Resistance Metrology Products Guide

True RMS Bench Multimeter 5492B

CHAPTER 6 DIGITAL INSTRUMENTS

A New Method for the Calibration of the mv Ranges of an AC Measurement Standard

Making Basic Strain Measurements

Use of the BVD for traceability of bipolar DC voltage scale from 1 mv up to 1200 V

ULTRA HIGH VOLTAGE DUAL OPERATIONAL AMPLIFIER

Micropower, Single-Supply, Rail-to-Rail, Precision Instrumentation Amplifiers MAX4194 MAX4197

MIC915. Features. General Description. Applications. Ordering Information. Pin Configuration. Pin Description. Dual 135MHz Low-Power Op Amp

Voltage-to-Frequency and Frequency-to-Voltage Converter ADVFC32

Fluke 8845A/8846A Digital Multimeters Extended Specifications

Op-Amp Specifications

Calibration Laboratory Assessment Service CLAS Certificate Number Page 1 of 10

Section3 Chapter 2: Operational Amplifiers

Model 4402B. Ultra-Pure Sinewave Oscillator 1Hz to 110kHz Typical Distortion of % Serial No. Operating Manual

CA330 RTD Calibrator: High-speed Response and High-resolution Resistance Simulator

EPAD OPERATIONAL AMPLIFIER

True RMS Bench Multimeter 5492B

An improvement for dual channel sampling wattmeter

High Speed FET-Input INSTRUMENTATION AMPLIFIER

An Improved Bandgap Reference (BGR) Circuit with Constant Voltage and Current Outputs

Operational Amplifiers

5790A Automated AC Measurement Standard

Dual Picoampere Input Current Bipolar Op Amp AD706

CERTIFICATE OF ACCREDITATION

MODEL INFORMATION. MODEL AccuBridge. Self Calibration Ratio Bridge

PHYS 536 The Golden Rules of Op Amps. Characteristics of an Ideal Op Amp

LINEAR IC APPLICATIONS

CERTIFICATE OF ACCREDITATION

5500A. Multi-Product Calibrator. Extended Specifications 2005

Calibration Specialists Ltd

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Fluke 5820A Oscilloscope Calibrator Specifications

Transcription:

Calibration of 100 MΩ Hamon resistor using current-sensing Wheatstone bridge Ivan Leniček 1, Roman Malarić 2, Alan Šala 3 1 Faculty of electrical engineering and computing, Unska 3, 10000 Zagreb, Croatia, phone:+385-1-6129-709, fax: +385-1-6129-616, e-mail: ivan.lenicek@fer.hr 2 Faculty of electrical engineering and computing, Unska 3, 10000 Zagreb, Croatia, phone:+385-1-6129-650, fax: +385-1-6129-616, e-mail: roman.malaric@fer.hr 3 Faculty of electrical engineering and computing, Unska 3, 10000 Zagreb, Croatia, phone:+385-1-6129-749, fax: +385-1-6129-616, e-mail: alan.sala@fer.hr Abstract - Primary electromagnetic laboratory in Zagreb maintains standards of four major electromagnetic units: volt, farad, ohm and second. Reference standards are periodically calibrated in PTB (Germany) and several methods has been realized for internal laboratory traceability check. RC-comparison method uses Hamon divider constructed of 100 equally balanced 1-megaohm resistors that connected in parallel or serially exhibit Hamon ratio of 100 MΩ : 10 kω with relative uncertainty of 10-8 [1]. Hamon resistance of 100 MΩ is directly compared with impedance of 100 pf capacitance reference at frequency of 15,873 Hz. Measurement of low-frequency voltage ratio is based on two digital voltmeters HP 3458A with improved algorithm for data processing [2]. Accuracy of reference standards comparison 10 kω 100 pf depends on consistency of Hamon divider that can be disturbed by aging of used metal-film components. In this paper a recently realized method for accurate measurement of 100 MΩ Hamon resistor by means of comparison with calibrated and stable 10 MΩ resistance standard at higher voltage levels has been presented. I. Introduction Method of RC-comparison begins with determination of 1:1 ratio between Hamon divider connected in parallel and 10 kω reference standard that has been previosly calibrated at PTB. This comparison is based on voltage ratio measurement using two digital voltmeters (DVs) positioned across each of 10 kω resistor and one voltage source of 20 V, hence each resistor has voltage drop of 10 V. Making two voltage measurements by interchanging voltemeter's positions, instability of voltage source and systematic errors of the DVs have no influence on resistance ratio uncertainty. The resistors of Hamon divider are balanced within 10-4, therefore the error of the 1:100 2 Hamon ratio should not exceed 10-8. However, aging of metal-film resistors during 12 years of Hamon resistor exploatation could significantly disturb Hamon ratio. This possibility launched development of modified Wheatstone bridge which can be used for comparison of Hamon divider, now in 100 MΩ mode, with another reference standard, such as 10 MΩ. II. DVM-based system for high-ohm resistance comparison Currently used method for high-ohm resistance comparison by means of two digital voltmeters and electrometer interface (E) is shown in Figure 1. Influence of input resistance instability of digital voltmeters has been lowered with much higher input resistance of electrometer interface [3,4]. Therefore voltage measurement on two resistors' positions gives exact ratio R 1 /R 2 without influence of DV2's input resistance paralleling effect. Entire measuring system is carefully shielded and uncertainty of ratio is practically caused only by nonlinearity of DV2 on 10 V measuring range, what can be find out separately. The best results are obtained for 1:1 resistance ratio when uncertainty of the method is only influenced by short-term instability of DV2's readings. The estimated uncertainty for the ratio 100 MΩ : 10 MΩ is less than 0,8 10-6 (coverage factor k = 1). Limitation of this method lies in maximal permitted input voltage of the electrometer interface of 10 V, what is the reason why comparison that incorporates interchanging positions of the resistors couldn t be done at 100 V. Since RC-comparison is performing on measuring voltage of 100 V (calibrating voltage for capacitance standard), there is possibility of influence of metal-film voltage coefficient on Hamon divider accuracy. Therefore a new method for resistance comparison at higher voltage levels has been investigated.

Figure 1: Measuring system for high-ohm resistance comparison uses two digital voltmeters and electrometer interface. III. Modified Wheatstone bridge-based system for high-ohm resistance comparison Modified Wheatstone bridge shown in Figure 2. consists of two voltage sources and current-sensing detector. The use of effective low-impedance current-sensing detector instead of high-impedance electrometer null-detector insures that the only one sensitive point of bridge (A) is continuously held at potential near ground (virtual zero), even when the bridge is unbalanced, thus eliminating parasitic currents paths from point "A" towards ground. Figure 2. Modified Wheatstone bridge realized with current-sensing detector as an indicator of bridge balance. A. Realization of current-sensing detector The current-sensing detector used in this method is transimpedance amplifier that converts current difference I = I 1 I 2 into voltage measured by sensitive null-detector (N), such as analog null-detector FLUKE 745AB or programmable digital voltmeter HP 3458A. The principle of operation is shown in Figure 3. The non-inverting input of operational amplifier (OPamp) is connected to the ground, keeping as well the inverting input at potential of "virtual zero". The difference of currents I 1 and I 2 that flow through compared resistors R 1 and R 2 respectively is converted to output voltage by factor R F V/A. The feedback resistor R F defines sensitivity of the detector, depending on values of compared resistors, applied voltages and Johnson noise of R F itself. The operational amplifier used in this design has been carefully chosen between commercially available OPamps having ultra-low input offset current and low input offset voltage with low temperature coefficient. This parameters has been measured on few OPamps of AD 549K type (Analog Devices) and the chosen one exhibited approximately 30 fa of input offset current and less than 10 µv of input offset voltage. The feedback resistor used for 100 MΩ : 10 MΩ comparison is epoxy encapsulated cermet type 1 GΩ (Meggit HB). Since the measured voltage I R F is considered to be an error of resistors ratio, the true value of R F, as well as its temperature stability, are not of great significance. The capacitor C F is added as a bypass capacitor for reduction of 50 Hz noise. With value of 1 nf, time constant of the detector is 1 second, therefore 50 Hz noise is reduced more than two orderds of quantity. This simple detector design also has a few components added for automation of measuring process. One is relay Re1 that bypasses feedback resistor during voltage sources activation.

After the voltage sources were set to approximate value needed for bridge balance and enabled by computer, the relay Re1 is opened and the voltage on detector output is being measured. On that way a current shock through R F caused by nonsynchronous voltage applying is avoided, decreasing also the setting time of the detector. Second are relays Re2 and Re3 that are used for shorting resistors R 1 and R 2 in step of "zero of the bridge" determination. After the balance of the bridge was adjusted, relays Re4 and Re5 are interchanging positions of the voltmeters DV1 and DV2 (digital voltmeters HP 3458A) during voltage ratio measurement in order to eliminate their systematic errors. Figure 3. Functional schematic of modified Wheatstone bridge for high-ohm resistance comparison. B. Principle of measuring procedure After the balance of the bridge was achieved, the ratio of compared resistors is determined by means of voltage ratio using simple following equation: R R 1 2 U1a U 2b =, (1) U U where indexes "1" and "2" designate the voltmeters DV1 and DV2, and indexes "a" and "b" their positions. Procedure of balancing modified Wheatstone bridge has two steps. For the period of first step resistors R 1 and R 2 are connected to the ground and output voltage U 0 of the detector is measured with DV3. This voltage is a consequence of input offset voltage of detector s amplifier and its input offset current flowing through the feedback resistor: U 2a 1b = + R. (2) F 0 U of 1 RF I of R1 R 2 Assuming that offset parameters of OPamp are constant, U 0 therefore represents "zero" that has to be established by bridge balancing. In second measuring step, when voltage sources are enabled, PC establishes feedback loop of bridge balancing with programmable voltage calibrator FLUKE 5700A controlled over IEEE-488 bus. The bridge is considered "in balance" when readings of DV3 are not showing deviation from U 0 more than previously set margins of error. Since 100 MΩ Hamon resistor has to be measured at voltage of 100 V, that will produce current flow in upper branch of bridge of 1 µa. The balance of bridge will be achieved with 10 V across reference standard of FLUKE 742A - 10 MΩ in lower branch, and for that purpose FLUKE 731B voltage standard is used. If relative current difference of 10-6 (1 pa) is sensed, the detector will provide output voltage of 10 9 V/A 1pA = 1 mv. IV. Results The measurement setup for the first step of balancing procedure includes 10 subsequent readings with DV3 (HP 3458A) to gain needed "zero" of the bridge. It was found to be U 0 = 0,51 mv with standard deviation of s U0 = ±70 µv. This "zero" measurement was repeated several times during two weeks and no significant changes were observed. Prior to the start of the balancing procedure the margins of bridge balance were set on U m = ±0,1 mv around measured U 0. In second measuring step, after initial voltages had been enabled, the mean value of 10 readings of DV3 was taken as a measure

of bridge unbalance. That "error" voltage has been used for correction of calibrator voltage via PC and balancing procedure was repeated. The balance was found for such voltage of calibrator that puts mean value and associated standard deviation of 10 subsequent readings of detector error voltage within calculated margins (U 0 ± U m ). That balancing algorithm is a result of number of bridge testing on sensitivity, influence of parasitic impedances and balancing speed, and is found to be satisfactory with regard to reliable zero finding. Table 1. shows results of 100 MΩ : 10 MΩ comparisons made in the same laboratory conditions within twelve days. Table 1: Results of comparison of 100 MΩ Hamon divider with 10 MΩ reference resistor. Day Number of Standard Value of 100 MΩ Measured ratio measurements deviation / 10-6 Hamon resistor / MΩ 2004-02-16 10 10,1074430 0,19 101,078928 2004-02-19 10 10,1074434 0,15 101,078932 2004-02-22 10 10,1074449 0,11 101,078947 2004-02-25 10 10,1074459 0,13 101,078957 2004-02-28 10 10,1074477 0,13 101,078975 Reference resistance: R 1 = 10,000445 MΩ 101,07900 R 1 / MΩ 101,07898 101,07896 +0,039 µω/ω per day 101,07894 101,07892 101,07890 16 18 20 22 24 26 28 Days of February 2004 Figure 4: Graphic analysis of 100 MΩ : 10 MΩ comparisons during 12 days of measurement. Error bars include only uncertainty of estimated resistance ratio. Assuming that 10 MΩ reference resistor is stable over measuring time, the calculated values for Hamon resistor given in Figure 4. are showing positive time drift, what is an expected behaviour thanks to series of RC-comparisons done in 1994 [1] and 1999. It must be emphasized that the exact value of Hamon resistor also depends on accuracy of 10 MΩ standard. Its value can be established with uncertainty of ±0,2 10-6 [4] using chain resistance comparisons down to 10 kω primary resistance standard. V. Uncertainty calculation for ratio measurement The uncertainty budget for the realized modified Wheatstone bridge includes several components described in Table 2. The combined relative standard uncertainty (k = 1) for 100 MΩ : 10 MΩ ratio was calculated u = 0,19 10-6. The first line in Table 2. is contribution due to mean standard deviation of five measuring results stated in Table 1. Other contributions are consequences of limitations of used components and measuring instruments.

Table 2: Uncertainty budget for 100 MΩ : 10 MΩ comparison on 100 V using realized currentsensing detector and modified Wheatstone bridge (k = 1). Quantity Relative standard Probability Uncertainty uncertainty / 10-6 distribution contribution / 10-6 Deviation of measuring results (type A) 0,14 normal 0,14 Nonlinearity of voltmeters on 100 V range 0,1 rectangular 0,06 Margins of bridge balance 0,1 rectangular 0,06 Stability of "zero" value including max. temperature coefficient of OPamp 0,08 normal 0,08 offset voltage Noise of feedback resistor < 0,01 normal 0,01 Resolution of voltage calibrator 0,1 rectangular 0,06 VI. Conclusion The described modified Wheatstone bridge with current-sensing detector is capable of measuring resistance ratios at voltage levels up to 1000 V. The sensitivity of detector can be simply altered by changing the value of feedback resistor. First measurements described in this paper confirm excellent repeatability of 100 MΩ : 10 MΩ ratio at 100 V with extended uncertainty (k = 2) better than 5 10-7. This is very good result having in mind simplicity and inexpensiveness of current-sensing detector realization. References [1] V. Bego, J. Butorac, D. Hrsan, Calibration of the standard resistors by means of the standard capacitors and precise digital voltmeters, CPEM Conference Digest, pp. 392-393, Braunschweig, Germany, 1996. [2] N. Hlupić, A sampling method for high accuracy AC measurement, Proceedings of the 12 th IMEKO TC4 International Symposium, part I, pp. 55-59, Zagreb, Croatia, 2002. [3] I. Leniček, Influence of DVM's Input Parameters on DVM-based High-ohm Resistance Comparison, Proceedings of the 12 th IMEKO TC4 International Symposium, Part I, pp. 91-95, Zagreb, Croatia, 2002. [4] I. Leniček, R. Malarić, A. Šala, An accurate comparison of high-ohm resistances using improved DVM-based system, Proceedings of the 17 th IMEKO World Congress, pp. 859-864, Cavtat, Croatia, 2003. [5] L.C.A. Henderson, A new technique for the automated measurement of high valued resistors, J. Phys., Electron., Sci. Instrum., vol. 20, pp. 492-495, Sept. 1987.