High Speed Parametric Test Using Agilent 4070 Series

Similar documents
Measuring CNT FETs and CNT SETs Using the Agilent B1500A

Achieving Maximum Throughput with Keithley S530 Parametric Test Systems

Keysight Technologies Accurate Capacitance Characterization at the Wafer Level

SOURCE MEASURE UNITS. Make Multiple Measurements Accurately Using a Single Instrument All While Saving Space, Time and Money

Techniques to Achieve Oscilloscope Bandwidths of Greater Than 16 GHz

KickStart Instrument Control Software Datasheet

MEMS On-wafer Evaluation in Mass Production Testing At the Earliest Stage is the Key to Lowering Costs

Keysight Technologies Improve the Accuracy and Efficiency for Organic-Thin Film Transistor (Organic-TFT) Characterization

Simplifying FET Testing with 2600B System SourceMeter SMU Instruments APPLICATION NOTE

Keysight Technologies Accurate NBTI Characterization Using Timing-on-the-fly Sampling Mode. Application Note

Using the Ramp Rate Method for Making Quasistatic C-V Measurements with the 4200A-SCS Parameter Analyzer APPLICATION NOTE

Keysight Technologies Making Field Effect Transistor Characterization Using SMU

Keysight Technologies MEMS On-wafer Evaluation in Mass Production

Simplifying DC-DC Converter Characterization using a 2600B System SourceMeter SMU Instrument and MSO/DPO5000 or DPO7000 Series Scope APPLICATION NOTE

Measuring Vgs on Wide Bandgap Semiconductors APPLICATION NOTE

Keysight Technologies Direct Power MOSFET Capacitance Measurement at 3000 V

S540 Power Semiconductor Test System Datasheet

S540 Power Semiconductor Test System Datasheet

Automated Frequency Response Measurement with AFG31000, MDO3000 and TekBench Instrument Control Software APPLICATION NOTE

Switching Between C-V and I-V Measurements Using the 4200A-CVIV Multi-Switch and 4200A-SCS Parameter Analyzer APPLICATION NOTE

Measuring Power Supply Switching Loss with an Oscilloscope

Using the 4200A-CVIV Multi-Switch to Make High Voltage and High Current C-V Measurements APPLICATION NOTE

Keysight Technologies 7 Hints That Every Engineer Should Know When Making Power Measurements with Oscilloscopes. Application Note

Keysight Technologies How to Take Fast, Simultaneous Measurements of Two or More Signals Using BenchVue Software. Application Note

7 Hints That Every Engineer Should Know When Making Power Measurements with Oscilloscopes.

Verifying Power Supply Sequencing with an 8-Channel Oscilloscope APPLICATION NOTE

Using the Model 4225-RPM Remote Amplifier/ Switch to Automate Switching Between DC I-V, C-V, and Pulsed I-V Measurements APPLICATION NOTE

Agilent U9397A/C FET Solid State Switches (SPDT)

Agilent 4070 Series Accurate Capacitance Characterization at the Wafer Level

Keysight Technologies Measuring Low Current Consumption with a Digital Multimeter

Keysight Technologies How to Perform QSCV (Quasi-Static Capacitance Voltage) Measurement

Performing Safe Operating Area Analysis on MOSFETs and Other Switching Devices with an Oscilloscope APPLICATION NOTE

Keysight Technologies Understanding the Importance of Maximum Power Point Tracking Efficiency for Solar Inverters.

Keysight Technologies

Agilent Maximizing Measurement Speed Using P-Series Power Meters

Keysight Technologies Network Analyzer Measurements: Filter and Amplifier Examples. Application Note

Keysight Technologies

Agilent U1881A and U1882A Power Measurement Application for Agilent InfiniiVision and Infiniium Oscilloscopes

Two-Way Radio Testing with Agilent U8903A Audio Analyzer

Keysight Technologies FFT and Pulsed RF Measurements with 3000T X-Series Oscilloscopes. Application Note

UWB Antenna Measurements with the 20 GHz E5071C ENA Network Analyzer

Isolation Addresses Common Sources of Differential Measurement Error

Keysight Technologies Essential Capabilities of EMI Receivers. Application Note

U1881A and U1882A Power Measurement Application for InfiniiVision and Infiniium Oscilloscopes

Making a S11 and S21 Measurement Using the Agilent N9340A

U1881A and U1882A Power Measurement Application for InfiniiVision and Infiniium Oscilloscopes

Keysight Technologies USB Preamplifiers

Essential Capabilities of EMI Receivers. Application Note

Keysight Technologies How to Read Your Power Supply s Data Sheet. Application Note

Keysight DSOXT3FRA/DSOX4FRA/DSOX6FRA Frequency Response Analyzer (FRA) Option

Keysight Technologies Resistance Measurements Using the B2900A Series of SMUs

Keysight Technologies DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement Options

Keysight Technologies Improving the Test Efficiency of MEMS Capacitive Sensors Using the E4980A Precision LCR Meter.

Agilent Correlation between TDR oscilloscope and VNA generated time domain waveform

Keysight Technologies Improving Test Efficiency of MEMS Electrostatic Actuators Using the E4980A Precision LCR Meter.

Keysight Measuring High Impedance Sources Using the U8903B Audio Analyzer. Application Note

Solutions for Solar Cell and Module Testing

A Time-Saving Method for Analyzing Signal Integrity in DDR Memory Buses

Keysight Technologies Making Current-Voltage Measurement Using SMU

EM Insights Series. Episode #1: QFN Package. Agilent EEsof EDA September 2008

Keysight Technologies P9400A/C Solid State PIN Diode Transfer Switches

Agilent N8973A, N8974A, N8975A NFA Series Noise Figure Analyzers. Data Sheet

Keysight Technologies Make Better AC RMS Measurements with Your Digital Multimeter. Application Note

Don t Let EMI/EMC Compliance Certification Slow You Down TUTORIAL

Comparing Contact Performance on PCBA using Conventional Testpads and Bead Probes

Evaluating Oscilloscopes for Low-Power Measurements

Trouble-shooting Radio Links in Unlicensed Frequency Bands TUTORIAL

Evaluating Oscilloscope Bandwidths for your Application

Agilent N8480 Series Thermocouple Power Sensors. Technical Overview

Keysight Technologies Precise Low Resistance Measurements Using the B2961A and 34420A

Keysight Technologies 1500 A and 10 kv High-Power MOSFET Characterization using the Keysight B1505A

Keysight Technologies Migrating Balanced Measurements from the

Keysight Technologies Precise Current Profile Measurements of Bluetooth Low Energy Devices using the CX3300. Application Brief

Keysight Technologies How to Measure 5 ns Rise/Fall Time on an RF Pulsed Power Amplifier Using the 8990B Peak Power Analyzer.

Keysight Technologies Photodiode Test Using the Keysight B2980A Series

Keysight Technologies 87405C 100 MHz to 18 GHz Preamplifier. Technical Overview

Keysight E5063A ENA Series Network Analyzer

Keysight Technologies 87405C 100 MHz to 18 GHz Preamplifier. Technical Overview

Process Control Calibration Made Easy with Agilent U1401A

Agilent Spectrum Visualizer (ASV) Software. Data Sheet

Keysight Technologies, Inc. UWB Antenna Measurements with the 20 GHz E5071C ENA Network Analyzer. Application Note

Agilent 8762F Coaxial Switch 75 ohm

In-circuit Measurements of Inductors and Transformers in Switch Mode Power Supplies APPLICATION NOTE

Power Analysis Application Module DPO4PWR MDO3PWR Datasheet

ADS-SystemVue Linkages

Keysight Technologies How to Perform QSCV (Quasi-Static Capacitance Voltage) Measurement

Keysight Technologies Using an External Trigger to Generate Pulses with the B2960A


Keysight Technologies Making Simpler DC Power Measurements with a Digital Multimeter

Keysight HMMC-1002 DC 50 GHz Variable Attenuator

Agilent 8761A/B Microwave Switches

Keysight Technologies Phase Noise X-Series Measurement Application

Keysight Technologies Measuring Group Delay of Frequency Converters with Embedded Local Oscillators. Application Note

Keysight Technologies U9400A/C Solid State FET Transfer Switches

Keysight Technologies Using a Network and Impedance Analyzer to Evaluate MHz RFID Tags and Readers/Writers

Keysight Technologies Power of Impedance Analyzer

Keysight Technologies Optimizing VNA Settings for Testing of LTE-A Wireless Components. Application Note

Solar Array Simulation System Integration

AN2842 Application note

Keysight Technologies 8490G Coaxial Attenuators. Technical Overview

Transcription:

High Speed Parametric Test Using Agilent 4070 Series Throughput Tuning Techniques for Parametric Test Agilent 4070 Series Semiconductor Parametric Tester Application Note 4070-6 Introduction Constant advances in semiconductor device scale and integration are leading to rapid increases in the number of devices in a single semiconductor chip. Quick process flaw detection and improved yield for semiconductor chips are becoming more important than ever for semiconductor device manufacturers. In process integration and process monitoring, more evaluation items using various test structures are required. The parametric test plays a crucial role in evaluating those increased test items. High throughput for evaluating the huge number of test items is strongly required. Agilent Technologies has been offering parametric testers in two generations the HP4062 series and the Agilent 4070 series. These series have been recognized as de- facto standard parametric testers and widely used by most semiconductor manufacturers around the world. In particular, the Agilent 4070 series has the fastest throughput in the industry and can perform a huge number of measurements with high speed and high accuracy. However, if a program developed for the HP 4062 series is used with the Agilent 4070 series without any modification, the throughput of the Agilent 4070 is not necessarily the best. Throughput can be improved dramatically by making a small change in the input parameter values or the TIS statements in an algorithm. This application note describes know- how and techniques to make high speed parametric testing using the Agilent 4070 series semiconductor parametric tester. The first chapter introduces four selected techniques, which can be mastered in a short period and can be implemented in a few hours. The subsequent chapters describe more techniques and know- how for making the further throughput improvement. [1] Four simple tuning techniques for dramatic throughput improvement A few slight adjustments to the measurement program can boost throughput tremendously. In the following, four simple techniques are described.

(1) Wait Time Optimization 1) Search and find WAIT and Wait_th commands in measurement algorithms. 2) If the current to measure is more than 100 na and the device response speed is supposed to be fast, change the wait time defined by WAIT or Wait_th command to zero. 3) If the current to measure is low current (< 100 na), make sure that the low current measurement port (SMU1 or SMU2) is used for the measurement. If not, change the port to a low current port. If a long wait time such as 1 second is used, then reduce it by a large margin unless the dielectric constant of the device is very large. 4) Search and find the Set_lsearch and Set_iv commands in the measurement algorithms. 5) Check the hold time and the delay time specified in those commands. If the current to measure is more than 100 na and the device response time is supposed to be fast, change those times to zero. (2) Integration Time Optimization 1) Check if the Set_smu command is used in the measurement algorithms. 2) If Set_smu is used, then check the integration time, which is specified by the first parameter. If the current to measure is more than 100 na and the integration time is set to medium or long, it should be changed to short. In the Agilent 4070 series, use the Set_smu_ch command to specify the ADC type for each measurement port and assign integration time with the second parameter of the Set_adc_i or Set_adc command. If the current to measure is more than 100 pa and less than 100 na and the integration time is set to long, then change the integration time to medium by using the Set_smu_ch and Set_adc_i commands. (3) Use of Limited Auto Range and Adjustment of Compliance 1) Search and find the Measure_i and Sweep_iv commands in measurement algorithms. 2) If the third parameter for these commands is set to zero, auto- ranging is used. If the expected current is not so small, then set the parameter to a higher current range such as 100 ua by considering the expected current value. In particular, if the measurement range is set to 1 na or lower, the internal wait time taken by the Agilent 4070 series becomes long. It is recommended to use a range higher than 10 na range unless ultra low current measurement with high accuracy is required. 3) Check the compliance value set by the Force_v or Set_iv command which is usually placed before the Measure_i or Sweep_iv command. (Compliance is set by a fourth parameter for Force_v and a ninth parameter for Set_iv.) Adjust the compliance to the appropriate value by considering the expected current. For instance, if the compliance is set to 100 ma when 150 ua is an expected current value, then the compliance should be lowered to the appropriate value such as several hundreds ua. (4) Application of Set_vth, Measure_vth and Set_lsearch commands to Vth 1) If there are many MOSFETs in the TEG (Test Element Group) and Vth measurement is required for them, then the intelligent search function is effective. In such a case, edit measurement algorithms, which include Vth measurement. 2) If the Vth is extracted by drawing a regression line on Id (or square root of Id) versus Vg characteristics from the point where the slope is the maximum, then the measurement and the Vth calculation portion should be replaced by the Set_vth and Measure_vth commands. 3) If the Set_vth command is used, three parameters (Id fine search start, Skip, Skip back) have to be assigned in order to use the intelligent search function. Ideally, Id fine search start should be as high as possible so long as the drain current does not reach the threshold current. Practically, you will obtain reasonable throughput improvement just by specifying a value 10 times larger than Idoff. Likewise, though it is desirable to find the Skip value by trying out several values for the best throughput, you will obtain reasonable throughput optimization by specifying a number around one fifth the number of total sweep points. Skip back value should always be 1. 4) If Set_lsearch is used, you will assign two input parameters, Skip and Skip back. Just like the explanation in 3), you get reasonable throughput improvement by specifying a number around one fifth the total sweep points for Skip. Skip should always be 1. 2

[2] Know-how and techniques for further throughput improvement on Agilent 4070 series Repeatability (%) Correlation (%) 96 97 98 99 100 Difference owing to technique Difference owing to hardware performance 30 35 40 45 50 55 60 Measurement Time (Min.) Figure 1. Relationship between speed and accuracy 95 96 97 98 2.24% less correlation 18.7 ms (x 5 times) 0.64% less correlation 59.3 ms (x 1.6 times) 99 100% Correlation - 70% 94.8 ms 100-40% 0 50 100 Throughput (ms) Figure 2. Throughput vs. Correlation (Idoff) Correlation (%) 99.5 99.6 99.7 0.23% less correlation 53.1 ms (x 12.9 times) 0.04% less 99.8 correlation 100% correlation 102 ms 99.9 (x 6.7 times) 683 ms 100-50% -85% 0 200 400 600 800 Throughput (ms) Figure 3. Throughput vs. Correlation (Vth) Defining Measurement Accuracy Target Acquisition of Reference Data Bottle-neck Analysis of Entire Measurement Tuning Figure 4. Flow of throughput tuning Throughput versus Accuracy In general, measurement throughput and measurement accuracy are in a trade- off relationship. For instance, faster speed is achieved by using shorter integration time. However, measurement accuracy (repeatability, correlation) may be degraded. On the other hand, longer integration time improves accuracy, but speed will be slower. Figure 1 shows a graphical relationship between speed and accuracy. As this trade- off curve is determined by the performance of the tester hardware, it shifts to the left as indicated by the dashed line if the tester hardware performance is superior. In spite of the general relationship described above, there are some special techniques for improving the throughput without sacrificing measurement accuracy. The curve shifts to further left if these special techniques are used. Figures 2 and 3 show the relationship between speed and accuracy plotted by using actual measurement results on a real device. These are examples for low and medium current measurements respectively. As you can see from the figures, you will get significant speed improvement by allowing a small deviation in correlation. Especially in the Vth measurement example, allowing only 0.2% correlation deviation produces nearly 13 times throughput improvement. In general, requirements for Repeatability target 1) For I measurement < 10pA (for calculated parameters raw data is used instead) Sigma (s) <=1pA 2) For capacitance measurements (for calculated parameters raw data is used instead) Repeatability <= 5.0% 3) For all other measurements Repeatability <= 1.0% Repeatability definition Repeatability = (s) / (mean result) Repeatability is defined using 10 measurements at the same device on the same site on the same wafer except for measurement items such as breakdown voltage measurement which may destroy or damage the device parametric tests differ for R&D and production. In R&D, measurement accuracy is prioritized over throughput. In production, the fastest throughput is required as long as enough accuracy is obtained for process monitoring. When actually adjusting throughput of measurement algorithms, it is important to set the target measurement accuracy limit. Speed is adjusted within the defined target limit. Flow of Throughput Tuning Figure 4 shows a flow of throughput tuning. When adjusting throughput, it is important to focus not only on the measurement algorithms but also on the bottlenecks found by looking over the entire sequence. For instance, bottlenecks might be prober movement or data handling routine. Defining Measurement Accuracy Target The measurement accuracy target can be defined by two characteristics, repeatability and correlation. Figure 5 shows an example definition of repeatability and correlation. Target values for those characteristics should be determined by considering the requirements for the measurement. Acquisition of Reference Data Reference data is necessary in order to evaluate the correlation of measurement results. Reference data has to be accurate for meaningful correlation evaluation. Therefore, reference Correlation target 1) For I measurement < 10pA (for calculated parameters raw data is used instead) Reference result Measured result <= 1pA 2) For capacitance measurements (for calculated parameters raw data is used instead), for I measurements > 10pA and <100nA, and for V measurement < 1V (both for calculated and noncalculated parameters) Correlation <= 3.0% 3) For all other measurements Correlation <= 1.0% Correlation definition [(reference result) - (measured result)] Correlation % = 100 x (reference result) Figure 5. Example definition of measurement accuracy target 3

Wafer in Wafer out Analysis data should be taken by using a trustworthy parametric tester or instrument with sufficient integration time. The Agilent 4070 series or the Agilent 4156C, both of which are de- facto standard measurement equipment, will be the right choice for this purpose. Bottle-neck analysis It is crucial to find the bottlenecks in the entire test sequence to achieve efficient throughput tuning. The bottleneck analysis can be done in two steps. In the first step, break down the entire execution time into smaller chunks such as execution time per algorithm or prober index time. Figure 6 shows the concept of the first step. By making the first step analysis, you can find items that account for a large percentage of the overall execution time. If you use the Agilent SPECS (a test shell for the Agilent 4070 series), you can easily log the execution time of each algorithm by the following method. Figure 6. Analysis of entire execution time Wafer alignment Probing index time Gate leakage Idon Vth Idoff Isubmax Bvdss Cap R Results display Test-shell execution overhead 1) Create a directory for the logging $ mkdir /SPECS/usr/spool/prof 2) Set the write permission to the above directory $ chmod 777 /SPECS/usr/spool/prof If you execute a test plan, a file is created under the directory for logging. Execution time for each algorithm is recorded in a millisecond in the file. If the Agilent SPECS is not used, execution time of each algorithm can be examined by inserting a time checking command right before and after the algorithm. In the second step, real bottlenecks are found. Tuning will have a big effect if some items take up longer execution time than others and the execution time of those items is expected to be reduced appreciably by applying some of the tuning techniques described later. For instance, a sweep measurement, which has a large number of measurement steps, can be speeded up significantly. If such measurement takes longer than other measurement items, then the priority for tuning the item should be set to high. Efficient tuning is possible by prioritizing the items to tune with the analysis method mentioned above. Measurement Tuning Measurement tuning is done by applying several tune- up techniques to algorithms and the test plan. The speed should be maximized as long as the measurement accuracy meets the defined repeatability and correlation target limits. When applying a tune- up technique, it is best to use the one that has less effect on measurement accuracy before the one that might cause accuracy degradation if overly tuned. It is also advisable to check the repeatability whenever applying one technique. Usually, if you improve the repeatability better, the correlation automatically becomes better. If you use the Agilent SPECS, a repeatability check can easily be done by using an inspection mode and a repetitive measurement function in the algorithm panel. When checking the correlation, you need to set identical conditions as much as possible. For instance, the same devices on the same wafer should be used. The environment such as temperature or prober should be identical. In the following, various throughput- tuning techniques are described in detail. Tune-up Techniques The following is the summarized list of tune- up techniques. 1) Use faster measurement function 2) Use intelligent search 3) Limit the range changes 4) Optimize wait/delay time 5) Use high- speed ADC 6) Optimize integration time 7) Avoid SMU filter 4

8) Use two integration time settings for wide range measurement These techniques can be categorized into two groups. Techniques in the first group are 1) and 2); these improve throughput without degrading measurement accuracy. Ones in the second group are 3) to 8); these might degrade measurement accuracy if overly tuned up. Even with the techniques in the second group, you can improve the measurement throughput without degrading measurement accuracy if the amount of adjustment is appropriate. 1) Use faster measurement function If the measurement program on the Agilent 4070 series was originally developed for the HP 4062 series, the repetitive loop of Force_v / Measure_i, (Measure_cmu) is often used for sweep measurement such as I- V and C- V measurements. Those Coarse measurement in the first sweep Id Goal Fine measurement in the second sweep Id Goal (1) Skip back commands were developed for a single spot measurement. Dedicated commands for sweep or search measurement are provided, Set_iv / Sweep_iv for I- V sweep, Set_cv / Sweep_cv for C- V sweep, Set_lsearch / Search for I- V search, and Set_bdv_search / Search for breakdown search. Sweep commands are faster than spot commands. Search related commands are even faster than sweep commands. Using faster measurement commands depending on the application will improve the throughput without degrading the measurement accuracy. Especially when the repetitive loop of Force_v and Measure_i is used for sweep measurement, just replacing it with sweep or search dedicated commands improves throughput improved substantially. 2) Use intelligent search The Agilent 4070 series provides you with so- called intelligent search TIS commands. As shown (3) Vth Vg (2) Measurement using fine voltage step Figure 7. Vth measurement using intelligent search Vg in Figure 7, it performs the sweep measurement twice. In the first sweep, coarse measurement is done to roughly find the target. In the second sweep, fine sweep is done only for the area close to the point found in the first measurement. In general, measurement for lower current takes longer in the parametric test because the internal wait time and integration time are longer in order to obtain good accuracy. However, for Vth or breakdown measurement, measurement points really required are only those around the target current. Intelligent search commands allow you to reduce the sampling of time consuming low current measurement while keeping the fine resolution for the middle to high current range, thus enabling very fast yet very accurate search measurement. 3) Limit range changes [3-1] Use of limited auto ranging You can perform the most accurate measurement with the maximum resolution if the smallest measurement range, that includes the actual current or voltage, is used. On the other hand, the most accurate measurement is not necessarily required in all situations. For instance, in Id- Vg measurement for extracting Vth of MOSFET, high accuracy is not required for the low current range because low current results do not contribute to the calculation of Vth, but high resolution is required for the middle to high current range. If you use auto ranging for this type of measurement, the range that gives the maximum resolution is automatically selected for each measurement. Because low current measurement takes longer than medium or high current measurement, auto range measurement wastes time making unnecessarily accurate low current measurement. If you use 5

limited auto ranging of the Agilent 4070 series, accurate measurement can be done just for the required current range without performing unnecessarily accurate measurement for the low current range. Limited auto ranging behavior is the same as auto ranging except the range does not lower below the specified minimum range. In other words, limited auto range performs measurement with coarse resolution for current lower than the specified range and with high resolution for current higher than the specified range. By using limited auto ranging, you can perform very fast measurement by avoiding unnecessarily high resolution measurement and unnecessary range changing while keeping good accuracy for the range where required. [3-2] Use of quasi-fixed range If the DUT characteristic is roughly known, you can perform high- speed measurement by specifying the measurement range, thus eliminating the time required for range changing. The Agilent 4070 series does not have an explicit fixed range function. However, if you specify the same value for measurement range and compliance, you can virtually perform the fixed range measurement. (Figure 8) Icomp = 1E-3 Irange = 1E-3 : Set_iv( High, 1,20,0,3,11,0,0, Icomp ) Sweep_iv( High,2,Irange,Im(*) ) Figure 8. Quasi-fixed range [3-3] Avoidance of unreasonably high compliance setting If you specify both a measurement range and a compliance, the current / voltage range that the SMU can use for the measurement can be determined. When the SMU searches the best measurement range in its limited auto ranging sequence, ranging operation always starts from the range specified by the compliance. If you specify unnecessarily high compliance, range changes more than necessary. The purpose of setting the compliance is to avoid damaging the DUT with high voltage or current. In this sense, specifying too high compliance is not desirable. Setting appropriate compliance leads to throughput improvement. Usually, the disable_port command is used at the end of each measurement algorithm to set the SMU to inactive status. Then the compliance is set back to the default value, which is 100 ua. Therefore, 100 ua is recommended as a compliance for the measurement unless higher current is expected. 4) Optimize wait/delay time Measurement wait and delay times are used for two purposes. One is to avoid the error caused by the transient response from the DUT. The other is to avoid undesirable influence from noise such as dielectric absorption or thermal drift caused by the measurement environment. When transient response of the DUT is slow, for instance evaluating leakage current of amorphous silicon MOSFET or dielectrics, you need to set a wait time or a delay time appropriate for the transient characteristics of the DUT. When avoiding noise caused by the measurement environment such as dielectric absorption of the cable, a wait time or a delay time is necessary so that measurement is performed after the noise settles. In both cases, appropriate wait time or delay time has to be chosen by performing a time domain (I- t) measurement. On the other hand, the Agilent 4070 series has a specially designed fully guarded measurement path which greatly reduces the noise current caused by the measurement equipment. Figure 9 shows a settling characteristics taken by the Agilent 4073A when applying 100 V to the measurement path. Even though a much larger than usual voltage is applied, the dielectric absorption current diminishes less than 1 pa in 60 ms. Since dielectric absorption current is roughly proportional to the applied voltage, if the usual voltage such as 2 V is applied, then the dielectric absorption current should go down to less than 1 pa within 2 ms. Note that this 2 ms is necessary only for very low current measurement under 1 pa. For medium or high current measurement, there is no need to wait for the settling because the noise is much smaller than the measurement range from the beginning. If a measurement program used for the HP 4062 series is used as it is for the Agilent 4070 series, you can expect drastic throughput improvement just by adjusting the wait/delay time to a much smaller value while maintaining the measurement accuracy. This is enabled simply by the remarkable tester performance improvement. 5) Use high-speed ADC The Agilent 4070 series has two types of ADC (Analog to Digital Converter). One is the high resolution ADC (HR- ADC) and is shared by all SMUs. The other is the high speed ADC (HS- ADC) equipped with each SMU. The HS- ADC can perform measurement faster than the HR- ADC though measurement accuracy is not as good as with the HR- ADC. Although its accuracy is slightly 6

Im (A) 1.E-09 1.E-10 1.E-11 1.E-12 1.E-13 1.E-14 1.E-15 1.E-16 0.01 0.1 1 10 100 Time (s) inferior, the HS- ADC has good enough accuracy for normal measurement. To cite a case, the HS- ADC gives you better repeatability than the HP 4062 series. By using the HS- ADC for most of the measurements and using the HR- ADC for only selected measurements such as very low current measurements that require high accuracy, you can improve the measurement throughput. You can use the HS- ADC as follows: [1] Set the optimization level to 2 or 3. [2] Specify 0 to a parameter for determining ADC type in the Set_smu_ch command for each SMU. Do not use the Set_smu command. 6) Optimize integration time Measurement integration time directly relates to measurement accuracy, especially on repeatability. If you choose a long integration time, more accurate measurement can be done because the noise is canceled out, but the throughput is degraded. With a probe card Without a probe card Figure 9. 4073A settling characteristics (when 100 V is applied) On the other hand, if you use a short integration time, throughput improves but the accuracy may be degraded. When tuning throughput by setting the target limit for measurement accuracy, integration time is mainly tweaked so that throughput becomes the maximum while meeting the target accuracy. If you compare the HP 4062 series and the Agilent 4070 series, hardware performance improvement in the Agilent 4070 series enables the same level of measurement accuracy with a shorter integration time. Therefore, if a measurement program written for the HP 4062 series is used for the Agilent 4070 series, just by shortening the measurement integration time, the measurement throughput can be improved without sacrificing measurement accuracy. You can set an integration time by using the Set_adc or Set_adc_i command. 7) Avoid SMU filter use The output filter equipped with each SMU is a LPF; this slows down the ramp speed of the SMU output and reduces spikes and noise. The SMU in the Agilent 4070 series is carefully designed to suppress spikes and noise. The possibility of spike and noise occurring even when the SMU filter is not used is very low. By setting the SMU filter to off, you can improve throughput and keep enough accuracy for most of the measurements. You can set the SMU filter to off as follows. [1] Set the optimization level to 2 or 3. [2] Assign 2 to the filter specifying parameter in the Set_smu_ch command for each SMU. 8) Use two integration time settings for wide range measurement The Agilent 4070 series has a function that allows you to change the integration time dynamically during measurement execution, depending on the level of current. You can perform the measurement in a shorter period by using this function. For instance, longer integration time such as medium is used for low current less than 100 na and short integration time for current more than 100 na. Then you can perform accurate measurement over the entire range of current in a shorter measurement time. The Agilent 4070 series has another useful function, which is called smart mode, to optimize the integration time for low current measurement. The smart mode function automatically finds out the appropriate integration time depending on the premeasurement for low current. You can improve the measurement with reasonable accuracy by using two integration times and applying a smart mode for low current. 7

Tester [Condition] Tester [Condition] 4071A [Tuned] 4071A [Original] 4072A [Tuned] 4072A [Original] 0.16 0.16 Tune-up Examples 0.44 Quantitative throughput improvement estimation requires thorough inspection of the measurement items, algorithms, and devices to measure. Therefore, it is hard to say how much improvement you will get without scrutinizing those conditions. Figure 10 shows two example results of a tune- up exercise at two actual customer sites using actually operated measurement programs. In both cases, the accuracy target shown in Figure 5 is complied with when tuning up the measurements. Case (A) shows an example using the Agilent 4071A. Approximately 3.2 times throughput improvement is achieved. Case (B) shows an example using the Agilent 4072A. About 4.1 times throughput improvement is achieved. 1.40 0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 Time (Min. / die) Figure 10. Example of throughput tuning Case (A) 0.04 0.04 0.25 1.03 Summary Probing Time Testing Time 0.00 0.20 0.40 0.60 0.80 1.00 1.20 Time (Min. / die) Figure 10. Example of throughput tuning Case (B) Probing Time Testing Time Implementing several easy tuning techniques allows you to perform very fast yet very accurate parametric testing. Parametric tests, which are becoming even more bulky and important, can be processed with the highest throughput in the industry by using the Agilent 4070 series and some appropriate tunings. For more information about Agilent and its products, go to www.agilent.com. For more information about Agilent Technologies semiconductor test products, applications, and services, visit our web site: www.agilent.com/ go/semiconductor or you can call one of the centers listed and ask to speak with a semiconductor test sales representative. Americas Brazil (11) 4197-3600 Canada (French) 1 877 894-4414 Canada (English) 1 800 447-8378 Mexico 33 134-5841 United States 1 800 447-8378 Asia/Asia Pacific Australia 1800 629-485 China 800 810 0509 Hong Kong 852 2599 7889 India 1600 112 626 Japan 0120 421 345 Malaysia 1800 880 780 New Zealand 0800 738 378 Philippines 1800 1651 0135 Singapore 1800 276 3059 South Korea 080 778 0011 Taiwan 0800 047 662 Thailand 1800 2758-5822 Europe Austria (0) 1 25125 7006 Belgium (0) 2 404 9340 Denmark 70 13 15 15 Finland (0) 10 855 2100 France (0) 825 010 710 Germany (0) 1805 24 6333 Greece (0) 20 547 9999 Ireland (0) 1 890 924 204 Italy 02 9260 8484 Netherlands (0) 20 547 2111 Poland 22 608 14 40 Russia (0) 20 547 9999 Spain 91 631 3300 Sweden 0200 88 22 55 Switzerland (Italian) (0) 2 92 60 8484 Switzerland (German) (0) 1 735 9200 Switzerland (French) 080081 5353 United Kingdom (0) 7004 666666 Middle East Israel 3 6892 500 Product specifications and descriptions in this document subject to change without notice. Agilent Technologies Inc. 2003 Printed in U.S.A September 24, 2003 5988-9964EN