AC/DC Current Probe GCP-100 QUICK START GUIDE ISO-9001 CERTIFIED MANUFACTURER

Similar documents
AC/DC Current Oscilloscope Probe Model SL261

SERIE E N. The E N series clamps use Hall-effect technology for the measurement of AC and DC currents from several milliamps to over 100 A.

10.00 (1/2) PAC series

K1 K2. User's manual AC/DC CURRENT PROBE C H A U V IN A R N O U X AC/DC CURRENT PROBE. 1 mv/ma OFF 300V CAT II

Current clamps for AC current

Current clamps for AC current

AC/DC current clamps (1/2) E N series

AC/DC Current Probes Model MR411 Model MR521

34134A AC/DC DMM Current Probe. User s Guide. Publication number April 2009

E Series 9.00 (1/2) E series

MR461 MR561 AC/DC OSCILLOSCOPE CURRENT PROBES. User Manual. AC/DC CURRENT PROBE Model MR 561. AC/DC CURRENT PROBE Model MR V CAT III - 600A

The table below lists the symbols used on the Clamp and/or in this manual. Important Information. See manual.

80i-110s AC/DC CURRENT PROBE

Flexible AC current probes

Agilent 1146A AC/DC Oscilloscope Current Probe

80i-110s. AC/DC Current Probe Instructions. Introduction

AC/DC CLAMP METER USER S MANUAL

USER MANUAL 600A AC Clamp Meter + NCV Model MA610

OPERATOR S INSTRUCTION MANUAL

CM605. User Manual AC/DC LOW CURRENT CLAMP-ON METER ENGLISH

For ultra-high precision measurement of current: DC, AC, pulsed..., with galvanic separation between primary and secondary. Applications.

FlexProbe FLEXIBLE AC CURRENT PROBE. User Manual. FlexProbe ON OL 1000V CAT III 300A 3000A OFF

Multi range AC current clamps (1000A / 100A / 5A / 0.5A) A 1281 Instruction manual Version 2.0, Code no

RIGOL. User s Guide. RP1000D Series High Voltage Differential Probe. Feb RIGOL Technologies, Inc

3B SCIENTIFIC PHYSICS

F 03. Pince multimètre Clamp multimeter Vielfachmesszange Pinza multimetro Pinza multimetrica

AMM-1022 Digital Multimeter USER`S MANUAL

Preface First of all, thank you for purchasing our products, this instruction manual is the description about the function, usage, operation attention

TA MHz oscilloscope probe TA MHz oscilloscope probe

MiniFlex MF MF FLEXIBLE AC CURRENT SENSOR. User Manual

For ultra-high precision measurement of current: DC, AC, pulsed..., with galvanic separation between primary and secondary. Applications.

For ultra-high precision measurement of current: DC, AC, pulsed..., with galvanic separation between primary and secondary. Applications.

BENNING. CM7 Electrical Tester. Service Information

CT238A. AC/DC Current Probe. User Manual. For detailed specifications and ordering info go to

High Performance Current Transducer IT 200-S ULTRASTAB = A. ε L

For ultra-high precision measurement of current: DC, AC, pulsed..., with galvanic separation between primary and secondary. Applications.

USER MANUAL. Model MA A AC Mini Clamp-on Meter. Additional User Manual Translations available at

USER'S MANUAL ACDC-100 TRMS ACDC-100. Versatile AC/DC Clamp-on Multimeter Series

1. SAFETY INFORMATION.1 2. DESCRIPTION SPECIFICATIONS.6 4. OPERATING INSTRUCTION Voltage measurement Current measurement 10

Model: PR-55. High Voltage Oscilloscope Probe 10 kv, 40 MHz USER MANUAL

Instruction Manual ICM 3091N Digital AC Clampmeter EN FR IT DE ES

TETRIS User's Guide. High Impedance Active Probe DO177-1

CONTENTS. SAFETY PRECAUTIONS: Before use, read the following safety precautions

MiniFlex MF HF MF HF FLEXIBLE AC CURRENT SENSOR. User Manual. High Frequency Response for Oscilloscopes

For ultra-high precision measurement of current: DC, AC, pulsed..., with galvanic separation between primary and secondary. Applications.

DIGITAL DUAL DISPLAY AC/DC CLAMP METER MODEL-860A OPERATION MANUAL

PT-2000series AC/DCcurrentprobe PT-2710 PT-2720 PT-2740 PT-2770 MANUAL INSTRUCTION

DC/AC Current clamp 2000A A 1636 Instruction manual Version 1.2, Code no

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL:

NOTE: Fully read and understand this manual before using this Digital Multimeter.

Model: &

I P. /dt. di p V S Applications. Standards 1) IEC : 2007; IEC : ) IEC : 2016; IEC : 2017

DM-45 Digital Multimeter

Digital Clamp Meter Model: &

3B SCIENTIFIC PHYSICS

User s Guide. Model MA A AC Mini Clamp-on Meter

Preface Shenzhen Zhiyong Electronics Co., Ltd. First of all, thank you for purchasing our products, this instruction manual is the description about t

AC / DC STROOMTANG MULTIMETER NI 30 AC / DC CLAMP MULTIMETER NI 30. Handleiding / Manual

Preface Shenzhen Zhiyong Electronics Co., Ltd. First of all, thank you for purchasing our products, the product specification is about the function, u

15B+/17B+/18B+ Calibration Manual. Digital Multimeter

Safety Warnings Features Specifications Instrument Layout Operation Preparation AC Current Measurement How to Use Peak Hold Function How to Use The

For ultra-high precision measurement of current: DC, AC, pulsed..., with galvanic separation between primary and secondary. Applications.

323/324/325 Clamp Meter

Model UT233 OPERATING MANUAL

AC/DC DIGITAL CLAMP METER OPERATION MANUAL

DC/AC CLAMP METER ISO-TECH 2000 USERS MANUAL

Electrical Safety Tester

LH240 LH1040 LH2040 LH2015

User s Manual. Miniature Passive Probe. IM EN 3rd Edition

For the electronic measurement of current: DC, AC, pulsed..., with galvanic separation between the primary and the secondary circuit.

Thank you again for choosing AstroAI, if you have any questions or concerns regarding your product, please contact us at

DVM645BI BENCH MULTIMETER TAFELMULTIMETER MULTIMETRE DE TABLE BANCO MULTÍMETRO TISCHMULTIMETER. User Manual. Gebruikershandleiding

374 FC/375 FC/376 FC Clamp Meter

Compact Autoranging Clamp Meters. Models (400 A AC), (400 A AC/DC)

400Amp True RMS AC Clamp Meter + NCV

MT /6 Smart Digital Clamp Meter. User s Manual 1 st Edition, 2016 Copyright by Prokit s Industries Co., Ltd.

UT207A/208A/209A Operating Manual. Table of Contents

User Manual Digital Clamp Multimeter. model no.: MSR-C600

MHz FUNCTION GENERATOR INSTRUCTION MANUAL

HZ 115. Differential probe for oscilloscopes. C E IEC 1010 Cat III

PKT 512A-RO High Impedance Passive Cable Divider

TETRIS 1000 High Impedance Active Probe. Instruction Manual

Mini Clamp Meter Model:

For the electronic measurement of current: DC, AC, pulsed..., with galvanic separation between the primary and the secondary circuit.

UT232. Operating Manual. Digital Power Clamp Meter

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: MODEL: Multimeter Service Information

Sound Level Meter C.A 832

HHMA2 DC / TRUE RMS AC NON-CONTACT MILLIAMMETER

Dawson DDM181. Pocket-Size Autorange Digital Meter User s Manual

TA MHz ±700 V Differential Probe User s Manual. This probe complies with IEC , IEC CAT III, Pollution Degree 2.

200Amp AC Clamp Meter + NCV Model MA250

User s Manual Current Probe. IM E 1st Edition. Yokogawa Electric Corporation

TRMS LEAKAGE CURRENT CLAMP-ON METER 565

BENNING. Service Information MODEL: MM1-1 / MM1-2 / MM1-3

Operator s Manual. PP016 Passive Probe

Analog Technologies. Multimeter 15B and17b

PEN TYPE DIGITAL MULTIMETER OPERATION MANUAL T8211D

Relion 605 series Self-Powered Feeder Protection REJ603 Product Guide

HAWK5000 Operators Manual

Two Line V-Network GLN-5040A USER MANUAL GW INSTEK PART NO. 99 Washington Street Melrose, MA Phone Toll Free

Transcription:

AC/DC Current Probe GCP-100 QUICK START GUIDE ISO-9001 CERTIFIED MANUFACTURER

This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will Corporation. The information in this manual was correct at the time of printing. However, Good Will continues to improve its products and therefore reserves the right to change the specifications, equipment, and maintenance procedures at any time without notice. Good Will Instrument Co., Ltd. No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.

SAFETY INSTRUCTIONS In this User s manual, failure to follow or carry out instructions preceded by this symbol may result in personal injury or damage to the device and the installations. Safety Symbols Warning Warning: Identifies conditions or practices that could result in injury or loss of life. Symbol 102 according to IEC-1010-2- 032(Application or withdrawal authorized on the drivers under dangerous tension) Do not dispose electronic equipment as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased. Thank you for purchasing GCP-100 AC/DC current probe. To obtain the best possible service from your device: Read this User s manual carefully, Comply with the precautions for use. PRECAUTIONS FOR USE Do not use the clamp on conductors in which the voltage is more than 600 V rms. Keep the jaw faces clean. Clean them if necessary with a lightly oiled soft cloth to prevent rusting. Keep the cable or the busbar centred in the clamp which should be perpendicular to the conductor. Avoid proximity to other conductors which may create fields of interference. The clamp must be totally disconnected in the OFF position when the battery is changed.

For your safety, ensure that the instruments used with your clamp are also in accordance with IEC 1010. ELECTRICAL SAFETY Protection From Electric Shocks Instrument with double insulation or strengthened insulation in the part that is hand held in normal use, and with single insulation or additional insulation between the primary and the secondary output. Maximum service voltage in accordance with IEC 1010 600V in installations of category III and degree of pollution 2 300V in installations of category IV and degree of pollution 2 Dielectric strength test voltage in accordance with IEC 1010 5550V 50/60Hz between the parts that are hand-held in normal use, 3250V 50/60Hz between the on one hand, and the primary primary on one hand, and the with the secondary on the other secondary on the other hand hand. Electromagnetic Compatibility Electrostatic discharge 4kV without disturbance, 8kV without destruction in accordance with IEC 801-2 Radiated field 3V/m in accordance with IEC 801-3 1kV class 1 without disturbance, 2kV class 2 with Rapid transients minor defects in accordance with IEC 801-4 Electric shocks 1kV - 0.5kA class 2 without disturbance, 2kV - 1kA class 3 with minor defects in accordance with IEC 801-5

FEATURES The GCP-100 is a current probe for oscilloscope which uses a Hall effect cell for the measurement of DC or AC current without modification of the installation (without switching off the circuit) It can measure currents from 50 ma to 100 A peak. It has 2 ranges and 2 lights indicating: ON OL correct power supply to the clamp overload of the range in use (saturation or peak) In addition a thumbwheel can be used to reset zero for adaptation to the measurement environment. This clamp adapts to all measurement instruments which have a BNC input and an impedance of 1MΩ, < 100 pf. DESCRIPTION 1. Switch The switch has 3 positions: Off The clamp is no longer supplied with power Measurement of DC or AC peak currents on the Range 10 mv/a basis of 10 mv per ampere. Measurement of DC or AC peak currents on the Range 100 mv/a basis of 100 mv per ampere. 2. Battery indicator This LED is unlit when the switch is OFF. In normal use, the green LED is lit. When the switch is no longer in the OFF position and the LED is unlit, this warns that the value of the battery voltage is too low or that the cell is no longer supplied with power, thus showing that there is a fault in the clamp.

3. Resetting zero on the probe The thumbwheel makes it possible to reset the output voltage to zero. It is thus possible to overcome the different errors due to thermal shifts, the Earth s magnetic field, the environment, and residual induction. 4. Indicator of range overload This indicator symbolized by «OL» shows as a red light when the measurement signal is outside the capabilities of the range. It can also indicate the presence of an impulse outside the capabilities of the range or that the measurement made on the clamp isn t valid. 5. Battery compartment To change the 9V battery, the clamp should be in the normal position (jaws closed) and disconnected from the oscilloscope, it must not clamp any conductor and the switch must be in the «OFF» position. Unscrew the tool release screw then pull off the cover in the handle extension to access the battery to be changed. OPERATING INSTRUMENT To make a current measurement, switch on the clamp by selecting the 100 mv/a range. Check that the battery indicator (green) is lit and that the «OL» indicator is not lit. Connect the clamp to the oscilloscope. With the clamp closed and without clamping a conductor, select the highest sensitivity (for example 1 mv/cm) on the oscilloscope and 100 mv/a on the clamp, then set zero on the clamp with the thumbwheel in relation to a reference chosen on the oscilloscope. Zero on the oscilloscope makes it possible to adjust this setting. Select the measurement sensitivities of the clamp and the oscilloscope. Choose the connection method which is the best adapted to measurement on the oscilloscope.

Note the direction of the primary current by means of the arrow marked on and under the case. Insert the conductor carrying the current to be measured in the clamp and take the measurement. If necessary, recheck the origin of the graph, with the jaws not clamped around the conductor, and make the measurement again. SPECIFICATIONS Mechanical Specifications Model GCP-100 Max. cable dimensions Ø 11.8 mm Jaw opening 12.5 mm maximum Dimension 231 x 67 x 36 mm win core lead (coaxial) 2m ended by a BNC plug insulated and moulded Weight 330g with battery Protection Degrees of protection: IP 20 to IEC 529 Protection from shocks: 100 G, 6 ms, halfcycle, to IEC 68-2-27 Drop height from all angles: 1 m Resistance to vibrations: 10/55/10 Hz 0.15 mm, to IEC 68-2-6 Bumps: 40 G, 6 ms, 4000 bumps, to IEC 68-2-29 29mm 67m 231mm 21m 36m

Electrical Specifications Range Measurement extent Intrinsic error Range 100mV/A 50mA to 10A peak 3% R + 50mA Range 10mV/A 50mA to 40A peak 4% R + 50mA Range 10mV/A 40A to 100A peak Graph value below R: reading Typical noise level at output (peak-peak value) These specifications are given for an ambient temperature of 23 ± 3 C, humidity 20 to 75% RH, frequency DC to 1 khz, load impedance: 1 MΩ/100 pf, conductor centred and parallel to the mark Frequency band DC-100kHz Range 10mV/A 480µV Range 100mV/A 3mV Pass band DC to 100 khz The pass band of the oscilloscope used depends on the frequency to be measured. A pass band of more than 4 times the frequency of the signal to be measured is sufficient. Operating DC to 20 khz (not causing an additional error of frequency more than 3% in relation to the reference range) Breaking - 3 db at 100 khz frequency Rise or fall time < 4µs General Specifications Environmental conditions Battery 9V type 6 LR 61 Consumption 8.6 ma typical, 12 ma max Service life 55 h typical, 40 h min. Indicator Green light goes out for a battery voltage < 6.5 V Class of protection Class II according to IEC 348 Dielectric strength 4 kv Leakage current < 0.5 ma

DISTORTION MAGNITUDES Battery voltage from 6.5 V to 10 V: ± 6 ma/v typical, ± 10 ma/v max. Temperature from 0 to 50 C: ± 2000 ppm/ C max Position of the conductor in the window (AC signal of frequency 1 khz): max ± 0.5% of the reading. Errors of linearity, accuracy, temperature shift and other specifications of the oscilloscope should be taken into account during measurement. External magnetic fields generated by an AC or DC current of 1A flowing in a conductor placed in immediate proximity MAINTENANCE For maintenance, use only specified spare parts. The manufacturer will not be held responsible for any accident occurring following a repair done other than by its After Sales Service or approved repairers. Jaw faces : It is necessary to always keep the jaw faces clean. Clean them and lightly oil them to avoid rust. Do not leave the clamp in very damp places, or directly exposed to water. Handles and case : Clean with a cloth or a sponge soaked with soapy water, rinse in the same way without ever getting water on the clamp. Dry with a cloth or in an air flow. To maintain the performance of the clamp, it is advisable to carry out a check or recalibration every year. Calibration For checking and calibration of your instrument, please contact our accredited laboratories or agent in your country.

Repairs Repairs under or out of guarantee: please return the product to your distributor. APPENDICE Typical response curves for frequency and phase Shift measurement current : 1A peak Measurement current : 1A peak Different limiting examples of the response of the response of the Clamp