Nature Protocols: doi: /nprot Supplementary Figure 1. Schematic diagram of Kőhler illumination.

Similar documents
EPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY

Supplementary Information. Stochastic Optical Reconstruction Microscopy Imaging of Microtubule Arrays in Intact Arabidopsis thaliana Seedling Roots

INTRODUCTION TO OPTICAL MICROSCOPY

Zeiss AxioImager.Z2 Brightfield Protocol

Zeiss LSM 880 Protocol

Imaging Introduction. September 24, 2010

Very short introduction to light microscopy and digital imaging

CCAM Microscope Objectives

Figure 3.4 Approximate size of various types of cells. ~10 um. Red Blood Cells = mm 1500 um. Width of penny Pearson Education, Inc.

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide

Nikon E800 Operating Instructions.

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal

Katarina Logg, Kristofer Bodvard, Mikael Käll. Dept. of Applied Physics. 12 September Optical Microscopy. Supervisor s signature:...

Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE

Instructions for Making On-Line Reservations for Microscopes in NB11-204

3. are adherent cells (ie. cells in suspension are too far away from the coverslip)

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

Education in Microscopy and Digital Imaging

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

Bio 407. Applied microscopy. Introduction into light microscopy. José María Mateos. Center for Microscopy and Image Analysis

Motorized Axio Observer Start-up instructions

Axio Zoom.V16 The Fluorescence Zoom Microscope for Large Fields

FLUORESCENCE MICROSCOPY. Matyas Molnar and Dirk Pacholsky

Optics Day 3 Kohler Illumination (Philbert Tsai July 2004) Goal : To build an bright-field microscope with a Kohler illumination pathway

Nikon E800 Microscope. Operating Instructions

Quick Operation Guide

2/4/15. Brightfield Microscopy! It s all about Magnification..! or is it?!

a) How big will that physical image of the cells be your camera sensor?

Microscopy Training & Overview

BASICS IN BIOIMAGING AND OPTICS PLATFORM EPFL SV PTBIOP LIGHT MICROSCOPY

Systems Biology. Optical Train, Köhler Illumination

MICROSCOPY MICROSCOPE TERMINOLOGY

Point Spread Function Estimation Tool, Alpha Version. A Plugin for ImageJ

Microscopic Structures

Variable microinspection system. system125

Zeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09)

Microscope anatomy, image formation and resolution

Low Voltage Electron Microscope

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands

Nikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON

Zeiss Axioskop II. The AIF's "routine" light microscope. (Installed 8/24/04)AxioCam installed July 11th 2005

Zeiss Axio Imager.A1 manual

Light microscopy. Part II

Zeiss LSM 780 Protocol

Widefield 1. Switching on

contents TABLE OF The SECOM platform Applications - sections Applications - whole cells Features Integrated workflow Automated overlay

Point Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy

Basics of Light Microscopy and Metallography

Microscopy: Fundamental Principles and Practical Approaches

Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement

Figure 1. Oil-immersion objectives available for use with the Lionheart FX.

Chapter 3 Op,cal Instrumenta,on

Light Microscopy. Upon completion of this lecture, the student should be able to:

Leica DMI 4000 tutorial

Tissue Preparation ORGANISM IMAGE TISSUE PREPARATION. 1) Fixation: halts cell metabolism, preserves cell/tissue structure

QUICKSTART GUIDE: WIDEFIELD HWF1 Zeiss Cell Observer Live Cell Imaging System (HAMMERSMITH, L BLOCK, ROOM 314) Imperial College London

ANSWER KEY Lab 2 (IGB): Bright Field and Fluorescence Optical Microscopy and Sectioning

The light microscope

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background

Introduction to Light Microscopy. (Image: T. Wittman, Scripps)

QUICKSTART GUIDE: WIDEFIELD WF3 Zeiss Cell Observer Live Cell Imaging System (SAF, ROOM 409) Imperial College London

SETTING UP OF A TOTAL INTERNAL REFLECTION FLUORESCENT MICROSCOPE (TIRFM) SYSTEM: A DETAILED OVERVIEW

Final Exam, 150 points PMB 185: Techniques in Light Microscopy

Ocular Lenses. Head. Arm. Objective Lenses. Slide Holder Stage. On / Off Switch. Condenser with Iris Diaphragm. Light Intensity Control

Bias errors in PIV: the pixel locking effect revisited.

VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands

Reichert Univar Manual

Invitation for a walk through microscopy. Sebastian Schuchmann Jörg Rösner

Practical Light Microscopy

Single-photon excitation of morphology dependent resonance

Low Voltage Electron Microscope

LVEM 25. Low Voltage Electron Mictoscope. fast compact powerful

Confocal Microscopy and Related Techniques

ECEN 4606, UNDERGRADUATE OPTICS LAB

Please do not hesitate to contact us if you have any questions or issues during installation or operation

Chapter 3 Op+cal Instrumenta+on

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

EXC500p-- PATHOLOGY MICROSCOPE. EXC500hd -- HD DIGITAL PATHOLOGY MICROSCOPE. EXC500r -- RESEARCH MICROSCOPE EXC500-LABORATORY SCOPE

Observing Microorganisms through a Microscope LIGHT MICROSCOPY: This type of microscope uses visible light to observe specimens. Compound Light Micros

Rates of excitation, emission, ISC

Microscope. Dr. Leena Barhate Department of Microbiology M.J.College, Jalgaon

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy,

MICROSCOPE LAB. Resolving Power How well specimen detail is preserved during the magnifying process.

Introduction: Why electrons?

microscopy A great online resource Molecular Expressions, a Microscope Primer Partha Roy

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Nikon E800 Operating Instructions.

CCAM s Selection of. Zeiss Microscope Objectives

Diskovery Spinning Disk Guide

Nature Structural & Molecular Biology: doi: /nsmb Supplementary Figure 1

Prof. Enrico Gratton - Lecture 6 Fluorescence Microscopy

Measuring incidence angle for throughthe-objective

User manual for Olympus SD-OSR spinning disk confocal microscope

Leica TCS SP8 Quick Start Guide

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005

Light microscopy BMB 173, Lecture 14, Feb. 21, 2018

Transcription:

Supplementary Figure 1 Schematic diagram of Kőhler illumination. The green beam path represents the excitation path and the red represents the emission path.

Supplementary Figure 2 Microscope base components and assembly process. Numbers correspond to assembly flow as indicated in Steps 1 3.

Supplementary Figure 3 Camera mount components and assembly process. Numbers correspond to assembly flow as indicated in Steps 4 7.

Supplementary Figure 4 Spirit level alignment. Example of using (a) bullseye spirit level [Step 8] and (b) tubular spirit level [Step 11] to align components.

Supplementary Figure 5 Tube lens mount components and assembly process. Numbers correspond to assembly flow as indicated in Step 9A.

Supplementary Figure 6 Dichroic cube mount components and assembly process for epifluorescence imaging feature. Numbers correspond to assembly flow as indicated in Step 9B.

Supplementary Figure 7 Motorized microscope objective stage components and assembly process. Numbers correspond to assembly flow as indicated in Steps 10 16.

Supplementary Figure 8 Microscope objective mounted on translation stage adapter. The base of the objective lens (above the treaded section) should be resting on the tube lens mount. Once the objective lens is properly secured to the stage mount, shift the lens mount up and down to ensure that the lens threading does not rub against the mount.

Supplementary Figure 9 Microscope specimen stage components and assembly process. Numbers correspond to assembly flow as indicated in Steps 17 26.

Supplementary Figure 10 Fabrication steps for microscope stage. (a) Two pairs of 1 x 3 glass slides are bonded to each other by applying superglue on the areas shown by blue arrows. Apply enough superglue so that the entire area of the base is covered. (b) Bond two additional glass slides on top and bottom of the two slide pairs and (c) wait until the glue has cured. (d) Mount the completed glass stage on the XY translation adapter, and tightly position it by sliding and locking the slider. The glass stage can then be secured by tightening four screws on the top of the XY translation adapter.

Supplementary Figure 11 Transillumination source components and assembly process. Numbers correspond to assembly flow as indicated in Step 27A.

Supplementary Figure 12 Epifluorescence illumination module components and assembly process. Numbers correspond to assembly flow as indicated in Step 27B.

Supplementary Figure 13 Schematic diagram showing lens and diaphragm location for Kőhler illumination. Beam path showing the approximate location of lenses and iris diaphragms, where f is the focal length and f b is the back focal length of the lens. The actual positions may vary during the setup; hence calibration must be performed to achieve good Kőhler illumination. For example, the first iris diaphragm (condenser diaphragm) must be located where the LED source is clearly visible after the condenser lens.

Supplementary Figure 14 Calibration of microscope magnifying power. To achieve the desired magnifying power, (a) acquire an image of a calibration grid and measure the grid length (in pixels) with ImageJ. The effective magnification power is given by M LLM =detector pixel size [μm/px] /(grid size [μm] /measured number of pixels [px]). (b) If needed, adjust camera position. Repeat steps if necessary.

Supplementary Figure 15 Imaging area comparison with and without epifluorescence illumination add-on. (a) Micrographs of 100 μm spaced target grid captured with and epi-fluorescence illumination module. Significant vignetting can be observed. (b) In comparison, the imaging area is larger if the epi-fluorescence module is omitted. Scale bar, 100 μm.

Supplementary Figure 16 Comparison between analog and digital scintillation images. Analog scintillation images captured with (a) no binning, 1200 EM gain, and 1 min exposure time and (b) 4 binning, ~700 EM gain, and 2 min exposure time. (c) Digitial scintillation image captured with 4 binning, 1200 EM gain, 30 ms exposure time per frame, and 20,000 frames. Hot spots (red arrows) are caused by stray ionizing radiation directly hitting the EMCCD detector, frequently appears in analog images, whereas such signals are rejected in digitally reconstructed images. Scale bar, 200 μm.

Supplementary Figure 17 Multimodal micrographs of confluent MDA-MB-231 cells. Brightfield, fluorescence, bioluminescence, and overlay images from MDA-MB-231/Luc cells. Top row. Nikon 20x objective lens (effective magnification of 4x). Scale bar, 200 μm. Bottom row. Olympus 40x oil-immersion objective lens (effective magnification of ~11x). Scale bar, 100 μm.

Supplementary Figure 18 Example of bioluminescence imaging captured under short exposure times for real-time imaging. (a) Brightfield image of MDA-MB-231/Luc cells and its corresponding bioluminescence images captured with different imaging parameters : (b) no binning, ~1060 EM gain, and 20 s exposure time; (c) no binning, 1200 EM gain, and 3 s exposure time; (d) 2 binning, 1200 EM gain, and 0.75 s exposure time. Scale bar, 200 μm.

Supplementary Table 1, Theoretical numerical aperture (NA) and brightness (B) for different microscope objectives. Make Model M LLM Eq.(2) NA LLM (NA orig ) 1 B LLM,trans B LLM,epi Nikon Olympus Zeiss CFI Plan Apo Lambda 2 0.5 Yes 0.1 400.0 4 CFI LU Plan Fluor EPI P 5 1.25 Yes 0.15 144.0 3.24 CFI S Fluor 10 2.5 No 0.45 (0.5) 330.6 68.3 CFI Plan Apo Lambda 20 5 No 0.71 (0.75) 204.1 104.1 CFI S Plan Fluor ELWD ADM 40 10 Yes 0.6 36 13.0 CFI Plan Fluor (oil) 40 10 Yes 1.3 169 285.6 CFI Plan Apo Lambda (oil) 60 15 Yes 1.4 87.1 170.7 CFI Plan Apo Lambda 100 25 Yes 1.45 33.6 70.7 MPLN 5 1.4 Yes 0.1 51.8 0.5 UPLSAPO 10 2.8 Yes 0.4 207.4 33.2 UPLSAPO 20 5.6 Yes 0.75 182.3 102.5 UPLSAPO 40 11.1 Yes 0.9 65.6 53.1 UPLFLN 40 11.1 Yes 1.3 136.9 231.3 UPLSAPO 60 16.7 Yes 1.35 65.6 119.6 UPLSAPO 100 27.8 Yes 1.4 25.4 49.8 C Epiplan-Apochromat 5 1.5 Yes 0.2 172.2 7.0 Epiplan-Apochromat 10 3.0 Yes 0.4 172.2 27.9 Plan-Apochromat 20 6.1 No 0.78 (0.8) 164.7 99.6 Plan-Apochromat 40 12.1 No 0.93 (0.95) 58.5 50.3 Plan-Apochromat 40 12.1 Yes 1.4 133.4 261.5 Plan-Apochromat 63 19.1 Yes 1.4 53.8 105.4 Alpha Plan-Apochromat 100 30.3 Yes 1.46 23.2 49.5 Table comparing theoretical performance of different microscope objective manufacturers. The theoretical transillumination brightness of standard microscopes ranges between 4-16 and 16-25 for 5 and 10 magnification, respectively. References 1. Kim, T. J., Tuerkcan, S., Ceballos, A. & Pratx, G. Modular platform for low-light microscopy. Biomed. Opt. Express 6, 4585-4598 (2015).

Supplementary Note 1, Basics on radioactivity. In a radioluminescence microscopy (RLM) experiment, the rate of radioactive decay must be considered because the amount of radioactivity cannot be regarded as constant over the duration of the experiment. The radioactivity A by definition is, A 1/2 dn dt ln 2, t N, (1a) where N is the time-decaying number of radioactive atoms, λ is the decay rate constant, and t 1/2 is the radioactive half-life. The user can refer to a table of nuclides for information about individual radionuclides (e.g. http://www.nndc.bnl.gov/chart/). Solving differential equation (1) yields, (1b) T N N0e, (2) where N 0 is the number of radioactive atoms at the beginning of the radioluminescence exposure and T is the total duration of the exposure. To quantify the molecular uptake of individual cells using the radioluminescence microscopy, two parameters should be considered, sensitivity and radioactive yield. The sensitivity accounts for the probability that a radioactive emission be detected by the microscope. It is a function of the isotope, scintillator, and light collection efficiency by the microscope. For quantitative estimation of radionuclide concentration, calibration must be performed, as explained in Steps 71-79. Also, the radioactive yield must be taken into consideration since the radioluminescence microscope only detects particulate radiation (α, β - or β + ). This is due to the fact that the main detectable signals will originate from particles that undergo collision cascades in the scintillator rather than from photons (γ-rays). Other modes of radioactive decay such as electron capture and internal transition cannot be detected. Accounting for the aforementioned parameters, the number of measured events D is related to the number of actual radioactive decays N B, D NB N0 N, (3) SY where S is the sensitivity calibration factor and Y is the radioactive yield for particulate radiation. Solving for equation (3), the number of molecules initially present per pixel is, N 0 D S Y 1 1 e T. (4) It should be noted that equation (4) can also be expressed in terms of activity ( A N).