CALIBRATION OF THE LH SYSTEMS ADS40 AIRBORNE DIGITAL SENSOR

Similar documents
ENMAP RADIOMETRIC INFLIGHT CALIBRATION, POST-LAUNCH PRODUCT VALIDATION, AND INSTRUMENT CHARACTERIZATION ACTIVITIES

SPECTRAL SCANNER. Recycling

High Resolution Sensor Test Comparison with SPOT, KFA1000, KVR1000, IRS-1C and DPA in Lower Saxony

Chapters 1-3. Chapter 1: Introduction and applications of photogrammetry Chapter 2: Electro-magnetic radiation. Chapter 3: Basic optics

EXAMPLES OF TOPOGRAPHIC MAPS PRODUCED FROM SPACE AND ACHIEVED ACCURACY CARAVAN Workshop on Mapping from Space, Phnom Penh, June 2000

Geometry of Aerial Photographs

Radiometric Solar Telescope (RaST) The case for a Radiometric Solar Imager,

Camera Calibration Certificate No: DMC III 27542

Real-Time Scanning Goniometric Radiometer for Rapid Characterization of Laser Diodes and VCSELs

1 st IFAC Conference on Mechatronic Systems - Mechatronics 2000, September 18-20, 2000, Darmstadt, Germany

Camera Requirements For Precision Agriculture

Airborne test results for a smart pushbroom imaging system with optoelectronic image correction

Camera Requirements For Precision Agriculture

Optoliner NV. Calibration Standard for Sighting & Imaging Devices West San Bernardino Road West Covina, California 91790

Kazuhiro TANAKA GCOM project team/jaxa April, 2016

Full Spectrum. Full Calibration. Full Testing. Collimated Optics, Software and Uniform Source Solutions

Infra Red Interferometers

Leica ADS80 - Digital Airborne Imaging Solution NAIP, Salt Lake City 4 December 2008

TEST RESULTS OBTAINED WITH THE LH SYSTEMS ADS40 AIRBORNE DIGITAL SENSOR

Be aware that there is no universal notation for the various quantities.

TELLS THE NUMBER OF PIXELS THE TRUTH? EFFECTIVE RESOLUTION OF LARGE SIZE DIGITAL FRAME CAMERAS

Chapters 1-3. Chapter 1: Introduction and applications of photogrammetry Chapter 2: Electro-magnetic radiation. Chapter 3: Basic optics

CALIBRATION OF OPTICAL SATELLITE SENSORS

Chapters 1 & 2. Definitions and applications Conceptual basis of photogrammetric processing

ENHANCEMENT OF THE RADIOMETRIC IMAGE QUALITY OF PHOTOGRAMMETRIC SCANNERS.

3.0 Alignment Equipment and Diagnostic Tools:

Aerial photography: Principles. Frame capture sensors: Analog film and digital cameras

Defense Technical Information Center Compilation Part Notice

SPECTRAL IRRADIANCE DATA

Determining MTF with a Slant Edge Target ABSTRACT AND INTRODUCTION

AIAA/USU Small Satellite Conference 2007 Paper No. SSC07-VIII-2

TSBB09 Image Sensors 2018-HT2. Image Formation Part 1

Comprehensive Vicarious Calibration and Characterization of a Small Satellite Constellation Using the Specular Array Calibration (SPARC) Method

CALIBRATION OF IMAGING SATELLITE SENSORS

Optical basics for machine vision systems. Lars Fermum Chief instructor STEMMER IMAGING GmbH

Norsk Elektro Optikk AS (NEO) HySpex Airborne Sensors System Overview

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

Leica RCD30 Calibration Certificate

ImageMaster. Worldwide Leader in MTF Testing

Compact High Resolution Imaging Spectrometer (CHRIS) siraelectro-optics

POTENTIAL OF LARGE FORMAT DIGITAL AERIAL CAMERAS. Dr. Karsten Jacobsen Leibniz University Hannover, Germany

UltraCam Eagle Prime Aerial Sensor Calibration and Validation

5 180 o Field-of-View Imaging Polarimetry

Optical design of a high resolution vision lens

Advanced Target Projector Technologies For Characterization of Staring-Array Based EO Sensors

AMIPAS. Advanced Michelson Interferometer for Passive Atmosphere Sounding. Concepts and Technology for Future Atmospheric Chemistry Sensors

APPLICATION AND ACCURACY POTENTIAL OF A STRICT GEOMETRIC MODEL FOR ROTATING LINE CAMERAS

Some of the important topics needed to be addressed in a successful lens design project (R.R. Shannon: The Art and Science of Optical Design)

Radiometric performance of Second Generation Global Imager (SGLI) using integrating sphere

Sensitivity analysis of phase diversity technique for high resolution earth observing telescopes

IMAGE FORMATION. Light source properties. Sensor characteristics Surface. Surface reflectance properties. Optics

ECEN 4606, UNDERGRADUATE OPTICS LAB

Chapter 5 Nadir looking UV measurement.

I-I. S/Scientific Report No. I. Duane C. Brown. C-!3 P.O0. Box 1226 Melbourne, Florida

BEAM HALO OBSERVATION BY CORONAGRAPH

Imaging Systems Laboratory II. Laboratory 8: The Michelson Interferometer / Diffraction April 30 & May 02, 2002

RADIATION BUDGET INSTRUMENT (RBI): FINAL DESIGN AND INITIAL EDU TEST RESULTS

Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI)

Miniaturized Spectroradiometer

Outline. Introduction. Introduction: Film Emulsions. Sensor Systems. Types of Remote Sensing. A/Prof Linlin Ge. Photographic systems (cf(

WaveMaster IOL. Fast and accurate intraocular lens tester

Eric B. Burgh University of Wisconsin. 1. Scope

Department of Mechanical and Aerospace Engineering, Princeton University Department of Astrophysical Sciences, Princeton University ABSTRACT

RADIOMETRIC CAMERA CALIBRATION OF THE BiLSAT SMALL SATELLITE: PRELIMINARY RESULTS

A Study of Slanted-Edge MTF Stability and Repeatability

Hyperspectral goes to UAV and thermal

Optical Performance of Nikon F-Mount Lenses. Landon Carter May 11, Measurement and Instrumentation

Astigmatism Particle Tracking Velocimetry for Macroscopic Flows

Dual-FL. World's Fastest Fluorometer. Measure absorbance spectra and fluorescence simultaneously FLUORESCENCE

LENSES. INEL 6088 Computer Vision

SYSTEM DESIGN ASPECTS OF A SPACEBORNE WIDE-ANGLE OPTOELECTRONIC STEREO SCANNER

ME 297 L4-2 Optical design flow Analysis

LECTURE NOTES 2016 CONTENTS. Sensors and Platforms for Acquisition of Aerial and Satellite Image Data

Congress Best Paper Award

Digital airborne cameras Status & future

Fast MTF measurement of CMOS imagers using ISO slantededge methodology

CRISATEL High Resolution Multispectral System

METimage an innovative imaging radiometer for Post-EPS

Test procedures Page: 1 of 5

Image Acquisition. Jos J.M. Groote Schaarsberg Center for Image Processing

Imaging Optics Fundamentals

Variable microinspection system. system125

WaveMaster IOL. Fast and Accurate Intraocular Lens Tester

Remote sensing image correction

ISUAL Imager Science Performance Test Report Author: Stewart Harris

A collection of hyperspectral images for imaging systems research Torbjørn Skauli a,b, Joyce Farrell *a

GEOMETRIC RECTIFICATION OF EUROPEAN HISTORICAL ARCHIVES OF LANDSAT 1-3 MSS IMAGERY

Observational Astronomy

Xenon-Diamond 2.9/106 With beam splitter

OPAL. SpotOptics. AUTOMATED WAVEFRONT SENSOR Single and double pass O P A L

Lecture Notes 10 Image Sensor Optics. Imaging optics. Pixel optics. Microlens

Criteria for Optical Systems: Optical Path Difference How do we determine the quality of a lens system? Several criteria used in optical design

An integral eld spectrograph for the 4-m European Solar Telescope

Open Access Structural Parameters Optimum Design of the New Type of Optical Aiming

IMAGE SENSOR SOLUTIONS. KAC-96-1/5" Lens Kit. KODAK KAC-96-1/5" Lens Kit. for use with the KODAK CMOS Image Sensors. November 2004 Revision 2

DIGITAL IMAGING. Handbook of. Wiley VOL 1: IMAGE CAPTURE AND STORAGE. Editor-in- Chief

ISO INTERNATIONAL STANDARD. Photography Electronic still-picture cameras Resolution measurements

APEX AIRBORNE PRISM EXPERIMENT A NEW CONCEPT FOR AN AIRBORNE IMAGING SPECTROMETER *

Puntino. Shack-Hartmann wavefront sensor for optimizing telescopes. The software people for optics

NIST Agency Report May 2012 OUTLINE. The case for traceability NMI capabilities A view to the future the HIP Current/recent NIST activities

Transcription:

CALIBRATION OF THE LH SYSTEMS ADS4 AIRBORNE DIGITAL SENSOR Reinhard SCHUSTER*, Bernhard BRAUNECKER** *Deutsches Zentrum für Luft- und Raumfahrt (DLR), Germany Institute of Space Sensor Technology and Planetary Exploration reinhard.schuster@dlr.de **Leica Geosystems AG, Switzerland bernhard.braunecker@leica-geosystems.com Working Group IC-4 KEY WORDS: Photogrammetry, Camera, CCD, Sensors, Calibration, Radiometric Qualities ABSTRACT High performance photogrammetric cameras require high resolution geometric, spectral and radiometric calibration. Instruments to meet the stringent specifications exist at LH Systems and DLR. The underlying principles of operation are discussed with respect to the new LH Systems ADS4 airborne digital sensor, and the results of the first cross measurement are reported. 1 INTRODUCTION The rapid progress of electronic sensors will undoubtedly lead to the substitution of film-based photogrammetric cameras by digital cameras. However, both analogue and digital cameras will be used in parallel by the photogrammetric community for at least one decade. With the advent of the ADS4, a joint project of LH Systems and Deutsches Zentrum für Luft- und Raumfahrt (German Aerospace Centre DLR), the issues of sensor calibration must be addressed. This paper reviews the calibration equipment currently in use at both locations. While the DLR equipment is favoured owing to its great flexibility and the possibility of modifying geometric, radiometric and spectral measurements in the development phase, the LH Systems equipment is optimised to industrial needs, especially through fast and automatic measurement modes. 2 THE CALIBRATION-FACILITY AT DLR - EXAMPLES OF CALIBRATION RESULTS The calibration facility located in the DLR Institute of Space Sensor Technology was originally developed for calibrating spaceborne sensors (Schuster, 1994) and was later extended to the infrared spectrum. The optical scheme of the calibration set up is shown in figure 1. cal-source blackbody halogen, hg or Xe filter lightmixing-rod slit-spot stepper vis-collimator Vis-Sensor rotation-tilt-table theodolite MWIR monochromator lightsource shutter optics autocollimationtelescope LWIR x-z-stage R-source IR-collimator Figure 1. Calibration facility at DLR (optical scheme) 288 International Archives of Photogrammetry and Remote Sensing. Vol. XXXIII, Part B1. Amsterdam 2.

The DLR facility (figure 2) features: Clean room class 1 Lamps: halogen (15W), xenon (15W), mercury (2W) Collimator f=12 mm, D=15 mm, apochromatic corrected 39-113 nm, with autocollimation device, adjusting slit or spot: aperture ±25 mm, accuracy ±1, resolution.1 Spectral resolution with grid-monochromator (35-8 nm, 7-15 nm); λ = 2...2 nm Two axis nodal bench azimuth axis 35, elevation axis ±5, accuracy ±2.5 arcsec, resolution.5 arcsec, load 5 kg PC-control of lamps, shutter, monochromator-drive, optical sensors, adjusting aperture and camera nodal bench Special software (MS Windows) for geometric and radiometric calibration tasks. Figure 2. View of the calibration laboratory at DLR-Berlin The main tasks for geometric calibration of a digital sensor are: Highly accurate measurement of the internal orientation of any active pixel for determination of the image coordinates Measurements of the system MTF for any pixel of the sensor. The first requirement is met by single pixel illumination along the CCD-line with a pinhole spot from the collimator focus. The direction of the illuminated pixel to the collimator axis is determined by the angles α (angle in line direction) and β (angle perpendicular to the line direction). These two angles can be exchanged in the case of horizontal mounting of the camera. For the transformation from the spatially fixed co-ordinate system (x,y,z) to the image co-ordinate system (x,y,z ) the simple transformation x' = f tanα / cosβ ; y' = f tan β ; z' = z (figure 3) must be applied (Ohlhof and Kornus, 1994). As example the result of calibration from the nadir-line of the first engineering model of the ADS4 (figure 4) is shown in figure 5 before and after transformation. Figure 3. Transformation from (x,y,z) to (x,y,z ) Figure 4. The engineering model of the ADS4 during the calibration procedure International Archives of Photogrammetry and Remote Sensing. Vol. XXXIII, Part B1. Amsterdam 2. 289

.2 6 4 Alpha i Alppix i 2 2.2 3 2 1 1 2 3 Beta i 4 6 2 4 6 8 1 1 4 Betpix i Figure 5. Result of geometric calibration in degree (1) before and in units of pixel-size (2) after transformation The second requirement is met by measurements of the Point Spread Function (PSF), from which the Modulation Transfer Function (MTF) can be calculated by a Fast Fourier Transform (FFT). It is useful to register at the same pixel the PSF and geometry. Because the PSF is a two-dimensional function, we measure in the two main directions corresponding to the flight direction and perpendicular in the line direction. From these measurements the MTF can be calculated at an arbitrary number of points along the line and used as input for the image post processing. Examples for PSF/MTF measurements with the engineering model of the ADS4 are shown in figure 6. The MTF curves characterise the optical performance of the digital camera system, which is mainly determined by the pixel size and the optics. The resolution always seems to be better in the flight direction than in the line-direction. flight-direction swath direction MTF x412n1 y412n1 x412t1 y412t1 1. 1. % 1 8-1.5.5 6 4 2.. 1 2 3 4 5 1 2 3 4 5 5 1 15 mm-1 x5938n1 y5938n1 x5938t1 y5938t1 1. 1. % 1 8.5.5 6 4 2.. 1 2 3 4 5 1 2 3 4 5 5 1 15 mm-1 x7479n1 y7479n1 x7479t1 y7479t1 1. 1. % 1 8 +1.5.5 6 4 2. 1 2 3 4 5. 1 2 3 4 5 5 1 15 mm-1 Figure 6. Examples of PSF/MTF measurements (nadir line of engineering model of the ADS4) The use of modern digital sensors together with a flexible electronic read-out channel also leads to new possibilities for the radiometry. If the total field of view (FoV) of a line camera is illuminated homogeneously we get a line-signal 29 International Archives of Photogrammetry and Remote Sensing. Vol. XXXIII, Part B1. Amsterdam 2.

similar to that in figure 7a. The shape of the curve reflects the intensity drop across the FoV, known as the cos 4 law, but the high frequency variation is caused by the variation of responsivity (photo response non-uniformity - PRNU) of the line-pixels. It seems useful to correct the signals pixel-wise to obtain constant signals with homogenous illumination. The correction can be done very simply online by subtraction (dark signal) or multiplication (PRNU and limb shadowing) by the digital part of the electronics. An alternative solution would be off-line correction during the image processing work. In both cases it is necessary to determine the radiometric correction values in a separate calibration process. After this correction the line signal must be free from PRNU and limb shadowing (figure 7c). a b c pk1a_14 pk1a_18 pk1a_22 pk1a_26 pk1a_3 pk1a_34 pk1a_38 pk1a_42 pk1a_46 pk1a_5 pk1a_54 7. KAF1A1K1 3.5 tesk1a14 tesk1a18 tesk1a22 tesk1a26 tesk1a3 tesk1a34 tesk1a38 tesk1a42 tesk1a46 tesk1a5 tesk1a54 7. 6. 5. 3. 6. 5. 4. 2.5 4. 3. 2. 3. 2. 1. 1.5 2. 1.. 1... 2. 4. 6. 8. 1. 12.. 2. 4. 6. 8. 1. 12.. 2. 4. 6. 8. 1. 12. Figure 7. Line-signal: a-uncorrected, b-prnu-correction values, c-corrected Normally CCD camera systems are sensitive in a relatively wide spectral band. Therefore it is not possible to determine the spectral behaviour of radiance of an unknown source. Nevertheless the absolute calibration can be done to an averaged value given by the convolution of the normalised responsivity with the spectral characteristics of the source. This absolute calibration is a part of the calibration program of the DLR facility. To proceed the following steps are necessary: 1. Measurements of the irradiance in front of the camera optics with separate detector dependent on wavelength. 2. Measurement of the spectral responsivity of the pixels and normalisation with the spectral behaviour of the facility. 3. Measurement of an absolute calibrated source (radiometric sphere) covering the total FoV of the camera. 4. Calculation of the absolute radiometric correction values. The method of absolute calibration was tested with a spaceborne camera system and gave a Noise Equivalent Radiation (NER) of 3-7 1 4 W/m³sr with a resolution of 7 1 4 W/m³sr. The performance of the PRNU correction was better than.2%. Figure 8 shows some sample results. RADI_LB2 6. BDS5_UW E+6 15 5. 4. 1 3. 2. 5 1.. 4 5 6 7 8 9 nm 4 5 6 7 8 9 nm Figure 8. Spectral responsivity of (a) a pixel, (b) radiance of the radiometric W/m³sr 3 PHOTOGRAMMETRIC CALIBRATION FACILITIES AT LH SYSTEMS The main optomechanical parts of the ADS4 are produced by LH Systems in Heerbrugg, Switzerland. Consequently, new instruments were developed for dual use at this location: to test lenses during the manufacturing process and to perform the final calibration of the integrated camera head at system level. 3.1 Geometric Calibration Geometric calibration of the ADS4 includes the quantitative determination of the image quality and of the registration geometry. To speed up measurement, both criteria should be measured simultaneously. This is accomplished by a coded International Archives of Photogrammetry and Remote Sensing. Vol. XXXIII, Part B1. Amsterdam 2. 291

vertical goniometer (CVG). The test programs are run automatically and the results directly linked to the LH Systems databases. 3.2 CVG Hardware The CVG was developed in 1998 by upgrading an existing vertical goniometer. Its main goal was the electronic testing of film-based analogue lenses under realistic operational conditions. Later, with the advent of digital cameras, hardware and software modifications were added to test the lenses together with the digital camera head (Pacey et al., 1999). Figure 9. Inserting optical systems into the CVG: a) objective UAGS for testing the optics b) engineering model of the ADS4 for calibration Figure 9 shows the installation of test lenses into the CVG, whereby the lens entrance pupil must coincide with the swivel axis of the goniometer. Inside the swivel arm an auxiliary optics images a small code pattern to infinity. The ADS4 lens under test refocuses the code pattern on to its digital sensors. Rotating the swivel performs a field scan along the CCD-line at nadir position. However, to address pixels outside the nadir line, a mirror scanner, moveable in the x-direction, is mounted on top of the goniometer arm. From the contrast of the transferred image we deduce values for the image quality; from the actual angles of swivel and x-scanner we gain the registration properties. The code pattern consists of a series of stochastically arranged black and white bars of width 3 pointing in two orthogonal directions. The code is designed to be sensitive to small lateral displacements to increase the accuracy for the registration measurement, and to possess sufficient spatial bandwidth for the image quality tests. 3.3 Measurement of the Image Quality As criteria for the image quality of a pixel the OTF (optical transfer function) across and along the CCD-lines is chosen. The OTF describes the complex variation of the transferred contrast through the optics with the spatial frequency. To ensure that the digital sensor lines are at best focus, the OTF must be measured at different depths of focus. This is easily accomplished by defocusing the small code pattern in the swivel arm. Figure 1a shows the results of a measurement: The ideal code pattern (red) has contrast one, while the measured code signals (blue) show smaller contrast values. The OTF is mathematically deduced from both functions. Its amplitude function MTF (modulation transfer function, upper curve in figure 1b, is already visible in the measured signals in figure 1a, where code sequences with low frequencies exhibit a larger contrast than those with high frequencies. Note further that the OTF s phase function PTF (lower curve in figure 1b) is nearly zero over the full frequency range, thus indicating a perfect, coma-free test lens. Again, this is nicely visualised in figure 1a, where the shape of all measured code elements is highly symmetric. 292 International Archives of Photogrammetry and Remote Sensing. Vol. XXXIII, Part B1. Amsterdam 2.

3.4 Measurement of the Geometric Registration The purpose of the registration calibration is to assign to each sensor pixel the angular position of its seen ground pixel element. OTF Measurement 1.8.6.4.2 5 1 15 2 25 3 35 4 45 5 position (pixel) 1 ) /.5 -.5 5 1 15 2 25 3 frequency (lp/mm) MTF>.6: MTF>.3: MTF>.1: ν< 22.8 lp/mm ν< 6.9 lp/mm ν< 15.7 lp/mm Figure 1. (a) code pattern before (red) and after (blue) imaging through a test lens (b) transfer functions MTF (blue) and PTF (green) deduced from (a) Unlike film cameras, where film shrinking requires the need of artificial fiducial marks, the situation is much easier for the digital camera. Each CCD-pixels is a fiducial per se. This allows to reduce the calibration task to assign to each pixel the value of its 'down to ground looking' direction vector a, mostly expressed by two polar angles, relative to a lens based co-ordinate system. Owing to the stringent accuracy requirements of 1 arcsec for both angles, special electro-optical means are foreseen in the CVG to control mechanical motions of swivel and x-scanner in real time. Powerful algorithms, specially adapted to the stochastic properties of the code structure, allow determination of the actual position of the code at the CCD-lines with sub- resolution. 3.5 Radiometric Calibration The radiometric calibration is done with a large Ulbricht sphere, located directly beside the CVG. The test bench is still under construction, but operates very similarly to the DLR equipment. At the time of writing the first test runs with the engineering model of the ADS4 are complete. 4 COMPARISON OF THE GEOMETRIC CALIBRATION MEASUREMENTS Photogrammetric cameras must be geometrically calibrated with an accuracy of parts of one pixel. Therefore very precise angular measurements are necessary to meet the stringent requirements. Using the first working ADS4 model, the engineering model, geometrical calibration was performed at both calibration facilities of DLR and LH Systems for the purpose of comparison. The outputs of the geometrical calibration process are two polar angular co-ordinates, assigned to each measured pixel. The first results indicate that it is sufficient to measure pixels every 2-5 degrees within the field of view, depending on the length of the CCD-line and on the focal length of the camera. The angular coordinates for pixels in between are interpolated numerically. The geometrical calibration results can be presented in different forms and units, e.g. in multiples of a pixel. The conversion to conventional units e.g. SI units, is then a simple multiplication with the pixel size. In the preceding chapters we outlined the different measurement modes at both locations. At DLR in Berlin the ADS4 is mounted horizontally on a rotation-tilt stage and the direction of the pixel through the optics to the fixed collimator International Archives of Photogrammetry and Remote Sensing. Vol. XXXIII, Part B1. Amsterdam 2. 293

axis is measured by moving the camera around two perpendicular axes. At LH Systems in Switzerland the camera is fixed in the vertical direction and the illumination device moves within the FoV of the camera. The direct comparison of the results for the nadir line of the engineering model at different times and places is presented in figure 11. The measured angles are converted to distortion values, expressed in pixel units. The difference between both curves is less than one pixel over the whole line of 12, pixels. This impressive result indicates the high level of precision and the high reliability of both devices. 1,5 1,5 distortion in pixels 2 4 6 8 1 DLR-stage -,5-1 CVG -1,5 number of pixel Figure 11. Comparison of the calibration results at DLR and LH Systems for the nadir line of the engineering model REFERENCES Ohlhof, T., Kornus, W., 1994. Geometric calibration of digital three-line CCD cameras. International Archives of Photogrammetry and Remote Sensing, Vol. 3, Part 1, Como, Italy, pp. 71-81. Pacey, R.E., Walker, A.S, Scheidt, M., 1999. Calibration of analog and digital airborne sensors at LH Systems. Proceedings of the 1999 ASPRS Annual Conference, Portland, Oregon, pp. 95-956. Schuster, R., 1994: Sensor calibration and geometric calibration of a three line stereo camera. International Archives of Photogrammetry and Remote Sensing, Vol. 3, Part 1, Como, Italy, pp. 265-271. 294 International Archives of Photogrammetry and Remote Sensing. Vol. XXXIII, Part B1. Amsterdam 2.