A THREE DIMENSIONAL REAL-TIME MEMS BASED OPTICAL BIOPSY SYSTEM FOR IN-VIVO CLINICAL IMAGING

Similar documents
Miniaturized probe using 2 axis MEMS scanner for endoscopic multiphoton excitation microscopy

Optical coherence tomography

Copyright 2009 Year IEEE. Reprinted from IEEE TRANSACTIONS ON ADVANCED PACKAGING. Such permission of the IEEE does not in any way imply IEEE

In vivo three-dimensional microelectromechanical endoscopic swept source optical coherence tomography

Figure 1: Layout of the AVC scanning micromirror including layer structure and comb-offset view

Temporal coherence characteristics of a superluminescent diode system with an optical feedback mechanism

NIH Public Access Author Manuscript J Biomed Opt. Author manuscript; available in PMC 2010 May 3.

A miniature all-optical photoacoustic imaging probe

TIP-TILT-PISTON ACTUATORS FOR HIGH FILL-FACTOR MICROMIRROR ARRAYS

Piezoelectric transducer based miniature catheter for ultrahigh speed endoscopic optical coherence tomography

Integrated photonic circuit in silicon on insulator for Fourier domain optical coherence tomography

Moving from biomedical to industrial applications: OCT Enables Hi-Res ND Depth Analysis

Out-of-plane translatory MEMS actuator with extraordinary large stroke for optical path length modulation in miniaturized FTIR spectrometers

LMT F14. Cut in Three Dimensions. The Rowiak Laser Microtome: 3-D Cutting and Imaging

60 MHz A-line rate ultra-high speed Fourier-domain optical coherence tomography

OPTICAL COHERENCE TOMOGRAPHY: OCT supports industrial nondestructive depth analysis

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science

MEMS enabled microscopes for in-vivo studies of cancer biology

Mode analysis of Oxide-Confined VCSELs using near-far field approaches

OCT mini-symposium. Presenters. Donald Miller, Indiana Univ. Joseph Izatt, Duke Univ. Thomas Milner, Univ. of Texas at Austin Jay Wei, Zeiss Meditec

Optical design of a dynamic focus catheter for high-resolution endoscopic optical coherence tomography

MEMS for RF, Micro Optics and Scanning Probe Nanotechnology Applications

Vixar High Power Array Technology

Fiber-optic Michelson Interferometer Sensor Fabricated by Femtosecond Lasers

Characterization and Micro-assembly of Electrostatic Actuators for 3-DOF Micromanipulators in Laser Phonomicrosurgery

MICROVISON-ACTIVATED AUTOMATIC OPTICAL MANIPULATOR FOR MICROSCOPIC PARTICLES

CHIRPED FIBER BRAGG GRATING (CFBG) BY ETCHING TECHNIQUE FOR SIMULTANEOUS TEMPERATURE AND REFRACTIVE INDEX SENSING

462 IEEE JOURNAL OF SELECTED TOPICS IN QUANTUM ELECTRONICS, VOL. 10, NO. 3, MAY/JUNE 2004

2D Asymmetric Silicon Micro-Mirrors for Ranging Measurements

OCT. Optical Coherence Tomography

3D Optical Motion Analysis of Micro Systems. Heinrich Steger, Polytec GmbH, Waldbronn

Photoacoustic imaging using an 8-beam Fabry-Perot scanner

Ultrahigh speed endoscopic optical coherence tomography using micro-motor imaging catheter and VCSEL technology

MICROACTUATED MICRO-XYZ STAGES FOR FREE-SPACE MICRO-OPTICAL BENCH

Characterization of Silicon-based Ultrasonic Nozzles

A novel tunable diode laser using volume holographic gratings

Adaptive optics two-photon fluorescence microscopy

POINTING ERROR CORRECTION FOR MEMS LASER COMMUNICATION SYSTEMS

Single camera spectral domain polarizationsensitive optical coherence tomography using offset B-scan modulation

Imaging the Subcellular Structure of Human Coronary Atherosclerosis Using 1-µm Resolution

MICROMACHINED INTERFEROMETER FOR MEMS METROLOGY

High-speed spectral-domain optical coherence tomography at 1.3 µm wavelength

A water-immersible 2-axis scanning mirror microsystem for ultrasound andha photoacoustic microscopic imaging applications

MEMS in ECE at CMU. Gary K. Fedder

FRAUNHOFER INSTITUTE FOR PHOTONIC MICROSYSTEMS IPMS. Application Area. Quality of Life

Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI)

Full-range k -domain linearization in spectral-domain optical coherence tomography

Isolator-Free 840-nm Broadband SLEDs for High-Resolution OCT

POCKET DEFORMABLE MIRROR FOR ADAPTIVE OPTICS APPLICATIONS

some aspects of Optical Coherence Tomography

Nature Neuroscience: doi: /nn Supplementary Figure 1. Optimized Bessel foci for in vivo volume imaging.

MEMS scanner enabled real-time depth sensitive hyperspectral imaging of biological tissue

attosnom I: Topography and Force Images NANOSCOPY APPLICATION NOTE M06 RELATED PRODUCTS G

Automation of Fingerprint Recognition Using OCT Fingerprint Images

Dynamic Phase-Shifting Microscopy Tracks Living Cells

Optical Characterization and Defect Inspection for 3D Stacked IC Technology

Fast Optical Form Measurements of Rough Cylindrical and Conical Surfaces in Diesel Fuel Injection Components

Optics and Lasers. Matt Young. Including Fibers and Optical Waveguides

Semi-resonant operation of a fiber-cantilever piezotube scanner for stable optical coherence tomography endoscope imaging

Optimization for Axial Resolution, Depth Range, and Sensitivity of Spectral Domain Optical Coherence Tomography at 1.3 µm

SENSOR+TEST Conference SENSOR 2009 Proceedings II

Adaptive multi/demultiplexers for optical signals with arbitrary wavelength spacing.

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background

Two-axis MEMS Scanning Catheter for Ultrahigh Resolution Three-dimensional and En Face Imaging

Visualization of human retinal micro-capillaries with phase contrast high-speed optical coherence tomography

NEW LASER ULTRASONIC INTERFEROMETER FOR INDUSTRIAL APPLICATIONS B.Pouet and S.Breugnot Bossa Nova Technologies; Venice, CA, USA

Customized Lasers for Specific Swept Source OCT Applications. Bill Ahern Axsun Technologies, Inc. June 20, 2013

Numerical analysis of gradient index lens based optical coherence tomography imaging probes

Spatial Investigation of Transverse Mode Turn-On Dynamics in VCSELs

Impact of the light coupling on the sensing properties of photonic crystal cavity modes Kumar Saurav* a,b, Nicolas Le Thomas a,b,

Optical beam steering using a 2D MEMS scanner

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy

Axsun OCT Swept Laser and System

Faculty Development Program on Micro-Electro-Mechanical Systems (MEMS Sensor)

Measurement of Surface Profile and Layer Cross-section with Wide Field of View and High Precision

Non-contact Photoacoustic Tomography using holographic full field detection

USER MANUAL VarioS-Microscanner-Demonstrators

attocfm I for Surface Quality Inspection NANOSCOPY APPLICATION NOTE M01 RELATED PRODUCTS G

Multimodal simultaneous photoacoustic tomography, optical resolution microscopy and OCT system

Fourier Domain (Spectral) OCT OCT: HISTORY. Could OCT be a Game Maker OCT in Optometric Practice: A THE TECHNOLOGY BEHIND OCT

Theory and Applications of Frequency Domain Laser Ultrasonics

A MEMS Based Visible-NIR Fourier Transform Microspectrometer

BMC s heritage deformable mirror technology that uses hysteresis free electrostatic

Confocal Microscopy and Related Techniques

Development of innovative fringe locking strategies for vibration-resistant white light vertical scanning interferometry (VSI)

Simultaneous acquisition of the real and imaginary components in Fourier domain optical coherence tomography using harmonic detection

:... resolution is about 1.4 μm, assumed an excitation wavelength of 633 nm and a numerical aperture of 0.65 at 633 nm.

Diffraction, Fourier Optics and Imaging

Exercise 8: Interference and diffraction

Modifications of the coherence radar for in vivo profilometry in dermatology

SUPPLEMENTARY INFORMATION

FIRST REPORTED in the field of fiber optics [1], [2],

Comparison of resolution specifications for micro- and nanometer measurement techniques

TWO-AXIS laser-beam steering systems have been used in

attocube systems Probe Stations for Extreme Environments CRYOGENIC PROBE STATION fundamentals principles of cryogenic probe stations

High-Coherence Wavelength Swept Light Source

Comparison of FMCW-LiDAR system with optical- and electricaldomain swept light sources toward self-driving mobility application

InP-based Waveguide Photodetector with Integrated Photon Multiplication

LARGE-SCAN-RANGE ELECTROTHERMAL MEMS MICROMIRRORS AND MICROLENSES AND THEIR BIOMEDICAL IMAGING APPLICATIONS

Lab Report 3: Speckle Interferometry LIN PEI-YING, BAIG JOVERIA

Transcription:

A THREE DIMENSIONAL REAL-TIME MEMS BASED OPTICAL BIOPSY SYSTEM FOR IN-VIVO CLINICAL IMAGING Daniel T. McCormick 1, Woonggyu Jung 2,3, Yeh-Chan Ahn 2, Zhongping Chen 2,3 and Norman C. Tien 4 1 Advanced MEMS Inc., Berkeley, CA, USA (Tel : +1-530-867-0905; E-mail: dmcc@advancedmems.com) 2 Beckman Laser Institute, Irvine, CA, USA 3 Department of Biomedical Engineering, University of California, Irvine, CA, USA 4 Department of Electrical Engineering and Computer Science, Case Western Reserve University, Cleveland, OH, USA TRANSDUCERS & EUROSENSORS 07 Abstract: In this work we present our latest optical biopsy system realized via MEMS based optical coherence tomography imaging. We report, for the first time, real-time 3-D optical biopsies of human internal organs employing a miniature MEMS based endoscopic probe in a clinical environment. We also present a variety of MEMS imaging probes, each designed for a particular application area. Leveraging the unique advantages of specific OCT systems with properly optimized MEMS probes we are able to achieve truly optimal results for individual clinical and research applications. Artifact free images with 5µm 5µm 5µm sampling resolution (10µm 10µm 10µm optical) at speeds of 33µs per voxel are achieved. Keywords: micro-electromechanical systems (MEMS), bio-photonics, 3D imaging, optical coherence tomography (OCT), minimally invasive imaging, in-vivo imaging I. INTRODUCTION Optical coherence tomography (OCT) is an emerging imaging modality capable of achieving high resolution, cross-sectional images in tissue and other materials. [1]-[4] Due to the nondestructive nature of low-power optical imaging technologies biophotonics has become a major area of research focus in both academia and industry. In particular OCT has received a great deal of attention in the fields of biology and medicine. This is due, in part, to the fact that in addition to providing high resolution, nondestructive tomographic imaging capabilities OCT provides the opportunity to simultaneously realize real-time visualization of tissue structure and blood flow in a non-invasive, or minimally invasive, manner. Currently imaging technologies such as ultrasound and Doppler ultrasound are employed to obtain minimally invasive, in-vivo, real-time images of tissue structure and blood flow profiles. However, the spatial resolution of these techniques is limited to approximately 100 m due to the relatively long wavelength of acoustics waves. OCT takes advantage of the short coherence length of broadband light sources in order to achieve cross-sectional images with micrometer (2-10 m) scale resolution. Due to the limited penetration depth of OCT systems in most biological tissue (approximately 2-3mm) miniaturized endoscopic probes are required for imaging of internal organs and tissue. Furthermore, small handheld probes with fast imaging speeds, excellent stability and ease of use are required for a wide range of clinical applications including ophthalmology and imaging of the skin as well as the oral cavity. Advances in silicon micromachining and microelectromechanical systems (MEMS) technologies provide an enabling technology, allowing a new approach to be taken in order to achieve highspeed, high resolution OCT imaging. Modern photolithography and micromachining technologies allow the realization of structures and devices that are capable of manipulating light with sub-wavelength precision. The small physical size of MEMS optical components is highly advantageous when building miniaturized, portable scanning heads, hand-held scanners and 203

minimally invasive endoscopic probes. In addition, MEMS technologies allow the realization of miniature optical systems exhibiting both high speed and low power operation. A large number of MEMS based optical scanners have been developed since the inception of MEMS technologies; applications have ranged from the creation of micro-silicon optical benches and all-optical photonic switches to free space communication, displays, imaging and adaptive optics. MEMS mirrors and scanning lenses have previously been employed to provide beam steering in biophotonic applications such as confocal microscopy [5], cell sorting [6] and florescence imaging [7]. Recently MEMS devices have also been developed for OCT applications and OCT systems utilizing a scanning MEMS device in the sampling arm have been presented [8]-[10]. Previously we demonstrated the first three-dimensional (3-D), high-resolution, video rate capable, MEMS based OCT system. [11] High speed, 3-D OCT provides several improvements and advantages compared to 2-D imaging. The ability to arbitrarily visualize and manipulate 3-D images of tissue containing lesions and tumors provides significant additional diagnostic information to both physicians and researchers. In this paper we present a fullyfunctional, clinical OCT system complete with a wide range of application specific, optimized probes for real-time, 3-D, in-vivo and in-vitro imaging. II. OCT SYSTEM AND METHODS A schematic diagram of a simplified 3D Time- Domain OCT system is provided in Fig. 1. Light from a low coherence source (in this work the source wavelength is 1310nm with a 70nm bandwidth at a power level of 10mW) is coupled into a fiber-optic Michelson interferometer. A visible aiming beam (typically a 633nm laser) is also coupled into the interferometer in order to allow simultaneous imaging and visual identification of the area of tissue being imaged. This is useful during general clinical imaging as well as for precise labeling of areas to be identified for post OCT histologies. In the reference arm of the interferometer a Fig.1. A simplified schematic diagram of the 3-D time-domain OCT system employing a 2-D scanning endoscopic probe. rapid scanning optical delay line (RSOD) is utilized to provide a variable optical path length. In the time-domain (TD-OCT) system the RSOD line utilizes a galvanometer to scan a grating, allowing separate control of the phase and group delays. In addition, an electro-optical phase modulator is incorporated into the reference arm to generate a stable carrier frequency. The MEMS based probe is placed at the end of the sampling arm of the interferometer. The MEMS scan-head is used to direct a focused beam onto the sample as well as to provide beam scanning during image acquisition. In the presented schematic (Fig. 1) the MEMS scanner is illustrated with a pre-scan lens. In many benchtop and hand-held probes both pre-scan and postscan optics are employed. However, in the case of endoscopic probes generally either a single prescan lens or lenslet is utilized. A 1-dimensional scanning device will allow a single cross-sectional (i.e. 2-D) image to be collected whereas a 2- dimensional or 2-axis scanning probe allows 3-D images to be obtained. The reflected beam entering the sample arm of the interferometer is recombined with the beam from the reference path and the interference signal is detected using a high speed photodetector. The sampled signal is then processed and displayed as an image to the operator. As previously described the MEMS based imaging probe serves as a means of delivering light to the sample being imaged as well as a beam steering and optical focusing device. The optical 204

elements in the probe determine the lateral resolutions of the imaging system while the coherence length of the source determines the axial (depth) resolution. III. MEMS PROBE DESIGN When designing the MEMS devices and imaging probes numerous trade-offs are made between probe size, imaging speed, achievable resolution and ultimately cost. It is essential to identify the goals, requirements and preferences of clinicians as well as the source of design limitations in order to optimize each probe for a given application. Furthermore, ease of use and reliability are two important concerns when designing the probes for clinical use. Finally, patient safety is the primary concern during design. The optical performance of the MEMS scanner must be sufficient as not to degrade the overall image quality. In general a scanner that achieves the largest angle and the highest operating speed is desirable. Minimum clinical line scan lengths range from approximately 1mm to 2mm for most applications. Imaging speed is a very important parameter if large areas are to be imaged quickly and also in order to minimize motion artifacts. In addition, for a mirror based scanner the aperture size and quality are also very important parameters. In order to maximize resolution the objective is to achieve a larger mirror. However, as the mirror is enlarged the inertia of the mirror is increased and for a given spring stiffness the resonant frequency will decrease. On the other hand if the suspension is stiffened to maintain a given speed the torque available from the actuator must be increased. For a given fabrication technology and set of critical-dimensions this equates to enlarging the size of the actuator and thereby the area occupied by the die and package. The mirror must have good reflectivity in the visible and near infrared ranges; this requires either metallization or a dielectric mirror. The mirror surface should be sufficiently flat (generally r curvature 1m) as to not distort the beam and must also have surface roughness less than 100nm at a minimum. It is also critical that the devices exhibit excellent accuracy and precision in terms of pointing stability and scan to scan repeatability. In our experience the angular accuracy must be, at a minimum, better than one percent of the angle step size. In some cases it is desirable to achieve maximum optical resolution; in others higher acquisition rates or larger fields of view are required. Many parameters of the MEMS device and optical probe must be selected during design, fabrication or assembly. Others can be adjusted on a case-by-case basis before an imaging session (this require recalibration). However, using our MEMS control system it is possible to adjust many operating parameters, such as scan mode (point-to-point or constant velocity), scan region, scan speed and sampling resolution, in real-time. Therefore, the clinician can select the best probe for a specific application and still make trade-offs between field-of-view, image resolution and acquisition speed continuously during an imaging session. The MEMS scanning actuators employed in this work are monolithic, single crystal silicon, 2- dimensional, gimbal-less, vertical comb-driven structures. The devices are designed and realized in a self-aligned DRIE fabrication process on SOI [12],[13]; for endoscopic packaging the area footprint of the die is 2.8mm 3.3mm. The mirrors are fabricated in a separate SOI process and metalized prior to being bonded to the actuator. This design approach allows the mirror and actuator to be independently optimized. An SEM and optical micrographs of scanners with bonded mirrors are presented in Fig. 2. Fig. 2. (left) SEM of an endoscopic scanner, the die is 2.8mm 3.3mm with a 1mm bonded mirror. (top right) A micrograph of a larger die and 800µm mirror. (bottom right) Scanners with different mirror apertures and a US dime for size reference. 205

The apertures used in this work are metalized, low-inertia, SCS structures with a thinned mirror plate (1.5-5µm), thick stiffening trusses (~20µm) and a standoff pedestal (~120µm). A variety of scanners with 800µm, 1mm, 1.2mm, 1.6mm and 2.0mm mirror diameters have been realized. The mirrors and optical components are integrated into bio-compatible packages which provide optical alignment, electrical connections, mechanical protection and optical windows. The precise design of the package depends on the specific application. In this work four generations of endoscopic probes ranging in diameter from 3.9mm to 5.5mm have been utilized. Handheld probes with an outer diameter of 10mm and 15mm are also reported. Fig. 3 provides photographs of fully functional 1st, 2nd and 4th generation endoscopic probes (3.9mm, 4.9mm and 5.0mm respectively.) Fig. 3. Photographs of completed fully-functional probes. (A, B) 1st (3.9mm) and 4th (5.0mm) generation probes are compared to a US dime and a US quarter respectively. (C) Magnified photograph of 2nd generation probes. The steel lens housing provides rigid as well as proper alignment with and spacing to the MEMS device; the outer diameter of this probe is 4.9 mm. The normal imaging mode for our clinical OCT system is to display 2-D cross-sectional images to the clinician on a monitor in real-time during a scan. The imaging frame rate is dependant on many factors, including the number of voxels in the image and the speed of the OCT hardware. There may also be a difference between the acquisition speed and the display rate as the clinician or physician may wish to slow down the visualization on the monitor. Therefore it is not meaningful to quote frame rates without describing the specific imaging conditions. In the presented experiments all imaging speeds are limited by the axial scanning rate or processing hardware, not the speed of the MEMS device. For the TD-OCT system the axial scan rate is 500Hz, while for frequency domain system rates as high as 40kHz are achievable. In a point-to-point scanning mode imaging times of 33 microseconds per point are achieved, and in constant velocity modes 4000 lateral scans per second are realized. The typical optical resolutions of our probes is between 10 m 10 m 10 m and 20 m 20 m 10 m and the imaged volumes nominally range from 1mm 1mm 1.4mm to 2mm 2mm 1.4mm. A standard endoscopic instrument and fiber based camera are used to guide the positioning of the OCT probe during imaging experiments. This allows surgeons and clinicians already familiar with standard biopsy and endoscopic tools to quickly become comfortable and proficient with the OCT biopsy system. Fig. 4 presents images of a probe prepared for esophageal imaging and a picture of the probe scanning the upper airways of a human patient. IV. IMAGE ACQUISITION AND RESULTS Employing the MEMS based optical biopsy system large numbers of in-vivo and in-vitro image sets have been collected of healthy and diseased tissue. Clinical in-vivo imaging has been performed in both animals and humans. Fig. 4. (left) A 4 th generation MEMS probe attached to a flexible fiber-optic bronchoscope. (right) An image from the endoscope camera of a MEMS optical biopsy probe in the upper airway of a human patient. The endoscope guides the positioning of the MEMS probe and the visible aiming beam allows precise identification of the imaged area. 206

In Fig. 5 in-vivo OCT images of healthy esophageal and rectal tissue collected from an anesthetized rabbit utilizing 4 th generation endoscopic probes and a spectral domain OCT (SD-OCT) system during clinical studies are presented. The optical resolution of these images is 20 m 20 m 10 m and the imaged volume is 1mm 1mm 1.4mm. At this resolution and scan volume the frame rate of the SD-OCT is approximately 8 frames per second. These images clearly show the important structural and morphological features of the tissue. Histologies of the imaged tissue taken to verify the OCT images are also presented in Fig. 5. Fig. 5. (top) An in-vivo OCT image and subsequent histology of healthy rabbit trachea, clearly visualizing the epithelium (e), glands (g), cartilage (c) and tunica muscularis (tm). (bottom) In-vivo OCT image of healthy rabbit rectal tissue showing the important structural features. visualization of the region to be imaged and also allows precise positioning of the probe. Utilizing this setup 3-D in-vivo images were obtained of the laryngeal surface of the epiglottis as well as the false and true vocal cords under endoscopic visualization. A volume image of the true vocal chords is presented in Fig. 6. If it is determined that a lesion must be removed the optical biopsy system enables precise identification of resection boundaries. Fig. 6. True vocal cord of human in-vivo. Important structures such as the stratified squamous epithelium (sse), basement membra (bm), and superficial lamina propria (slp) are clearly visualized. Employing a hand-held probe in-vivo images of human skin have also been obtained and important features such as the epithelial layer, epidermal layer and sweat ducts are clearly visualized. In Fig. 7. a 3-D image of human skin tissue is shown. Clinical studies have also been completed with human patients. In addition to providing the ability to image large regions of tissue in a short period of time it is also possible to image organs and tissue that can be permanently damaged by traditional excissional biopsies, such as the vocal chords. For imaging of the human vocal chords the optical biopsy probe is attached to a rigid telescope type endoscope, which provides direct Fig. 7. In-vivo image of a human finger, the image has been thresholded to clearly show the spiral structure of the sweat ducts. The presented image was obtained with the SD- OCT system and has been thresholded to emphasize the spiral structure of the sweat ducts, 207

which are approximately 20-40 m in diameter. False coloring can also readily be applied during post-scan image processing in order to simultaneously emphasize various structures of different optical density in the tissue. V. CONCLUSION We have realized a 3-D MEMS based OCT optical biopsy system capable of providing realtime volume images with optical resolutions of 20 m or less both in-vitro and in-vivo. Bench-top, handheld and endoscopic probes have been employed to image healthy and diseased tissue in clinical studies with animal as well as human subjects. Previously the system has also been used to diagnose early stage cancer in an animal model. It is believed that this optical biopsy system will find use, and have significant impact, in a wide range of clinical and research environments. As neither the speed limits nor the optical resolution limits of our MEMS devices have been reached we will continue to develop faster imaging systems with higher resolution capabilities. We are also continuing to work with physicians and clinicians to optimize individual imaging probes for use in a variety of specific applications. ACKNOWLEDGMENT The authors thank David S. Mukai for assistance with animal preparation, regulatory compliance, and sample processing. We would also like to acknowledge the clinicians and physicians who performed the clinical imaging experiments and provide feedback regarding probe optimization: Ali Sepehr, William B. Armstrong and Matt Brenner. We also wish to thank V. Milanovi at the Adriatic Research Institute for mirror fabrication and wire-bonding. This work was partially supported by research grants from the National Institutes of Health (NCI-91717.) D. McCormick s email address is dmcc@advancedmems.com. REFERENCES [1] D. Huang, E.A. Swanson, C. P. Lin, J. S. Schuman, W. G. Stinson, W. Chang, M. R. Hee, T. Flotte, K. Gregory, C. A. Puliafito and J. G. Fujimoto, "Optical coherence tomography," Science, 254, 1178-1181 (1991). [2] A. F. Fercher, Optical coherence tomography, J. Biomed. Opt., 1, 157-173 (1996) [3] A. M. Rollins, M. D. Kulkarni, S. Yazdanfar, R. Ungarunyawee and J. A. Izatt, "In vivo video rate optical coherence tomography," Opt. Express, 6, 219-229 (1998),. [4] J. M. Schmitt, Optical coherence tomography (OCT): A review, IEEE J. Sel. Top. Quant., 7, 931-935 (2001). [5] S. Kwon, G.L. Liu, K.H. Jeong and L.P. Lee, Micro confocal line scanning system for high density microfluidics, in Proceedings IEEE/LEOS Optical MEMS, Aug 2003. [6] Y.C. Pei, W. Wilson, J.C. Liao, M.C. Wu, Cell addressing and trapping using novel optoelectronic tweezers, in Proceedings Micro Electro Mechanical Systems, pp. 21-24, 2004. [7] W. Piyawattanametha, R.P.J. Barretto, T.H. Ko, B.A. Flusberg, E.D. Cocker, H. Ra, D. Lee, O. Solgaard and M.J. Schnitzer, Fast-scanning two-photon fluorescence imaging based on a microeletromechanical systems twodimensional scanning mirror, Optics Letters, vol. 31, no. 13, pp. 2018-2020, 2006. [8] Y. Pan, H. Xie, G. K. Fedder, Endoscopic optical coherence tomography based on a microelectromechanical mirror, Opt. Lett., 26, 1966-1968 (2001) [9] D.T. McCormick and N.C. Tien, A mems based optical fiber scanning probe, in Proceedings of IEEE/LEOS Optical MEMS August 2002, pp/ 207-208. [10] W. Jung, D.T. McCormick, N.C. Tien and Z. Chen, Optical coherence tomography based on high speed scanning mems mirror, in Proceedings of SPIE: Coherence Domain Optical Methods and OCT in Biomedicine IX, 2005, vol. 5690. [11] D.T. McCormick, W. Jung, Z. Chen and N.C. Tien, 3-D MEMS Based Minimally Invasive Optical Coherence Tomography, in Proceedings of the 13th Annual International Conference on Sensors, Actuators and Microsystems, 2005. [12] D.T. McCormick, V. Milanovic and K. Castelino, A modular custom aperture technology for optimization of mems scanners, in Proceedings of IEEE/LEOS Optical MEMS August 2005. [13] V. Milanovic, G.A. Matus and D.T. McCormick, Gimbal-less Monolithic silicon actuators for tip-tiltpiston micromirror applications, J. of Selected Topics in Quantum Electronics, vol. 10, no. 3, pp. 462-471, 2004. 208