EmbryoCellect. RHS Scanning and Analysis Instructions. for. Genepix Pro Software

Similar documents
In our previous lecture, we understood the vital parameters to be taken into consideration before data acquisition and scanning.

Instructions for Mapping * µarray Images using GenePix 5.0

NPTEL VIDEO COURSE PROTEOMICS PROF. SANJEEVA SRIVASTAVA

Scan slides (Axon Genepix 4200AL)

ScanArray Overview. Principle of Operation. Instrument Components

GenePix Application Note

Instructions for Howto Scan µarrays

NPTEL VIDEO COURSE PROTEOMICS PROF. SANJEEVA SRIVASTAVA

Products - Microarray Scanners - Laser Scanners - InnoScan 900 Series and MAPIX Software

Quick Guide. NucleoCounter NC-3000

GenePix Application Note

Product Information. Introduction

Arcturus XT Laser Capture Microdissection System AutoScanXT Software Module. User Manual

Volocity Tutorial Creating a Measurement Protocol in Volocity Software

Microarray Data Pre-processing. Ana H. Barragan Lid

Scanning Procedure Using SilverFast Ai Software (6.6.1r2b) 1

Guidance on Using Scanning Software: Part 5. Epson Scan

Scanning and Image Processing -by Steve Clough

6.12. SilverFastJobManager. SilverFast JobManager for Film Scanner. Overview

Developed by BioDiscovery, Inc. DualChip evaluation software User Manual Version 1.1

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

Brightness and Contrast Control Reference Guide

Before you start, make sure that you have a properly calibrated system to obtain high-quality images.

Scanning Setup Guide for TWAIN Datasource

Using the TWAIN Datasource

3. When you import the scanner for the first time make sure you change it from Full Auto Mode to that of Professional Mode.

dlsoft Barcode Analyser By dlsoft

Spotxel 1.7 Microarray Image and Data Analysis Software User s Guide

Using the TWAIN Datasource

Color and More. Color basics

Managing images with NewZapp

Learning Photo Retouching techniques the simple way

User Guide V10 SP1 Addendum

Scanning Setup Guide for the TWAIN Datasource

Monaco ColorWorks User Guide

i1800 Series Scanners

COPYRIGHT. Limited warranty. Limitation of liability. Note. Customer remedies. Introduction. Artwork 23-Aug-16 ii

Extreme Makeovers: Photoshop Retouching Techniques

Using Adobe Photoshop to enhance the image quality. Assistant course web site:

DataCapture Transcript Module Getting Started Guide

Using the TWAIN Datasource

Chapter 6: TVA MR and Cardiac Function

Chlorophyll Fluorescence Imaging System

Steps involved in microarray analysis after the experiments

Technical Bulletin. Guide for Using BD Cytometric Bead Array (CBA) Flex Sets with the BD Accuri C6 Flow Cytometer. Introduction

User Manual Veterinary

Importing and processing gel images

TOON BOOM HARMONY Advanced Edition - Compositing and Effects Guide (Server)

FLIR Tools for PC 7/21/2016

Stitching MetroPro Application

CALIBRATION MANUAL. Version Author: Robbie Dowling Lloyd Laney

Contents Foreword 1 Feedback 2 Legal information 3 Getting started 4 Installing the correct Capture One version 4 Changing the version type 5 Getting

AUTODESK INVENTOR Trial Projects

Océ Colour Copy 7.0. User Manual

Operating Rausch ScanCam within POSM.

UN-SCAN-IT gel TM Version 6.3. User's Manual for. Macintosh (OS X)

HP Designjet HD Scanner and T1200 HD Multifunction Printer

SCANNING METHODS. Edited 7/2014 JKH REPRESENTING ORIGINAL IMAGE CONDITION - 2 SCANS. 1) Open Adobe Photoshop.

Machinery HDR Effects 3

DOCUMENT SCANNER INSTRUCTIONS. Space. Backup. Count Only. New File. Scanner. Feeding Option Manual Auto Semi-Auto

EVOS M5000 Imaging System

GALILEO TMA CK 4500 HTS Tissue Microarray Platform

Drawing output in imos ix 2017 SR2. Date of creation: March 2018; last changes Version used: ix 2017 SR2

PASS4TEST. IT Certification Guaranteed, The Easy Way! We offer free update service for one year

We recommend downloading the latest core installer for our software from our website. This can be found at:

W i n d o w s. ScanGear CS-S 4.3 for CanoScan FB1200S Color Image Scanner. User's Guide

ScanGear CS-U 5.3 for CanoScan FB630U/FB636U Color Image Scanner User s Guide

RAGE TOOL KIT FAQ. Terms and Conditions What legal terms and conditions apply to the RAGE Tool Kit?

Quality control of microarrays

ColorPony User Manual

Image and Data Acquisition

An Introduction to Lasercut 5.3 Preparing the Artwork

Definiens. Tissue Studio 4.2. Tutorial 1: Composer and Nuclear Markers

Configuring User Settings at the Scan Station

Locating Molecules Using GSD Technology Project Folders: Organization of Experiment Files...1

Image Capture TOTALLAB

Center for Microscopy and Image Analysis Axio Scan.Z1 Operating Manual

Label Studio Quick Start Guide

UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY

Printer Software Guide

The ideal K-12 science microscope solution. User Guide. for use with the Nova5000

ScanMate. i920 Scanner. Scanning Setup Guide for TWAIN Applications A-61733

Version 4.2 For WINDOWS

Making an Architectural Drawing Template

Mullingar Camera Club Basic introduction to Digital Printing using Photoshop CC.

Collage Images. By Rob MacIlreith For Use with GIMP 2.8 Last Updated July 2013

Manual. Cell Border Tracker. Jochen Seebach Institut für Anatomie und Vaskuläre Biologie, WWU Münster

Adobe Photoshop PS2, Part 3

SKF TKTI. Thermal Camera Software. Instructions for use

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST

Drawing Layouts Paper space & Model Space

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope

Positive Pixel Count Algorithm. User s Guide

IncuCyte ZOOM Fluorescent Processing Overview

Océ Colour Copy 5.0. User Manual

ab Firefly Cytometer Setup Particles

Signal Integrity Analyzer

SilverFast. Colour Management Tutorial. LaserSoft Imaging

Toon Boom Harmony Reference Guide 3. Chapter 1: Dialog Boxes 7. Add Column Dialog Box 7. Add Drawing Layer Dialog Box 9

ISO Verification Fact Sheet

Transcription:

EmbryoCellect RHS Scanning and Analysis Instructions for Genepix Pro Software EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October 2015 1

Copyright Reproductive Health Science Limited 40-46 West Thebarton Road Thebarton, South Australia, 5031 EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October 2015 Reproductive Health Science Limited (RHS) makes no representations and gives no warranties of any kind in relation to the contents of this document and all warranties, conditions and other terms implied by statute or common law are, to the fullest extent permitted by law, hereby excluded. In particular, RHS assumes no responsibility for any errors or omissions that may appear in this document and makes no commitment to update or keep current the information contained in this document. RHS retains the right to make changes to this document (including any specifications contained herein) at any time without notice. This document is confidential. No part of this document may be modified, copied, reproduced, republished, published, transmitted or distributed in any form or by any means without the prior written consent of RHS. The contents of this document are to be used solely for the purpose for which they are provided by RHS and for no other purpose. All content, text, graphics and all other materials contained in this document are owned by RHS, and all proprietary and intellectual property (names, logos and trademarks) wherever arising in relation to this document, rest in RHS (or its licensors) and all such rights are reserved. EmbryoCellect is a Research Use Only product and is not to be used for diagnostic procedures EmbryoCellect is for Research Use Only and should not be used in diagnostic procedures. You are responsible for ensuring that you accurately follow the protocols provided in this Genepix Pro Scanning and Analysis Technical Data Sheet (TDS) and analysing and interpreting the results you obtain. RHS does not guarantee any results obtained. EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October 2015 2

Contents 1. Genepix Pro Software Compatibility... 4 2. General Instructions... 4 2.1 Slide format... 4 2.2 Scanner requirements... 4 2.3 Prescan... 4 2.4 Scanning... 5 2.5 Scan parameters... 5 2.6 Saving TIFF images... 6 3. Scanning Instructions... 6 3.1 Check the settings... 7 3.2 Perform and save a Preview Scan of the slide... 8 3.3 Scan each array individually... 8 3.4 Array quality control... 13 4. Quantifying the Data... 14 5. Analysing the Data using the RHS Macro... 16 6. Final Result Quality Control... 17 7. Macro Troubleshooting... 20 EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October 2015 3

1. Genepix Pro Software Compatibility The instructions described in this technical data sheet for the analysis of the EmbryoCellect TM array using the Genepix Pro software (Molecular Devices, USA) are compatible with version 6.1 and should be compatible with other versions. These instructions require the use of the Genepix Pro.gal files (Genepix_A, Genepix_B, Genepix_C, Genepix_D) and Macro 6.2, which are available from the RHS dropbox or the RHS website. The RHS analysis macro only works when the Excel decimal separator is a period (.) not a comma (,). While every effort has been made to ensure the protocol steps are compatible across all versions of Genepix Pro software, there may be a version where the user may experience a run error during the analyses of array data. If an error is encountered, please contact RHS immediately by emailing support@rhsc.com.au and we will provide you with assistance. 2. General Instructions 2.1 Slide format Each EmbryoCellect TM slide contains four microarrays with identical probes in each (Figure 1). A B C D Figure 1: Microarray slide template for sample loading displaying the locations of individual arrays Four sample batching is possible, making EmbryoCellect TM ideal for processing small numbers of samples. Slides cannot be reused so the recommended minimum number of samples to be hybridised is four. If all four microarrays are not hybridised, the unused microarrays will not be able to be utilised in subsequent hybridisations. 2.2 Scanner requirements A dual channel scanner equipped with a green and red laser (532nm and 635nm wavelength, respectively) is required to scan EmbryoCellect TM slides. 2.3 Prescan A prescan allows low-resolution visualisation of the entire slide surface, showing the array locations (see Figure 2). Most scanners will also read the slide barcode. When the prescan has completed, the scan area can be adjusted independently for each microarray for the higher resolution analysis scan. Figure 2: Example of a prescan image illustrating individual array locations EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October 2015 4

2.4 Scanning Some microarray scanners have automatic scanning settings which adjust the power of the laser and/or gain of the photomultiplier tube (PMT). The effects of these settings can be viewed on a histogram (see Figure 3). If required, these settings can usually be changed manually and should be adjusted until the red and green lines on the histogram overlap as much as possible (see Figure 4). Figure 3: Example of Histogram before scanning parameter optimisation. The high peak close to the Y axis represents the background signal. The secondary smaller peaks represent the microarray spot signal. The green line represents results from the green laser (532 wavelength test) and the red line represents results from the red laser (635 wavelength - reference). This image shows poor overlap between the green and red signals. Figure 4: Example of Histogram with ideal scan parameters. The red and green lines for the test and reference signal are overlapping. 2.5 Scan parameters In addition to the laser and PMT settings, other scanner software parameters should be adjusted for ideal EmbryoCellect TM slide scanning. Spot signals depend on several parameters, from sample DNA quality to hybridization and washing process success. Scan parameters should be adapted to each slide or array in order to fit within the following image quality control parameters; Pixel Saturation: EmbryoCellect TM slides must be scanned with minimal saturation because saturated pixels are excluded from the image. GenePix Pro s Auto-PMT option allows you to set the percentage of saturated pixels allowed. Set the option Saturation tolerance (%) to 1e-4 Pixel size: Set the Pixel Size (µm) to 10µm EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October 2015 5

2.6 Saving TIFF images Images should be saved with the following settings as a TIFF file containing both colour images (this is usually the default scanner setting). RHS recommends that files are saved using the slide barcode plus the array suffix as a minimum. E.g. RHS001234_A. Figure 5: Save settings 3. Scanning Instructions Switch on Scanner Start GenePix Pro Software Allow scanner to warm up for ~15 minutes before use Insert Slide Ensure that the slide will be scanned from the top to the bottom array by placing the slide in the correct orientation (the barcode is at the bottom of the slide) EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October 2015 6

3.1 Check the settings Open the options box Open the Workflow tab: Check the box: Search for Barcode after each Data Scan or Image load. Check the box: Register Images after a Data Scan. Open the Analysis tab: In the Ratio Formulation select 532/635. Open the Alignment tab: Change value for Minimum Diameter % to 50. Change value for Maximum Diameter % to 200. Change value for Composite Pixel Intensity (CPI) threshold to include a pixel in a feature during alignment to 500. Open the File save tab: In the Image and Results File Naming section, check the Prefix-Barcode box. Select OK EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October 2015 7

3.2 Perform and save a Preview Scan of the slide Open the Hardware Settings dialog box: Set PMT levels at 600 for both 635 and 532. Leave the Power % setting at 100% Set the desired Pixel size to 10µm Leave the Lines to Average setting at 1. Leave the desired Focus Position setting at 0. Press Preview Scan Save the Preview Scan Select File/save Select Export Images Save the preview scan in a desired location as a JPEG 3.3 Scan each array individually Define the region of interest: Select the Scan Area tool Press F10 to bring up the Scan Area options (alternatively right click with the mouse and select Scan Area Options). Array Left Top Width Height A 7000 B 21000 C 7080 35080 7800 4200 D 49200 Enter the following coordinates to select the scan areas for each of the RHS arrays, starting with array A. Press OK EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October 2015 8

Load the Array Layout (.gal file) Click the File button Select the Load Array List command. (Shortcut: Alt + Y) Select the corresponding.gal file for the array you are analysing. E.g. Use the.gal file Genepix_A when scanning and analysing the A array. Use AUTO PMT to set the PMT Gain automatically Click the Auto-PMT button in the Hardware Settings box. Set the Saturation tolerance (%) to 1e-4. Click the Start button. Once finished, press Apply button. Perform a Data Scan Set the resolution for the data scan to 10µm in the Pixel Size field in the Hardware Settings box Press Data Scan EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October 2015 9

Check for pixel saturation If there are any saturated spots; decrease PMT by 20 for both 635 and 532 then repeat the data scan. This image shows saturation in the 532 laser for some spots. The microarray should be rescanned with the PMT lowered by 20 for both lasers until no saturation is present. Align Template Line up the grid to overlap with the array by pressing F8 or click on the Gridding button and select the Find Array, Find Blocks, Align Features option. Ensure that GenePix Pro has correctly identified the position of the array and the blocks. If necessary due to incorrect spot positioning, incorrect flag, presence of debris or a scratch; If the array or the blocks are not correctly identified, move the grid manually into place with the mouse while in Block Mode. Move, resize, include or exclude spots by selecting Feature Mode Click once on a spot to select the spot + Use Ctrl + arrow keys to resize a grid spot Use the arrow keys to move a grid spot EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October 2015 10

Grid spots should be adjusted to avoid dust or debris If necessary; right click on a selected spot and select Flag as Not Found or Flag as Bad to flag a spot and exclude it from the dataset Right click on selected spot and select Clear flags except Absent or Clear flags to include a spot that was previously flagged as not found EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October 2015 11

If a halo effect occurs you may need to resize the spots to the inner ring of strong hybridisation otherwise it may be excluded from the dataset by the RHS macro. Save Image Select file, save images (Shortcut: Ctrl + S) In the Filename field, the slide name should already be filled in. Add the position of the array behind the barcode number (e.g. _A; _B; _C or _D Choose the correct folder for saving your image Click the Save button Proceed to steps 3.4 and 4 below before scanning other arrays on the same slide or another slide. EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October 2015 12

3.4 Array quality control Each array has four replicate blocks. Spots are printed in duplicate in each of the four blocks to make a total of 8 spots for each chromosome (for example in the image to the right each spot marked with a * is the same chromosome). Verify that each of the replicates are consistent. * * * * * * * * In this example there was a bubble which caused uneven hybridisation of the spots in the middle of the array. You can see two of the replicates encompassed in this bubble hybridised differently. These spots should be flagged as bad and excluded from the data analysis. In this example some spots in the bottom left quadrant have been scratched and they are no longer consistent between the 8 spot replicates. These spots should be flagged as bad and excluded from the data analysis. EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October 2015 13

Verify that there is consistent hybridisation across the whole array An array which appears like the above is considered less reliable and the sample should be re-hybridised Verify that there is consistent hybridisation across each spot 4. Quantifying the Data Press the Analyze button Arrays that have spots that have dark areas of hybridisation within them, as in the image above, are considered less reliable and the sample should be rehybridised EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October 2015 14

When the analysis is finished, it will tab to the Results tab located at the top. The raw data is presented. Click the Save Results as... button (Shortcut: Alt + U) to save the results as a.gpr file. Save this in the same location as the scanned images Use the following save parameters: Repeat from 3.3 Scan each array individually with the remaining arrays Repeat from 3.2 Perform and save a Preview Scan of the slide to scan additional slides EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October 2015 15

5. Analysing the Data using the RHS Macro Ensure macros are enabled Select enable macros when prompted or change settings in Excel Options Trust Centre Trust Centre Settings Macro Settings Open the RHS Mapix Analysis Macro The macro will open in the background of your raw data or will appear as an empty excel file Open the.gpr file with Microsoft Office Excel With the raw data file and the macro open Press Ctrl + Shift + E + + EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October 2015 16

6. Final Result Quality Control The Analysis tab displays the resultant Karyotype and Array SD Check the Array SD - Results where SD > 0.10 are less reliable Check the number of spots analysed for each chromosome - Results where less than 4 spots are analysed per chromosome are less reliable If a small number of spots were analysed (eg 4 or less), check the reason for exclusion. Common reasons for exclusion include: - Flagged by Mapix as bad - Flagged by user as bad - Flagged by Macro as bad This can be due to saturation, the spot being too small or too big, the spot not being uniform (eg having a halo (which is an area of strong hybridisation in the middle of the spot surrounded by an area of weaker hybridisation), donuts (which is an area of weak hybridisation in the middle of a spot surrounded by strong hybridisation) or poor hybridisation. It is recommended that the sample is rehybridised. Contact support@rhsc.com.au for additional information. EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October 2015 17

The ArrayData Graph tab displays the raw signal overlap between test and reference for each chromosome. The 2 coloured lines corresponding to the raw signal of both lasers should be as close to overlapping as possible. Where the test and reference lines do not overlap, this suggests: - A difference in the labelling efficiency between test and reference samples - Poor scanning parameters Results where the lines do not overlap are less reliable. 60000.00 50000.00 40000.00 30000.00 20000.00 10000.00 0.00 50000.00 40000.00 30000.00 20000.00 10000.00 0.00 Good overlap 1 3 5 7 9 11 13 15 17 19 21 X (Inv) Poor overlap 1 3 5 7 9 11 13 15 17 19 21 X (Inv) Ensure there is sufficient fluorescent signal by checking the signal intensity on the Y axis. RHS generally sees intensities in the range of 30,000-50,000 units Weak signals may be due to: - Poor scanning parameters i.e. not enough laser power or the PMT detection gain is too low, or; - Samples being poorly labelled. This can be seen in the spectrophotometer readings and on the agarose gel of the labelled products - Loss of labelled sample during purification and/or precipitation Results with low intensities are less reliable. 60000.00 50000.00 40000.00 30000.00 20000.00 10000.00 0.00 15000.00 10000.00 5000.00 0.00 Normal signal intensities 1 3 5 7 9 11 13 15 17 19 21 X (Inv) Weak signal intensities 1 3 5 7 9 11 13 15 17 19 21 X (Inv) EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October 2015 18

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 X (Inv) Yp Negative Mean of normalized ratios 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 X (Inv) Y Negative Mean of normalized ratios The Array Graph tab displays the mean normalized ratio for each chromosome; along with the standard deviation across each of the 8 spots analysed. Check the error bars on the Array Graph. Large error bars indicate: - inconsistent hybridisation across the array replicates This may be caused by debris, smearing of the microarray or bubbles, all of which are evident on the scanned image (jpeg file). Results with large error bars are less reliable than array results with smaller standard deviations per chromosome (see example) 2.00 1.75 1.50 1.25 1.00 0.75 0.50 0.25 0.00 2.00 1.75 1.50 1.25 1.00 0.75 0.50 0.25 0.00 Large error bars Chromosome Small error bars Chromosome EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October 2015 19

7. Macro Troubleshooting The RHS analysis macro only works when the Excel decimal separator is a period (.) not a comma (,). You can change these settings in Excel Options under the advanced tab. Please contact support@rhsc.com.au for technical assistance. EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October 2015 20

Reproductive Health Science Ltd (ABN 84 010 126 708 40-46 West Thebarton Road THEBARTON SA AUSTRALIA 5031 E: info@rhsc.com.au Ph: +61 88152 9383 ASX:RHS EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October 2015 21