DiMod 307 and SAE J211-1 Compliance. Dipl.-Ing. (FH) Jochen Schulz, Service Engineer, Kistler Automotive GmbH, Heidelberg, Germany

Similar documents
TDAS PRO and SAE J211 Compliance

Kistler View2017. Understanding in order to change. Page 3 Measurement solutions for the challenges of our times. Page 4 Direct path to test results

Kistler Instrumente GmbH has new German headquarters.

Kistler View01/2013. Kistler measurement technology reduces emission of pollutants in an increasingly mobile world. Get Better. With Kistler.

Kistler View2015. Smart Measuring Solutions for Self-Regulated Production Processes. Get Better. With Kistler.

DIGITAL FILTERING OF MULTIPLE ANALOG CHANNELS

The new dimension in torque measurement

EDS-400A COMPACT RECORDER. New KYOWA ELECTRONIC INSTRUMENTS CO., LTD.

Kistler LabAmp for DynoWare

Application Note (A12)

Chapter 2 Analog-to-Digital Conversion...

USB Dynamic Signal Acquisition

Optimizing vibration assisted drilling

DT9838 Strain Measurement Module

Kistler LabAmp. Electronics & Software. Charge Amplifier and Data Acquisition Unit for Dynamic Measurements. Type 5165A...

DT9838. Strain- and Bridge-Based Measurement Module. Key Features: Bridge Configurations. Analog Input Features

Production Analyzer Rev (USB + FireWire) Rev. 2.x (USB only)

Pulse-Width Modulation (PWM)

Optimized testing of electric drives

Electronic Counters. Sistemi Virtuali di Acquisizione Dati Prof. Alessandro Pesatori

Vibration Transducer Calibration System

Ethernet-Based Temperature, Voltage and Strain Measurement Modules

Annex to the Accreditation Certificate D-K according to DIN EN ISO/IEC 17025:2005

Isolated, Frequency Input 5B45 / 5B46 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

ISO Rules Part 500 Facilities Division 502 Technical Requirements Section Aggregated Generating Facilities Technical Requirements

HD Radio FM Transmission. System Specifications

Windfreak Technologies SynthHD v1.4 Preliminary Data Sheet v0.2b

P2042A LCD Panel EMI Reduction IC

80 khz Cylindrical Ultrasound Transducer

OPERATION MANUAL LDM-1000

PXI Timing and Synchronization Control

DIS-5010A, DIA-512A. Data Acquisition System for Crash Test AUTOMOTIVE TEST EQUIPMENT

SC-4/M User manual 4-channel measuring amplifier

THE BENEFITS OF DSP LOCK-IN AMPLIFIERS

HD Radio FM Transmission System Specifications

3D Distortion Measurement (DIS)

SmartCrash Overlap Segment

Agilent 81980/ 81940A, Agilent 81989/ 81949A, Agilent 81944A Compact Tunable Laser Sources

RX1032 PRECISION TEMPERATURE MEASUREMENTS FOR RUGGED ENVIRONMENTS FEATURES. 32-Channel Thermocouple Measurement System

Agilent Spectrum Visualizer (ASV) Software. Data Sheet

EX FEATURES. Stand-alone 48-channel unit with built-in Ethernet controller. Built-in bridge completion and Excitation

Racal Instruments. Product Information

SPIDER-80SG. General Data Acquisition Device with Strain Gage Measurement. Spider Measurement Solution

Guide Version Five techniques for fast, accurate power integrity measurements

ni.com Sensor Measurement Fundamentals Series

DIGEM f 96 x 48 EK. Applications. Description

Figure 1: Functional Block Diagram

TIME AND FREQUENCY ACTIVITIES AT THE CSIR NATIONAL METROLOGY LABORATORY

Contents. ZT530PCI & PXI Specifications. Arbitrary Waveform Generator. 16-bit, 400 MS/s, 2 Ch

Differential Two-Wire Hall Effect Sensor IC TLE 4942 TLE 4942 C

ATA-2001 Analog LVDT/RVDT Signal Conditioner with digital calibration

Passive Current Transducers for Sinusoidal Alternate Currents From A/0...5 A MCR-SLP-1/5-UI-0(-SW)

CERTIFICATE OF CALIBRATION

Positioning a new product brand

ASK Cards. Features Analog signal isolation Digital signal isolation. Applications

MiniSense 100 Analog PCB

easypll UHV Preamplifier Reference Manual

24-Bit, ks/s Dynamic Signal Acquisition and Generation

Agilent 8657A/8657B Signal Generators

Ideal for high dynamic range measurements from compression to noise floor

APPLICATION NOTE. Atmel AVR127: Understanding ADC Parameters. Atmel 8-bit Microcontroller. Features. Introduction

VT1419A Multifunctional Plus Measurement and Control Module

Precise, simultaneous data acquisition on rotating components Dx telemetry: from single channels to complex multi-component systems

Technical Datasheet UltraScope USB

EMX-4350 APPLICATIONS FEATURES A SMART PXI EXPRESS 625 KSA/S 4-CHANNEL DIGITIZER

7 Hints That Every Engineer Should Know When Making Power Measurements with Oscilloscopes.

High Pressure OEM Sensor RHU50

RCTrms Technical Notes

Measurement Setup for Phase Noise Test at Frequencies above 50 GHz Application Note

Features EX1629. High-performance Remote Strain Gage Measurement Unit. Overview. 48 programmable channels in a single unit

Multipoint measuring units UPM100 and Centipede100

ATA-2001 Analog LVDT/RVDT Signal Conditioner with digital calibration

Ground vibration testing: Applying structural analysis with imc products and solutions

Analog Signal Conditioning Accessories

Blade Tip Timing Frequently asked Questions. Dr Pete Russhard

GFT bit High Speed Digitizer

D94 Series. 1 Hz to 400 khz* Low Noise Fixed Frequency. 4 - Pole Single Power Supply Anti-Aliasing Low-Pass Filters

CAEMAX D x telemetry. digital modular convenient. Modular, multi-channel telemetry system for a variety of applications. productive testing

DC/DC-Converters in Parallel Operation with Digital Load Distribution Control

Transcom Instruments. Product Brochure TRANSCOM INSTRUMENTS. Product Brochure

CHAPTER 6 EMI EMC MEASUREMENTS AND STANDARDS FOR TRACKED VEHICLES (MIL APPLICATION)

Analog Modulation Analysis (AM/FM/φM) Specifications

HART Modem DS8500. Features

ISOLATION AMPLIFIER. 20 SERIES Indispensable in highly noisy environments! Designed with State-of-the-art Analog Technology

Models FSW-0010 FSW-0020

TS V adjustable shunt voltage reference. Description. Features. Applications

CL55C CLICK ANALYSER The cost competitive, PC-driven, automatic, multi-channel discontinuous interference analyser

D98 Series. 1 Hz to 400 khz* Low Noise Fixed Frequency. 8- Pole Single Power Supply Anti-Aliasing Low-Pass Filters

D92 Series. 1 Hz to 400 khz* Low Noise Fixed Frequency. 2 - Pole Single Power Supply Anti-Aliasing Low-Pass Filters

PXIe Contents CALIBRATION PROCEDURE. 10 GHz or 20 GHz RF Analog Signal Generator

U8 I1. Linearity ±0.10 % f. s. (standard); optional ±0.05 % WB61

National Instruments Flex II ADC Technology The Flexible Resolution Technology inside the NI PXI-5922 Digitizer

AN-1011 APPLICATION NOTE

Data Sheet. Agilent M9185A PXI Isolated D/A Converter. DISCOVER the Alternatives... Agilent MODULAR Products. 8/16-Channel 16-bit, ±16 V

Keysight Technologies N1918A Power Analysis Manager and U2000 Series USB Power Sensors. Demo Guide

INDUSTRIAL VIBRATION SENSOR SELECTION MADE EASY

CoMo Injection/CoMo Injection Basic

SYNC separator IC with AFC

EX1000 Series EX1000A EX1000A-TC EX1016A EX1032A EX1048A EX10SC EX1000A-TCDC RELIABLE DATA FIRST TIME EVERY TIME.

CALYS 150. Advanced documenting multifunction calibrator thermometer

Transcription:

DiMod 307 and SAE J211-1 Compliance Dipl.-Ing. (FH) Jochen Schulz, Service Engineer, Kistler Automotive GmbH, 69126 Heidelberg, Germany

Contents 1 Performance Requirements for a Data Channel...3 2 Linearity Error...3 3 Frequency Response of the Filter...4 4 Frequency Response of the Digital Measuring Chain...4 5 Time Base...5 6 Relative Time Delay Error...5 7 Time of Contact......5 8 Dynamic Shunt Stimulation...6 9 Overview...8 2

DiMod-307 and SAE J211-1 Compliance Dipl.-Ing. (FH) Jochen Schulz, Service Engineer, Kistler Automotive GmbH, 69126 Heidelberg, Germany This document compares the technical measurement characteristics of the DTI measuring system with the requirements arising from SAE J211-1, Rev. MAR 2014, subsequently referred to as SAE J211-1. 1 Performance Requirements for a Data Channel SAE J211-1 describes a data channel in the form of a measuring chain comprising the transducer, data acquisition system, internal and external wiring and all the analytical processes that can influence the frequency and time behavior of the measuring chain. The DTI measuring system is designed for use with an extensive range of active as well as passive sensor technology. Therefore, the following study can only examine the behavior of a DiMod with the use of a DTI data recorder. To arrive at a complete assessment of the SAE compliance of the measuring chain, consideration must be given to the manufacturer's specific information on the transducer and the software used. On commissioning, any DiMod undergoes a complex and extensive series of tests that aim to ensure the correct functioning of individual electronic components and compliance with various specification limits. Where available, the aforementioned specification limits are defined in accordance with the requirements of SAE J211-1. 2 Linearity Error According to SAE J211-1, the linearity error must not exceed a maximum value of 2,5 % (in relation to the measuring range). The DTI measuring system provides multiple and simultaneous verifications of this parameter's compliance: Test Requirement in SAE J211-1 Specification of the DTI measuring system Commissioning/specification test 2,5 % in relation to the measuring range, measured between F L und F H For target gains of 10, 100 und 1 000 <0,2 % (3 different test cycles) Final test on the DiMod after final processing 2,5 % in relation to the measuring range, measured between F L und F H For fixed/defined gain <0,2 % Final test on the DiMod after final processing 2,5 % in relation to the measuring range, measured between F L und F H Dependent on the transducer type; typ. <1 % Table 1: Comparison between the requirement for the linearity error as per SAE J211-1 and the specification limits of the DTI measuring system 3

3 Frequency Response of the Filter At the input of every DTI measuring channel, there is a 5-pole filter with a 3-dB cut-off frequency of 4 khz. This filter is implemented with hardware, and its primary purpose is to provide analog protection against foldover (anti-aliasing filter). The measured phase response can now be used to define the corresponding time delays. SAE J211-1 requires a time variance <100 μs in the 30 Hz to 1 000 Hz frequency range (CFC1 000). As can be seen in the following graph (Graph 3), the DTI measurement technology can meet these requirements. Graph 1 shows the amplitude response of a filter of this sort, in relation to the signal frequency: Attenuation / db Graph 3: Time variance over signal frequency Graph 1: Amplitude response: anti-aliasing filter SAE J211-1 also describes another important parameter, namely the phase delay. The following graph (Graph 2) shows the phase response in degrees over the signal frequency (sampling rate 20 khz; 3 db cut-off frequency 4 khz): 4 Frequency Response of the Digital Measuring Chain To achieve filtering compliant with the requirement in SAE J211-1, digital post-filtering of the measurement signal is required, depending on the Channel Frequency Class (CFC) in each case. The following graph shows the prefiltered amplitude response of the digital output of a DiMod (blue). The red curve profile shows the associated digitally post-filtered signal (CFC1 000), which remains within the specification limits as per SAE (dashed lines). Attenuation / db Graph 2: Phase response over signal frequency Frequency / Hz Graph 4: Amplitude response of the digital output (blue) against the post-filtered signal (red) 4

5 Time Base SAE J211-1 requires a resolution for the time base of 1/100 seconds with a maximum error of <0,1ms. A 1-kHz central clock pulse is used to synchronize various DTI data recorders or other types of measuring and/or video equipment. This means that all devices used in the test receive a uniform time stamp in relation to the t 0 trigger signal. This pulse is generated by a very narrow-band phase-locked loop (abbreviated as PLL) with a capture range of ±200 ppm. The frequency accuracy of the PLL used for this purpose is stated as ±100 ppm, giving a value of 0,2 μs for the maximum time error. 6 Relative Time Delay Error According to SAE J211-1, the maximum relative time delay error of the respective measuring channels in relation to one another must not exceed a value of 1 ms. If the channels are mathematically correlated, moreover, a maximum relative time delay error of 0,1 ms is required. 7 Time of Contact Forwarding of the t 0 trigger signal to all components used in the test is an essential task which is accomplished by the 'ComBox' when DTI measurement technology is used. According to SAE J211-1, the contact time, i.e. the precise time that elapses between detection of the trigger contact and actuation of the trigger event (forwarding of the trigger) must not exceed a value of 0,4 ms. The following chart shows the determination of the ComBox contact time. A t 0 (yellow) is simulated with the help of a contact switch. Actuation of the trigger event is recorded at a t 0 output of the ComBox (turquoise): the time of contact for the DTI measuring system is less than 5 μs and is therefore significantly below the requirements of SAE J211-1. The relative time delay error is verified with the help of the DiMod's own oscillator pulse and a counter; this verification is completed for each measuring channel after final processing. The deviation between the ascertained counter value and the specified setpoint must be in the ±0,5 % range. This corresponds to a maximum inaccuracy of ±0,75 μs. Graph 5: Time of contact for the ComBox 5

8 Dynamic Shunt Stimulation The purpose of the dynamic shunt stimulation is to verify the filter coefficients after final processing and fixed wiring of the DiMod with the relevant analog transducer. For this purpose, use is made of the shunt switches located at the input which are now switched with corresponding frequencies. A square wave signal is generated due to the positive or negative transducer bridge imbalance; this signal is digitally captured and evaluated. The following analog signal forms can be recorded during the dynamic shunt stimulation (Stim R = 22 100 Ω; Gain R = 18 200 Ω): Frequency Unfiltered Filtered 100 Hz 1 khz Graph 6: Dynamic Shunt Stimulation 6

Frequency Unfiltered Filtered 2 khz 3,33 khz 5 khz Graph 7: Dynamic shunt stimulation In this series of tests, the different amplitude attenuations of the signals listed here are determined in relation to the 100 Hz signal. The values determined are stored in the memory of the relevant measuring channel, so they can be reviewed at any time by repeating this test. Typically, the following amplitude attenuations are determined for an unswitched (open) DiMod: 1 000 Hz typ. -0,6 db 2 000 Hz typ. -1,3 db 3 330 Hz typ. -3 db 5 000 Hz typ. -10 db 7

9 Overview Section in SAE J211-1 Description Requirements from SAE J211-1 Specifications of the DTI measuring system 4.1 Linearity error 2,5 % in relation to the measuring range between F L und F H 4.2 Amplitude response See the graphs in SAE: Figure 1: CFC600 and CFC1000 Figure 2 : CFC60 and CFC180 <0,2 % in the specification test Typ. <1 % for the calibration Filter test with sinusoidal excitation for specification test Dynamic shunt stimulation for specification test and calibration 4.3 Phase response Max. variation in the time delay of <100 μs in the range from 30 Hz to 1 000 Hz 4.4.1 Time base Min. resolution of 1/100 seconds for an error <0,1 ms 4.4.2 Relative time delay error <1 ms between all channels <0,1 ms between mathematically correlated channel <30 μs in the 30 to 4 000 Hz range Central synchronization pulse of 1 khz For a maximum error <1 μs <1 μs between all channels 4.6.4 Calibration of the frequency response Measurement of a known input signal from DC to 3 khz Filter test with sinusoidal excitation for specification test from DC to 4 khz Dynamic shunt stimulation for specification test and calibration from 100 Hz to 5 khz 8.1 Presample filtering In or above the range of CFC1000 4 khz hardware filtering compliant with CFC1000 corridor ranges 8.2 Sampling rate Min. 10 khz sampling rate per channel 20 khz sampling rate per channel Simultaneous capture of all channels 8.3 Resolution Min. 12 Bit and LSB <0,2 % of measuring range 16 Bit 8.4.1 Digital filtering CFC filters must comply with the limits in point 4.2 8.4.2 Scaling Software is utilized to ensure use of the correct scaling Attained with Butterworth 4-pole phaseless filter algorithms Definition of the gain factor during final processing of the DiMod Determination of the digital transmission factor by calibration Digital transmission factor in the memory of the DiMod 8.4.3 Zeroing Software is used for zeroing of the signal Several options for signal zeroing (depending on the evaluation software used) 10 Time of contact 0,4 ms 5 μs Table 2: Comparison between the requirements as per SAE J211-1 and the specification limits of the DTI measuring system 8

9

10

About us Sales Center Tech Center Production Center Kistler, the originator of piezoelectric measuring technology, is a leader in dynamic pressure, force, torque and acceleration measurement. Cutting-edge technologies provide the basis for Kistler's modular systems and services. Customers in industry, research and development benefit from Kistler's experience as a development partner, enabling them to optimize their products and processes so as to secure sustainable competitive edge. This owner-managed Swiss corporation plays a key part in the evolution of automobile production and industrial automation, and its innovative sensor technology also helps foster the development of many newly emerging sectors. Drawing on its extensive application expertise, and always with an absolute commitment to quality, Kistler drives innovations ahead in lightweight construction, vehicle safety, emission reduction and Industry 4.0. Some 1 500 employees at 56 facilities across the globe are dedicated to the development of new measurement solutions, and they offer individual application-specific support at the local level. Ever since it was founded in 1959, the Kistler Group has grown hand-in-hand with its customers and in 2015, it posted sales of CHF 329 million. About 10% of this figure is reinvested in innovation and research with the aim of delivering better results for every customer. 11

920-689e-04.16 2016, Kistler Group Kistler Group Eulachstrasse 22 8408 Winterthur Switzerland Tel. +41 52 224 11 11 Kistler Group includes the Kistler Holding AG and all its subsidiaries in Europe, Asia, Americas and Australia. Find your local contact on